Download Critikon Dinamap S Service manual

Transcript
™
DINAMAP Compact
Vital Signs Monitor
Portable Adult/Pediatric
and
Neonatal Monitoring
Service Manual
Part No: 776-856
1
List of Effective Pages
Part No./Rev
776856
776856B
Page No.
All
All
U.S. Patent 5,579,776
U.S. Patent 5,170,795
U.S. Patent 4,349,034
U.S. Patent 5,052,397
U.S. Patent 4,360,029
U.S. Patent 4,543,962
Patents Pending
Date of Latest Revision
Original (May. 2000)
June 2001
U.S. Patent 4,754,761
U.S. Patent 4,501,280
U.S. Patent 4,638,810
U.S. Patent 4,546,775
U.S. Patent 4,638,810
U.S. Patent 4,546,775
U.S. Patent 5,518,000
European Patents:
104771 104772 217918 225256 335357
CAUTION: Federal (U.S.A.) law restricts this device to sale by or on order of a health care
practitioner.
The contents of this document, including all figures and drawings, is proprietary information of
GE Medical Systems, provided solely for purposes of operation, maintenance, or repair of
Dinamap™ Compact Monitors.
Dissemination for other purposes or copying thereof without the prior written consent of GE
Medical Systems, Tampa, Florida, is prohibited. Illustrations may show design models;
production units may incorporate changes.
©GE Medical Systems Information Technologies 2001 TAMPA, FL 33614
Printed in the U.S.A. All rights reserved.
Reissues and Updates
Changes occurring between issues are addressed through Change Information
Sheets and replacement pages. If a Change Information Sheet does not
accompany this manual, it is correct as printed.
Errors & Omissions Excepted
If, in the normal use of this manual, errors, or incorrect data are found, please notify:
United States
GE Medical Systems Information Technologies.
4502 Woodland Corporate Boulevard
Tampa, FL 33614
United Kingdom
Monitor House
Unit 3 Cherrywood
Chineham Business Park
Basingstoke
Hants RG24 8WF
2
™
DINAMAP Compact Vital Signs Monitor Model TS
3
TABLE OF CONTENTS
1 INTRODUCTION
1.1
Scope of Manual
7
1.2 Warranty and Service
1.2.1 Packing Instructions
1.2.2 Service Loan Units
1.2.3 Repair Parts
1.2.4 Replacement Accessories
7
8
8
8
9
1.3
Service Centers
9
1.4
Installation and Operation
10
1.5
Warnings & Precautions
10
1.6 Symbol Definitions
1.6.1 Safety Symbols
1.6.2 Key and Display Symbols
1.6.3 Connector Symbols
10
10
10
11
1.7
11
Electrical, Mechanical & Environmental Specifications
2 PRODUCT DESCRIPTION
15
2.1
Introduction
15
2.2
Physical Description
15
2.3
Overall Principles of Operation
19
2.4 Functional Description
2.4.1 System Processor
2.4.2 Temperature System (TS & T models only)
2.4.3 NIBP System
2.4.4 Pneumatics System
2.4.5 SpO2 System (TS & S models only)
2.4.6 Power Supply
2.4.7 Audio Amplifier
2.4.8 Alarm and Communication Interfaces
2.4.9 User Controls
2.4.10
Displays
2.4.11
Printer
3 MAINTENANCE
3.1 General Maintenance
3.1.1 Replacing Fuses
3.1.2 Periodic Maintenance
3.1.3 Care of the Storage Battery
4
7
19
19
20
21
21
23
23
24
24
25
25
26
27
27
27
27
28
3.2 Service Modes
3.2.1 The Clinician Menu (Service mode 1 2 3 4)
3.2.2 The Service Menu (Service mode 2 2 1 3)
29
31
32
3.3 Calibration - Checking & Re-calibrating
3.3.1 Checking of NIBP Calibration
3.3.2 NIBP Calibration
3.3.3 NIBP Pop Off Checking
3.3.4 NIBP Pop Off Pressure Calibration
3.3.5 Predictive Temperature Calibration Check
3.3.6 Predictive Temperature Calibration
36
36
37
38
39
40
40
3.4 Monitor Disassembly
3.4.1 Removal of the Front Panel
3.4.2 Removal of the Switch board, Display board and Liquid Crystal Display module
3.4.3 Removal of the Chassis
3.4.4 Removal of the Main Board and Power Supply board from the chassis
3.4.5 Removal of the Pump and Pressure Transducer
3.4.6 Removal of the Printer
3.4.7 Reassembling the Monitor
42
42
44
46
47
48
49
50
4 MANUAL RELEASE TEST PROCEDURE
51
4.1 General
4.1.1 Test Equipment
4.1.2 Test Conditions
4.1.3 Service mode
51
51
51
51
4.2 Test Procedure
4.2.1 Power Up Tests
4.2.2 External DC Supply and Battery Indication Operation
4.2.3 Temperature Requirements
4.2.4 Pulse Oximeter Requirements
4.2.5 NIBP Requirements
4.2.6 Printer Operation
4.2.7 Calibration Time Stamp
52
52
52
52
53
54
55
55
4.3
56
Manual Release Test Results Sheets
5 TROUBLE SHOOTING.
58
6 TECHNICAL INFORMATION
64
6.1
Host Port Connector (rear panel)
64
6.2
Remote Alarm Connector (rear panel)
64
6.3
Accessory Part Numbers
65
6.4
DINAMAP™ Compact Service Spares List
69
6.5
Procedural & Error Alarm Code Table
75
5
7 GLOSSARY OF TERMS AND ABBREVIATIONS
77
8 SERVICE DIAGRAMS
78
6
1 Introduction
1.1 Scope of Manual
This Service Manual provides service and parts repair information for the DINAMAP™ Compact Vital Signs
Monitor.
This manual is intended for use by service technicians who are familiar with electromechanical devices and
digital and analog circuit techniques.
!
WARNING
To reduce the risk of electric shock, do not remove the instrument’s covers.
Refer servicing to qualified service personnel.
Only trained service technicians should perform repairs on this equipment.
Take particular care when servicing the power supply assembly.
For information about operating the monitor in a clinical environment, refer to the DINAMAP™ Compact Vital
Signs Monitor Operation Manual (p/n 776-980.)
1.2 Warranty and Service
The warranty for the product is enclosed with the product in the shipping carton. All repairs on products
under warranty must be performed by or approved by Technical Service personnel. Unauthorized repairs
will void the warranty. Qualified electronic service personnel should repair products not covered by
warranty.
Extended warranties may be purchased for most products. Contact your sales representative for details and
pricing.
If the product fails to operate correctly, or if assistance, or service is required, contact the Technical Support
Department at Critikon. Before doing so, it is helpful to attempt to duplicate the problem and to confirm the
correct operation of all the accessories to ensure that they are not the cause of the problem.
Prior to calling, please be prepared to provide:
??
??
the product name and model number
a complete description of the problem
and if repair parts or service are necessary:
??
??
??
??
the product serial number
your business name and address, and Critikon Company account number
a purchase order number if parts or service are to be purchased
the part number of the required spare parts
7
If your monitor requires warranty, extended warranty, or non-warranty service, call Technical Support and a
representative will assist you. To expedite service where the product has external chassis or case
mechanical damage, please advise the Technical Service representative when you call.
The representative will record all the necessary details and will provide the information, which is required for
all returned products.
Prior to returning the monitor, contact GE Medical Systems Information Technologies:
1-877-274-8456
This service is available Monday to Friday between the hours of 8 am and 7 pm EST excluding holidays.
1.2.1 Packing Instructions
??
??
??
??
??
??
Remove all hoses, sensors, power cables and the battery from the monitor.
Only pack the accessories, which you are requested to return, and pack these in a separate bag
within the same carton as the monitor.
Use the original shipping carton and packing materials, if available.
Place the monitor in a plastic bag and tie or tape it shut to prevent loose particles entering the
product.
Pack 4 to 6 inches of padding around all sides of the monitor.
Use packing tape to securely close the carton.
Insurance is at the customer’s discretion. Any claims for damage must be initiated by the sender.
1.2.2 Service Loan Units
A loan unit is provided on request, at no charge during the service life of the product, while Critikon is
performing the repair. The loaner equipment will be shipped to your site within 2 business days.
??
??
??
Critikon will pay for shipping the loaner unit to the customer during warranty repairs.
The customer must pay shipping charges when a loaner unit is issued for a non-warranty repair.
The customer must pay shipping charges when a loaner unit is returned.
All loaner units must be returned within the time specified otherwise a rental fee may be incurred.
1.2.3 Repair Parts
Repair parts can be ordered from Customer Care via telephone or fax. Complete assemblies such as the
power supply board are available on an exchange basis. Contact Technical Support for details.
Fax: 813-887-2403
Please allow one working day for confirmation of faxed orders. All such orders must contain the following
information:
??
??
??
??
8
Your business name, address and telephone number
Your fax number
Your purchase order number
Your account number.
1.2.4 Replacement Accessories
Replacements such as hoses and sensors must be purchased from GE Medical Systems.
Tel: (877) 274-8456
Please have your account number and the reorder / product code available for the item you wish to order.
A table of accessories and replacement part numbers appears in section 6.3.
1.3 Service Centers
For service, repairs or parts information, contact either the following Critikon office or your local Critikon
Sales Representative
North America
GE Medical Systems Information Technologies
4502 Woodland Blvd
Tampa, FL, 33614
877-274-8456
9
1.4 Installation and Operation
For information on the installation and/or operation of the DINAMAP™ Compact Vital Signs Monitor,
reference must be made to the DINAMAP™ Compact Vital Signs Monitor Operation Manual. This instrument
is to be operated and serviced by authorized personnel only, and only in accordance with the Warnings and
Precautions given in both the DINAMAP™ Compact Vital Signs Monitor Operation Manual and this
document.
1.5 Warnings & Precautions
Refer to the safety warnings and precautions detailed in the DINAMAP™ Compact Vital Signs Monitor
Operation Manual.
This manual is intended for use by authorized personnel who are familiar with digital and analog electronic
principles and who are also familiar with the operation of the DINAMAP™ Compact Vital Signs Monitor.
Disconnect all the power sources before removing covers.
1.6 Symbol Definitions
1.6.1 Safety Symbols
ATTENTION: Read accompanying
documents
Equipment complies with IEC601-1 (1988)
Type B.F.
Defibrillator protected.
1.6.2 Key and Display Symbols
SpO2
Silence Alarms
Predictive
Temperature **
NIBP
Battery Power
STAT
External Power
MAP
Beats Per Minute
* = TS & S models only
** = TS & T models only
10
*
Power Off/On
1.6.3 Connector Symbols
External Alarm Connector
External Comms Port Connector
Fuse
External Power
1.7 Electrical, Mechanical & Environmental Specifications
NIBP
CUFF PRESSURE RANGE:
Adult / Pediatric:
Neonate:
0 mmHg to 290 mmHg
0 mmHg to 140 mmHg
DEFAULT TARGET
CUFF INFLATION:
Adult / Pediatric:
Neonate:
178 ± 15 mmHg
110 ± 15 mmHg
TARGET CUFF INFLATION
ADJUSTMENT RANGE:
Adult / Pediatric:
Neonate:
100 to 250 mmHg; 5 mmHg increments
100 to 140 mmHg; 5 mmHg increments
BLOOD PRESSURE
MEASUREMENT RANGE:
Adult / Pediatric:
Neonate:
BLOOD PRESSURE ACCURACY:
Systolic
mmHg
30 - 245
40 - 140
MAP mmHg
15 - 215
30 - 115
Diastolic
mmHg
10 - 195
20 - 100
Meets AAMI/ANSI SP-10 Electronic or Automated
Sphygmomanometer Standard for Accuracy. (AAMI/ANSI
standard :± 5 mmHg mean error; 8 mmHg standard deviation;
intra-arterial method.)
BLOOD PRESSURE
DETERMINATION TIME:
20 seconds to 45 seconds typical
Adult:
120 seconds maximum,
Neonate:
85 seconds maximum.
PULSE RATE DETERMINATION:
Adult / Pediatric:
Neonate:
PULSE RATE ACCURACY:
± 3.5 percent
OVERPRESSURE CUT-OFF:
Adult / Pediatric
Neonate
30 - 200 BPM
30 - 200 BPM
300 - 330 mmHg
150 - 165 mmHg
Predictive Temperature Determination (TS & T models)
RANGE:
Max.
Min
PROBE ACCURACY:
? 0.1oC
42.2 ? Celsius
31.1? Celsius
108.0° Fahrenheit
88.0° Fahrenheit
± 0.2oF
11
PREDICTIVE TEMPERATURE
ACCURACY :
? 0.6oC, 36.1 - 39.4 oC ± 1oF, 97 - 103 oF
UNSPECIFIED < 36.1OC, > 39.4 OC
UNSPECIFIED < 97 OF, > 103 OF
DETERMINATION TIME:
30 seconds typical; 60 seconds maximum
SpO2 (TS & S Model)
OXYGEN SATURATION
RANGE:
SATURATION ACCURACY:
0 to 100%
0% to 69%: unspecified
ADULT ACCURACY (70-100%)
Nellcor Puritan Bennett™ Sensor
Accuracy
OXICLIQ-P Pediatric sensor
OXIBAND Pediatric/infant sensor
DURA-Y ear clip
OXISENSOR II D-20 Pediatric sensor
OXICLIQ-N neonatal/adult sensor
REFLECTANCE sensor
DURASENSOR adult
OXIBAND adult/neonatal sensor
DURA-Y multisite sensor
OXISENSOR R-15 adult nasal sensor
OXISENSOR II D-25 adult sensor
OXICLIQ-A adult sensor
OXISENSOR II N-25 neonatal/adult sensor
OXISENSOR II I-20 infant sensor
OXISENSOR II D-25L adult sensor, long cable
2.5 digits
3.0 digits
3.5 digits
3.5 digits
2.5 digits
3.5 digits
3.0 digits
3.0 digits
3.0 digits
3.5 digits
2.0 digits
2.5 digits
2.0 digits
2.0 digits
2.0 digits
Neonatal Accuracy:
When sensors are used on neonatal subjects as recommended, the specified accuracy range is increased
by ? 1 digit to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal
blood (e.g., N –25 accuracy on neonates is ? 3, rather than ? 2.)
Note: Refer to NELLCOR PURITAN BENNETT sensor specifications.
PULSE RATE DETERMINATION:
20 BPM - 250 BPM +/- 3 beats
SATURATION PITCH INDICATOR:
Pitch changes with saturation.
Volume selectable from 0 (off) to 9.
WAVEFORMS:
Pulse Plethysmograph waveform on LCD gain compensated
SENSOR DISCONNECT /
DISCONNECT FROM PATIENT:
The monitor will detect the attachment or disconnection
of a sensor from the patient within 5 seconds.
PULSE DETECTION:
The monitor will detect a pulse or enter a no signal state within 15
seconds of being attached to the patient.
12
LOSS OF PULSE:
The monitor will detect loss of pulse from patient and
enter a no signal state within 10 seconds.
Mechanical
DIMENSIONS:
Height:
Width:
Depth:
9.1 in (23.0 cm)
7.3 in (18.5 cm)
6.9 in (17.5 cm)
WEIGHT including battery:
8.3 lb (3.75 kg)
MOUNTINGS:
Self-supporting on rubber feet or pole mountable.
PORTABILITY:
Carried by recessed handle or pole mounted.
CLASSIFICATION INFORMATION:
• Mode of Operation - Continuous
• Degree of Protection against harmful ingress of water: (See
IPX1 definition).
Power Requirements:
POWER CONVERTER:
Protection against electrical shock - Class 1
INPUT VOLTAGE:
115 / 230 VAC, 50 / 60 Hz (nominal),
90 ~ 264 VAC, 47 ~ 63 Hz (range)
MONITOR:
Protection against electrical shock - Class 1
INPUT VOLTAGE:
24 VDC (nominal), 12-30 VDC from supplied power converter
EXTERNAL DC LINE FUSE:
T3.15A 250 VAC
BATTERY:
12 volt, 2.3 amp-hours.
Minimum operation time: 2 hours (5 minute auto cycle with adult
cuff at 25°C with power save mode enabled) from full charge.
Time for full recharge: 8 hours from full discharge
Environmental
OPERATING TEMPERATURE:
+ 5° C to + 40° C
OPERATING ATMOSPHERIC
PRESSURE RANGE:
700 to 1060 hectoPascal
STORAGE TEMPERATURE:
–20° C to + 50° C
(+ 41° F to + 104° F)
(– 4° F to + 122° F)
STORAGE / TRANSPORTATION
ATMOSPHERIC PRESSURE:
500 to 1060 hectoPascal
HUMIDITY RANGE:
0 % to 95 % non-condensing
RADIO FREQUENCY
Complies with IEC Publication 801-3 (Draft 7, second edition)
ELECTROMAGNETIC IMMUNITY:
3.0V/m, modulation 80% AM @ 1kHz, frequency range
26.0~1000.0Mhz
13
0086
IPX1
14
This product conforms with the essential requirements of the Medical Device Directive.
Accessories without the CE Mark are not guaranteed to meet the Essential
requirements of the Medical Device Directive.
The DINAMAP™ Compact monitor is protected against vertically falling drops of water
and conforms to the IEC 529 standard at level of IPX1. No harmful effects will come of
vertically falling drops of water making contact with the monitor.
2 PRODUCT DESCRIPTION
2.1 Introduction
The DINAMAP™ Compact series of monitors are portable devices, provide non-invasive determination
systolic and diastolic blood pressure, mean arterial pressure (MAP), pulse rate, predictive temperature and
blood oxygen saturation for neonatal and adult/Pediatric patients. Certain model variants are not equipped
with all functions. The monitors are mains or battery operated, and are primarily intended for use in hospital
acute care settings such as Day Surgery, Accident & Emergency, ITU, High Dependency Units, Labor and
Delivery, GI/Endoscopy, and Medical/Surgical Units.
2.2 Physical Description
Each monitor is supplied with an accessory pack. The contents of the pack vary according to model and are
listed below:
Model: DINAMAP™ Compact BP Monitor
Qty
1
1
2
1
1
1
2
1
Description
DINAMAP™ Compact Monitor with integral printer
AC-DC Power Converter
Mains power leads
Rechargeable Battery
Cuff, Standard Adult
Pneumatic Hose, 12 ft., Standard Adult
Printer paper rolls
Operating Instruction Manual
Model: DINAMAP™ Compact TS & T Monitors are supplied as above, with the addition of:
Qty
1
1
Description
Nellcor Puritan Bennett™ Finger sensor
Nellcor Puritan Bennett™ sensor extension cable
Models TS and T are supplied with the Predictive Temperature Kit. This is comprised of the following
items:
Qty
1
1
2
1
Description
Oral Temperature Probe
Probe holder accessory
Pack of 20 probe covers
Instruction sheet
15
Rear Panel Connections
2
1
6
5
T3.15A 250V
4
3
16
1
POLE CLAMP
Used to clamp the monitor to a pole or stand.
2
BATTERY COVER
Securely retains and protects the internal battery.
3
EXTERNAL POWER
Connection socket for the supplied external power
converter ONLY.
4
FUSE HOLDER
5
DATA INTERFACE
CONNECTOR
Host communications port: 15 way D-type RS232 serial
port for use with equipment conforming to IEC-601 only.
6
REMOTE ALARM
CONNECTOR
Remote Alarm port with both normally open and
normally closed isolated contacts for use with equipment
conforming to IEC-601 only. (if so equipped)
External power source line fuse holder.
Front Panel Controls & Indicators (TS Model shown)
17
15
19
10
16
20
21
18
22
11
14
13
8
12
7
9
23
27
26
24
25
7
This push button switch controls the ON/
OFF state of the monitor. Push for power
ON and again for power OFF.
18
8
This green LED indicates the external
power and battery charging status of the
monitor.
19
9
This yellow LED indicates operation and
charge status of the internal battery.
20
This yellow LED flashes, indicating
real-time pulse rate measurements
are derived from the SpO2 signals.
This is used to highlight and select items in
the LCD menus. If the monitor is OFF,
pushing it will switch the monitor ON.
21
This 4 digit red LED Display
indicates the measured
Temperature.
This panel displays all alarms, user
interface messages & configuration
options.
22
°C °F DISPLAY
These indicate that Temperature is
being displayed in degrees Celsius
or Fahrenheit.
12
Press this key to begin or halt an NIBP
determination. The key also cancels STAT
mode.
23
TEMP. PROBE
CONNECTOR
Plug the predictive temperature
probe cable into this socket.
13
Press this key to begin or halt operation of
the STAT mode of NIBP monitoring.
24
SpO2 SENSOR
CONNECTOR
Attach the SpO2 sensor extension
=
cable to this socket.
14
This switch alternately mutes & enables
the audio alarm. When pressed once
(SILENCE ON) the switch lights indicating
audible alarms have been silenced.
25
CUFF
CONNECTOR
Insert the NIBP Cuff hose into this
connector.
10
ROTOR
CONTROL
11
LCD
This 3 digit LED indicates SpO?
=
Oxygen Saturation in %.
PULSE BPM
DISPLAY
This 3-digit yellow LED display
shows pulse rate in Beats Per
Minute.
15
SYSTOLIC
PRESSURE
This 3 digit red LED display indicates the
measured systolic B.P. in mmHg.
26
LIGHT SENSOR
Automatically measures ambient
light to set LED display intensity.
16
DIASTOLIC
PRESSURE
This 3 digit red LED display indicates the
measured diastolic B.P. in mmHg.
27
PRINTER DOOR
Open this door to gain access to the
paper.
17
This red LED display indicates the
measured mean arterial B.P. in mmHg. It
also shows instantaneous cuff pressure
during an NIBP determination.
17
18
= TS & S models only
2.3 Overall Principles of Operation
This section of the manual describes the principles of operation of the DINAMAP™ Compact monitor. The
section is arranged to give an overall description of the instrument, the following sections then detail each of
the functional systems. For full assembly drawings, circuit diagrams and parts lists of the component circuit
boards, refer to section 8 of this manual. The system block diagram of the DINAMAP™ Compact monitor is
shown in drawing 8600EB.
Patient vital signs can be measured by a variety of electronic sensors, including oscillatory pressure sensors
for Non-Invasive Blood Pressure measurement. Both the DINAMAP™ Compact T and TS models are
provided with a predictive temperature parameter determination. The Compact TS model is also equipped
with a Nellcor Puritan Bennett™ Blood Oximetry module enabling oxygen saturation and heart rate to be
determined from information received via a finger sensor.
Operator access is via buttons and the rotor located on the front panel. A liquid crystal graphics display
module provides the user with menu, alarm and status information as well as a plethysmograph display or a
record of previous measurements. Parameter information is presented by multiplexed seven segment LED
displays and discrete LED indicators. A thermal printer provides hard copies of patients vital signs either
automatically or on demand. A built-in speaker and piezo-electric sounder advise the user of pulse signals
and alarm conditions.
The monitor is designed to operate from AC mains via the supplied power converter, or from an internal
rechargeable battery. The external AC power converter rectifies the mains to produce a raw DC supply
which is regulated internally to provide the DC operating power. The external raw DC source is employed to
charge the internal batteries. The real time clock and user monitor settings are maintained by an internal
rechargeable NiCd backup battery.
The monitor includes a pneumatics system required for NIBP operation. NIBP determinations are made by
pumping up the air pressure in a restrictive cuff and monitoring the oscillatory signals in the system as the
pressure is released. Large deflation steps are employed for NIBP and the system processor interpolates
measurements between steps. The pneumatic system then regulates operation of the pneumatic pump and
valves. An overpressure sensor provides independent protection against over-inflation. Protection against
deflation failure is inherent in the design; the valve and pneumatic system being designed to auto-deflate
the system in the event of power fail or alarm situations.
Blood Oximetry (SpO2) determinations (model TS only) are made using a proprietary module from Nellcor
Puritan Bennett™ , which communicates the parameter measurements to the system processor via an
internal serial port.
Predictive temperature measurements are based on a 3 or 12-minute mercury in glass reference. Predictive
Temperature probes can be connected to the monitor to provide an analog signal representing the
measured temperature. An A/D converter then digitizes this analog signal; this information is then
interpreted by the system processor and displayed for viewing.
2.4 Functional Description
The functions of the DINAMAP™ Compact monitor can be separated into eleven discrete parts. The
following paragraphs describe each of the Monitors, with full circuit diagrams provided in section 8.
2.4.1 System Processor
The system processor is based on 68302 microprocessor IC22 on the Main Board. A master 19.6608 MHz
clock generated by XL1 exits the system processor at pin 98 and is divided by 16 inside binary ripple
19
counter IC36 to provide the 1.23 MHz ADC clock. Communication between microprocessor IC22 and all bus
devices is provided by a 24-bit address bus, 8-bits data bus and the three wire serial peripheral interface
(SPI) bus. The system processor is equipped with a watchdog timer and reset circuit.
The system program is stored in a 4 Mbit FLASH memory IC28 as 512k by 8 bit words. Program memory is
provided by a low power 1 Mbit SRAM IC32 formatted as 128k by 8 bit words. Non-volatile settings are
stored in an EEPROM IC26 which talks to a dedicated serial port on the system processor. All bus devices
are powered from the +5VD digital supply, except for the RAM that is also powered from the VRAM supply
when the system is in “standby”mode.
Should the software detect a system fault or fail to reset its watchdog timer, the system processor will output
a logic low level on its watchdog output WDOG. This is latched as FAILSAFE by the output of the bistable
formed by IC31, allowing the output of the relaxation oscillator IC1 to pass through IC4 to sound the piezo
sounder LS1. The FAILSAFE signal is also used to disable the isolated power supply module PM100.
The power supply board battery backup voltage BATT_BACKUP, derived from either NiCd battery or an
external supply, appears on the supervisory circuit IC35. This supervisory circuit has two functions. First, it
generates Reset and Halt signals for the system processor when the 5 volt power supply powers on and off
respectively. Secondly, it controls transistor TR8, allowing the 5 volt line to power the RAM chip IC32 and
PIC when the system is “on”, or the BATT_BACKUP to power the RAM and the PIC when the system is
“off”. This allows values to be saved in the otherwise volatile RAM and to maintain the low power standby
mode of the PIC.
Additional control signals from the system processor are decoded by IC16, which generates seven chip
select lines and IC17 which generates latch outputs for the LED display drivers. The data bus D0 ~ D7 is
filtered by inductors L12~19 to produce a buffered data bus DB0 ~ DB7 for the display drivers.
There is also a PIC micro-controller IC5 on the power supply board 8620AB, which has its own independent
watchdog, reset circuit and provides the real-time clock function. The PIC provides control of PSU status,
battery charging, printer and alarm/annunciator audio output. Communications between the system
processor and the PIC are continually monitored in order to provide additional fail-safe security for the
monitor. All measurement, alarm and communications ports of the instrument are isolated to IEC601.
2.4.2 Temperature System
The temperature system for the Model DINAMAP™ Compact Monitor for reading predictive thermistor probes
is located on the Main Board.
If a thermistor probe is used, the resistance of the probe varies in relation to temperature. The probe
resistance is effectively in parallel with R63 and therefore a change in temperature results in a change in
voltage applied to input selector IC19. The selected input voltage from IC19 is connected to the voltage to
frequency converter (VFC) IC18. The frequency of the square wave output from the VFC varies
proportionately to the input voltage. The output signal is isolated through opto-coupler PC2 and enters the
system processor, as PRED_TEMP_F. The processor receives the square wave signal and determines the
temperature by counting the number of edges received per unit interval.
Through the application of a software algorithm, the system processor provides two select lines TEMP_SEL0
and TEMP_SEL1. These are isolated by opto-isolators PC3 and PC4 and used to control the input selector
IC19, allowing the appropriate resistor tree to be selected. There are four selectable channels, two providing
calibrations points, a third used for probe type detection and the fourth for the thermistor measurement.
The probe type is determined by measurement of the frequency produced by the VFC from a DC input
derived from resistor tree R59-R61; for an oral probe this produces a frequency equivalent to 42 degrees
Celsius, Rectal probes short out R61 and hence change the frequency slightly. The software is capable of
determining which probe type is attached and hence selects the appropriate measurement mode.
20
Compensation for drift in accuracy of the system is achieved by the unit recalibrating itself on switch on and
after each measurement and at ten-minute intervals. Calibration is achieved by measuring two points on the
high precision resistor chain of RV3, R70, and R74-77 which equate to 10°C and 37°C. Any changes in
VREFI or voltage to frequency transfer characteristic can therefore be accommodated within reason.
The temperature circuit is powered from an isolated 5-volt and ? 9 volt supply (5 and ? 15V on revision 04
and earlier), these supplies are generated by the isolated power supply module PM100. Calibration of the
voltage and hence frequency in the temperature reference circuit is achieved by the use of resistor trees,
attached to a stable reference voltage. This reference, VREFI is 3.24 volts (nominal) and is generated from
the isolated 5-volt line by D2, adjustment provided by RV2. Thus RV2 will adjust the overall calibration of
voltage ranges X0, X1 and X2. Additional adjustment of the calibration points X0 and X1 is provided by RV3
in their resistor tree. Together RV2 and RV3 are used to calibrate the temperature circuit.
The VREFI may be measured across TP3 and TP2, and the PRED_TEMP_F signal may be measured
across TP9 and TP6.
2.4.3 NIBP System
The non-invasive blood pressure system measures blood pressure and pulse rate. The NIBP measurement
circuitry is located on the Main Board.
To measure blood pressure, a reference voltage generated by IC2 supplies a bridge pressure transducer
ML1 to enable it to convert air pressure from the pneumatics system into an analog voltage. The
instrumentation amplifier IC5 amplifies this voltage. The unfiltered signal PT1 is routed to into an eight
channel analog multiplexer, IC20. When selected, the output signal is buffered by IC21 as signal
BUF_ANA_MUX and enters the ADC IC23. The digitized output of the ADC is presented to IC24, an octal
buffer/line driver, from where it is transmitted to the data bus and the system processor. Pressure
transducer offset PT1_REF from the output of IC2 also enters the analog multiplexer IC20. When selected,
it provides an off-set value to the system processor. The processor then measures the offset value and
subtracts the offset from PT1 measurements. The control signals for the multiplexer (ASEL0 ~ ASEL2) are
generated by the system processor.
The pulse rate component (typically one percent) of the cuff pressure signal is filtered and separated from
the PT1 pressure signal. The low pass filter formed around IC9a allows the low frequency pulse signal
through whilst attenuating any higher frequency noise. The signal is then ac coupled by the high pass filter
of C20 & R41 in order to strip off the small pulse rate signal from the cuff pressure signal. IC9b then
amplifies the pulse signal before being applied to high pass filter IC12. The output from IC12 FPT1 is routed
to the analog multiplexer IC20 for selection by the system processor. The system processor controls the
analog switch IC7, it is used to clamp and apply offset signals into the filtered pressure measurement
channel.
The system software includes an internal on-line self-test feature for the FPT channel at power-up and after
each non-STAT mode determination.
2.4.4 Pneumatics System
The pneumatics system provides air pressure for the NIBP cuff, manifolding for control of cuff pressure, and
an overpressure signal to the system processor and pneumatic control logic GAL if the pressure exceeds
300 mmHg in adult mode or 158 mmHg in neonate mode. The system is comprised of an air pump, pressure
transducer, overpressure detection circuitry, power control circuit, and two solenoid-operated pneumatic
valves. Air from the pump is routed through the dump valve to the cuff output manifold.
The pneumatics system provides control of the pump by commands received from the system processor.
The PUMP_ON signal from latch IC3 on the Main Board is passed to the pneumatic control logic GAL IC1 on
the power supply board. The GAL produces a pump drive signal which turns on FET TR4b, TR13 and TR8,
pulling down the ground return side of the pump motor through current sensing resistor R31. Both the
21
Deflate and Dump valve control signals are also generated by latch IC3 (Main Board) and passed to the GAL
IC1 (power supply board). Pump motor current is sensed in R31 and amplified by IC4. This is routed to the
system processor as PUMPC and also compared to a reference to produce an over current input signal to
the GAL. If pin 9 of GAL IC1 goes high, pin 12 (signal LATCHED_OVC) will be latched high and the pump
enable output will go high switching off the pump. The latched over current condition can be reset by the
Main Board processor asserting PNEURESET high.
The Deflate valve vents the system to atmosphere in its energized state. When the pump is operating, the
deflate valve is de-energized by the GAL IC1 turning off FET TR1b in the ground return side of the valve
circuit. If a failsafe, overpressure or deflate condition signal arrives on the GAL, the valve will be energized
and the system will rapidly deflate. The over-pressure signal is produced by the pump pressure transducer
and associated electronics, while the failsafe can be derived from the system processor’s watchdog output
as well as from the PIC IC5.
Conversely, the Dump valve vents the system to atmosphere in its de-energized state. When the pump is
operating during a determination cycle, the Dump valve remains energized, the GAL receiving a high level
DUMP signal from the latch IC3 on the Main Board. When the determination cycle is complete, or if an alarm
condition arises, the DUMP signal from latch IC3 will go low causing the GAL to turn off FET TR4a and deenergize the dump valve, opening the system to atmosphere.
Because the action of deflation can be initiated by energizing one valve or de-energizing another, a power
failure during a determination will always result in de-pressurization of the system.
The action of operating the Dump and Deflate valves produces a voltage across R20/21 and R66. This
signal is labeled VALVESENSE and is routed to the multiplexer IC20 on the Main Board from where it can
be selected by the system processor. The four possible conditions of the Dump and Deflate valves can be
ascertained from the VALVESENSE signal.
Once the system processor has determined which cuff type is being used, it instructs latch IC3 to output an
ADULT signal to TR1. In the pump pressure transducer circuit on the Main Board, IC8 generates a high
precision 5 volt reference, which is made available to dual op-amps IC10 and IC21. This reference is
monitored by the system processor as PT2_REF. The buffered reference from IC21 is applied to excite the
pump output pressure transducer ML2 on the Main Board (Transducer mounted on Pump assembly on
revision 04 and earlier). The output of the transducer varies in proportion to the air pressure in the system.
The transducer ML2 output pins connect to IC11, an instrumentation op-amp in differential mode
configuration. An offset voltage derived from the voltage reference IC8 is buffered by IC10b and fed to IC11
it is then summed with the amplified difference signal from the transducer. IC11’s output is passed to the
analog multiplexer IC20 as PT2 and the inverting input of comparator IC15a where it is compared to the
over pressure limit.
Note: For revision 04 units and earlier the reference to IC25 on the Main Board in the following paragraphs
relates to IC10.
The system processor produces control signals POT_INC, POT_U/D and POT_CS to drive a digital
potentiometer IC13 which sets the gain of amplifier IC25a and hence the over pressure limit. The “high”end
of the digital pot is connected to the output of the output of IC25a while the “low”end of the pot connects to
the precision resistor chain formed by R47, 48, & 50. The “wiper”of the pot connects to the inverting input of
IC25a. During factory set-up, the over pressure alignment is trimmed by selecting IC13 and incrementing its
output up or down. After alignment is complete, jumper LK1 is removed, thus preventing chip select of the
digital pot.
The output of IC25b applies an offset to the Adult/Neonate threshold input of IC15a. The output of IC15a,
labeled UNFIL_OVP, is normally at a high logic level but changes to low when overpressure occurs. This
signal is available as an input to the PSU board as unfiltered overpressure (UNFIL_OVP). UNFIL_OVP is
then filtered by R49 and C27 and fed in to comparator IC15b to produce filtered overpressure signal
FLT_OVP and is made available to the PSU board. The overpressure threshold voltage reference is
22
buffered by IC12 to produce TH_VREF which can be measured by the system processor via multiplexer
IC20 and ADC IC23.
Upon clearance of the overpressure condition the Main Board system processor can reset the overpressure
latch in GAL IC1 on the PSU Board, by asserting PNEU_RESET signal high, in order that another
determination can be carried out.
2.4.5 SpO2 System (TS & S models only)
Blood oxygen saturation and pulse determinations can be made (DINAMAP™ Compact TS and S only)
using a Nellcor Puritan Bennett™ finger sensor. This connects to the SpO2 module fitted inside a shielded
case mounted in the monitor. This module is supplied as a complete replaceable part, no service being
possible.
The SpO2 module communicates with the system processor via an isolated two wire serial interface. Data
signals from the system processor (SPO2_TX) are isolated by opto-coupler PC6. From here they enter
connector PL13 to the SpO2 module. Likewise, data transmitted from the SpO2 module are isolated by optocoupler PC5, from where it is routed to the system processor as signal SPO2_RXD.
The SpO2 module (and the temperature determination circuit) is powered via an isolated power supply
module PM100. This is powered from the +12VV supply and generates the isolated +5VI, +9VI and -9VI
supplies (+5VI, +15VI and -15VI supplies on revision 04 units and earlier). On revision 06 units the isolated
power supply module is switched on by the system processor SPO2_PSU_ON signal, turning TR11 off and
setting the control input to ML3 high. On revision 04 and earlier units the isolated power supply module ML3
is turned on by grounding its -ve supply connection through FET TR12. In the event of a FAILSAFE
condition arising, the control signal of ML3 is set low by TR10 switching on (06 unit), or by switching TR12
off (on revision 04 units and earlier).
2.4.6 Power Supply
The power supplies are generated and controlled on the power supply board.
The power supply is designed to operate from both an external DC line source and from a 12-volt
rechargeable lead-acid battery. The supplied mains power converter provides 24 volt DC output. When both
supplies are present, the power supply will operate from the line supply if it is greater than 12 volts. If the
line supply is greater than 16 volts the battery will be charged both when the monitor is switched on and off.
The power supply sequencing functions are controlled by the PIC IC5.
Incoming line power from the external power converter has any high voltage spikes snubbed by transient
suppresser D4 and over-voltage protection is provided by D6, R10 and CSR1. Should the line input voltage
rise above 34 volts, zener diode D6 conducts, turning on thyristor CSR1. This causes a short circuit which
blows the external DC line fuse. Reverse polarity protection for the power supply is provided by blocking
diodes D8 and D29. A sample of the line input voltage is taken by R3 and R6 (DC_INPUT), and routed to
the external DC detection circuit IC16 and the PIC.
The switcher IC3, inductor L1 and diode D10 form a flyback converter with a nominal output of 15.1 volts
and VRAW of 14.8 volts through the battery blocking diode D11. A sample of the smoothed output is taken
by R29 & R30 to provide a feedback voltage.
The PIC uses the DC_BATTERY, DC_INPUT and EXT_DC_ON signals to determine the available power
sources. If valid power supply conditions exist and a battery is detected, the PIC will turn on TR7, which
turns on p-channel FET TR6, allowing the +15VI line to charge the battery via D9, R17 and FS1. When
charging is taking place, a voltage will appear across R17, which will be proportional to the charge current.
This voltage is sampled by R14 & R15 to produce the BATTERY_C signal.
23
If there is no valid external DC line supply applied to the monitor, the monitor will default to use the battery.
DC from the rechargeable battery arrives on thermal fuse FS1, from where it is routed to the source of TR2,
a p-channel FET under the control of TR3. The battery supplies power to VRAW via transistor TR2 and
diode D28. When operating from a battery, VRAW will be in the range 10.4 - 13.5 V depending on the
battery charge. A sample of the battery’s terminal voltage is taken by R1 & R2 and applied to the ADC input
of the PIC (DC_BATTERY). From this, the PIC can determine the charge state of the battery.
Diodes D24 and D25 provide a power supply line VDC_OP which is present whenever a supply is available.
This supply is used exclusively for IC16, a programmable voltage regulator with internal comparator. IC16
performs two functions. First, its internal comparator compares the sampled line input voltage DC_INPUT to
an internal reference level. When the DC_INPUT is below approximately 10.4 volts, IC16 turns off TR14 to
indicate to the PIC that there is no valid external DC input. The second function of IC16 is to generate a
precision voltage supply from VDC_OP, set by R5 & R9 to 4.5 volts. This voltage is connected to the
BATT_BACKUP line by D13, D23 preventing the voltage from entering the NiCd backup battery. This
battery, B1, has a nominal terminal voltage of 3.6 volts and is trickle charged form the +15.1 volt supply
when the system is powered.
The VRAW supply, which is nominally 14.8 volts when operating from an external line voltage or 10.4 to
13.6 volts when operating from battery, is used to generate the other power supply lines. The monitor’s
power on switch is routed directly to the PIC, which controls the set up sequencing of the power rails. The
PIC switches on the +5VD line then the +VDISP line, the 12-volt supplies are controlled by the system
processor which turns them on after the +5VD supply is stable.
VRAW enters IC7, a step down DC-to-DC converter, which is used to produce the +5VD digital line. The 5
volt logic supply IC7 is controlled by the PSU_INH signal from the PIC. VRAW also connects to IC17 to
produce the +4.5 volt LED display supply +VDISP, which is controlled by the PIC signal ENABLE_VDISP.
The 12 volt line and other supplies are controlled by the 12V_ON signal from the system processor on the
Main Board. When this pin is at a high logic level, TR9 is turned on, as is p-channel FET TR10. This causes
the VRAW supply to enter IC8, a boost-switching regulator. R41 & R42 set the output voltage of the flyback
converter to 14.5 volts, which is further regulated by linear regulator IC9 to produce the pump and valve
supply +12VV line and linear regulator IC18 to produce the analog supply +12VR line. The output of TR10
also feeds the inverting switching regulator IC10, which produces a -14 volt supply to linear regulator IC19
that in turn produces the analog supply -12VR line. A sample of the +12VV line is taken by R44 & R45,
producing the PUMP_V signal, while a sample is taken across +12VR and -12VR lines by R47 & R48
producing the ANALOG_V signal. These are measured by ADCs inside the PIC to confirm the operation of
the power supplies.
2.4.7 Audio Amplifier
User warning sounds are generated by the power supply board PIC as a pulse width modulated digital
signal AUDIO_FREQ. This signal is fed to all the inputs of the quad bilateral switch IC14, their outputs being
connected together via a resistor ladder to produce a switched gain-summing amplifier IC6. The gain
elements are operated by the PIC signal A_GAIN0-3, together providing a 4-bit (16 level) volume control.
The output of the summing amplifier IC6 also applies band-pass filtering to the audio signal. The circuit
around IC6b sets the upper breakpoint of the filter, while the circuit on the input to the final amplifier IC15
sets the lower breakpoint and provides amplification for the speaker.
2.4.8 Alarm and Communication Interfaces
The remote alarm signal originates from the system processor on the Main Board, from where it is logic
ORed with the FAILSAFE signal, inverted and sent as the REMOTE_ALARM signal to the power supply
board. The signal controls n-channel FET TR11 to drive the coil of the relay RL1/2. The switch side of the
relay contains both normally open and normally closed contacts, which are separated from the monitor
electronics via an isolation barrier. A dual pole dual throw relay is used, but its switch elements are
connected in parallel to provide a single pole dual throw action with increased current capacity.
24
Communications signals use the RS232 bi-directional serial format. Digital TTL level signals from the
system processor are converted to RS232 level signals on the power supply board by IC13, which
generates the necessary positive and negative voltage swings. In addition, inverted TTL data transmit and
receive lines are provided. A 5-volt supply is also made available to the serial port, this being protected by a
0.1 amp thermal resetable fuse.
2.4.9 User Controls
The user controls are comprised of four push buttons; power, stat, NIBP start/stop and alarm silence, and
the rotor control. When the rotor control is turned, it operates an optical encoder producing a Gray code
output. Pressing the rotor operates a push button switch. These rotor control signals and the four push
buttons are encoded together on the switch board by a register latch IC2. The register is periodically polled
by the system processor to determine if any of the switches have been operated. IC2 is controlled by the
chip select signal KEY_CS and the read signal LD_RD. KEY_CS is produced by the Main Board chip select
decoder IC16 under control of the system processor, while LD_RD comes from the processor.
The ambient light detection circuit comprises of the light dependant resistor LDR1 and a bias resistor in a
potential divider configuration. A change in the level of incident light on the LDR1 causes its resistance to
alters, changing the voltage applied to the PIC ADC which determines the ambient light level. The system
processor regularly requests the ambient light level measurements from the PIC, and in turn controls the
brightness of the LED displays on the display board.
2.4.10 Displays
All displays, except for line power indicator, are multiplexed, and are split into three separate control circuits,
each with its own display driver. The power / charging green LED is controlled by the PIC software.
The three LED display drivers (IC1, 2 & 3 on the display board) are Maxim MAX7219’s, which are controlled
by the system processor via the SPI bus. The LED parameter displays are divided into functional groups on
the display board. IC1 drives the systolic and diastolic displays; IC2, the Temperature and MAP displays;
IC3, the Heart Rate and SpO2 displays as well as the battery, °C, °F and Heart Rate discrete indicators. A
current limiting resistor attached to iSET input of each driver IC sets the maximum segment current in the
display; each current limiting resistor is different because each driver IC controls different sized LEDs. IC3
drives red and amber seven segment displays as well as red and amber discrete displays, additional low
value resistors or links R1 to R11 are inserted in series with the anode of each diode to provide different
LED currents within the same multiplex group.
The display drivers are all controlled by the system processor via the SPI bus data signal SPTXD. This
transmits the display data and controls the brightness of the displays by varying the multiplexer’s PWM
ratio. Data is latched into each driver ICs when the respective load signal LED_LOAD0-2 is pulled high by
decoder IC17 on the Main Board. This is synchronized with the LED SPI clock signal LED_CLK from the
system processor.
Text and graphical data is displayed on the LCD graphics module, which contains its own on-board driver
electronics. The module is controlled by the system processor via filtered data bus DB0 ~ DB7, address
signal A4, RESET and the LCD_CS, LD_WE and LD_RD lines. The LCD_CS line is produced by the Main
Board chip select decoder IC16, while the LD_WE and LD_RD lines are derived from the system
processor’s R/W and CS and logic ICs IC1 & IC27.
Backlight illumination of the LCD is provided, the green LED source being under software control via the
system processor. The BACKLIGHT signal is decoded on the Main Board and used to turn on an N-channel
FET TR1 on the display board. A resistor sets the current and hence the constant backlight brightness.
The LCD module requires the +5VD and -VLCD supplies to function. The -VLCD supply is derived from the
-12VR supply and can be trimmed by potentiometer RV1 on the switchboard. This adjusts the effective
25
viewing angle or contrast of the LCD display. The trimmed voltage is nominally set to -9.6V and may be
measured across TP1 and TP2 of the switchboard.
2.4.11 Printer
The monitor employs an Axiohm HPT4050 thermal printer module, fitted to the printer board, which contains
the interface electronics. The PIC provides stepper motor control signals M_OE, M_PHASE1 and
M_PHASE2 to IC3 on the printer board. This converts these to sequential drive signals MP1_1, MP1_2,
MP2_1 and MP2_2 for the stepper motor.
The PIC also sends a PRINTER_TRIGGER signal to initiate printer action; this triggers a timer IC4 which
generates the printer enable signal PRINTER_OE. To optimize print speed, print quality and the life of the
print head, a thermistor constantly measures the temperature of the head. As the system heats up, the print
head thermistor is used by the timer circuit to reduce the width of the PRINTER_OE pulse. The print head
voltage +VP is also employed to alter the width of the PRINTER_OE pulse. A typical print trigger pulse
width is 6 ms. An opto-transistor detects the presence of printer paper and a micro switch detects the status
of the printer door, open or closed. These two devices are combined to produce the PRINTER_READY
signal, which inhibits the printer action when clear and is sent back to the PIC as PRINTER_BUSY. Data for
printing is sent from the system processor to the printer board via the SPI bus.
Because the instantaneous power demand of the printer is quite high, a dedicated 6.1-volt supply +VP is
generated from VRAW by IC1 and extra reservoir capacity is provided by C16 and C17.
26
3 Maintenance
This section of the manual contains four parts:
??
??
??
??
The first covers general maintenance functions, which can be performed without
disassembly of the monitor.
The second covers use of the built-in Service Software.
The third describes how to calibrate the NIBP and overpressure detection of the
monitor, or check their calibration.
The fourth describes maintenance, which requires removal of the monitor’s covers,
disassembly of the components and internal alignments.
Procedures that pertain to SpO2, the printer and temperature components apply only to suitably equipped
DINAMAP™ Compact Vital Signs Monitors.
3.1 General Maintenance
3.1.1 Replacing Fuses
The Monitor contains three fuses:
??
??
??
Replaceable fuse protecting the external line input
Thermal auto resetting fuse protecting the battery input
Thermal auto resetting fuse protecting the 5V output on the Host Communications Port
UK models will also be protected by a fuse fitted in the mains plug of the power converter input cable.
To replace the External Line Fuse, first remove the line power cord from the rear of the Monitor, if
connected. Insert a flat bladed screwdriver or small coin into the slot in the fuse holder cap and carefully
rotate anti-clockwise several revolutions. The fuse-holder cap and fuse will be released.
Pull the fuse from the cap and inspect it for a burned or broken filament. If the filament appears to be intact,
check the fuse for continuity with an ohmmeter. If damaged, replace with a fuse of the specified rating and
type.
To reassemble, push the fuse into the cap and insert both into the holder. With the flat bladed screwdriver or
small coin, rotate the cap clockwise until it is flush with the holder. Do not over tighten. Apply the external
line power and confirm that the monitor operates correctly.
3.1.2 Periodic Maintenance
Periodic maintenance tasks include cleaning the Monitor, checking pressure calibration, checking for
pneumatic leaks, checking the overpressure cut-out, and checking the temperature calibration where fitted.
Cleaning the Monitor
The exterior of the Monitor may be wiped clean with a cloth slightly dampened with mild detergents or
normal hospital bactericides.
??
??
Do not immerse unit.
Do not clean with isopropyl alcohol or other solvents.
27
??
??
??
??
The adult DURA-CUF™ range supplied for use with this Monitor may be cleaned by hand washing
in warm soapy water. Care should be exercised, however, to ensure that no water enters the cuff
or cuff hoses at any time. In the event water accidentally enters the cuff, passing air through the
cuff will dry it.
The neonatal cuffs are supplied for single use and should be discarded if they become soiled.
Do not immerse hoses.
Do not immerse cuffs without prior application of cuff hose caps.
CAUTION
??
??
??
??
Moisture or foreign substances introduced into the pneumatic system can cause damage to the
unit. Calibration equipment should always be kept dry and free of particulate matter.
The probe holder may be removed and cleaned with a mild detergent. The probe holder may be
immersed during cleaning.
Do not immerse temperature calibration plug.
The probe may be cleaned with an alcohol solution. Use a cloth or sponge, just damp, not wet,
and avoid getting any liquid into the interior of the probe.
Checking of Calibrations
Refer to section 3.3 for instruction on the procedures required to check each of the parameter calibrations.
3.1.3 Care of the Storage Battery
It is best to keep the battery charged as fully as is practical and never store the monitor with the battery in a
discharged condition. When the battery will no longer hold a charge, remove and replace with one of the
same part number.
To charge the battery, connect the supplied power converter to the monitor's rear panel, then plug the
converter’s mains lead into an appropriate AC receptacle. The battery will charge regardless of the position
of any other monitor switches.
A battery that is completely discharged can be fully recharged by the monitor within eight hours. Charging
will take place with the monitor switched on or off, as long as it remains connected to an external AC power
source via the supplied power converter.
The expected lifetime of the battery largely depends on the way in which the instrument is used. If the
battery is allowed to completely discharge before being fully recharged, the battery should survive around
two hundred recharge cycles. If the battery is used in such a way that it never becomes more than one third
discharged and is fully recharged whenever possible, it can survive up to twelve hundred cycles. This
means that by thoughtful usage, the lifetime of the battery can be extended up to six times.
PRECAUTIONS
??
To ensure that the battery will be ready for portable operation, keep the unit connected to a mains
supply whenever possible.
??
Repeated failure to fully charge the battery will result in a significant reduction in battery life.
Avoid storing batteries at temperatures above 77° F (25° C). High storage temperatures can dramatically
increase the rate of self-discharge of the battery.
28
3.2 Service Modes
There are three different service modes, each offering different levels of access to the monitor’s set-up and
calibration. These modes can be accessed using the rotor control and display panel. The modes are:
1. The “Clinician Menu”mode, allowing:
? ? setting of temperature units,
? ? checking NIBP and temperature calibration dates,
? ? setting the alarms into “Permanent Silence”mode.
(This mode is described in the DINAMAP™ Compact Monitor Operating Manual).
2. The “Main Service Mode”provides the features allowed from the clinician menu (described
above) plus the following:
? ? Checking NIBP calibration,
? ? NIBP re-calibration,
? ? Pop Off (overpressure detection) calibration,
? ? Changing temperature algorithm,
? ? Set temperature calibration date,
? ? Configuration for use with Central Station,
? ? Changing language of operation,
? ? Monitor “Set-up”- specifying whether SpO2, printer, and temperature are fitted.
3. The “Super Service Mode”is used to allow the calibration of the Predictive Temperature circuitry and
Neonatal Overpressure checking procedure. Other menu options within this mode are for
manufacturing test purposes only.
Note : Certain menu functions may not be present or appearance may be slightly different than depicted in
the following section if unit is fitted with software earlier than 8613-V4.1.
To enter the service mode from the Main Menu, use the rotor control and LCD.
First select More…
then Service
29
then sequentially enter the access code for the required Service mode.
The code choices are:
1 2 3 4 ............. for access to the Clinician Menu.
2 2 1 3 ............. for access to the Main Service Menu.
30
3.2.1 The Clinician Menu (Service mode 1 2 3 4)
3.2.1.1
Press
Sets default target pressure for first NIBP cycle.
3.2.1.2
Temp
Selection of this button allows the temperature units of the monitor to be toggled between degrees Celsius
and degrees Fahrenheit.
When Celsius is selected, the °C indicator lights. When Fahrenheit is selected, the °F indicator lights. Click
on OK to clear and save the selection. Note that changing temperature units will clear the trend readings.
The user is requested to confirm or cancel this.
3.2.1.3
Info
Selection of this button causes the most recent calibration dates of the NIBP and Temperature (not
displayed) systems to be displayed. Click on OK to clear the menu.
31
3.2.1.4
Silence
Selection of this button will cause all alarms other than failsafe to be muted until either the monitor is
powered off and on again, or the Alarm Silence button is pressed.
A confirmation menu will appear on the display. Selection of either Yes or No will exit the menu.
If silence is confirmed, the monitor’s Alarm Silence button illuminates.
3.2.1.5
Main
Selection of this button exits the user service menu and returns to the Main Menu.
3.2.2 The Service Menu (Service mode 2 2 1 3)
3.2.2.1
NIBP
Selection of this button causes the NIBP service menu to appear. This menu is used for calibrating, and
checking the calibration of, both the NIBP transducers and the over pressure (Pop Off) circuitry.
Check
Selection of this button allows the NIBP calibration of the monitor to be checked. In this mode, the monitor
functions as a digital pressure gauge, the systolic and diastolic displays showing the current pressure as
measured by PT1 (main pressure transducer) and PT2 (pressure transducer on pneumatic assembly)
respectively. With no hose attached, both displays should indicate “000”. See section 3.3.1 for a more
detailed description of using this mode.
32
Cal
Selection of this button initiates the calibration procedure. See section 3.4.1 for a more detailed description
of using this mode.
Pop Off
Selection of this button initiates the Pop Off test. See section 3.4.3 for a more detailed description of using
this mode.
OK
Selection of this button returns the display to the Service Menu.
3.2.2.2
Temp
Selection of this button causes the Temperature service menu to appear.
C or F
Selection of this button allows the temperature measurement units to be toggled between Celsius and
Fahrenheit. A warning will appear to inform the user that this change will result in the loss of any existing
trend information.
Cal
Selection of this button causes the current date to be saved for the Info display. This should only be done
when the temperature calibration has been set as described in section 3.3.6.
3 Mins
Selection of this button sets the temperature mode to “3 minute predictive”i.e. the results obtained are
those which would be obtained by a mercury-in-glass thermometer after 3 minutes.
12 Mins
Selection of this button sets the temperature mode to “12 minute predictive”i.e. the results obtained are
those which would be obtained by a mercury-in-glass thermometer after 12 minutes.
OK
Selection of this button returns the display to the Service Menu.
3.2.2.3
Info
Selection of this button displays the last calibration date information for both the NIBP and Temperature
systems.
Selection of OK returns the display to the Service Menu.
33
3.2.2.4
System
Selection of this button causes the System service menu to appear.
Error Log
When selected will produce a printout of the system errors that have occurred on unit. The Main Board and
PSU board software revision is printed out.
Language
Selection of this button allows the display language to be pre-set.
Selection of any language button removes all other language buttons from the screen, indicating that the
remaining button is the chosen language.
Selection of Clear will restore all language buttons, allowing the user to select again.
Selection of OK saves the setting to the chosen language. If no language was selected when OK was
pressed the language will be stored as undefined. The monitor will then prompt the user to select a
language on every power up until a successful selection has occurred.
Comms
Selection of this button causes the Communications menu to appear.
Selection of the Remote Operation check box toggles remote operation on and off. When on, a tick
appears next to the box, denoting that remote operation (i.e. from the central station) is possible.
Selection of the Address box allows the monitor’s address to be set within the range 1 to 255.
Selection of OK returns to the Service Menu.
System
Selection of this button causes the hardware System menu to appear.
34
Selection of any of the check boxes causes that monitor feature to be toggled on or off. When on, a tick
appears next to the box, denoting that the option is available. Monitors without certain features are supplied
without the necessary hardware for that feature and thus will have that option unselected. Selection of OK
saves changes and returns to the Service menu.
OK
Selection of this button returns the display to the Service Menu.
3.2.2.5
Silence
Selection of this button will cause all patient alarms to be muted until either the monitor is powered off and
on again, or the Alarm Silence button is pressed.
A confirmation menu will appear on the display. Selection of either Yes or No will exit the menu.
If silence is confirmed, the monitor’s Alarm Silence button illuminates.
35
3.3 Calibration - Checking & Re-calibrating
This section details the following procedures:
??
??
??
??
??
??
Checking NIBP calibration
Recalibrating NIBP
Checking NIBP overpressure detection (Pop Off)
Re-calibrating NIBP overpressure detection
Predictive temperature calibration checking
Re-calibration of predictive temperature
3.3.1 Checking of NIBP Calibration
The following section describes how to verify the NIBP calibration on a DINAMAP™ Compact Vital Signs
Monitor. Refer also to the next section for details on re-calibration of a DINAMAP™ Compact.
Checking of NIBP calibration is possible without disassembling the DINAMAP™ Compact.
Set up the DINAMAP™ Compact and calibration equipment as shown on page 38.
Switch on the DINAMAP™ Compact and allow 10 seconds to settle.
Verify that the calibration pressure measurement system reads 0 mmHg (0 mBar)
If the DINAMAP™ Compact displays the language choice menu, choose “English”, then “OK”.
Reboot the DINAMAP™ Compact by turning monitor off, and then back on.
V. Select the “Mode”button from the main menu. The Mode menu will be displayed
VI. Select the “Service”button from the mode menu
VII. Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
VIII. Select “NIBP”from the service menu. The NIBP Service menu will be displayed.
IX. Select “Check”from the NIBP service menu.
X. The DINAMAP™ Compact now displays pressures on the systolic and diastolic displays.
XI. Check calibration of the DINAMAP™ Compact unit is within specification at 0mmHg and
200mmHg (267mBar) applied pressure using a reference pressure measurement system.
I.
II.
III.
IV.
Display
Systolic
Diastolic
†
Meaning
Pressure measured at PT1
(main pressure transducer, on
Main Board)
Pressure measured at PT2 (on
pump assembly)
Units
mmHg
mmHg
System Range
0-300
0-330†
Pop Off will occur
before 330
†
Normal operation of the DINAMAP™ Compact is not possible until powered off and back on.
36
DINAMAP™ COMPACT Calibration check or re-calibration
3.3.2 NIBP Calibration
The following section describes how to perform NIBP re-calibration on a DINAMAP™ Compact Vital Signs
Monitor. Refer also to the previous section for details on checking of the DINAMAP™ Compact calibration.
NIBP calibration is possible without disassembling the DINAMAP™ Compact.
I. Set up the DINAMAP™ Compact and calibration equipment as shown in Section 3.3.2.
II. Switch on the DINAMAP™ Compact and allow 10 seconds to settle.
III. Verify that the calibration pressure measurement system reads 0 mmHg (0 mBar)
™
IV. If the DINAMAP Compact displays the language choice menu, choose “English”, then “OK”.
™
Reboot the DINAMAP Compact by turning monitor off, and then back on.
V. Select the “Mode”button from the main menu. The Mode menu will be displayed
VI. Select the “Service”button from the mode menu
VII. Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
VIII. Select “NIBP”from the service menu. The NIBP Service menu will be displayed.
IX. Select “Cal”from the NIBP service menu.
-- Note -From this point, the timing is critical.
It is recommended to read through
the following steps before proceeding.
37
The DINAMAP™ Compact requests that the pressure be set to 0 mmHg. No action should be
necessary other than to verify that the calibration pressure measurement system still displays
0 mmHg (0 mBar).
XI. Approximately 6 seconds later, the DINAMAP™ Compact beeps and prompts for the pressure
to be set to 200 mmHg. Apply 200 mmHg and maintain that pressure.
XII. Approximately 12 seconds after prompting for 200 mmHg, the DINAMAP™ Compact will beep
3 times, then will vent the system to atmosphere. Immediately, prior to venting, the monitor will
record the measured pressure reading. This reading and the zero reading are the only points
used for calibration, so it is important to ensure that the correct pressures are applied at these
two points.
XIII. After venting to atmosphere, the DINAMAP™ Compact will display either “Calibration
Complete”or “Calibration Failed”.
XIV. If the calibration failed, turn off the DINAMAP™ Compact, check the equipment set up and
repeat the process. Refer to service if calibration is still unsuccessful.
X.
If the calibration was successful, the DINAMAP™ Compact will enter its “Failsafe Alarm” mode. The
DINAMAP™ Compact has now been calibrated and can be switched off.
Check calibration of unit by repeating section 3.4.3.
3.3.3 NIBP Pop Off Checking
The following section describes how to verify the pressure at which the DINAMAP™ Compact
“Overpressure”detection system prevents further cuff inflation and generates an alarm.
Neonate Mode
I.
II.
III.
IV.
V.
VI.
VII.
VIII.
IX.
X.
Switch on the DINAMAP™ Compact and wait for the main menu to be displayed on the LCD.
Select the “Mode”button from the main menu. The Mode menu will be displayed.
Select the “Service”button from the mode menu. The service entry panel will be displayed.
Enter the Super Service Mode PIN code, 8 3 7 8. This will display the super service menu.
Select NIBP from the super service main menu.
Select VSEN from sub menu followed by VAL2.
Connect calibration kit to the DINAMAP™ Compact as shown on page 41.
Inflate system and observe the calibration pressure measurement system - the pressure will
be seen to rise and, at the “Pop Off Pressure”, will immediately deflate, and the pump will stop.
Verify that the maximum pressure, as shown on the calibration pressure measurement system,
was in the range 150-165 mmHg (200-220mBar) and that the system pressure falls to less
than 20mmHg (26.6mBar) within 4 seconds.
Switch off unit
CAUTION
Ensure that the calibration pressure measurement
system used
for this test is capable of displaying
pressures up to 350 mmHg safely.
Adult Mode
I.
II.
III.
IV.
V.
38
Turn on the DINAMAP™ Compact, and allow 10 seconds for initialization.
Select the “Mode”button from the main menu. The Mode menu will be displayed
Select the “Service”button from the mode menu
Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
Select “NIBP”from the service menu. The NIBP Service menu will be displayed.
VI. Select “Check”from the NIBP service menu.
VII. Inflate the system and observe the calibration pressure measurement system - the pressure
will be seen to rise and, at the “Pop Off Pressure”, will immediately deflate, with the pump
stopping and DINAMAP™ Compact reporting an “Overpressure”alarm.
VIII. Verify that the maximum pressure, as shown on the calibration pressure measurement system,
was in the range 300 to 330 mmHg (400-440mBar).
3.3.4 NIBP Pop Off Pressure Calibration
This section describes how to set the Pop Off pressure - that is, the pressure at which the DINAMAP™
Compact hardware automatically opens the valves to atmosphere and stops the pump.
Only the “Neonatal”overpressure needs to be set; the adult pressure will be double that of the neonatal.
Calibrating the Pop Off pressure is not possible without disassembling the DINAMAP™ Compact, since a
jumper (LK1) needs to be added on the Main Board to allow the adjustment. The jumper is removed after
calibration during product manufacture.
I.
II.
III.
IV.
Ensure LK1 is fitted on the Main Board
Set up the DINAMAP™ Compact and calibration equipment as shown on page 41.
Switch on the DINAMAP™ Compact and allow 10 seconds to settle.
Verify that the calibration pressure measurement system reads 0 mmHg (0 mBar)
a. If the DINAMAP™ Compact displays the language menu, choose “English”, then “OK”.
b. Reboot the DINAMAP™ Compact by turning monitor off, then back on.
V.
Select the “Mode”button from the main menu. The Mode menu will be displayed
VI.
Select the “Service”button from the mode menu
VII.
Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
VIII.
Select “NIBP”from the service menu. The NIBP Service menu will be displayed.
-- Note -From this point, the timing is critical.
It is recommended to read through
the following steps before proceeding.
X. Select “Pop Off”from the NIBP service menu.
XI. Apply exactly 157.0 mmHg of pressure using the bulb pump or syringe, and hold the pressure.
XII. Approximately 12 seconds after the “Pop Off”option was selected, the monitor will bleep 3
times and then start calibration. Rigidly maintaining the pressure at this point is crucial.
IMPORTANT: The monitor requires a 157.0 mmHg pressure setting. Failure to maintain
157.0 mmHg could potentially cause an erroneous pressure reading to be stored in the
EEPROM, thereby rendering the monitor inoperable.
XIII. When calibration is complete, the monitor will enter its Failsafe alarm mode.
XIV. Turn off the DINAMAP™ Compact, and remove LK1.
XV. Check calibration by repeating test 3.4.3.
XVI. Perform the Manual Release Test Procedure as per Section 5.4 prior to release of unit
for clinical usage.
™
Note: If overpressure occurs when applying pressure in point (XII) above, turn off the DINAMAP Compact
and re-start the sequence.
39
3.3.5 Predictive Temperature Calibration Check
The predictive temperature calibration of the DINAMAP™ Compact monitor should be checked every 12
months or when there is doubt about the validity of the temperature readings. The temperature calibration
plug (part number 320-635) may be used to check the calibration of the temperature detection circuits within
the monitor.
-- CAUTION -DO NOT twist the calibration plug when inserting
or removing it since this may damage the precision
resistors contained within the plug.
To check the temperature calibration:
I.
Insert the plug into the temperature probe connector on the front of the monitor and switch the
monitor on.
II. Switch on the DINAMAP™ Compact and wait for the main menu to be displayed on the LCD.
III. If UUT ? C LED is lit, select MODE followed by SERVICE from the sub menu. Enter service
code 2 2 1 3.
IV. Select TEMP followed by ‘C OR F’from the sub menu.
V. When LCD displays message ‘Trend data will be lost’, select YES and confirm ? F LED is lit.
VI. Switch unit off then on.
VII. Select the TEMP button from the main menu.
VIII. After determination, the Temperature display should show 98.6 ? 0.1 ? F.
IX. If the display is off by more than ? 0.1 degrees, the monitor is out of calibration and should be
calibrated.
Note: The calibration plug contains an internal resistor that must be verified every time the monitor is
calibrated. Using a calibrated voltmeter, the resistance between pins 1 and 2 of the calibration plug
should be 6090 ? 7 ? . If the plug is not within this range, replace it.
3.3.6 Predictive Temperature Calibration
This section describes how to enter the temperature calibration code and how to calibrate the predictive
temperature circuitry. Calibration shall be conducted with an ambient temperature of 25 ? C ? 5 ? C.
Calibration of the predictive temperature is not possible without disassembling the DINAMAP™ Compact,
since adjustment of potentiometers RV2 and RV3 on the Main Board may be required.
Note: Entry to the super service menu disables all the temperature calibration error reporting.
3.3.6.1
Calibration Requirements
I.
II.
III.
40
A Counter/Timer with NAMAS approved calibration with a resolution of at least 0.1? S and an
accuracy of equal or greater than 0.1? S is required for set up of units with Main Board
software version earlier than 8613-V4.1.
A temperature calibration plug (part number 320-635) is required for the procedure, check that
the resistance between pins 1 and 2 of the temperature calibration plug is 6090 ? 7 ? .
Adjustments of calibration settings must be carried out in sequence and with no appreciable
time delay between steps. This is due to the procedure being reliant on no changes occurring
to the temperature of the UUT during the set up.
IV.
3.3.6.2
If UUT is a T or a TS model with Main Board software revision of 8613-V4.1 or greater,
proceed to the next section.
Temperature Calibration Procedure (Software Version 8613-V4.7 and subsequent)
I.
Remove screws in recessed handle of the UUT and drop the front cover forward to reveal
potentiometers RV2 and RV3 on the Main Board.
II.
III.
IV.
Connect AC Adapter to UUT DC input socket and switch on UUT.
Connect Temperature Calibration Plug to the temperature probe socket on UUT.
While depressing STAT, START and Alarm Mute buttons on front panel, switch on UUT.
Confirm UUT LCD is showing the Super Service Menu.
V. Select PTEMP from Super Service Menu, followed by PLUG.
VI. Adjust RV2 on UUT Main Board to obtain an edge count for the Calibration PLUG of 957.
VII. Remove calibration resistor from UUT temperature socket
Note: Test paragraphs VIII to IX must be carried out immediately after the preceding paragraph
VIII. Select 98 on the menu corresponding to the 98°F internal calibration point. Adjust RV3 until
the number of edges reported for the 98 point is 948? 1 edges.
IX. Select 54 on the menu corresponding to the 54°F internal calibration point. Confirm number of
edges reported is 57? 6 edges.
X. Turn unit off then on.
XI. Ensure that the Temperature Calibration Plug (Part No 320-635) is connected to the
temperature probe socket on the UUT.
XII. If UUT is set to read temperature in °C enter service mode PIN CODE 2 2 1 3. Select TEMP
followed by °C/°F, when unit displays message ‘Confirm Trend data will be lost Y/N’select
YES. Switch UUT off then on and confirm °F LED is lit.
XIII. Select TEMP from the main menu on the UUT LCD and confirm that reading displayed after
approximately 60 seconds is 98.6°F.
XIV. If reading is not 98.6°F then repeat from paragraph I.
3.3.6.3
Calibration Time Stamp
I.
II.
III.
IV.
V.
VI.
VII.
Enter Service Mode PIN code 2 2 1 3.
From the sub menu select TEMP.
To record the current time and date as the time and date of calibration select CAL.
Switch off UUT.
Switch on UUT
Enter Service Mode PIN code 2 2 1 3.
Select INFO and confirm date of NIBP and time/date of TEMP calibration are correct.
Carryout Manual Release Test Procedure as per section 5.4 prior to release of unit for clinical use.
41
3.4 Monitor Disassembly
Most alignment and repair operations can only be performed with the monitor disassembled. This section
describes how to remove the covers, disassemble the chassis and remove the circuit boards. The
description refers to fully equipped DINAMAP™ Compact TS models. DINAMAP™ Compact T models do not
have SpO2 functions and therefore the SpO2 related components are not fitted.
Before beginning disassembly, disconnect the line power cord from the power socket and remove the
rechargeable lead-acid battery. Be careful to observe anti-static handling precautions throughout.
Note:
Pictures shown are from a revision 06 unit, appearance may differ slightly on earlier units.
3.4.1 Removal of the Front Panel
The front panel contains the Display board, Liquid Crystal Display module and switchboard. The Rotor,
Temperature probe socket and SpO2 socket (model TS) are also mounted on the inside of the panel.
The front panel must be removed if access to any of the boards or modules mounted on it or behind it is
required. To remove the panel, release the two Pozidriv™ screws mounted inside the hand recess of the
instrument’s rear enclosure.
Front panel
retaining screws
Then carefully pull the top of the front panel away from the rear enclosure. The panel will tilt forwards but is
still retained by two plastic clips at its bottom edge. Unhook the front panel from these clips, taking care not
to damage them. This is particularly important when refitting the front panel later.
42
Before the panel can be separated from the case, it is necessary to disconnect the switchboard cable,
temperature probe cable and SpO2 sensor cable if fitted.
The switchboard ribbon cable plugs into socket PL10 on the Main Board via a ferrite, which clamps the
cable to the front of the chassis metalwork. Detach the ferrite from the chassis front plate by separating the
Dual-Lock™ fastener.
With the ferrite detached, the ribbon cable can be unplugged from socket PL10.
43
The cable from the Temperature probe socket plugs into connector PL14 on the Main Board and is easily
removed.
The SpO2 cable runs from the small front panel connector printed circuit board to another small PCB with
integral connector attached to the Nellcor Puritan Bennett™ module. To remove the PCB from the front
panel, locate and remove the two small Pozidriv™ retaining screws, which fix the PCB metal support bracket
to the front panel. The cable and PCB assembly can then be lifted away.
Attachment
points for hoses
SpO2 bracket
retaining screws
Finally remove the two rubber hoses, disconnect the pump hose at the pump and pressure sensor hose
from the front panel, noting their routing. The front panel can now be lifted away.
3.4.2 Removal of the Switch board, Display board and Liquid Crystal Display
module
™
The switch board, display board and LCD module are held in place with Pozidriv screws which, in some
cases, also hold the RFI earthing fingers in position. When removing the boards, note the orientation of the
copper RFI earthing fingers so that these can be replaced in the same position later.
44
RFI earthing
finger
Rotor control
Display board
8630AB
LCD backlight
cable
LCD panel
Switch board
Temperature
probe connector
cable
SpO2 connector (Model TS)
Blanking plate (Model T)
Display board
ribbon cable
Hose connector
Switch board
ribbon cable
45
3.4.3 Removal of the Chassis
The chassis contains the Main board (revision 04 and earlier), the Power Supply board and the Pump
assembly. SpO2 equipped models also have the Nellcor Puritan Bennett™ module fitted in a shield case
fixed to the front plate of the chassis. This is connected to the Main Board via the ribbon cable to SK13.
Unplug this connector. Note that there are no serviceable parts inside this module.
With the front panel removed, take off the front chassis plate by removing the two small Pozidriv™ retaining
screws.
Chassis front
retaining screws
SK13 for SpO2
equipped models
Lift off the front plate complete with SpO2 module if fitted. This will give access to the topside of the Main
Board including all the test points and the Flash memory.
The remainder of the chassis can now be released from the rear enclosure by removing a single large
Pozidriv™ screw located between the printer molding and the pump assembly.
Chassis retaining
screw
Once the screw has been removed, tilt the chassis forwards and remove the three plugs at the top of the
power supply board. From left to right, these plugs connect to the host port, the power input connector and
the remote alarm. Disconnect the printer cable from the Printer board. The chassis assembly may now be
lifted slightly and eased forwards, so that the speaker plug can be removed from the bottom of the power
supply board.
The chassis assembly may now be lifted clear of the rear enclosure.
46
3.4.4 Removal of the Main Board and Power Supply board from the chassis
To remove the Main and Power Supply boards from the chassis, the valve control and pump connector
must first be removed. This is visible through the hole in the rear of the chassis.
The chassis spade terminal must also be removed. This is located between the top of the chassis and the
Power Supply board (Unit revision 06 only).
Disconnect pneumatic hose from pump assembly to Main Board secondary pressure transducer on revision
06 units. (Disconnect pressure sensor cable on earlier revisions)
The boards can now be slid together up and out of the pcb guides built into the metalwork of the chassis.
47
To separate the boards, use a fine nosed pliers to pinch the tip of the nylon pcb spacer while gently pulling
the boards apart. Note that this will result in the loss of the user settings.
To avoid damage to the PCB interconnection, ensure that the boards are not tilted or twisted relative to
each other as they are separated.
3.4.5 Removal of the Pump and Pressure Transducer
To remove the Pump assembly, take out the two Pozidriv™ screws holding the metal pump retaining strap.
Rear chassis
metalwork
Pump assembly
Pump retaining
strap screws
The pump assembly may now be simply removed from the chassis by tilting it forward and pulling.
To remove the pressure transducer from the pump (revision 04 and earlier units), peel back the selfadhesive foam covering the transducer.
Then use a flat bladed screwdriver to pry out the U-clip. The transducer may now be lifted away from the
pump.
48
3.4.6 Removal of the Printer
The printer motor assembly and PCB are mounted upside-down inside the base of the rear enclosure.
Before removing the printer, take out the paper roll if fitted and leave the printer door open. To remove the
printer assembly, withdraw the three small Pozidriv™ screws. Support the printer with one hand as the
screws are removed.
Printer retaining
screws
When removed from the enclosure, the printer’s ribbon cable may be disconnected from the PCB by lifting
the outer part of the plastic connector. To refit the ribbon, first lift the outer part of the connector, insert the
ribbon, and then push the outer part of the connector down.
49
Printer module
ZIF socket
Printer board
8650AB
3.4.7 Reassembling the Monitor
Re-assembly is a direct reversal of disassembly, however additional care must be taken to ensure that all
cables and hoses are refitted correctly. The main points to note when reassembling are:
??
??
??
??
??
??
50
Pay particular attention to the wire dressing and ensure that all ferrite components are refitted
correctly.
When reassembling the chassis, remember to reconnect the valve control from the pump to the
power supply board.
When refitting the front panel, ensure that the copper earthing fingers of the front-panel engage
correctly inside the rear enclosure.
Ensure that the hoses to the front panel are not trapped or kinked while refitting the panel.
Keep your work area clean. It is quite easy for the plastic display panel cover to become
scratched if the front panel is pressed against a hard object such as a screw.
Be sure to re-calibrate and test the monitor as per section 3.4.1 and 3.4.6 before releasing
it for clinical use.
4
Manual Release Test Procedure
The following procedure can be used to check the primary operation of the DINAMAP™ Compact monitor in
order to release the unit for clinical use.
4.1 General
The DINAMAP™ Compact unit is to be tested by applying various stimuli to the units’interfaces or
measuring unit parameters. The unit test procedure utilizes functions from the operational software and also
a service test mode attached to the operational software. A guide to the UUT Rear Connector and Front
Panel layout can be found in section 9.
4.1.1 Test Equipment
a.) Pressure Gauge 0-375mmHg, 0.2% accuracy, Digitron P200L or similar
b.) Nellcor Puritan Bennett™ SpO2 Simulator Model SRC-2, PT2500 or N1290 or similar
c.) NIBP calibration kit part number 320-685
d.) Temperature Calibration Plug part number 320-635
4.1.2 Test Conditions
a.) Testing shall be conducted with an ambient temperature of 25 ? C ? 5 ? C
4.1.3 Service mode
4.1.3.1 Service Mode Selection
During the test procedure the operator will be requested to enter the Service Mode.
The UUT can be switched into its Service Mode by using the following procedure.
From the main menu select MODE
From the sub menu select SERVICE.
Enter the following code 2 2 1 3.
The UUT should then display the Service Menu.
4.1.3.2 Super Service Test Mode Selection
During the test procedure the operator will be requested to enter the Super Service Test Mode.
The UUT can be switched into its Super Service Test mode by using the following procedure:
From the main menu select MODE
From the sub menu select SERVICE.
Enter the following code 8 3 7 8.
The UUT should then display the Super Service Test Menu.
51
4.2 Test Procedure
4.2.1 Power Up Tests
4.2.1.1
LED Display
Press and release the main power on switch.
Check that all LED segments are illuminated momentarily.
Press and release the main power on switch.
Ensure unit shuts down.
4.2.1.2
LCD Display
Press and release ROTOR push button.
Confirm that all the Pixels of the LCD display are lit momentarily and that a series of three long beeps
followed by three short beeps can be heard.
Confirm that the display indicates the Monitor’s software revision momentarily.
If the instrument is not in calibration, the LCD will display an NIBP calibration request prompt.
Press and release the main power switch.
If the UUT requires calibration refer to section 3 of the Service Manual.
4.2.2 External DC Supply and Battery Indication Operation
4.2.2.1
UUT Supply Indicator Test
Switch on UUT wait for UUT to initialize and confirm that the External DC LED is lit and the battery LED is
extinguished.
Remove Battery and confirm that External DC LED on UUT Flashes.
Fit Battery and switch off External DC supply.
Confirm the External DC LED has extinguished and the Battery LED is lit.
Switch on External DC supply.
4.2.3 Temperature Requirements
Proceed to the next section if temperature circuitry has already been calibrated and checked according to
section 3.4.2 and 3.4.6
If TEMP option is unavailable on main operational menu of the UUT LCD:
Enter Service Mode as detailed in section 3.3.2.
Select TEMP option box on UUT LCD using rotor switch, press rotor push button and confirm it has a tick
against it.
Switch off UUT.
4.2.3.1
Predictive Temperature Accuracy Check
Switch on unit.
IF UUT ? C LED is lit, select MODE followed by SERVICE from the sub menu.
Select TEMP followed by ‘C OR F’.
When LCD displays message ‘Trend data will be lost’, select YES and confirm ? F LED is lit.
Fit predictive oral temperature calibration plug (part no 320-635) in to unit.
Using the ROTOR control select TEMP from the main menu.
Confirm that the Temperature LED displays (starting from the left) show the following sequence
52
‘? ’then ‘? ? ’, and ‘? ? ? ’.
Confirm first digit shows ‘-’indicating an oral probe is fitted.
Confirm LED display indicate 98.6°F (+/- 0.1°F).
4.2.4 Pulse Oximeter Requirements
The operation of the SpO2 module shall be tested if fitted (TS Model only). If no module is fitted continue
from section 4.2.5.
Switch on UUT.
Enter Service Mode as detailed in section 3.3.2
Select SYSTEM from sub menu.
Select SYSTEM from sub menu again.
If SpO2 option box on UUT LCD does not have a tick against it, select box using rotor control, press rotor
push button and confirm it has a tick alongside it.
Switch off UUT.
4.2.4.1
Pulse data and saturation data verification
Connect a Nellcor Puritan Bennett™ Pulse Simulator either a model SRC-2, PT-2500 or N-1290 to simulate
the oximeter sensor.
Switch on UUT and allow it to initialize.
Select ALARMS from main menu.
Set BPM upper limit to 250.
Set BPM lower limit to 10.
Set SpO2 upper limit to OFF.
Set SpO2 lower limit to 50.
Plug in Nellcor Simulator and switch on (N-1290 only).
Test DINAMAP™ Compact unit using the full complement of SpO2 and pulse rate settings available on your
simulator.
Confirm unit is within the tolerances given below.
Saturation Tolerance:
Rate Tolerance:
4.2.4.2
70 - 100%
+2 digits
50 - 69%
+3 digits
+ 3 beats per minute from 20 - 250 beats per minute.
Pulse Search verification
Set pulse rate to zero on Nellcor Puritan Bennett™ simulator (N1290 only) or disconnect simulator, verify
the "No SpO2 Signal" message is displayed on the UUT LCD. Acknowledge message by pressing rotor
switch.
4.2.4.3
Probe Disconnected Verification
After a period of normal operation with a pulse input of 90 applied, disconnect SpO2 simulator from UUT,
verify that the "No SpO2 Signal" message is displayed on the LCD.
Acknowledge message by pressing rotor switch.
Verify that the "No SpO2 Sensor" message is displayed on the LCD from the module.
Acknowledge message by pressing rotor switch.
Switch off UUT.
53
4.2.5 NIBP Requirements
4.2.5.1
NIBP Calibration Check
Note: Proceed to section 4.2.6 if NIBP system has already been calibrated and checked in accordance with
section 3.4.1.
The linearity and calibration for both pressure transducers (PT1 and PT2) shall be verified from 0 to 290
mmHg, the tolerance shall be ? 3 mmHg across the whole range.
Set up unit and calibration kit as per diagram on page 41.
Switch UUT on.
Enter Service mode as detailed in section 3.3.2.
Select NIBP from sub menu.
Select CHECK from sub menu.
Apply the following pressures as measured by your Pressure Gauge and confirm that the DINAMAP™
Compact readings are as per the following table for both PT1 and PT2 channels.
PT1 & PT2 Linearity Check Table
Applied Pressure mmHg (mBar)
0 (0)
58 (77.3)
116 (155)
174 (232 )
232 (309)
290 (387)
DINAMAP™ Compact UUT Measured Pressure
mmHg
0 -0,+3
58 ? 3
116 ? 3
174 ? 3
232 ? 3
290 ? 3
If the Calibration Check is satisfactory continue from section 5.2.6 otherwise refer to section 3.4.2 of this
Manual if calibration is required.
4.2.5.2
Pressure System Leakage Test.
Apply a pressure of 200mmHg (267mBar) to the DINAMAP™ COMPACT unit using the NIBP calibration kit.
Confirm that after 36 seconds the pressure has not dropped by more than 5mmHg (6.66mBar).
4.2.5.3
Overpressure Limit Tests
Proceed to section 4.2.6 if NIBP system has already been calibrated and checked in accordance with
section 3.4.2.
4.2.5.3.1
Neonate Overpressure Limit Check
Enter Super Service Mode as detailed in section 3.3.2.
Select NIBP from Super Service Test menu.
Select VSENSE from sub menu.
Select VAL2 from sub menu.
Increase applied pressure until overpressure failsafe occurs.
Confirm the following:
Pressure at failsafe point is between 150mmHg to 165mmHg (200mBar to 220mBar)
The system pressure falls to less than 20mmHg (26.6mBar) within 4 seconds.
Switch off UUT.
54
If UUT fails re-calibrate unit as per section 3.4.2.
4.2.5.3.2 Adult Overpressure Limit Check
Switch on UUT.
Enter Service mode as detailed in section 3.3.2.
Select NIBP and then CHECK.
Increase applied pressure until overpressure failsafe occurs.
Confirm the following:
Pressure at failsafe point is between 300mmHg to 330mmHg (400mBar to 440mBar)
The system pressure falls to less than 20mmHg (26.6mBar) within 4 seconds.
Switch off UUT and disconnect calibration kit.
If UUT fails re-calibrate unit as per section 3.4.2
4.2.6 Printer Operation
Switch on UUT.
The operation of the internal printer shall be tested if fitted, otherwise ignore this section.
If printer is fitted and the PRINT option is unavailable on main operational menu of the UUT LCD:
Enter Service Mode as detailed in section 3.3.2.
Select PRINTER option box on UUT LCD using rotor switch, press rotor push button and confirm it has a
check mark beside it.
Switch off UUT and insure a paper roll is fitted in the printer and the printer door is shut.
4.2.6.1
Print Quality
Confirm printer status on LCD indicates ‘Man’.
Select PRINT option from main menu followed by NOW. Check printer output and confirm that all characters
are readable and have an even contrast.
4.2.6.2
Printer Error
Open the printer door.
From the main menu select PRINT and NOW.
Confirm that the printer does not start and LCD printer status display indicates ‘No Paper’…
Remove paper and close printer cover.
Confirm that the printer does not start.
Replace paper and close the printer door.
Select PRINT and NOW and confirm printer starts.
4.2.7 Calibration Time Stamp
If all temperature readings correct switch off UUT and switch on again.
Enter Service Mode as detailed in section 3.3.2.
From the sub menu select TEMP.
To record the current time and date as the time and date of calibration select CAL.
Switch off UUT.
Switch on UUT
Enter Service Mode as detailed in section 3.3.2.
Select INFO and confirm date of NIBP calibration and time and date of TEMP calibration are correct.
55
4.3 Manual Release Test Results Sheets
Test Section
4.2.1.1
Test Description
LED Display
4.2.1.2
LCD Display
4.2.2.1
4.2.3.1
4.2.4.1
All Pixels Lit
UUT Software Revision
UUT Supply Indicator
Test
-
External DC LED
Operation
Battery LED Operation
Predictive Temperature
Accuracy Check
-
Oral/98.6? C
Pulse data and
saturation data
verification
SAT/PULSE
4.2.4.2
4.2.4.3
56
Test Results
-
Pulse Search
verification
Probe Disconnected
Verification
-
PASS
FAIL
Test Section
4.2.5.1
4.2.5.2
4.2.5.3.1
4.2.5.3.2
Test Description
NIBP Calibration Check
(Applied Pressure)
Test Results
PT1
(mmHg)
PASS
FAIL
PT2
(mmHg)
0mBar
77.3mBar
155mBar
232mBar
309mBar
387mBar
Pressure System
Leakage Test.
Target Pressure
Pressure after 36s
Neonate Overpressure
Limit Check
Overpressure Limit
Final Pressure
Adult Overpressure
Limit Check
4.2.6.1
Overpressure Limit
Final Pressure
Print Quality
4.2.6.2
Printer Error
-
No paper
Door open
4.2.7
Calibration Time Stamp
Unit Serial No:
Product Code:
Test Completed by :
Date:
57
5 Trouble Shooting.
UUT Problem
UUT will not switch on
from EXT DC source.
UUT will not switch on
when powered by
Battery.
Possible Causes
EXT DC adapter o/p out of spec.
Further Checks
Measure o/p.
EXT DC Fuse Blown.
Check Fuse.
PSU Board fault.
Swap or test PSU Board.
If unit can be powered from a
battery source then a fault exists
in the external DC detect
circuitry on the PSU Board.
Measure terminal voltage of
battery.
Battery discharged.
PSU Board fault.
Unit will not switch on
from either Battery or
EXT DC source.
+5VD supply low.
Measure +5VD supply on Host
Comms Connector +VE Pin 4, VE Pin 1 and confirm is in spec.
PIC Processor not running.
No Keyswitch response.
Switch Board failure.
Reset PIC by linking Pin 8 to Pin
1 on the Host Comms port
momentarily. Then attempt to
power on UUT.
Swap or test Switch Board.
No Rotor response.
Main Board failure.
Rotor failure.
Swap or test Main Board.
Test Rotor operation.
Switch Board failure.
Swap or test Switch Board.
Main Board failure.
Rotor failure.
Swap or test Main Board.
Swap or test Rotor Switch.
Switch Board Failure.
PSU Board.
Swap or test Switch Board.
Swap or test PSU Board.
Switch Board.
Swap or test Switch Board.
(Probable D3 O/C)
When Rotor is revolved
menus are skipped.
Pressing Rotor Switch
fails to switch on unit but
menu selection is
possible.
58
On/Off switch faulty.
Swap or test PSU Board.
If unit can be powered from
mains then fault lies on PSU
Board Between the Battery and
the input to IC7. Check
DC_Battery Signal is available to
the PIC and confirm operation of
USE_Battery switch TR2.
Try turning unit on by pressing
the rotor switch. If UUT switches
on check On/Off signal from
Switch Board to PSU Board.
UUT Problem
Incorrect or No LCD
output.
Possible Causes
LCD Board fault.
Addressing fault either:Switch Board or
Main Board (will usually be
accompanied by an audible
alarm if Main Board Processor
has stopped running).
No LCD Backlight.
Incorrect or No LED
display output.
Check Negative Supply reaches
LCD Module.
Display Board fault.
Check Operation of backlight
switch TR1 and confirm
+Vbacklight is present on PL31
Pin 2 w.r.t Pin 1.
Addressing fault either:Switch Board.
Main Board.
Note: If a partial loss of the
LED’s is found the fault is most
likely to reside on the LCD
Display Board.
Addressing fault either:Switch Board or
Main Board (will usually be
accompanied by an audible
alarm if Main Board Processor
has stopped running).
No or low +4.5V VDISP supply.
SpO2 No signal error
(P55 Error).
Swap or test Switch Board.
Swap or test Main Board.
Negative supply to LCD module,
potted down on the Switch Board
from the -12v supply, out of
spec.
LCD Board fault.
LED Display Board fault.
SpO2 Probe
disconnected error
reported on LCD
(P00 Error).
Further Checks
Swap or test LCD Board.
Faulty Finger Probe.
Swap or test LCD Board.
Swap or test Switch Board.
Swap or test Main Board.
Swap or test LED Display Board.
Swap or test Switch Board.
Swap or test Main Board.
Swap or test PSU Board.
Check supply continuity from
PSU Board to Display Board.
Test with known good sensor.
Faulty SpO2 Module to Front
Panel cable.
Swap or check cable continuity.
SpO2 Module fault.
Swap module and test unit with
Nellcor SpO2 Simulator.
Test with known good sensor.
Faulty Finger Probe.
Faulty SpO2 Module to Front
Panel cable.
Swap or check cable continuity.
SpO2 Module fault.
Swap module and test unit with
Nellcor SpO2 Simulator.
59
UUT Problem
After UUT power up
sequence is complete
UUT Alarm Sounds with
the following error
message displayed on
LCD.
System Error, P105
or
System Error, P110
or
System Error, P115
When the pump starts
an NIBP Overpressure
warning is displayed on
LCD (even without cuff
attached to unit) .
When the pump starts
an NIBP Overpressure
warning displayed on
LCD when a cuff is
attached.
When an NIBP
determination is started
pump is laboring
i.e. pump sound is lower
in frequency than
normal.
Fault may be
accompanied with either
of the following warning
messages:‘N33-NIBP Inflation
Timeout’
or
‘N00-NIBP
Overpressure’
60
Possible Causes
SpO2 Module fault.
Further Checks
Swap SpO2 module and power
up UUT.
Isolated Power Supply Module
failure on Main Board.
Check +5VI (+/-0.25V), -15VI &
+15VI (+/-1V) isolated supplies
are in specification.
Main Board Processor SpO2
RS232 comms failure.
Swap Main Board and power up
UUT.
Blocked or kinked hose from
pump output to front panel.
Examine hose for kinks or
blockage.
Pneumatic Assembly fault.
Check Pneumatic Assembly for
blocked output. Swap or test
Pneumatic Assembly.
Kinked hose to cuff or faulty cuff.
Test unit with known good cuff.
Blocked or kinked hose from
Front panel to Main Board
pressure sensor ML1.
Examine hose for kinks or
blockage.
Main Board fault.
Kinked hose to cuff or faulty cuff.
Swap or test Main Board
(ML1, IC5 or IC7 may be
defective).
Test unit with known good cuff.
Blocked or kinked hose from
pump output to front panel.
Examine hose for kinks or
blockage.
Pneumatic Assembly fault.
Check Pneumatic Assembly for
blocked output. Swap or test
Pneumatic Assembly.
UUT Problem
N12 Displayed on
Systolic display and
alarm sounding. LCD
display is blank.
Possible Causes
State of pneumatic system does
not concur with the unit’s
expected software state.
e.g. pressure in system exceeds
ambient level despite system
being set open to atmosphere.
If a failure condition arises in the
valve control circuitry that does
not allow the cuff to deflate an
N12 error will be reported.
Further Checks
Determine status of valve drive
and hence locate any valve drive
problem.
Most likely to be a fault on the
Pneumatics Control Gal IC1 or
TR1 or TR4 on the PSU board.
Therefore swap or test PSU
Board.
Fault could also be caused by
Main Board Failure with IC3
being the most likely cause.
N13 Displayed on
Systolic display and
alarm sounding. LCD
display is blank.
If pump runs continuously on
power up or remains on at any
time when it is supposed to be
off a N13 fault condition will be
set.
Most likely to be a fault on the
Pneumatics Control Gal IC1 on
the PSU board. Therefore swap
or test PSU Board.
Fault could also be caused by
Main Board Failure with IC3
being the most likely cause.
Fault could also be due to
defective pump current
measurement circuitry on PSU
Board.
When an NIBP
measurement is
attempted unit alarms
with the following
message on LCD.
Leak in pneumatic system.
Test unit with known good cuff.
‘N33-NIBP: Timeout
Inflation’
When an NIBP
measurement is
attempted unit alarms
with the following
message on LCD.
‘N44-NIBP: Timeout’
Total Time’
or
‘N55-NIBP: Timeout
Pressure’
or
If unit is leaking check integrity
of internal hoses if ok swap or
test pump assembly.
N44 and N55 warnings may be
caused by excessive movement
of patient (‘Motion Artifact’).
Ensure ‘Motion Artifacts’are not
the cause of the failure.
Main Board NIBP circuitry fault
or
PSU Board pneumatic control
fault or Pneumatic Assembly
fault
Carryout an NIBP calibration
check this may indicate if their is
a problem with either pressure
measurement channels PT1
(shown on systolic display) or
PT2 (shown on diastolic display).
If calibration is satisfactory fault
may be due to a pneumatic
problem with cuff deflation.
The fault may also reside on the
NIBP pulse determination
circuitry on the Main Board.
‘N99- No determination’
61
UUT Problem
Unit alarms with the
following message on
LCD
Possible Causes
A pump over current condition
has been signaled to the Main
Board processor.
Further Checks
‘Pump Over Current
0,(code line failure),0,0
Pump failure.
Swap or test Pump Assembly.
PSU Board failure of pump
current sense circuitry.
Swap or test PSU Board.
Main Board failure
Unit has failed temperature
calibration for its 54°F point.
Swap or test Main Board.
Main Board failure most likely.
Check output of isolated supply
module on Main Board +5VI,
+/-15VI and VREFI.
Unit alarms with the
following message on
LCD
‘TSH: HE_54CP’
Unit alarms with the
following message on
LCD
Unit has failed temperature
calibration for its 98°F point.
Main Board failure most likely.
‘TSH: HE_98CP’
Unit alarms with the
following message on
LCD
Check output of isolated supply
module on Main Board +5VI,
+/-15VI and VREFI.
During temperature calibration
an error greater than can be
compensated for has been
detected.
‘TSH: SE_TACF’
No response to Host
Communications.
Remote Alarm Output
not operating.
Main Board failure is most likely.
PSU Board
Check VREFI supply.
Swap or test PSU Board.
Main Board fault.
Swap or test Main Board.
PSU Board or Main Board fault.
Check continuity of PSU Board
to Host Comms Port cable on
rear panel.
Stimulate an alarm condition by
adjusting a parameter alarm to a
minimum level and applying a
higher stimulus to the unit.
Check remote alarm output as
shown in section 0.
Swap or test PSU Board.
Swap or test Main Board.
Check continuity of cable
between PSU Board and
Remote Alarm socket on rear
panel.
62
UUT Problem
Unit alarms with the
following message on
LCD:
Possible Causes
The PSU Board PIC software
has detected that the +12VR or 12VR rails are operating out of
spec.
‘PIC: +/-12V Out of
Spec.’
(Filtered Voltage),(Code
Line Failure), 0,0
Unit alarms with the
following message on
LCD:-
The software has detected an
incorrect PIC software version to
run with Main Board software.
Further Checks
Check and isolate faulty supply.
Fault is most likely to be on the
PSU Board or the Main Board.
Fault could also be caused by
input voltage to IC18 or IC19 on
PSU board going to low and
therefore causing the regulators
to dropout. This could be caused
by excessive current drawn on
the +12VV supply or
interruptions to the input supply
caused by erroneous operation
of supply switch TR10.
Check Main Board and PSU
Board PIC software compatibility
‘PIC: Incompatible SW
Version
0, (code line failure),0,0
63
6 TECHNICAL INFORMATION
6.1 Host Port Connector (rear panel)
Pin 15
Pin 1
Important! For use ONLY with equipment conforming to IEC-601.
Pin
Function
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Ground
Inverted TTL Transmit Data
Inverted TTL Receive Data
Fused +5 volts
No connection
No connection
Ground
No connection
RS232 Request to Send (RTS)
RS232 Clear to Send (CTS)
RS232 Transmit Data (TxD)
No connection
RS232 Receive Data (RxD)
No connection
No connection
6.2 Remote Alarm Connector (rear panel) – if so equipped
Pin 1
Important! For use ONLY with equipment conforming to IEC-601.
Relay contact Maximum rating: 25V AC / 60V DC @ 1 A
Pin
1
2
3
4
Normal Condition
Contact Closed
Common
Contact Open
No Connection
Alarm Condition
Contact Open
Common
Contact Closed
No Connection
Note: Pins are numbered anti-clockwise when viewed from rear of monitor.
64
6.3 Accessory Part Numbers
Description of Compatible Part
Code
™
SOFT-CUF CUFFS
Infant, Orange/White
Child, Green/White
Child Long, Green/White
Small Adult, Lt Blue/White
Small Adult Long, Lt Blue/White
Adult, Navy/White
Adult Long, Navy/White
Large Adult, Rose/White
Large Adult Long, Rose/White
Thigh, Brown/White
2500
2501
2506
2502
2507
2503
2604
2504
2644
2505
CLASSIC-CUF™ CUFFS
Infant
Child
Small Adult
Adult
Large Adult
Thigh
2618
2613
2608
2603
2643
2648
CLASSIC-CUF™ CUFFS, Isolation Use Yellow
Adult
Small Adult
Large Adult
2602
2607
2642
DURA-CUF® CUFFS
Infant, Rust
Child, Green
Small Adult, Royal Blue
Adult, Navy Blue with Hanger
Adult, Navy Blue
Adult Long, Navy Blue
Large Adult, Wine
Large Adult Long, Wine
Thigh, Brown
2783
2781
2779
2771
2774
2772
2791
2784
2796
Air Hose, Gray, Adult Pediatric, Screw Connector, 12ft.
Air Hose, Gray, Adult Pediatric, Quick Disconnect, 12ft.
Air Hose, Gray, Adult Pediatric, Screw Connector, 24ft.
Air Hose, Teal, Neonatal, 8ft.
107365
88847
107366
107368
DINAMAP® Compact Operation Manual
DINAMAP® Compact Service Manual
776980
776856
Pole Stand
3210
Basket
3211
Temperature Calibration Kit
320635
65
Description of Compatible Part
AC/DC Power Converter
300978
Printer Paper, 10 rolls per box
89100
SpO2 Extension cable
EC-8
Nellcor Puritan Bennett™ Finger Sensor
66
Code
DS100A
Oral Temperature Probe, Blue, 21”Coiled
8978
Rectal Temperature Probe, Red
8976
Temperature Probe Covers (5000 per case)
8815
DINAMAP™ Compact Vital Signs Monitor
6.4 DINAMAP™ Compact Service Spares List
Items available from Service Center:
Pfx
BAT
BAT
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CAP
CBL
CBL
GE Medical
Systems P/N
633150
633132
604194
604195
604169
604197
604198
604199
604200
604201
604202
604204
604205
606114
606189
606115
606162
606151
606116
606117
606226
602253
602254
602255
605445
605736
605446
605628
605920
605534
605535
605157
605536
605158
605159
605626
605627
605447
604206
604207
604208
680388
680367
Description/Nomenclature
Battery, 3V6, 100mAh, NiCd, PCB
BATTERY 12V 2.3AH LEAD ACID
CAP, 470uF, 20%, 6V3, ELECT, SMT, 8.0 DIA x 10.0
CAP, 2200uF, 20%, ELECT, AXIAL
CAP, 1000uF, 20% 10V, ELECT, Radial, 7.5SP
Cap, 470uF, 16V, Elect, Radial, 5.0SP
Cap, 10uF, 16V, Elect, Radial, 1.5SP
CAP, 220uF, 20%, 25V, ELECT, LOZ, RADIAL, 3.5SP
CAP, 470uF, 20%, 25V, ELECT, LOZ, RADIAL, 3.5SP
CAP, 220uF, 20%, 10V, ELECT, LOZ, RADIAL, 5.0SP
CAP, 150uF, 20%, 16V, ELECT, LOZ, RADIAL, 3.5SP
CAP, 100uF, l20%, 16V, ELECT, RADIAL, 2.5SP
CAP, 330uF, 20%, 16V, ELECT, RADIAL, 3.5SP
CAP, 1u0, 20%, 16V, TANT, SMT, Case A
CAP, 4u7, 20% 35V TANT SMT CASE C
Cap, 4u7, 20%, 16V, TANT, SMT, Case B
CAP, 2u2, 20% 25V TANT SMT CASE A
CAP, 2u2, 20%, 25V, TANT, SMT, Case A
CAP, 10uF, 20%, 16V, TANT, SMT
CAP, 68uF, 16V, TANT, SMT
Cap, 47u 20%, 10V, TANT, SMT, Size D
Cap, 470p, 3kV, CER
CAP, 100pF, 2%, 100V, Low K, CER PL, 2.5SP
Cap, 0ul, 20%, 100V, MULTI CER, Radial, 2.5SP
Cap, 47pF, 5%, Multi CER, SMT, 0805 Size
CAP, 330upF, 5%, 50V, MULTI CER, SMT, 0603 Size
CAP, 0u001, 5%, 50V, MULTI CER, SMT, 0805 Size
CAP, 1u0, +80-20%, 16V, MULTI CER, SMT, 0805 Size
CAP, 22pF, 5%, 50V, MULTI CER, SMT, 0603 Size
Cap, 0u01, 10%, 50V, Multi CER, SMT, 0805 Size
Cap, 0u033, 10%, 50V, Multi CER, SMT, 0805 Size
CAP, 0u39, 10%, 50V, MULTI CER, SMT, 1206 Size
Cap, 0u001, 10%, 50V, Multi CER, SMT, 0805 Size
Cap, 1u0, 10%, 16V, Multi CER, SMT, 1206 Size
CAP, 0u33, 10%, 25V, MULTI CER, SMT, 1206 Size
CAP, 0ul, 20%, 50V, Multi CER, SMT, 0805 Size
CAP, 0u47, 20%, 50V, MULTI CER, 0805
CAP, 330pF, 5%, 100V, MULTI CER, SMT, 0805 Size
Cap, 0ul, 5%, 63V, Polyester, Radial, 5.08SP
Cap, 0ul5, 5%, 63V, Polyester, Radial, 5.08SP
Cap, 1u0, 5%, 63V, Polyester, Radial, 5.08SP
Wire, 24AWG , 0.56mm, BTC
Wire, 1/0.6mm, (0.28mm), PVC, BK
67
DINAMAP™ Compact Vital Signs Monitor
Pfx
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CBL
CHM
CHM
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
68
GE Medical
Systems. P/N
680387
680368
680370
680371
680376
680377
680378
680380
683211
683212
683213
683214
683215
683216
683217
752304
756136
756137
742147
742148
742149
742151
601186
782243
782242
608240
608241
608242
608244
704734
607978
607979
607980
607981
752290
607982
607983
607984
607985
607986
607987
607988
607989
607990
607991
607992
607994
607995
Description/Nomenclature
Wire, 30AWG (0.25mm), Teflon green wrap
Wire, 16/0.2mm (0.5mm), PVC, BK
Wire, 16/0.2mm (0.5mm), PVC, R
Wire, 16/0.2mm (0.5mm), PVC, OR
Wire, 7/0.127mm (28AWG), PVC, OR
Wire, 7/0.127mm (28AWG), PVC, Y
Wire, 7/0.127mm (28AWG), PVC, V
Wire, 7/0.127mm (28AWG), PVC, W
Cable, 9 Way, 7/36 AWG, IDC, GY
Cable, 14 Way, 7/36 AWG, IDC, GY
Cable, 16 Way, 7/36 AWG, IDC, GY
Cable, 20 Way, 7/36 AWG, IDC, GY
Cable, 34 Way, 7/36 AWG, IDC, GY
Cable, 6 Way, LO NSES, Nellcor, 044385
Cable, 2 Way, 13/0.2mm, FIG 8
GROMMET STRIP, 1.3-2.2mm (17-14AWG) PANEL
Cable Tie, 100L x 2.5W, 16 DIA MAX, NYLN, NATL
Cable Tie, 100L x 2.5W, 18 DIA MAX, M3, NYLN, NATL
Sleeving, H/S, 1/8 (3.2mm), BK
Sleeving, H/S, 3/16 (4.8mm), BK
Sleeving, H/S, 1/4 (6.3mm), BK
SLEEVING, H/S 1" (25.4MM) BK
CABLE, EXTENSION
ADH, RUBBER/PLASTIC LOCTITE PRISM 406
ASHESIVE, ETHYL CYANOACRYLATE, RBR, 4105
Pin, Test, Eye Terminal
Pin, Crimp, 'Molex' 4809, Anti Fishhooking
Pin, Crimp, 'Molex', 2478 Anti Fishhooking
Pin, Crimp, Female, 'Hirose', AWG 26-30
Contact, Battery
HDR, 2 Way, STRT, 2.54SP
HSG, 2 Way, STRT, Lock, 'Molex', 2.54SP
HDR, 2 Way, RT ANGL, Lock, 'Molex', 2.54SP
SKT, 2 Way, JMPR, 2.5SP, Black
PLG, 2 Way, DC PWR
HSG, 3 Way, STRT, Lock, 'Molex', 2.54SP
HDR, 3 Way, RT ANGL, Lock, 'Molex', 2.54SP
HDR, 4 Way, RT ANGL, Lock, 'Molex', 3.96SP
HSG, 4 Way, Free, STRT, Lock, 'Molex', 3.96SP
SKT, 4 Way, Pan MT, STRT, CIRC
HDR, 6 Way, STRT, Lock, 'Molex', 2.54SP
HSG, 6 Way, STRT, Lock, 'Molex', 2.54SP
HDR, 6 Way, STRT, 'HIROSE', SMT, 1.25SP
HDR, 2x3 Way, STRT, 'Hirose', SMT, 2.0SP
HDR, 6 Way, STRT, Shrouded, 2.0SP
HSG, 6 Way, Free, STRT, 'Hirose', 1.25SP
SKT, 6 Way, Pan MT, STRT, CIRC
SKT, 9 Way, 'D', RT ANGL, PCB
DINAMAP™ Compact Vital Signs Monitor
Pfx
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CON
CUS
CUS
CUS
ENC
ENC
ENC
ENC
FIX
FIX
FIX
FIX
FSE
FSE
FSE
FSE
HWR
HWR
HWR
HWR
HWR
HWR
HWR
HWR
HWR
ICS
ICS
GE Medical
Systems. P/N
607996
607997
607998
607999
608322
608323
608351
608324
608325
608326
608327
608328
608329
608330
608352
608331
608332
608333
608334
608335
608336
608337
633148
669141
320692
759331
759332
732174
732175
723136
715123
723133
723134
628177
628174
628175
617121
735326
735327
735328
722234
736216
736217
736230
727166
740201
694143
623123
Description/Nomenclature
SKT, 2x5 Way, STRT, 2.5SP
HDR, 2x7 Way, Shrouded, 2.54SP
Transition, 2x7 Way, IDC, 2.54SP
HDR, 2x7 Way, RT ANGL, Shrouded, 'Molex', 2.5SP
SKT, 2x7 Way, IDC, 2.54SP
SKT, 15 Way, 'D', IDC, PAN MT
HDR, 2 X 8 WAY SHROUDED 2.54SP
HDR, 2x8 Way, RT ANGL, Shrouded, 'Molex', 2.5SP
SKT, 2x8 Way, IDC, 2.54SP
HDR, 2x10 Way, Shrouded, 2.54SP
Transition, 2x10 Way, IDC, 2.54SP
SKT, 2x10 Way, IDC, 2.54SP
SKT, IC, 20 Way, PLCC, SMT
SKT, 20 Way, STRT, PCB, 'Harwin', 2.54SP
SKT, 26 WAY STRT MOLEX FFC, 1.25SP
SKT, IC, 32 Way, PLCC, SMT
HDR, 2x17 Way, Shrouded, 2.54SP
SKT, 2x17 Way, IDC, 2.54SP
Transition, 2x17 Way, IDC, 2.54SP
HDR, 2x20 Way, Shrouded
SKT, 2x20 Way, PCB, 2.54SP
SKT, IC, 44 Way, PLCC, SMT
PSU, Isolated (633148)
Transformer (669141)
Pneumatic Manifold Assembly
Case, Back, Metal, Sp02, Module
Case, Front, Plastic, Sp02, Module
FOOT, SQUARE, 20MM SELF ADHESIVE
FOOT, ROUND, 12.7 DIA X 3.5H SELF ADHESIVE
WASHER, M2.5 PLN, STL ZNPL
Nut, M3, Full, STL, ZnPL
Washer, M3, PLN, STL, ZnPL
WASHER, M4 PLN STL ZNPL
FUSE, 3.15A (T) 20MM, GLASS
Fuse, 100mA, Resetable, PTC RES, 5.1SP
Fuse, 2.5A, Resetable, PTC RES, SMT, 9.5x6.7
FSEHLDR, 20x5mm, PAN MT, Snap
CARD GUIDE, PCB 114.3L X 6.2W, 4IN CRS
CARD GUIDE, PCB 75.8L X 6.2W, 2.5IN CRS
SPACER, PCB, LOCK, 10.0L
SCREW, KIT, LOCKING, FEMALE DCONN 12.7MM
Clip, Spring, Knob
CIRCLIP, 10.5 DIA SHAFT 8.0 DIA GROOVE E TYPE
CLIP, SPRING, PUSH-ON, 4.0 DIA
SPRING, TORSION 10.0 DIA 4.6 LG 5 COILS 180 DEG
TUBING, AYX42007, 1/4 X 1/8
MC68302FC20 Microprocessor, 16 Bit, 20MHz, 132PQF
PIC16LC74-04-L MICROCONT 8 BIT ADC 44PLCC
69
DINAMAP™ Compact Vital Signs Monitor
Pfx
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
ICS
IND
IND
IND
70
GE Medical
Systems. P/N
694126
692252
619279
623124
692233
693171
692256
693112
693126
693172
693176
693177
693178
693179
693167
693180
693181
693182
693183
693184
693185
691110
691128
691129
691139
691147
691109
691117
693111
691148
692253
692254
692139
692112
692255
692250
692113
691145
691146
694140
692208
693170
692251
694141
694142
669218
669219
669220
Description/Nomenclature
IC, 128x8 HIGH SPEED CMOS SRAM SMT
NM93C46M8 EEPROM, CMOS, 1kBIT, Serial, SO8
AM29F040A-70JC FLASH CMOS, 4MBIT 5V 32PLCC
GAL16V8A-15J EE ARRAY PROG 64 X 32, 20PLCC
MAX163BCWG ADC< 12BIT, Track & Hold, SO24W
AD654JR Voltage to Freq Converter, SO8
BU4S71, or Gate, SNGL, S05
DG444DY SW, Analog, SPST, Quad, SO16
DG508ADY Multiplexer, 4 CHAN, Analog, SO16N
MAX7219CNG Driver, 8 Digit LED, DP24.3
L293DDWP Driver, Push Pull, 600mA, QUAD, SO28W
78L12ACM Volt REG, +12V, 100mA, SO8
LT1086-12CT Volt REG, +12V, 1.5A, TO220
MAX666CSA Volt REG, Dual Mode, +5V, PROG, SO8
79L12ACD Volt REG, -12V, 100mA, SO8
LM2577T-ADJ Volt REG, ADJ, +11.6/12.4V, 3A, TO226
L4960H Volt REG, ADJ, +5V/40V, 2.5A, Heptawatt
MAX758ACPA Volt REG, ADJ, Current MDE, 750mA, DP8
MAX758ACSA Volt REG, ADJ, Current MDE, 750mA, SO16
MAX744ACWE Volt CONV, +6V/16V to 5V, 2A, SO16W
MAX774CSA DC-DC CONT, INV, -5V/ADJ, 1A, SO8
TL082CD OPAMP, JFET, Dual, SO8
CMP04FS Comparator, PRECN, Quad, SO14
AD620AR OPAMP, Instrumentation, LO PWR, SO8
AD822AR OPAMP, FET, SINGL/Dual SUPP, Dual, SO8
LM392M OPAMP, FET, SNGL, Dual, SUPP, Dual, SO8
LM358D OPAMP, HI GAIN, Dual, SO8
IC, LM386 AUDIO POWER AMP SUR MOUNT
4051BD Multiplexer, 8 CHAN, Analog, SO16
4066BM Switch, Analog, Bilateral, QUAD, SO14
74HC32D or Gate, 2 Input, QUAD SO14
74HC14D Inverter, CMOS, Schmitt TRIG, HEX, SO14
74HC03D NAND, 2 Input POS, Open DRN, QUAD SO14
74HC259D Addressable Latch, 8 Bit SO16
74HC138D Decoder, 3 to 8 Line, SO16
74HC574DW Flip Flop, D Type, POS Edge, OCTAL, SO20
74HC541DW Buffer, Tristate, OCTRAL NON INV, SO20W
74HC00D NAND Gate, 2 Input POS, QUAD, SO14
74HC4020D Counter, Binary, 14 Stage, SO16
MAX809TCUR Supervisory, Microprocessor, SOT23
X9312US EEPOTR, 50k, Digitally Controlled, SO8
AD586JR Volt REF, PRECN, 5V, SO8
NE555D Timer, SNGL, SO8
MAX232ACWE TxRx, Dual, RS232, +5V, PWRD, SO16W
MAX691ACSE Supervisory, Microprocessor, SO16
Inductor, 100uH, 1.4A, 0.213 OHM, 5.0SP
Inductor, 47uH, 1.6A 0.1 OHM, 10.0 DIA, 5.0SP
Inductor, 100uH, 1.2A, 0.22 OHM, 10.0 DIA, 5.0SP
DINAMAP™ Compact Vital Signs Monitor
Pfx
IND
IND
IND
IND
IND
IND
IND
IND
IND
IND
IND
IND
IND
LAB
LAB
LAB
LAB
MET
MET
MSC
MSC
MSC
MSC
OPT
OPT
OPT
OPT
OPT
OPT
OPT
OPT
OPT
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
GE Medical
Systems. P/N
669221
669222
669223
669224
669225
669226
618111
669231
669232
669227
669233
669234
669228
774172
774173
774175
774176
614207
690173
662208
770130
621262
645189
614215
614208
614210
614211
614212
614213
614214
625111
625112
652386
652387
652388
652389
652390
652391
659227
659228
658125
650258
685632
685633
685634
685635
685636
685637
Description/Nomenclature
Inductor, 100uH, 1.2A, 0.23 OHM, 13.0x9.4, SMT
Inductor, BLM21A05, NSE Suppression, SMT, 0805 Size
Inductor, BLM32A07, NSE Suppression, SMT, 1206 Size
Inductor, BLM41A01, NSE Suppression, SMT, 1206 Size
Choke, 120uH, 1.7A, 0.095 OHM, 16L x 7.0DIA, Axial
Choke, Wide Band 2.5 t, J8FE-1153-NC
Filter, EMI, 28uH, 100 OHM, 100mA, 3 Line
FERRITE, SLEEVE SNAP SS28B2031
FERRITE, SLEEVE SNAP, SS28B2033
Ferrite, CYL, EMI ATTEN, H8FE-1111-NC
FERRITE, EMI ATTEN H8FE-1134-WC W/CLIPS
FERRITE SLEEVE, EMI ATTEN, H8FE-1135-NC
Ferrite, CYL. EMI Atten. H8FE-1137-NC
Tape, Copper, Adhesive, EMI Shielding 25.4W
Tape, Adhesive, Double Sided 25.4W
Tape, Tinned copper, ADH, EMI Shielding
TAPE, ADHESIVE, D/S, 19W
Display, Graphics, LCD MGL(S)-24064
Printer, HTP-4050, 60.0W Paper, 5V
MPX7050GP Gauge, Pressure, Side Port, 0-7.3psi
PAPER, PRINTER PTP ROLL
CONV, AC-DC 90-240V 47-63HZ TO 24VDC 1.5A CON B5
Module, Sp02, Nellcor MP204P
HLMP4719, LED, Yellow, TI-3/4, 5.0 DIA.
HLMP4740 LED, Green, TI-3/4, 5.0 DIA.
HLMP-T200 LED, RED, Rectangular, 5.7x3.2, 2.54SP
HDSP-A153 LED, RED, 7 SEG, COM Cathode, 7.6mm
HDSP-H153 LED, RED, 7 SEG, COM Cathode, 14.2mm
HDSP-5703 LED, YELLOW, 7 SEG, COM Cathode, 14.2mm
MPY54C569, Cell, Photoconductive, CdS, 3.4SP
HCPL2601 Optocoupler, High CMR, SNGL, DP8, GULL
4N35 Optocoupler, GaAs, NPN, SNGL, SO6, GULL
RES, 15RO, 1%, 1/8W, MF, Axial
RES, 1k00, 1%, 1/8, MF, Axial
RES, 3k0, 1%, 1/8W, MF, Axial
RES, 9k1, 1%, 1/8W, MF, Axial
RES, 16k, 1%, 1/8W, MF, Axial
RES, 24k, 1%, 1/8W, MF, AXIAL
RES, 221k, 1%, 1/4W, MF, SMT, 1206 Size
RES, 332k, 1%, 1/4W, MF SMT, 1206 Size
RES, 4R7, 5%, 1/2W, CF, Axial
RES, ZERO OHM 0805 SIZE
RES, 0R0, SMT, 0805 Size
RES, 1RO, 1%, 1/10W, SMT, 0805 Size
RES, 2R7, 1%, 1/10W, SMT, 0805 Size
RES, 10R0, 1%, 1/10W, SMT, 0805 Size
RES, 47R, 1%, 1/10W, SMT, 0805 Size
RES, 100R, 1%, 1/10W, SMT, 0805 Size
71
DINAMAP™ Compact Vital Signs Monitor
Pfx
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
72
GE Medical
Systems. P/N
685638
652392
652396
685640
685641
685642
652393
685643
652394
652397
685645
685646
685647
685648
685649
685650
652395
685651
685652
685653
685654
685655
685656
685657
685658
685659
685660
685661
685662
685663
685664
685665
685666
685667
685668
685669
685670
658671
685672
685673
685674
685675
685676
685677
685678
685679
655137
655135
Description/Nomenclature
RES, 150R, 1%, 1/10W, SMT, 0805 Size
RES, 220R, 1% 1/10W SMT 0805 SIZE
RES, 510R, 1% 1/10W SMT 0805 SIZE
RES, 510R, 1%, 1/10W, SMT, 0805 Size
RES, 750R, 1%, 1/10W, SMT, 0805 Size
RES, 1k00, 1%, 1/10W, SMT, 0805 Size
RES, 1K10, 1% 1/10W SMT 0805 SIZE
RES, 1K2, 1%, 1/10W, SMT, 0805 Size
RES, 1K3, 1% 1/10W SMT 0805 SIZE
RES, 1K8, 1% 1/10W SMT 0805 SIZE
RES, 2K00, 1%, 1/10W, SMT, 0805 Size
RES, 2k2, 1%, 1/10W, SMT, 0805 Size
RES, 3k3, 1% 1/10W, SMT 0805 Size
RES, 4K30, 1%, 1/10W, SMT, 0805 Size
RES, 4K7, 1%, 1/10W, SMT, 0805 Size
RES, 5k1, 1% 1/10W, SMT 0805 Size
RES, 5K6, 1% 1/10W SMT 0805 SIZE
RES, 6k2, 1% 1/10W, SMT 0805 Size
RES, 7k50, 1%, 1/10W, SMT, 0805 Size
RES, 9K1, 1%, 1/10W, SMT, 0805 Size
RES, 10k, 1% 1/10W, SMT 0805 Size
RES, 10k7, 1% 1/10W, SMT 0805 Size
RES, 12K, 1%, 1/10W, SMT, 0805 Size
RES, 13k, 1% 1/10W, SMT 0805 Size
RES, 15k0, 1%, 1/10W, SMT, 0805 Size
RES, 16k9, 1% 1/10W, SMT 0805 Size
RES, 18K, 1%, 1/10W, SMT, 0805 Size
RES, 20K0, 1%, 1/10W, SMT, 0805 Size
RES, 24K, 1%, 1/10W, SMT, 0805 Size
RES, 27K, 1%, 1/10W, SMT, 0805 Size
RES, 30K, 1%, 1/10W, SMT, 0805 Size
RES, 49k9, 1% 1/10W, SMT 0805 Size
RES, 43K, 1%, 1/10W, SMT, 0805 Size
RES, 47K, 1%, 1/10W, SMT, 0805 Size
RES, 51K, 1%, 1/10W, SMT, 0805 Size
RES, 68K, 1%, 1/10W, SMT, 0805 Size
RES, 82K, 1%, 1/10W, SMT, 0805 Size
RES, 100k, 1% 1/10W, SMT 0805 Size
RES, 150K, 1%, 1/10W, SMT, 0805 Size
RES, 200K, 1%, 1/10W, SMT, 0805 Size
RES, 510K, 1%, 1/10W, SMT, 0805 Size
RES, 820K, 1%, 1/10W, SMT, 0805 Size
RES, 1MO, 1% 1/10W, SMT 0805 Size
RES, 1M43, 1%, 1/10W, SMT, 0805 Size
RES, 2M00, 1%, 1/10W, SMT, 0805 Size
RES, 10M0, 1%, 1/10W, SMT, 0805 Size
RES, 3R3, 5%, 3W W/W VIT AXL
RES, OR1, 5%, 2.5W, W/W, Cemented, Axial
DINAMAP™ Compact Vital Signs Monitor
Pfx
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
RES
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCD
GE Medical
Systems. P/N
685680
685681
685682
685683
685684
685685
685686
685687
685688
685689
685690
685691
658692
685693
685694
685695
685696
685697
685698
630301
630302
630303
610145
610146
611149
611150
611151
611152
611153
611154
611155
611156
610147
610148
611157
611158
612152
612149
612150
611159
611160
673120
674102
674142
674143
674144
674126
Description/Nomenclature
RES, 11OR, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 511R, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 549R, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 1k07, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 750R, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 1k21, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 1k78, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 1k82, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 2k15, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 4k99, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 6k81, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 10k0, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 13k7, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 15k0, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 19k6, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 30k1, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 44k2, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 68k1, 0.1% 1/10W, MF, SMT, 0805 Size
RES, 499k, 0.1%, 1/10W, MF, SMT, 0805 Size
POTR, 1k0, 10%, 1/2W, CER, Top ADJ, 20t, 3/8 SQ
POTR, 20k, 10%, 1/2W, CER, Top ADJ, 20t, 3/8 SQ
POTR, 2k2, 20%, 0.15W, Top ADJ, 1t, 10x5 SP
30VF20F Diode, Fast RECVRY, 4.7A, 200V, TO252
DE3L20U Diode, Fast RECVRY, 3A, 200V, E Pack
EC10DS2 Diode, RECTR, 1A, 200V, SMT, 5.0x2.5
DIF20 Diode, RECTR, 1A, 200V, SMT, 5.0x2.5
D2F20 Diode, RECTR, 1.4A, 200V, SMT, 57.6x4.0
NSD03A20 Diode, RECTR, 3A, 200V, SMT, 8.0x4.0
Bar 43, Diode, Silicon, Schottky, 100mA, 30V, SOT23
1N5817M Diode, Schottky, 1A, 14V, MELF, 5.0x2.4
1N5819M Diode, Schottky, 1A, 28V, MELF, 5.0x2.4
LLBAT43 Diode, Silicon, Schottky, 200mA, 30V
SS14 DIODE, SILICON SCHOTTKY, 1A 40V
SS34 DIODE, SILICON SCHOTTKY, 3A 40V
SM4T36C Diode, Suppressor, Transient Volt, SOD6
LL4148 Diode, Silicon, Small Signal, SMT
BZX55C5V1 ZRN, 5V1, 500MW DIODE
BZX84C5V1 Diode, Zener, 5.1V, 350mW, SOT23
BZX84C33 Diode, Zener, 33V, 350mW, SOT23
TL431CLP Diode, Volt REF, ADJ, PRECN, TO92
TICP106M RECTR, Silicon CONT, 2A, 600V, TO92
IMH3A NPN, Base RES, Dual, SO6
2N4401 NPN, TO92
MMBT2222A NPN, SOT23
BC857C NPN, SOT23
BC212 PNP, MED PWR, TO92
MMBT2907A PNP, SOT23
73
DINAMAP™ Compact Vital Signs Monitor
Pfx
SCD
SCD
SCD
SCD
SCD
SCD
SCD
SCR
SCR
SCR
SCR
SCR
SCR
SCR
SDR
SDR
SWS
SWS
SWS
SWS
SWS
TDC
XTL
XTL
XTL
74
CRITIKON U.S.
P/N
676105
676168
676140
676129
676153
676156
676169
719324
719322
719325
721123
719326
719327
719328
600121
600122
665127
665128
665129
666119
648122
662207
609137
609138
609139
Description/Nomenclature
VN10KM NFET, Enhancement Mode, TO237
BUK581-60A NMOS, SOT223
2N7002 NMOS, Enhancement Mode, SOT23
2N7000 NMOS, Enhancement Mode, TO92
Si941ODY NMOS, Enhancement Mode, SO8
Si9955DY NMOS, Enhancement Mode, Dual, SO8
ME4P06F PMOS, TO252
SCREW, M2.5 X 8MM, PAN, SDRV, STL, ZNPL
Screw, M3x6mm, Pan, SDRV, STL, ZnPl
SCREW, M3 X 10MM, PAN, SDRV, STL, ZNPL
Screw, M3x8mm, CHS, SLTD, NYLN
SCREW, M4X6 PAN SDRV STL ZNPL
SCREW, M4X10 PAN SDRV STL ZNPL
SCREW, M3X12MM, INSERT, PSM
Speaker, M/C, 8 OHM, 0.2W, 1.5 DIA (38.1mm)
Sounder, Piezoelectric, 4kHz, 20.00DIA
Switch, SPST, Key, RF15, Non ILLUM, 15.0SQ
Switch, SPST, Key, RF19, Non ILLUM, 19.0SQ
Switch, SPST, Key, RF15, ILLUM, Red, 15.0SQ
Encoder, OPT CPLD, 16 Way, Rotary, W/Pushbutton
Relay, DPDT, 2A, Coil 12V, 400 OHM
MPX7050D, Sensor, Pressure, 0-7.3psi, Case 344
Crystal, 19.6608MHz, HC49/SD3
Crystal, 32.768kHz, 4 PIN, SMT, 9x3
Crystal, 4MHz, HC49, SMT, Short
DINAMAP™ Compact Vital Signs Monitor
6.5
Procedural & Error Alarm Code Table
Alarm
Code
LED
Display
LCD Display
N99
No
change
N99-NIBP
FAILED
Audio Tone &
Volume
Effect of
Alarm
Silence
switch
Effect of
Clear
via
Rotor
Probable Cause
Unable to make an NIBP determination
due to insufficient signal.
N55
N55TIMEOUT:
PRESS
N44
High priority
alarm. Volume
adjustable.
2 minutes
silence
Clear
N33
Determination time exceeds 2 minutes.
Motion artifact.
N44TIMEOUT:
TOTAL
N00
One cuff pressure for > 1 minute. Motion
artifact.
Inflation time > 40 seconds or air leak
detected.
N33TIMEOUT:
INFLATE
Overpressure detected.
N00- OVER
PRESSURE
N12
Systolic
Display
‘N12’
N0 Display
Steady tone,
maximum
volume
No effect
No effect
N13
Systolic
Display
‘N13’
No Display
Steady tone,
maximum
volume
No effect
No effect
P55-SpO2: NO
SIGNAL
High priority
alarm. Volume
adjustable.
2 minutes
silence
Clear
P55
P00
No
change
P00-SpO2: NO
SpO2
SENSOR
E33
E11
E00
E33-Temp:
FAIL
No
change
E11-Temp:
FAIL
E00-Temp:
FAIL
System pressure whilst idle 5 or
15mmHg above ambient for 30 seconds
for a neonate or an adult cuff
respectively
Pump on/off status determined by
measurement of pump current does not
concur with Main board software
expected status. Unit will power down
within 30 seconds of detected failure..
No or very low SpO2 signal. Check or
reposition sensor.
SpO2 sensor not connected. No sensor
code detected. Sensor failure.
Clear
High priority
alarm. Volume
adjustable.
2 minutes
silence
Clear
Clear
Temperature probe not connected or
inoperable.
Predictive temperature loss of tissue
contact > 30 sec.
Removal of probe before temperature
determination complete.
Predictive temperature determination
exceeds 60 sec.
75
DINAMAP™ Compact Vital Signs Monitor
Procedural & Error Alarm Code Table continued
Alarm
Code
LED
Display
LCD Display
Audio Tone &
Volume
No
change
Low Battery
Flashing
battery icon
3 beeps every
10 sec,
adjustable
volume.
Blank
No code
No
change
Blank
Low Battery System
Disabled
Printer - No
Paper
NIBP Range
error
Other:
N,P,E,I,S.
76
Blank
Error code,
description
Steady tone,
maximum
volume
High priority
alarm
Volume
adjustable
High priority
alarm
Volume
adjustable
Steady tone,
maximum
volume
Effect of
Alarm
Silence
switch
2 minutes
silence
Effect of
Clear via
Rotor
Probable Cause
No effect
No effect
No effect
Replace or recharge battery. From
onset of alarm, 5 NIBP
measurements available. Beep
linearly increases rate as battery
discharges.
2 minute
silence
Clear
2 minute
silence
Clear
Replace or recharge battery. NIBP
measurement disabled.
Paper expired or printer door open.
NIBP algorithm returned value
outside specified accuracy rage.
No effect
No effect
Internal system fault
DINAMAP™ Compact Vital Signs Monitor
7 Glossary Of Terms and Abbreviations
AC
ADC
ADU
: Alternating Current
: Analog to Digital Converter
: Analog to Digital Units:Main Board ADC has 4096 steps which equates to 5V full scale (1.2207mV/Bit)
PSU PIC processor ADC has 256 steps which equates to 5V full scale (19.531mV/Bit)
AM
BPM
DC
DMM
FET
FPT
GAL
Hz
ITU
LCD
LED
MAP
NIBP
NiCd
PSU
SPI
TTL
UUT
VAC
VDC
VFC
: Amplitude Modulation
: Beats Per Minute
: Direct Current
: Digital Multi-Meter
: Field Effect Transistor
: Filtered Pressure Transducer
: Gate Array Logic
: Hertz
: Intensive Therapy Unit
: Liquid Crystal Display
: Light Emitting Diode
: Mean Arterial Pressure
: Non Invasive Blood Pressure
: Nickel Cadmium
: Power Supply Unit
: Serial Peripheral Interface
: Transistor to Transistor Logic
: Unit Under Test
: Volts Alternating Current
: Volts Direct Current
: Voltage to Frequency Converter
77
DINAMAP™ Compact Vital Signs Monitor
8 Service Diagrams
This section include the part lists, assembly drawings and circuit diagrams for the following boards for
unit revision levels 02, 04, 06:
I.
II.
III.
IV.
V.
VI.
78
Compact NIBP Monitor Block Diagram 8600EB
Main Board 8610 (8760 on unit revision 06)
Power Supply Board 8620
Display Board 8630
Switch Board 8640
Printer Board 8650