Download - JAL Innovation (Singapore)

Transcript
Match ΙΙ
Blood Glucose
Monitoring System
For in vitro diagnostic use.
Please consult instructions
for use.
Temperature limitation / Store at.
Do not reuse.
Manufacturer.
Lot number.
Caution, consult accompanying
document.
SIDE
2. Prepare For Blood Sampling
Use by /Expiry date.
Performance Characteristics:
A : 1. Getting To Know Your System
SIDE
B :
ACCURACY: More than ninety-five percent (95 %) of the individual glucose results fell within ± 0.83 mmol/L (15 mg/dL) of the
results of the glucose analyzer YSI 2300 measurement procedure at glucose concentrations < 4.2 mmol/L (< 75 mg/dL) and
within ± 20 % at glucose concentrations ≧ 4.2 mmol/L (≧ 75 mg/dL).
The following basic safety precautions should always be taken.
1. Close supervision is necessary when the device is used by, on, or near children,
handicapped persons or invalids.
2. Use the device only for the intended use described in this manual.
3. Do not use test strips and control solutions which are not supplied by the
manufacturer.
4. Do not use the device if it is not working properly, or if it has suffered any damage.
5. Before using any product to test your blood glucose, read all instructions
thoroughly and practice the test. Do all quality control checks as directed and
consult with a diabetes healthcare professional.
6. Do not use this meter in a dry environment, especially if synthetic materials are
present. Synthetic clothes, carpets, etc., may cause damaging static discharges in
a dry environment.
7. Do not use this meter near cellular or cordless telephones, walkie talkies, garage
door openers, radio transmitters, or other electrical or electronical equipment that
are sources of electromagnetic radiation, as these may interfere with the proper
operation of the meter.
8. KEEP THESE USER GUIDE WITH YOU.
Intended Use
The system is intended for use outside the body (in vitro diagnostic use only). It
should be used only for testing blood glucose (blood sugar) and with fresh capillary,
venous, arterial whole blood samples. The system is intended for use in the home and
Getting To Know Your System
10-61-3112-0103 V2-JAN13
The OKmeter Match II Blood Glucose Monitoring System.
The OKmeter Match II system uses the latest technology to provide you with easy and
comfortable testing. The system requires only a 0.7 μL of blood sample to complete
the testing in only 6 seconds.
The OKmeter Match II system consists of
1.OKmeter Match II Blood Glucose Meter
2.OKmeter Match II Blood Glucose Test Strips
3.OKmeter Control Solution
Important : Use only OKmeter Match II test strips and control solutions with your
OKmeter Match II meter. Use other test strips and control solutions with this meter
can produce inaccurate results.
SET BUTTON :
STRIP SLOT :
Located in the battery compartment;
press button to enter meter setting.
Refer to "SETTING TIME AND DATE "
section for details.
Insert the test strip here.
The meter will turn on
automatically.
DISPLAY :
Shows test results
and messages.
BATTERY
COMPARTMENT :
Holds ONE 3v Lithium
battery (battery type
CR2032). Please install
battery into meter before
you start to test.
Caution
1. Venous and arterial tests are for professional use only. Do not try them at
home!
2. The user should not take any decision of medical relevance without first consulting
his or her medical practitioner.
3. Call your doctor immediately if you experience symptoms that are not consistent with
your blood glucose test results.
4. High altitudes above than 3,402 meter (11,161 ft) may affect the test results.
5. Temperatures outside the range of 10°C to 40°C (50°F to 104°F) may affect the test
results. Do not test beyond of temperature range.
IMPORTANT HEALTH-RELATED INFORMATION
1. Apply only capillary, venous, arterial whole blood sample to test your blood glucose.
Applying other substances or plasma, serum will cause wrong results.
2. Severe dehydration and excessive water loss may cause false low results. If
you believe you are suffering from severe dehydration, consult your healthcare
professional immediately.
3. Test results below 60 mg/dL (3.3 mmol/L)*1 indicates low blood glucose
1. AVERAGE : Appears when the meter is in the
memory mode while recalling 7/14/21/28 day test
averages.
2. BLOOD DROP SYMBOL : Flashes when sample
4
13
5
should be applied.
12
3. STRIP SYMBOL : Appears when the meter is
6
11
7
turned on and waiting for inserting a test strip.
10
8
4. TIME : The meter is programmed for a 24 hour
9
period format.
5. MEMORY SYMBOL : Appears when in the
memory mode.
6. : Appears when test result is within the range of 70 ~120 mg/dL (or 3.9 ~ 6.7 mmol/L)
7.
: Appears when test result is lower than 70 mg/dL (or 3.9 mmol/L) or higher than
180 mg/dL (or 10.0 mmol/L)
8. DELETE SYMBOL : when it flashes, press “SET” BUTTON for 2 seconds to delete of
selected test result.
9. UNIT OF MEASURE : Appears with the test result in mg/dL or mmol/L.
10. CONTROL SOLUTION TEST SYMBOL : Appears when you are in Control Solution
Mode. Your test result will not be stored in meter memory.
11. BATTERY SYMBOL: Appears when battery is weak.
12. THERMOMETER SYMBOL : Appears when ambient temperature is above or below
the acceptable range needed for testing.
13. DATE : Display year, month and date.
2
3
REPLACING THE BATTERY
The OKmeter Match II comes with ONE Lithium batteries (battery type CR2032) .
Battery life will vary depending on usage, so always keep a spare battery on hand. The
battery should last about 1000 tests or 12 months when testing 3 times a day.
When the battery symbol appears on the meter display, battery is getting low. You will
still be able to test with low battery, but you should replace it as soon as possible.
When battery symbol and E-b shows up in the meter display, the meter will no longer
give results and you must replace the battery immediately. Please always have one
spare battery with you to ensure that you can replace the battery anytime.
How to replace the battery
01
0100
OK1G
Blood glucose is measured by an electrical current that is produced when a blood
samples mixes with the reagent (special chemicals) of the test strip. The electrical
current changes with the amount of glucose in the blood sample. The OKmeter Match
II meter measures the strength of the electrical current, calculates your blood glucose
level and then displays your result in either milligrams of glucose per deciliter (mg/dL) or
millimoles of glucose per liter (mmol/L).
1
ΙΙ
MadtGlchucosestem
Blooitoring Sy in Taiwan
Mon -1B Made
METER LABEL :
63-0003
10-62-31 N13
V2-JA
1. Make sure the meter is turned off. Let the front of the meter
rest in the palm of your hand. Slide battery cover open.
Each meter has its unique
number on it. Do not alter
or tear the label off.
M BUTTON :
perform other functions
described in User Guide.
2. Use the edge of battery cover to pull out battery from the
battery holder.
Meter
3. Insert new battery (battery type CR2032), being sure to
align the plus (+) side up. You should hear a beep to indicate
the battery installed correctly. If not, please reinsert the battery
correctly.
TEST STRIP HANDLE :
hold this part to insert the test strip
into the meter slot.
4. Close battery cover.
ABSORBENT
CHANNEL:
CONTACT BARS:
Insert this end into strip slot.
apply a drop of
blood here.
NOTE:
1.Replacing the battery does not affect the test result stored in memory. However the
time and date may need to re-set.
2.As with all small objects, the battery should be kept away from small children as a
safety precaution. If the battery is swallowed, seek medical assistance immediately.
3.Batteries might leak chemicals if not used for a long time. Remove the batteries if
you are not going to use the device for extended (i.e., 3 months or more ).
4. Please discard the product or the batteries properly according to the regulations
of your country.
CONFIRMATION WINDOW:
this is where we confirm if enough blood is
applied for has been drawn into strip.
Test Strip
2.
Prepare For Blood Sampling
Adjustable Lancing Device
Your lancing device and lancets are used for obtaining
blood samples from the puncture site.
Release Button
Do NOT share them with anyone including other family members !
Do NOT use on multiple patients !
Compounds
Gentisic Acid
Hydroxyurea
L-dopa
Maltose
Methyldopa
Tolbutamide
Triglycerides
Uric Acid
Higest concentrations tested
at which no interference
occured
≦ 5.0 mg/dL (0.32 mmol/L)
≦ 3.0 mg/dL (0.39 mmol/L)
≦ 10 mg/dL (0.51 mmol/L)
≦ 900 mg/dL (26.3 mmol/L)
≦ 3.0 mg/dL (0.13 mmol/L)
≦ 400 mg/dL (14.8 mmol/L)
≦ 2,000 mg/dL (22.6 mmol/L)
≦ 8.0 mg/dL (0.48 mmol/L)
REFERENCE:
* 1: Kahn, R. and Weir, G.: Joslinis Diabetes Mellitus, 13thed Philadelphia : Lea and Febiger (1994), 489.
* 2: Krall, L.P. and Beaser, R. S.: Joslin Diabetes Manual. Philadelphia : Lea and Febiger(1989), 261-263.
SETTING TIME AND DATE
Your OKmeter Match II meter comes with a new battery (Battery Type CR2032).
Please install battery first and set correct time and date before you begin to test.
STEP 1. Enter Setting Mode
Start with the meter off. Open the battery compartment, and press
SET BUTTON FOR 3 SECONDS. The meter will turn on and enter
the setting mode.
STEP 2. Set the Year
The year appears with the number flashing. Press and release the M
BUTTON to obtain the desired year. Press SET BUTTON to confirm
and move to next MONTH setting.
STEP 3. Set the Month
The month appears with the number flashing. Press and release
the M BUTTON to obtain the desired month. Press SET BUTTON to
confirm and move to next DATE setting.
STEP 4. Set the Date
The date appears with the number flashing. Press and release the M
BUTTON to obtain the desired date. Press SET BUTTON to confirm
and move to next HOUR setting.
STEP 5. Set the Hour
The hour appears with the number flashing. Press and release the M
BUTTON to obtain the desired hour. Press SET BUTTON to confirm
and move to next minutes setting.
STEP 6. Set the Minutes
The minute appears with the number flashing. Press and release
the M BUTTON to obtain the desired minute. Press SET BUTTON to
confirm and the meter will enter to next unit setting.
STEP 7. Set the Unit of Measurement
The existing unit (for example mg/dL) will appear and flash. Press
and hold the M BUTTON for 3 seconds to change the unit to mmol/L.
Press SET BUTTON to confirm and the meter will enter to next delete
all memory setting.
Lancet Ejector
STEP 7. Delete Memory
When the dEL symbol and the flashing memory symbol appear on
the display, you can choose to clear the memory. If you do not want
to clear the memory, press the SET BUTTON again to skip this step.
If you want to clear ALL memory, press and hold M BUTTON for 3
seconds. The “---“ image will appear on the LCD screen to indicate
that all memory has been deleted.
STEP 8. Complete Setting
After delete memory, the meter will display “OFF” before shut down.
The meter setting is now completed.
IMPORTANT :
1.The time, date and unit of measurement can ONLY be changed in the setting
mode. Therefore, when you perform a blood glucose testing, it is not possible to
change those parameters.
2.Your meter displays 7, 14, 21 and 28-day averages which you can access from
the meter memory. These averages are calculated from the date of your latest
result to 7, 14, 21, and 28 days before.
3.Your meter displays test results in milligram per deciliter (mg/dL). Use of the
wrong unit of measure may cause you to misinterpret your blood glucose level,
and may lead to incorrect treatment. Please always consult with your healthcare
professionals before you reset the unit of measure.
4.While the meter is in the setting mode, if no button is pressed for 30 seconds,
the meter will turn off automatically.
2. Twist off the protective cap until it separates from the
lancet.
3. Replace the lancing device cap and set the puncture
depth to the desired number.
NOTE: THE DEPTH SELECTOR OFFERS 5 LEVELS OF SKIN
1-2 : FOR SOFT OR THIN SKIN
4-5 : FOR THICK OR CALLOUSED SKIN
Setting your Lancing Device
Lancet Holder
Protective cap
Acetaminophen
Ascorbic Acid
Aspirin
Bilirubin
Cholesterol
Creatinine
Dopamine
Galactose
Higest concentrations tested
at which no interference
occured
≦ 8.0 mg/dL (0.53 mmol/L)
≦ 5.0 mg/dL (0.28 mmol/L)
≦ 60 mg/dL (3.33 mmol/L)
≦ 90 mg/dL (1.54 mmol/L)
≦ 500 mg/dL (12.9 mmol/L)
≦ 5.0 mg/dL (0.44 mmol/L)
≦ 2.0 mg/dL (0.11 mmol/L)
≦ 900 mg/dL (50 mmol/L)
3 : FOR AVERAGE SKIN
Cocking Control
Lancet
Compounds
Important Lancing Device and Lancets Information
1. Lancet is for single use only.
2. Keep lancing device and lancets clean.
3. Use caution when removing the used lancet from the device and when disposing the
used lancet.
4. The lancing device and lancets are in conformity with MDD 93/42/EEC. Refer to
product labels for contact information of manufacturer and CE marking.
IMPORTANT : The meter and lancing device are for single patient use.
Depth Selector
Lancing Device Cap
(hypoglycemia). Test results greater than 240 mg/dL (13.3 mmol/L)*2 indicates
high blood glucose (hyperglycemia).If your results are below 60 mg/dL(3.3 mmol/L)
or above 240 mg/dL(13.3 mmol/L), repeat the test, and if the results are still below
60 mg/dL(3.3 mmol/L) or above 240 mg/dL(13.3 mmol/L), consult your healthcare
professional immediately.
4. Inaccurate results may occur in severely hypotensive individuals or patients
in shock. Inaccurate low results may occur for individuals experiencing a
hyperglycemic-hyperosmolar state, with or without ketosis. Critically ill patients
should not be tested with blood glucose meters.
5. Abnormal red blood cell counts (hematocrit level below 20% or above 60%) may
cause false results. Please consult your healthcare professional if you do not know
your hematocrit level.
6. Interference: Reducing substances occurring in the blood naturally (uric acid,
bilirubin) or from therapeutic treatments (ascorbic acid, acetaminophen) will not
significantly affect OKmeter Match II test results. However, elevated concentrations
of these substances may affect test results. The compounds listed in the tables
were found to have no affect at the concentration indicated.
: OK
Name
g.
lf-testin
.
For se
Co., Ltd
om
otech
OK Biokbiotech.c
www.
Main button, press button
to turn the meter on or
A
PLEASE READ THIS BEFORE USING.
9. Display Messages And Problem-Solving Guide
MDSS GmbH
Schiffgraben 41, D-30175
Hannover, Germany
Principle of Measurement
SIDE
Before You Begin
8. About Alternative Site Testing (AST)
in clinical settings. It should not be used for the diagnosis of diabetes or for the testing of
newborns.
Model
6. Caring For Your Meter And Test Strip
7. System Specifications
OK Biotech Co., Ltd.
4F-1, No. 83, Sec. 2, Gongdao 5th Rd.,
Hsinchu City, 30070, Taiwan
TEL: +886-3-5160258
FAX: +886-3-5160028
www.okbiotech.com
E-mail: [email protected]
1.
4. Memory Recall
5. Control Solution Testing
PRECISION: CVs (%) of intermediate precision and repeatability were less than 5%.
User GUIDE
3. Performing Blood Test
1. Screw off the cap of lancing device. Insert a lancet
into the lancet holder and push down until it is fully
seated.
4. Pull back the cocking control until it makes a click,
and then release. If it does not click, the device may
have been cocked when the lancet was inserted.
3. Performing Blood Test
Figure 1
1. Wash Your Hands and the Puncture Site : Wash your
hands in warm, soapy water. Rinse and dry completely.
Warm your fingers to increase blood flow.
Figure 2
Figure 3
9. Discard the Used Lancet : Slide the lancet ejector
forward and dispose the lancet. Discard the lancet and
test strip according to your safety regulations.
*Do not reuse lancets.
1. Insert Test Strip : Insert a new test strip into the strip slot,
the meter will activate.
2. Mark as a Control Solution Test : After the blood
symbol( ) appears, press M BUTTON and " " appears
on the display indicating you are in the Control Solution
Mode. The meter will not store your test result in the
memory when you preset the test as a control solution
test. If you decide not to perform a control solution test,
press M BUTTON again and the " " will disappear.
1. To ensure that your meter and test strip are working properly together.
2. To allow you to practice testing without using your own blood.
When should the control solution test be performed?
1. When you first get your OKmeter Match II meter. Before use this system to test
your blood, you can practice the procedure by using control solution. When you
can do three tests in a row that are within the expected range, you are ready to
test your blood.
2. Once a week (to make sure that you continue to get accurate results)
3. When you begin using a new vial of test strips.
4. Whenever you suspect that the meter or test strips are not working properly.
5. When your blood glucose test results are not consistent with how you feel, or
when you think your results are not accurate.
6. When test strips are exposed to extreme environmental conditions.
7. If you drop the meter.
3. Squeeze a drop of Control Solution : Shake control
solution bottle well. Remove the cap. Squeeze bottle,
discard the first drop and wipe off the dispenser tip with
a clean tissue paper or cotton swab. Squeeze a drop on
clean bottle cap/fingertip /non-absorbent surface.
4. Apply Control Solution : Apply the drop to the opening
of the strip absorbent channel until the confirmation
window is filled. The meter begins to count down.
Important Control Solution Information
1. Check the expiration date on the control solution bottle. Do not use if expired.
2. Control solution, meter, and test strips should come to room temperature (68-77
ºF/20-25ºC) before testing.
3. Shake the bottle before use, discard the first drop of control solution after
squeezing, wipes off the dispenser tip to avoid contaminations. These steps
ensure you will get a good sample and an accurate result.
4. The discard date for control solution is 30 days after first opening. Record the
discard date on the bottle, when you open a new bottle of control solution.
NOTE : 1. There are two levels of control solution (normal and high) available to
purchase. Please contact with your local distributor when required.
2. The control solution range printed on the test strip vial is for OKmeter
Match Control Solution only. It is used to test meter and strip
performance. It is not recommended range for your blood glucose level.
8. About Alternative Site Testing (AST)
There are important limitations for doing AST. Please consult your healthcare
professional before you perform AST.
What is AST?
Alternative Site Testing (AST) means you can use parts of the body other than your
fingertips to check your blood glucose levels. The system allows you to test from the
palm, forearm, upper arm, calf or thigh, with equivalent results to fingertip testing.
What is the advantage?
Fingertips feel pain more readily because they are full of nerve endings (receptors).
At other body sites, nerve endings are not so numerous and you will not feel as much
pain as you will experience at the fingertip.
When to use AST?
Food, medication, illness, stress and exercise can affect blood glucose levels.
Capillary blood at fingertip reflects these changes faster than capillary blood at other
sites. Therefore, if you are testing blood glucose level during or immediately after
meal, physical exercise or stressful event, take the blood sample from your fingertip
only.
5. Check if the test result is in range: After the meter
counts down from 6 to 1, the test result shows up.
Compare the test result with the range printed on the test
strip vial. The result should fall within the printed range.
NOTE : 1. DO NOT APPLY THE CONTROL SOLUTION DIRECTLY TO THE TEST
STRIP! Overdosed solution may give inaccurate result.
2. Repeat test if test result falls outside the control range stated on the test
strip label. If subsequent test remains to produce unacceptable result,
the meter or test strip may be faulty. DO NOT use the system. Contact
us or your local distributor for help.
The following is a summary of some display messages and symbols. These messages
help to identify certain problems but do not appear in all cases when a problem has
occurred.
Improper use may cause an inaccurate result without producing an error message.
In the event of a problem, refer to information under ‘‘action to take’’
ACTION TO TAKE
If some parts of the display are not
working. Contact your local distributor
for help.
Use AST only:
1. In a pre-meal or fasting state (more than 2 hours since
the last meal).
2. Two hours or more after taking insulin.
3. Two hours or more after exercise.
4. During steady state blood glucose conditions.
Do NOT use AST if:
1. You have reason to believe you have hypoglycemia or
hyperglycemia.
2. Your routine glucose results are often fluctuating.
3. You are pregnant.
How to increase the accuracy?
Stimulating blood perfusion by rubbing the puncture site prior to blood extraction has
a significant influence on the glucose value obtained.
Blood from the site without rubbing exhibits a measurably different glucose
concentration than blood from the fingertip. When the puncture site was rubbed prior
to blood extraction, the difference was significantly reduced.
IMPORTANT :
To increase the accuracy when using AST, rub the puncture site more than 20
seconds before extracting blood.
Test result with
Appears when result is lower than
70mg/dL (3.9 mmol/L) or higher than
180 mg/dL (10 mmol/L).
sign.
Test Result with No
sign.
or
Blinking “del” with test
result
The meter is waiting for test strip be
insert.
Moving Strip
Deleting memory.
The meter is ready for blood applying
into test strip.
Blinking Blood.
Test result with
sign.
Appears when result falls in 70-120
mg/dL (3.9 - 6.7 mmol/L).
STEP 4. Exit the Memory Mode
After reaching the last set of result, the meter will display
"End" and turn off. Anytime in the memory mode, you can
press and hold M BUTTON for 3 seconds to turn off the
meter.
Deleting the Individual Memory
6. Caring For YourAndMeterTest Strip
To avoid the meter and test strips getting dirt, dust or other contaminants, please
wash and dry your hands thoroughly before use.
Cleaning
Your meter does not require special maintenance. As long as no blood or control
solution comes in direct contact with the meter, there is no special cleaning
required.
To clean the meter exterior, wipe with a cloth moistened with tap water or a mild
cleaning agent, then dry the device with a soft and dry cloth. Do not flush with
water.
Do not use organic solvents to clean the meter. Your meter is a precision
instrument. Please handle it with care.
Storage
1. Meter Storage
* Storage condition: -20ºC~50ºC (-4ºF~122ºF), below 90% relative humidity.
* Avoid dropping and strong impact.
* Avoid direct sunlight and humidity.
2. Strip Storage
* Storage condition: 4ºC~40ºC (39ºF~104ºF), and 10~85% relative humidity. Do
not refrigerate.
* Store your test strips in their original vial only. Do not transfer to other
container.
* Store test strip packages in a cool and dry place. Keep away from direct
sunlight and heat.
* After removing a test strip from the vial, immediately replace the vial cap and
close it tightly.
* You may touch the test strip anywhere with clean, dry hands when removing it
from the vial or inserting it into the meter.
* Use each test strip immediately after removing it from the vial.
* Do not bend, cut, or alter a test strip in any way.
* Keep the strip vial away from children since the cap and the test strip can be
a potential choking hazard. If swallowed, please seek medical assistance
immediately.
3. Control solution storage
* Storage condition: Store the control solution tightly closed at temperatures
below 86ºC (30ºC). Do not refrigerate.
Display Messages And Problem-Solving Guide
Display check
STEP 3. Recalling Individual Test Results
After 28-day average, the most recent test result with date
and time will be shown. Press M BUTTON once and the next
most recent test result will appear. Each time you press and
release the M BUTTON, the meter will recall up to your last
250 test results in order. When the memory is full, the oldest
result is dropped as the newest is added.
1.When you are in the memory mode and recall the individual memory, select the
test result you wish to delete and display it on the screen.
2.Press SET BUTTON. The deleting symbol ”dEL” appears on the display which
confirms that the selected test result has been deleted successfully.
3.The meter will return to the next individual memory recall.
How to Perform a Control Test
Why perform a control solution test?
DESCRIPTION
STEP 2. Recalling Average Test Results
When entering the memory mode, the 7-day average will
appear. If you continue to press the M BUTTON, the 14-day,
21-day, and 28-day averages will display in order.
Correct
Control Solution Testing
DISPLAY
STEP 1. Enter the Memory Mode
PRESS M BUTTON to turn on the meter, and press M bottom
again to enter memory mode.
*blood glucose value without or sign
If your test result falls in 120-180 mg/dL (6.7-10 mmol/L),
there will be no or sign. See Figure 3.
B
Incorrect
OKmeter control solutions contain a known amount of glucose that reacts with
OKmeter Match test strips. By testing your control solution and comparing the test
results with the expected range printed on the test strip vial label, you can make sure
that the meter and the test strips are working properly together as a system and that
you are performing the test correctly. It is very important that you do this simple check
routinely to make sure you get accurate results.
9.
Recall the Memory
8. Secure the Used Lancet: Twist off the lancing device
cap, and push the exposed tip of the lancet into its
protective cap.
5. Apply Blood Sample : Apply the blood sample to the
opening of absorbent channel of test strip until the
confirmation window is fully covered with blood. Then
the meter begins to count down and displays the test
result in 6 seconds.
5.
The OKmeter Match II automatically stores 250 test results, letting you review them
in order from the most recent to the oldest. The meter also calculates and displays 7,
14, 21 and 28-day averages. You can review the individual or average test result by
entering the memory mode.
7. Remove Strip to Turn Meter Off : Your blood glucose
result is automatically stored in the meter memory. Turn
the meter off by removing the test strip.
3. Select and Lance a Puncture Site
*For Fingertip
Hold the prepared lancing device firmly against the side
of your fingertip. Press release button.
*For AST
Please refer to the “About Alternative Site Testing (AST)”
Section. Please consult your healthcare professional
before obtaining blood from site other than your fingertip.
4. Obtain a Blood Sample : Gently massage your finger
or puncture site to obtain the required blood volume.
To perform the test, you need only 0.7 μL of blood
sample. Do not smear the blood sample. To obtain
best accurate result, wipe off the first drop of blood and
gently squeeze another drop of blood.
*blood glucose value with sign
The meter will display a
sign and following 4 warning
buzzer if your test result is lower than 70 mg/dL (3.9
mmol/L) or higher than 180 mg/dL (10 mmol/L). See
Figure 2.
4. Memory Recall
SIDE
2. Insert Test Strip : Remove a new test strip from vial. Be
sure to tightly replace vial cap after removing test strips.
Insert test strip immediately into strip slot as illustrated.
The meter turns on automatically. When the blood
symbol blinking, you are ready to perform a test.
6. Read Your Result :
*blood glucose value with sign
The meter will display a symbol if the result falls in the
range of 70-120 mg/dL (3.9 - 6.7 mmol/L). See Figure 1.
This default value is only for your reference, please
consults your healthcare professional to find out your
target blood glucose value.
Test result is higher than
600 mg/dL (33.3 mmol/L).
Test result is lower than
20 mg/dL (1.1 mmol/L).
Appears when result is within 120-180
mg/dL (6.7 - 10 mmol/L).
Appears every time after doing test, if
you want to delete this record, press
the SET BUTTON (in the battery
compartment) for 3 seconds to delete
the selected test.
7.
System Specifications
Model Name
Assay Method
Test Sample
Test Result
Alternative Site Testing
Sample Size
Measuring Time
Measuring Range
Acceptable Hematocrit Range
Operating Condition
Storage/Transportation Condition
Memory Capacity
Average Calculation
Power Supply
Battery Life
Automatic shut-off
Dimensions
Weight
Battery is dead.
8.
Maybe:
1.Used strip or
moistened strip
2.Defective meter.
Temperature is below the
operating range.
Temperature is above the
operating range.
Deleting is complete.
High or low blood glucose levels can
indicate a possibly serious medical
condition. If this is not confirmed by
the way you feel, review proper testing
procedure and perform a control
test. Repeat blood test, if the display
still appears, please call medical
assistance immediately.
OK-1B
Electrochemical biosensor
Capillary, venous, and arterial whole Blood
Referenced to plasma glucose value
YES (palm, forearm, upper arm, calf, or thigh)
0.7 μL
6 seconds
20 – 600 mg/dL
20~60%
10°C~40°C(50°F~104°F), between 10-85% R. H.
4°C~40°C(39°F~104°F), between 10-85% R. H.
250 test results with time and date
7, 14, 21, and 28 days
One 3-volt Lithium Battery (battery type CR2032)
Approximately 1,000 tests
In 3 minutes
99 x 62 x 20 mm
56g
Replace battery now.
You have to:
1.Repeat test with a new test strip.
2.Contact your local distributor for
help.
The meter is not working. Move to
an area with temperature between
10ºC to 40ºC (50ºF - 104ºF) and wait
at least 30 minutes. Do not artificially
heat or cool the meter.
No responses
when the
test strip is
inserted into
the meter.
Maybe:
1.Battery is dead.
2.Wrong test strip is
inserted.
3.Meter is defective.
You have to:
1.Replace battery
2.Insert the test strip correctly.
3.Contact your local distributor for
help.
No responses
when blood
sample is
applied to the
test strip.
Maybe:
1.Blood sample is not
sufficient.
2.Meter is defective.
You have to:
1.Repeat test with sufficient sample.
2.Contact your local distributor for
help.