Non-Touch Thermometer
User Manual
ARC Devices Ltd.
Copyright ARC Devices Ltd. 2015
PT-100153-IMD Rev 00
INTRODUCTION
The ARC InstaTemp MD™ is a sophisticated,
technologically advanced electronic medical
device that has been carefully engineered to
quickly and accurately measure human body
temperature without requiring any contact with
the human body. To ensure the best results with
your ARC InstaTemp MD™ thermometer, please
be sure to read the following operational and
procedural manual carefully. If you give this
device to another adult, it is vital that you
include this user manual.
Intended Use
The ARC InstaTemp MD™ is intended for use by a
Health Care Professional, in a clinical environment,
for the measurement and display of human body
temperature for patients of all ages. This device
is not intended for home use.
The CE Mark of this device refers to the
0123
EC Directive MDD 93/42/EEC.
page 2 of 36
KEY INFORMATION
Only use ARC InstaTemp MD™ in ambient
temperatures ranging from 59–95˚F (15–35˚C).
The thermometer and patient must remain in a
stable temperature environment and away from
ventilation sources and heat sources for at least
20 minutes before measuring temperature.
Always protect the sensor, which is in the recess
at the top of the device (see illustration on page
4) from shocks or contamination. See Precautions
(page 22) and Cleaning (page 25) for more details.
Package Contents
The package contains:
• The ARC InstaTemp MD™
• This instruction booklet
• Wall mount
• Retractable tether
page 3 of 36
GETTING TO KNOW
The ARC InstaTemp MD™
ARC InstaTemp MD™
Infrared Sensor
Backlit LCD
Display Screen
(remove protective label prior to
commencing operation)
Temperature Range
Indicator Light
Activation Button
Hole for tether/wall mount
Internal, non-replaceable
batteries are included
page 4 of 36
GETTING TO KNOW
The ARC InstaTemp MD™
Features
1
2
Accurate body temperature measured
from the forehead in 2.5 seconds
­
LCD screen with backlight for ease of
use in low-light conditions
3
Temperature shown in either degrees
Fahrenheit (˚F) or degrees Celsius (˚C)
4
The temperature reading, when
ready, is displayed on backlit LCD
screen for 4 seconds
5
Turns off automatically when not
in use
6
Fast temperature readings that meet
the ASTM 1965-98 and IEC80601-2-56
standards
7
LED display temperature range indicator system: Green, Orange, Red.
8
Low battery indicator
9
Test mode for LCD screen check
page 5 of 36
DEVICE SETUP
The ARC InstaTemp MD™
Before using the device for the first time:
1
REMOVE FROM PACKAGING
2
READ ALL SECTIONS OF THE IFU.
3
REMOVE THE PROTECTIVE LABEL FROM
THE LCD SCREEN
4
SET THE UNITS
The ARC InstaTemp MD™ can display
temperatures in degrees Fahrenheit or degrees
Celsius. Follow the steps page on 7 to toggle
between °F or °C and select the preferred
measurement units for display.
5
AFFIX THE WALL MOUNT
While not in use, the ARC InstaTemp MD™ should
be stored using the wall mount. In an operational
environment make sure to secure the wall mount
away from any direct heating, cooling or
ventilation systems. Otherwise the ARC
InstaTemp MD™ should be stored in a manner
which protects the sensor from contamination or
damage. [See Carrying and Storage, page 21]
6
ATTACH THE RETRACTING TETHER
The retracting tether can be used to safely carry
the ARC InstaTemp MD™ between uses. While
not in regular use, the ARC InstaTemp MD™
should be returned to the wall mount.
page 6 of 36
OPERATIONS
Changing Between Fahrenheit and Celsius:
The ARC InstaTemp MD™ is set to Fahrenheit by
default. To change to Celcius, follow these simple
steps:
1
Press and hold the activation
button to get a reading, which
can be a temperature reading,
Lo, Hi or “---”.
2
Once the reading appears on
screen in Fahrenheit, press and
release the activation button again
—the reading will automatically
change to Celsius.
Repeat steps 1 and 2 to change back.
!
Note: Once the display units are set (°F or °C),
they remain as the default display units until
changed again
page 7 of 36
OPERATIONS
1
Prepare To Take A Reading
Prepare the patient.
Move hair if present and all
obstructions away from the path of
the sensor
Wipe perspiration from the forehead,
if apparent
Ensure that the person has been in a
stable temperature environment for
20 minutes
Do not touch the lens of the sensor or
the area surrounding the front of the
device. It is the most delicate part of
the ARC InstaTemp MD™
page 8 of 36
4
Protect from
rain
Keep Dry
5
Not to be
used
in case
package is
damaged
6
7
LOT
8
Do not measure the person’s
temperature immediately after
he/she has been exposed to really
Caution
consult
documents
cold or hot climates, including
any heating or cooling systems
in the operational environment,
Caution
consult
if the person has been wearing a
documents
hat, or has consumed hot or cold
food or liquids as it may
temporarily change the surface
Lot Code
See
Base of
temperature
of the skin.
device
In these instances, wait 20 minutes before taking a temperature
reading.
page 9 of 36
OPERATIONS
Taking A Temperature
2
0.5” - 1.25”
(1-3cm)
1
NON
STERILE
A
2
Storage and
transport
temperature
Hold for
2.5 seconds
Non Sterile
Do not
to
Point
the
thermometer at the patient,
3 expose
sunlight
Keep away
from heat
without
touching them, as shown above.
Ensure the device is approximately 0.5-1.25”
4 Protect from
rain
Keep Dry
(between
1 and 3cm) from the forehead,
above the eyebrows and between the eyes.
5
Not to be
used
in case
package is
damaged
Thermometer must be held steady,
Caution
perpendicular
to and pointing directly at
consult
documents
the center of the forehead at the area above
the eyebrows and between the eyes when
taking a temperature.
7 Caution
6
consult
documents
LOT
8
Lot Code
See Base of
device
page 10 of 36
OPERATIONS
Hold for
2.5 seconds
™
NON
STERILE
B
™
1
Storage and
transport
temperature
2
Non Sterile
™
™
Hold the thermometer in position. Press and
Do not
hold
the
activation button until the display
to
3 expose
sunlight
Keep away
from heat
is illuminated
and a temperature reading
is displayed on the screen. This will take
approximately 2.5 seconds.
4
Protect from
rain
Keep Dry
The activation button should be released
when
the display back light turns on and the
5 Not to be
used
in case
package is
result
damagedis displayed on the screen.
Keep
the device as steady as possible
Caution
consult
documents
during
the measurement process. The LCD
backlight is off and displays “188.8” during
the measurement process. The backlight of
7 the
CautionLCD comes on when the measurement
consult
documents
process is complete.
6
LOT
8
Lot Code
See Base of
device
page 11 of 36
OPERATIONS
Reading the Temperature Indicator
3
The temperature will be displayed for 4 seconds
on the display screen. The LCD display will show a
temperature in ˚F or ˚C.
1
NON
STERILE
2
1
3
NON
STERILE
2
Storage and
transport
temperature
Non Sterile
Storage and
transport
temperature
Do not
expose to
sunlight
Keep away
from heat
Non Sterile
Do not from
4 Protect
expose to
3 rain
sunlight
Keep
Dry
Keep away
from heat
5
4
Not to be
used
in case
package is
damaged
Protect from
rain
Keep Dry
The temperature
range indicator will
be green, orange or
red, depending on
how high or low the
temperature is from
normal. [See the
Operations section on
page 13 for a detailed
explanation of the
temperature ranges.]
Should you wish to take a second reading, wait for
6 Caution
the
LCD display to turn off first. Wait 15 seconds
consult
documents
to be
5 Not
used
between
measurements.
in case
package is
damaged
Holding the device in your hand for prolonged
7 Caution
consult
6 documents
Caution of time may affect the readings. It is not
periods
consult
documents
recommended to hold the device in your hand
between measurements.
LOT
8
7
Lot Code
Caution
See Base of
consult
device
documents
page 12 of 36
LOT
8
Lot Code
See Base of
3
Do not
expose to
sunlight
Keep away
from heat
OPERATIONS
Understanding the Temperature Range Indicator Color
4
Protect from
rain
Keep Dry
The temperature range indicator will display
5 Not to becolors: green, orange, or red.
one of three
used
in case
package is
damaged
6
LOT
The
temperature range indicator
Caution
consult
documents
colors
indicate the degree to which the
patient’s temperature is outside the
“normal” range.
7
Caution
consult
documents
8
Lot Code
See Base of
device
Green: Temperatures within
normal range. (see tables on
pages 14 & 15)
Orange: Temperature is mildly
outside of normal range
(whether hot or cold)
(see tables on pages 14 & 15)
Red: Temperature is excessively
outside of normal range
(whether hot or cold)
(see tables on pages 14 & 15)
page 13 of 36
OPERATIONS
D
or
ca
t
di
In
Temperature Range
is
pl
ay
Li
FAHRENHEIT (˚F)
Re
ad
gh
t
ou
t
Understanding The Temperature Range
>109.1°F
HI
1 Storage and
≥100.5⁰F
- ≤109.1⁰F
transport
Temp in ˚F
≥99.0⁰F - <100.5⁰F
Temp in ˚F
temperature
NON
STERILE
1
2
Non Sterile
Storage
and
transport
temperature
≥96.7⁰F - <99.0⁰F
3
Do not
expose to
sunlight
Keep away
from Sterile
heat
Non
Temp in ˚F
2
≥94.9⁰F
- <96.7⁰F
Temp in ˚F
Do not
expose
to
≥89.5⁰F
- <94.9⁰F
43 Protect
from
sunlight
rain
Temp in ˚F
<89.5⁰F
5 Not to be
used
LO
NON
STERILE
KeepDry
away
Keep
from heat
4
in case
package is
Protect
from
damaged
rain
Keep Dry
Note:
6 Caution a red indicator light may mean “very
consult
to be
5 Not
documents
used
hot”
in caseor “very cold”.
package is
damaged
Outside the measurable range, 89.5˚ - 109.1°F,
the screen shows “HI” or “LO” as applicable.
LOT
7
6
Caution
consult
Caution
documents
consult
documents
7
8
Caution
Lot Code
consult
See Base of
documents
device
page 14 of 36
8
Lot Code
OPERATIONS
is
pl
ay
D
In
Temperature Range
di
ca
t
or
Li
CELSIUS (˚C)
Re
ad
gh
t
ou
t
Understanding The Temperature Range
>42.8°C
HI
1 Storage and
≥38.0⁰C
- ≤42.8⁰C
transport
temperature
Temp in ˚C
≥37.2⁰C - <38.0⁰C
Temp in ˚C
NON
STERILE
1
2
Non Sterile
Storage
and
transport
temperature
≥36.0⁰C - <37.2⁰C
3
Do not
expose to
sunlight
Keep away
from Sterile
heat
Non
≥35.0⁰C - <36.0⁰C
NON
STERILE
2
expose
to
43 Protect
from
≥32.0⁰C
- <35.0⁰C
sunlight
rain
Keep away
Do not
Temp in ˚C
Temp in ˚C
Temp in ˚C
Keep Dry
from heat
LO
5 Not to be
<32°C
used
4
in case
package is
Protect
from
damaged
rain
Keep Dry
Note:
a red indicator light may mean “very
Caution
consult
Not
to be
documents
used
hot”
in case or “very cold”.
package is
6
5
damaged
Outside the measurable range, 32˚ - 42.8˚°C,
the screen shows “HI” or “LO” as applicable.
LOT
7
6
Caution
consult
Caution
documents
consult
documents
7
8
Caution
Lot Code
consult
See Base of
documents
device
page 15 of 36
8
Lot Code
OPERATIONS
Temperature Range “HI” & “LO”
A “HI” Reading
The LCD will display ‘HI’. This
indicates that the temperature
which the thermometer has
measured is outside the
device’s measurement range
(i.e. >109.1°F/ >42.8°C).
Above 109.1º F / 42.8˚ C
A “LO” Reading
The LCD will display ‘LO’. This
indicates that the temperature
which the thermometer has
measured is outside the
device’s measurement range
(i.e. <89.5° F/ <32° C).
Below 89.5º F / 32˚ C
page 16 of 36
OPERATIONS
1
Storage and
transport
temperature
Other Screen Indications
NON
STERILE
Indication
2
Non Sterile
MeaningDo not
3
expose to
sunlight
Keep away
from heat
After 4 seconds the
4 Protect LCD
from
backlight,
display and
rain
Keep Dry
temperature range indicator
turn off
and the thermome5 Not to be
1 used
Storage and
intransport
case
ter powers
temperature
package
isdown.
damaged
NON
STERILE
If the display remains
6 Caution
in2 consult
this
documents
Non Sterilestate after the
button is pressed then
the battery is no longer
Do not
7 Caution
working.
expose to
3 consult
sunlight
documents
Keep away
from heat
An Error message indicates
an internal problem with
LOT
8
4
Lot Code
Protect
from
See Base of
rain
deviceDry
Keep
the thermometer
(see pg.
26
to befor error codes).
5 Not
used
in case
package is
damaged
If an Error message
6 Caution
is
displayed, please
consult
documents
contact ARC Devices or
your distributor.
7
Caution
consult
documents
Contact info on page 36 of this document.
LOT
8
Lot Code
See Base of
device
page 17 of 36
OPERATIONS
Other Screen Indications
Indication
Meaning
This indicates an invalid
reading. This is due to incorrect
usage, such as:
• trying to measure a temperature when the ambient temperature is outside the operating
temperature range.
• an indeterminate reading
because the activation button
was not held until the display
was illuminated.
• temperature measurement
failed due to too much
movement by the operator or
patient. Wait 15 seconds and
repeat the temperature.
Repeat the temperature
measurement as described in
the section titled “Taking the
Temperature.” (page 10)
page 18 of 36
transport
temperature
OPERATIONS
Non Sterile
Other Screen2 Indications
NON
STERILE
Indication
3
Do not
expose to
sunlight
Keep away
from heat
Meaning
Flashing
battery
symbol will
4 Protect
from
rain
Keep Dry
start being displayed when
there is battery power for
51 Not to be
Storage and
used
intransport
case
temperature
approx. 500
left.
package measurements
is
damaged
NON
STERILE
Contact ARC Devices or
6 Caution
2 consult
Non Sterile
your
distributor for a
documents
replacement unit.
Do not
expose to
73 Caution
sunlight
consult
Keep away
documents
from heat
Steady battery symbol & red
indicator light: insufficient
4 Protect
from
Lot Code
rain
See Base
of
LOT 8 power
battery
for a
Keep
deviceDry
measurement. Replace your
5 Not to be
ARC InstaTemp
MD™
used
in case
package is
damaged
Contact
ARC Devices or
6 Caution
consult
documents
your
distributor for a
replacement unit.
LOT
7
Caution
consult
documents
8
Lot Code
See Base of
device
page 19 of 36
NON
STERILE
CARRYING & STORAGE
2
Non Sterile
3
Do not
expose to
sunlight
Keep away
from heat
The wall mount is provided to help protect the
sensor in its recess position by mounting the
ARC InstaTemp
MD™ on the wall mount with the
4 Protect from
rain
Keep Dry
sensor facing toward the floor. The retractable
tether is provided as means of carrying the ARC
5 Not to be
used
InstaTemp
MD™ and keeping the device at
in case
package is
damaged
ambient temperature.
Caution
consult
Note:
You should choose to use either the
documents
wall mount or the retractable tether, as
they do not work together. Do not place
the wall mount in any location with direct
7 Caution
cooling
or ventilation.
consult
documents
6
Wall Mount
LOT
8
Lot Code
See Base of
device
™
™
The wall mount can be attached to any
smooth vertical surface using its adhesive pad
on the rear of the wall mount. Because it is
magnetic, the wall mount can also be attached
to any metal surface. The internal magnet
also keeps the ARC InstaTemp MD™ gently
attached to the wall mount.
page 20 of 36
CARRYING & STORAGE
ARC InstaTemp MD™ should be attached
with its sensor facing downwards, as shown
in the illustration on page 20.
Retractable Tether
The extending tether is attached by threading
the lanyard through the hole in the body of the
ARC InstaTemp MD™, as shown in the
illustration
below.
The retractable tether can
1 Storage
and
transport
temperature
then be attached
to a suitable exterior pocket,
waistband or belt.
™
NON
STERILE
2
Non Sterile
3
Do not
expose to
sunlight
Keep away
from heat
4
Protect from
rain
Keep Dry
5
Not to be
used
in case
package is
damaged
Body
heat can raise the ambient
temperature of the thermometer or
contaminants
may lodge in the sensor,
6 Caution
consult
documents of which can affect the ability to
either
make accurate measurements.
Therefore, do not carry the ARC InstaTemp MD™ under your clothing or in
7 Caution
consult pocket.
your
documents
LOT
8
Lot Code
See Base of
device
page 21 of 36
PRECAUTIONS
95˚
1
35˚
Storage and
transport
temperature
Only use the ARC InstaTemp MD™
in climate conditions ranging from
59-95°F (15-35°C). Outside of this
range, a "---" will be displayed when
a measurement is taken. Move the
15˚C device and patient to an area inside
the recommended temperature range,
wait 20 minutes for temperatures to
stabilize, then take a patient temperature measurement.
NON
STERILE
59˚F
The thermometer and the patient
must remain in a stable temperature
environment for at least 20 minutes
before a reading can be taken. Do not
measure the patient’s temperature
immediately after he/she has been
exposed to really cold or hot climates,
or has has consumed hot or cold food
or liquids. In these instances, wait 20
minutes before taking a temperature
reading. Do not apply any hot or cold
compress, cloth or gel, on the patient’s
forehead prior to the measurement
for at least 20 minutes.
2
Non Sterile
3
Do not
expose to
sunlight
Keep away
from heat
4
Protect from
rain
Keep Dry
5
Not to be
used
in case
package is
damaged
6
7
If the ARC InstaTemp MD™ touches or
comes into contact with the patient follow
your local procedure for cleaning devices
before using with another patient.
Caution
consult
documents
Caution
consult
documents
page 22 of 36
LOT
8
Lot Code
See Base of
device
PRECAUTIONS
Do not use or store outside or close
to a heat or cooling source. Using the
thermometer in these conditions may
affect the accuracy of the temperature
reading.
Do not drop, tap or shake the ARC
InstaTemp MD™ thermometer. It is a
precision device and any damage may
result in an inaccurate reading.
NON
STERILE
1
Storage and
transport
temperature
2
Non Sterile
Do not allow the ARC InstaTemp MD™
thermometer to become damp or immersed in water or other liquids.
3
1
NON
STERILE
Do not
expose to
sunlight
Storage and
Keep away
transport
from
heat
temperature
Do not touch the lens of the infrared
sensor, any damage or scratches may
affect the accuracy or function of the
ARC InstaTemp MD™.
4 Protect from
2 rain
Non Sterile
Keep Dry
Not to be
Do not
used
case to
3 inexpose
sunlight is
package
Keep away
damaged
from heat
5
7
5
Do not attempt to take a patient
temperature measurement if there is
any evidence of damage to the device
or if it has been exposed to extreme
conditions.
6
8
Storage temperature range 45 to 90°F
(7-32°C)
6
4
LOT
7
Caution
consult
documents
Protect from
rain
Keep Dry
Caution
Not to be
consult
used
documents
in case
package is
damaged
Caution
consult
Lot Code
documents
See Base of
device
Caution
consult
documents
page 23 of 36
LOT
8
Lot Code
See Base of
43
NON
STERILE
2
Do not
Protect
exposefrom
to
rain
sunlight
Keep
KeepDry
away
from heat
Non Sterile
PRECAUTIONS
Not to be
used
in case
package is
Do not
damaged
4 Protect
exposefrom
to
3 rain
sunlight
Keep
KeepDry
away
from heat
1 Storage and
transport
temperature
6 Caution
consult
to be
5 Not
documents
used
in case
package is
damaged
4 Protect
from
rain
Keep Dry
2 Non Sterile
5
Do not use in oxygen rich
environments.
NON
STERILE
7
6
LOT
78
6
LOT
78
4
5
LOT
8
6
7
LOT
Do not use in or close to strong
magnetic fields or X-Ray sources.
Caution
consult
Caution
documents
consult
Not
to be
documents
used
in case
package
Do not is
damaged
expose to
3 sunlight
Keep away
from heat
5
8
Failure to understand these precautions and warning statements provided in this section of the document
could lead to device damage and or
inaccurate temperature measurements.
Caution
Lot Code
consult
See Base of
documents
Caution
device
consult
documents
Protect from
rain
Keep Dry
Lot Code
Caution
See Base of
consult
device
documents
Not to be
used
in case
package is
damaged
If the informational content of the
LCD display is in question then the
LCD display test mode (see troubleshooting section, page 26), should be
used to check the display.
Lot Code
See Base of
Caution
device
consult
documents
Caution
consult
documents
Lot Code
See Base of
device
page 24 of 36
CARE OF DEVICE
Do not touch the lens of the sensor.
It is the most delicate part of the ARC
InstaTemp MD™ thermometer and
must be protected from dirt and
damage.
Do not immerse the ARC InstaTemp
MD™ in water or other liquids. It is
not water resistant.
Take care not to scratch the lens
of the sensor, the display or the
housing.
Never use aggressive cleaning
agents.
When cleaning is required, the
external surfaces of the thermometer, Wall Mount and Retractable
tether can becan be safely wiped
with a cloth moistened with a 70%
IPA (isopropyl alcohol) solution.
WARNING: DO NOT immerse or
soak the thermometer in any type
of fluid, or use any heat or steam
sterilization process.
page 25 of 36
TROUBLESHOOTING
Error Codes
The table below lists the error codes which
the ARC InstaTemp MD™ can display. If the
device continues to display one of these
errors, the device should be returned to the
manufacturer.
Error Code
Explanation
Er1
Sensor Voltage Error
Er2
Sensor Configuration Error
Er3
Sensor Self-Test Error
Er4
Sensor Read Error
Er5
Core Processor Frequency
Er6
System Configuration Memory
page 26 of 36
TROUBLESHOOTING
LCD/LED Display Testing
The LCD can be tested for consistency at
regular intervals, or if any LCD figures seem
incomplete, by activating the display test
mode in the following manner:
Press the activation button and hold until
after the display period is complete. The ARC
InstaTemp MD™ then enters test mode and
runs the display test once.
During Test Mode, the device behaves as
follows:
The display turns off for 1 second,
1
then turns on with all segments active.
2
The LCD then cycles through the
following sequence to exercise all the
segment combinations:
1. The tens and units values of the
display will increment from 0 to 9.
2. The hundreds unit will toggle
between the upper and lower
segment.
page 27 of 36
TROUBLESHOOTING
3. The ˚F and ˚C symbols will be
display alternately.
4. The battery symbol and decimal
point will both turn on and off
alternately.
5. The green and red LEDs will be
display alternately.
If any of the above steps fail, the device
may be faulty and should not be used.
Should this occur, please contact ARC
Devices customer service (see page 35).
page 28 of 36
WARNINGS
Do not attempt to fix or repair the
thermometer yourself; there are no
user-serviceable parts in the
ARC InstaTemp MD™.
1
NON
STERILE
Storage and
transport
temperature
Do not cover the sensor aperture
wholly or partially, or insert any
objects or material into the aperture.
2
Non Sterile
Do not attempt to replace the internal
batteries. The internal batteries are
non-removable, non-replaceable and
non-serviceable.
3
4
5
6
7
Do not
expose to
sunlight
Keep away
from heat
Protect from
rain
Keep Dry
Do not open the thermometer casing.
If the case is not correctly closed, do
not use the device.
Not to be
used
in case
package is
damaged
Avoid Strong magnetic fields.
Avoid X-ray.
Avoid oxygen rich environment.
Caution
consult
documents
DISPOSAL
Caution
consult
documents
The ARC InstaTemp MD™ contains batteries:
doLOT
not dispose in fire, or domestic waste. Dispose
!
in accordance
with local regulations at local
authority collection points
8
Lot Code
See Base of
device
For more information about disposal please contact
ARC Devices customer service (see page 35).
page 29 of 36
ICON IDENTIFICATION
This instruction manual belongs to
this device. It contains important
information about operating the ARC
InstaTemp MD™ non-touch thermometer. Read the instruction manual
carefully as non-observance of these
instructions can result in inaccurate
temperature readings or damage to
the device
!
Manufacturer
!
Caution; consult section titled
‘Understanding the Temperature
Range Indicator System’ and ‘Taking
a Temperature’.
!
!
Degree of protection against electrical
shock; there is no conductive connection to the person; Type BF Protection
page 30 of 36
ICON IDENTIFICATION
Device to be disposed of in
segregated waste (WEEE Directive
2012/19/EU)
Protect from liquids
!
!
!
!
Device conforms with RoHS
directive (2002/95/EC)
0123
The CE Mark of this device refers to
the EC Directive MDD 93/42/EEC
page 31 of 36
TECHNICAL SPECIFICATIONS
Item
Value
Model
ARC InstaTemp MD™
Display System
Digital segmented display
(LCD) with backlight
Reported
measurement
range
89.6 - 109.1°F (32 - 42.8°C)
Operating
temperature /
humidity range
59 - 95 °F (15 - 35°C)
Operational up to 95%
humidity non-condensing
Storage/ Transport
Temperature range
Temperature
measurement
accuracy
Temperature
display
45°F - 90°F (7°C - 30°C)
+/- 0.4° F and +/- 0.2° C in the
measurement range 89.6 109.1°F (32 - 42.8°C) whilst
inanambienttemperature
range of 59 - 95 °F (15 - 35°C)
(In accordance with ASTM
E1965-98andISO80601-2-56
pt 2-56)
Units °F or °C, user
selectable
page 32 of 36
TECHNICAL SPECIFICATIONS
Item
Value
Power supply
Contains two AAAA
alkaline batteries, nonreplaceable
Battery life
Approximately 13,000
measurements. Low
battery warning displays
when approximately 500
measurements left.
Thermometer
Dimensions
1.37” x 0.55” x 5.08”
(3.5cm x 1.4cm x 12.9cm)
Weight
1.94 oz (55 gm)
Wall Mount &
Retractable Tether
Wall Mount:
• 1.24” x 2.125” x 0.42”
(31.5mm x 54mm x 10.7mm)
Tether:
• 1.4” x 2.12” x 0.56”
(35.3mm x 53.8mm x14.2mm)
• Retractable cord length:
27’’ (69cm)
Compliance with
standards
Electromagnetic
compatibility:
IEC60601-1-2 Clinical
thermometer:
ASTME1965-98(Reapproved
2009), ISO 80601-2-56
page 33 of 36
LIMITED TWO YEAR WARRANTY
ARC Devices warrants the product to be free of defects in material and workmanship and to perform in accordance with manufacturer’s specifications for the period of two years from the
date of purchase from ARC Devices, its authorized distributors or
agents. The warranty period shall start on the date of purchase.
The date of purchase is: 1) the invoiced ship date if the device was
purchased directly from ARC Devices, 2) the date specified during
product registration, 3) the date of purchase of the product from
an ARC Devices authorized distributor as documented from a receipt from said distributor. This warranty does not cover damage
caused by: 1) handling during shipping, 2) use or maintenance
contrary to labeled instructions, 3) alteration or repair by anyone
not authorized by ARC Devices, and 4) accidents.
Shipping cost to return a device to ARC Devices or an ARC Devices
Service center is not included.
A service notification number must be obtained from ARC
Devices prior to returning any products or accessories to ARC
Devices’ designated service centers for repair. To obtain a service
notification number, contact ARC Devices Technical Support.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE. ARC DEVICES’S OBLIGATION
UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS CONTAINING A DEFECT. ARC DEVICES IS
NOT RESPONSIBLE FOR ANY INDIRECT OR CONSEQUENTIAL
DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY
THE WARRANTY
page 34 of 36
LIMITED TWO YEAR WARRANTY
How to return your thermometer
Should service be required during or after the
warranty period, you must call ARC InstaTemp
MD™ Customer Service Department to obtain
a return authorization number and shipping
address. Repackage the thermometer carefully
in its original packaging or securely pack to avoid
damage during shipping. Include the original
sales slip indicating the date of purchase, a note
describing the problem, your return authorization
number, and your return address.
Send the ARC InstaTemp MD™ thermometer
prepaid and insured.
For Customer Service Department
contact:
USA and Canada: +1 561 287 5487
Monday - Friday 8:00 AM to 5:00 PM ET
EMEA and RoW: +353 91 886 002
Monday - Friday 8:00 AM to 5:00 PM GMT
page 35 of 36
CONTACT ARC DEVICES
Designed, manufactured and distributed in all
territories except USA by:
ARC Devices Ltd
2 Deerpark Industrial Estate
Oranmore
Co Galway
!
Ireland
Phone: +353 (91) 398 191
EMail: [email protected]
Web: www.arcdevices.com
Distributed in USA by:
ARC Devices USA, Inc.
1200 N. Federal Hwy, Ste 207,
Boca Raton, FL 33432
USA
Phone: 561.282.6074
0123
PT-100153-IMD Rev 00
page 36 of 36