Download ThunderScientificPsy.. - Chicago Classic Computing
Transcript
I f TH u ND ER s c IE NTI FI c CORPORATION . I ELECTRONIC PSYCHROMETE R MODEL 4AI - MP RELATIVE HUMIDTY · AND TEMPERATURE MEASUREMENT SYSTEM OPERATING AND SERVICE MANUAL _J DATE 623 WYOMING, S.E. ~ : j . ALBUQUERQUE , NEW MEXICO 87123 June 17, 1987 TEL. (505) 265-8701 TABLE OF CONTENTS PAGE WARRANTY . . . . . . . SPECIFICATION S/FEATURES GENERAL DESCRIPTION . . . . . . . . . • . . ... . . PRE-OPERATIONAL INSPECTION . SET-UP/TURN-O N PROCEDURES . . . . . 1 2 3 4 ·- . . . . . . . 5 PROGRAMMING THE SYSTEM FOR: MANUAL OPERATIO~ . . . . . 6 . . . HARDCOPY (PRINTER) OUTPUT . AUTO MODE OPERATION .•. ' 8 9 •· CHANGING THE: ( \ " i ...._,, I . HARDCOPY TIME INTERVAL 10 PRESSURE 11 • • TIME . . 12 DATE . . 12 SWITCHING TO: 0 c . . o F • • 13 . ·• . 13 SUMMARY OF COMMANDS: DISPLAY FUNCTIONS 14 ON/OFF FUNCTIONS 15 SET FUNCTIONS 16 PROBE ~ICK CHANGEOUT INSTRdCTIONS FACTORY CERTIFICATIO N SERVICE 17 . DRAWINGS L) M83Cl29 FRONT PANtL DETAIL DE83C023 PSYCHROMETER PROBE DETAIL DRAWING S86D016 PSYCHROMETER SCHEMATIC DIAGRAM S83Al30 DISPLAY BOARD SCHEMATIC DIAGRAM S83Bl31 PROBE AND CABLE WIRING DIAGRAM --\_) i (4A~lMP) . . 19 ; l \._ j u WARRANTY Thunder Scientific Corp0ration warrants each instrument of its manufacture to be free from. defect in ma_terials and workmanship. The limit of liability under this warranty is to repair or replace any part thereof which shall within one year after delivery t0 the original purchaser, and returned by the original purchaser, prove to be defective after examination by us. This warranty specifically excludes the probe sensors, which may become damaged or broken due to wick changes. Shipping damage is covered by insurance and will be honored if report is made within one week of receipt of the instrument. Warranty returns mu.st be authorized by the fact0ry. ( ) u u ·u WARNING -- Calibration seals have been applied to the instrument case, cable, and probe. Should the instrument and/or associated components be opened or tampered with, or if the seals are broken, Thunder Scientific Corporation accepts no responsibili ty for calibration and this warranty is void. (j v.· i 4A-1MP - MICROPROCESSOR BASErr WET BULB, DRY BULB PSYCHROMETER SPECIFICATIONS Relative Humidity (RH) Measurement Range: Guaranteed Accuracy: Typical Accuracy: 1% RH to 100% RH +1% indicated RH +,5% indicated RH Temperature Measurement Range: Guaranteed Accuracy: Typical Accuracy: 0~100°c / 32-212°F +.3°C / +.5°F +.1°C / +.2°F Readout: Digital Display: Printer Interface: c or °F (user selectable) ,01%RH and .01° Resolution Trendcom Parallel (std) Centronics Parallel (opt) RS-232C Serial (opt) Trendcom 200 High Speed Intelligent Printer 0 Printer (optional) FEATURES ( ) u * Z-80 Central Processing Unit, automatically computes percent relative humidity and both temperature and dewpoint in ° C and 0 1''. * 3 computer/printer interface options: 2 parallel, 1 serial. * Battery backed-up real time clock. * Timed printer output mode allowing user selectable intervals from 1 minute to 24 hrs in 1 minute increments. * Automatic Mode capability enables probe aspirator minute prior to each printout. * Full function display capability. * Not susceptible to contamination by chemicals or long term exposure to high temperature and relative humidity. * Certification of Traceability direct to the National Bureau of Standards. u u 2 for 1 , \.. ) ' (;· GENERAL DESCRIPTION The Thunder Scientific Model 4A-l-MP is a microproc essor based Laboratory Standard Psychrome ter designed for precise measureme nt of air temperatu re and relative humidity. The 4A-l-MP uses a 2-bulb measureme nt system employing wet and dry matched pair calibrated Platinum Resistanc e Thermome ters (PRT's) operating under a controlled air velocity. One PRT, referred to as the dry bulb, is used for sensing air temperatu re. The other PRT, the wet bulb, uses a specially weaved wick (or sock) to facilitate an unimpeded capillary flow of distilled water from a reservoir below. The wet bulb measures a depression temperatu re. The signals from the PRT's are converted to digital format, read by the micro~rocessor, linearized , and then converted to actual temperatu re data in both degrees c and degrees F. The data is then used to calculate Percent Relative Humidity and Dew Point Temperatu re. Front panel display of any of the above data is selected by user input at the keyboard. The upper display is the data readout device, and the lower display indicates which data or other function is currently enabled. Upon selection of data output to this real-time LED display, current readings are updated 4 to 5 times each second, allowing for fast system response to any changes in the sampled environme nt. Printer output, enabled from the keyboard, allows for hardcopy printouts of all current psychrome tric data. Printer output is at user selectable timed intervals ranging from a 1 minute minimum to a 24 hour maximum, or at any 1 minute increment in between. An Auto Mode feature is also included which, when enabled, will put the probe in synchrono us operation with the printer. In this mode, the probe aspirator will automatic ally run prior to any sample for hardcopy output. After printout, the aspirator will be disabled and will remain off until the next sample time is reached. The Model 4A-l-MP is a precise laboratory instrumen t and, if handled according ly, will provide years of excellent service. Q) Q_/ 3 u u ' \ PRE-OPERATIONAL INSPECTION The 4A-l-MP Psychrome ter is shipped from Thunder Scientifi c Corp (TSC) calibrated and basically ready for operation . Any damage from shipment should be reported to TSC within one week of receipt. Prior to each use, the following inspection should be performed : 1) Inventory . . enclosed. Ensure all necessary parts and component s are 2) Inspect the probe for cl~anliness. Pay particula r attention to the wick on the wet bulb. If the wick requires changing see the section titled PROBE WICK CHANGEOUT INSTRUCTI ONS. 3) Inspect the main psychrome ter unit. Ensure the system is clean, that all screws and connettin g hardware are tightened , and that all k~ys operate smoothly without binding or sticking. 4) Inspect cables for wear. nicks, cuts, or frays, easily without binding. Cables should be free from any and all connector s should mate -,/.- ( \ v 4 SYSTEM SET-UP/TURN-ON PROCEDURES ( 1) Remove the black probe housing from its styrofoam holder in the display panel. This housing contains the Platinum Resistance Thermometers (PRT's). A clean wick. is installed from the factory on the wet bulb sensor. To the rear of the housing is the aspirator motor which draws air over the PRT's. 2) Remove the housing. 3) Fill the reservoir with pure distilled water. tap water as this will contaminate the w,ick.). 4) Replace the filler cap and finger tighten, then gently shake excess water from inside the white probe cylinder. Check for leaks around the bottom edge of the housing. Finger tighten the knurled knob as necessary. 5) Place the probe in the environment to be tested and then route and connect one end of the white 12 foot power/signal cable to the connector on the front of the probe housing. Connection to the probe should be smooth, as excessive force could misalign and damage the connectors. 6) Connect on the smooth. 7) Connect printer (optional) to directly above the probe cable. 8) Connect the 6 foot AC power cable to the display panel and then to a 120 VAC source. 9) Turn on the main power switch. 10) The upper display should show the time in military format, with a blinking decimal point in the middle (the seconds indicator). 11) The lower display should show Program 7, the time). 12) Push the LED TEST button and ensure that all segments all 4 decimal points of each display light up. ) \..._;' u u knurled filler cap on the bottom of the probe (N.ever use the other end of the cable to the mating connector display panel. Connection here should also be TURN ON IS COMPLETE. 5 the P--7 multipin connector jack. (meaning in the Display and \.__,;; PROGRAMMING THE SYSTEM FOR MANUAL OPERATION \_j The 4A-l-MP is a. versatile instrument and therefore requires some input from the user. The user mu~t perform at least the following minimum requirements to operate in manual mode: * Ensure the 4A-l-MP displays the correct time. * Ensure the 4A-l-MP displays the corr~ct date. * Input the atmospheric pressure at the probe environment. * Turn on the aspirator motor. * Select the desired display~ ie Temp, Dew Point, RH, etc. * Turn off the aspirator when measurements are complete. Detailed descriptions follow: 1) Display the TIME with the following key strokes: If the time shown on the upper display is incorrect, follow the procedure on CHANGING THE TIME. \_,,· 2) Display the DATE with the following key strokes: If the date shown on the upper display is incorrect, follow the procedure on CHANGING THE DATE. 3) Input the correct barometric pressure in inches of mercury. For example, a pressure of 26.53 in Hg would ~e input with the following key strokes: For more detailed information, CHANGING THE PRESSURE. 4) refer to the section titled Turn on the aspirator with the following key strokes: The motor should turn on and the left decimal point of the lower display should light. 6 5) Select the desired display output with these key strokes: Where 6) X is one of the following: 1 - % RH 2 3 4 5 6 7 8 Dew Point Temperature Air Temperature Wet-bulb Temperature Resistance of Dry-bulb in Ohms Resistance of Wet-bulb in Ohms Time Date - Once measurement is complete, turn off aspirator with: ( ) ( ; \_/ Take additional measurements by repeating steps 4 thru 6. IMPORTANT NOTE The fan mounted in the probe housing is designed with a maximum operating life of 300 hours. The motor need not be run for more than 10 minutes before taking Temperature, Dew Point, or Relative Humidity Measurements. v I ' 7 (-,,__/' PROGRAMMING THE SYSTEM FOR HARDCOPY OUTPUT If a printer is connecte d to the 4A-l-MP via the multipin connecto r on the display panel, hardcopy outputs of all pertinen t psychro metric data are availabl~. The followin g sequence s are used to control the printer output. Informa tion is sent to the printer once each minute unless the user changes the time interval (see CHANGING THE HARDCOPY TIME INTERVA LS). * Turn on the printer function with the followin g kaystro~es: The second decimal point from the left on the lower display should light. * Turn off the printer function with the followin g keystrok es: ,' \ ) i._! The second decimal point should go out. When using the printer, the aspirato r should also be turned on. To enable and/or disable the aspirato r, refer to the previous section on PROGRAMMING THE SYSTEM FOR MANUAL OPERATION. Any one of availab le: 1) 2) 3) the followin g printer/c omputer Trendcom compati ble paralle l Centron ics compati ble paralle l RS~232C serial, 300 to 9600 Baud Consult TSC for your particu lar requirem ents. 8 interfac es are u { \ \,_,) PROGRAMMING THE SYSTEM FOR 'AUTO MODE' OPERATION In 'Auto Mode' operat ion, the aspira tor motor functio n is automa tic and is under comput er contro l. If the hardcop y output is enable d, (as previo usly describ ed), the comput er will turn on the aspira tor precis ely 60 second s prior to print time, and will turn the motor off immed iately after print. This feature extend s the service of the aspira tor motor which has a 300 hour maximum run-lif e. It also provid es for accurac y and repeat ability of measur ements , as the motor is run for equal amount s of time prior ta each and every printo ut. * To enable the Auto Mode, type the followi ng keystro kes: Now anytim e the printe r is enable d, (see PROGRAMMING THE SYSTEM FOR HARDCOPY OUTPUT), the probe will automa tically turn on and off at the correc t times. ( ) u * Disabl e the Auto Mode functio n with the followi ng k.ey stroke s: The Auto Mode is now disable d and contro l of the aspira tor is solely under user keyboa rd contro l. The printe r functio n is not affecte d. Note - Defaul t interv al is 1 minute betwee n printo uts. This means that in the Auto Mode, when the printe r functio n is enable d, the motor will stay on continu ously unless the time interv al is changed tsee CHANGING HARDCOPY TIME INTERVALS). Note - If the motor happen s to be on when either the Auto Mode o-r--th e printe r functio n is turned off from the keyboa rd by the user~ the aspira tor motor may also need to be turned off in the same manner . u u 9' i \.._ . _// \ CHANGING THE HARDCOPY TIM E INTERVAL \._ ) Th e 4A -l-M P is ca pa ble oth er de vic e, suc h as a of sen din .g da ta to a pr in ter or some Th e inf orm ati on se nt co nsco mp ute r, at sp ec ifi ed tim e in ter va ls. and De wp oin t tem pe rat ure ist s of Da te, Tim e, Dry Bu lb, Wet Bu lb, Va po r Pr ess ure ov er wa ter s, Re lat ive Hu mi dit y, and sa tu rat io n A he ad er is se nt be for at the Dry and Wet Bu lb tem pe rat ure s. e the fir st lin e of ta and ev ery 23 lin es th er ea fte r. The tim e be tw een eac h pr da int ou t de fau lts to the minimum in ter va l of 1 on , bu t is ke yb oa rd se lec mi nu te eac h tim e the sys tem is tur ne d tab le up to 24 ho urs . * To ch an ge the INTERVAL, exa mp le ch an ge s the in ter vafol low the sam ple be low . Th is can use yo ur own in ter va l l to 1 ho ur 25 mi nu tes (bu t you ins tea d) . ( \ J (_./ No te th at the num ber on when the SET key is pu she the low er dis pla y be gin s fla sh ing d. When the EN'l'ER key is pu she the num ber sto ps fla sh ing , ind ica tin g tha t the co mp d, ack no w,l edg es the key and ute r is now inp ut. Th e in ter va l nu mb ers wi aw ait ing the in ter va l ll ap pe ar on the up pe r dis pla y, sh ift in g the old nu mb ers ou t to the le ft as the new nu mb ers are pu nch ed in. (T his sam e pro ce du re oc cu rs for al l SET com ma nds ; ie, Pr es su re, Tim e, and Da te. ) 10 l l \._/ CHANGING THE PRESSURE The 4A-l-MP has been programmed from the factory with a barometric pressure of 29. 92 inches of mercury (in HG). The instrument uses the barometric pressure for all calculations of Relative Humidity and D.ewpoint temperature. rt. is recommended that the absolute pressure be measured in the environment where the probe is to be set up and used. The pressure should also be measured frequently during the operation of the system, as the accuracy of the calculations can differ as much as 2%RH for every inch of mercury. To change the PRESSURE, follow the sample example changes the pressure to 25.75 in HG. \, ) u below. This You must type in all 4 digits including the decimal point. For an explanation of the upper and lower display readings during this k~yboard sequence, refer to CHANGING THE HARDCOPY TIME INTERVAL. u \...._,,..i 11 u \_) CHANGING THE TIME/DATE The 4A-l-MP contains a battery backed-up Real Time Clock. The clock operates on a four year calendar, meaning that it makes no adjustmen t for leap year. Should it require setting for any reason, refer to the below procedure s. * To change the TIME, follow the sample below. bhanges the time to 3:40 PM. This eKample Note that the time is input in MILITARY FORMAT, and that no decimal point is needed between the hours and minutes. Far an explanatio n of the display readings during this keyboard sequence, refer to CHANGING THE HARDCOPY TIME INTERVAL. ( ) L; * To change the DATE, follow the sample below. change the date to June 4th. This would Note that June was input as '06' and that the day was input as '04'. Both digits of each must be typed. 12 u u SWITCHING TO °C/~F The 4A-l-MP is capable of both °C and °F operation. The user may at any time select which mode to use. When switched, all temperature displays, in6luding hardcopy output, will be affected. Th€ temperature data sent to the hardcopy device will be followed by C or F as applicable. * To switch to °C, type the following * To switch to 0 ~eystrokes: 'F 1 type these keystrokes: NOTE - The pow.er up default mode is °F. u 13 u u SUMMARY OF DISPLAY COMMANDS [8J8~B ( ) u % Relative Humidity ssrn6 Dewpoint Temperature B8GJ6 Dry Bulb Temperature [8]8~~ Wet Bulb Temperature s:s~.[i]· Dry Bulb Resistance in Ohms BBm~ Wet Bulb Resistance in Ohms 88~·~· Time B88B Da. te 14 u u SUMMARY OF ON/OFF COMMANDS BB~~ BB~.~ BB~~ ( u u ) ( 88, [[][i ] 88[ ;][i ] Enable Auto Mode Disable Auta Mo<le Turn Probe ON Manually Turn Probe OFF Manually Enable Printer Output / sram~ Disable Printer Output BB.[[]~ switch to srBm~ Switch to °F 1 u 15 0 c . \_. ) (_) SUMMARY OF SET COMMANDS BBGJBGJGJBGJ6 LGJBLQJBGJGJDGJGJm 8BGBGJGJBBliJ 8BGJBGJDGJGJ[i] ; \ ; v : ; ~ I, i u ; u'· . . 16 Interval Pressure 1 'ime Date u u PROBE WICK CHANGEOUT INSTRUCTIONS The CK-1 Wick Changeout Kit is provided to easily change wicks in the sensing probe. Wick changes should be performed after forty hours of operation, or more often, particularly if the instrument is in a dirty environment. A replacement wic~ supply should be ordered from Thunder Scientific Corporation in advance of the need and kept on hand. Order PW~lOl Probe Wick $10.00 each Suggested Stock: Minimum Order: F.O.B. ( I v u u 25 $50.00 Albuquerque, New Mexico 87123 lf Place probe upside down upon. two blocks of wood or material, one on each Side of the probe handle. 2) Remove reservoir plug and rubber feet. 3) Gently pry reservoir cover with plastic tool and remove. 4) Carefully remove the four housing retainer screws located in the corners of the probe; They join the two halves of the probe. 5) Gently remove bottom half of probe housing, feeding the wick through the teflon wick tube. Keep the teflon a.ir sleeve in place during removal. NOTE - Blower motor is also loose once housing is removed. 6,) Lift tef lon air sleeve up and off the sensor rnoun ts, taking care to feed the wick through the air sleeve avoiding any lateral movement and undue strain upon the sensor. 7) The sensor and wick are now in the open where the wick may be removed. using the dropper provided, apply several drops of ethyl alcohol to the portion of the wick covering the sensor. Allow to soak for approximately one minute, replenishing the alcohol as evaporation occurs. 8) Using one of the picks provided, covering the sensor to loosen. 9) Using the curved tweezers provided, very gently push the wick up on itself or up on the sensor, assisting with the other hand from the reservoir side. (The wick construction is very similar to a "Chinese finger puzzle" in that pushing tends to release it where pulling only other . l ..., gently probe the wick \_) u tightens the wick and compoun ds the problem of removal ). The wi.ck should expand. When this occurs, use the tweezers to slide the wick from the sensor. IMPORTANT! - If wick is frozen to the sensor with deposits due to contami nated water, DO NOT FORCE. This will break the sensor leads, causing expensiv e repair costs. If alcohol does not loosen the deposit , return to Thunder Scienti fic Corpora tion for removal. Thunder Scienti fic is not respons ible for lead break.age and/or sensor replacem ent if such occurs. If a wick changeo ut schedule is maintain ed, will not occur. (. ) this problem 10) Apply finger cots supplied to the thumb and index finger of each hand. soak a clean cloth with alcohol and remove all traces of talcum from the outer surfaces . If this is not done, the wick becomes contami nated. 11) The replacem ent wicks are supplied in a sealed alcohol. Remove wick from the vial. 12) Using the blue taper tool provided , push the wick onto the tool approxim ately one inch or more, causing expansio n by pushing on the wick. 13.) Remove from tool and clip frayed end square. 14) Slide wick shoulde r. 15.) Using the dropper , apply several drops of w.ater on the wick at the sensor. Note wick expands slightly , indicati ng water retentio n. Gently form the wick around end of sensor and flatten between the sensor and wick tube to offer minimum wind resistan ce. 16) Feed wick thru teflon air sleeve, slide sleeve over teflon sensor mounts, and seat it on the probe half. 17) Feed wick thru wick tube in bottom probe half and mate the probe halves. Coil wick inside reservo ir and replace cover. 18) Replace plug. u four vial gently onto sensor and over the teflon to corner screws, 18 rubber feet, and of the reservo ir FACTORY CERTIFICATION & RECERTIFICATION Factory certifica tion is initially ~rovided with each DBW instrumen t on a. no-charge basis. Certifica tion is good for one year or longer: However, the instrumen t should be recertifie d by the factory on a re9ular yearly basis. This is done for a calibratio n charge of $200.00 per temperatu re and humidity point. Generally , three temperatu re and humidity points are sufficien t. If the instrumen t requires more than genBral cleaning, a reconditio ning charge of $150.00 is required. NBS CERTIFICATION & RECERTIFICATION If it is desired that the instrumen t be certified by the National Bureau of Standards (NBS), all NBS charges are passed net to the customer, plus F.O.B. charges to and from, plus insurance coverage. ( ) In the case of NBS certifica tion, Thunder Scientific Corporatio n does require that the instrumen t be shipped direct from Thunder to NBS and returned prior to shipment to the customer. The same is required for recertific ation. The instrumen t should be returned to Thunder Scientific for inspection and reconditio ning prior to recertific ation by the Bureau. u ******** SUMMARY Your Thunder Scientific psychrome ter is a precise laboratory standard and should be treated according ly. Good maintenan ce will result in many years of service. Should any problems or question arise relative to your instrumen t, direct all correspond ence to the following address: Thunder scientific Corporatio n 623 Wyoming S.E. Albuquerq ue, New Mexico 81123 Or call collect (505) 265-8701 1Q i !. '-) PRINTER/COMPUTER\ INTERFACE OPTIONS There ar:-e three computer/p rinter interfac:e options cur:-r:-ently available for:- the 4A--1-MP Psychr;-orneter:-. Based upon which option' your;- unit has, the 25 pin output connector w.i ll match one of the below 1 isted pinouts. DB,...25 Pins 1 2 3 4 5 6 7 8 9 Trend~om Parallel Gr;-ound D1 03 D5 07 11 15· 16 17 18 19 20. 21 22 23 24 25 Strobe 01 02 03 D4 05 06 D7 XMIT Data * RTS * CTS ** DSR Ground Busy 1,2 13 RS-232 Sel'.'i al D0 10· 14 Centronics PClr;-allel Busy Strobe 00 D2 04 D6 Ground Ground Gi'i'ound• Ground Gi'i'ound Ground Gi'i'ound ** DTR No special cables shou ld be required. Each, of the output options al':'e configured so as to make the proper connections with the peripheral· device using onJ.y a flat ribbon cable and the appropdate insulation displacemen t connectors. ( ie DB-25, 36 pin· Centr:-onks, 20 pin Dual Row IDC, etc. ) 1 Interface cables ar:-e available frorn Thunder Scientific Corporation . Serial option notes: * ** DB-25 pins 4 and 5 ar:-e tied together inside the 4A"-1-MP~ DB-25 pins 6 and 20. are tied together inside the 4A-1-MP. Either of these 1 ines rnay be used: as a BUSY/READY handshake. lir1e. If handshaking ' is used, the receiving device shou:ld be able to accept one more fuill character in addition to the one ct..trrentlyb eingsent due to the da.uble-buff ered nature of the UART used within: the 4A'-1-MP. Signals g!'l'eater than +5 volts Signals greater than -5 volts ( u = READY = BUSY lf pins 6 and 20 are leflt ur;iterminate d, READY is assurned. - -- If (1) J co ' OJ an II "" I .€9. ( J ) ~· E = C\J • - 01 ' ,._ I VI ,\ - - c--, - ·--·-· RESERVOIR (REMOVABLE COVER FOR ACCESS) THREADED INSERT .5DO" DEEP 1/4-20 SCREW (4 PLCS.I RUBBER FEET (4 PLCS.I PRT ELEMENT (WET BULB) PRT ELEMENT (ORY BULB! KNURLED KNOB (REMOVABLE I .y· ., , ==· ."' . . :l-""""" FAN --.... >z I l!~!J FRONT VIEW AIRFLOW I I''. I >•'< ~ _,,. ~rl \._l. SIDE VIEW AIRFLOW BOTTOM VIEW THUNDER SCIENTIFIC CORP 61] WYOMING S.I. fou:R"..;[j ::~'.:2··· I _"."•v•••O"'•----- 1-1 .... I .. I AUUQUUQUI. NIW '41 •ICO 171'Jl ---- --psYCHROMETEll -- I MODEL 4A-IMP ... NA ............- --- DE83G023 11:!n-eon -11•·1111~11·11r· - 0z ,, § ~ ~ I• >r ::s 6~ I• [ !! c ·, \_J ; ~~ "'~ :1.L-~----i ,,:! ~~1 - ~ ( v ; . t 1 ~· st1 ', ,. I #~ r:lJ -_..• . -~ -- :t ~;f·"-"' "'~ t• 4 'J".11 : f ., :~~~~:.~ ~1~= ~.t.' :.?: ' "• ' ..... I ~ s; I ~ c . .. % 0 ID < er [ ...er O' 0 ;:: :i: w tj, ... Q ;:: . J ~ 5 " .. rs"' 1' <~ "' % .. I l:. er ' w c :z ~ .,i .J ~ .... 0 I! N' ·•> ~ .,r:oJ. =o! O! ~_..._0!--4 ~i··1r1JJ~J _,a·~ °'' '......... / ., , 0 "/ W·Ct,01 / - ~ ~ ( .i I I "' < 0 O' G : 8 Q ~ !: j ., 0 Q J~ ( ', \..__/ .., .., 0' ~1 r-EF-i .~~ ~~ ?~ ~1 I i l_,,; I I .tl .. ' 4~ -u" 1 I I ~ .$ '!~ .':2 " "~II ... ~Q ,i~rJI s~ JQ ,," u ~ if -- ! - - --ol .. - - "' --- - - ~'S .." ~~ ~~ . 1---9 ~~ ~"' 'tr~ .. s~ ~ f: -.:-::.:: .~ =--=------: :-_:-_-_-------- - - -- . :l !c g3 ' c c_= c_:_ ...----- .------1 REVISION.;;.S_ _ LTR 4 II 12 - 12 - - -- ~- 10 , __ 9 ------ 1----- 3 -- -- -- - - --- - -- - a 13 b 3 14 c 6 -- - 4 II 5 -- 13 l'I - -- 2 I ---- -- --- --- - -- IQ -- 7 • 28 ...,;._. -- --- - - - APPROVtD I II 3 8 13 12 16 15 f -- - 27 -- - -- ---- I ---- ----- --- --- 2 -- I 24 -- -- -- - - - - Q 26 4 19 5 -- d 25 UI 8 - -- 10 14 15 16 3 23 I 2 4 20 --DAf( LI 7 l - DESCRIPTION 8 7 16 21 - ll 17 - 622 - -- ., - I l 7 10 I 4 I II -8 3 13 12 16 15 L2 14 NOTES:: , 1. u1·1s AN -· LED DRIVER SYSTEM. .- 2. L 1 ANO L2 ARE - --- THUND ER SClEN TlFIC CORP LED DISPLA.vs: Ul WYOMING I.I, , 3. J1 IS A 14-PIN CONNECTOR, SHOlJU) COf\l\ECT· WITH J3 OF tv'0Tl-ER BOARD BY~ 4. Pro A .1 tv'F .CAfl. BETWEEN PIN 9 & 19 Of=U1, . . tOlc1U.-t1CCI- UHUll ----OTHUWll[ ll'lCIPllD fAACTIOHI DCC AHOLU :I: :I: :I: Al'PAOVALS - - ----- - - - DR.AW'D·A· V.Zi.5·63 --- -- -- -- NEW DISPLAY SCHEMATIC CtUCRlO - - 4A ... ·1-MP ---- DATE - f\ .-•\MO~ -C-lllA.~MiC1 IHC: II •tOllOI.• litO )0\11 AllUOUllO!JI , NIW MllllCO 11111 - SCAl[ -r\JA ----1 8 1DRAs'a3°A 130 5 oo Nor scAu 011Aw11m -- --- 1utm 1 or 1 C ~·, C. ,__ .--· (_ REVISIONS -Cl(SCR11'110H - LTR PROBE 1/2 CABLE 112 I I +26V RETUR I I 1 A' ~ I I I I I I I I +26V I Di ~ . I I ~ I I I I I I B' ~ - I 1 (B . I WET I BULB I I I 4-< I I o,_( I I I Bl ( r "I::: ' I I -(E'I - I I I I I I _I I I I I (61I . I : -! I I lI I I I <~ I I I I - - - I I I I I I I -- I I I I I I I I I I I I I I 1I A ~I . I I I I I ,.F 1 I ~ . I ( I I I I DRY BULB I I I I I I I I I I I A'I I F I< I . I I I I I Et, - ( I I I I <F1 . j I I I I I I II ( E!::.a1-.,...-=- - . I I I I I -THUND ER SCIENT IFIC CORP I I I 6ll WYOMING S,l, fMCTK>HI ;I: fn'\·11~-"°" ~11....,.,..•CI llritC. 4A+·MP ;I: SCEMEMATIC OAAW"t:>R.h CHtC•LD 'AllUOUUOUI, HM MIXICO 1711l MODEL Of..C ;I: AF'PAOVAlS "rlll ALO•OU JIO, 111>\JI - APP!IOvtO MAINFRAME I A <I . OATl- SCAlC- - - DO PROBE CONNECTOR IDRAWIHGSB3Bl31 -'8 NOT SCALE OR-AWING - - I SHCCT ... v i I - i i i . \_) I N WI ;i~ 1. ,,., i 1!1,,., ' . -~'~ llC~o I . . •••••••• •••••••• • u 8 .... .. _ . • • =O O" • • • g • •••••••• •••••••• • 0 + • • • .., ·D v I e :n ······· ········ ~· -ooz 2!'0 ,,i;·, 002•~ -· ~ ~-_,._e<>_,,i;·,~s~ i . . . I ~- ·I~ .·- .,._ " _ ~: --0-20 5· . • • .f;;§ • . • ~~ §El . !--G!l • , . ,• .'; ·· ····· ···· ···· lI ( 18 .) \ . .g.-. " i I: I'I I''111 -'. i 11 I .. 1~1~ . ,i: :,II . : ; 1111 . .:i• • I I I'"! .~ ;~ ;I i I Iiii . I I I ' I Q l ~ • • • • ·liill:l ...~ · - ;i ~ .•. ·····:· ~tR r·····:· ···:··· ... .o o· !~ •""R,.. fj Q o <•I!_-~ :; • • • • o, ~f I:> • u, - • . :: o J i!J ~ • ... : ... ········ lltt•L.!._.• Z.)b 8tt1•t-...!,_JI' • "" v u • e -u·voc e 41llYOC e ." N ::> • ,.., ......•• ........ • • ....•..• ........ .·o o. ~ei o.. ........ ....... Citi•:_..L-• lttf~;«E ........ ........ • :- • 0 • • •• •• • ••• • • • • • •• • e •l'VDC e IWITCKED 14V lltE:TUfl.N a a .. ,. a~ 2 0 u ~l - ·. ~~:::~:::::1. . . Q •• UY,COLMOW IV COl.IMION t9 :~:::·:::i. . a~~/:~1G UY·llETUlllN YDC