Download Agilent Technologies 2000 X-Series Technical data

Transcript
Agilent InfiniiVision 2000 X-Series Oscilloscopes
Programmer's Guide
9
:CALibrate Commands
Utility commands for viewing calibration status and for starting the user
calibration procedure. See "Introduction to :CALibrate Commands" on
page 207.
Table 59 :CALibrate Commands Summary
Command
Query
Options and Query Returns
n/a
:CALibrate:DATE? (see
page 209)
<return value> ::=
<year>,<month>,<day>; all in NR1
format
:CALibrate:LABel
<string> (see
page 210)
:CALibrate:LABel?
(see page 210)
<string> ::= quoted ASCII string
up to 32 characters
:CALibrate:OUTPut
<signal> (see
page 211)
:CALibrate:OUTPut?
(see page 211)
<signal> ::= {TRIGgers | MASK |
WAVEgen}
n/a
:CALibrate:PROTected?
(see page 212)
{PROTected | UNPRotected}
:CALibrate:STARt (see
page 213)
n/a
n/a
n/a
:CALibrate:STATus?
(see page 214)
<return value> ::=
<status_code>,<status_string>
<status_code> ::= an integer
status code
<status_string> ::= an ASCII
status string
n/a
:CALibrate:TEMPeratur
e? (see page 215)
<return value> ::= degrees C
delta since last cal in NR3
format
n/a
:CALibrate:TIME? (see
page 216)
<return value> ::=
<hours>,<minutes>,<seconds>; all
in NR1 format
Introduction to
:CALibrate
Commands
The CALibrate subsystem provides utility commands for:
s1
207