USER MANUAL Download

Transcript
USER MANUAL
19
__________________________________________________________________________________________________________________
15.3. Pieces counting function (PCS)
12.05 g
T
0.00 g
5
23.04 g
PCS
MENU
T
PCS oFF
PCS on
5
0 pcs
12
12 pcs
T
T
The function allows counting
identical pieces, e.g. pills or buttons
included into weighed portion.
Measurement is performed in two
phases:
- first phase – calculating the mass of
single item basing on a sample
containing defined amount of
pieces: 5, 10, 20, 50, 100, 200 or
500 pieces,
- second phase – counting pieces in
weighed portion.
The function has the following
options:
-PCS OFF – switch the function off
-PCS on – switch the function on
-PCS ... – recover last used unitary
mass,
-PCS 5, 10, 20, ... , 500 – amount of
pieces in a sample,
-PCS SEt – set any amount of pieces
in a sample,
-PCS MJ – set unitary mass directly,
-out – leave menu without changes.
Comments:
1. It is recommended that mass of
single piece is greater than reading
unit and mass of sample used in
first phase is bigger than 100
reading units.
2. Err-PCS message indicates that a
sample was not put on a pan or a
mass of single piece is less than one
reading unit (it is possible to count
pieces knowing that errors will be
bigger).