Download Agilent Technologies 2000 X-Series Technical data

Transcript
4
Commands Quick Reference
Table 3
Root (:) Commands Summary (continued)
Command
Query
Options and Query Returns
:DIGitize
[<source>[,..,<source
>]] (see page 159)
n/a
<source> ::= {CHANnel<n> |
FUNCtion | MATH} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | POD{1 | 2} | BUS{1 |
2} | FUNCtion | MATH} for MSO
models
<source> can be repeated up to 5
times
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
:MTEenable <n> (see
page 161)
:MTEenable? (see
page 161)
<n> ::= 16-bit integer in NR1
format
n/a
:MTERegister[:EVENt]?
(see page 163)
<n> ::= 16-bit integer in NR1
format
:OPEE <n> (see
page 165)
:OPEE? (see page 165)
<n> ::= 15-bit integer in NR1
format
n/a
:OPERregister:CONDiti
on? (see page 167)
<n> ::= 15-bit integer in NR1
format
n/a
:OPERegister[:EVENt]?
(see page 169)
<n> ::= 15-bit integer in NR1
format
:OVLenable <mask>
(see page 171)
:OVLenable? (see
page 172)
<mask> ::= 16-bit integer in NR1
format as shown:
Bit Weight Input
--- ------ ---------10
1024 Ext Trigger Fault
9
512 Channel 4 Fault
8
256 Channel 3 Fault
7
128 Channel 2 Fault
6
64 Channel 1 Fault
4
16 Ext Trigger OVL
3
8 Channel 4 OVL
2
4 Channel 3 OVL
1
2 Channel 2 OVL
0
1 Channel 1 OVL
n/a
:OVLRegister? (see
page 173)
<value> ::= integer in NR1
format. See OVLenable for <value>
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
69