Digital Retinal Camera
CX-1
Operation Manual
Before using the instrument, be sure to read this manual thoroughly.
Keep the manual where it is easily accessible.
PLEASE NOTE
1. Please contact a Canon representative or distributor to have the instruments installed.
2. Use the computer, monitor, and other equipment that conforms to the system standard IEC 60601-1/
UL 60601-1 or IEC 60950-1/UL 60950-1 for the Digital Retinal Camera CX-1. Be sure that the entire system
conform to IEC 60601-1-1/UL 60601-1-1. Be sure to also use an isolation transformer conforming to
IEC 60601-1/UL 60601-1 when a computer or monitor conforming to IEC 60950-1/UL 60950-1 is used. For
details, consult a Canon representative or distributor.
3. To maintain compliance to IEC 60601-1-2, it is recommended that the computer, monitor, or other
equipment configured with the CX-1 be evaluated to IEC 60601-1-2 or FCC Part15 Subpart B Class B and
CISPR 24.
4. The user is responsible for the use and maintenance of the product. We suggest that a member of the user’s
staff be designated as being in charge of maintenance so as to ensure that the product is kept in a safe and
good condition. Also, medical products must be used only by a qualified person.
5. This product may malfunction due to electromagnetic waves caused by cellular phones, transceivers, radiocontrolled toys, etc. Be sure to avoid having objects such as these, which affect this product, brought near
the product.
6. Disposal of this product in an unlawful manner may have a negative impact on health and on the
environment. When disposing of this product, therefore, be absolutely sure to follow the procedure which is
in conformity with the laws and regulations applicable in your area.
7. In no event will Canon be liable for direct or indirect consequential damage arising out of the use of this
product. Canon will not be liable for loss of image data due to any reason.
8. Reading of image and storage of data must be performed in accordance with the law of the country where
the product is being used. Also, the user is responsible for maintaining the privacy of image data.
9. The power cable supplied is designed to be used solely with this camera. Do not use it for any other
products.
10.Canon reserves the right to change the specifications, configuration and appearance of the product without
prior notice.
European Union (and EEA*) only.
This symbol indicates that this product is not to be disposed of with your household waste,
according to the WEEE Directive (2002/96/EC) and your national law. This product should
be handed over to a designated collection point, e.g., on an authorized one-for-one basis
when you buy a new similar product or to an authorized collection site for recycling waste
electrical and electronic equipment (EEE). Improper handling of this type of waste could
have a possible negative impact on the environment and human health due to potentially
hazardous substances that are generally associated with EEE. At the same time, your
cooperation in the correct disposal of this product will contribute to the effective usage of
natural resources. For more information about where you can drop off your waste
equipment for recycling, please contact your local city office, waste authority, approved
WEEE scheme or your household waste disposal service.
For more information regarding return and recycling of WEEE products, please visit
www.canon-europe.com/environment.
* EEA : Norway, Iceland and Liechtenstein
For California, USA Only
Included battery contains Perchlorate Material—Special handling may apply.
See http://www.dtsc.ca.gov/hazardouswaste/perchlorate/ for details.
© CANON INC. 2009
All rights reserved.
Under copyright laws, this manual may not be copied, in whole or in part, without the written consent
of Canon.
Safety Information
Regulations
This instrument conforms to IEC 60601-1-2:2001.
For U.S.A. and Canada
• When the CX-1 is going to be operated at a voltage of 240V in USA or Canada, be sure to connect the instrument to a center tapped voltage source.
• This instrument is a CLASS I EQUIPMENT according to UL 60601-1 and CAN/CSA C22.2
No. 601.1.
MEDICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 60601-1 and CAN / CSA C22.2 No. 601.1
41C4
• Grounding reliability can only be achieved when the equipment is connected to an equivalent
receptacle marked “Hospital Only” or “Hospital Grade”.
• The following mark indicates that the instrument is a Type B Applied Parts.
• The degree of protection against ingress of water is IPX0.
• This equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air
or with oxygen or nitrous oxide.
• The mode of operation is Continuous Operation.
For EU Countries
• The following mark shows compliance of the instrument with Directive 93/42/EEC.
• This instrument is a CLASS I EQUIPMENT according to EN 60601-1.
• The following mark indicates that the instrument is a Type B Applied Parts according to EN
60601-1.
• The degree of protection against ingress of water is IPX0.
• This equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air
or with oxygen or nitrous oxide.
• The mode of operation is Continuous Operation.
Für Deutschland
• Während des Betriebs liegt der Schalldruckpegel dieses Instruments bei 70 dB(A) oder weniger
gemäß ISO 7779.
(1)
Safety Information
General Safety Information
Follow the safety instructions in this manual and all warnings and cautions printed on the warning labels.
Ignoring such cautions or warnings while handling the product may result in injury or accident. Be sure to
read and fully understand the manual before using this product.
Keep this manual for future reference.
Meaning of Caution Signs
To protect the safety of users and others and to prevent accidents, this operation manual uses the symbols and
text shown below in warnings and cautions. Read the meanings of these caution signs and the Safety
Precautions (see page (3)), and follow the safety instructions.
! WARNING
This indicates a potentially hazardous situation which, if not heeded, could result in
death or serious injury to you or others.
! CAUTION
This indicates hazardous situation which, if not heeded, may result in minor or
moderate injury to you or others, or may result in machine damage.
NOTE
This is used to emphasize essential information.
Be sure to read this information to avoid incorrect operation.
(2)
Safety Information
Safety Precautions
Be sure to follow the safety instructions below to ensure correct operation of the instrument.
Installation and Environment of Use
! WARNING
Do not use or store the instrument near any flammable chemicals such as alcohol,
thinner, benzine, etc. If chemicals are spilled or evaporate, it may result in fire or
electric shock through contact with electric parts inside the instruments. Also, some
disinfectants are flammable.
Be sure to take care when using them.
! CAUTION
Do not use or store the instrument in a location with the conditions listed below.
Otherwise, it may result in failure or malfunction, fall or cause fire or injury.
• Close to facilities where water is used.
• Where it will be exposed to direct sunlight.
• Close to air-conditioner or ventilation equipment.
• Close to heat source such as a heater.
• Prone to vibration.
• Insecure place.
• Dusty environment.
• Saline or sulfurous environment.
• High temperature or humidity.
• Freezing or condensation.
! CAUTION
Do not cover the vent holes on the cover.
Otherwise, the temperature in the instrument may rise and cause fire.
! CAUTION
Place the instrument on a firm table.
Do not place it extremely near the edge of the table in order to avoid damage or
injury due to falling.
Installation Operation
! WARNING
Do not connect the instrument with anything other than specified.
Otherwise, it may result in fire or electric shock.
Also, when other equipment is going to be connected to the instrument using the
connector for interface, be sure to check after the connection that leakage current is
within the tolerable value. For details, please contact a Canon representative or
distributor.
Power Supply
! WARNING
Only operate the instrument with the type of power supply indicated on the rating
label. Otherwise, it may result in fire or electric shock.
! WARNING
Be sure to turn OFF the power before plugging or unplugging the cables as indicated
in this manual. Also, do not handle it with wet hands. Otherwise, you may get an
electric shock that may result in death or serious injury.
! WARNING
Securely plug in the power cable into the AC outlet.
If contact failure occurs, or if dust or metal object comes in contact with the exposed
metal prong of the plug, fire or electric shock may result.
(3)
Safety Information
Power Supply
! WARNING
Be sure to hold the plug to disconnect the power cable.
If you pull the cable, the core wire may be damaged, resulting in fire or electric
shock.
! WARNING
Do not cut or process the cables. Also, do not place anything heavy, including the
instrument on it, step on it, pull it, bend it, or bundle it.
Otherwise, the cable may be damaged, which may result in fire or electric shock.
! WARNING
Do not get the power for more than one instrument from the same AC outlet.
Otherwise, it may result in fire or electric shock.
! CAUTION
Before connecting or disconnecting the cables, be sure to hold the instrument firmly
in order to ensure safety.
Otherwise, the main unit may fall over, possibly causing injury.
! CAUTION
The instrument is shipped with a grounding type (three-core) power cable.
To reduce the risk of electric shock, always plug the cable into a grounded power
outlet.
! CAUTION
To make it easy to disconnect the plug at any time, avoid putting any obstacles near
the outlet.
Handling
! WARNING
Never disassemble or modify the product as it may result in fire or electric shock.
Also, since the instrument incorporates high-voltage parts that may cause electric
shocks and other hazardous parts, touching them may cause death or serious injury.
! WARNING
Do not place anything on top of the instrument.
Otherwise, the object may fall and cause injury.
Also, if metal objects such as needle or clip falls into the instrument, or if liquid is
spilled, it may result in fire or electric shock.
! WARNING
When the instrument is going to be moved, be sure to turn OFF the power, unplug
the power cable from the AC outlet, and disconnect other cables.
Otherwise, the cable may be damaged, which may result in fire or electric shock.
! WARNING
When the instrument is going to be carried, be sure to tighten the stage unit lock,
hold the carrying indentations at the left and right of the bottom panel and hold the
instrument horizontally.
Do not hold it by the digital camera or the face rest poles or other parts, as they may
come off and result in injury.
! WARNING
Do not hit or drop the instrument. The instrument may be damaged if it receives a
strong jolt, which may result in fire or electric shock if the instrument is used without
it being repaired.
! CAUTION
Wipe the forehead rest with ethanol or glutaraldehyde solution to disinfect them
each time a different patient uses the instrument in order to prevent infection.
Please consult a specialist for the procedure for disinfection.
The forehead rest can be damaged if using a disinfectant other than those above or
if another disinfectant is added to ethanol. Please consult a specialist in this case as
well.
(4)
Safety Information
Handling
! CAUTION
To ensure cleanliness, replace the chin rest paper whenever changing patients.
! CAUTION
When adjusting the forward position of the main unit, be sure to move the main unit
slowly toward the patient while looking from the side of the patient, so as to prevent
accidental contact of the objective lens with the patient.
! CAUTION
Do not place your hand or fingers between the stage and base. Also, ensure that the
patient does not place his/her hand or fingers there either. Otherwise, fingers may be
hurt.
! CAUTION
When the instrument is not going to be used, turn OFF the power.
Also, unplug the power cable from the AC outlet when it is not going to be used for a
long time.
When Problem Occurs
! WARNING
Should any of the following occur, immediately turn OFF the power of each
instrument, unplug the power cable from the AC outlet, and contact a Canon
representative or distributor.
• When there is smoke, an odd smell or abnormal sound.
• When liquid has been spilled into the instrument or a metal object has entered
through an opening.
• When the instrument has been dropped and it is damaged.
Maintenance and Inspection
! WARNING
For safety reasons, be sure to turn OFF the power when the inspections indicated in
this manual are going to be performed. Otherwise, it may result in electric shock.
! WARNING
When the instrument is going to be cleaned, be sure to turn OFF the power, and
unplug the power cable from the AC outlet.
Never use alcohol, benzine, thinner or any other flammable cleaning agents.
Otherwise, fire or electric shock may result.
! WARNING
Clean the plug of the power cable periodically by unplugging it from the AC outlet
and removing dust or dirt from the plug, its periphery and AC outlet with a dry cloth.
If the cable is kept plugged in for a long time in a dusty, humid or sooty place, dust
around the plug will attract moisture, and this could cause insulation failure which
could result in a fire.
! WARNING
The instrument must be repaired by a qualified engineer only.
If it is not repaired properly, it may cause fire, electric shock, or accident.
! CAUTION
For safety reasons, be sure to inspect the instrument before using it.
(5)
Safety Information
System Use
! WARNING
Do not place multiple portable socket-outlets on the floor. Otherwise, it may result in
fire or electric shock.
! WARNING
Do not connect an additional multiple portable socket-outlet or extension cord to the
system. Otherwise, it may result in fire or electric shock.
! WARNING
Do not connect instruments which are not specified as part of the system.
Otherwise, it may result in fire or electric shock.
! WARNING
For the system, use a power transformer of 1000 VA of the maximum permitted load.
Otherwise, it may result in fire or electric shock.
! WARNING
Use multiple portable socket-outlets only for supplying power to equipment which is
intended to form part of the system. Otherwise, it may result in fire or electric shock.
! WARNING
Do not connect non-medical electrical equipment which has been supplied as a part
of the system to any power socket other than the multiple socket-outlets for the
system. Otherwise, it may result in electric shock.
! WARNING
Do not connect any electrical equipment which has not been supplied as a part of
the system to the multiple portable socket-outlets for the system. Otherwise, it may
result in fire or electric shock.
! WARNING
Do not simultaneously touch a patient and non-medical electrical equipment with
which you may get contact. Otherwise, it may result in electric shock.
! CAUTION
For cleaning equipment forming part of the system, follow the instruction manual for
the equipment. Otherwise, it may cause failure, an accident, or fire.
! CAUTION
Install the system in a way that enables the user to achieve optimal use.
(6)
Safety Information
Label on the Instrument
The CX-1 has a label on it.
The label contents and its position are indicated below.
(7)
Components
Digital Retinal Camera CX-1............................................. 1
Digital camera .....................................1
Camera cover...................................... 1
Power cable
(3 m, Non-shielded) ............................1
* Plug shapes differ by power
specifications.
External eye fixation lamp.................. 1
Chin rest paper ................................100
Video cable ......................................... 1
Camera mount cap...............................1
Objective lens cap............................... 1
Dust cover ...........................................1
Body cap ............................................. 1
CD-ROM (Retinal imaging control
software)..............................................1
Operation Manual (CX-1/control
software) ..................................... 1 each
Optional Units
Internal eye fixation target
Stereo unit
Chin rest paper (500 sheets)
(8)
Contents
Safety Information ..............................................................................................................(1)
Components .......................................................................................................................(8)
1. Features .......................................................................................................................... 1
2. Notes for Using the Instrument........................................................................................ 2
3. Names of Parts................................................................................................................ 4
3.1 Main Unit ................................................................................................................................4
3.2 Operation Panel .....................................................................................................................6
3.3 Observation Monitor and Viewfinder ......................................................................................7
4. Preparation ...................................................................................................................... 8
4.1 Power Cable Connections ......................................................................................................8
4.2 Connecting to a Computer .....................................................................................................9
4.3 Carrying the Instrument ........................................................................................................10
5. Basic Photography ........................................................................................................ 11
5.1 Before Taking Photographs ..................................................................................................11
5.1.1 Turn On the Power ...................................................................................................................... 11
5.1.2 Move the Stage Unit to Get Started ............................................................................................ 12
5.1.3 Adjust the Viewfinder .................................................................................................................. 12
5.2 Photography Flow ................................................................................................................13
5.3 Non-Mydriatic Mode .............................................................................................................14
5.3.1 Color Photography Mode ............................................................................................................ 14
5.3.2 Red-free Photography Mode ....................................................................................................... 20
5.3.3 Cobalt Photography Mode .......................................................................................................... 21
5.3.4 Fluorescein Angiography Mode .................................................................................................. 21
5.3.5 Fundus Autofluorescence Mode ................................................................................................. 23
5.4 Mydriatic Mode .....................................................................................................................24
5.4.1 Color Photography Mode ............................................................................................................ 24
5.4.2 Red-free Photography Mode ....................................................................................................... 30
5.4.3 Cobalt Photography Mode .......................................................................................................... 30
5.4.4 Fluorescein Angiography Mode .................................................................................................. 30
5.4.5 Fundus Autofluorescence Mode ................................................................................................. 32
6. Photography Auxiliary Functions ................................................................................... 33
6.1 Using the Focus Indicator .....................................................................................................33
6.2 ×2 photography function .......................................................................................................33
6.3 Small pupil photography function .........................................................................................34
6.4 Diopter Compensation Function ...........................................................................................34
6.5 Panning Function and Tilting Function .................................................................................35
6.5.1 Panning Function ........................................................................................................................ 35
6.5.2 Tilting Function ............................................................................................................................ 35
6.6 Using the External Monitor ...................................................................................................36
(9)
Contents
7. Daily Inspection and Maintenance................................................................................. 38
7.1 Inspection .............................................................................................................................38
7.1.1 Before Turning ON the Power ..................................................................................................... 39
7.1.2 Turning ON the Power ................................................................................................................ 40
7.2 Before Calling a Service Technician ....................................................................................41
7.2.1 Troubleshooting .......................................................................................................................... 41
7.3 Cleaning and Disinfection .....................................................................................................44
7.3.1 Objective Lens ............................................................................................................................ 44
7.3.2 Viewfinder ................................................................................................................................... 45
7.3.3 Forehead Rest ............................................................................................................................ 45
7.3.4 Retinal Camera and Digital Camera Outer Covers ..................................................................... 46
7.4 Refilling the Chin Rest Paper ...............................................................................................47
7.5 Removing and Installing the Digital Camera ........................................................................48
7.5.1 Removing the Digital Camera ..................................................................................................... 48
7.5.2 Installing the Digital Camera ....................................................................................................... 50
7.5.3 Cleaning the Digital Camera (imaging sensor) ........................................................................... 52
8. Main Specifications and Performance ........................................................................... 53
8.1 Main Specifications ..............................................................................................................53
8.2 Functions in the Observation and Photography Modes .......................................................54
8.3 Guidance and Manufacturer’s Declaration for EMC Directive ..............................................55
9. Service Information........................................................................................................ 59
(10)
1. Features
The Digital Retinal Camera CX-1 is used to observe the retinas of patients’ eyes and take photographs of
them. The camera has the following features.
Both mydriatic and non-mydriatic modes in a single unit
The CX-1 includes a mydriatic mode and non-mydriatic mode, and each mode can be selected with a
single switch.
In the mydriatic mode, the optical viewfinder is used for observation. In the non-mydriatic mode, the
observation monitor on the special digital camera is used for observation.
Fundus autofluorescence photography
Fundus autofluorescence photography is possible in both mydriatic and non-mydriatic modes.
Special digital camera
The CX-1 includes a special digital camera using EOS digital technology as a standard accessory.
This enables the CX-1 to fully maximize Canon technology for shooting high-quality, high-resolution
diagnostic images.
Compact, user-friendly design
Panning and tilting functions are provided as standard options in a compact, lightweight design.
The compact body makes it easier to assist patients and provides significantly improved ease-of-operation.
Vertical movement of the main body by a motorized chin rest and an operation lever significantly improves
alignment operability.
Small pupil photography function
The CX-1 can photograph dilated pupils with a diameter of at least 5.1 mm in the mydriatic mode, with
50° of their field angle. It can photograph non-dilated pupils with the diameter of at least 4.3 mm in the
non-mydriatic mode, with 45° of their field angle.
The CX-1 provides the small pupil photography function for patients whose eyes are difficult to dilate.
Using the function, it can take photographs dilated pupils with a diameter of at least 4.3 mm and nondilated pupils with a diameter of at least 3.8 mm.
System optimized for digital photography
With the Retinal imaging control software (called “control software” below), images photographed by the
CX-1 can be browsed, processed, stored, output to the DICOM storage server, exported to DICOM files,
and output to a printer.
The settings for the lighting system filter, flash intensity, and digital camera are automatically optimized
for each photography mode – color, red-free, cobalt, fluorescein angiography, or fundus autofluorescence.
-1-
2. Notes for Using the Instrument
Digital camera
•
The digital camera installed on the Digital Retinal Camera CX-1 is designed for the CX-1.
•
The specifications differ from the typical commercially-available Canon digital cameras. It cannot be
used for purposes other than observing and taking photographs of peoples’ eyes. Do not use the digital
camera removing from the retinal camera body. This can damage or harm the performance of the
instrument.
•
If any problems with the digital camera occur or repairs are needed, please contact a Canon
representative or distributor.
Before use
•
Check and clean the objective lens before taking an image, as any stains or scratches on it will appear
as white spots.
•
Sudden heating of the room in cold areas will cause condensation to form on the objective lens or on
optical parts inside the instrument. In this case, wait until condensation disappears before performing
photography.
•
Do not touch the lens part of the retinal camera and the mirror part of the digital camera when attaching
and detaching the digital camera from the retinal camera body. If any dirt, fingerprints, dust, and other
foreign objects attaches to the lens or mirror part, you cannot take a good image.
•
Be sure to adjust the diopter of the viewfinder so it matches the diopter of photographer’s eye.
Otherwise, focusing will not be able to be performed correctly.
After use
•
Turn OFF the power of the instrument. Place the cap over the objective lens, and place the dust cover
•
If the digital camera is removed and left unattached, dust and other foreign objects can enter the retinal
over the instrument. You cannot take a good picture when the objective lens is dirty.
camera and the digital camera. Be sure to always attach the caps to the respective mounts.
Cleaning and disinfection
•
Do not let the tip of a blower touch the objective lens.
•
Do not wipe off or rub the objective lens when there is dust or other substances on it.
This could scratch the lens surface.
•
Never wipe the objective lens with disinfecting ethanol, eyeglass lens cleaner, or cleaning paper
containing silicon. The lens surface could be damaged or the surface may not be completely wiped off.
•
Do not clean the cover of the instrument with lens cleaner. The cover of the instrument could be
damaged.
•
Never use alcohol, benzine, thinner, or other solvents to clean the cover of the instrument. This could
damage the cover of the instrument.
-2-
2. Notes for Using the Instrument
•
Never use disinfecting ethanol, glutaraldehyde or other solvents to clean the cover of the instrument,
except the forehead rest and the chin rest. This could damage the cover of the instrument.
•
If the chin rest paper will not be used, be sure to disinfect the chin rest in the same way as the forehead
rest whenever the patient changes.
Environment of use
•
Use the camera in an environment that satisfies the following requirements.
Use
Temperature:
10°C to 35°C
Humidity:
30% to 80% RH (no condensation)
Atmospheric pressure:
800 hPa to 1060 hPa
Storage and Transportation
•
Temperature:
–30°C to 60°C
Humidity:
10% to 60% RH (no condensation)
Atmospheric pressure:
700 hPa to 1060 hPa
Do not install or store the camera or leave it standing in an adverse environment such as where the
temperature and humidity levels are high. Doing so may cause trouble and/or malfunction. Be sure to
observe the points of Installation and Environment of Use (see page (3)) when selecting the
installation location.
•
Dust in the air will not only attach to the objective lens, but will also attach onto the optical parts inside
the instrument. You cannot take a good image when dust is on them. Please keep the room clean.
Installation
•
Please ask a Canon representative or distributor to perform installation of this product.
•
Please handle this product carefully. The adjustment can be altered if it is subjected to a strong shock or
jolt.
•
If this product will be transported in an automobile or shipped a long distance, protective measures need
to be taken for vibrations and shocks. Ask a Canon representative or distributor for more information.
Others
For U.S.A.
Rx only-Caution:
Federal law restricts this device to sale by or on the order of a licensed practitioner.
Intended Use:
The device is intended to be used for taking digital images of retina of human eye.
For European Union
Intended Use:
This medical device is intended to observe image and record retinal fundus through the pupil without
contact with subject’s eye for the purpose of diagnosis by way of producing fundus image information.
-3-
3. Names of Parts
3.1 Main Unit
1
2
3
7
4
5
8
6
9
10
11
1
External eye fixation lamp (see page 17,
7
Diopter compensation lever
(see page 34)
27)
2
Face rest
8
Focus knob (see page 17, 27)
3
Forehead rest (see page 45)
9
Pan Tilt grip (see page 35)
4
Objective lens (see page 44)
10 Finger guard
5
Objective lens cap
11 Carrying indentations (see page 10)
6
Chin rest (see page 15)
-4-
3. Names of Parts
12
20
13
21
14
15
16
22
17
18
23
24
25
19
26
27
12 Viewfinder (see page 7)
20 Height adjustment mark (see page 15)
13 Digital camera (see page (8))
21 Tilting lock lever (see page 35)
14 Digital camera cover (see page 48)
22 Panning lock knob (see page 35)
15 Observation monitor (see page 7)
23 Operation panel (see page 6)
16 Digital camera power switch
24 Stage unit lock button (see page 19)
25 USB connector (see page 9)
(see page 11)
17 Shutter release button (see page 18, 21,
26 AC power connector (see page 8)
27 Retinal camera power switch
22, 28)
18 Operation lever (see page 12, 15, 16)
19 Stage unit (see page 12, 15, 19, 29)
-5-
3. Names of Parts
3.2 Operation Panel
9
10
1
2
3
4
5
11
12
13
6
14
7
8
15
11 Photography mode switches/lamps
1
Brightness adjuster (see page 16)
2
Focus indicator switch/lamp
From left to right;
(see page 33)
COLOR (see page 14, 24)
3
×2 switch/lamp (see page 33)
RED FREE (see page 20, 30)
4
Small pupil photography switch/lamp
COBALT (see page 21, 30)
(see page 34)
FLUO (see page 21, 30)
5
BA switch/lamp (see page 30)
FAF (see page 23, 32)
6
Flash intensity indicator (see page 19)
7
Flash intensity switches (see page 19)
8
TIMER/C switch/lamp (see page 22)
13 Blink switch (see page 17)
9
Observation mode lamp
14 POWER lamp (see page 11)
Myd (see page 24)
15 CHIN REST switch (see page 15)
Non-Myd (see page 14)
10 Observation mode switch
(see page 14, 24)
-6-
12 FIX TARGET (internal eye fixation)
switches (see page 17)
3. Names of Parts
3.3 Observation Monitor and Viewfinder
Observation monitor (with x1.3 magnification)
1
R
COLOR
5
6
2
7
8
3
2
9
4
1
2
3
4
Photography mode indication
COLOR:
Color (see page 14)
RED FREE: Red free (see page 20)
COBALT:
Cobalt (see page 21)
FLUO:
Fluorescein angiography
(see page 21)
FAF:
Fundus autofluorescence
(see page 23)
Focus indicator (see page 17)
×2 photography indicator (see page 33)
Small pupil photography indicator
(see page 34)
5
6
7
8
9
Right/Left eye indicator
R: Right eye
L: Left eye
Small pupil photography effective area
(see page 34)
Internal eye fixation lamp position
(see page 17)
×2 photography range (see page 33)
Photography ready lamp (see page 18)
Viewfinder
1
2
3
2
4
1
×2 photography range (see page 33)
3
Photography ready lamp (see page 18)
2
Focus indicator (see page 27)
4
×2 photography range (see page 33)
-7-
4. Preparation
Please ask a Canon representative or distributor to perform installation of this instrument.
4.1 Power Cable Connections
WARNING
Be sure to operate this instrument using the power supply described in the
specifications. Otherwise, it may result in fire or electric shock.
WARNING
Be sure to turn OFF the power before plugging or unplugging the cables as
indicated in this manual. Also, do not handle it with wet hands. Otherwise,
you may get an electric shock that may result in death or serious injury.
CAUTION
Before connecting or disconnecting the cables, be sure to hold the
instrument firmly in order to ensure safety.
Otherwise, the main unit may fall over, possibly causing injury.
CAUTION
Always use a grounded AC outlet for a 3-pin plug as the power outlet.
1
Check that the power switch of the retinal camera is set to the O position (OFF).
2
Connect the power cable to the retinal camera.
Insert the power cable connector firmly into the AC
power connector while holding down the retinal
camera with your hand.
ON
OFF
3
Insert the power plug into the AC outlet.
Insert the plug all the way into the AC outlet.
-8-
4. Preparation
4.2 Connecting to a Computer
Connect the retinal camera to a computer where the supplied software has been installed.
1
Check that the power switches for the retinal camera and digital camera are set to the OFF
2
Connect the USB cables to the main unit.
position.
Insert the USB cable connectors firmly into the USB
connectors while holding down the main unit with your
hand.
3
Connect the USB connectors at the other end to the computer.
Connect the connectors at the other end of the USB cables firmly into USB connectors of the
computer.
Note: Use the following type of USB cables. Communication may not be possible if any other
kind of cable is used.
• Cable with type AB connector plug supporting USB 2.0 Hi-Speed, maximum length of
3 meters
For further details, contact a Canon representative or the distributor from which you
purchased the unit.
Note: When connecting the USB cable, pass it underneath the cables of the camera main unit.
If it is passed over the main unit cables, the USB cable may become disconnected as the
main unit moves for taking images.
-9-
4. Preparation
4.3 Carrying the Instrument
This is the procedure when carrying the instrument by hand.
1
WARNING
When the instrument is going to be moved, be sure to turn OFF the power,
unplug the power cable from the AC outlet, and disconnect other cables.
Otherwise, the cable may be damaged, which may result in fire or electric
shock.
Also, when the instrument is going to be carried, be sure to tighten the
stage unit lock, the panning lock knob and the tilting lock lever, hold the
indentations at the left and right of the bottom panel, and hold the
instrument horizontally.
Do not hold it by the digital camera or the face rest poles or other parts, as
they may come off and result in injury.
WARNING
Do not hit or drop the instrument. The instrument may be damaged if it
receives a strong jolt, which may result in fire or electric shock if the
instrument is used without it being repaired.
Turn off the power of the retinal camera.
Press the power switch of the retinal camera to the O side (OFF).
2
Remove the connected cables.
Remove all power cables and USB cables.
3
Secure the parts in place.
Tighten the panning lock knob, tilting lock lever and the stage unit lock.
4
Carry the retinal camera main unit.
Hold the retinal camera main unit by holding the carrying indentations at the left and right of the
bottom panel, and carrying the unit kept horizontally.
To remove the digital camera before carrying the retinal camera
See section 7.5.1 Removing the Digital Camera (page 48).
Note: If this product will be transported in an automobile or shipped a long distance, protective
measures need to be taken for vibrations and shocks. Ask a Canon representative or
distributor for more information.
-10-
5. Basic Photography
5.1 Before Taking Photographs
5.1.1 Turn On the Power
1
Remove the dust cover and the objective lens cap.
2
Turn on the computer.
Turn on the computer, and start the control software.
Concerning control software operations
For details on the control software operations, refer to the operation manual of the control
software.
3
Turn on the power switch of the digital camera.
Set the power switch of the digital camera to
or
.
4
Turn on the power of the retinal camera.
Press the power switch of the retinal camera to the I
side (ON).
ON
OFF
Concerning the power-saving function
The power-saving function is activated under the following circumstances.
• When none of the switches on the operation panel has been operated for 10 or more minutes
while power is supplied to the camera
• When none of the switches on the operation panel has been operated for 1 or more minutes
after the operation to stop the timer has been performed during fluorescein angiography
photography
When the power-saving function has been activated, the POWER lamp on the operation panel
blinks.
To restore normal operation, press any switch on the operation panel, move the stage to the left or
right or operate the diopter compensation lever.
-11-
5. Basic Photography
5.1.2 Move the Stage Unit to Get Started
Press the stage unit lock button to release the lock.
Grasp the operation lever, and pull the stage all the way to the
right front, and then press the stage lock button to lock it in
place.
5.1.3 Adjust the Viewfinder
Note: Be sure to adjust the diopter of the viewfinder so it matches the diopter of photographer’s
eye before taking photographs in the mydriatic mode. Otherwise, focusing will not be able
to be performed correctly.
1
Turn the viewfinder diopter compensation ring
counterclockwise (to the positive end) until it
stops.
2
Look into the viewfinder with one eye.
3
Gradually turn the diopter compensation ring
clockwise (toward negative direction) until the
crosshairs in the field of view appear sharpest.
-12-
5. Basic Photography
5.2 Photography Flow
This instrument includes non-mydriatic and mydriatic observation modes. In each observation mode,
photographs can be taken using color, red-free, cobalt, fluorescein angiography, or fundus
autofluorescence. The observation monitor is used in non-mydriatic mode, and the viewfinder is used in
mydriatic mode. The flow of photography operations is shown in the photography flowchart below.
For a list of the functions in the observation modes and photography modes, refer to 8.2 Functions in the
Observation and Photography Modes (page 54).
Enter study information
Select observation mode
Non-mydriatic mode
Mydriatic mode
Select photography mode
Adjust viewfinder diopter
Select photography mode
Set flash intensity
Set flash intensity
Check pupil diameter
Check mydriatic condition
Set observation light intensity
Set observation light intensity
Adjust int./ext. eye fixation lamp position
Adjust external eye fixation lamp position
Align positions and focus image
Align positions and focus image
Select diopter compensation
Select diopter compensation
Press shutter release button
Press shutter release button
Check images
Edit and save images
Finish photographing
Concerning control software operations
Use the control software for operations in
in the chart. For details on the control software
operations, refer to the Operation Manual of the control software.
-13-
5. Basic Photography
5.3 Non-Mydriatic Mode
The Non-Mydriatic mode is designed for photographing the
patient eye without using a mydriatic.
To switch from the mydriatic mode to the non-mydriatic
mode, hold down the observation mode switch for approx.
2 or more seconds.
The observation mode lamp Non-Myd turns on.
The observation monitor is used for retinal observation.
Note: Overexposure may result when photographs of patient eyes are taken using a mydriatic in
the non-mydriatic mode.
5.3.1 Color Photography Mode
1
Enter the study information into the computer.
Enter the study information with the control software.
2
Select the color photography mode.
Press the COLOR switch.
Or select the photography mode with the control
software.
COLOR lamp turns on.
COLOR is displayed on the observation monitor.
3
Disinfect the forehead rest and replace the chin rest paper.
! CAUTION
To prevent the risk of infection, wipe the forehead rest with a disinfectant
such as ethanol or glutaraldehyde whenever the patient is changed.
For details on the disinfection procedure, consult a specialist.
To ensure cleanliness, replace the chin rest paper whenever the patient is
! CAUTION changed.
If chin rest paper is not used, disinfect the chin rest.
4
Have the patient sit down.
If the patient is wearing glasses or contact lenses, instruct the patient to remove them.
Have the patient place his or her chin against the chin rest and his or her forehead against the
forehead rest.
-14-
5. Basic Photography
5
Align the height of the patient’s eye.
Press the CHIN REST switches to adjust the chin rest up or down so that the patient’s eye is aligned
with the height adjustment mark.
Height
adjustment
mark
6
Move the stage position.
! CAUTION
Do not put your fingers or hands between the sliding part and base of the
stage unit or on the panning and tilting sliding parts.
Similarly, do not allow patients to put their hands in these places.
Otherwise, your or their hands or fingers may be pinched, possibly
resulting in injury.
If the stage is locked at the operator’s side, press the
stage lock button to unlock the stage.
Hold the operation lever, pull it towards you, and then
move the stage toward the eye to be photographed.
7
Align the retinal camera to the eye to be photographed.
! CAUTION
When adjusting the position of the retinal camera in the front-back
direction, slowly bring the retinal camera closer to the patient while looking
at the patient’s eye from the side.
You may injure the patient’s eye if the objective lens makes contact with it.
Slowly bring the stage closer to the patient’s eye so that
a retinal image becomes visible on the observation
monitor.
-15-
5. Basic Photography
Direction
Height
Side movement
Forward/backward
Operation
Turn the operation lever.
Hold the operation lever, and pull it to the left and right to perform
rough adjustment, and tilt it to the left and right to perform fine
adjustment.
Hold the operation lever, and pull it forward and backward to perform
rough adjustment, and tilt it forward and backward to perform fine
adjustment.
The right and left dots (working distance dots) appear
sharpest when the retinal camera is at the optimum
COLOR
R
position.
Working distance dots
Split lines
When the split lines are not visible
Part of the split lines may not be visible if the pupil diameter is small.
In this case, press the small pupil photography switch to turn on the small pupil photography
function. (see page 34)
Concerning the linkup function of the working distance dots
With a setting in the control software, the working distance dot and split line operations can be
linked up.
For details, refer to the operation manual of the control software.
8
Adjust brightness in the field of view for observation.
Turning the brightness adjuster clockwise increases the
brightness for observation, and turning it counterclockwise reduces the brightness for observation.
Concerning the brightness for observation
Standard brightness is the position 4 on the scale.
Considering the strain on the patient’s eye, select the lowest possible brightness level.
-16-
5. Basic Photography
9
Determine the area to be photographed.
Have the patient stare at the green internal eye fixation lamp. If necessary, press a FIX TARGET
switch to move the internal eye fixation lamp to guide the patientís eye. If you press the blink switch,
the internal eye fixation lamp blinks to make the lamp more noticeable to the patient and helps to
guide the patient’s eye. If the eye cannot be guided by the internal eye fixation lamp, use the external
eye fixation lamp.
Concerning the external eye fixation lamp function
In the non-mydriatic mode, you can use either the internal eye fixation lamp or external eye
fixation lamp. The lamp to be used can be selected in the control software.
For details, refer to the Operation Manual of the control software.
Resetting the internal eye fixation lamp position
If you hold down the blink switch for approx. 2 or more seconds, the internal eye fixation lamp
blinks for 2 seconds and can be returned to its initial position.
10
Focus on patient’s eye.
Turn the focus knob to make the split lines into a single
line.
While the focus knob is being turned, the image
appears larger.
COLOR
NG
R
Good
Focus is not being adjusted
During focus adjustment
NG
When the split lines cannot be aligned
If the patient diopter is not in the range of -10 to +15 (D), focusing cannot be performed using the
focus indicator.
In this case, use the diopter compensation function (see page 34) to focus the image so that it
appears sharpest in the observation monitor.
-17-
5. Basic Photography
11
Take the image.
Check that the photography ready lamp lights green.
Check again that the working distance dots appear
COLOR
R
sharpest and the split lines form a single line, and then
press the shutter release button.
Photography ready lamp
Concerning the photography ready lamp
The ready status of the fundus camera is indicated by the color of the photography ready lamp.
This status differs depending on the photography mode and image-taking conditions.
Color
Description
Green
Images can be taken. (Up to 10 images can be taken continuously when in the
fluorescein angiography mode.)
Orange
Images can be taken, but if image-taking is continued, the getting ready status will
soon be established.
Red
The camera is in the getting ready status (no images can be taken). Wait until the
lamp lights up green or orange.
-18-
5. Basic Photography
Concerning the adjustment of the flash intensity (for color, red free, cobalt photography
mode, and fluorescein angiography mode)
The flash intensity is already set to the standard level for each photography mode.
To adjust the intensity, press the flash intensity switches.
Each time one of these switches is pressed, the intensity is increased or reduced by steps of 0.3.
The flash intensity selected is shown by the lighted area of the flash intensity indicator.
Standard
Increased by 87% (equivalent to a step of 0.9)
of the standard
Increased by 52% (equivalent to a step of 0.6)
of the standard
Increased by 23% (equivalent to a step of 0.3)
of the standard
Reduced by 19% (equivalent to a step of 0.3)
of the standard
Reduced by 34% (equivalent to a step of 0.6)
of the standard
Reduced by 47% (equivalent to a step of 0.9)
of the standard
Reduced by 56% (equivalent to a step of 1.2)
of the standard
Reduced by 65% (equivalent to a step of 1.5)
of the standard
Note: For details on the flash intensity adjustment in the fundus autofluorescence mode,
refer to Concerning the adjustment of the flash intensity (for fundus
autofluorescence mode) (page 23).
12
Review the image.
The fundus images and study information are displayed in the control software. Check the images.
To take images of the other eye, repeat steps 7 and the following.
13
End the study.
End the study with the control software.
14
Lock the stage unit.
Press the stage unit lock button to lock the stage unit.
-19-
5. Basic Photography
Concerning locking the stage unit
When the stage unit lock button is pressed, the stage unit can be locked or unlocked in the
following two places.
Lock position
Description
Locked at center
Center position of stage unit (The unit is locked here upon completion of
the examination.)
Position where the stage unit has been drawn to the front right.
(This protects the patient’s eyes from unintended operations either while
Lock at front right
an examination is being undertaken or after an examination has been
completed.)
15
Turn off the power of the retinal camera.
5.3.2 Red-free Photography Mode
1
Enter the study information into the computer.
2
Select the red-free photography mode.
Press the RED FREE switch.
Or select the photography mode with the control
software.
The RED FREE lamp turns on. RED FREE is
displayed on the observation monitor.
3
Perform the procedure starting from step 3 in 5.3.1 Color Photography Mode (see page 14).
-20-
5. Basic Photography
5.3.3 Cobalt Photography Mode
In this mode, photography using the exciter filter for fluorescein angiography is possible.
1
Enter the study information into the computer.
2
Select the cobalt photography mode.
Press the COBALT switch.
Or select the photography mode with the control
software.
The COBALT lamp turns on. COBALT is displayed on
the observation monitor.
3
Perform the procedure starting from step 3 in 5.3.1 Color Photography Mode (see page 14).
5.3.4 Fluorescein Angiography Mode
If taking multiple photographs, a mydriatic must be used.
1
Enter the study information into the computer.
2
Select the fluorescein angiography mode.
Press the FLUO switch.
Or select the photography mode with the control
software.
The FLUO lamp turns on. FLUO is displayed on the
observation monitor.
3
Disinfect the forehead rest and replace the chin
4
Have the patient sit down.
5
Align the height of the patient’s eye.
6
Move the stage position.
7
Align the retinal camera to the eye to be photographed.
8
Adjust brightness in the field of view for observation.
rest paper.
-21-
5. Basic Photography
9
Roughly determine the area to be photographed.
10
Focus on the patient’s eye.
11
Start the timer.
Press the TIMER/C switch as soon as the fluorescein
medium is injected. The buzzer sounds each second,
the lamp lights brightly and the elapsed time is
displayed in the control software.
12
Take the image.
Check again that the working distance dots appear
sharpest and the split lines form a single line, and then
press the shutter release button.
To take images continuously, hold down the shutter
release button. Images can be taken continuously for up
to 10 seconds at a rate of one frame per second.
13
Review the image.
14
Stop the timer.
Press the TIMER/C switch. The elapsed time display is reset.
15
End the study.
16
Lock the stage unit.
Press the stage unit lock button to lock the stage unit.
17
Turn off the power of the retinal camera.
-22-
5. Basic Photography
5.3.5 Fundus Autofluorescence Mode
1
Enter the study information into the computer.
2
Select the fundus autofluorescence mode.
Press the FAF switch.
Or select the photography mode with the control
software.
The FAF lamp turns on. FAF is displayed on the
observation monitor.
3
Perform the procedure starting from step 3 in 5.3.1 Color Photography Mode (see page 14).
Concerning the adjustment of the flash intensity (for fundus autofluorescence mode)
The flash intensity is already set to the standard level for each photography mode.
To adjust the intensity, press the flash intensity switches.
Each time one of these switches is pressed, the intensity is increased or reduced by steps of 0.3.
The flash intensity selected is shown by the lighted area of the flash intensity indicator.
Standard
Reduced by 19% (equivalent to a step of 0.3)
of the standard
Reduced by 34% (equivalent to a step of 0.6)
of the standard
Reduced by 47% (equivalent to a step of 0.9)
of the standard
Reduced by 56% (equivalent to a step of 1.2)
of the standard
Reduced by 65% (equivalent to a step of 1.5)
of the standard
Note: For details on the flash intensity adjustment in the color, red free, cobalt
photography mode, and fluorescein angiography mode, refer to Concerning the
adjustment of the flash intensity (for color, red free, cobalt photography mode, and
fluorescein angiography mode) (page 19).
-23-
5. Basic Photography
5.4 Mydriatic Mode
The Mydriatic mode is designed for photographing the
patient’s eye that a mydriatic is applied. To switch from the
non-mydriatic mode to the mydriatic mode, hold down the
observation mode switch for approx. 2 or more seconds.
The observation mode lamp Myd turns on.
The viewfinder is used for retinal observation.
Check that the diopter of the viewfinder is aligned (see
page 12).
5.4.1 Color Photography Mode
1
Enter the study information into the computer.
Enter the study information with the control software.
2
Select the color photography mode.
Press the COLOR switch.
Or select the photography mode with the control
software.
The COLOR lamp turns on.
3
Disinfect the forehead rest and replace the chin rest paper.
! CAUTION
To prevent the risk of infection, wipe the forehead rest with a disinfectant
such as ethanol or glutaraldehyde whenever the patient is changed.
For details on the disinfection procedure, consult a specialist.
To ensure cleanliness, replace the chin rest paper whenever the patient is
! CAUTION changed.
If chin rest paper is not used, disinfect the chin rest.
4
Have the patient sit down.
Have the patient place his or her chin against the chin rest and his or her forehead against the
forehead rest.
-24-
5. Basic Photography
5
Align the height of the patient’s eye.
Press the CHIN REST switches to adjust the chin rest up or down so that the patient’s eye is aligned
with the height adjustment mark.
Height
adjustment
mark
6
Move the stage position.
! CAUTION
Do not put your fingers or hands between the sliding part and base of the
stage unit or on the panning and tilting sliding parts.
Similarly, do not allow patients to put their hands in these places.
Otherwise, your or their hands or fingers may be pinched, possibly
resulting in injury.
If the stage is locked at the operator’s side, press the
stage lock button to unlock the stage.
Hold the operation lever, pull it towards you, and then
move the stage toward the eye to be photographed.
7
Align the retinal camera to the eye to be photographed.
! CAUTION
When adjusting the position of the retinal camera in the front-back
direction, slowly bring the retinal camera closer to the patient while looking
at the patient’s eye from the side.
You may injure the patient’s eye if the objective lens makes contact with it.
Slowly bring the retinal camera closer to the patient
eye.
-25-
5. Basic Photography
Direction
Height
Side movement
Forward/backward
Operation
Turn the operation lever.
Hold the operation lever, and pull it to the left and right to perform
rough adjustment, and tilt it to the left and right to perform fine
adjustment.
Hold the operation lever, and pull it forward and backward to perform
rough adjustment, and tilt it forward and backward to perform fine
adjustment.
The right and left dots (working distance dots) appear
sharpest when the retinal camera is at the optimum
Split line
position.
Working distance dots
When the split lines are not visible
Part of the split lines may not be visible if the pupil diameter is small.
In this case, press the small pupil photography switch to turn on the small pupil photography
function. (see page 34)
Concerning the linkup function of the working distance dots
With a setting in the control software, the working distance dot and split line operations can be
linked up.
For details, refer to the operation manual of the control software.
8
Adjust brightness in the field of view for observation.
Turning the brightness adjuster clockwise increases the
brightness for observation, and turning it counterclockwise reduces the brightness for observation.
Concerning the brightness for observation
Standard brightness is the position 4 on the scale.
Considering the strain on the patient’s eye, select the lowest possible brightness level.
-26-
5. Basic Photography
9
Determine the area to be photographed.
Use the external eye fixation lamp to guide the patient’s eye.
When it is difficult to guide a patient’s eye using the external eye fixation lamp
Patients’ eyes can also be guided using the internal eye fixation target (sold separately).
For details, contact a Canon representative or distributor.
10
Focus on patient’s eye.
Turn the focus knob to focus the image.
When using the focus indicator, ensure that the split
lines are aligned into a single straight line.
Split line
CX-1
NG
Good
NG
When the split lines cannot be aligned
If the patient diopter is not in the range of -10 to +15 D, focusing cannot be performed using the
focus indicator.
In this case, use the diopter compensation function (see page 34) to focus with the cross hair
line in the viewfinder.
When bringing the target into exact focus
When bringing the target into focus, the focus indicator serves as the general guideline for
focusing.
To bring it into focus even more exactly, focus on the target in such a way that the cross hair lines
in the viewfinder and fundus image can be clearly seen at the same time.
-27-
5. Basic Photography
11
Take the image.
Check that the photography ready lamp lights green.
In a flare-free condition, check the focus again, and press the shutter release button.
Concerning the photography ready lamp
The ready status of the fundus camera is indicated by the color of the photography ready lamp.
This status differs depending on the photography mode and image-taking conditions.
Color
Description
Green
Images can be taken. (Up to 10 images can be taken continuously when in the
fluorescein angiography mode.)
Orange
Images can be taken, but if image-taking is continued, the getting ready status will
soon be established.
Red
The camera is in the getting ready status (no images can be taken). Wait until the
lamp lights up green or orange.
Concerning the adjustment of the flash intensity (for color, red free, cobalt photography
mode, and fluorescein angiography mode)
The flash intensity is already set to the standard level for each photography mode.
To adjust the intensity, press the flash intensity switches.
Each time one of these switches is pressed, the intensity is increased or reduced by steps of 0.3.
The flash intensity selected is shown by the lighted area of the flash intensity indicator.
Standard
Increased by 87% (equivalent to a step of 0.9)
of the standard
Increased by 52% (equivalent to a step of 0.6)
of the standard
Increased by 23% (equivalent to a step of 0.3)
of the standard
Reduced by 19% (equivalent to a step of 0.3)
of the standard
Reduced by 34% (equivalent to a step of 0.6)
of the standard
Reduced by 47% (equivalent to a step of 0.9)
of the standard
Reduced by 56% (equivalent to a step of 1.2)
of the standard
Reduced by 65% (equivalent to a step of 1.5)
of the standard
Note: For details on the flash intensity adjustment in the fundus autofluorescence mode,
refer to Concerning the adjustment of the flash intensity (for fundus
autofluorescence mode) (page 32).
-28-
5. Basic Photography
12
Review the image.
The fundus images and study information are displayed in the control software. Check the images.
To take images of the other eye, repeat steps 7 and the following.
13
End the study.
End the study with the control software.
14
Lock the stage unit.
Press the stage unit lock button to lock the stage unit.
Concerning locking the stage unit
When the stage unit lock button is pressed, the stage unit can be locked or unlocked in the
following two places.
Lock position
Description
Locked at center
Center position of stage unit (The unit is locked here upon completion of
the examination.)
Position where the stage unit has been drawn to the front right.
(This protects the patient’s eyes from unintended operations either while
Lock at front right
an examination is being undertaken or after an examination has been
completed.)
15
Turn off the power of the retinal camera.
-29-
5. Basic Photography
5.4.2 Red-free Photography Mode
1
Enter the study information into the computer.
2
Select the red-free photography mode.
Press the RED FREE switch. Or select the photography mode with the control software.
The RED FREE lamp turns on.
3
Perform the procedure starting from step 3 in 5.4.1 Color Photography Mode (see page 24).
5.4.3 Cobalt Photography Mode
1
Enter the study information into the computer.
2
Select the cobalt photography mode.
Press the COBALT switch. Or select the photography mode with the control software.
The COBALT lamp turns on.
3
Perform the procedure starting from step 3 in 5.4.1 Color Photography Mode (see page 24).
5.4.4 Fluorescein Angiography Mode
1
Enter the study information into the computer.
2
Select the fluorescein angiography mode.
Press the FLUO switch. Or select the photography mode with the control software.
The FLUO lamp turns on.
3
Disinfect the forehead rest and replace the chin rest papar.
4
Have the patient sit down.
5
Align the height of the patient’s eye.
6
Move the stage position.
7
Align the camera to the eye to be photographed.
Concerning the barrier filter
During retinal observations in the fluorescein angiography mode, each time the BA switch is
pressed, the barrier filter is inserted or disengaged. However, while images are being taken, the
filter is inserted automatically.
While the barrier filter is inserted, the lamp above the button lights brightly.
8
Adjust brightness in the field of view for observation.
9
Determine the area to be photographed.
-30-
5. Basic Photography
10
Focus on patient’s eye.
11
Start the timer.
Press the TIMER/C switch as soon as the fluorescein medium is injected. The buzzer sounds each
second, the lamp lights brightly and the elapsed time is displayed in the control software.
12
Take the image.
In a flare-free condition, check the focus again, and press the shutter release button.
To take images continuously, hold down the shutter release button. Images can be taken continuously
for up to 10 seconds at a rate of one frame per second.
13
Review the image.
14
Stop the timer.
Press the TIMER/C switch. The elapsed time display is reset.
15
End the study.
16
Lock the stage unit.
Press the stage unit lock button to lock the stage unit.
17
Turn off the power of the retinal camera.
-31-
5. Basic Photography
5.4.5 Fundus Autofluorescence Mode
1
Enter the study information into the computer.
2
Select the fundus autofluorescence mode.
Press the FAF switch.
Or select the photography mode with the control
software.
The FAF lamp turns on.
3
Perform the procedure starting from step 3 in 5.4.1 Color Photography Mode (see page 24).
Concerning the adjustment of the flash intensity (for fundus autofluorescence mode)
The flash intensity is already set to the standard level for each photography mode.
To adjust the intensity, press the flash intensity switches.
Each time one of these switches is pressed, the intensity is increased or reduced by steps of 0.3.
The flash intensity selected is shown by the lighted area of the flash intensity indicator.
Standard
Reduced by 19% (equivalent to a step of 0.3)
of the standard
Reduced by 34% (equivalent to a step of 0.6)
of the standard
Reduced by 47% (equivalent to a step of 0.9)
of the standard
Reduced by 56% (equivalent to a step of 1.2)
of the standard
Reduced by 65% (equivalent to a step of 1.5)
of the standard
Note: For details on the flash intensity adjustment in the color, red free, cobalt
photography mode, and fluorescein angiography mode, refer to Concerning the
adjustment of the flash intensity (for color, red free, cobalt photography mode, and
fluorescein angiography mode) (page 28).
-32-
6. Photography Auxiliary Functions
6.1 Using the Focus Indicator
The focus indicator switch on the operation panel can be used
to select whether to use the focus indicator or not.
Focus indicator lamp
When the focus indicator is being used, the focus indicator
lamp lights brightly.
Focus indicator switch
6.2 ×2 photography function
Press the ×2 switch.
If the ×2 photography function is enabled, the ×2 lamp is
turned on, and the ×2 photography indication is displayed in
×2 lamp
the viewfinder or on the observation monitor.
If a photograph is taken in this state, the image in the ×2
photography range is saved. The enlarged image is displayed
on the window in the control software.
×2 switch
Viewfinder
×2
×2 indication
Observation monitor
COLOR
R
2
×2 indication
-33-
6. Photography Auxiliary Functions
6.3 Small pupil photography function
The small pupil photography function can be used in the
color, red free or cobalt photography mode.
Small pupil photography indicator
To activate this function, press the small pupil photography
switch on the operation panel.
The small pupil photography lamp turns on brightly.
During the non-mydriatic mode, the small pupil photography
indication (
) and effective photography range are
displayed on the observation monitor. For the details on the
Small pupil photography switch
required pupil diameter, refer to 8.2 Functions in the
Observation and Photography Modes (page 54).
Small pupil photography range
COLOR
R
Note: Do not use the small pupil photography function when the patient’s eyes have been
thoroughly dilated. The images will be overexposed.
Note: The images may contain flare around the edges of the images outside the photography
range.
6.4 Diopter Compensation Function
If the target cannot be brought into focus by turning the focus
knob, use the diopter compensation lever to insert the diopter
compensation lens, and take the photographs. Given below is
Diopter
compensation
lever
the diopter range in which images can be photographed.
0 position: –10 to +15 D
– position: –7 to –31 D
+ position: +11 to +33 D
Note that the focus indicator (see page 33) cannot be used
when the diopter compensation lever is set to the – or +
position. To bring the target into focus in the non-mydriatic mode, turn the focus knob until the fundus
image is clearly visible on the observation monitor; similarly, to bring it into focus in the mydriatic mode,
turn the focus knob until the cross hair lines and the fundus images are clearly visible in the viewfinder.
Note: When operating the diopter compensation lever, set it precisely to the 0, – or + position.
If the lever is set to a position midway between two settings, fundus images will not be
taken properly.
-34-
6. Photography Auxiliary Functions
6.5 Panning Function and Tilting Function
6.5.1 Panning Function
This function is used to take photographs of peripheral areas
in the side-to-side direction.
First, turn the panning lock knob to the
position, and then
adjust the angle of the retinal camera in the side-to-side
To lock
direction.
Before carrying the retinal camera around, turn the panning
lock knob to the
position.
To unlock
6.5.2 Tilting Function
This function is used to take photographs of peripheral areas
in the up/down direction.
First turn the tilting lock lever to the
position, and then
adjust the angle of the retinal camera in the vertical direction.
To unlock
To lock
The lever clicks in at the horizontal position so you can easily
find the position.
Before carrying the retinal camera around, turn the tilting
lock lever to the
position.
Note: Hold the pan-tilt grip when adjusting the camera angle by panning or tilting.
Note: To lock the camera at the tilt angle, turn the tilting lock lever precisely to the
and check that the camera is completely locked.
If it is not completely locked, perform these steps again.
-35-
position,
6. Photography Auxiliary Functions
6.6 Using the External Monitor
The image on the observation monitor can be viewed on an external monitor. Please ask your Canon
representative or distributor about which external monitors can be used.
1
Before connecting or disconnecting the cable, turn off the power for the retinal camera.
2
Connect the supplied video cable to the digital camera.
Connect the video cable to the VIDEO OUT terminal
on the digital camera.
Firmly insert the cable up to the plug base.
3
Put the cable into the grooves in the camera cover.
Fit the core of the video cable into the space in the
cover.
Be careful not to damage the cable while putting it into
the groove.
4
Attach the camera cover.
Place the camera cover hook onto the strap mount of
the digital camera. Fit the cover screw into the threaded
socket for the tripod on the bottom of the camera while
taking care that the cable is not pinched, and secure the
camera cover.
-36-
6. Photography Auxiliary Functions
5
Connect the video cable to the external monitor.
Connect the video cable to the video input terminal on
the external monitor.
Turn on the external monitor, and get the input selected
R
on the monitor.
AUDIO
L
Yellow plug
VIDEO
6
Turn on the power for the retinal camera and digital camera.
The image is displayed on the connected external monitor (No image is displayed on the digital
camera observation monitor).
Note: Do not use any video cable other than the one provided. Images might not be displayed if
a different video cable is used.
Note: Dress the video cable for smooth operation of the retinal camera. It can be bundled with
the retinal camera cables.
The cables may be disconnected if they are forcibly pulled.
-37-
7. Daily Inspection and Maintenance
! WARNING
Repairs must be performed by a technician qualified for the specific
repair work.
If it is not repaired properly, it may cause fire, electric shock, or
accident.
! CAUTION For safety reasons, be sure to inspect the instrument before using it.
7.1 Inspection
In order to ensure that the instrument is used safely and normally, please be sure to inspect the instrument
before use.
If the inspection finds a fault, and you are unable to correct the problem, please contact a Canon
representative or distributor.
It is recommended that you either copy the tables of 7.1.1 Before Turning ON the Power (see
page 39) and 7.1.2 Turning ON the Power (see page 40), or create a separate checklist for recording
the inspection results.
-38-
7. Daily Inspection and Maintenance
7.1.1 Before Turning ON the Power
! WARNING
Clean the plug of the power cable periodically by unplugging it from the
AC outlet and removing dust or dirt from the plug, its periphery and AC
outlet with a dry cloth. If the cable is kept plugged in for a long time in a
dusty, humid or sooty place, dust around the plug will attract moisture, and
this could cause insulation failure which could result in a fire.
Main unit
Power cable
Location
! WARNING
For safety reasons, be sure to turn OFF the power when the following
inspections are going to be performed.
Otherwise, it may result in electric shock.
Result
Inspection
Check that the power cable is
not damaged and the covering
is not torn.
Month
Month
Month
Remedy
Day
Day
Day
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there is any
problem.
Check that the power cable is
fully inserted into the
connector on the main unit
and the AC outlet.
Also check the connections of
the cables between the
equipment.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Fully insert the power cable into
the connector and the AC outlet
with the power turned OFF.
Also, be sure to plug in cables
between equipments as far as they
will go.
Check that the cover and parts
are not damaged and are not
loose.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if the cover and parts
are damaged or are loose.
Grasp the operation lever and
tilt it back and forth and from
left to right, checking that the
stage unit moves smoothly.
Check that the locking of the
stage unit operates properly.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Turn the operation lever, and
check that the retinal camera
main unit moves smoothly up
and down as far as the upper
and lower limits.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Check that the panning and
tilting operations can be
performed smoothly, and that
there are no unusual sounds.
Also check that the locking of
the panning and tilting
operates properly.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Check that there is no dust or
other substances on the
viewfinder.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Clean the viewfinder if it is dirty
(see page 45).
Disinfect the forehead rest. If
the chin rest paper will not be
used, be sure to disinfect the
chin rest as well.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Disinfect the forehead rest
whenever changing patients (see
page 45).
-39-
Contact a Canon representative or
distributor if there is any
problem.
Digital camera Location
7. Daily Inspection and Maintenance
Result
Inspection
Check that the digital camera
is firmly attached and the
cables are connected
correctly.
Month
Month
Month
Day
Day
Day
Pass/
Fail
Pass/
Fail
Pass/
Fail
Remedy
Contact a Canon representative or
distributor if there is any
problem.
7.1.2 Turning ON the Power
Main unit
Location
Turn on the power of all the equipment which has been connected as a system.
Result
Inspection
Month
Month
Month
Day
Day
Day
Remedy
Check whether the POWER
lamp lights.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if the POWER lamp
fails to light even though the
power cable is connected
properly.
Check that there is no dirt or
scratches on the objective
lens.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Clean the lens if it is dirty (see
page 44).
Contact a Canon representative or
distributor if it is scratched.
Place a sheet of paper in front
of the objective lens, turn the
brightness adjuster, and check
that the brightness for
observation changes (on the
retinal observation screen).
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there are no
changes.
Place a sheet of paper in front
of the objective lens and press
the shutter release button to
check that the flash is emitted.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if no flash is emitted.
Set to fluorescein
angiography mode and press
the TIMER/C switch to check
that the timer starts and stops
properly.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there are any
problems.
Check that the chin rest
moves smoothly up and down
when the CHIN REST switch
is pressed.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there are any
problems.
-40-
7. Daily Inspection and Maintenance
7.2 Before Calling a Service Technician
Perform the appropriate remedy when any of the problems below appear in inspection or during operation
or a warning is displayed in the control software.
If performing the remedy still does not fix the problem, or the warning is still displayed, turn OFF the
power, and contact a Canon representative or distributor.
Please be sure to describe the problem or warning message in detail.
7.2.1 Troubleshooting
Problem
When the retinal camera
main unit is raised and
lowered, its movement is
stiff at the upper limit.
Cause
Remedy
The camera was subjected to
severe vibration while it was in
transit.
Turn the operation lever, and move the
main unit to the upper level by raising
and lowering it several times.
The power cable is disconnected.
Turn OFF the power to the main unit, and
connect the cable properly.
An indoor circuit breaker has
been tripped.
Find and resolve the cause of the tripped
circuit breaker, and reset the circuit
breaker.
The cable is cut.
Turn OFF the power to the main unit, and
contact a Canon representative or
distributor.
When the power is turned
on, the buzzer sounds
(beep, beep beep), the
display lamps on the
operation panel blink.
A system error has occurred.
Turn off the main unit’s power, and
contact a Canon representative or
distributor.
The external eye fixation
lamp does not light up or
flicker.
The external eye fixation lamp is
not attached properly.
Insert the external eye fixation lamp
fully into the connector on top of the face
rest.
The external eye fixation
lamp does not stop at the
specified position.
The holding power has become
weak.
Contact a Canon representative or
distributor.
The observation monitor
screen appears dim and is
hard to view.
The brightness of the observation
monitor is not appropriately
adjusted.
Refer to the operation manual of the
control software and adjust the monitor.
The digital camera detects a
malfunction.
Turn off and on the power of the retinal
camera.
Note: If the same error keeps
appearing, there may be a
problem. Contact a Canon
representative or distributor.
The digital camera is turned off.
Turn on the digital camera.
The mydriatic mode is selected.
Select the non-mydriaric mode (see
page 14).
An external monitor is connected.
Disconnect the external monitor (see
page 36).
The POWER lamp does
not light up when the
power is turned ON.
An error (Err xx) appears
on the observation
monitor.
The observation monitor
does not turn on.
-41-
7. Daily Inspection and Maintenance
Problem
Cause
Remedy
The working distance dots The observation light is too
are difficult to view.
bright.
Lower the observation light intensity.
The working distance dots
The positioning is not correct.
are not visible.
Perform positioning (see page 15).
The retinal image is still
not visible even when
increasing the brightness
for observation with the
brightness adjuster.
Contact a Canon representative or
distributor.
The lamp for observation is burnt
out.
It is hard to see the split
The brightness for observation is
lines and working distance
too strong.
dots.
Reduce the brightness for observation.
The working distance dots
are not visible.
The positioning is not correct.
Flare is visible.
Perform positioning.
The split lines are not
visible.
The split lines do not
align.
The split bar is not displayed.
Press the focus indicator switch to
display the split bar.
The pupil is not dilated enough.
Make the patient’s pupil dilate. If the
pupil is still too small, press the small
pupil photography switch to activate the
small pupil function.
Eyebrows or eyelashes are
blocking the pupil.
Instruct the patient to open his or her eye
wider.
The patient’s vitreous body is
unclear.
Focusing cannot be performed with the
split lines.
The diopter compensation lever is
Set to the no compensation position.
set to the - side or + side.
The Internal Eye Fixation Target
(sold separately) is used.
The split lines cannot be displayed.
In a case like this, turn the focus knob to
adjust to the position where the cross hair
lines and fundus image are clearly
visible at the same time inside the
viewfinder.
The LED lamp for the split lines
display does not light.
Contact a Canon representative or
distributor.
Turn the focus knob to focus so that the
The diopter of the patient’s eye is cross hair lines in the field of view and
outside the range of -10 to +15 D. the fundus image can be seen clearly at
the same time.
-42-
7. Daily Inspection and Maintenance
Problem
Sharp images cannot be
taken.
Black points appear on
photographed images in
the same area.
White spots appear on the
images.
Cause
Remedy
The pupil is not dilated enough.
Make the patient’s pupil dilate. If the
pupil is still too small, press the small
pupil photography switch to activate the
small pupil function.
The positioning is not correct.
Perform positioning.
The image is unfocused.
Focus the image.
Eyebrows or eyelashes are
blocking the pupil.
Instruct the patient to open his or her eye
wider.
The patient’s eye is opaque.
A sharp image cannot be taken.
The patient blinked when the
image was taken.
Take another image.
The flash intensity, diopter
compensation lever, and
viewfinder diopter compensation
ring have not been adjusted.
Adjust the settings.
The objective lens is dirty.
Clean the lens (see page 44).
The patient is wearing contact
lenses.
Instruct the patient to remove his or her
contact lenses.
Dust or other substance is
adhering to the surface of the
objective lens.
Clean the lens (see page 44).
Dust or other substance is
adhering to the surface of the
imaging sensor of the digital
camera.
Perform a self-cleaning (see page 52).
Note: The imaging sensor is very
sensitive. Contact a Canon
representative or distributor for
cleaning it.
The objective lens is dirty.
Clean the lens (see page 44).
A patient’s eyelashes have
appeared on the images.
Instruct the patient to open his or her eye
wider. Take another image.
-43-
7. Daily Inspection and Maintenance
7.3 Cleaning and Disinfection
7.3.1 Objective Lens
Note: Do not wipe off or rub the objective lens when there is dust or other substances on it.
This could scratch the lens surface.
Note: Never wipe the objective lens with disinfecting ethanol, eyeglass lens cleaner, or cleaning
paper containing silicon.
The lens surface could be damaged or the surface may not be completely wiped off.
Note: Do not let the tip of a blower touch the objective lens.
Use the procedure below to clean the objective lens when it is dirty.
Please order the lens cleaning paper, lens cleaner, and blower from a Canon representative or distributor.
1
Turn ON the power of the retinal camera.
Press the power switch of the retinal camera to the I
side (ON).
ON
OFF
2
Adjust the brightness.
Set the brightness adjuster in the 8 to MAX range.
3
Check for any dust or dirt.
Look at the objective lens from the patient side to check for any dust or dirt.
4
Blow away any dust or dirt.
Remove the brush from the blower brush, and use the
blower to blow away any dust or dirt on the objective
lens.
-44-
7. Daily Inspection and Maintenance
5
Wipe off the objective lens.
Gently wipe the lens with Canon-designated lens
cleaning paper that has been slightly wet with the
Canon-designated lens cleaner.
Starting from the center of the lens, wipe the lens in
increasingly larger spirals toward the circumference.
Change the paper, and wipe several times until the dirt
is removed and the cleaner is completely wiped off.
7.3.2 Viewfinder
Note: Do not let the tip of a blower touch the viewfinder lens.
If the viewfinder is dirty, blow off any dirt or dust on the lens
surface with a blower.
7.3.3 Forehead Rest
CAUTION
To prevent the risk of infection, wipe the forehead rest with disinfecting
ethanol, glutaraldehyde solution, or other disinfectant whenever changing
patients.
For more details on the disinfection procedure, please consult a specialist.
Please also consult a specialist when using a disinfectant other than those
specified above or when using ethanol with another disinfectant added
since this could damage the forehead rest.
Disinfect the forehead rest whenever changing patients.
-45-
7. Daily Inspection and Maintenance
7.3.4 Retinal Camera and Digital Camera Outer Covers
! WARNING
Before cleaning the instrument, be sure to turn OFF the power, and
unplug the power cable from the AC outlet.
Never use alcohol, benzine, thinner or any other flammable cleaning
agents. Otherwise, fire or electric shock may result.
Note: Do not clean the cover of the instrument with lens cleaner. The cover of the instrument
could be damaged.
Use the procedure below to clean the cover when it is dirty.
1
Turn OFF the power of the retinal camera.
Press the power switch of the retinal camera to the O side (OFF).
2
Unplug the power cable.
3
Wipe the instrument with cleanser.
Soak a soft cloth with diluted neutral cleanser and wring it out.
4
Wipe off with a rinsed cloth.
Wipe with a cloth that has been dipped in water and wringed out.
-46-
7. Daily Inspection and Maintenance
7.4 Refilling the Chin Rest Paper
Note: If the chin rest paper will not be used, be sure to disinfect the chin rest whenever the
patient is changed in the same way as the forehead rest.
Note: The chin rest paper is a consumable product (sold separately). To purchase the chin rest
paper, please contact a Canon representative or distributor.
1
Pull out the chin rest right and left holding pins.
2
Insert the holding pins in the chin rest paper.
Insert the holding pins in the right and left holes of the chin rest paper.
About 100 sheets of paper can be mounted.
3
Attach the chin rest paper to the chin rest.
Attach the chin rest paper with holding pins attached to
the chin rest.
-47-
7. Daily Inspection and Maintenance
7.5 Removing and Installing the Digital Camera
It is not necessary to remove and re-install the digital camera during normal usage.
The procedure below is used to remove and install the digital camera during maintenance and inspection of
the digital camera or other work.
Note: Do not touch the lens part of the main unit and the mirror part of the digital camera when
attaching and detaching the digital camera from the retinal camera body. If any dirt,
fingerprints, dust, and other foreign objects attaches to the lens or mirror part, you cannot
take a good image.
7.5.1 Removing the Digital Camera
Note: If the digital camera is removed and left unattached, dust and other foreign objects can
enter the main unit and digital camera. After removing the digital camera, be sure to also
attach the caps to each mount.
1
Turn off the power for the retinal camera and digital camera.
Set the power switch of the digital camera to the OFF position.
Press the power switch of the retinal camera to the O side (OFF).
2
Remove the camera cover.
Loosen the screws of the camera cover to remove the
cover.
3
Disconnect the cables.
Remove the respective cables from the digital terminal,
PC terminal, and remote control terminal.
-48-
7. Daily Inspection and Maintenance
4
Remove the digital camera.
While pressing down on the lens unlock button of the digital camera
, slowly turn
counterclockwise so that the camera mount index I on the top of the digital camera is aligned with
the seam of the cover of the retinal camera body
.
Seam of cover
Release the lens unlock button
.
While holding the digital camera, pull the knob of the
mount lock lever
UNLOCK position
, and pull up the lever to the
.
Note: Hold the digital camera securely so that it does not drop.
Remove the digital camera.
-49-
Digital camera
mount index
7. Daily Inspection and Maintenance
5
Remove the DC coupler.
Open the battery compartment cover on the digital
camera, and then slide the battery lock lever to remove
the DC coupler.
After removing the DC coupler, close the battery
compartment cover.
6
Attach the cap to the camera mount.
Attach the mount cap to the retinal camera mount.
Also attach the body cap to the digital camera.
Note: The digital retinal camera CX-1 is already installed with an digital camera for this
instrument. Please take careful note of the following points if the digital camera must be
removed or installed.
• Perform the work quickly in a clean room.
• If the digital camera will be removed and stored, be sure to attach the supplied mount
cap to the retinal camera body and the body cap to the digital camera.
• Wipe off any dust or other substances on the caps before attaching them.
7.5.2 Installing the Digital Camera
1
Turn off the power of the retinal camera.
Press the power switch of the retinal camera to the O side (OFF).
2
Attach the DC coupler to the digital camera.
Open the battery compartment cover on the digital
camera, and attach the DC coupler.
After attaching the DC coupler, pass the cord through
the DC coupler cord hole, and then close the battery
compartment cover.
-50-
7. Daily Inspection and Maintenance
3
Mount the digital camera to the retinal camera.
Set the mount lock lever to the UNLOCK position
.
Fit the digital camera lens mount onto the camera
mount of the retinal camera body
.
Seam of cover
When fitting them together, be sure that the digital
camera mount index mark
is aligned with the cover
seam on the retinal camera body.
Hold the knob of the mount lock lever, and push down
the lever to the LOCK position
.
Turn the digital camera clockwise until it clicks into
place
.
Check that the digital camera is firmly secured in place.
“Clicks”
-51-
Digital camera
mount index
7. Daily Inspection and Maintenance
4
Attach the cable to the digital camera.
Open the terminal cover of the digital camera, and
connect the respective cables to the digital terminal, PC
terminal, and remote control terminal.
5
Attach the camera cover to the digital camera.
Place the camera cover hook onto the strap mount of
the digital camera. Fit the cover screw into the threaded
socket for the tripod on the bottom of the camera, and
secure the camera cover.
7.5.3 Cleaning the Digital Camera (imaging sensor)
In this digital camera, a self-cleaning sensor unit is activated when the power switch is set to the ON or
position and to the OFF position for automatically removing dust and other substances that are
adhering to the imaging sensor front surface. The user does not need to pay particular attention to this
function in normal operation.
The self-cleaning sensor unit is capable of removing most dust and substances that can appear in images.
Use the operation below to perform automatic cleaning.
1
Turn on the power of the retinal camera body.
2
Set the power switch of the digital camera from ON to OFF, or from
to OFF.
The self-cleaning sensor unit on the front surface of the CMOS sensor (low-pass filter) is activated
and performs cleaning.
3
Return the main body to its original state, and check that the dust and other substances
have been removed.
Note: If the above procedure does not remove the dust and other substances, the cause may be
something else. Please contact a Canon representative or distributor.
-52-
8. Main Specifications and Performance
8.1 Main Specifications
Field angle
Mydriatic mode
Non-Mydriatic mode
50°
45°
Magnification view
1.1× (Image size on the sensor)
Image magnification of observation
(viewfinder)
10×
Diameter of pupil required
Mydriatic mode
Non-Mydriatic mode
ø 5.1 mm or more; ø 4.3 mm or more when SP function works
ø 4.3 mm or more; ø 3.8 mm or more when SP function works
Working distance
35 mm
Focus adjustment range
-10 to +15D (Without compensation lens)
-31 to -7D (when using negative compensation lens)
+11 to +33D (when using positive compensation lens)
(Split-line aligning type focusing)
Viewfinder diopter compensation
range
±5 D
Flash intensity
Linked to the photography mode
Can be set manually
Camera
Digital camera
Eye fixation lamp
External eye fixation lamp
Internal eye fixation lamp, LED
Internal Eye Fixation Target (sold separately)
Moving range
Stage horizontal movement
110 mm side to side, 65 mm front and back
Main unit vertical movement
30 mm
Chin rest vertical movement
60 mm
Panning
30° to the right and left
Tilting
15° up and 10° down
Rated power supply
AC 100 V to 240 V, 50/60 Hz, 8 A to 4 A
Dimensions
320 (W) × 531 (D) × 577 (H) mm
Mass
26 kg (including the digital camera (0.8 kg))
Note: In the International System of Units (SI), the expression 1D (diopter) = 1m-1.
-53-
8. Main Specifications and Performance
8.2 Functions in the Observation and Photography Modes
The following table shows the availability of functions when each of observation modes is used with each
photography mode. Available options are indicated with the check mark “9”. Unavailable options are
indicated with “N/A.” It also shows the specifications in those observation and photography modes.
Non-Mydriatic Mode
Function
COL*1
RF*2
Viewfinder
RF*2
COB*3 FLUO FAF
9
9
N/A
Near-infrared
Visible light
45°
50°
Internal or external eye fixation lamp
External eye fixation lamp
9
9
9
9
4.3 mm
5.1 mm
9
9
Monitor
Observation
Illumination
Field Angle
Fixation Lamp
COB*3 FLUO FAF COL*1
N/A
Observation
Used Eye
Mydriatic Mode
Focus Indicator
Working Distance
Dot
Minimally
required pupil
diameter
x2 Photography
Small Pupil
Photography
Function
9 with pupil size of
ø 3.8 mm
Timer
Barrier Filter
N/A
N/A
9
N/A
*1 COL: COLOR mode
*2 RF:
RED FREE mode
*3 COB: COBALT mode
-54-
N/A
9 with pupil size of
ø 4.3 mm
N/A
N/A
9
N/A
N/A
9
N/A
8. Main Specifications and Performance
8.3 Guidance and Manufacturer’s Declaration for EMC Directive
Electromagnetic Emissions
The CX-1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CX-1 should assure that it is used in such an
environment.
Emission Test
Compliance
RF emissions
CISPR 11
GROUP 1
RF emissions
CISPR 11
Class B
Electromagnetic Environment – Guidance
The CX-1 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electromagnetic equipment.
Harmonic emissions
The CX-1 is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Class A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Compliances
IEC 61000-3-3
-55-
8. Main Specifications and Performance
Electromagnetic Immunity
The CX-1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CX-1 should assure that it is used in such an
environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electrostatic
discharge (ESD)
±(2, 4, 6) kV
contact
±(2, 4, 6) kV
contact
IEC 61000-4-2
±(2, 4, 8) kV air
±(2, 4, 8) kV air
±2 kV for power
supply lines
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV for input/
output lines
±1 kV
differential
mode
±1 kV
differential
mode
±2 kV common
mode
±2 kV common
mode
<5% UT
(>95% dip in
UT) for 0.5 cycle
<5% UT
(>95% dip in
UT) for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in
UT) for 5 sec
<5% UT
(>95% dip in
UT) for 5 sec
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
3 A/m
3 A/m
IEC 61000-4-8
Electromagnetic Environment –
Guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the CX-1 requires continued
operation during power mains
interruptions, it is recommended
that the CX-1 be powered from an
uninterruptible power supply.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
-56-
8. Main Specifications and Performance
Immunity Test
IEC 60601
Test Level
Compliance
Level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
Electromagnetic Environment –
Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the CX-1, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site surveya, should be
less than the compliance level in each
frequency rangeb.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflections from structures, object and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. For assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the CX-1 is used exceeds
the applicable RF compliance level above, the CX-1 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the CX-1.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
-57-
8. Main Specifications and Performance
Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the CX-1
The CX-1 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the CX-1 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the CX-1 as recommended
below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum
output power of
transmitter
Conducted RF
W
150 kHz ~ 80 MHz
80 MHz ~ 800 MHz
800 MHz ~ 2.5 GHz
d = 1.2 P
d = 1.2 P
d = 2.3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Radiated RF
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
NOTE 1: To maintain the optimum EMC performance, do not use cables other than supplied, nor
extend the cable lengths more than specified.
Name
Type
Length
Remarks
AC Power Cable
Non-Shielded
3.0 m fixed-length
Supplied
USB Cable
Type AB connector plug
supporting USB 2.0 Hi-Speed
Max. 3.0 m
Not supplied
USB Cable
Type AB connector plug
supporting USB 2.0 Hi-Speed
Max. 3.0 m
Not supplied
NOTE 2: Do not install the CX-1 adjacent to or stacked with other electric/electronic equipment.
NOTE 3: Essential Performance (against Electromagnetric Disturbances)
• No unintentional raise/lower movements of Chin Rest shall occur.
• No unintentional flashing shall occur.
• No degradation of Photographic quality, which could lead erratic diagnose.
• No change in operation status, or stored data.
-58-
9. Service Information
Repair
If problem cannot be solved even after taking the measures indicated in Section 7, contact a Canon
representative or distributor for repair.
When requesting repair, please provide us with the following information by referring to the rating label on
the main unit.
Name of the instrument:
CX-1
Serial number of the instrument:
6-digit number indicated on the rating label
Phenomenon:
In detail.
Limit for Supplying Performance Parts for Repair
Performance parts (parts required to maintain the functioning of the product) of this product will be
stocked for eight years after discontinuance of production, to allow for repair.
Expendable Parts Replaced by Service Person
Following parts are apt to become worn out or to deteriorate due to the characteristics of the material or
structure.
These parts cannot be replaced by the user. If these parts are found to be worn out or to have deteriorated
during daily or regular inspection, contact a Canon representative or distributor for repair.
•
Halogen lamp for observation
•
Xenon flash tube for photography
•
Base board (USB connector)
-59-
CANON INC. Medical Equipment Group
30-2, Shimomaruko 3-chome, Ohta-ku, Tokyo, Japan
Telephone: (81)-3-3758-2111
CANON U.S.A., INC.
CANON MEDICAL SYSTEMS Eye Care Systems Department
15975 Alton Parkway, Irvine, CA 92618, U.S.A.
Telephone: (1)-949-753-4000
CANON EUROPA N.V. Medical Products Division
Bovenkerkerweg 59-61, 1185 XB Amstelveen, The Netherlands
Telephone: (31)-20-545-8926
CANON (CHINA) CO., LTD. Medical Equipment Division
15F. Jinbao Building, No.89 Jinbao Street,
Dongcheng District, Beijing, 100005, China
Telephone: (86)-10-8513-9999
CANON SINGAPORE PTE. LTD. Medical Equipment Department
1 Harbour Front Avenue, #04-01 Keppel Bay Tower, Singapore 098632
Telephone: (65)-6799-8888
CANON AUSTRALIA PTY. LTD. Optical Products Division
1 Thomas Holt Drive, North Ryde, NSW 2113, Australia
Telephone: (61)-2-9805-2000
L-IE-5169