Download Agilent Technologies 2000 X-Series Technical data

Transcript
20 :MEASure Commands
Table 70 :MEASure Commands Summary (continued)
Command
Query
Options and Query Returns
:MEASure:PREShoot
[<source>] (see
page 357)
:MEASure:PREShoot?
[<source>] (see
page 357)
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= the percent of
preshoot of the selected waveform
in NR3 format
:MEASure:PWIDth
[<source>] (see
page 358)
:MEASure:PWIDth?
[<source>] (see
page 358)
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>} for
DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | FUNCtion | MATH |
WMEMory<r>} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<return_value> ::= width of
positive pulse in seconds in NR3
format
:MEASure:RISetime
[<source>] (see
page 359)
:MEASure:RISetime?
[<source>] (see
page 359)
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= rise time in
seconds in NR3 format
:MEASure:SHOW {1 |
ON} (see page 360)
:MEASure:SHOW? (see
page 360)
{1}
336
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide