About this Manual
P/N: 01.54.455422-10
Release Date: November 2011
© Copyright EDAN INSTRUMENTS, INC. 2011. All rights reserved.
Statement
This manual will help you understand the operation and maintenance of the product better. It is
reminded that the product shall be used strictly complying with this manual. User’s operation
failing to comply with this manual may result in malfunction or accident for which EDAN
INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
materials contained in this manual shall not be photocopied, reproduced or translated into other
languages.
Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information are contained in this manual, the user shall
not disclose such information to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of EDAN.
EDAN holds the rights to modify, update, and ultimately explain this manual.
Product Information
Product Name: Electrocardiograph
Model: SE-1200, SE-1200 Express
Responsibility of the Manufacturer
EDAN only considers itself responsible for any effect on safety, reliability and performance of
the equipment if:
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by
persons authorized by EDAN, and
I
The electrical installation of the relevant room complies with national standards, and
The instrument is used in accordance with the instructions for use.
Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other
information to help qualified technician to maintain and repair some parts, which EDAN may
define as user serviceable.
Terms Used in this Manual
This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
NOTE
A NOTE provides useful information regarding a function or a procedure.
II
Table of Contents
Chapter 1 Safety Guidance ........................................................................................................... 1
1.1 Intended Use ........................................................................................................................... 1
1.2 Warnings and Cautions ........................................................................................................... 1
1.2.1 Safety Warnings ............................................................................................................... 2
1.2.2 Li-ion Battery Care Warnings .......................................................................................... 5
1.2.3 General Cautions.............................................................................................................. 5
1.3 List of Symbols....................................................................................................................... 6
Chapter 2 Introduction................................................................................................................ 10
2.1 Top Panel .............................................................................................................................. 11
2.2 Keyboard and Keys .............................................................................................................. 12
2.3 Front Panel............................................................................................................................ 14
2.4 Rear Panel............................................................................................................................. 15
2.5 Right Panel ........................................................................................................................... 15
2.6 Bottom Panel ........................................................................................................................ 19
2.7 Function Features ................................................................................................................. 21
Chapter 3 About the Electrocardiograph Application.............................................................. 22
3.1 Selecting Menu Functions .................................................................................................... 22
3.2 Entering Data........................................................................................................................ 23
3.3 Selecting an Item .................................................................................................................. 24
3.4 About the main screen .......................................................................................................... 25
3.5 About the System Setup screen ............................................................................................ 27
3.6 About the File Manager screen............................................................................................. 29
Chapter 4 Operation Preparations............................................................................................. 32
4.1 Power and Earthing .............................................................................................................. 32
4.2 Loading/Replacing Recorder Paper...................................................................................... 34
4.3 Preparing the Patient............................................................................................................. 38
4.3.1 Instructing the Patient .................................................................................................... 38
4.3.2 Preparing the Skin .......................................................................................................... 38
4.4 Connecting the Patient Cable to the Electrocardiograph and Electrodes ............................. 39
4.5 Attaching Electrodes to the Patient....................................................................................... 40
4.5.1 Lead Placement .............................................................................................................. 40
4.5.2 Electrode Placement....................................................................................................... 42
4.6 Inspection before Power-On ................................................................................................. 44
Chapter 5 Turning on the Electrocardiograph.......................................................................... 46
III
Chapter 6 Entering Patient Information ................................................................................... 47
6.1 Entering Patient ID ............................................................................................................... 47
6.2 Entering Other Information .................................................................................................. 48
6.3 Using a Bar Code Reader ..................................................................................................... 50
Chapter 7 Freezing ECG Waves ................................................................................................. 54
Chapter 8 Printing ECG Reports ............................................................................................... 56
8.1 Auto Mode ............................................................................................................................ 56
8.2 Manual Mode........................................................................................................................ 60
8.3 Rhythm Mode ....................................................................................................................... 62
8.4 R-R Analysis Mode .............................................................................................................. 64
8.5 Copy Printing........................................................................................................................ 66
8.6 ECG Reports......................................................................................................................... 67
8.6.1 ECG Reports in the Auto Mode ..................................................................................... 67
8.6.2 ECG Reports in the Rhythm Mode ................................................................................ 72
8.6.3 ECG Reports in the Manual Mode................................................................................. 73
8.6.4 ECG Reports in the R-R Analysis Mode........................................................................ 74
8.6.5 ECG Reports Printed by the USB Printer ...................................................................... 78
8.6.6 ECG Report of Patient with Pacemaker......................................................................... 79
8.6.7 ECG Report of Nehb Lead............................................................................................. 80
Chapter 9 Transmitting ECG Data Automatically.................................................................... 82
9.1 Transmitting ECG Data to the PC ........................................................................................ 82
9.1.1 Transmitting ECG Data in DAT Format through Serial Cable ...................................... 82
9.1.2 Transmitting ECG Data in DAT Format through Net Cable.......................................... 84
9.1.3 Transmitting ECG Data in SCP/FDA-XML/PDF Format through Net Cable............... 87
9.2 Exporting ECG Data to the U Disk Automatically............................................................... 89
9.3 Real-time Transmission to PC ECG ..................................................................................... 90
Chapter 10 Managing Files ......................................................................................................... 92
10.1 Transmitting Files to the PC ............................................................................................... 94
10.1.1 Transmitting DAT Files through Serial Cable.............................................................. 95
10.1.2 Transmitting DAT Files through Net Cable ................................................................. 96
10.1.3 Transmitting SCP/FDA-XML/PDF Files through Net Cable ...................................... 97
10.2 Exporting Files to the U Disk ............................................................................................. 98
10.3 Deleting Files...................................................................................................................... 99
10.4 Searching Files ................................................................................................................. 100
10.5 Importing Files ................................................................................................................. 101
10.6 Editing Patient Information .............................................................................................. 101
IV
10.7 Printing Files..................................................................................................................... 102
10.8 Previewing a File .............................................................................................................. 102
Chapter 11 System Setup........................................................................................................... 105
11.1 Work Mode Setup ............................................................................................................. 105
11.1.1 Selecting Work Mode ................................................................................................. 106
11.1.2 Selecting Manual Style............................................................................................... 107
11.1.3 Selecting Rhythm Style .............................................................................................. 107
11.1.4 Selecting Auto Display Style...................................................................................... 107
11.1.5 Selecting Sampling Mode .......................................................................................... 108
11.1.6 Setting Interval and Duration ..................................................................................... 109
11.1.7 Selecting Preview Function........................................................................................ 109
11.1.8 Selecting Auto Extend Record ................................................................................... 109
11.2 Filter Setup........................................................................................................................ 109
11.2.1 Setting AC Filter......................................................................................................... 110
11.2.2 Setting DFT Filter ...................................................................................................... 110
11.2.3 Setting EMG Filter ..................................................................................................... 111
11.2.4 Setting Lowpass Filter................................................................................................ 111
11.3 Record Info Setup ............................................................................................................. 112
11.3.1 Setup 1........................................................................................................................ 112
11.3.2 Setup 2........................................................................................................................ 116
11.4 Patient Information Setup ................................................................................................. 117
11.4.1 Selecting First/Last Name / Gender / D.O.B / Height / Weight / BP / Race / Pacemaker
/ Medication / Room No. / Department / Physician / Technician / Ref-Physician................ 118
11.4.2 Selecting ID Mode ..................................................................................................... 118
11.4.3 Selecting ID Hint........................................................................................................ 119
11.4.4 Setting Prompt............................................................................................................ 119
11.4.5 Selecting H/W Unit .................................................................................................... 119
11.4.6 Selecting BP Unit ....................................................................................................... 119
11.4.7 Selecting Next Patient ................................................................................................ 120
11.4.8 Setting the Time to Clear Patient Information ........................................................... 121
11.4.9 Inputting Customized Information ............................................................................. 121
11.5 Transmission Setup........................................................................................................... 121
11.5.1 Setting Device No. ..................................................................................................... 122
11.5.2 Selecting Auto Transmission...................................................................................... 123
11.5.3 Selecting Transmission Mode .................................................................................... 123
11.5.4 Setting FTP Information............................................................................................. 123
11.5.5 Setting IP Addresses ................................................................................................... 123
11.6 Lead Setup ........................................................................................................................ 124
V
11.6.1 Setting Rhythm Lead1/2/3 ......................................................................................... 125
11.6.2 Setting Lead Sequence ............................................................................................... 126
11.6.3 Selecting Nehb ........................................................................................................... 126
11.7 Display&Sound Setup ...................................................................................................... 126
11.7.1 Setting Brightness (Optional)..................................................................................... 127
11.7.2 Selecting Display Colors (Only for SE-1200 Express).............................................. 127
11.7.3 Selecting Antialiasing................................................................................................. 127
11.7.4 Setting Key Volume.................................................................................................... 128
11.7.5 Setting Hint Volume ................................................................................................... 128
11.7.6 Setting QRS Volume .................................................................................................. 128
11.8 Date&Time Setup ............................................................................................................. 128
11.8.1 Setting Date and Time ................................................................................................ 129
11.8.2 Setting Date Mode...................................................................................................... 129
11.8.3 Setting Time Mode ..................................................................................................... 130
11.8.4 Setting Power-Off Time ............................................................................................. 130
11.8.5 Setting LCD Off Time................................................................................................ 130
11.9 File Setup .......................................................................................................................... 130
11.9.1 Selecting Auto Save ................................................................................................... 131
11.9.2 Setting Saving Path .................................................................................................... 131
11.9.3 Setting File Format..................................................................................................... 132
11.9.4 Selecting Deleting Files After Transmission or Export.............................................. 132
11.9.5 Selecting Replacing When Memory is Full ............................................................... 132
11.9.6 Selecting SCP File Compression................................................................................ 132
11.10 System Maintenance Setup............................................................................................. 133
11.10.1 Exporting System Settings to U Disk....................................................................... 133
11.10.2 Loading System Settings from U Disk..................................................................... 134
11.10.3 Backing Up Settings................................................................................................. 134
11.10.4 Loading Backup Settings ......................................................................................... 134
11.10.5 Printing System Settings .......................................................................................... 134
11.10.6 Loading Factory Settings ......................................................................................... 134
11.10.7 Setting System Password ......................................................................................... 140
11.11 Other Setup ..................................................................................................................... 140
11.11.1 Entering Institution................................................................................................... 141
11.11.2 Selecting a Language................................................................................................ 141
11.11.3 Selecting Caps Lock ................................................................................................. 141
11.11.4 Setting External Input............................................................................................... 141
11.11.5 Setting External Output ............................................................................................ 142
11.11.6 Setting Pacemaker Detection Sensitivity ................................................................. 142
VI
Chapter 12 Turning off the Electrocardiograph ..................................................................... 143
Chapter 13 Hint Information.................................................................................................... 144
Chapter 14 Cleaning, Care and Maintenance ......................................................................... 145
14.1 Cleaning............................................................................................................................ 145
14.1.1 Cleaning the Main Unit and the Patient Cable........................................................... 145
14.1.2 Cleaning the Reusable Electrodes.............................................................................. 145
14.1.3 Cleaning the Print Head ............................................................................................. 146
14.2 Disinfection ...................................................................................................................... 146
14.3 Care and Maintenance ...................................................................................................... 147
14.3.1 Recharge and Replacement of Battery ....................................................................... 147
14.3.2 Recorder Paper ........................................................................................................... 148
14.3.3 Visual inspection ........................................................................................................ 149
14.3.4 Maintenance of the Main Unit and the Patient Cable ................................................ 149
Chapter 15 Accessory List ......................................................................................................... 152
Chapter 16 Warranty & Service ............................................................................................... 154
16.1 Warranty............................................................................................................................ 154
16.2 Contact information .......................................................................................................... 154
Appendix 1 Technical Specifications ........................................................................................ 155
A1.1 Safety Specifications ....................................................................................................... 155
A1.2 Environment Specifications............................................................................................. 156
A1.3 Physical Specifications .................................................................................................... 156
A1.4 Power Supply Specifications ........................................................................................... 156
A1.5 Performance Specifications ............................................................................................. 157
Appendix 2 EMC Information.................................................................................................. 160
Appendix 3 Abbreviation........................................................................................................... 165
VII
SE-1200 Series Electrocardiograph User Manual
Safety Guidance
Chapter 1 Safety Guidance
This chapter provides important safety information related to the use of the 12-channel
electrocardiograph.
1.1 Intended Use
The intended use of the 12-channel electrocardiograph is to acquire ECG signals from adult and
pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended
to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The
cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart
disease. However, the interpreted ECG with measurements and interpretive statements is offered
to clinicians on an advisory basis only.
WARNING
1. This equipment is not designed for internal use or direct cardiac application.
2. This equipment is not intended for home use.
3. This equipment is not intended for treatment or monitoring.
4. This equipment is intended for use on adult and pediatric patients only.
5. The results given by the equipment should be examined based on the overall clinical
condition of the patient, and they can not substitute for regular checking.
1.2 Warnings and Cautions
In order to use the electrocardiograph safely and effectively, and avoid possible dangers caused
by improper operation, please read through the user manual and be sure to be familiar with all
functions of the equipment and proper operation procedures before use.
Please pay more attention to the following warning and caution information.
-1-
SE-1200 Series Electrocardiograph User Manual
Safety Guidance
1.2.1 Safety Warnings
WARNING
1. The electrocardiograph is intended to be used by qualified physicians or personnel
professionally trained. They should be familiar with the contents of this user manual
before operation.
2. Only qualified service engineers can install this equipment, and only service
engineers authorized by the manufacturer can open the shell.
3. EXPLOSION HAZARD - Do not use the electrocardiograph in the presence of
flammable anesthetic mixtures with oxygen or other flammable agents.
4. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet.
5. Make sure that the power is turned off and the power cord is disconnected from the
AC socket before connecting or disconnecting equipment. Otherwise, electrical
shock or other injuries may happen to the patient or operator.
6. If the integrity of the external protective conductor is in doubt, the equipment should
be powered by an internal rechargeable battery.
7. Do not use this equipment in the presence of high static electricity or high voltage
equipment which may generate sparks.
8. Only the patient cable and other accessories supplied by the manufacturer can be
used. Or else, the performance and electric shock protection can not be guaranteed.
The electrocardiograph has been safety tested with the recommended accessories,
peripherals and leads.
9. Make sure that all electrodes are connected to the patient correctly before operation.
10. Ensure that the conductive parts of electrodes and associated connectors, including
neutral electrodes, do not come in contact with earth or any other conducting objects.
11. Electrodes with defibrillator protection should be used while defibrillating.
12. If
reusable
electrodes
with
electrode
gel
are
used
during
defibrillation,
electrocardiograph recovery will take more than 10 seconds. The manufacturer
recommends the use of disposable electrodes at all times.
13. Electrodes of dissimilar metals should not be used; otherwise it may cause a high
polarization voltage.
-2-
SE-1200 Series Electrocardiograph User Manual
Safety Guidance
WARNING
14. The disposable electrodes can only be used for one time.
15. No hazard is found when the electrocardiograph is operated with cardiac
pacemakers or other stimulators.
16. Do not touch the patient, bed, table or the equipment while using the ECG together
with a defibrillator.
17. Do not touch accessible parts of non-medical electrical equipment and the patient
simultaneously.
18. The use of equipment that applies high frequency voltages to the patient (including
electrosurgical equipment and some respiration transducers) is not supported and
may produce undesired results. Disconnect the patient data cable from the
electrocardiograph, or detach the leads from the patient prior to performing any
procedure that uses high frequency surgical equipment.
19. If the wireless AP technology is used, in order to maintain compliance with the FCC
RF exposure guidelines, the wireless AP should be installed and operated with a
minimum distance of 20cm between the radiator and the human body. There should
be no shield in or around the room where the wireless AP is used.
20. Fix attention on the examination to avoid missing important ECG waves.
21. SHOCK HAZARD - Don’t connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical
equipment is intended to be supplied by a multiple portable socket-outlet with an
isolation transformer.
22. SHOCK HAZARD - Don’t connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlets supplying the system.
23. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1-1 approved to the electrocardiograph.
The operation or use of non-approved equipment or accessories with the
electrocardiograph is not tested or supported, and electrocardiograph operation and
safety are not guaranteed.
24. Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.).
25. Do not exceed the maximum permitted load when using multiple portable
socket-outlets to supply the system.
-3-
SE-1200 Series Electrocardiograph User Manual
Safety Guidance
WARNING
26. Multiple portable socket-outlets shall not be placed on the floor.
27. Do not use the additional multiple portable socket-outlet or extension cord in
the medical electrical system, unless it’s specified as part of the system by
manufacturer. And the multiple portable socket-outlets provided with the system shall
only be used for supplying power to equipment which is intended to form part of the
system.
28. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data
processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
Therefore anybody, who connects additional equipment to the signal input or output
connector to configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC/EN 60601-1-1. If in
doubt, consult our technical service department or your local distributor.
29. Parts and accessories used must meet the requirements of the applicable IEC/EN
601 series safety standards, and/or the system configuration must meet the
requirements of the IEC/EN 60601-1-1 medical electrical systems standard.
30. Connecting any accessory (such as external printer) or other device (such as the
computer) to this electrocardiograph makes a medical system. In that case,
additional safety measures should be taken during installation of the system, and the
system shall provide:
a) Within the patient environment, a level of safety comparable to that provided by
medical electrical equipment complying with IEC/EN 60601-1, and
b) Outside the patient environment, the level of safety appropriate for non-medical
electrical equipment complying with other IEC or ISO safety standards.
31. If multiple instruments are connected to a patient, the sum of the leakage currents
may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard.
Consult your service personnel.
32. The potential equalization bar can be connected to that of other equipment when
necessary. Make sure that all these devices are connected to the potential
equalization terminal.
-4-
SE-1200 Series Electrocardiograph User Manual
Safety Guidance
1.2.2 Li-ion Battery Care Warnings
WARNING
1. Improper operation may cause the internal li-ion battery (hereinafter called battery) to
be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It
is necessary to read the user manual carefully and pay more attention to warning
messages.
2. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and
specification should be used.
3. DANGER OF EXPLOSION -- Do not reverse the anode and the cathode when
installing the battery.
4. Do not heat or splash the battery or throw it into fire or water.
5. Do not destroy the battery; do not pierce battery with a sharp object such as a needle;
do not hit with a hammer, step on or throw or drop to cause strong shock; do not
disassemble or modify the battery.
6. When leakage or foul smell is found, stop using the battery immediately. If your skin
or cloth comes into contact with the leakage liquid, cleanse it with clean water at once.
If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with
clean water first and go to see a doctor immediately.
7. Properly dispose of or recycle the depleted battery according to local regulations.
8. Only when the device is off can the battery be installed or removed.
9. Remove the battery from the electrocardiograph when the electrocardiograph is not
used for a long time.
10. If the battery is stored alone and not used for a long time, we recommend that the
battery should be charged at least once every 6 months to prevent overdischarge.
1.2.3 General Cautions
CAUTION
1. Avoid liquid splash and excessive temperature. The temperature must be kept
between 5 ºC and 40 ºC during operation, and it should be kept between -20 ºC and
55 ºC during transportation and storage.
-5-
SE-1200 Series Electrocardiograph User Manual
Safety Guidance
CAUTION
2. Do not use the equipment in a dusty environment with bad ventilation or in the
presence of corrosive.
3. Make sure that there is no intense electromagnetic interference source around the
equipment, such as radio transmitters or mobile phones etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment etc. is likely to bring electromagnetic
interference.
4. Ruptured fuse must only be replaced with that of the same type and rating as the
original.
5. The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose of
them together with house-hold garbage. At the end of their lives hand the batteries
over to the applicable collection points for the recycling of waste batteries. For more
detailed information about recycling of this product or battery, please contact your
local Civic Office, or the shop where you purchased the product.
6. Federal (U.S.) law restricts this device to sale by or on the order of a physician.
1.3 List of Symbols
External output
External input
Equipment or part of CF type with defibrillator proof
Caution
Consult Instructions for Use
-6-
SE-1200 Series Electrocardiograph User Manual
Potential equalization
Mains supply
USB socket
Battery indicator
Battery recharging indicator
Delete key
Enter key
Esc key/Reset key
Shift key
Fn key
Power On/Off key
COPY key
MODE key
SLEEP/WAKE UP key
-7-
Safety Guidance
SE-1200 Series Electrocardiograph User Manual
Safety Guidance
START/STOP key
Tab key/Feed paper key
UP/DOWN/LEFT/RIGHT Arrow key
Recycle

Part Number

Serial Number
Date of Manufacture
Manufacturer
Authorized Representative in the European Community
The symbol indicates that the device complies with the
European Council Directive 93/42/EEC concerning medical
devices.
With respect to electrical shock, fire and mechanical hazards
only in accordance with UL 60601-1and CAN/CSA C22.2
No. 601.1
-8-
SE-1200 Series Electrocardiograph User Manual
Rx only (U.S.)
Safety Guidance
Federal (US) law restricts this device to sale by or on the
order of a physician.
It indicates that the device should be sent to the special
agencies according to local regulations for separate collection
after its useful life.
-9-
SE-1200 Series Electrocardiograph User Manual
Introduction
Chapter 2 Introduction
The 12-channel electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays
the operation menu, ECG parameters as well as electrocardiograms.
The 12-channel ECG waves can be viewed on the LCD screen and printed out by using a
high-quality thermal recorder. The sampled ECG data can be saved, transmitted and exported.
The manual, auto, rhythm, R-R analysis or off mode can be chosen freely.
The 12-channel electrocardiograph can be powered by the mains supply or battery.
With a high resolution thermal recorder, a 32-bit processor and a large-capacity memorizer, the
12-channel electrocardiograph has advanced performance and high reliability. The compact size
makes it suitable for clinic and hospital uses.
The 12-channel electrocardiograph has two models: SE-1200 and SE-1200 Express.
SE-1200 adopts 320×240 dot single color LCD screen; SE-1200 Express adopts 800×600
multicolor LCD screen.
Configuration: main unit, power cord, patient cable, chest electrodes, limb electrodes, disposable
electrodes, alligator clips, thermal recorder paper, fuses, battery.
NOTE: The pictures and windows in this manual are for reference only.
- 10 -
SE-1200 Series Electrocardiograph User Manual
Introduction
2.1 Top Panel
Indicators
LCD Screen
Recorder
Keyboard
Figure 2-1 SE-1200 Express
Indicators
LCD Screen
Recorder
Keyboard
Figure 2-2 SE-1200
- 11 -
SE-1200 Series Electrocardiograph User Manual
Introduction
2.2 Keyboard and Keys
A
B
F
C
G
H
D
E
I
J
L
K
M
N
Figure 2-3 SE-1200 Express/SE-1200 Keyboard
Name
Explanation
A
Function Key
Selecting menu functions on the screen
B
Delete
Deleting characters
C
Enter
Confirming operation
Moving the cursor
Pressing Tab can move the cursor forward, and pressing
Shift + Tab can move the cursor backward.
D
Tab/Feed paper
E
Fn
F
Spacebar
G
Shift
Feed paper: when the main screen is displayed, if Paper
Marker is set to Yes, you can press the Tab key to
advance the recorder paper to the next black marker; if
Paper Marker is set to No, you can press the Tab key to
advance the paper for 2.5cm. Press the Tab key again to
stop advancing the paper.
Inputting special characters. Press Fn + a to type è.
Adding a space between typed
selecting/deselecting a checkbox
characters
Pressing Shift + Tab can move the cursor backward.
- 12 -
or
SE-1200 Series Electrocardiograph User Manual
Introduction
Press Shift and a number key to input the special character
in the top right corner of the key.
If Caps Lock is set to Off, pressing Shift + P can type a
capital P.
If Caps Lock is set to On, pressing Shift + P can type a
lowercase p.
Canceling operation
H
Esc/Reset
I
Power On/Off
NOTE: A large polarization voltage may cause
baseline drift. On the main screen, pressing
the ESC key can decrease the polarization
voltage and draw the baseline to zero quickly.
Power-on/Power-off
Press this key to select a working mode among the auto,
manual, rhythm, R-R analysis, and off modes.
J
MODE
K
SLEEP/WAKE UP
L
COPY
M
START/STOP
NOTE: Only if a working mode is ticked in the Work
Mode Setup window, can the working mode
be selected by pressing the MODE key when
the main screen is displayed.
Resting/wakening the electrocardiograph
In the auto mode, pressing the Copy key can print the
ECG report which was printed out last time.
Start/Stop printing reports
Moving the cursor (Up, Down, Left, Right)
N
Arrow Keys
In the manual mode, press the Up or Down arrow to
switch the lead groups.
- 13 -
SE-1200 Series Electrocardiograph User Manual
Introduction
2.3 Front Panel
Figure 2-4 SE-1200 Express Front Panel
Figure 2-5 SE-1200 Front Panel
Symbol
A
B
C
Name
Explanation
Mains supply
indicator
When the device is powered by the mains supply,
this indicator is lit.
Battery indicator
When the device is powered by battery, this
indicator is lit.
Battery recharging
indicator
When the battery is being recharged, this indicator
is lit.
- 14 -
SE-1200 Series Electrocardiograph User Manual
Introduction
2.4 Rear Panel
Figure 2-6 SE-1200/SE-1200 Express Rear Panel
Name
A
Explanation
Mains Supply Socket
AC SOURCE: alternating current supply socket
Potential Equalization
Conductor
Potential equalization conductor provides a connection
between the unit and the potential equalization bus bar
of the electrical installation.
B
C
Handle
Part for people to hold
D
Heat Emission Hole
Path for internal heat emission
2.5 Right Panel
C
A
D
B
E
G
F
Figure 2-7 SE-1200 Express/SE-1200 Right Panel
- 15 -
SE-1200 Series Electrocardiograph User Manual
Introduction
Name
Explanation
A
Patient Cable Socket
Connecting to the patient cable
B
Serial Port 1
Connecting to a PC
C
USB Socket 1 (Optional)
Standard USB socket, connecting to a PC
USB Socket 2 (Optional)
Standard USB socket, connecting to a U disk, a
bar code reader or a USB printer recommended
by the manufacturer
E
External Input / Output Socket
Connecting to the external signal device
F
Serial Port 2
Reserved
G
Net port
Standard net port, connecting to a PC
D
1) Patient Cable Socket
: Applied part of type CF with defibrillator proof
: Caution
Definitions of corresponding pins:
Pin
Signal
Pin
Signal
Pin
Signal
1
C2 / V2
6
SH
11
F / LL
2
C3 / V3
7
NC
12
C1 / V1
3
C4 / V4
8
NC
13
or NC
C1 / V1
4
C5 / V5
9
R / RA
14
RF (N) /RL
or NC
5
C6 / V6
10
L / LA
15
RF (N) / RL
NOTE: The left side of “/” is European standard, and the right side is American standard.
- 16 -
SE-1200 Series Electrocardiograph User Manual
Introduction
2) Serial Port 1
WARNING
The isolated intensity of serial port 1 is 1500V AC and the maximum voltage applied
should not exceed +15V DC.
Definitions of corresponding pins:
Pin
Signal
Pin
Signal
Pin
Signal
1
NC
4
NC
7
NC
2
RxD (input)
5
GND
8
NC
3
TxD (output)
6
NC
9
NC
3) Serial Port 2
6
8
2
7
1
5
3
4
Definitions of corresponding pins:
Pin
Signal
Pin
Signal
1
TxD (output)
5
NC
2
RxD (input)
6
NC
3
NC
7
GND
4
NC
8
GND
- 17 -
SE-1200 Series Electrocardiograph User Manual
Introduction
4) External Input/Output Socket
Definitions of corresponding pins:
Pin
Signal
Pin
Signal
1
GND
4
GND
2
GND
5
ECG Signal (input)
3
GND
6
ECG Signal (output)
5) USB Socket 1/USB Socket 2 (Optional)
CAUTION
Only the USB equipment recommended by the manufacturer can be connected to the
USB interface.
Definitions of corresponding pins:
Pin
Signal
Pin
Signal
1
+5V
3
D+
2
D-
4
GND
WARNING
1. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment).
- 18 -
SE-1200 Series Electrocardiograph User Manual
Introduction
WARNING
Furthermore all configuration shall comply with the valid version of the standard
IEC/EN 60601-1-1. Therefore anybody, who connects additional equipment to the
signal input or output connector to configure a medical system, must make sure that it
complies with the requirements of the valid version of the system standard IEC/EN
60601-1-1. If in doubt, consult our technical service department or your local
distributor.
2. If multiple instruments are connected to a patient, the sum of the leakage currents may
exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult
your service personnel.
2.6 Bottom Panel
D
C
D
B
E
A
Figure 2-8 SE-1200 Express/SE-1200 Bottom Panel
Name
Explanation
A
Speaker Hole
Path for sound from speaker
B
Battery Compartment
Compartment for battery
C
Fuse
The specification is: T1AL250VP Ø5×20
- 19 -
SE-1200 Series Electrocardiograph User Manual
D
Heat Emission Hole
Path for internal heat emission
E
Label
Position for product information label
Introduction
1) Battery Compartment
The rated voltage and the rated capacity of the battery pack are as follows:
SE-1200 Express: Rated Voltage: 14.8V; Rated Capacity: 4400mAh
SE-1200: Rated Voltage: 14.8V; Rated Capacity: 2200mAh
WARNING
1. Improper operation may cause the battery to be hot, ignited or exploded, and it may
lead to the decrease of the battery capacity. Therefore, it is necessary to read the
user manual carefully and pay more attention to warning messages.
2. When leakage or foul smell is found, stop using the battery immediately. If your skin
or cloth comes into contact with the leakage liquid, cleanse it with clean water at once.
If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with
clean water first and go to see a doctor immediately.
3. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and
specification must be used.
4. Only when the device is off can the battery be installed or removed.
NOTE: If the battery has not been used for two months or more, you should recharge it
before using it again.
2) Fuse
There are two fuses of the same specification installed on the bottom of the main unit. The
specification is: T1AL250VP Ø5×20.
WARNING
Ruptured fuses must only be replaced with those of the same type and rating as the
original.
- 20 -
SE-1200 Series Electrocardiograph User Manual
Introduction
2.7 Function Features
♦
Supporting multi-language
♦
Full alphanumeric keyboard (For SE-1200 Express, touch screen is available)
♦
Battery with large capacity
♦
ECG signals of 12 leads are gathered and amplified simultaneously, 12-channel waves
are displayed and recorded simultaneously
♦
High resolution thermal recorder, recording frequency response ≤150Hz
♦
Supporting external USB printer
♦
Flexible printing formats
♦
The auto, manual, rhythm, R-R analysis and off modes can be chosen freely
♦
Automatic baseline adjustment for optimal printing
♦
Convenient operation of system setup and file management
♦
Multiple file formats: DAT/SCP/FDA-XML/PDF
♦
Measurement function and interpretation function
♦
Hint information of lead off, lack of paper, low battery capacity etc.
♦
Supporting bar code reader
♦
ECG data can be transmitted to the PC software through the serial cable, net cable, or
wireless AP (optional).
♦
Real-time transmission to PC ECG
- 21 -
SE-1200 Series Electrocardiograph User Manual
About the Electrocardiograph Application
Chapter 3 About the Electrocardiograph
Application
The following sections provide an overview of the main operations and functions.
You can operate the electrocardiograph by using the keyboard. For SE-1200 Express, you can
also carry out the operations by using the touch screen.
CAUTION
Do not touch the LCD screen with sharp things such as pencils or pens; otherwise, it will
be damaged.
3.1 Selecting Menu Functions
To select Patient, press the function key F1 below Patient.
To select Setup, press the function key F6 below Setup.
- 22 -
SE-1200 Series Electrocardiograph User Manual
About the Electrocardiograph Application
3.2 Entering Data
1. When the main screen is displayed, press the function key F1 below Patient to open the
Patient Information window.
2. Press Tab or Shift + Tab to move the cursor to the Name textbox.
3. Press
on the keyboard to erase the typed information.
4. Press the letter or number keys on the keyboard to input patient name. Press Fn and a letter
key to input the special character in the top right corner of the key. Press Shift
and a
number key to input the special character in the top right corner of the key. For example, press
Fn + a to input è, and press Shift + 3 to input #. If Caps Lock is set to Off, pressing Shift
and a letter key can input a capital letter. If Caps Lock is set to On, pressing Shift
and a letter key can input a lowercase letter.
5. Press
and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the OK button,
to confirm.
6. Press ESC to cancel the operation, or press Tab or Shift + Tab to move the cursor to the
Cancel button, and then press
to cancel the operation.
- 23 -
SE-1200 Series Electrocardiograph User Manual
About the Electrocardiograph Application
3.3 Selecting an Item
When the main screen is displayed, press the function key F6 below Setup to open the System
Setup window. The cursor is on Work Mode. Press
to open the Work Mode Setup
window.
1. Press Tab or Shift + Tab to move the cursor to R-R, press Spacebar to select it, and then a
check mark √ appears in the box before R-R.
2. Press Tab or Shift + Tab to move the cursor to the Rhythm Style list box. Press the Up or
Down arrow to highlight Single Lead or Three Leads.
3. Press
and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the OK button,
to confirm.
4. Press ESC to cancel the operation, or press Tab or Shift + Tab to move the cursor to the
Cancel button, and then press
to cancel the operation.
- 24 -
SE-1200 Series Electrocardiograph User Manual
About the Electrocardiograph Application
3.4 About the main screen
After you turn on the device, the main screen pops up.
Figure 3-1 SE-1200 Express Main Screen
B
D
F
E
C
G
H
I
J
A
R
K
Q
P
O
N
M
L
Figure 3-2 SE-1200 Main Screen
- 25 -
SE-1200 Series Electrocardiograph User Manual
A
About the Electrocardiograph Application
Name
Explanation
Name
Patient Name: within 60 ASCII characters or the equivalent
number of other characters that can be supported by equivalent
memory used by 60 ASCII characters
When ID Mode is set to Manual, the patient ID is within 30
ASCII characters.
B
ID
When ID Mode is set to Auto, the patient ID is 0~1999, 999, 999.
When ID Mode is set to Time, the patient ID is within 10
characters.
Patient Age
C
Age
D
Gender
The patient age value and the age unit can be set in the Patient
Information window.
Patient Gender (Male/Female/Empty)
Hint Information 1
Including DEMO, Module Error, Overload, Lead Name (When the
leads are off, the lead names will be shown in black on a yellow
background in SE-1200 Express, while in SE-1200, they will be
shown in black on a white background.)
F
Hint Information 2
Including No Paper, Paper Error, Battery Weak, Sampling,
Analyzing, Recording, Testing, Learning, Transmitting, Transmit
Fail, Detecting, Memory Full, U Disk, USB Printer, Lead Off, USB
Scanner
G
Work Mode
Manual, Auto, Rhythm, R-R Analysis or Off
H
Heart Rate
Actual Heart Rate
I
Current Time
J
Battery Symbol
K
File
Press the function key F7 below File to open the File Manager
screen. For details, please refer to Chapter 10, “Managing Files”.
Setup
Press the function key F6 below Setup to display the System
Setup screen. For details, please refer to Chapter 11, “System
Setup”.
E
L
Current examination time. Refer to Section 11.8.1, “Setting Date
and Time”.
Identify the current capacity of the battery
EMG Filter: 25Hz, 35Hz or 45Hz
M
Filter
Lowpass Filter: 75Hz, 100Hz or 150Hz
NOTE: This setup can’t be saved if it is modified on the main
screen.
- 26 -
SE-1200 Series Electrocardiograph User Manual
N
O
Gain
Speed
About the Electrocardiograph Application
Gain: 2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV, 10/5
mm/mV or AGC
NOTE: This setup can’t be saved if it is modified on the main
screen.
In the manual mode, you can set Speed to 5mm/s, 6.25mm/s,
10mm/s, 12.5mm/s, 25mm/s or 50mm/s. In the auto and rhythm
modes, only 25mm/s and 50mm/s are available. In the R-R
analysis mode, only 25mm/s is available.
NOTE: This setup can’t be saved if it is modified on the main
screen.
Freeze
Freezing ECG waves. For details, please refer to Chapter 7,
“Freezing ECG Waves”.
Q
Patient
Press the function key F1 below Patient to open the Patient
Information window. For details, please refer to Chapter 6,
“Entering Patient Information”.
R
ECG waveform
P
Display ECG waveform
3.5 About the System Setup screen
When the main screen is displayed, press the function key F6 below Setup to display the System
Setup screen.
NOTE: If you set the system password in the System Maintenance window, you need
enter the password before opening the System Setup screen. For details, refer
to Section 11.10.7, “Setting System Password”.
- 27 -
SE-1200 Series Electrocardiograph User Manual
About the Electrocardiograph Application
On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to
move the cursor. When the cursor is on an item, press
to open the setup window of the
item.
Take the Work Mode Setup window for example:
Check
Box
Textbox
Dropdown
List
Box
1. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor among different
check boxes. Press Spacebar to select a check box, and a check mark √ appears in the box.
- 28 -
SE-1200 Series Electrocardiograph User Manual
About the Electrocardiograph Application
2. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a list box or
dropdown. Press the Up or Down arrow to highlight an option.
3. In the Work Mode Setup window, press Tab or Shift + Tab to move the cursor to a textbox.
Fill in the selected textbox.
4. Press
and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the OK button,
to confirm.
5. Press ESC to cancel the operation, or press Tab or Shift + Tab to move the cursor to the
Cancel button, and then press
to cancel the operation.
3.6 About the File Manager screen
When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1. Press the Up or Down arrow to highlight a file on the File Manager screen1,
and then press the function key F4 below Select to select the file and display the File Manager
screen2. Then press the function key F7 below Return to return to the File Manager screen1.
Press the function key F7 below Return on the File Manager screen1 to return to the main
screen.
- 29 -
SE-1200 Series Electrocardiograph User Manual
A
B
C
D
About the Electrocardiograph Application
E
F
G
M
N
Figure 3-3 File Manager screen1
H
I
J
K
L
Figure 3-4 File Manager screen2
- 30 -
SE-1200 Series Electrocardiograph User Manual
About the Electrocardiograph Application
Name
Explanation
A
Trans All
Press the function key F1 below Trans All to transmit all the files
to the PC.
B
Export All
Press the function key F2 below Export All to export all the files
from the electrocardiograph to the U disk.
C
Del All
Press the function key F3 below Del All to delete all the files from
the electrocardiograph.
D
Select
Press the Up or Down arrow to highlight a file on the File Manager
screen1, and then press the function key F4 below Select to select
the file and display the File Manager screen2.
E
Search
Press the function key F5 below Search to open the SearchInfo
Setup window.
F
Import
Press the function key F6 below Import to import files from the U
disk to the electrocardiograph.
G
Return
Press the function key F7 below Return to return to the main
screen.
H
Edit
Press the function key F1 below Edit to open the Patient
Information window. Then you can edit the patient information.
I
Record
J
Trans
Press the function key F3 below Trans to transmit the selected file
to the PC.
K
Export
Press the function key F4 below Export to export the selected file
from the electrocardiograph to the U disk.
L
Delete
Press the function key F5 below Delete to delete the selected file
from the electrocardiograph.
M
Preview
Press the function key F6 below Preview to open the file preview
screen.
N
Return
Press the function key F7 below Return to return to the File
Manager screen1.
Press the function key F2 below Record to print the selected file.
- 31 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
Chapter 4 Operation Preparations
WARNING
Before use, the equipment, patient cable and electrodes should be checked. Replace
them if there is any evident defectiveness or aging which may impair the safety or the
performance, and make sure that the equipment is in proper working condition.
4.1 Power and Earthing
WARNING
If the integrity of the external protective conductor is in doubt, the equipment should be
powered by battery.
Power Supply
The electrocardiograph can be powered by either the mains supply or battery.
♦ Mains Supply
The mains socket is on the rear panel of the unit. If the mains supply is used, connect the
power cord to the socket first, and then connect the power cord to the hospital grade outlet.
Operating voltage:
100V-240V~
Operating frequency:
50Hz/60Hz
Input current:
0.9-0.4A
Make sure that the mains supply meets the above requirements before power-on, and then
press
on the keyboard to power on the unit. The mains supply indicator ( ) is lit.
If the battery is weak when the mains supply is used, it will be recharged automatically at
the same time. Both the mains supply indicator ( ) and the battery recharging indicator
(
) will be lit.
- 32 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
♦ Battery
When the battery is used, turn on the unit by pressing
battery indicator (
on the keyboard. Then the
) will be lit and the battery symbol will be displayed on the LCD
screen. Because of the consumption during the storage and transport course, the battery
capacity may not be full. If the symbol
and the hint information Battery Weak are
displayed, which means the battery capacity is low, please recharge the battery first.
CAUTION
1. If the electrocardiograph is turned off automatically because of low battery capacity,
the settings may not be saved.
2. If the battery has been fully charged and requires recharging after printing only a few
ECGs, consider replacement.
3. The use of electrocardiograph accessories (such as barcode reader) will deplete
battery power at a faster rate. The battery will require more frequent charging if these
accessories are used with the electrocardiograph.
When the battery is fully charged, SE-1200 can work normally about 4 hours, and print
about 300 ECG reports of 3×4+1R in the auto mode; SE-1200 Express can work normally
about 5 hours, and print about 350 ECG reports of 3×4+1R in the auto mode.
Please refer to the maintenance section for how to recharge the battery. When the battery is
being recharged, the electrocardiograph can be powered by the mains supply.
WARNING
Potential equalization conductor of the unit should be connected to the potential
equalization bus bar of the electrical installation when necessary.
- 33 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
4.2 Loading/Replacing Recorder Paper
Two kinds of recorder paper can be used. One is the folded thermal paper, and the other is the
rolled thermal paper.
NOTE:
1. When the folded thermal paper is used, the paper roller is unnecessary and must be
taken out.
2. When using the paper of 216mm in width, the two movable parts should be removed.
3. The exit edge can help you tear the recorder paper.
CAUTION
Make sure that the recorder paper, especially the rolled paper, is installed in the center of
the recorder, and the paper edge is parallel with the casing edge in the direction of
advancing paper, in order to avoid paper deviation or damage to the paper edge.
When the recorder paper runs out or is not loaded, the hint message No Paper will appear on the
screen. Then you should load or replace the recorder paper immediately.
Movable
Groove
Part 1
Paper Tray
Exit Edge
(Edge for
tearing paper)
Casing Button
Exit
Paper Roller
Casing
- 34 -
Movable
Part 2
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
Loading/Replacing Process of Folded Paper:
1) Press the casing button downwards and remove the casing to open the recorder.
2) Remove the remainder paper from the paper tray if necessary.
3) Take off the wrapper of the new folded paper, and then put it in the paper tray.
4) Pull the paper out with the grid side facing the thermal print head, and replace the casing
on the recorder.
- 35 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
5) Press down the recorder casing firmly.
6) To select Paper Marker.
Press
on the keyboard to turn on the unit. After the main screen is displayed, press
the function key F6 below Setup to display the System Setup screen. Press Tab or Shift
+ Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on
Record Info on the System Setup screen, press
to open the Record Info
Setup window. Press Tab or Shift + Tab to move the cursor to the Paper Marker list
box, and then press the Up or Down arrow to highlight No or Yes. Press
to
confirm, and then the System Setup screen appears. Press ESC to return to the main
screen.
7) Advance the recorder paper.
When the main screen is displayed, if Paper Marker is set to Yes, you can press
to advance the recorder paper to the next black marker; if Paper Marker is
set to No, you can press
to advance the paper for 2.5cm. Press
again to stop advancing the paper.
Loading/Replacing Process of Rolled Paper:
1) Press the casing button downwards and remove the casing to open the recorder.
2) Take out the paper roller, and remove remainder paper from the roller if necessary.
3) Take off the wrapper of the new thermal paper roll, and then put the paper roll through
the roller.
- 36 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
4) Place the paper and the roller gently in the recorder with the roller pin clicking into the
groove.
Groove
5) Pull the paper out with the grid side facing the thermal print head, and replace the casing
on the recorder.
6) Press down the recorder casing firmly.
7) To select Paper Marker.
Press
on the keyboard to turn on the unit. After the main screen is displayed, press
the function key F6 below Setup to display the System Setup screen. Press Tab or Shift
+ Tab, or the Up, Down, Left or Right arrow to move the cursor. When the cursor is on
Record Info on the System Setup screen, press
to open the Record Info
Setup window. Press Tab or Shift + Tab to move the cursor to the Paper Marker list
box, and then press the Up or Down arrow to highlight No. Press
to confirm,
and then the System Setup screen appears. Press ESC to return to the main screen.
- 37 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
8) Advance the recorder paper.
When the main screen is displayed, you can press
2.5cm. Press
to advance the paper for
again to stop advancing the paper.
4.3 Preparing the Patient
4.3.1 Instructing the Patient
Before attaching the electrodes, greet the patient and explain the procedure. Explaining the
procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless.
Privacy is important for relaxation. When possible, prepare the patient in a quiet room or area
where others can’t see the patient. Make sure that the patient is comfortable. The more relaxed
the patient is, the less the ECG will be affected by noise.
4.3.2 Preparing the Skin
Thorough skin preparation is very important. The skin is a poor conductor of electricity and
frequently creates artifacts that distort the ECG signals. By performing methodical skin
preparation, you can greatly reduce the possibility of noise caused by muscle tremor and baseline
drift, ensuring high-quality ECG waves. There is natural resistance on the skin surface due to dry,
dead epidermal cells, oils and dirt.
To Prepare the Skin
1. Shave hair from electrode sites, if necessary. Excessive hair prevents a good connection.
2. Wash the area thoroughly with soap and water.
3. Dry the skin with a gauze pad to increase capillary blood flow to the tissues and to remove the
dead, dry skin cells and oils.
- 38 -
SE-1200 Series Electrocardiograph User Manual
4.4
Connecting
the
Patient
Electrocardiograph and Electrodes
Operation Preparations
Cable
to
the
WARNING
The performance and electric shock protection can be guaranteed only if the original
patient cable and electrodes of the manufacturer are used.
The patient cable includes the main cable and lead wires which can be connected to electrodes.
Connecting to
Lead Wires
Electrocardiograph
Main Cable
Connecting
to
Electrodes
1. Connecting the Patient Cable to the Electrocardiograph
Connect the patient cable to the patient cable socket on the right side of the main unit, and then
secure them with two screws.
2. Connecting the Patient Cable to Electrodes
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the
reusable electrodes or the alligator clips. Firmly attach them.
- 39 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
4.5 Attaching Electrodes to the Patient
4.5.1 Lead Placement
4.5.1.1 Standard 12-Lead Placement
The identifiers and color codes of electrode connectors used comply with IEC/EN requirements.
In order to avoid incorrect connection, the identifiers and color codes are specified in Table 4-1.
Moreover the equivalent codes according to European requirements are given in Table 4-1 too.
Table 4-1 Electrode Connectors and Their Identifiers and Color Codes
European
American
Electrode Connectors
Identifier
Color Code
Identifier
Color Code
Right arm/Right deltoid
R
Red
RA
White
Left arm/Left deltoid
L
Yellow
LA
Black
N or RF
Black
RL
Green
F
Green
LL
Red
Chest 1
C1
White/Red
V1
Brown/Red
Chest 2
C2
White/Yellow
V2
Brown/Yellow
Chest 3
C3
White/Green
V3
Brown/Green
Chest 4
C4
White/Brown
V4
Brown/Blue
Chest 5
C5
White/Black
V5
Brown/Orange
Chest 6
C6
White/Violet
V6
Brown/Violet
Right leg/Upper leg as
close to torso as possible
Left leg/Upper leg as
close to torso as possible
As the following figure shows, the positions of chest electrodes on the body surface are
C1: Fourth intercostal space at the right border of the sternum
C2: Fourth intercostal space at the left border of the sternum
C3: Fifth rib between C2 and C4
C4: Fifth intercostal space on the left midclavicular line
C5: Left anterior axillary line at the horizontal level of C4
C6: Left midaxillary line at the horizontal level of C4
- 40 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
4.5.1.2 NEHB Placement
The electrodes’ positions on the body surface are shown in the following table and figure.
American European
Electrode Placement
Label
Label
A1
Nst
Attachment point of the second rib to the right sternal edge
A2
Nax
Fifth intercostal space on the left posterior axillary line
V4
Nap
Left mid-clavicular line in the fifth intercostal space
RA
R
Right arm
LA
L
Left arm
RL
N or RF
Right leg
LL
F
Left leg
- 41 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
4.5.2 Electrode Placement
Two kinds of electrode can be used, one is the reusable electrode, and the other is the disposable
electrode.
4.5.2.1 Reusable Electrodes
Reusable electrodes include limb electrodes and chest electrodes, as the following figures show.
Suction Bulb
Connecting to a Lead Wire
Metal Cup
Chest Electrode
Connecting to a Lead Wire
Reed
Clamp
Limb Electrode
Chest Electrode Connection:
1)
Ensure that the electrodes are clean;
2)
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to
the corresponding electrodes according to the colors and identifiers;
3)
Clean the electrode area on the chest surface with 75% alcohol;
4)
Daub the round area of 25mm in diameter on each electrode site with gel evenly;
5)
Place a small amount of gel on the brim of the chest electrode’s metal cup;
6)
Place the electrode on the chest electrode site and squeeze the suction bulb. Unclench it
and the electrode is adsorbed on the chest;
7)
Attach all chest electrodes in the same way.
- 42 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
NOTE: Long-time measurement with a strong negative pressure on the suction bulb may
cause reddening of the skin. When using the electrode on small children or
patients with delicate skin, squeeze the suction ball lightly.
Limb Electrode Connection:
1)
Ensure that the electrodes are clean;
2)
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to
the corresponding electrodes according to the colors and identifiers;
3)
Clean the electrode area which is a short distance above the ankle or the wrist with 75%
alcohol;
4)
Daub the electrode area on the limb with gel evenly;
5)
Place a small amount of gel on the metal part of the limb electrode clamp;
6)
Connect the electrode to the limb, and make sure that the metal part is placed on the
electrode area above the ankle or the wrist;
7)
Attach all limb electrodes in the same way.
4.5.2.2 Disposable Electrodes
CAUTION
The disposable electrodes can only be used for one time.
Disposable Electrode
Alligator Clip
Disposable electrodes must be used together with alligator clips.
- 43 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
Disposable Electrode Connection
1) Align all lead wires of the patient cable to avoid twisting, and connect the alligator clips to
the lead wires.
2) Clean the electrode areas on the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Clip the disposable electrodes with the alligator clips.
The quality of ECG waveform will be affected by the contact resistance between the patient and
the electrode. In order to get a high-quality ECG, the skin-electrode resistance must be minimized
while connecting electrodes.
WARNING
1. Make sure that all electrodes are connected to the patient correctly before operation.
2. Ensure that the conductive parts of electrodes and associated connectors, including
neutral electrodes, do not come in contact with earth or any other conducting objects.
4.6 Inspection before Power-On
In order to avoid safety hazards and get good ECG records, the following inspection procedure is
recommended before power-on and operation.
1) Environment:
♦
Make sure that there is no electromagnetic interference source around the equipment,
especially large medical electrical equipment such as electrosurgical equipment,
radiological equipment, magnetic resonance imaging equipment etc. Turn off these
devices when necessary.
♦
Keep the examination room warm to avoid muscle tremor voltages in ECG signals
caused by cold.
2) Power Supply:
♦
If the mains supply is used, please check whether the power cord is connected to the
unit well. The grounded three-slot outlet should be used.
♦
When the battery capacity is low, recharge the battery before use.
- 44 -
SE-1200 Series Electrocardiograph User Manual
Operation Preparations
3) Patient Cable:
♦
Make sure that the patient cable is connected to the unit firmly, and keep it far away
from the power cord.
4) Electrodes:
♦
Make sure that all electrodes are connected to lead wires of the patient cable
correctly.
♦
Ensure that the chest electrodes do not contact.
5) Recorder Paper:
♦
Ensure that there is enough recorder paper loaded correctly.
6) Patient:
♦
The patient should not come into contact with conducting objects such as earth, metal
parts etc.
♦
Ensure that the patient is warm and relaxed, and breathes calmly.
WARNING
The electrocardiograph is intended to be used by qualified physicians or personnel
professionally trained, and they should be familiar with the contents of this user manual
before operation.
- 45 -
SE-1200 Series Electrocardiograph User Manual
Turning on the Electrocardiograph
Chapter 5 Turning on the Electrocardiograph
♦ When the mains supply is used, connect the power cord, and the mains supply indicator ( )
is lit. Then press
on the keyboard to turn on the unit. The logo will be displayed on the
LCD screen after self-test. Then the electrocardiograph is ready for use.
♦ When battery is used, press
battery indicator (
on the keyboard directly to turn on the unit, and then the
) is lit. The logo will be displayed on the LCD screen after self-test. Then
the electrocardiograph is ready for use.
- 46 -
SE-1200 Series Electrocardiograph User Manual
Entering Patient Information
Chapter 6 Entering Patient Information
After the electrocardiograph is turned on, the main screen appears.
6.1 Entering Patient ID
1. By default, the system generates the patient ID automatically. The range of the patient ID is 0
~ 1999, 999, 999.
2. Or the system can generate the patient ID according to the time when you press the
START/STOP key to print an ECG report.
The procedures are as follows:
1) Press the function key F6 below Setup to open the System Setup screen.
2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When
the cursor is on Patient Info, press
.
3) Press Tab or Shift + Tab to move the cursor to ID Mode, and then press the Up or Down
arrow to select Time.
4) Press
press
, or press Tab or Shift + Tab to move the cursor to the OK button, and then
.
5) Press ESC to return to the main screen.
Then the system will generate the patient ID according to the time when you press the
START/STOP key to print an ECG report.
3. Or you can enter the patient ID manually.
The procedures are as follows:
1) Press the function key F6 below Setup to open the System Setup screen.
2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When
the cursor is on Patient Info, press
.
- 47 -
SE-1200 Series Electrocardiograph User Manual
Entering Patient Information
3) Press Tab or Shift + Tab to move the cursor to ID Mode, and then press the Up or Down
arrow to select Manual.
4) Press
press
, or press Tab or Shift + Tab to move the cursor to the OK button, and then
.
5) Press ESC to return to the main screen.
6) Press the function key F1 below Patient to open the Patient Information window. Press
Tab or Shift + Tab to move the cursor. When the cursor is in the ID textbox, enter the
patient ID manually.
7) Press
press
, or press Tab or Shift + Tab to move the cursor to the OK button, and then
.
NOTE: In the auto, rhythm or off mode, when ID Mode is set to Manual and ID Hint is set
to On, if you do not input the patient ID before pressing the START/STOP key, a
hint will pop up to remind you to input the patient ID.
6.2 Entering Other Information
When the main screen is displayed, press the function key F1 below Patient to open the Patient
Information window.
- 48 -
SE-1200 Series Electrocardiograph User Manual
Entering Patient Information
First Name
Within 30 ASCII characters
Last Name
Within 30 ASCII characters
Age
Age Unit: Years, Months, Weeks or Days
Gender
Patient Gender (Male/Female/Empty)
Weight (kg)
Patient Weight (Range 0~999)
Height (cm)
Patient Height (Range 0~999)
Pacemaker
If you select Pacemaker in the Patient Information Setup window,
Pacemaker appears in the Patient Information window. If Pacemaker is
set to Yes, the pacemaker signals are easy to be detected. If Pacemaker is
set to No, the pacemaker signals are not easy to be detected.
BP (mmHg)
Patient Systolic Blood Pressure/Diastolic Blood Pressure
Race
Patient Race (unknown/ Oriental/ Caucasian/ Black/ Indian/ Mongolian/
Hispanic/ Asian/ Pacific/ other)
Medication
The medication which patient took previously (Within 30 ASCII
characters)
Room No.
Room Number (Within 10 ASCII characters)
Department
Within 30 ASCII characters
Physician
Physician Name (Within 35 ASCII characters)
Technician
Technician Name (Within 35 ASCII characters)
Ref-Physician
Referring Physician Name (Within 35 ASCII characters)
NOTE: The total number of supported characters may be fewer if either special Latin
characters or Chinese characters are entered.
Operation Procedures are as follows:
1. Press Tab or Shift + Tab to move the cursor to a textbox. Fill in the selected textbox. For
details on inputting data, please refer to Section 3.2, “Entering Data”.
2. Press Tab or Shift + Tab to move the cursor to a list box. Press the Up or Down arrow to
select an option.
- 49 -
SE-1200 Series Electrocardiograph User Manual
3. After inputting all the information, press
Entering Patient Information
to confirm; or press Tab or Shift + Tab to
move the cursor to the OK button, and then press
to confirm.
NOTE: If you select D.O.B in the Patient Information Setup window, the D.O.B textbox
appears and the Age textbox becomes unavailable in the Patient Information
window, you can enter the birthday of the patient, and the system will calculate
the patient age automatically.
6.3 Using a Bar Code Reader
1. Configure the bar code
1) When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen.
2) Press F1 to display the Service Password window. Then enter the password and press
to open the Advanced Setup screen Press Tab or Shift + Tab, or the Up, Down,
Left or Right arrow to move the cursor to Barcode, and then press
Barcode Setup window.
- 50 -
to open the
SE-1200 Series Electrocardiograph User Manual
Entering Patient Information
3) Enter the start and end addresses, the male and female codes, the vendor ID and the product
ID, and then press
to confirm.
View the vendor ID and product ID on the PC:
a) Connect the bar code reader to the PC
b) Click on start, right-click on My Computer, and then select Manage in the pop-up
menu to open the Computer Management window.
- 51 -
SE-1200 Series Electrocardiograph User Manual
Entering Patient Information
c) In the Computer Management window, click on Device Manager. Right-click on HID
Keyboard Device, and then select Properties in the pop-up menu.
d) Click on the Details tab in the HID Keyboard Device Properties window. View the
vector ID (VID) and the product ID (PID). Take the following figure for example, the
vector ID is 04B4, and the product ID is 0100.
- 52 -
SE-1200 Series Electrocardiograph User Manual
Entering Patient Information
4) Press ESC to return to the main screen.
2. Connect the bar code reader to USB socket 2 on the right panel of the electrocardiograph.
3. When the main screen is displayed, scan the patient’s bar code with the bar code reader, and
then the patient information will appear in the corresponding box.
NOTE: Only bar code readers recommended by the manufacturer can be used.
- 53 -
SE-1200 Series Electrocardiograph User Manual
Freezing ECG Waves
Chapter 7 Freezing ECG Waves
You can freeze the ECG waves displayed on the main screen.
Operation Method:
1) Press the function key F2 below Freeze to display the freezing screen.
NOTE: Within ten seconds after returning to the main screen, pressing the function
key F2 can not display the freezing screen.
Figure 7-1 freezing screen
2) Press the function key F3 to set the paper speed, press the function key F4 to set the
gain, press the function key F5 to set the filter, and press the function key F6 to set the
printing style.
NOTE: When you press the function key F7 below Return to exit the freezing
screen, these settings will not be saved.
3) Press the function key F2 below Replay to replay the frozen ECG waves on the Replay
screen.
- 54 -
SE-1200 Series Electrocardiograph User Manual
Freezing ECG Waves
Figure 7-2 Replay screen
4) Press F1 below Pre Page or F4 below Next Page to turn pages.
5) Press F2 below Pre Sec or F3 below Next Sec to view the ECG waves of the previous
or next second.
6) Press the function key F5 below Play to display ECG waves continuously.
7) Press the function key F6 below Pause to stop playing ECG waves.
8) Press the function key F7 below Return to return to the main screen.
- 55 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
Chapter 8 Printing ECG Reports
There are four modes to print ECG reports.
In the auto mode, the lead groups are switched automatically according to the lead sequence
during the printing course. After the ECG waves of one lead group are printed within a certain
time, the system switches to print ECG waves of another lead group automatically. A 1mV
calibration mark will be printed at the beginning of an ECG report.
In the manual mode, you can determine the lead group to be displayed and printed by pressing the
Up or Down arrow.
In the rhythm mode, you can print 60s rhythm-lead ECG waveform of one lead in the Single
Lead style or 20s rhythm-lead ECG waveform of three leads in the Three Leads style.
In the R-R analysis mode, you can select a lead to print its R-R histogram, R-R trend chart, 180s
compressed ECG waveform and all the R-R interval values.
NOTE:
1. The working mode can not be changed during the printing course. Stop printing reports
before changing the working mode.
2. Within three seconds after returning to the main screen, if you press the START/STOP
key to print an ECG report in the auto quick mode or the manual mode, the recorder
will not respond.
3. The electrocardiograph can not print ECG reports when transmitting data through USB
socket 2.
8.1 Auto Mode
Operation Method:
1. To set Mode Options, Auto DispStyle, Sampling Mode, Preview, Record Style, Rhythm
Lead1/2/3, or Lead Sequence (Optional)
When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen. The cursor is on Work Mode. Press
to open the Work
Mode Setup window.
1) Press Tab or Shift + Tab to move the cursor to AUTO, press Spacebar to select it, and
then a check mark √ appears in the box before AUTO.
- 56 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
2) Press Tab or Shift + Tab to move the cursor to the Auto DispStyle list box, and then
press the Up or Down arrow to highlight a style.
3) Press Tab or Shift + Tab to move the cursor to the Sampling Mode list box, and then
press the Up or Down arrow to highlight a sampling mode. For details of the sampling
mode, refer to Section 11.1.5, “Selecting Sampling Mode”.
4) Press Tab or Shift + Tab to move the cursor to Preview, and then press the Up or Down
arrow to highlight Off or On.
5) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
6) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right
arrow to move the cursor. When the cursor is on Record Info, press
to open
the Record Info Setup window.
7) Press Tab or Shift + Tab to move the cursor to the Record Style list box, and then press
the Up or Down arrow to select a style.
- 57 -
SE-1200 Series Electrocardiograph User Manual
8) Press
Printing ECG Reports
, or press Tab or Shift + Tab to move the cursor to the OK button, and
then press
.
9) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right
arrow to move the cursor. When the cursor is on Lead, press
to open the Lead
Setup window.
10) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box, and then
press the Up or Down arrow to select a lead.
11) Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box, and then
press the Up or Down arrow to select a sequence.
12) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
- 58 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
After setup, press ESC to return to the main screen.
2. When the main screen is displayed, press the MODE key to select the auto mode. Press F3 to
select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the
Lowpass filter.
3. Press the START/STOP key to print an ECG report. It will stop automatically after printing
a complete ECG report of 12 leads. Or press the START/STOP key again to stop printing the
report.
4. If Preview is set to On in the Work Mode Setup window, press the START/STOP key to
display the preview screen.
Figure 8-1 Waveform Preview Screen
- 59 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
1) Press the function key F1 below Waveform to display the Waveform Preview Screen.
2) Press the function key F2 below Template to display the template preview screen.
3) Press the function key F3 below Measure to display the measure preview screen.
4) Press the function key F4 below Diagnosis to open the diagnosis preview screen.
5) Press the function key F5 below Continue to print an ECG report.
6) Press the function key F6 below Cancel to stop printing the ECG report.
8.2 Manual Mode
Operation Method:
1. To set Mode Options, Manual Style or Lead Sequence (Optional)
When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen. The cursor is on Work Mode. Press
to open the Work
Mode Setup window.
1) Press Tab or Shift + Tab to move the cursor to MANU, press Spacebar to select it, and
then a check mark √ appears in the box before MANU.
2) Press Tab or Shift + Tab to move the cursor to the Manual Style list box, and then press
the Up or Down arrow to highlight a style.
3) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
- 60 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
4) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right
arrow to move the cursor. When the cursor is on Lead, press
to open the Lead
Setup window.
5) Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box, and then
press the Up or Down arrow to select a sequence.
6) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
After setup, press ESC to return to the main screen.
- 61 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
2. When the main screen is displayed, press the MODE key to select the manual mode. Press
F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the
Lowpass filter.
3. Press the Up or Down arrow to select the lead group to be displayed and printed.
4. Press the START/STOP key to print an ECG report.
5. Press the START/STOP key to stop printing the ECG report.
8.3 Rhythm Mode
Operation Method:
1. To set Mode Options, Rhythm Style, Rhythm Lead1/2/3 or Lead Sequence (Optional)
When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen. The cursor is on Work Mode. Press
to open the Work
Mode Setup window.
1) Press Tab or Shift + Tab to move the cursor to RHYT, press Spacebar to select it, and
then a check mark √ appears in the box before RHYT.
2) Press Tab or Shift + Tab to move the cursor to the Rhythm Style list box, and then press
the Up or Down arrow to highlight a style.
3) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
- 62 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
4) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right
arrow to move the cursor. When the cursor is on Lead, press
to open the Lead
Setup window.
5) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box, and then
press the Up or Down arrow to select a lead.
6) Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box, and then
press the Up or Down arrow to select a sequence.
7) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
- 63 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
After setup, press ESC to return to the main screen.
2. When the main screen is displayed, press the MODE key to select the rhythm mode. Press
F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the
Lowpass filter.
3. Press the START/STOP key to begin to count the sampling time. When the sampling time
reaches 60s in the Single Lead style or 20s in the Three Leads style, it begins to print an
ECG report.
4. It will stop automatically after printing a complete report of rhythm-lead ECG waveforms. Or
press the START/STOP key again to stop printing the ECG report.
8.4 R-R Analysis Mode
Operation Method:
1. To set Mode Options or Rhythm Lead1
When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen. The cursor is on Work Mode. Press
to open the Work
Mode Setup window.
1) Press Tab or Shift + Tab to move the cursor to R-R, press Spacebar to select it, and then
a check mark √ appears in the box before R-R.
- 64 -
SE-1200 Series Electrocardiograph User Manual
2) Press
then press
Printing ECG Reports
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
3) On the System Setup screen, press Tab or Shift + Tab, or the Up, Down, Left or Right
arrow to move the cursor. When the cursor is on Lead, press
to open the Lead
Setup window.
4) Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1 list box, and then
press the Up or Down arrow to select a lead.
5) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
- 65 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
After setup, press ESC to return to the main screen.
2. When the main screen is displayed, press the MODE key to select the R-R analysis mode.
Press F4 to switch the gain. Press F5 to set the EMG filter or the Lowpass filter.
3. Press the START/STOP key to begin to count the sampling time. When the sampling time
reaches 180s, it begins to print an ECG report.
4. It will stop automatically after a complete R-R analysis report is printed, or press the
START/STOP key to stop printing the ECG report.
NOTE: In the R-R analysis mode, you can not set the speed. The constant speed is
25mm/s and the printing speed is 5mm/s, because in the R-R analysis mode, the
ECG wave length is compressed to one fifth of the original wave length.
8.5 Copy Printing
In the auto mode, pressing the COPY key can print the ECG report which was printed out last
time. Pressing the START/STOP key can stop printing the ECG report.
- 66 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
8.6 ECG Reports
8.6.1 ECG Reports in the Auto Mode
8.6.1.1 Example of 6×2+1R
Save Paper Mode
The above figure shows an ECG report in the auto mode. The style is 6×2+1R. The ECG report
includes:
Patient Information, Measure Information, Diagnosis Information, Report Confirmed by
Current Date and Current Time
6×2+1R ECG waves
0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter)
AC50 (50Hz AC Filter)
25mm/s (Paper Speed)
10mm/mV (Gain)
2*5s+1r (12 leads are printed in 2 groups of 6, and every group is printed for about 5s, with one
rhythm lead at the bottom simultaneously)
- 67 -
SE-1200 Series Electrocardiograph User Manual
Institution Name
SE-1200 Express (Electrocardiograph Model)
V1.8 (Software Version Number)
SEMIP V1.5(Algorithm Version Number)
Quick Mode
(a)
- 68 -
Printing ECG Reports
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
(b)
The above figure (a) and (b) show an ECG report in the auto mode. Template is selected, and
Record Style is set to 6×2+1R.
Figure (a) shows:
02-06-2010 16:52:40 (Current Date & Current Time)
ID: 001 (Patient ID)
Name: Marry, Johnson (Patient Name)
Age: 25 Years (Patient Age)
(1mV calibration mark)
І, II, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (12 standard leads) and ECG waves
II on the bottom (Rhythm Lead) and rhythm wave
0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter)
AC50 (50Hz AC Filter)
25mm/s (Paper Speed)
10mm/mV (Gain)
- 69 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
2*5s+1r (12 leads are printed in 2 groups of 6, and every group is printed for about 5s, with one
rhythm lead at the bottom simultaneously)
Institution Name
SE-1200 Express (Electrocardiograph Model)
V1.8 (Software Version Number)
SEMIP V1.5(Algorithm Version Number)
Figure (b) shows:
Patient Information, Measure Information, Diagnosis Information, Average Template, Report
Confirmed by, Current Date and Current Time, 0.67Hz (DFT Filter), 100Hz (Lowpass Filter),
AC50 (50Hz AC Filter), 25mm/s (Paper Speed), 10mm/mV (Gain), Institution Name, SE-1200
Express (Electrocardiograph Model), V1.8 (Software Version Number), SEMIP V1.5(Algorithm
Version Number).
Patient Information includes:
ID, Name, Age, Gender, Height, Weight, Room No., Department
Measure Information includes:
HR (Heart Rate)
P Dur----P wave duration: the average P-wave duration from several selected dominant beats;
PR int----P-R interval: the average P-R interval from several selected dominant beats;
QRS Dur----QRS complex duration: the average QRS complex duration from several selected
dominant beats;
QT/QTC int----Q-T interval: the average Q-T interval from several selected dominant beats /
Normalized QT interval;
P/QRS/T axis----Dominant direction of the average integrated ECG vectors;
RV5/SV1 amp----The maximum of the amplitude of R or R’ wave of one selected dominant beat
from lead V5 / The maximum absolute value of the amplitude of S or S’ wave
of one selected dominant beat from lead V1;
RV5+SV1 amp---- Sum of RV5 and SV1;
RV6/SV2 amp---- The maximum of the amplitude of R or R’ wave of one selected dominant beat
from lead V6 / The maximum absolute value of the amplitude of S or S’ wave
of one selected dominant beat from lead V2;
Diagnosis Information:
Diagnosis information shows the auto diagnosis result.
- 70 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
Average Template:
Average template shows the average value of 10s sampled ECG signals of every lead.
The broken lines on the template are position markers. They respectively mark the start and end
points of the P and QRS waves, and the end point of the T wave.
8.6.1.2 Example of 3×4+1R
The above figure shows an ECG report in the auto mode. The style is 3×4+1R. The ECG report
includes:
Patient Information, Measure Information, Diagnosis Information, Report Confirmed by
02-06-2010 16:53:52 (Current Date and Current Time)
3×4+1R ECG waves
0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter)
AC50 (50Hz AC Filter)
25mm/s (Paper Speed)
10mm/mV (Gain)
4*2.5s+1r (12 leads are printed in 4 groups of 3, and every group is printed for about 2.5s, with
one rhythm lead at the bottom simultaneously)
Institution Name
- 71 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
SE-1200 Express (Electrocardiograph Model)
V1.8 (Software Version Number)
SEMIP V1.5(Algorithm Version Number)
NOTE: Please refer to Section 8.6.1.1 for the explanation of the patient information,
measure information and diagnosis information.
8.6.2 ECG Reports in the Rhythm Mode
The above figure shows an ECG report in the rhythm mode, and Rhythm Style is set to Three
Leads.
02-06-2010 16:54:43 (Current Date & Current Time)
ID: 001 (Patient ID)
Age: 25 Years (Patient Age)
Name: Marry, Johnson (Patient Name)
00:00, 00:10 (Timer)
60 (Heart Rate)
(1mV calibration mark)
II, V1, V5 (Rhythm Lead Name)
- 72 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
20s rhythm waveform of lead II / V1 / V5
0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter)
AC50 (50Hz AC Filter)
25mm/s (Paper Speed)
10mm/mV (Gain)
Institution Name
SE-1200 Express (Electrocardiograph Model)
V1.8 (Software Version Number)
SEMIP V1.5(Algorithm Version Number)
8.6.3 ECG Reports in the Manual Mode
The above figure shows an ECG report in the manual mode, and Manual Style is set to 12
channels.
02-06-2010 16:56:30 (Current Date & Current Time)
ID: 001 (Patient ID)
Age: 25 Years (Patient Age)
Name: Marry, Johnson (Patient Name)
(1mV calibration mark)
І, II, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (12 standard leads)
- 73 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
ECG waveform of 12 standard leads
0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter)
AC50 (50Hz AC Filter)
25mm/s (Paper Speed)
60 (Heart Rate)
10mm/mV (Gain)
Institution Name
SE-1200 Express (Electrocardiograph Model)
V1.8 (Software Version Number)
SEMIP V1.5(Algorithm Version Number)
8.6.4 ECG Reports in the R-R Analysis Mode
(a)
- 74 -
SE-1200 Series Electrocardiograph User Manual
(b)
- 75 -
Printing ECG Reports
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
(c)
The above figure (a), (b) and (c) show an ECG report in the R-R analysis mode.
Figure (a) shows:
Current Date & Current Time
Patient Information (Name, ID, Gender, Age, Height, Weight)
Measuring Time
Total Beat Number
HR (Heart Rate)
RR Avg Interval (Average RR interval)
RR Max Interval (Maximum RR interval)
RR Min Interval (Minimum RR interval)
Max/Min (Ratio of Maximum RR interval to Minimum RR interval)
SDNN (Standard Deviation of Normal to Normal Intervals)
RMSSD (Root Mean Square Successive Difference)
RR Histogram
RR Trend
- 76 -
SE-1200 Series Electrocardiograph User Manual
Figure (b) shows:
Current Date & Current Time
Patient Information (Name, ID, Age)
Gain, Speed, Filter
(1mV calibration mark)
II (Lead name)
180s compressed ECG wave of Lead II
Figure (c) shows:
All the RR interval values within the measuring time
- 77 -
Printing ECG Reports
SE-1200 Series Electrocardiograph User Manual
8.6.5 ECG Reports Printed by the USB Printer
- 78 -
Printing ECG Reports
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
As the above figure shows, the ECG report printed by the USB printer includes:
Institution Name
Date and Time
Patient Information
Measurement Information
Diagnosis Information;
Report Confirmed by;
ECG waves of 12 leads
DFT Filter, Lowpass Filter, AC Filter, Paper Speed, Gain;
SE-1200 Express (Electrocardiograph Model)
V1.8 (Software Version Number)
SEMIP V1.5(Algorithm Version Number)
8.6.6 ECG Report of Patient with Pacemaker
The above figure shows the ECG report of a patient with a pacemaker. Pacemaker Detection
Sens. is set to High, and the Lowpass filter is set to 150Hz.
- 79 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
NOTE:
1. To the patient with a pacemaker, the EMG filter should be set to Off and the Lowpass
filter should be set to 150Hz; or else, the amplitude of the pulses may decrease, the
width may increase, and the diagnosis results may not include the pacemaker
information.
2. To the patient with a pacemaker, the results given by the equipment may be invalid.
8.6.7 ECG Report of Nehb Lead
ECG report of Nehb lead includes:
30-06-2010 17:45:16 (Current Date & Current Time)
ID: 001 (Patient ID)
Age: 25 Years (Patient Age)
Name: Marry, Johnson (Patient Name)
(1mV calibration mark)
І, II, Ш, ND, NA, NI (Nehb leads)
- 80 -
SE-1200 Series Electrocardiograph User Manual
Printing ECG Reports
0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter)
AC50 (50Hz AC Filter)
10mm/mV (Gain)
25mm/s (Paper Speed)
60 (Heart Rate)
Institution Name
SE-1200 Express (Electrocardiograph Model)
V1.8 (Software Version Number)
SEMIP V1.5(Algorithm Version Number)
- 81 -
SE-1200 Series Electrocardiograph User Manual
Chapter
9
Transmitting ECG Data Automatically
Transmitting
ECG
Data
Automatically
9.1 Transmitting ECG Data to the PC
NOTE:
1. To transmit ECG data in DAT format to the PC, the Smart ECG Viewer software of the
manufacturer must be installed in the PC. You should log into the Smart ECG Viewer
software before transmission.
2. To transmit ECG data in SCP/FDA-XML/PDF format to the PC, the FTP receiving
software must be installed in the PC. You should log into the FTP receiving software
before transmission.
3. By default, the ECG file format is DAT. You can activate SCP/FDA-XML/PDF function
on the Advanced Setup screen For details, refer to Section 9.1.3, “Transmitting ECG
Data in SCP/FDA-XML/PDF Format Through Net Cable”.
9.1.1 Transmitting ECG Data in DAT Format through Serial Cable
1. Connect the RS232 socket of the PC to the RS232 socket of the electrocardiograph with an
RS232 cable.
2. Or if the PC has no RS232 socket, connect the USB socket of the PC to the RS232 socket of
the electrocardiograph by using the RS232-USB assembly.
- 82 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
Figure 9-1 RS232-USB Assembly
3. Set Auto Transmission to On and Transmission Mode to UART in the Transmission Setup
window.
1) When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen.
2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to
to open the Transmission Setup window.
Transmission, press
3) Press Tab or Shift + Tab to move the cursor to the Auto Transmission list box, and then
press the Up or Down arrow to select On. Press Tab or Shift + Tab to move the cursor to
the Transmission Mode list box, and then press the Up or Down arrow to select UART.
4) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
5) Press ESC to return to the main screen.
4. Log into the Smart ECG Viewer software.
- 83 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
5. In the auto or rhythm mode, ECG data will be transmitted through the serial cable
automatically after an ECG report is printed out. In the off mode, ECG data will be
transmitted through the serial cable automatically after the START/STOP key is pressed.
9.1.2 Transmitting ECG Data in DAT Format through Net Cable
1. Connect the electrocardiograph to the network of the PC with an Ethernet cable recommended
by the manufacturer.
If the wireless AP transmission is used, connect the PC to the server wireless AP, and connect
the electrocardiograph to the client wireless AP. Only the wireless AP recommended by the
manufacturer can be used.
NOTE:
1. For details on configuring the wireless AP, please refer to the user manual
delivered with the wireless AP.
2. There should be no shield in or around the room where the wireless AP is used,
or else the wireless transmission may fail.
2. Set Auto Transmission to On and Transmission Mode to Ethernet in the Transmission
Setup window. Set the IP addresses.
1) When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen.
2) Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to
Transmission, press
to open the Transmission Setup window.
3) Press Tab or Shift + Tab to move the cursor to the Auto Transmission list box, and then
press the Up or Down arrow to select On.
4) Press Tab or Shift + Tab to move the cursor to the Transmission Mode list box, and then
press the Up or Down arrow to select Ethernet.
5) Press Tab or Shift + Tab to move the cursor to the Server IP item. Set the Server IP
item to the IP of the PC.
- 84 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
6) Press Tab or Shift + Tab to move the cursor to the Local IP item.
For the cross-network transmission, set the first two sections of the Local IP item to the first
two sections of the IP of the PC. Set the third section of the Local IP item to the network
segment of the electrocardiograph which depends on the configuration of Router. The last
section of the Local IP item can be set at random.
For the same network transmission, set the first three sections of the Local IP item to the first
three sections of the IP of the PC. The last section of the Local IP item can be set at random,
but it can’t be the same as the last section of the IP of the PC.
7) Press Tab or Shift + Tab to move the cursor to the Gateway item. Set the first three
sections of the Gateway item to the first three sections of the IP of the electrocardiograph.
The last section of the Gateway item must be set to 1.
NOTE: For more information on configuring network settings, see your Network
Administrator.
Take the same network transmission for example:
View the IP of the PC.
- 85 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
In the Transmission Setup window of the electrocardiograph, set Transmission Mode to
Ethernet. Set the Server IP item to the IP of the PC. Set the first three sections of the Local
IP item to the first three sections of the IP of the PC. The last section of the Local IP item is
not the same as the last section of the IP of the PC. Set the first three sections of the Gateway
item to the first three sections of the IP of the electrocardiograph. The last section of the
Gateway item must be set to 1. Commonly, Subnet Mask is set to 255.255.255.0.
8) Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
9) Press ESC to return to the main screen.
3. Log into the Smart ECG Viewer software.
4. In the auto or rhythm mode, ECG data will be transmitted through the net cable automatically
after an ECG report is printed out. In the off mode, ECG data will be transmitted through the
net cable automatically after the START/STOP key is pressed.
- 86 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
9.1.3 Transmitting ECG Data in SCP/FDA-XML/PDF Format
through Net Cable
1. Activate SCP/FDA-XML/PDF function on the Advanced Setup screen
1) When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen.
2) Press F1 to display the Service Password window. Then enter the password and press
to open the Advanced Setup screen Press Tab or Shift + Tab, or the Up,
Down, Left or Right arrow to move the cursor to Function, and then press
open the Function Configuration window.
- 87 -
to
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
3) Press Tab or Shift + Tab to move the cursor to the Activate button after the function to
be activated. Press
password and press
to open the Activate Password window. Enter the
to activate the function.
NOTE: Please contact local distributors to get the service password and the activating
password.
2. Set file format to SCP/FDA-XML/PDF
When the main screen is displayed, press the function key F6 below Setup to open the System
Setup screen. Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the
cursor to File, press
to open the File Setup window. Press Tab or Shift + Tab to
move the cursor to File Format, and then press the Up or Down arrow to select a format.
3. Transmit ECG data in SCP/FDA-XML/PDF format through the net cable
1) Connect the electrocardiograph to the network of the PC with an Ethernet cable
recommended by the manufacturer.
If the wireless AP transmission is used, connect the PC to the server wireless AP, and
connect the electrocardiograph to the client wireless AP. Only the wireless AP
recommended by the manufacturer can be used.
2) Input the correct FTP path, FTP user name and password. Set Auto Transmission to On
and Transmission Mode to Ethernet. Set the IP addresses. For details, please refer to
Section 9.1.2, “Transmitting ECG Data in DAT Format Through Net Cable”.
- 88 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
3) Log into the FTP receiving software. The FTP port is 21.
4) In the auto or rhythm mode, ECG data will be transmitted through the net cable
automatically after an ECG report is printed out. In the off mode, ECG data will be
transmitted through the net cable automatically after the START/STOP key is pressed.
9.2 Exporting ECG Data to the U Disk Automatically
1. Insert the U disk recommended by the manufacturer to USB socket 2 of the
electrocardiograph.
2. When the main screen is displayed, press the function key F6 below Setup to open the System
Setup screen. Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the
cursor to File, press
to open the File Setup window.
3. Press Tab or Shift + Tab to move the cursor to the Auto Save list box, and then press the Up
or Down arrow to select On. Press Tab or Shift + Tab to move the cursor to the Saving Path
list box, and then press the Up or Down arrow to select U Disk. Press Tab or Shift + Tab to
move the cursor to File Format, and then press the Up or Down arrow to select
DAT/SCP/FDA-XML/PDF. For details on activating SCP/FDA-XML/PDF function on the
Advanced
Setup
screen, refer to Section 9.1.3, “Transmitting ECG Data in
SCP/FDA-XML/PDF Format Through Net Cable”.
Press
, or press Tab or Shift + Tab to move the cursor to the OK button, and then press
. When the System Setup screen is displayed, press ESC to return to the main screen.
- 89 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
4. In the auto or rhythm mode, ECG data will be automatically exported to the directory of
ECGDATA\ECG-X\Store\Examination Date of the U disk after an ECG report is printed out.
In the off mode, ECG data will be automatically exported to the directory of
ECGDATA\ECG-X\Store\Examination Date of the U disk after the START/STOP key is
pressed.
NOTE: X can be set in the Device No. textbox in the Transmission Setup window.
9.3 Real-time Transmission to PC ECG
NOTE: For the real-time transmission, the PC ECG software of the manufacturer must
be installed in the PC.
1. Run the PC ECG software and configure it for the real-time transmission.
Double-click on the shortcut icon
on the desktop to start up the PC ECG software.
Click on the System Setting button to open the System Setting window. Click on the Device
tab. Select SE serial Machine from the Sampling Device pull-down list on the device setup
window. After setup, click on the OK button to confirm.
- 90 -
SE-1200 Series Electrocardiograph User Manual
Transmitting ECG Data Automatically
2. Connect USB socket 1 of the electrocardiograph to the USB socket of the PC by using the
high-speed USB cable
3. Start the real-time transmission
The electrocardiograph transmits to the PC the ECG signals acquired from the patient.
Acquisition and transmission are simultaneous. The ECG signals are displayed on the PC
monitor and eventually analyzed. For more details, refer to the user manual of the PC ECG
software.
In the real-time transmission, the function of the electrocardiograph is the same as that of the
ECG sampling box.
- 91 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
Chapter 10 Managing Files
If you want to save the ECG data in the electrocardiograph, you should set Auto Save to On and
Saving Path to ECG. By default, Auto Save is On and Saving Path is ECG. Then the ECG
data in the auto, off or rhythm mode will be saved on the File Manager screen automatically.
To Set Auto Save to On and Saving Path to ECG
1. When the main screen is displayed, press the function key F6 below Setup to open the
System Setup screen.
2. Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to
File, press
.
3. Press Tab or Shift + Tab to move the cursor to the Auto Save list box, and then press the
Up or Down arrow to select On. Press Tab or Shift + Tab to move the cursor to the
Saving Path list box, and then press the Up or Down arrow to select ECG.
4. Press
then press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
.
5. When the System Setup screen is displayed, press ESC to return to the main screen.
To Switch to the File Manager screen 1/2
When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1. Press the Up or Down arrow to highlight a file on the File Manager screen1,
and then press the function key F4 below Select to select the file and display the File Manager
screen2. Then press the function key F7 below Return to return to the File Manager screen1.
Press the function key F7 below Return on the File Manager screen1 to return to the main
screen.
- 92 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
Figure 10-1 File Manager screen1
Figure 10-2 File Manager screen2
On the File Manager screen, files can be printed, transmitted, exported, displayed, edited
searched or deleted. SE-1200 can accommodate 100 files, and SE-1200 Express can
accommodate 200 files.
If there is no file on the File Manager screen, the following dialog box will pop up when you
press function keys.
- 93 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
CAUTION
1. When files are being printed, transmitted, deleted or exported, you can not turn off the
electrocardiograph.
2. Do not cut off the mains supply directly when no battery is installed in the device, or
else, the stored data may be lost.
10.1 Transmitting Files to the PC
NOTE:
1. To transmit DAT files to the PC, the Smart ECG Viewer software of the manufacturer
must be installed in the PC. You should log into the Smart ECG Viewer software
before transmission.
2. To transmit SCP/FDA-XML/PDF files to the PC, the FTP receiving software must be
installed in the PC. You should log into the FTP receiving software before
transmission.
3. By default, the ECG file format is DAT. You can activate SCP/FDA-XML/PDF function
on the Advanced Setup screen For details, refer to Section 9.1.3, “Transmitting ECG
Data in SCP/FDA-XML/PDF Format Through Net Cable”.
4. If you select Delete After Trans. Or Export in the File Setup window, the files will be
deleted from the File Manager screen after they are transmitted.
CAUTION
It is forbidden to connect or disconnect a U disk or a USB printer during the transmission
course.
- 94 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
10.1.1 Transmitting DAT Files through Serial Cable
1. Connect the RS232 socket of the PC to the RS232 socket of the electrocardiograph with an
RS232 cable.
2. Or if the PC has no RS232 socket, connect the USB socket of the PC to the RS232 socket of
the electrocardiograph by using the RS232-USB assembly. For details of the RS232-USB
assembly, please refer to Section 9.1.1, “Transmitting ECG Data in DAT Format Through
Serial Cable”.
3. Log into the Smart ECG Viewer software.
NOTE: If you don’t connect the RS232 socket of the electrocardiograph to the PC, or you
do not log into Smart ECG Viewer before transmission, the following hint will pop
up.
4. Set Transmission Mode to UART in the Transmission Setup window. For details, refer to
Section 9.1.1, “Transmitting ECG Data in DAT Format Through Serial Cable”.
5. Press the function key F7 below File to display the File Manager screen1.
6. Press the function key F1 below Trans All to transmit all the files to the PC. After all the files
are transmitted, the following hint will pop up.
7. Or press the Up or Down arrow to highlight a file on the File Manager screen1, and then
press the function key F4 below Select to select the file and display the File Manager
screen2. Press the function key F3 below Trans to transmit the selected file to the PC.
- 95 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
10.1.2 Transmitting DAT Files through Net Cable
1. Connect the electrocardiograph to the network of the PC with an Ethernet cable recommended
by the manufacturer.
If the wireless AP transmission is used, connect the PC to the server wireless AP, and connect
the electrocardiograph to the client wireless AP. Only the wireless AP recommended by the
manufacturer can be used.
NOTE:
1. For details on configuring the wireless AP, please refer to the user manual delivered
with the wireless AP.
2. There should be no shield in or around the room where the wireless AP is used, or
else the wireless transmission may fail.
2. Log into the Smart ECG Viewer software.
NOTE: If you don’t connect the electrocardiograph to the network of the PC (or if you
don’t connect the electrocardiograph/PC to the wireless AP), or you do not log
into Smart ECG Viewer before transmission, the following hint will pop up.
3. Set Transmission Mode to Ethernet in the Transmission Setup window. Set the IP addresses.
For details on setting the transmission mode and the IP addresses, please see Section 9.1.2,
“Transmitting ECG Data in DAT Format Through Net Cable”.
4. Press the function key F7 below File to display the File Manager screen1.
5. Press the function key F1 below Trans All to transmit all the files to the PC. After all the files
are transmitted, the following hint will pop up.
- 96 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
6. Or press the Up or Down arrow to highlight a file on the File Manager screen1, and then
press the function key F4 below Select to select the file and display the File Manager
screen2. Press the function key F3 below Trans to transmit the selected file to the PC.
10.1.3 Transmitting SCP/FDA-XML/PDF Files through Net Cable
1. Activate SCP/FDA-XML/PDF function on the Advanced Setup screen For details, refer to
Section 9.1.3, “Transmitting ECG Data in SCP/FDA-XML/PDF Format Through Net Cable”.
2. Set file format to SCP/FDA-XML/PDF
When the main screen is displayed, press the function key F6 below Setup to open the System
Setup screen. Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the
cursor to File, press
to open the File Setup window. Press Tab or Shift + Tab to
move the cursor to File Format, and then press the Up or Down arrow to select a format.
3. Transmit SCP/FDA-XML/PDF files through the net cable
1) Connect the electrocardiograph to the network of the PC with an Ethernet cable
recommended by the manufacturer.
- 97 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
If the wireless AP transmission is used, connect the PC to the server wireless AP, and
connect the electrocardiograph to the client wireless AP. Only the wireless AP
recommended by the manufacturer can be used.
2) Log into the FTP receiving software. The FTP port is 21.
3) Input the correct FTP path, FTP user name and password. Set Transmission Mode to
Ethernet. Set the IP addresses. For details, please see Section 9.1.3, “Transmitting ECG
Data in SCP/FDA-XML/PDF Format to the PC Through Net Cable”.
4) Press the function key F7 below File to display the File Manager screen1.
5) Press the function key F1 below Trans All to transmit all the files to the PC. After all the
files are transmitted, the following hint will pop up.
6) Or press the Up or Down arrow to highlight a file on the File Manager screen1, and then
press the function key F4 below Select to select the file and display the File Manager
screen2. Press the function key F3 below Trans to transmit the selected file to the PC.
10.2 Exporting Files to the U Disk
1. Insert the U disk recommended by the manufacturer to USB socket 2 of the
electrocardiograph.
2. When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1.
3. Press the function key F2 below Export All to export all the files to the directory of
ECGDATA\ECG-X\Export\Examination Date and Time of the U disk.
4. Or press the Up or Down arrow to highlight a file on the File Manager screen1, and then
press the function key F4 below Select to select the file and display the File Manager
screen2. Press the function key F4 below Export to export the selected file to the directory
of ECGDATA\ECG-X\Export\Examination Date and Time of the U disk.
- 98 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
If the U disk is not connected well, the hint U disk is not ready! will be displayed on the LCD
screen of the electrocardiograph, and then you should connect the U disk again. If the
electrocardiograph fails to export files, the hint Exporting fails! will pop up. After the
electrocardiograph successfully exports files, the hint Exporting succeeds! will be displayed.
NOTE:
1. Please insert the U disk or the SD card reader recommended by the manufacturer.
Please set the format to FAT or FAT32 when formatting the U disk.
2. The format of the exported files can be set to DAT/SCP/FDA-XML/PDF. For details,
refer to Section 9.2, “Exporting ECG Data to the U Disk Automatically”.
3. X can be set in the Device No. textbox in the Transmission Setup window.
4. If you select Delete After Trans. Or Export in the File Setup window, the files will be
deleted from the File Manager screen after they are exported.
10.3 Deleting Files
1. When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1.
2. Press the function key F3 below Del All to display the following hint, press Tab or Shift +
Tab to move the cursor to the OK button, and then press
to delete all the files from
the electrocardiograph.
3. Or press the Up or Down arrow to highlight a file on the File Manager screen1, and then
press the function key F4 below Select to select the file and display the File Manager
screen2. Press the function key F5 below Delete to display the following hint, press Tab or
Shift + Tab to move the cursor to the OK button, and then press
selected file from the electrocardiograph.
- 99 -
to delete the
SE-1200 Series Electrocardiograph User Manual
Managing Files
10.4 Searching Files
To Open the SearchInfo Setup Window
1. When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1.
2. Press the function key F5 below Search to open the SearchInfo Setup window.
To Search Files
1. Press Tab, Shift + Tab, or the Up or Down arrow to select Name, and then input the patient
name. Press
, or press Tab or Shift + Tab to move the cursor to the OK button, and
. Then all the files with the input patient name will be displayed on the
then press
File Manager screen1.
2. Press Tab, Shift + Tab, or the Up or Down arrow to select ID, and then input the patient ID.
Press
, or press Tab or Shift + Tab to move the cursor to the OK button, and then
press
. Then all the files with the input patient ID will be displayed on the File
Manager screen1.
- 100 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
3. Press Tab, Shift + Tab, or the Up or Down arrow to select Time, and then input the
examination time range. Press
, or press Tab or Shift + Tab to move the cursor to
the OK button, and then press
. Then all the corresponding files from the start time
to the end time will be displayed on the File Manager screen1.
NOTE: The time mode is year-month-day in the SearchInfo Setup window.
10.5 Importing Files
1. Insert the U disk recommended by the manufacturer to USB socket 2 of the
electrocardiograph.
2. When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1.
3. Press the function key F6 below Import to import files from the directory of
ECGDATA/ECG-X/Import of the U disk to the electrocardiograph.
NOTE:
1. X can be set in the Device No. textbox in the Transmission Setup window.
2. Only the ECG files in DAT format produced by the electrocardiograph of the
manufacturer can be imported.
3. For SE-1200, a maximum of 100 files can be imported at one time. For SE-1200
Express, a maximum of 200 files can be imported at one time.
10.6 Editing Patient Information
When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1. Press the Up or Down arrow to highlight a file on the File Manager screen1,
and then press the function key F4 below Select to select the file and display the File Manager
screen2. Press the function key F1 below Edit to open the Patient Information window.
- 101 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
NOTE: The Address item is customized.
1. Press Tab or Shift + Tab to move the cursor to a textbox. Fill in the selected textbox.
2. Press Tab or Shift + Tab to move the cursor to a list box. Press the Up or Down arrow to
select an option.
3. After inputting all the information, press
to confirm; or press Tab or Shift + Tab to
move the cursor to the OK button, and then press
to confirm.
For details on inputting data, please refer to Section 3.2, “Entering Data”.
10.7 Printing Files
When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1. Press the Up or Down arrow to highlight a file on the File Manager screen1,
and then press the function key F4 below Select to select the file and display the File Manager
screen2. Press START/STOP or the function key F2 below Record to print the selected file.
Pressing START/STOP or the function key F2 below Stop can stop printing the file.
10.8 Previewing a File
To Open the File preview screen
1. When the main screen is displayed, press the function key F7 below File to open the File
Manager screen1.
2. Press the Up or Down arrow to highlight a file on the File Manager screen1, and then press
the function key F4 below Select to select the file and display the File Manager screen2.
3. Press the function key F6 below Preview to open the file preview screen.
- 102 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
To Preview the File
1. Press the function key F1 below Waveform to display the waveform screen.
2. Press the function key F2 below Template to display the template screen.
3. Press the function key F3 below Measure to display the measure screen which shows Patient
Name, Patient ID, Heart Rate, P Duration, PR interval, QRS Duration, QT/QTC interval,
P/QRS/T axis, RV5/SV1 amplitude, RV5+SV1 amplitude, and RV6/SV2 amplitude.
- 103 -
SE-1200 Series Electrocardiograph User Manual
Managing Files
4. Press the function key F4 below Diagnosis to show Minnesota Code and Diagnosis
Information.
5. Press START/STOP or the function key F5 below Record to print the file. Pressing
START/STOP or the function key F5 below Stop can stop printing the file.
6. Press the function key F7 below Return to return to the File Manager screen2.
NOTE: Rhythm data can not be previewed.
- 104 -
SE-1200 Series Electrocardiograph User Manual
System Setup
Chapter 11 System Setup
After you turn on the device, the main screen pops up. Press the function key F6 below Setup to
display the System Setup screen.
Figure 11-1 System Setup screen
After setup,
1. Press ESC to return to the main screen.
2. Or press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor to
Return, and then press
to return to the main screen.
11.1 Work Mode Setup
When the cursor is on Work Mode on the System Setup screen, press
Work Mode Setup window.
- 105 -
to open the
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-2 Work Mode Setup window
After setup, press
OK button, and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the
to confirm. Then the System Setup screen appears.
11.1.1 Selecting Work Mode
Press Tab or Shift + Tab to move the cursor to a work mode, press Spacebar to select it, and then
a check mark √ appears in the box before the work mode.
NOTE: Only if a work mode is selected in the Work Mode Setup window, can the work
mode be selected by pressing the MODE key when the main screen is displayed.
MANU: In the manual mode, you can determine the lead group to be displayed and printed.
AUTO: In the auto mode, the lead groups are switched automatically according to the lead
sequence during the printing course. After the ECG waves of one lead group are printed
within a certain time, the system switches to print ECG waves of another lead group
automatically.
RHYT: In the rhythm mode, you can select rhythm leads to print 60s or 20s rhythm-lead ECG
waves.
- 106 -
SE-1200 Series Electrocardiograph User Manual
System Setup
R-R: In the R-R analysis mode, you can select a lead to print its R-R histogram, R-R trend chart,
180s compressed ECG waveform and all the R-R interval values.
OFF: In the off mode, the lead groups are switched automatically according to the lead sequence.
When the main screen is displayed, after pressing the START/STOP key, the sampled
ECG data can be saved and transmitted, but can not be printed.
11.1.2 Selecting Manual Style
Press Tab or Shift + Tab to move the cursor to the Manual Style list box in the Work Mode
Setup window, and then press the Up or Down arrow to highlight 3 channels, 6 channels or 12
channels.
When Manual Style is set to 3 channels, ECG waves of 3 leads are displayed and printed in the
manual mode.
When Manual Style is set to 6 channels, ECG waves of 6 leads are displayed and printed in the
manual mode.
When Manual Style is set to 12 channels, ECG waves of 12 leads are displayed and printed in
the manual mode.
11.1.3 Selecting Rhythm Style
Press Tab or Shift + Tab to move the cursor to the Rhythm Style list box in the Work Mode
Setup window, and then press the Up or Down arrow to highlight Single Lead or Three Leads.
When Rhythm Style is set to Single Lead, in the rhythm mode, 60s ECG waves of the appointed
single rhythm lead will be printed.
When Rhythm Style is set to Three Leads, in the rhythm mode, 20s ECG waves of three
appointed rhythm leads will be printed.
11.1.4 Selecting Auto Display Style
Press Tab or Shift + Tab to move the cursor to the Auto DispStyle list box in the Work Mode
Setup window, and then press the Up or Down arrow to highlight an option.
When the auto display style is set to 3×4, ECG waves of 12 leads are displayed in 4 groups of 3.
- 107 -
SE-1200 Series Electrocardiograph User Manual
System Setup
When the auto display style is set to 3×4+1R, ECG waves of 12 leads are displayed in 4 groups
of 3, with the ECG wave of one rhythm lead on the bottom.
When the auto display style is set to 3×4+3R, ECG waves of 12 leads are displayed in 4 groups
of 3, with ECG waves of three rhythm leads on the bottom.
When the auto display style is set to 6×2, ECG waves of 12 leads are displayed in 2 groups of 6.
When the auto display style is set to 6×2+1R, ECG waves of 12 leads are displayed in 2 groups
of 6, with the ECG wave of one rhythm lead on the bottom.
When the auto display style is set to 12×1, ECG waves of 12 leads are displayed on one screen
simultaneously.
11.1.5 Selecting Sampling Mode
Press Tab or Shift + Tab to move the cursor to the Sampling Mode list box in the Work Mode
Setup window, and then press the Up or Down arrow to highlight an option.
When Sampling Mode is set to Pre-Sample, 10s ECG data sampled before pressing the
START/STOP key will be printed out.
NOTE: When Sampling Mode is set to Pre-Sample, if you press the START/STOP key
before the electrocardiograph samples for 10s, the recorder will not respond.
When Sampling Mode is set to Real-time Sample, 10s ECG data sampled after pressing the
START/STOP key will be printed out.
When Sampling Mode is set to Triggered Sample, after pressing the START/STOP key, if
Arrhythmia ECG data, including Asystole, Ventricular Fibrillation/Ventricular Tachycardia,
5>PVCS>=3, Paired PVCS, Bigeminy, Trigeminy, R ON T, single PVC and Missed Beat, is
detected during the learning course, the printing will be triggered automatically.
When Sampling Mode is set to Periodic Sample, first you should set Duration and Interval.
For example, if Interval is set to 2 min, and Duration is set to 24 min, after pressing the
START/STOP key, the printing will be performed every two minutes and come to 12 times.
NOTE:
1. There is no saving, exporting or transmitting in the Periodic Sample mode.
2. ID and patient information will not be changed while carrying out the periodic printing.
- 108 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.1.6 Setting Interval and Duration
Press Tab or Shift + Tab to move the cursor to the Duration and Interval textboxes in the Work
Mode Setup window, and then input the duration and the interval manually.
In the auto mode, when Sampling Mode is set to Periodic Sample, if Interval is set to 2 min,
Duration is set to 24 min, after pressing the START/STOP key, the printing will be performed
every two minutes and come to 12 times.
11.1.7 Selecting Preview Function
Press Tab or Shift + Tab to move the cursor to Preview in the Work Mode Setup window, and
then press the Up or Down arrow to highlight Off or On.
When Preview is set to On, the preview screen will appear after you press the START/STOP
key in the auto or off mode.
When Preview is set to Off, the preview screen will not appear after you press the
START/STOP key.
11.1.8 Selecting Auto Extend Record
Press Tab or Shift + Tab to move the cursor to Auto Extend Record in the Work Mode Setup
window, and then press the Up or Down arrow to highlight Off or On.
When Auto Extend Record is set to On, if arrhythmia is detected in the auto or off mode, a hint
will pop up to ask you whether to print an extra rhythm report after the 12-lead ECG report.
11.2 Filter Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Filter on the System Setup screen, press
to open the Filter Setup
window.
In the Filter Setup window, you can set AC Filter, EMG Filter, DFT Filter, and Lowpass
Filter.
- 109 -
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-3 Filter Setup window
After setup, press
to confirm, or press Tab or Shift + Tab to move the cursor to the
OK button, and then press
to confirm. Then the System Setup screen appears.
11.2.1 Setting AC Filter
Press Tab or Shift + Tab to move the cursor to the AC Filter list box in the Filter Setup window.
Press the Up or Down arrow to highlight Off or On.
AC Filter suppresses AC interference without attenuating or distorting the ECG signals. Select
On to turn on the function and select Off to turn it off.
NOTE: AC frequency can be set to 50Hz or 60Hz on the Advanced Setup screen
according to local mains supply specifications.
11.2.2 Setting DFT Filter
Press Tab or Shift + Tab to move the cursor to the DFT Filter list box in the Filter Setup window.
Press the Up or Down arrow to highlight an option.
- 110 -
SE-1200 Series Electrocardiograph User Manual
System Setup
DFT Filter greatly reduces the baseline fluctuations without affecting the ECG signals. The
purpose of this filter is to keep the ECG signals on the baseline of the printout. The set value is
the low limit of the frequency range, including 0.05Hz, 0.15Hz, 0.25Hz, 0.32Hz, 0.5Hz and
0.67Hz.
11.2.3 Setting EMG Filter
Press Tab or Shift + Tab to move the cursor to the EMG Filter list box in the Filter Setup
window. Press the Up or Down arrow to highlight an option.
EMG Filter suppresses disturbance caused by strong muscle tremor. The cutoff frequency can be
set to 25Hz, 35Hz or 45Hz. Select Off to turn off the function.
11.2.4 Setting Lowpass Filter
Press Tab or Shift + Tab to move the cursor to the Lowpass Filter list box. Press the Up or
Down arrow to highlight an option.
Lowpass Filter restricts the bandwidth of input signals. The cutoff frequency can be set to 150Hz,
100Hz or 75Hz. All the input signals whose frequency is higher than the set cutoff frequency will
be attenuated.
NOTE:
1. Only when EMG Filter is set to Off, can the setting of Lowpass Filter be effective.
2. To the patient with a pacemaker, the EMG filter should be set to Off and the Lowpass
filter should be set to 150Hz; or else, the amplitude of the pulses may decrease, the
width may increase, and the diagnosis results may not include the pacemaker
information.
- 111 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.3 Record Info Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Record Info on the System Setup screen, press
to open the Record Info
Setup window.
You can press F1 or F2 to switch between Setup 1 and Setup 2.
After setup, press
OK button, and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the
to confirm. Then the System Setup screen appears.
11.3.1 Setup 1
Press F1 or F2 to switch to the Setup 1 tab.
Figure 11-4 Record Info Setup1
11.3.1.1 Selecting Auto Recording Style
Press Tab or Shift + Tab to move the cursor to the Record Style list box on the Setup 1 tab, and
then press the Up or Down arrow to highlight an option.
- 112 -
SE-1200 Series Electrocardiograph User Manual
System Setup
In the auto mode, when Record Style is set to 3×4, ECG waves of 12 leads are printed in 4
groups of 3.
In the auto mode, when Record Style is set to 3×4+1R, ECG waves of 12 leads are printed in 4
groups of 3, with the ECG wave of one rhythm lead on the bottom of the ECG reports.
In the auto mode, when Record Style is set to 3×4+3R, ECG waves of 12 leads are printed in 4
groups of 3, with the ECG waves of three rhythm leads on the bottom of the ECG reports.
In the auto mode, when Record Style is set to 6×2, ECG waves of 12 leads are printed in 2
groups of 6.
In the auto mode, when Record Style is set to 6×2+1R, ECG waves of 12 leads are printed in 2
groups of 6, with the ECG wave of one rhythm lead on the bottom of the ECG reports.
In the auto mode, when Record Style is set to 12×1, ECG waves of 12 leads are printed
simultaneously.
11.3.1.2 Selecting Recording Mode
Press Tab or Shift + Tab to move the cursor to the Record Mode list box on the Setup 1 tab, and
then press the Up or Down arrow to highlight Save Paper or Quickly.
When Record Mode is set to Save Paper, 10s after pressing the START/STOP key on the main
screen, an ECG report is printed. The patient information, measure information, interpretation
and ECG waves are printed on the same paper.
When Record Mode is set to Quickly, pressing the START/STOP key on the main screen to
begin printing an ECG report immediately, the patient information, measure information,
interpretation and ECG waves are printed on two pieces of paper.
NOTE:
1. In the auto mode, only when Sampling Mode is set to Real-time Sample, Quickly is
available.
2. When Record Style is set to 3×4, 3×4+1R or 3×4+3R, only Save Paper is available.
When Record Style is set to 12×1, only Quickly is available.
11.3.1.3 Selecting Recording Sequence
Press Tab or Shift + Tab to move the cursor to the Record Sequence list box on the Setup 1 tab,
and then press the Up or Down arrow to highlight Sequential or Synchronous.
When Record Sequence is set to Sequential, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the previous lead group.
- 113 -
SE-1200 Series Electrocardiograph User Manual
System Setup
When Record Sequence is set to Synchronous, all leads are printed simultaneously. The start
time of each group is the same.
11.3.1.4 Setting Gain
Press Tab or Shift + Tab to move the cursor to the Gain list box on the Setup 1 tab, and then
press the Up or Down arrow to highlight an option.
You can set the indicated height of 1mV ECG on the paper.
You can set Gain to 10mm/mV, 20mm/mV, 10/5mm/mV, AGC, 2.5mm/mV or 5mm/mV.
AGC means auto gain control. When ECG signals vary greatly, AGC can be selected to adjust
the gain automatically according to actual signals.
10/5mm/mV means that the gain of limb leads is set to 10mm/mV, while the gain of chest leads
is set to 5mm/mV.
11.3.1.5 Selecting Recording Device
Press Tab or Shift + Tab to move the cursor to the Record Device list box on the Setup 1 tab.
Press the Up or Down arrow to select an option.
You can set Record Device to Thermal, HP-DJ-D2468, HP-DJ-D5568, HP-DJ-D2668,
HP-DJ-D4368 or HP-LJ-P2035.
HP-DJ-D2468, HP-DJ-D5568, HP-DJ-D2668, HP-DJ-D4368 and HP-LJ-P2035 are the types
of the USB printer, when HP-DJ-D2468, HP-DJ-D5568, HP-DJ-D2668, HP-DJ-D4368 or
HP-LJ-P2035 is selected, you should connect the corresponding USB printer to USB socket 2 of
the electrocardiograph with a special cable. Press the START/STOP key to print ECG reports by
using the USB printer.
WARNING
If the printer used is not the type listed above, additional safety measures (such as
applying an isolation transformer to supply the medical system) should be taken when the
safety of the medical system has not been evaluated. If in doubt, consult our technical
service department or your local distributor
- 114 -
SE-1200 Series Electrocardiograph User Manual
Options
HP-DJ-D2468
HP-DJ-D5568
HP-LJ-P2035
System Setup
Corresponding USB Printers
HP Deskjet 3938
HP Deskjet D2368
HP Deskjet D2468
HP Deskjet D2568
HP Deskjet D5568
HP Laserjet P2015
HP Laserjet P2035
NOTE:
1. During the USB printing course,pressing the START/STOP key again can not stop
printing ECG reports.
2. For details of the ECG report printed by the USB printer, please refer to Section 8.6.5,
“ECG Reports Printed by the USB Printer”.
3. USB printing is ineffective in the auto periodic sampling mode, auto 11~24s sampling
mode, manual mode and R-R analysis mode.
4. Make sure that paper is installed in the USB printer before printing. Error may occur if
no paper is loaded in the USB Printer.
CAUTION
It is forbidden to frequently connect or disconnect the U disk or the USB printer.
11.3.1.6 Setting Speed
Press Tab or Shift + Tab to move the cursor to the Speed list box on the Setup 1 tab, and then
press the Up or Down arrow to highlight an option.
In the manual mode, you can set Speed to 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s or
50mm/s. In the auto and rhythm modes, only 25mm/s and 50mm/s are available. In the R-R
analysis mode, only 25mm/s is available.
11.3.1.7 Selecting Paper Marker
Press Tab or Shift + Tab to move the cursor to the Paper Marker list box on the Setup 1 tab,
and then press the Up or Down arrow to highlight No or Yes.
Paper Marker is used to identify the start point of each page of the recorder paper.
- 115 -
SE-1200 Series Electrocardiograph User Manual
System Setup
When the paper with black markers on the bottom is used and Paper Marker is set to Yes, the
device can identify the start point of each page of the recorder paper while printing ECG reports.
When Paper Marker is set to No, the device can not identify the start point of each page of the
recorder paper while printing ECG reports.
11.3.1.8 Setting Sampling Time
If Record Style is set to 12x1, you can set the time period. Press Tab or Shift + Tab to move the
cursor to the Sample Time textbox, and then input the time period manually.
If the time period is set from 11 to 24 seconds, in the auto mode, after pressing the
START/STOP key, the ECG waves of the set time period are printed.
NOTE: If the time period is set from 11 to 24 seconds, the ECG data sampled will not be
analyzed automatically or stored.
11.3.2 Setup 2
Press F1 or F2 to switch to the Setup 2 tab.
Figure 11-5 Record Info Setup2
Press Tab or Shift + Tab to move the cursor to an item, and then press Spacebar to select or
deselect the item.
- 116 -
SE-1200 Series Electrocardiograph User Manual
System Setup
Auto Record Info includes Patient Info, Position Marker, Analysis, Template, Measure,
Minnesota Code and Baseline Adjustment.
When Patient Info, Position Marker, Analysis, Template, Measure or Minnesota Code is
selected, it will be printed in the ECG reports.
When Baseline Adjustment is set to Auto, the baselines of the lead groups are adjusted
respectively.
When Baseline Adjustment is set to Horizontal, the baselines of the lead groups are adjusted
simultaneously, and the baselines of the leads in the same row are on the same line.
R-R Record Info includes R-R Histogram, R-R Waveform and R-R Interval List.
When R-R Histogram, R-R Waveform or R-R Interval List is selected, it will be printed in the
ECG reports.
Grid of Report includes Thermal Report and USB Report.
When Thermal Report is selected, the grid will be printed while printing ECG reports with the
thermal recorder.
When USB Report is selected, the grid will be printed while printing ECG reports with a USB
printer.
NOTE:
1. The items of Auto Record Info are available only in the auto mode, and Template and
Position Marker do not work in the auto style of 3×4, 3×4+1R and 3×4+3R, and the
style of 6x2 and 6x2+1R in the Save Paper mode.
2. To get more information about the above contents, please refer to Section 8.6.1, “ECG
Reports in the Auto Mode”.
11.4 Patient Information Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Patient Info on the System Setup screen, press
Information Setup window.
- 117 -
to open the Patient
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-6 Patient Information Setup window
After setup, press
to confirm, or press Tab or Shift + Tab to move the cursor to the
OK button, and then press
to confirm. Then the System Setup screen appears.
11.4.1 Selecting First/Last Name / Gender / D.O.B / Height / Weight
/ BP / Race / Pacemaker / Medication / Room No. / Department /
Physician / Technician / Ref-Physician
Press Tab or Shift + Tab to move the cursor to an item in the Patient Information Setup window,
and then press Spacebar to select or deselect the item.
When First/Last Name, Gender, D.O.B, Height, Weight, BP, Race, Pacemaker, Medication,
Room No., Department, Physician, Technician, or Ref-Physician is selected, the item will be
displayed in the Patient Information window.
11.4.2 Selecting ID Mode
Press Tab or Shift + Tab to move the cursor to ID Mode in the Patient Information Setup
window, and then press the Up or Down arrow to select a mode.
- 118 -
SE-1200 Series Electrocardiograph User Manual
System Setup
You can set ID Mode to Auto, Time or Manual.
When ID Mode is set to Auto, the patient ID can be automatically generated. The patient ID
range is 0~1999, 999, 999.
When ID Mode is set to Time, the patient ID can be automatically generated according to the
time when you press the START/STOP key to print an ECG report.
When ID Mode is set to Manual, you can enter the patient ID manually in the Patient
Information window. (Only 30 ASCII characters can be input.)
11.4.3 Selecting ID Hint
Press Tab or Shift + Tab to move the cursor to ID Hint in the Patient Information Setup window,
and then press the Up or Down arrow to select On or Off.
In the auto, rhythm or off mode, when ID Mode is set to Manual and ID Hint is set to On, if
you do not input the patient ID before pressing the START/STOP key, a hint will pop up to
remind you to input the patient ID.
11.4.4 Setting Prompt
Press Tab or Shift + Tab to move the cursor to Prompt in the Patient Information Setup window,
and then press the Up or Down arrow to select Confirmed By or Unconfirmed.
When Prompt is set to Confirmed By, if Physician is selected and the physician name is input
in the Patient Information window, the physician’s name is printed in the ECG reports.
When Prompt is set to Unconfirmed, Unconfirmed Report is printed in the ECG reports.
11.4.5 Selecting H/W Unit
Press Tab or Shift + Tab to move the cursor to H/W Unit in the Patient Information Setup
window, and then press the Up or Down arrow to select cm/kg or inch/lb..
11.4.6 Selecting BP Unit
Press Tab or Shift + Tab to move the cursor to BP Unit in the Patient Information Setup window,
and then press the Up or Down arrow to select mmHg or kpa.
- 119 -
SE-1200 Series Electrocardiograph User Manual
System Setup
When BP Unit is set to kpa, two extra edit boxes will be displayed in the Patient Information
window for inputting decimal fraction.
11.4.7 Selecting Next Patient
Press Tab or Shift + Tab to move the cursor to Next Patient in the Patient Information Setup
window. Press the Up or Down arrow to highlight Off or On.
When Next Patient is set to On, in the auto mode, press the START/STOP key to print an ECG
report, and the system will automatically open the next patient screen after a complete ECG
report is printed.
On the next patient screen, pressing the function key F1 below Same Patient can return to the
main screen, all the patient information will keep the same; pressing the function key F2 below
Next Patient can return to the main screen, all the patient information will be changed except
gender; pressing F3 can switch the auto recording style; pressing F4 can switch the paper speed,
and pressing F5 can switch the gain. Press the START/STOP key to print the previous ECG
report again, according to the settings of the auto style, speed and gain which are shown on the
bottom of the screen.
When Next Patient is set to Off, in the auto mode, the system will not automatically display the
next patient screen after a complete ECG report is printed.
- 120 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.4.8 Setting the Time to Clear Patient Information
Press Tab or Shift + Tab to move the cursor to Patient Info Cleared in the Patient Information
Setup window. Press the Up or Down arrow to highlight an option.
If Patient Info Cleared is set to After Recording, the patient information will be cleared after
the ECG report is printed out.
If Patient Info Cleared is set to After Recording And Lead Off, the patient information will be
cleared after the ECG report is printed out and all the leads are off.
NOTE: This function is unavailable in the manual mode.
11.4.9 Inputting Customized Information
Press Tab or Shift + Tab to move the cursor to the User Defined textbox in the Patient
Information Setup window, and then manually input customized information such as Address,
which will be displayed in the Patient Information window.
11.5 Transmission Setup
NOTE:
1. To transmit ECG data in DAT format to the PC, the Smart ECG Viewer software of the
manufacturer must be installed in the PC. You should log into the Smart ECG Viewer
software before transmission.
2. To transmit ECG data in SCP/FDA-XML/PDF format to the PC, the FTP receiving
software must be installed in the PC. You should log into the FTP receiving software
before transmission.
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Transmission on the System Setup screen, press
Transmission Setup window.
- 121 -
to open the
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-7 Transmission Setup window
After setup, press
to confirm, or press Tab or Shift + Tab to move the cursor to the
OK button, and then press
to confirm. Then the System Setup screen appears.
11.5.1 Setting Device No.
Press Tab or Shift + Tab to move the cursor to the Device No. textbox, and then fill in the
textbox.
If you input 1 in the Device No. textbox, after you save data to the U disk, the data will be in the
directory of ECGDATA\ECG-1\Store\Examination Date of the U disk; after you export files from
the
electrocardiograph
to
the
U
disk,
the
files
will
be
in
the
directory
of
ECGDATA\ECG-1\Export\Examination Date and Time of the U disk; after you import files from
the U disk to the electrocardiograph, the files in the directory of ECGDATA\ECG-1\Import of the
U disk will be imported.
- 122 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.5.2 Selecting Auto Transmission
Press Tab or Shift + Tab to move the cursor to Auto Transmission, and then press the Up or
Down arrow to select On or Off.
When Auto Transmission is set to On, in the auto or rhythm mode, ECG data will be transmitted
automatically after an ECG report is printed out; in the off mode, ECG data will be transmitted
automatically after the START/STOP key is pressed.
11.5.3 Selecting Transmission Mode
Press Tab or Shift + Tab to move the cursor to Transmission Mode, and then press the Up or
Down arrow to highlight Ethernet or UART.
When Transmission Mode is set to Ethernet, first connect the net port of the electrocardiograph
to the net port of the PC by using an Ethernet cable recommended by the manufacturer. Then set
all the items in the Transmission Setup window. In the auto, rhythm or off mode, ECG data can
be transmitted through the net cable automatically.
When Transmission Mode is set to UART, first connect serial port 1 of the electrocardiograph to
the serial port of the PC by using a serial cable recommended by the manufacturer. In the auto,
rhythm or off mode, ECG data can be transmitted through the serial cable automatically.
11.5.4 Setting FTP Information
To Enter FTP Path, FTP User Name and Password
Press Tab or Shift + Tab to move the cursor to the FTP Path, FTP User Name, and Password
textboxes, and then fill in these textboxes.
11.5.5 Setting IP Addresses
To Set Server IP
Set the Server IP item to the IP of the PC.
- 123 -
SE-1200 Series Electrocardiograph User Manual
System Setup
To Set Local IP
For the cross-network transmission, set the first two sections of the Local IP item to the first two
sections of the IP of the PC. Set the third section of the Local IP item to the network segment of
the electrocardiograph which depends on the configuration of Router. The last section of the
Local IP item can be set at random.
For the same network transmission, set the first three sections of the Local IP item to the first
three sections of the IP of the PC. The last section of the Local IP item can be set at random, but
it can’t be the same as the last section of the IP of the PC.
To Set Gateway
Set the first three sections of the Gateway item to the first three sections of the IP of the PC. The
last section of the Gateway item must be set to 1.
To Set Subnet Mask
You can set Subnet Mask in the Transmission Setup window. Commonly, Subnet Mask is set to
255.255.255.000.
11.6 Lead Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Lead on the System Setup screen, press
window.
- 124 -
to open the Lead Setup
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-8 Lead Setup window
After setup, press
OK button, and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the
to confirm. Then the System Setup screen appears.
11.6.1 Setting Rhythm Lead1/2/3
Press Tab or Shift + Tab to move the cursor to the Rhythm Lead1/2/3 list box in the Lead Setup
window, and then press the Up or Down arrow to highlight an option.
The rhythm lead can be one of 12 standard leads: І, П, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5,
or V6.
In the auto mode, when Record Style is set to 3×4+1R or 6×2+1R, the rhythm lead selected in
the Rhythm Lead1 list box will be printed in the ECG reports; when Record Style is set to
3×4+3R, 3 rhythm leads selected respectively in the Rhythm Lead1/2/3 list box will be printed
in the ECG reports.
In the rhythm mode, when Rhythm Style is set to Single Lead, 60s wave of the rhythm lead
selected in the Rhythm Lead1 list box will be printed in the ECG reports; when Rhythm Style is
set to Three Leads, 20s waves of three rhythm leads selected respectively in the Rhythm
Lead1/2/3 list box will be printed in the ECG reports.
- 125 -
SE-1200 Series Electrocardiograph User Manual
System Setup
In the R-R analysis mode, the R-R analysis report of the rhythm lead selected in the Rhythm
Lead1 list box will be printed.
11.6.2 Setting Lead Sequence
Press Tab or Shift + Tab to move the cursor to the Lead Sequence list box in the Lead Setup
window, and then press the Up or Down arrow to highlight Standard or Cabrera.
Lead Sequence
Lead group 1
Lead group 2
Lead group 3
Lead group 4
Standard
І, II, Ш
aVR, aVL, aVF
V1, V2, V3
V4, V5, V6
Cabrera
aVL, І, -aVR
II, aVF, Ш
V1, V2, V3
V4, V5, V6
11.6.3 Selecting Nehb
Press Tab or Shift + Tab to move the cursor to the Nehb list box in the Lead Setup window, and
then press the Up or Down arrow to highlight Off or On.
Lead Sequence: I, II, III, ND, NA, NI
NOTE: If you set Nehb to On, the working mode is fixed to be manual.
11.7 Display&Sound Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Display&Sound on the System Setup screen, press
to open the
Display&Sound Setup window.
After setup, press
OK button, and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the
to confirm. Then the System Setup screen appears.
- 126 -
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-9 Display&Sound Setup window
11.7.1 Setting Brightness (Optional)
Press Tab or Shift + Tab to move the cursor to the Brightness progress bar in the
Display&Sound Setup window, and then press the Left or Right arrow to adjust the brightness of
the LCD screen. You can set the brightness within 0~20.
11.7.2 Selecting Display Colors (Only for SE-1200 Express)
Press Tab or Shift + Tab to move the cursor to the Display Colors list box in the Display&Sound
Setup window, and then press the Up or Down arrow to highlight Monochrome, Option1 or
Option2.
11.7.3 Selecting Antialiasing
Press Tab or Shift + Tab to move the cursor to the Antialiasing list box in the Display&Sound
Setup window, and then press the Up or Down arrow to highlight Off or On.
When Antialiasing is set to On, the system will automatically make the waveform smooth.
When Antialiasing is set to Off, the system will not make the waveform smooth.
- 127 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.7.4 Setting Key Volume
Press Tab or Shift + Tab to move the cursor to the Key Volume list box in the Display&Sound
Setup window, and then press the Up or Down arrow to highlight an option.
When you press keys on the keyboard, the electrocardiograph gives a short sound. You can set
Key Volume to Low, Medium, High or Off. When it is set to Off, there is no sound when you
press keys.
11.7.5 Setting Hint Volume
Press Tab or Shift + Tab to move the cursor to the Hint Volume list box in the Display&Sound
Setup window, and then press the Up or Down arrow to highlight an option.
When the electrocardiograph displays a hint such as Lead Off, Overload, Battery Weak etc., there
is a sound. You can set Hint Volume to Low, Medium, High or Off. When it is set to Off, there
is no hint sound.
11.7.6 Setting QRS Volume
Press Tab or Shift + Tab to move the cursor to the QRS Volume list box in the Display&Sound
Setup window, and then press the Up or Down arrow to highlight an option.
When an R wave is detected, there is a sound. You can set QRS Volume to Low, Medium, High
or Off. When it is set to Off, there is no sound when an R wave is detected.
11.8 Date&Time Setup
NOTE: Please set DATE&TIME correctly when it's the first time you use the
electrocardiograph.
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Date&Time on the System Setup screen, press
Setup window.
- 128 -
to open the Date&Time
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-10 Date&Time Setup window
After setup, press
to confirm, or press Tab or Shift + Tab to move the cursor to the
OK button, and then press
to confirm. Then the System Setup screen appears.
11.8.1 Setting Date and Time
Press Tab or Shift + Tab to move the cursor to the Date or Time textbox in the Date&Time Setup
window, and then input the date or the time manually.
The time will be displayed on the main screen, and the date and the time will be printed in the
ECG reports.
11.8.2 Setting Date Mode
Press Tab or Shift + Tab to move the cursor to the Date Mode list box in the Date&Time Setup
window, and then press the Up or Down arrow to highlight DD-MM-YYYY, MM-DD-YYYY or
YYYY-MM-DD.
NOTE: After setup, press
to confirm, or press Tab or Shift + Tab to move the
cursor to the OK button, and then press
will become effective.
- 129 -
to confirm. Then the new setup
SE-1200 Series Electrocardiograph User Manual
System Setup
11.8.3 Setting Time Mode
Press Tab or Shift + Tab to move the cursor to the Time Mode list box, and then press the Up or
Down arrow to highlight 24 Hours or 12 Hours.
11.8.4 Setting Power-Off Time
Press Tab or Shift + Tab to move the cursor to the Power Off textbox in the Date&Time Setup
window, and then input the power-off time manually.
When Power Off is not set or is set to 0 Minutes, this function will not be effective.
NOTE:
1. Power-off time is counted from the time when you last press the keys on the keyboard.
2. Only when the device is powered by battery, can the set automatic power-off time be
effective.
11.8.5 Setting LCD Off Time
Press Tab or Shift + Tab to move the cursor to the LCD Off textbox in the Date&Time Setup
window, and then input the LCD off time manually.
When LCD Off is not set or is set to 0 Minutes, this function will not be effective.
NOTE: LCD Off time is counted from the time when you last press the keys on the
keyboard.
11.9 File Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on File on the System Setup screen, press
- 130 -
to open the File Setup window.
SE-1200 Series Electrocardiograph User Manual
System Setup
Figure 11-11 File Setup window
After setup, press
to confirm, or press Tab or Shift + Tab to move the cursor to the
OK button, and then press
to confirm. Then the System Setup screen appears.
11.9.1 Selecting Auto Save
Press Tab or Shift + Tab to move the cursor to the Auto Save list box in the File Setup window,
and then press the Up or Down arrow to highlight Off or On.
When Auto Save is set to On, ECG data in the auto, rhythm or off mode will be saved in the
ECG or the U disk automatically.
When Auto Save is set to Off, ECG data will not be saved.
11.9.2 Setting Saving Path
Press Tab or Shift + Tab to move the cursor to the Saving Path list box in the File Setup window,
and then press the Up or Down arrow to highlight ECG or U Disk.
When Saving Path is set to ECG, if Auto Save is set to On, ECG data in the auto, rhythm or off
mode will be saved in the ECG automatically.
When Saving Path is set to U Disk, if Auto Save is set to On, ECG data in the auto, rhythm or
off mode will be saved in the U disk automatically.
- 131 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.9.3 Setting File Format
Press Tab or Shift + Tab to move the cursor to File Format in the File Setup window, and then
press the Up or Down arrow to highlight an option.
If you want to set File Format to SCP/FDA-XML/PDF, first activate the SCP/FDA-XML/PDF
function on the Advanced Setup screen For details on activating the SCP/FDA-XML/PDF
function, please refer to Section 9.1.3, “Transmitting ECG Data in SCP/FDA-XML/PDF Format
Through Net Cable”.
11.9.4 Selecting Deleting Files After Transmission or Export
Press Tab or Shift + Tab to move the cursor to Delete After Trans. Or Export in the File Setup
window, and then press the Up or Down arrow to highlight Off or On.
When Delete After Trans. Or Export is set to On, the files will be automatically deleted from
the File Manager screen after they are transmitted to the PC or exported to the U disk.
11.9.5 Selecting Replacing When Memory is Full
Press Tab or Shift + Tab to move the cursor to Replace When Memory Full in the File Setup
window, and then press the Up or Down arrow to highlight Off or On.
When Replace When Memory Full is set to On, if there is no space on the memory, the files
will replace the earliest ones automatically.
11.9.6 Selecting SCP File Compression
Press Tab or Shift + Tab to move the cursor to SCP File Compression in the File Setup window,
and then press the Up or Down arrow to highlight On or Off.
If SCP File Compression is set to On, the SCP file will be compressed.
After the SCP function is activated, SCP File Compression appears in the File Setup window.
For details on activating the SCP function, please refer to Section 9.1.3, “Transmitting ECG Data
in SCP/FDA-XML/PDF Format Through Net Cable”.
- 132 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.10 System Maintenance Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Maintenance on the System Setup screen, press
to open the System
Maintenance window.
Figure 11-12 System Maintenance window
After setup, press
OK button, and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the
to confirm. Then the System Setup screen appears.
11.10.1 Exporting System Settings to U Disk
1. Connect a U disk to the electrocardiograph.
2. Press Tab or Shift + Tab to move the cursor to the Export Settings To U disk button in the
System Maintenance window, and then press
disk.
- 133 -
to export the system settings to the U
SE-1200 Series Electrocardiograph User Manual
System Setup
11.10.2 Loading System Settings from U Disk
1. Connect the U disk containing the system settings to the electrocardiograph.
2. Press Tab or Shift + Tab to move the cursor to the Load Settings From U disk button in the
System Maintenance window, and then press
to load the system settings from the
U disk to the electrocardiograph.
11.10.3 Backing Up Settings
Press Tab or Shift + Tab to move the cursor to the Back up Settings button in the System
Maintenance window, and then press
to back up the system settings to the ECG.
11.10.4 Loading Backup Settings
Press Tab or Shift + Tab to move the cursor to the Load Backup Settings button in the System
Maintenance window, and then press
to load the backup settings from the ECG.
11.10.5 Printing System Settings
Press Tab or Shift + Tab to move the cursor to the Record System Settings button in the System
to print the system settings. Pressing this button
Maintenance window, and then press
again can stop printing system settings.
11.10.6 Loading Factory Settings
Press Tab or Shift + Tab to move the cursor to the Load Factory Settings button in the System
Maintenance window, and then press
to restore the factory settings shown in the
following table.
- 134 -
SE-1200 Series Electrocardiograph User Manual
System Setup
Table 11-1 System Setup Defaults
Setup
Default
1
Mode Options
Auto, Manual, Rhythm
2
Manual Style
6 channels
3
Auto Display Style
6×2
4
Rhythm Style
Three Leads
5
Sampling Mode
Real-time Sample
6
Duration (Periodic Sample)
60 min
7
Interval (Periodic Sample)
1 min
8
Preview
Off
9
Auto Extend Record
Off
10
AC filter
On
11
EMG filter
Off
12
DFT filter
0.67Hz
13
Lowpass filter
100Hz
14
Record Style
6×2
15
Record Mode
Save Paper
16
Record Sequence
Sequential
17
Gain
10mm/mV
18
Record Device
Thermal
19
Speed
25mm/s
20
Paper Marker
Yes
21
Sample Time
10s
22
Patient Information
On
23
Template
On
24
Position Marker
On
25
Measure
On
26
Analysis
On
- 135 -
SE-1200 Series Electrocardiograph User Manual
System Setup
27
Minnesota Code
Off
28
Baseline Adjustment
Horizontal
29
R-R Histogram
On
30
R-R Interval List
On
31
R-R Waveform
On
32
Grid of Thermal Report
Off
33
Grid of USB Report
On
34
Gender\Height\Weight
On
35
First/Last Name \ D.O.B \ BP \ Race
\ Pacemaker \ Medication \ Room
No. \ Department \ Physician \
Technician \ Ref-Physician
Off
36
ID Hint
On
37
Next Patient
Off
38
ID Mode
Auto
39
Patient Information Cleared
After Recording
40
H/W Unit
cm/kg
41
BP Unit
mmHg
42
Prompt
Confirmed By
43
User Defined
Cleared
44
Auto Transmission
Off
45
Transmission Mode
Ethernet
46
FTP User Name/FTP Password/FTP
Path
Cleared
47
Lead Sequence
Standard
48
Nehb
Off
49
Rhythm Lead 1
II
50
Rhythm Lead 2
V1
51
Rhythm Lead 3
V5
52
Screen Colors(Only for SE-1200
Express)
Option 1
- 136 -
SE-1200 Series Electrocardiograph User Manual
System Setup
53
Antialiasing
Off
54
QRS Volume
Off
55
Hint Volume
Medium
56
Key Volume
Medium
57
Date Mode
DD-MM-YYYY
58
Time Mode
24 Hours
59
Power Off
0
60
LCD Off
0
61
Auto Save
On
62
Saving Path
ECG
63
File Format
DAT
64
Delete After Trans. Or Export
Off
65
Replace When Memory Full
Off
66
SCP File Compression
On
67
System Password
Cleared
68
Caps Lock
Off
69
Pacemaker Detection Sensitivity
Low
Table 11-2 System Setup Defaults (only in the U.S.)
Setup
Default
1
Mode Options
Auto, Manual, Rhythm, OFF, R-R
2
Manual Style
6 channels
3
Auto Display Style
3×4+1
4
Rhythm Style
Three Leads
5
Sampling Mode
Real-time Sample
6
Duration (Periodic Sample)
60 min
7
Interval (Periodic Sample)
1 min
8
Preview
Off
9
Auto Extend Record
Off
- 137 -
SE-1200 Series Electrocardiograph User Manual
System Setup
10
AC filter
On
11
EMG filter
Off
12
DFT filter
0.67Hz
13
Lowpass filter
100Hz
14
Record Style
3×4+1
15
Record Mode
Save Paper
16
Record Sequence
Sequential
17
Gain
10mm/mV
18
Record Device
Thermal
19
Speed
25mm/s
20
Paper Marker
Yes
21
Sample Time
10s
22
Patient Information
On
23
Template
On
24
Position Marker
On
25
Measure
On
26
Analysis
On
27
Minnesota Code
Off
28
Baseline Adjustment
Horizontal
29
R-R Histogram
On
30
R-R Interval List
On
31
R-R Waveform
On
32
Grid of Thermal Report
Off
33
Grid of USB Report
On
34
Gender\Height\Weight\D.O.B
On
35
First/Last Name \ BP \ Race \
Pacemaker \ Medication \ Room No.
\ Department \ Physician \
Technician \ Ref-Physician
Off
36
ID Hint
On
- 138 -
SE-1200 Series Electrocardiograph User Manual
System Setup
37
Next Patient
Off
38
ID Mode
Manual
39
Patient Information Cleared
After Recording
40
H/W Unit
inch/lb
41
BP Unit
mmHg
42
Prompt
Confirmed By
43
User Defined
Cleared
44
Auto Transmission
Off
45
Transmission Mode
Ethernet
46
FTP User Name/FTP Password/FTP
Path
Cleared
47
Lead Sequence
Standard
48
Nehb
Off
49
Rhythm Lead 1
II
50
Rhythm Lead 2
V1
51
Rhythm Lead 3
V5
52
Screen Colors(Only for SE-1200
Express)
Option 1
53
Antialiasing
Off
54
QRS Volume
Off
55
Hint Volume
Medium
56
Key Volume
Medium
57
Date Mode
DD-MM-YYYY
58
Time Mode
24 Hours
59
Power Off
0
60
LCD Off
0
61
Auto Save
On
62
Saving Path
ECG
63
File Format
PDF
64
Delete After Trans. Or Export
Off
- 139 -
SE-1200 Series Electrocardiograph User Manual
System Setup
65
Replace When Memory Full
Off
66
SCP File Compression
On
67
System Password
Cleared
68
Caps Lock
Off
69
Pacemaker Detection Sensitivity
Low
11.10.7 Setting System Password
Press Tab or Shift + Tab to move the cursor to the System Password textbox in the System
Maintenance window, and then enter the password.
If you set the system password, after you press the function key F6 below Setup, the System
Password window will pop up. After you enter the correct password, the System Setup screen
will be displayed.
11.11 Other Setup
Press Tab or Shift + Tab, or the Up, Down, Left or Right arrow to move the cursor. When the
cursor is on Other on the System Setup screen, press
to open the Other Setup
window.
Figure 11-13 Other Setup Window
After setup, press
OK button, and then press
to confirm, or press Tab or Shift + Tab to move the cursor to the
to confirm. Then the System Setup screen appears.
- 140 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.11.1 Entering Institution
Press Tab or Shift + Tab to move the cursor to the Institution textbox on the Other Setup
Window, and then manually input the institution name within 40 ASCII characters.
NOTE: The total number of supported characters may be fewer if either special Latin
characters or Chinese characters are entered.
11.11.2 Selecting a Language
Press Tab or Shift + Tab to move the cursor to Language on the Other Setup Window, and then
press the Up or Down arrow to highlight a language.
11.11.3 Selecting Caps Lock
Press Tab or Shift + Tab to move the cursor to Caps Lock on the Other Setup Window, and
then press the Up or Down arrow to highlight Off or On.
When Caps Lock is set to On, the letters entered will be capital. Pressing Shift
and a
letter key can input a lowercase letter.
When Caps Lock is set to Off, the letters entered will be lowercase. Pressing Shift
and a
letter key can input a capital letter.
11.11.4 Setting External Input
Press Tab or Shift + Tab to move the cursor to the External Input list box on the Other Setup
Window, and then press the Up or Down arrow to highlight Off or On.
The extern input socket is equipped in the electrocardiograph, through which the
electrocardiograph can receive signals from the external equipment.
NOTE: When External Input is set to On, the touch screen becomes ineffective, and
then you should operate the device by using the keyboard.
- 141 -
SE-1200 Series Electrocardiograph User Manual
System Setup
11.11.5 Setting External Output
Press Tab or Shift + Tab to move the cursor to the External Output list box on the Other Setup
Window, and then press the Up or Down arrow to highlight an option.
The extern output socket is equipped in the electrocardiograph, through which the
electrocardiograph can send signals to the external equipment.
If External Input is set to On, and External Output is set to Standard or Triggered, the
electrocardiograph sends the signals which it receives.
If External Input is set to Off, and External Output is set to Standard, the electrocardiograph
sends ECG signals of rhythm lead I.
If External Input is set to Off, and External Output is set to Triggered, the electrocardiograph
sends pulses with the height of 4V and the width of 40ms, based on the data of rhythm lead I.
11.11.6 Setting Pacemaker Detection Sensitivity
Press Tab or Shift + Tab to move the cursor to Pacemaker Detection Sens. on the Other Setup
Window, and then press the Up or Down arrow to highlight Low or High.
When Pacemaker Detection Sens. is set to High, the pacemaker signals are easy to be detected.
When Pacemaker Detection Sens. is set to Low, the pacemaker signals are not easy to be
detected.
For more information about the detectable parameter ranges, please refer to Appendix 1,
“Technical Specifications”.
NOTE: If you select Pacemaker in the Patient Information Setup window, the pacemaker
sensitivity can only be set in the Patient Information window, and the setting of
Pacemaker Detection Sens. on the Other Setup Window is ineffective.
- 142 -
SE-1200 Series Electrocardiograph User Manual
Turning off the Electrocardiograph
Chapter 12 Turning off the Electrocardiograph
When the battery is used, hold down the
key to display the hint System is shutting down…
on the screen. Then the device will be off a few seconds later.
When the mains supply is used, hold down the
key to display the hint System is shutting
down… on the screen. Then the device will be off a few seconds later. Remove the plug from the
outlet.
NOTE:
1. When turning off the device, follow the above sequence strictly, or else there will be
something wrong on the screen.
2. Do not hold down the
key when the device displays the hint information System
is shutting down… on the screen.
- 143 -
SE-1200 Series Electrocardiograph User Manual
Hint Information
Chapter 13 Hint Information
Hint information and the corresponding causes provided by the electrocardiograph are listed in
Table 13-1.
Table 13-1 Hint Information and Causes
Hint Information
Causes
Lead off
Electrodes fall off the patient or the patient cable falls off the unit,
or a high polarization voltage occurs.
Battery Weak
The battery is weak.
No Paper
Recorder paper runs out or is not loaded.
Paper Error
When Paper Marker is set to Yes, the electrocardiograph
advances the recorder paper to the next black marker. If it
advances the paper for 300mm and can not find the next black
marker, the hint Paper Error is displayed.
Testing
The ECG data is being sampled periodically.
Sampling/Analyzing/
ECG signals are being sampled / analyzed / recorded.
Recording
Learning
The self-study process of arrhythmia arithmetic in the Trigger
Sample mode
Detecting
The examining process of arrhythmia data in the Trigger Sample
mode
Transmitting
ECG data is being transmitted from the electrocardiograph to the
PC through the net or serial cable in the auto, rhythm or off mode.
Transmit Fail
ECG data fails to be transmitted from the electrocardiograph to the
PC through the net or serial cable in the auto, rhythm or off mode.
Memory Full
The amount of files on the File Manager screen of SE-1200
exceeds 100.
The amount of files on the File Manager screen of SE-1200
Express exceeds 200.
Module Error
There is something wrong with the signal sample module.
DEMO
The system is in the demonstration mode.
Overload
The direct current offset voltage on an electrode is too high.
U Disk / USB Printer A U disk, a USB printer or a bar code reader is connected to the
/ USB Scanner
USB interface.
- 144 -
SE-1200 Series Electrocardiograph User Manual
Cleaning, Care and Maintenance
Chapter 14 Cleaning, Care and Maintenance
CAUTION
1. Turn off the power before cleaning and disinfection. The mains supply must be
switched off if it is used.
2. Prevent the detergent from seeping into the equipment.
14.1 Cleaning
CAUTION
Any remainder of detergent should be removed from the unit and the patient cable after
cleaning.
14.1.1 Cleaning the Main Unit and the Patient Cable
To clean the electrocardiograph:
1.
Unplug the AC power cord.
2.
Wipe the external surfaces of the electrocardiograph with a soft cloth dampened in any of the
approved cleaning solutions listed below.
To clean the patient cable:
1.
Dampen a soft cloth with one of the disinfectants or cleaning agents listed below.
2.
Wring excess moisture from the cloth before cleaning.
Recommended Cleaning Solutions
♦ Mild soap and water
♦ 75% alcohol
14.1.2 Cleaning the Reusable Electrodes
To clean reusable electrodes:
1.
Remove the remainder gel from the electrodes with a clean soft cloth first.
2.
Take suction bulbs and metal cups of chest electrodes apart, and take clamps and metal parts
of limb electrodes apart.
- 145 -
SE-1200 Series Electrocardiograph User Manual
Cleaning, Care and Maintenance
3.
Clean them in warm water and make sure there is no remainder gel.
4.
Dry the electrodes with a clean dry cloth or air dry naturally.
CAUTION
Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
14.1.3 Cleaning the Print Head
To clean the print head:
1.
Open the recorder casing and remove the paper.
2.
Wipe the print head gently with a clean soft cloth damped in 75% alcohol.
For stubborn stain, soak it with a little alcohol first and wipe it off with a clean soft cloth.
3.
After air drying, load the recorder paper and shut the recorder casing.
Dirty and soiled thermal print head will deteriorate the printing definition. So it should be cleaned
at least once a month regularly.
CAUTION
Prevent the detergent from seeping into the main unit while cleaning. Do not immerse the
unit or the patient cable into liquid under any circumstances.
14.2 Disinfection
Disinfection of the main unit is not necessary need in daily maintenance, it is only
necessary in operating room. In that case, please use hospital standard disinfectant.
NOTE: Clean and disinfect the chest and limb electrodes after each use.
CAUTION
1. Do not use high-temperature, high-pressure vapour or ionizing radiation as
disinfection methods.
2. Do not use chloric disinfectant such as chloride, sodium hypochlorite etc.
- 146 -
SE-1200 Series Electrocardiograph User Manual
Cleaning, Care and Maintenance
14.3 Care and Maintenance
CAUTION
Operate the electrocardiograph, charge the battery, and store the battery at a
temperature of 40℃(104ºF) or lower. Exposure to higher or lower temperature may
reduce battery life, damage the battery, and degrade overall electrocardiograph
performance.
14.3.1 Recharge and Replacement of Battery
1) Capacity Identification
The current capacity of the battery can be identified according to the battery symbol in the top
right corner of the LCD screen.
: Full capacity;
: 2/3 capacity;
: 1/3 capacity;
: Empty
2) Recharge
The 12-channel electrocardiograph is equipped with the recharge control circuit together with
battery. When the unit is connected to the mains supply, the battery will be recharged
automatically. Then the battery recharging indicator (
) and the mains supply indicator (
will be lit at the same time. During the recharging course, the symbol
)
flashes in the top
right corner of the LCD screen. After the battery is fully recharged, the symbol stops flashing, and
the battery recharging indicator (
) is black.
Because of the capacity consumption during the storage and transport course, the battery capacity
is not full when it is used for the first time. Battery recharge should be considered before the first
use.
CAUTION
Repeated undercharging of the battery will damage the battery and reduce battery life.
3) Replacement
When the useful life of the battery is over, or foul smell and leakage are found, please contact the
manufacturer or the local distributor for replacement.
- 147 -
SE-1200 Series Electrocardiograph User Manual
Cleaning, Care and Maintenance
WARNING
1. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and the battery of the same model and
specification provided by the manufacturer must be used.
2. Danger of explosion -- Do not reverse the anode and the cathode when installing the
battery.
3. Remove the battery from the electrocardiograph when the electrocardiograph is not
used for a long time.
4. If the battery is stored alone and not used for a long time, we recommend that the
battery should be charged at least once every 6 months to prevent overdischarge.
5. When the battery’s useful life is over, contact the manufacturer or the local distributor
for disposal or dispose of the battery according to local regulations.
CAUTION
If the battery has been fully charged and requires recharging after printing only a few
ECGs, consider replacement.
14.3.2 Recorder Paper
NOTE: Recorder paper provided by the manufacturer should be used. Other paper may
shorten the life of the thermal print head. The deteriorated print head may lead to
illegible ECG reports and block the advance of the paper.
Storage Requirements:
♦
Recorder paper should be stored in a dry, dark and cool area, avoiding excessive
temperature, humidity and sunshine.
♦
Do not put the recorder paper under fluorescence for a long time.
♦
Make sure that there is no polyvinyl chloride or other chemicals in the storage
environment, which will lead to color change of the paper.
♦
Do not overlap the recorder paper for a long time, or else the ECG reports may
trans-print each other.
- 148 -
SE-1200 Series Electrocardiograph User Manual
Cleaning, Care and Maintenance
14.3.3 Visual inspection
Perform a visual inspection of all equipment and peripheral devices daily. If you notice any items
that need repair, contact a qualified service engineer to make the repairs.
♦
Check the case and display screen for cracks or other damage.
♦
Regularly inspect all plugs, cords, cables, and connectors for fraying or other damage.
♦
Verify that all cords and connectors are securely seated.
♦
Inspect keys and controls for proper operation.
14.3.4 Maintenance of the Main Unit and the Patient Cable
CAUTION
Besides the maintenance requirements recommended in this manual, comply with local
regulations on maintenance and measurement.
The following safety checks should be performed at least every 12 months by a qualified person
who has adequate training, knowledge, and practical experience to perform these tests.
a) Inspect the equipment and accessories for mechanical and functional damage.
b) Inspect the safety related labels for legibility.
c) Inspect the fuse to verify compliance with the rated current and circuit-breaking
characteristics.
d) Verify that the device functions properly as described in the instructions for use.
e) Test the protection earth resistance according to IEC/EN 60601-1: Limit: 0.1 ohm.
f) Test the earth leakage current according to IEC/EN/UL 60601-1: Limit: NC/SFC 300μA.
g) Test the enclosure leakage current according to IEC/EN/UL 60601-1: Limit: NC 100μA,
SFC 300μA.
h) Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c.
10μA; SFC a.c. 50μA, d.c. 50μA.
- 149 -
SE-1200 Series Electrocardiograph User Manual
Cleaning, Care and Maintenance
i) Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10μA, d.c.
10μA; SFC a.c. 50μA, d.c. 50μA.
j) Test the patient leakage current under single fault condition with mains voltage on the
applied part according to IEC/EN 60601-1: Limit: 50μA (CF).
The leakage current should never exceed the limit. The data should be recorded in an equipment
log. If the device is not functioning properly or fails any of the above tests, the device has to be
repaired.
WARNING
Failure on the part of the responsible individual hospital or institution employing this
equipment to implement a satisfactory maintenance schedule may cause undue
equipment failures and possible health hazards.
1) Main Unit
♦
Avoid excessive temperature, sunshine, humidity and dirt.
♦
Put the dustproof coat on the main unit after use and prevent shaking it violently when
moving it to another place.
♦
Prevent any liquid from seeping into the equipment; otherwise the safety and the
performance of the electrocardiograph can not be guaranteed.
2) Patient Cable
♦
Integrity of the patient cable, including the main cable and lead wires, should be
checked regularly. Make sure that it is conductible.
♦
Do not drag or twist the patient cable with excessive stress while using it. Hold the
connector plug instead of the cable when connecting or disconnecting the patient cable.
♦
Align the patient cable to avoid twisting, knotting or crooking in a closed angle while
using it.
♦
Store the lead wires in a big wheel to prevent any people from stumbling.
♦
Once damage or aging of the patient cable is found, replace it with a new one
immediately.
- 150 -
SE-1200 Series Electrocardiograph User Manual
Cleaning, Care and Maintenance
3) Reusable Electrodes
♦
Electrodes must be cleansed after use and make sure there is no remainder gel on them.
♦
Keep suction bulbs of chest electrodes away from sunshine and excessive temperature.
♦
After long-term use, the surfaces of electrodes will be oxidized because of erosion and
other causes. By this time, electrodes should be replaced to achieve high-quality ECG
records.
CAUTION
The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for
recycling or proper disposal.
- 151 -
SE-1200 Series Electrocardiograph User Manual
Accessories
Chapter 15 Accessory List
WARNING
Only the patient cable and other accessories supplied by the manufacturer can be used.
Or else, the performance and electric shock protection can not be guaranteed.
Accessory
Part Number
01.57.107581 (Snap Style)
ECG Cable (European)
01.57.107583 (Grabber Style)
01.57.107582 (Snap Style)
ECG Cable (American)
01.57.107584 (Grabber Style)
Patient Cable (American)
01.57.107048
Patient Cable (European)
01.57.106902
ECG Chest Electrodes
02.04.110842
ECG Limb Electrodes
02.04.110843
Pediatric
Electrodes
Disposable
Adhesive
11.57.40024
Disposable Neonatal Electrode
11.57.40109
Adult Disposable Adhesive Electrodes
11.57.040159
Disposable Resting Tab electrodes
(1 piece)
11.57.040189
Power cord (European)
01.13.36014
Power cord(American)
01.13.036106
Power cord (American)
11.13.36015
Rechargeable Li-ion Battery
21.21.064146
Snap/Banana Socket Adapters
01.13.107449
- 152 -
SE-1200 Series Electrocardiograph User Manual
Accessories
Clip/Snap/Banana Socket Adapter
01.57.040172
Alligator Clip/Banana Socket Adapters
01.57.040173
Paper Roller
01.51.30340-01
Thermal Recorder Paper
01.57.107371
Fuse
11.21.64073
Input/Output Connector
11.13.19907
Grounding Wire
11.13.114214
03.24.38952
Smart ECG Viewer Software
03.24.38953
Snap Clip
01.57.040172
Electrode Gel
11.25.78047
RS232 Cable
11.13.20117
Ethernet Cable
11.13.20096
Smart
ECG
viewer
Software
Key/TINY-SPRO
01.18.47116
Smart ECG viewer Software Key/USB
12.01.47194
DeskJet/HP2368
11.18.52168
U Disk
11.18.78105
ECG Bag
11.56.78118
SD Card Reader
11.17.047324
MT-801 Trolley
03.28.107634
12V Vehicle-carried Inverter
11.21.64056
Bar Code Reader
11.23.068003
Wireless AP
11.17.047338
The 12-channel electrocardiograph and accessories are available by contacting the manufacturer
or your local distributor.
NOTE:The chest electrodes, limb electrodes, pediatric chest electrodes and pediatric
limb electrodes are not available in the U.S.
- 153 -
SE-1200 Series Electrocardiograph User Manual
Warranty & Service
Chapter 16 Warranty & Service
16.1 Warranty
EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be
free from defects in materials and workmanship that occur within warranty period.
The warranty is void in cases of:
a) Damage caused by mishandling during shipping.
b) Subsequent damage caused by improper use or maintenance.
c) Damage caused by alteration or repair by anyone not authorized by EDAN.
d) Damage caused by accidents.
e) Replacement or removal of serial number label and manufacture label.
If a product covered by this warranty is determined to be defective because of defective materials,
components, or workmanship, and the warranty claim is made within the warranty period, EDAN
will, at its discretion, repair or replace the defective part(s) free of charge. EDAN will not provide
a substitute product for use when the defective product is being repaired.
16.2 Contact information
If you have any question about maintenance, technical specifications or malfunctions of devices,
contact your local distributor.
Alternatively, you can send an email to EDAN service department at: [email protected].
- 154 -
SE-1200 Series Electrocardiograph User Manual
Technical Specifications
Appendix 1 Technical Specifications
A1.1 Safety Specifications
Comply with:
IEC/EN 60601-1+A1+A2, IEC/EN 60601-1-2+A1, IEC/EN
60601-2-25, ANSI/AAMI EC11, IEC/EN 60601-2-51, UL
60601-1, CAN/CSA C22.2 No.601.1
Anti-electric-shock type:
Class І with internal power supply
Anti-electric-shock degree:
CF type with defibrillation-proof
Degree of protection against
Ordinary equipment (Sealed equipment without liquid proof)
harmful ingress of water:
Disinfection/sterilization
method:
Refer to the user manual for details
Degree
of
safety
of
Equipment not suitable for use in the presence of flammable
application in the presence of
gas
flammable gas:
Working mode:
Continuous operation
EMC:
Group І, Class A
Patient
NC
<10μA (AC) / <10μA (DC)
Current:
SFC
<50μA (AC) / <50μA (DC)
Patient
NC
<10μA (AC) / <10μA (DC)
SFC
<50μA (AC) / <50μA (DC)
Leakage
Auxiliary
Current:
- 155 -
SE-1200 Series Electrocardiograph User Manual
Technical Specifications
A1.2 Environment Specifications
Temperature:
Transport & Storage
Working
-20℃ (-4ºF) ~ +55℃ (+131ºF)
+5ºC (+41ºF) ~ +40ºC (+104ºF)
25%~93%
25%~80%
Non-Condensing
Non-Condensing
700hPa ~1060hPa
860hPa ~1060hPa
Relative Humidity:
Atmospheric Pressure:
A1.3 Physical Specifications
Dimensions
Weight
Display
420mm×330mm×105mm (16.5in×13.0in×4.1in)
Approx. 5.0kg (11.0 lbs)
(Excluding recorder paper and battery)
320×240 dot single color LCD Screen
800×600 multicolor LCD Screen
A1.4 Power Supply Specifications
Operating Voltage = 100V-240V~
Mains Supply:
Operating Frequency = 50Hz/60Hz
Input Current = 0.9-0.4A
Internal Li-ion Battery Pack:
SE-1200 Express: Rated voltage = 14.8V
SE-1200: Rated voltage = 14.8V
SE-1200 Express: Rated capacity = 4400mAh
SE-1200: Rated capacity = 2200mAh
When the battery is fully charged, SE-1200 can work
normally about 4 hours, and it can continually print about 1.5
hours in the manual mode or print about 300 ECG reports of
3×4+1R in the auto mode; SE-1200 Express can work
normally about 5 hours, and it can continually print about 2.5
hours in the manual mode or print about 350 ECG reports of
3×4+1R in the auto mode.
- 156 -
SE-1200 Series Electrocardiograph User Manual
Technical Specifications
Charge mode: Constant current/voltage
SE-1200 Express: Charge current (standard) = 0.14 C5A
(600mA)
SE-1200: Charge current (standard) = 0.28 C5A (600mA)
Charge voltage (standard) = (16.8-0.1V)
Necessary Charge time:
SE-1200 Express: 9.5 hours
SE-1200: 5 hours
Cycle life ≥ 300 times
Fuse:
T1AL250VP Ø5×20
A1.5 Performance Specifications
Recording
Recorder:
Printing Density
Thermal dot-matrix recorder
8 dots per mm / 200 dots per inch (amplitude axes)
40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s)
Folded thermal paper: 210mm×295mm×100pages
Recorder Paper:
Folded
thermal
paper:
215mm×280mm×100pages
(Optional)
Rolled thermal paper: 210mm×30m (Optional)
Effective Width:
Paper Speed:
Accuracy of data:
203mm
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s
(±3%)
±5% (x-axis), ±5%(y-axis)
HR Recognition
Technique:
Peak-peak detection
HR Range:
30 BPM ~300 BPM
Accuracy:
±1 BPM
ECG Unit
Leads:
12 standard leads
- 157 -
SE-1200 Series Electrocardiograph User Manual
Acquisition Mode:
simultaneously 12 leads
A/D Resolution:
24 bits
Time Constant:
≥3.2s
Frequency Response:
0.05Hz ~ 150Hz (-3dB)
Gain:
2.5, 5, 10, 20, 10/5, AGC (mm/mV)
Input Impedance:
≥50MΩ (10Hz)
Input Circuit Current:
≤0.01μA
Input Voltage Range
≤±5 mVpp
Calibration Voltage:
1mV±2%
DC Offset Voltage:
±600mV
Noise:
≤12.5 μVp-p
Multichannel crosstalk
≤0.5mm
Technical Specifications
AC Filter: On/Off
Filter
DFT Filter: 0.05Hz/0.15Hz/0.25Hz/0.32Hz/0.5Hz/0.67Hz
EMG Filter: 25Hz/35Hz/45Hz/OFF
LOWPASS Filter:150Hz/100Hz/75Hz
CMRR
≥115dB
Sampling Frequency
1000 Hz
Pacemaker Detection
Amplitude
±2 to ±700 mV
Width
0.1 to 2.0 ms
Sampling Frequency
10,000/sec/channel
External Input/Output (Optional)
Input
Output
≥100kΩ; Sensitivity 10mm/V±5%;
Single ended
≤100Ω; Sensitivity 1V/mV±5%;
Single ended
- 158 -
SE-1200 Series Electrocardiograph User Manual
Technical Specifications
NOTE:
1. Test the accuracy of input signal reproduction according to the methods described in
clause 4.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001, and the result complies with the
clause 3.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001.
2. Overall System Error is tested using the method described in AAMI EC11 3.2.7.1.
Overall System Error is not more than ±5%.
3. Frequency Response is tested using the method described in AAMI EC11 3.2.7.2
methods A and D.
- 159 -
SE-1200 Series Electrocardiograph User Manual
EMC Information
Appendix 2 EMC Information
Guidance and manufacture’s declaration - electromagnetic emissions - for all EQUIPMENT
and SYSTEMS
Guidance and manufacture’s declaration - electromagnetic emission
The 12-channel electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or the user of the 12-channel electrocardiograph should assure
that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
The 12-channel electrocardiograph uses RF
RF emissions
energy only for its internal function.
CISPR 11
Group 1
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emission
Class A
CISPR 11
The 12-channel electrocardiograph is suitable
for use in all establishments, other than
Harmonic emissions
Class A
domestic and those directly connected to the
IEC/EN 61000-3-2
public low-voltage power supply network that
Voltage fluctuations/
supplies buildings used for domestic purposes.
Complies
flicker emissions
IEC/EN 61000-3-3
- 160 -
SE-1200 Series Electrocardiograph User Manual
EMC Information
Guidance and manufacture’s declaration - electromagnetic immunity - for all
EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration - electromagnetic immunity
The 12-channel electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or the user of 12-channel electrocardiograph should assure that
it is used in such an environment.
IEC/EN 60601 test
Electromagnetic
Immunity test
Compliance level
level
environment - guidance
Electrostatic
Floors should be wood,
6 kV contact
6 kV contact
discharge (ESD)
concrete or ceramic tile. If
8 kV air
8 kV air
IEC/EN 61000-4-2
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical
fast 2 kV for power
Mains
power
quality
2 kV for power
transient/burst
supply lines
supply lines
should be that of a typical
IEC/EN 61000-4-4
commercial or hospital
environment.
Surge
Mains
power
quality
1 kV line to line
1 kV line to line
IEC/EN 61000-4-5 2 kV line to ground 2 kV line to
should be that of a typical
ground
commercial or hospital
environment.
Power frequency
3A/m
3A/m
Power frequency magnetic
(50Hz/60Hz)
fields should be at levels
magnetic field
characteristic of a typical
IEC/EN 61000-4-8
location in a typical
commercial or hospital
environment.
Voltage dips, short <5% UT
<5% UT
Mains
power
quality
interruptions
and (>95% dip in UT)
(>95% dip in UT)
should be that of a typical
voltage variations for 0.5 cycle
for 0.5 cycle
commercial or hospital
on power supply
environment. If the user of
input lines
40% UT
40% UT
the 12-channel
IEC/EN 61000-4-11 (60% dip in UT)
(60% dip in UT)
electrocardiograph requires
continued operation during
for 5 cycles
for 5 cycles
power mains interruptions,
it is recommended that the
70% UT
70% UT
(30% dip in UT)
(30% dip in UT)
12-channel
for 25 cycles
for 25 cycles
electrocardiograph
be
powered
from
an
<5% UT
<5% UT
uninterruptible
power
- 161 -
SE-1200 Series Electrocardiograph User Manual
NOTE
EMC Information
(>95% dip in UT)
(>95% dip in UT)
supply or a battery.
for 5 sec
for 5 sec
UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration - electromagnetic immunity - for EQUIPMENT
and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration - electromagnetic immunity
The 12-channel electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or the user of the 12-channel electrocardiograph should assure
that it is used in such an environment.
Immunity
IEC/EN 60601 test Compliance
Electromagnetic environment test
level
level
guidance
Portable
and
mobile
RF
communications equipment should be
used no closer to any part of the
12-channel
electrocardiograph,
including
cables,
than
the
recommended separation distance
Conducted RF 3 Vrms
3Vrms
calculated from the equation applicable
IEC/EN
150 kHz to 80 MHz
to the frequency of the transmitter.
61000-4-6
Recommended separation distance
Radiated RF
IEC/EN
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d  1 .2 P
d  1.2 P 80 MHz to 800 MHz
d  2.3 P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according
to
the
transmitter
manufacturer
and
d
is
the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,a should be
less than the compliance level in each
frequency range.b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
- 162 -
SE-1200 Series Electrocardiograph User Manual
EMC Information
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the 12-channel
electrocardiograph is used exceeds the applicable RF compliance level above, the
12-channel electrocardiograph should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the 12-channel electrocardiograph.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
- 163 -
SE-1200 Series Electrocardiograph User Manual
EMC Information
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM - for EQUIPMENT or
SYSTEM that is not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the 12-channel
Electrocardiograph
The 12-channel electrocardiograph is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the 12-channel
electrocardiograph can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
12-channel electrocardiograph as recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to frequency of transmitter
Rated
(m)
maximum
150 kHz to 80 MHz 80 MHz to 800 MHz
800 MHz to 2.5 GHz
output power of
d  1 .2 P
d  1 .2 P
d  2 .3 P
transmitter
(W)
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
- 164 -
SE-1200 Series Electrocardiograph User Manual
Appendix 3 Abbreviation
Abbr
English
LCD
Liquid Crystal Display
BP
Blood Pressure
ECG
Electrocardiogram/Electrocardiograph
HR
Heart Rate
aVF
Left Foot Augmented Lead
aVL
Left Arm Augmented Lead
aVR
Right Arm Augmented Lead
LA
Left Arm
LL
Left Leg
RA
Right Arm
RL
Right Leg
ID
Identification
AC
Alternating Current
USB
Universal Serial Bus
AGC
Auto Gain Control
NC
Normal Condition
SFC
Single Fault Condition
- 165 -