SENSe:SWEep:OFFSet:POINts
This command defines the trigger point relative to the start of the returned data record when an acquire
trigger is used. The values can range from -4095 to 2E9. When the values are negative, the values in the
beginning of the data record represent samples taken prior to the trigger.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
SENSe:SWEep:OFFSet:POINts <NRf+>
4096 through 2E9 | MINimum | MAXimum
0 (zero)
SENS:SWE:OFFS:POIN -2047
SENSe:SWEep:OFFSet:POINts?
<NR3>
SENS:SWE:TINT
MEAS:ARR
SENSe:SWEep:TINTerval
This command defines the time period between samples when voltage and current digitization is
controlled by the acquire trigger sequence. The sample period can be programmed from 25 to 250
microseconds in 25 microsecond increments.
NOTE:
All the MEASure commands use the ACQuire trigger sequence implicitly. These
commands always set the sample period to 25 microseconds.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
SENSe:SWEep:TINTeral <NRf+>
25.037 through 250.37 (microseconds) | MAXimum | MINimum
25.037 µs (Agilent 6814B/6834B)
25.049 µs (Agilent 6811B/6812B/6813B/6843A)
SENS:SWE:TINT 100E-6
SENSe:SWEep:TINTerval?
<NR3>
SENS:SWE:OFFS:POIN
MEAS:ARR
SENSe:WINDow
Phase Selectable
This command sets the window function which is used in harmonic measurements. KBESsel is the
preferred window and should be used for most measurements. RECTangular is available for making
harmonic measurements that comply with regulatory requirements for quasi-stationary harmonics.
When RECTangular is selected, the output frequency is constrained to frequencies that give an integer
number of cycles in the acquired waveform buffers, and the measurement acquisition time is set to 0.1
seconds. Any programmed output frequency will be routed to the closest frequency that has this attribute.
These frequencies are exact multiples of 10.000207Hz
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
SENSe:WINDow[:TYPE] <type>
RECTangular | KBESsel
KBESsel
SENS:WIND KBES
SENSe:WINDow?
<CRD>
61