The Honeywell Precision Barometer (HPB) offers outstanding value to instrument builders requiring accurate
and stable barometric measurements in real-world conditions. The HPB uses proven silicon sensor technology
with microprocessor-based signal compensation, eliminating the need to insulate or temperature-regulate the
barometer. The HPB has a pressure range of 500 to 1200 hPa. The HPA, intended for altimeter applications,
provides a pressure range of 0 to 17.6 psia.
APPLICATIONS:

AWOS Weather Systems

Remote Meteorological Stations

Ocean Data Buoys

Environmental Data Logging

Secondary Air Data

Altimeters
Outstanding
Value
Small,
Rugged
Design
±0.4 hPa
Accuracy
From -40 to 85ºC
FEATURES AND BENEFITS
ISO-9001
ISO-14001

High Accuracy
0.4 hPa max from -40 to 85C
0.03% FS max from -40 to 85C
 Two-tiered accuracy including temperature errors over -40 to 85C

Multiple Interface Options
Interface, Plug-and-Play for your system requirements.
 Easy
TTL – for lowest power consumption (33 milliwatts)
– HPB, 0.4 hPa or 0.8 hPa; HPA, 0.03% or 0.06% FS Max.
Simplifies System Design – there is no need to insulate, temperatureregulate or provide additional signal compensation.
RS-232 – receives commands and sends data to a single serial port of
a computer.
RS-485 – up to 89 units can be connected to a two-wire multidrop bus.

Proven Honeywell Technology
and Reliable – Honeywell has been building the world’s highest
 Stable
performance silicon pressure sensors for over thirty years.
HPB
SPECIFICATIONS
CASE OUTLINE
(1)
Performance Specifications
Total Error Band : (from -40 to 85ºC)
HPB200: ±0.4 hPa maximum
HPB100: ±0.8 hPa maximum
HPA200: ±0.03% FS maximum
HPA100: ±0.06% FS maximum
Temperature: ±1ºC (at sensing element)
Temperature Range:
Operating -40 to 85ºC (-40 to 185ºF)
Storage: -55 to 90ºC (-67 to 194ºF)
Sample Rate(3): 8.33ms to 51.2 min
Resolution:
0.0011% FS(4)
Response Delay:
(1000/update rate) +1ms, minimum 17ms
Long Term Stability(6):
HPB: 0.25 hPa max per year
HPA: 0.02% FS max per year
Mechanical Specifications
Pressure Ranges:
HPB: 500 to 1200 hPa (1 hPa = 1 mbar)
HPA: 0 to 17.6 psia
Pressure Units(3): atm, bar, cmwc, ftwc, hPa, inHg,
inwc, kg/cm², KPa, mBar, mmHg, MPa, mwc, psi,
user, Icom, pfs
Media Compatibility: Suitable for non-condensing,
non-corrosive, and non-combustible gases.
Weight: 5 oz. (142 gm) without fittings
Electrical Specifications
Output: TTL, RS-232, RS-485
Power Requirements:
TTL Supply Voltage: 6 to 26 VDC
RS-232, RS-485 Supply Voltage: 5.5 to 30 VDC
Operating Current:
RS-232/RS-485: 17-30mA; TTL: 6-9 mA
Baud Rate(3): 1200, 2400, 4800, 9600,
14400, 19200, 28800
Bus Addressing(3): Address up to 89 units.
Environmental Features(2)
Humidity Sensitivity: Negligible
Acceleration Sensitivity: Negligible
Mechanical Shock: 1500g, 0.5ms half sine
Temp Shock: 24 1-hour cycles, -40 to 85ºC
Vibration: 0.5in or 20G’s, 20Hz – 2K Hz
(1) Accuracy is the sum of worst case linearity, repeatability,
hysteresis, thermal effects and calibration errors from -40 to
85ºC. Calibration is traceable to NIST. (2) Mechanical Shock
tested per MIL-STD-883D, M2002.3, Cond. B. Vibration tested
per MIL-STD-883D, M2007.2, Cond. A. (3) User Configurable.
(4) Best resolution obtained with PFS (percent full scale)
pressure units. (5) Demonstration kit includes unit, power
supply/data cable (120V), demonstration software, TTL-to-RS232 converter (TTL only), and user manual. (6) Beyond max.
total error band when continuously powered at 25±10°C,
<90%RH and 28 to 32 inHg atmospheric pressure.
Pin#
Signal Name
1
RS-232(TD)/RS-485(B)
2
RS-232(TD)/RS-485(A)
3
Case Ground
4
Common Ground(GND)
5
DC Power In(+)
6
N/C
Pin #
Signal Name
1
TTL XMIT
2
TTL RCV
3
Case Ground
4
Common Ground(GND)
5
DC Power In(+)
6
Power Control
ORDERING INFORMATION
Honeywell Precision Barometer
APPLICATION TYPE
HPB Barometer Application, 500 to 1200 hPa (1 hPa = 1 mbar)
HPA Altimeter Application, 0 to 17.6 psia
ACCURACY
BAROMETER
ALTIMETER
200
±0.4 hPa max
±0.03% FS max
100
±0.8 hPa max
±0.06% FS max
PRESSURE CONNECTION
W
Brass barbed (1/8 inch ID tubing)
X
Brass SwagelokTM (1/8 inch female)
R
Brass barbed, right angle (1/8 inch ID tubing)
F
Filter (blocks debris)
HPB 200
W
OUTPUTS
2D
RS-232
5D
RS-485
TT
TTL
ELECTRICAL CONFIGURATION AND CONNECTION
A 6-pin plastic connector
OPTIONS
A Demonstration Kit(5) (RS-232, TTL only)
B Mating Connector (See Below)
C Power Supply/Data Cable (RS-232, TTL with Option G
only, See Below)
G TTL to RS-232 Converter (TTL only)
2D A -B
Power Control
(TTL Only)
Find out more
For more information on Honeywell’s Precision Pressure Transducers visit us online at www.pressuresensing.com or
contact us at 1-800-601-3099 (International: 1-602-365-3099). Customer Service Email: D&[email protected].
Honeywell reserves the right to make changes to improve reliability, function or design. Honeywell does not assume any liability arising out of the application or use of any
product or circuit described herein; neither does it convey any license under its patent rights nor the rights of others.
Honeywell
Honeywell International Inc.
12001 Highway 55
Plymouth, MN 55441
www.pressuresensing.com
ADS-14181
July 2014
©2012 Honeywell International Inc.