28 :TIMebase Commands
:TIMebase:POSition
(see page 790)
Command Syntax
:TIMebase:POSition <pos>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
The :TIMebase:POSition command sets the time interval between the
trigger event and the display reference point on the screen. The display
reference point is either left, right, or center and is set with the
:TIMebase:REFerence command. The maximum position value depends on
the time/division settings.
NOTE
Query Syntax
This command is an alias for the :TIMebase:DELay command.
:TIMebase:POSition?
The :TIMebase:POSition? query returns the current time from the trigger to
the display reference in seconds.
Return Format
<pos><NL>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
See Also
• "Introduction to :TIMebase Commands" on page 578
• ":TIMebase:REFerence" on page 582
• ":TIMebase:RANGe" on page 581
• ":TIMebase:SCALe" on page 583
• ":TIMebase:WINDow:POSition" on page 585
• ":TIMebase:DELay" on page 748