Agilent Technologies 2000 X-Series Technical data Download

Transcript
4
Commands Quick Reference
Table 2
Common (*) Commands Summary (continued)
Command
Query
Options and Query Returns
n/a
*OPT? (see page 133)
<return_value> ::= 0,0,<license
info>
<license info> ::= <All field>,
<reserved>, <MSO>, <reserved>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<reserved>,
<Power Measurements>,
<reserved>, <Segmented Memory>,
<Mask Test>, <reserved>,
<Bandwidth>, <reserved>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<reserved>,
<Waveform Generator>,
<reserved>, <reserved>
<All field> ::= {0 | All}
<reserved> ::= 0
<MSO> ::= {0 | MSO}
<Power Measurements> ::= {0 |
PWR}
<Segmented Memory> ::= {0 | SGM}
<Mask Test> ::= {0 | MASK}
<Bandwidth> ::= {0 | BW10 | BW20}
<Waveform Generator> ::= {0 |
WAVEGEN}
*RCL <value> (see
page 135)
n/a
<value> ::= {0 | 1 | 4 | 5 | 6 |
7 | 8 | 9}
*RST (see page 136)
n/a
See *RST (Reset) (see page 136)
*SAV <value> (see
page 139)
n/a
<value> ::= {0 | 1 | 4 | 5 | 6 |
7 | 8 | 9}
*SRE <mask> (see
page 140)
*SRE? (see page 141)
<mask> ::= sum of all bits that
are set, 0 to 255; an integer in
NR1 format. <mask> ::= following
values:
Bit Weight Name Enables
--- ------ ---- ---------7
128 OPER Operation Status Reg
6
64 ---- (Not used.)
5
32 ESB Event Status Bit
4
16 MAV Message Available
3
8 ---- (Not used.)
2
4 MSG Message
1
2 USR User
0
1 TRG Trigger
66
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide