BRUGSANVISNING · USER MANUAL
XXL-Rehab Low Air Loss madras
XXL-Rehab Low Air Loss mattress
MAX 450 KG
Kan med fordel anvendes til:
- Bariatriske brugere med et længerevarende
sengeforløb.
- Bariatriske brugere, der har det meget varmt
og som på grund af deres overvægt eller anden
sygdom kun kan lejres i én stilling.
- Til den bariatriske bruger, der har brug for
et meget tørt og rent sengemiljø.
- Intensivpatienter.
Ideal use for:
- Bariatric users having to stay in bed.
- Bariatric users who are very hot and because
of their obesity or other disease only can be
positioned in one position.
- For the bariatric user who needs a very dry
and clean bed environment.
- Intensive care patients.
BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING ·
Brugsanvisning for XXL-Rehab Low Air Loss madras
ADVARSEL
ADVARSEL
– Dette produkt må ikke anvendes uden grundig instruktion fra uddannet
personale.
– Bruges produktet til personer, der vejer mere end den maksimale brugervægt,
er der risiko for at beskadige produktet.
– XXL-Rehab påtager sig intet ansvar for eventuelle skader på produktet eller
personskader forårsaget af forkert installation, forkert montering eller forkert
anvendelse af dette produkt.
– Hvis dele er beskadiget eller mangler, kontakt XXL-Rehab med det samme.
Brug IKKE erstatnings reservedele.
– XXL-Rehab kan ikke holdes ansvarlig for fejl eller ulykker, der opstår efter
reparationer uden skriftlig tilladelse fra XXL-Rehab.
Indikationer:
XXL-Rehab Low Air Loss helmadras er udviklet til forebyggelse og behandling
af tryksår. Systemet faciliterer blodcirkulation og nedsætter trykket i kontaktområdet. Systemet bør benyttes i samråd med læge eller terapeut. Dette
system egner sig til brug for brugere med høj risiko for at udvikle sår, samt til
brugere med sår til og med grad 4.
Kontraindikationer:
Brugere med flg. diagnoser må ikke benytte systemet:
Ustabile columna frakturer.
Ustabile frakturer generelt.
Konsulter altid læge eller terapeut vedr. brugeren før systemet tages i brug.
SIKKERHEDSVEJLEDNINGER
For at sikre korrekt brug af XXL-Rehab Low Air Loss madras skal brugsanvisningen læses grundigt før brug. Der skal rettes opmærksomhed specielt
mod advarslerne mærket med .
Desuden må madrassen kun betjenes af uddannet personale eller af personer,
der har fået instruktion i dens brug, da der ellers er fare for, at brugeren kan
komme til skade.
INDIKATIONER
Vær opmærksom! Læs omhyggeligt brugsvejledningen
Dobbelt isoleret (klasse II)
Overensstemmelses erklæring ifg direktivet for medicinsk udstyr
International elektronik godkendelse
IPX0 Nedsænk ikke apparatet i væske eller sprøjt væske direkte
på enheden
Type BF udstyr
XXL-Rehab Low Air Loss madras
Scan koden og få flere
produktinformationer.
2
For at sikre optimal udnyttelse af madrassen, check da at alle dele er ordentligt monteret og at madrassen er fæstnet til liggefladen. Sæt / læg ikke
noget ovenpå pumpe enheden. Se til at ledningen er under sengerammen
og ikke i vejen for hjulene eller andet der kan skade ledningen.
Red kun madrassen op med 1 lag for at tillade at vanddamp kan trænge
igennem betrækket. Kun ”åndbare” inkontinens stykker bør bruges sammen
med dette system.
Undgå at benytte systemet nær flammer, lightere eller cigaretter. Risiko
for antændelse er til stede. Systemet trækker luft fra omgivelserne og derfor
kan cigaret røg eller lignende skade delene i pumpen.
Systemet bør desinficeres grundigt ved overgang til anden patient/
bruger for at undgå krydskontamination.
Brugerens vægt må aldrig overstige madrassens max brugervægt på
450 kg.
Brug denne madras med den korrekte højde på sengehesten.
Er sengehestens højde ikke korrekt kan der ske alvorlige skader på
brugeren af madrassen.
· BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING
Dette produkt er IKKE AP/APG beskyttet.
Venderegimer er stadig relevante og nødvendige ved brug af dette
system.
Krav til omgivelserne ved opbevaring og transport:
Temperatur: 5~45℃
Luftfugtighed: 15%~60%
Atmosfærisk tryk: 700 – 1060 hPA
Der må ikke være fremtrædende objekter / genstande, skarpe
genstande eller sengefjedre under madrassen da disse kan punktere
luftcellerne.
ANVENDELSE
XXL-Rehab Low Air Loss helmadras kan med fordel anvendes på sygehuse, på
institutioner og i hjemmeplejen.
Madrassen er ideel til den bariatriske bruger, der vejer op til 450 kg, og som
har tryksår stadie 0-4, eller som er i høj risiko for at udvikle tryksår.
Hvis brugere vejer mere end 450 kg er der risiko for at madrassen vil gå i
stykker eller tage skade.
BESKRIVELSE
Low Air Loss systemet er helt unik til den bariatriske bruger, fordi Low Air
Loss med sin konstante luftcirkulation under bruger sikrer et tempereret og
meget tørt miljø. Derudover giver madrassen maksimal trykaflastning ved at
give brugeren støtte over den størst mulige liggeflade med det lavest mulige
tryk, uden at gå på kompromis med stabilitet og komfort.
CHECK-LISTE TIL UDPAKNING
Pumpe enhed:
Strømenhed x 1
Ledning x 1
Brugsvejledning x 1
Madras pakke:
Hel madras med betræk x 1
Funktioner
Kontrolpanelets funktioner
Tænd
Systemet er i stand-by funktionen og den orange LED lampe vil lyse når
vippe kontakten 0/1 på enhedens venstre side er tændt. Tryk på POWER
knappen for at starte enheden. LED lampen vil lyse grønt. Se vedr. CPR –
tømning af madras.
Patient vægt
Tryk pilene op / nedfor at justere madrassen til den individuelle patient. Angivelsen er kun omtrentlig. Tilpas evt også mængden af luft efter komfort.
Mode
Standard – er til patienter der vejer mellem 35-140 kg.
Bariatric – er til patienter der vejer mellem 160-450 kg.
Tryk og hold trykket i ca. 10 sekunder for at skifte mellem Standard og
Bariatric.
Upright
Gør at patienten ikke ”går igennem” når bruger sidder op. Funktionen skal
tages væk når brugeren lejres liggende igen.
Panel lock
Låser panelet så kun Power og max inflate kan benyttes. For at låse op tryk
på knappen i 3 sekunder.
Max inflate
Tryk på knappen for at fylde madrassen maksimalt. Tryk på knappen igen for
at returnere til oprindelige indstillinger. Sker det ikke vil systemet automatisk
efter 20 min returnere til de oprindelige indstillinger.
Enhedens højre side
2 Quick release CPC koblinger til at fæstne madrassens luftslanger i.
Enhedens venstre side
Vippe kontakt: Ved aktivering går systemet i stand-by.
Strøm indgang: Sæt stikket fast i modtageren.
Enhedens bagside
Krog: Kroge til brug når systemet skal hænge på en sengegavl. Krogene er
med fjeder hvorfor de slår ind når systemet tages ned.
Sikring: Kan åbnes for vedligehold.
Luftfilter og hætte: Det anbefales at filteret renses eller skiftes hver måned
for at sikre optimal drift af systemet. Filteret skal mindst skiftes en gang om
året.
Plejepersonalet / terapeuten bør altid kontrollere luften under brugeren
ved en håndprøve under brugerens bækken. Det gøres ved at føre hånden ind
i madrassen ovenpå den celle der er under brugerens tungeste del (bækken).
Kontrollér at brugeren ikke sidder igennem madrassen = brugeren skal være
sunket ca. halvvejs ned i madrassen.
XXL-Rehab Low Air Loss madras
Pumpen må KUN skilles ad af kvalificerede teknikere. Kontakt forhandleren for udførelse af alle former for service.
3
BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING ·
Brugsanvisning for XXL-Rehab Vekseltryksmadras
FØR BRUG
UNDER BRUG
Opsætning af madras og pumpe
1. Fjern den eksisterende madras fra sengen.
2. Placér XXL-Rehab Low Air Loss madrassen på liggefladen med logo i
fodenden. Fastgør madrassen til sengen i hver side, ved hjælp af stropperne. Efterse at alle sengens funktioner kan bruges når madrassen er
spændt fast på liggefladen, før næste trin udføres.
3. Hæng pumpeenheden på sengegavlen ved hjælp af krogene.
4. Tilslut madrassens luft slanger til quick release koblingerne på pumpe
enhedens udtag.
5. Sæt stikket i en kontakt og tænd for pumpe enheden på vippekontakten på enhedens højre side. Den orange LED lampe lyser for at vise at
systemet er på stand-by.
6. Tryk Power knappen for at starte luftpåfyldning af madrassen. Ved
aktivering af MAX INFLATE afkorter man fyldningstiden markant.
Patient transfer og transport
Transfer: Aktivér MAX INFLATE. Se efter at madrassen er fastspændt før
transfer.
TILPASNING TIL BRUGER
For at opnå normal funktion af madrassen igen, sæt da luftslangerne i igen
og POWER knappen igen, tænd systemet og genindstil brugerens vægt.
1. Brugeren skal ligge midt på madrassen. Tilpas madrassens interne tryk
alt efter brugerens estimerede vægt ved at trykke på weight (kg) knappen
på pumpeenhedens kontrolpanel. Hvis madrassen føles for blød - eller
for fast – så forøg eller formindsk madrassens interne tryk. Dog kun et
niveau ad gangen så systemet kan nå at stabilisere sig før den næste
ændring foretages. Dette fortsætter til et acceptabelt niveau er opnået.
2. Plejepersonale / terapeuter bør altid foretage en håndtest – som nævnt
under Funktioner. Det gøres ved at føre hånden ind i madrassen imellem
cellerne under brugerens tungeste del (bækken). Kontrollér at brugeren
ikke ”går igennem” madrassen = brugeren skal være sunket ca. halvvejs
ned i madrassen.
VIGTIGT: Madrassen må ikke redes hårdt op med et lagen. Det vil
reducere effekten af madrassen markant.
Transport: I tilfælde af en transport situation slukkes pumpeenheden og
ledning, pumpe og madras følger patienten til det nye sted, hvor pumpe­
enheden straks tilsluttes strøm.
XXL-Rehab Low Air Loss madrassen har en sikkerheds skumbase under
luftcellerne, men den er ikke beregnet til at ligge på i lang tid.
Når systemet startes op skal man begynde forfra og tilpasse madrassen til
bruger.
CPR/Hjertestop funktion
I tilfælde af hjertestop tryk på Standby / CPR og kobl evt. slangerne fra
pumpeenheden. Brugerens vægt vil tømme madrassen for luft.
Undersøg jævnligt madrassen for at sikre at den ikke er våd eller defekt
på andre måder.
CHECKLISTE FOR PROBLEMLØSNING
PROBLEM
INSPEKTIONS PROCEDURE
MULIGE LØSNINGER
Pumpe enheden virker ikke.
Tjek om stikket er sat korrekt i stikkontakten.
Sæt stikket korrekt i stikkontakten.
Tjek om vippekontakten er på ON position.
Sæt vippekontakten på ON position.
Tjek om HFI relæet er slået.
Sluk for enheden. Tjek den relevante sikring. Skift hvis det er
nødvendigt. Tjek pumpe enhedens sikring. Start pumpeenheden
igen som normalt.
Pumpe enheden reagerer ikke på mulige løsninger.
Kontakt forhandleren for assistance.
Tjek luftfilter for støv.
Rens eller udskift filteret.
Tjek om brugerens vægtangivelse er korrekt.
Øg eller formindsk vægt angivelsen til det passende tryk opnås,
eller skift til anden madras hvis brugerens vægt overstiger 450 kg.
Tjek om luftcellerne er i stykker, revnet eller lign.
Og om der kommer luft igennem manifold. *
Erstat med originale reservedele eller kontakt forhandleren.
Tjek at slangerne er korrekt sat sammen med pumpe enhedens
quick release koblinger.
Isæt luftslangerne korrekt til klik lyd høres.
XXL-Rehab Low Air Loss madras
Brugeren ”går igennem”!
4
Pumpe enheden kører, men madrassen
bliver ikke pustet op.
* Manifold er den slange på siden af madrassen, som alle celler får luft igennem.
· BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING · BRUGSANVISNING
SPECIFIKATIONER
Af madras og pumpe
Madras og pumpeenhed skal rengøres omhyggeligt ved skift mellem
brugere for at undgå kryds kontaminering. Følgende er en god retningslinje,
men lokale beslutninger vedr. rengøring bør ligeså følges.
Brugervægt max.
450 kg
Længde
200 cm
ADVARSEL! Tag altid stikket ud af væggen når madrassen skal rengøres.
Madrassens vægt
Daglige rengøringer kan udføres ved sengen med desinfektions midler og
vand samt en ren og tør klud. Alle ikke-phenolholdige midler kan bruges.
Bredde
90 / 100 / 120 cm
Højde
26 cm
12 / 16 kg
Pumpens vægt
4,6 kg
Pumpens mål
36 x 30 x 17 cm
Sektionsopdelt madras
Ja
Madrassen kan evt. desinficeres med Virkon fra KeboMed hvis der er spor af
blod, urin eller andre kropsvæsker.
Alarm ved lufttab
Ja
Daglig rengøring af pumpeenheden kan udføres med en hårdt opvredet klud
– enten opvredet i vand eller i vand med neutral sæbe. Sprøjt ALDRIG væske
direkte på enheden.
Batteribackup
Kun desinficerede og rengjorte systemer må lagres. Rul madrassen fra
hovedgærdet og ned, brug stropperne til at fixere den.
Af betræk
Maskinvask med varmt vand ved maksimum 90°C.
Må IKKE tumbles i tørretumbler. Må ikke tørres i tørreskab.
Må IKKE stryges.
Vær omhyggelig med at madras og betræk er HELT tørre før brugeren ligger
på overfalden igen.
Af luftfilter
Fjern luftfilteret fra undersiden af pumpe enheden ved at åbne filterhætten.
Filteret kan enten renses og genindsættes eller det kan udskiftes.
Filteret kan renses med mild sæbe og vand. Genindsæt det nu tørre og rene
filter og se efter at hætten er lukket korrekt.
Udskift filteret en gang om året og kompressoren hvert andet år for at
forlænge systemets levetid.
Ved returnering til service
Dette system er ikke service frit. Service og reparationer skal udføres af en
autoriseret tekniker eller repræsentant. Alle returnerede dele SKAL være
rengjorte og desinficerede før de sendes af sted til service. Ikke rengjorte
systemer og tilsølede systemer vil blive returneret uden service.
GARANTI
Garantiperioden for XXL-Rehab Low Air Loss madras er 1 år fra købsdato.
Fakturaen er “garantibevis”.
Produkter, der er blevet anvendt til andre formål end den tilsigtede anvendelse; eller som har været genstand for forsømmelighed, misbrug, forkert
opbevaring eller håndtering, fejlbetjening, uautoriserede modifikationer eller
skader uden normal slitage, som er afgjort af XXL-Rehab, er ikke omfattet af
denne garanti.
Hvis max brugervægt kapaciteten overskrides, vil garantien blive annulleret.
Enhver uautoriseret reparation af produktet / del af produktet, samt mani­
pulation med nogle af komponenterne, vil ugyldiggøre garantien.
Alarm ved strømsvigt
Ja
Nej
Lyntømningsfunktion
Ja
Statisk funktion
Ja
Anvendelig direkte på sengebund
Ja
Anvendelig til indstillelig sengebund
Ja
Brandhæmmende betræk
Ja
Inkontinens betræk
Ja
Allergivenligt betræk
Ja
Vask- og desinficerbart betræk
Ja
Tryksårsforebyggende stadie
0-4
Betræk
100% nylon med PU bagside
Celler
100% nylon med TPU laminering
Pumpe
ABS flamme resistent UL
AC100 - 120V/ 50 - 60 Hz 5A
AC220 - 240V/ 50 Hz 5A
Strøm input
Strøm forbrug
Normal brug: Max 350W
Ledning
H05VV-F3x1.00mm2
Sikring
T1A/ 250V
Krav til brugs-, transport- og
lagerforhold
Temperatur: 5°C~45°C
Fugtighed: 15%~60%
Atmosfærisk tryk: 700 - 1060 hPA
Elektrisk klassificering
Class II Type BF, dobbelt isoleret med /
uden funktionel jording.
IPX0, Nedsænk ikke apparatet i væske eller
sprøjt væske direkte på enheden Systemet er
ikke AP/APG beskyttet. Kontinuerlig operation.
Godkendt i henhold til
EN60601-1, EN60601-1-2, EN61000-3-2,
EN61000-3-3, UL2601-1
Elektromagnetisk kompatibilitet
EN60601-1-2:1993
Type BF Tillæg
ISO - klassifikation
PRODUCENT
Carilex
Turmbergstr. 9A
D-76227 Karlsruhe
Tyskland
ISO 18 12 18 - 28
FORHANDLER
XXL-Rehab
Fuglebækvej 1D
DK-2770 Kastrup
Danmark
Tlf.: +45 70 25 25 22
Fax: +45 70 25 25 44
www.xxl-rehab.com
11/2014/JP
XXL-Rehab Low Air Loss madras
VEDLIGEHOLDELSE OG RENGØRING
5
USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL ·
Instructions for using XXL-Rehab Low Air Loss mattress
CAUTION
CAUTION
– This product should not be used without proper instruction from a healthcare
professional.
– Using the product for individuals weighing more than the maximum weightbearing capacity may pose a risk for damaging the product.
– XXL-Rehab assumes no responsibility for any damage or injury caused by
improper installation, assembly or use of this product.
– If components are damaged or missing, contact XXL-Rehab immediately.
Do NOT use substitute parts.
– XXL-Rehab cannot be held liable for faults or accidents occurring after
repairs by service staff without written authorisation from XXL-Rehab.
To ensure proper use of the XXL-Rehab Low Air Loss Mattress, please
read these instructions thoroughly before using the product. In doing so,
pay particular attention to the warning notes marked with .
Also remember that in order to prevent the risk of injury to the user, the
mattress may be operated only by trained caregivers or aides or by individuals who have been instructed in its proper use.
INDICATIONS
Attention! Please read enclosed document thoroughly
Double insulated (Class II)
Declaration of Conformity to Medical Device Directive
International Electronics Approval
IPX0 Do not immerse power unit in liquid or spray liquids directly
on power unit
Type BF Applied Parts
Indications
This low air loss mattress replacement system is designed for patients who
endure pressure ulcer and potential patients who wish to reduce the likelihood of pressure ulcer. This device is intended to treat and prevent pressure
ulcers by facilitating blood circulation and decreasing pressure of each tissue’s contact area. Always consult a physician or health professional before
using this mattress system. This anti-decubitus mattress system is mainly for
high risk patient group.
Contraindications
Certain patient conditions are not suitable for using this type of device such
as fracture of instable vertebrae and illness of instable vertebrae.
Always consult a physician or health professional before using this device.
The use of this system does not replace the regular repositioning, monitoring, and nursing of the patient.
SAFETY NOTES
To ensure proper operation, please inspect and verify all parts are set up
properly and anchor securely. Do not place anything on top of the power
unit. Make sure power cord is underneath bed frame and free from hazard.
It is recommended to limit bed linens to a single layer in order to allow
the moisture escape efficiently through the coverlet. Only ”breathable”
incontinent pads are recommended for use with this mattress system.
Avoid using this device near open flames, lighters, or cigarettes. Flammability hazard exists. This device draws air from the surrounding. Thus,
cigarette smoking may damage internal component.
This system should be disinfected thoroughly between patients to avoid
cross contamination.
Verify patient weight does not exceed bed frame, bed rails, and this mattress system’s weight capacity.
XXL-Rehab Low Air Loss mattress
Use this mattress with proper side rails to ensure the gap between
the side rail and top of the mattress is small enough to prevent patient
from getting his or her head or neck into the gap. Fail to do so could
result in serious patient injury.
6
· USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL
This product is NOT AP/ APG protected.
Re-position the patient once awhile is still necessary when using this
mattress.
Control Panel Features
Power (on led/ standby-cpr led)
The system is in STANDBY-CPR mode and the orange LED will light when
the main power rocker switch on the left side panel is turned on. Press the
POWER button to turn on the system and the green LED will light.
Operating, transportation, and storage conditions:
Temperature: 5-45°C
Humidity: 15%-60%
Atmospheric pressure: 700-1060 hPA
Patient weight
Simply press the or button to adjust the patient weight according to
each individual. The scale is only an approximation. Please adjust the weight
setting if the mattress is too soft or firm
to suit each patient.
CAUTION: ENSURE THAT THERE ARE NO PROTRUDING OBJECTS,
SHARP POINTS OR BED SPRINGS UNDER THE MATTRESS AS THESE
COULD PUNCTURE THE AIR CELLS.
Mode
STANDARD mode is for patient who weights in the range of 35 kg - 140 kg.
BARIATRIC mode is for patient who weights in the range of 160 kg - 450 kg.
USE
Press and hold the desired mode button for 10 seconds to switch between
Standard and Bariatric mode.
The XXL-Rehab Low Air Loss mattress can be used in hospitals, institutions
and nursing homes.
The XXL-Rehab Low Air Loss mattress is ideal for bariatric users weighing up
to 450 kg, with a long history of pressure sores, with high risk of developing
new pressure sores or with existing pressure sores up to grade 4.
If users are weighing more than 450 kg there is a risk that the mattress
will break or take damage.
DESCRIPTION
Very immobile and bedridden patients benefit a lot from the features of the
XXL-Rehab Low Air Loss mattress. The constant flow of air helps reduce the
maceration of the skin due to moisture/sweat. The blower makes it easy to
max inflate the mattress within seconds, making it easy to help the patients
with hygiene, help turn the patient or provide stability for the patient so he
or she can mobilise himself/herself.
UNPACKING CHECKLIST
Power Unit Package:
Power unit x 1
Power cord x 1
Operation manual x 1
Mattress Package:
Mattress replacement unit with coverlet x 1
FUNCTIONS
Upright mode
Is used to prevent patient from bottoming out in sit up position. Make sure
UPRIGHT mode is switch off when the patient is laying down to avoid incorrect pressure setting.
Panel lock
Press PANEL LOCK button to lock out function buttons besides max inflate
and power functions from unauthorized changes. If you wish to cancel panel
lock, simply press and hold the PANEL LOCK button for 3 seconds.
Max inflate
Press the MAX INFLATE button to inflate the mattress rapidly to maximum
pressure. Press the button again to restore to the original setting or this
function will disable automatically after 20 minutes if unattended.
Right Side Panel Features
2 Quick release CPC couplers are used to secure mattress air hoses to power
unit.
Left Side Panel Features
Power unit’s main power switch. The power unit automatically goes into
STANDBY-CPR mode after the switch is turned on.
Insert power cord firmly into receptacle.
Rear Panel Features
Hanging hooks are designed to hang the power unit on almost any foot
board. Hooks are spring loaded and will fold against the power unit when
not in use.
Fuse holder can be opened for maintenance.
Air Filter and Filter Cap: XXL-Rehab recommends that the filter to be cleaned
or replaced once a month to ensure optimal performance of the power unit.
Caregivers should always perform a hand check by placing their hands
underneath patient’s pelvis area to check if there is sufficient air support to
ensure the patient is not bottoming out.
XXL-Rehab Low Air Loss mattress
DO NOT disassemble the power unit if you are not a qualified technician.
Please contact your dealer for all service.
7
USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL ·
Instructions for using XXL-Rehab Low Air Loss mattress
BEFORE USE
DURING USE
Mattress and Power Unit Installations
1. Remove existing mattress from bed frame.
2. Place XXL-Rehab Low Air Loss mattress replacement on the bed frame
with the logo at foot end. Secure the mattress at each side using anchor
strap. Please verify all bed functions are working properly without interference before proceeding to the next step.
3. Secure the power unit onto the foot board using hanging hooks.
4. Firmly connect the mattress’ air hose couplings to the CPC quick release
couplers on the power unit’s air outlet.
5. Plug the power unit into a wall outlet and turn on the main power rocker
switch on the left side panel. The STANDBY-CPR orange LED on control
panel will light to acknowledge the power unit is in standby mode.
6. Press the POWER button on the control panel to inflate the mattress
system. MAX INFLATE may be used to shorten the initial inflation time if
needed.
Patient Transfer and Transport
Transfer: It is recommended to have the mattress system in MAX INFLATE
mode during transfer. Make sure the bed is secured before proceeding.
Transport: If a transport occasion comes up, then let the tubes stay in the
pump and disconnect the power supply.
XXL-Rehab Low Air Loss mattress system has a safety foam base underneath
the air cells to support the patient for a short period of time.
To resume normal operation, go through the steps of user customisation.
Emergency CPR Deflation
In the case of emergency, press the POWER button on the control panel to
turn the system into STANDBY-CPR mode to deflate the mattress rapidly in
order to perform CPR. Press the POWER button again and reset the patient
weight to resume therapy
Inspect regularly the mattress to make sure it is not wet or damaged in
other ways.
USER CUSTOMISATION
1. Center the patient on the mattress. Adjust the mattress’ internal pressure according to the patient weight by using the or button on the
control panel of the power unit. If the patient feels the mattress is too
soft or firm, increase or decrease the pressure one increment at a time
and wait for the system to stabilize before making another change until a
comfortable state is achieved.
2. Caregivers should always perform a hand check by placing their hands
underneath patient’s pelvis area to check if there is sufficient air support
to ensure the patient is not bottoming out.
IMPORTANT: tucking the sheet in tightly reduces the effectiveness of
the system.
CHECK-LIST FOR TROUBLESHOOTING
PROBLEM
INSPECTION PROCEDURE
POSSIBLE SOLUTIONS
Power unit does not function.
Check if power cord is firmly plugged into wall outlet.
Secure power cord into wall outlet.
Check if power switch is in ON position.
Turn power switch to ON position.
Check if power surge shuts down the power unit.
A power surge may overload the circuitry temporarily. Turn the
unit off and check the fuse for damage. Turn the unit ON again
with normal procedure.
Power unit does not respond to possible solutions.
Please contact your dealer for assistance.
Inspect air filter for dust.
Clean or replace air filter.
Verify if patient weight setting is correct.
Increase or decrease weight setting until appropriate pressure is
reached.
Check if air is flowing smoothly through air hoses and mattress
manifolds*.
Look for cuts and cracks.
Air hoses or manifolds may need to be readjusted to prevent
obstruction. Air hoses or manifolds may need to be replaced if
cuts or cracks are observed.
Check if mattress’ air hose couplings are properly connected to
power unit’s CPC quick release couplers.
Secure air hose couplings firmly into place.
Check if control panel’s ON green LED is on.
Press POWER button on and ON green LED will light.
XXL-Rehab Low Air Loss mattress
Bottoming out is occurring.
8
Power unit is working, but mattress
replacement is not inflating.
* Mattress Manifold - a side panel within the mattress which serves as air chamber to direct air flow into each individual air cell.
· USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL · USER MANUAL
Of mattress and pump
Air mattress and power unit must be cleaned thoroughly between patients
to avoid cross contamination. The following is a suggested guideline, but
local infection control policies should be followed as well.
WARNING! Always unplug the power unit before cleaning.
Regular cleaning can be performed at bedside with disinfectant and water
followed by drying with a clean dry cloth.
SPECIFIKATIONER
Maximum weight capacity is
450 kg
Length
200 cm
Width
90 / 100 / 200 cm
Height
26 cm
Weight
12 / 16 kg
Weight of pump
4.6 kg
Pump measurement
36 x 30 x 17 cm
Mattress divided into sections
Yes
Use only mild detergents and water to clean the coverlet and the mattress.
Any appropriate NON-PHENOLIC cleaning agent may be used for heavy soiling with urine, blood or other body fluids.
Air loss alarm
Yes
Power failure alarm
Yes
Battery back up
No
Routine cleaning of power unit can be done by wiping down with damp
cloth using disinfectant and water or mild neutral detergent. Never spray
liquids directly on the unit itself.
CPR control
Yes
Static function
Yes
Place directly on bed base
Yes
Only disinfected and dry systems are to be stored. Disconnect the air
hoses from the power unit. Roll up the mattress starting from head end and
working down toward the foot end. Use the straps to fix it.
Fire resistant cover
Yes
Allergenic cover
Yes
Of cover
Machinewash using warm water at maximum 90°C.
Do not use electric or tumble dryers.
Do not iron.
Please ensure air mattress and coverlet are completely dry before letting the
patient laying on the surface again.
Of air filter
Remove air filter from the real panel of the power unit by opening up the
filter cap to clean or replace the air filter.
Inspect the filter for dirt or dust and clean it with mild soap and water. Reinsert the dried air filter after cleaning and ensure the cap is secure.
Washable and disinfectable cover
Yes
Pressure ulcer prevention grade
0-4
Cover
100% nylon with PU on the back
Cells
100% nylon with TPU lamination
Pump
AC100 - 120V/ 50 - 60 Hz 5A
AC220 - 240V/ 50 Hz 5A
Power consumption
Normal operation: Max 350W
Power cord
WARRANTY
The XXL-Rehab Low Air Loss mattress is covered by one year warranty from
the invoice date for materials and workmanship.
Products which have been used for purposes other than the intended uses;
or which have been subject to negligence, abuse, improper storage or handling, improper operation, unauthorized modifications or damages beyond
normal wear and tear as determined by XXL-Rehab, are not covered by this
warranty.
If weight capacity is exceeded, the warranty will be void.
Any unauthorized repairs to product / part, as well as tampering with any
components, will void the warranty.
H05VV-F3x1.00mm2
Fuse rating
T1A/ 250V
Operating & transportation
& storage conditions
Electrical classification
Replace with genuine air filter once a year is recommended and compressor every two year is strongly suggested to prolong system lifetime.
Returns for Service
This device is not self-serviceable. Service and repair must be performed
by an authorized technician or representative. All returned device must be
cleaned and disinfected prior to shipping. Unsanitary or soiled systems will
be returned without servicing.
ABS flame retardant UL
Power input
Agency approval
Temperature: 5°C~45°C
Humidity: 15%~60%
Atmospheric pressure: 700-1060 hPA
Class II Type BF, double insulated with or
without functional ground wire.
IPX0, do not immerse power unit in any liquid
or spray any liquids directly on the power unit.
EN60601-1, EN60601-1-2,
EN61000-3-2, EN61000-3-3, UL2601-1
Electromagnetic compatibility
EN60601-1-2:1993
Type BF Appendix
ISO – classification
MANUFACTURER
Carilex
Turmbergstr. 9A
D-76227 Karlsruhe
Tyskland
ISO 18 12 18 -28
DEALER
XXL-Rehab
Fuglebækvej 1D
DK-2770 Kastrup
Denmark
Tel.: +45 70 25 25 22
Fax: +45 70 25 25 44
www.xxl-rehab.com
11/2014/JP
XXL-Rehab Low Air Loss mattress
MAINTENANCE AND CLEANING
9
Strongly engineered and functionally designed exclusively for bariatric users!
10
XXL-Rehab Minimaxx
XXL-Rehab Minimaxx
Push Motor
XXL-Rehab Manual
Wheelchair - The Eclipse
XXL-Rehab Manual
Wheelchair - The Eclipse Tilt
XXL-Rehab
HD Electric Wheelchair
XXL-Rehab Wheelchair Cushions
XXL-Rehab Shower Bench
XXL-Rehab
Bedside commode
XXL-Rehab
Shower Commode
XXL-Rehab
Shower Commode Wheeler
XXL-Rehab
Shower Commode Rise N Recline
XXL-Rehab
Shower Commode Tilt
XXL-Rehab Barton Transferstol
XXL-Rehab Patient/Dinner chair
XXL-Rehab Rollator King
XXL-Rehab Rollator Support
XXL-Rehab Stand Tall
XXL-Rehab Canes
XXL-Rehab Multifunktional Scale
XXL-Rehab Lift and Slings
XXL-Rehab 400 bariatric hospital bed
XXL-Rehab Home Care Bed
XXL-Rehab Floor raiser
XXL-Rehab
Comfort Foam Mattress
XXL-Rehab
Low Air Loss mattress
XXL-Rehab
Alternating mattress
XXL-Rehab Bariatric walking frame
XXL-Rehab Bariatric suit
11
mauryathome.dk
11-2014 XXL-Rehab Low Air Loss madras / XXL-Rehab Low Air Loss mattress
BRUGSANVISNING · USER MANUAL
XXL-Rehab · Fuglebækvej 1D · DK-2770 Kastrup · Denmark
Telephone: +45 7025 2522 · Fax: +45 7025 2544
[email protected] · www.xxl-rehab.com