3 - Language Dictionary
CALibrate:CURRent:MEASure
Agilent 6811B, 6812B, 6813B, 6843A Only
This command is used to initiate the calibration of the current metering circuits and the peak current limit
circuits. It can only be used in the calibration mode.
Command Syntax
Parameters
Examples
Related Commands
CALibrate:CURRent:MEASure
None
CAL:CURR:MEAS
CAL:STAT CAL:SAV CAL:LEV
CALibrate:DATA
Phase Selectable
This command is only used in calibration mode. It enters a calibration value that you obtain by reading an
external meter. You must first select a calibration level (with CALibrate:LEVel) for the value being entered.
These constants are not stored in nonvolatile memory until they are saved with CALibrate:SAVE. If
CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous calibration constants are
restored.
Command Syntax
Parameters
Unit
Examples
Related Commands
CALibrate:DATA <NRf>
<external reading>
A (amperes)
CAL:DATA 3222.3 MA
CAL:STAT CAL:SAV
CAL:DATA 5.000
CALibrate:IMPedance
Agilent 6811B, 6812B, 6813B, 6843A Only
This command can only be used in calibration mode. It calibrates the output impedance circuits. The
automatically performs the calibration and stores the impedance constant in nonvolatile memory.
CALibrate:IMPedance is a sequential command that takes several seconds to complete.
Command Syntax
Parameters
Examples
Related Commands
CALibrate:IMPedance
None
CAL:IMP
CAL:STAT CAL:SAV
CALibrate:LEVel
Phase Selectable
This command can only be used in calibration mode. It is used to advance to the next state in the
calibration sequence.
Command Syntax
Parameters
Examples
Related Commands
30
CALibrate:LEVel <level>
P1 | P2 | P3 | P4
CAL:LEV P2
CAL:STAT CAL:SAV