BS1015C
USER’S MANUAL
MA13-02/B
12/09/2011
www.davicia.com
USER’S MANUAL
MA13-02/B
INDEX
Page
3
1. DESCRIPTION
2. INSTALLATION
2.1. Scale assembly
2.2. Location and setting
2.3. Startup
4
4
6
6
3. OPERATION
3.1. LCD symbol description
3.2. Key functions
3.3. Weighing
3.4. Taring
3.5. Hold
7
7
8
8
8
9
4. SETUP
4.1. Connecting with PC
10
10
5. MAINTENANCE AND CLEANING
5.1. Instructions for charge and connect
5.2. Instructions for replacing batteries
5.3. Environmental
5.4. Equipment disposal
12
12
12
12
12
6. TROUBLESHOOTING
13
7. HEIGHT ROD ACCESSORIES
7.1. Height rod assembly
7.2. Battery installation
7.3. Key functions
7.4. Calibrating the height rod
7.5. Using the height rod
7.6. Troubleshooting
7.7. Specifications
14
14
15
15
16
16
18
18
8. TECHNICAL SPECIFICATIONS
19
9. WARRANTY CERTIFICATE
20
10. CE DECLARATION OF CONFORMITY
21
2
USER’S MANUAL
MA13-02/B
1. DESCRIPTION
1
2
3
4
5
(1)
(2)
(3)
(4)
(5)
Display unit
Display mounting column
Tray
Scale platform
Adjustable feet
3
USER’S MANUAL
MA13-02/B
2. INSTALLATION
Before putting the device into use, please read with care the information given in the Operating Instructions.
They contain important instructions for installation, proper use and maintenance of the device.
The manufacturer shall not be liable for damages arising out of failure to heed the following instructions:
o
o
o
o
o
o
When using electrical components under increased safety requirements, always comply with the appropriate
regulations.
Improper installation will render the warranty null and void.
Ensure the voltage marked on the power supply unit, matches your main power supply.
This device is designed for use indoors.
Observe the permissible ambient temperatures for use.
The device meets the requirements for electromagnetic compatibility. Do not exceed the maximum values
specified in the applicable standards.
2.1.
2.1. SCALE ASSEMBLY
Indicator can be installed in FIXED and ADJUSTABLE mode
Route the cable through the tube
FIXED
MODE
Screw the hole below
ADJUSTABLE
MODE
Screw the hole above
4
USER’S MANUAL
MA13-02/B
1. Route the cable through the
column.
2. Route the cable through the base
and screw column to base.
3. Tighten the three base screws by
turning clockwise.
4. Connecting the cable
5. Align the bracket on the base and
tighten the screws.
6. Installing the adjustable foot
NOTE: Make sure
that the weighing
tray is installed
properly and at the
right position.
7. Install the tray to the proper location
5
USER’S MANUAL
MA13-02/B
8. Install the tray at the proper location and tighten the screws on both sides.
9. Assembly finished.
10. Adjust four rubber feet to center bubble level.
2.2.
2.2. LOCATION AND SETTING
We strongly recommend you use this device on flat and hard surface. Any soft surface, like carpet will cause
inaccuracy.
2.3.
2.3. STARTUP
o
o
o
Connect the equipment to the power supply using the AC adaptor.
Press the ON/OFF button in the indicator.
Wait until “0” appears on display.
6
USER’S MANUAL
MA13-02/B
3. OPERATION
Before reading detailed instructions on how to use all the weighing functions that are built into your scale, please
read the following important guidelines:
o
o
Always be sure that the display shows `Zero` before use, if it does not then please press the ZERO key.
The equipment is designed to detect when a stable weight is achieved, your reading should be taken at this
point.
3.1.
3.1. LCD SYMBOL DESCRIPTION
DESCRIPTION
PRE-TARE Indication
HOLD Indication
NET WEIGHT Indication
WEIGHT Indication
BMI Indication
Stable symbol
kg
Minor weight
Zero symbol
Low Battery
Display
[→ 0 ←]
HOLD
NET
WEIGHT
Description
Zero symbol: If the scale does not show exactly zero value although the scale pan is
unloaded, press the [→0
0←]] key. After a short waiting time, the scale will be zeroed again.
Stable symbol: If the display shows the stabilization display [ο], the scale is in the stable
condition. When the scale is in the unstable condition, the stabilization display [ο]
disappears.
Hold function / saving function is active.
The net weight is displayed.
The present weight value is displayed.
7
USER’S MANUAL
MA13-02/B
3.2.
3.2. KEY FUNCTIONS
Designation
ON/OFF-switch
HOLD button
ZERO setting key
TARE button
Function
Turn ON/OFF the scale
To hold weighing result on the display / determine stable weighing value
To reset the display to 0.0kg display / Zero the scale (±2% of full capacity)
Tare weight during weighing process
3.3.
3.3. WEIGHING
Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is performed and then the software
version is displayed. The scale is ready for weighing when the “0,00
0,00 kg”
kg weight display is shown.
NOTE: If “0,00 kg” won’t display on the screen, press [→0←] key to zero the scale, [→0←] can be used any time
to zero the scale.
Place a person in the middle of the scale. Wait until the stability display (ο) is shown and then read the weighing
result.
DIRECTION: If a person is heavier than the weighing range, the display will show the “Err” symbol (= overload).
3.4.
3.4. TARING
Tare allows the user to zero the instrument to cancel the weight of a container/ clothes from the reading of the
instrument, thus giving the true weight of the product/ person being tested.
o
o
o
o
o
o
Turn on the scale using [ON/OFF] key.
When“0,00
0,00 kg”
kg displays on the screen and scale beeps twice, place the object (clothes) to be tared on the tray.
Press [TARE] key after the weight stabilizes and stable sign appears on the display.
Display will return to “0,00
0,00 kg”
kg
Place the item/ baby to be weighed (without removing the tare object). The weighing result is the Net weight
of the item / baby.
To delete the saved tare value, remove the tare object from the tread platform and press [TARE] key.
8
USER’S MANUAL
MA13-02/B
3.5.
3.5. HOLD
The scale is provided with the integrated hold function (determination of average value). It enables people to be
weighed accurately although they are not still on the scale plate.
NOTE: Determination of average value is not possible when a person moves too much.
o
o
o
o
o
o
Switch on the scale with the [ON/OFF] key. The scale self-check diagnostic is performed. The scale is ready
for weighing when the “0,00
0,00 kg”
kg weight display is shown.
Place a person in the centre of the scale plate.
Press the [HOLD] key. When the triangle is flashing on the display, the scale takes some measuring values and
then the calculated average value is displayed.
Remove the object/ baby from the scale platform. The object/ baby’s weight reading will remain on the
display.
Press the [HOLD] key again to return the scale to the normal weighing mode.
HOLD function can be activated before or after putting the weight on the platform. In case of weighing babies
that move a lot, it is recommended to press [HOLD] key before the person moves on the platform.
9
USER’S MANUAL
MA13-02/B
4. SETUP
Switch ON the scale and long press [ZERO] key for 3 seconds. First “SETUP” and then “A.OFF” will display
successively.
Auto-off time setup
This enables operator to select the auto turn OFF time of the device.
Auto off time: 120 sec/180 sec/240 sec/300 sec/Off
Buzzer on/off setup
This enables operator to select beep sound ON/OFF.
Buzzer:
Buzzer: On/Off
NOTE: To confirm the settings, please press HOLD when “END” displays on the display.
4.1.
4.1. CONNECTING WITH PC
The equipment can be connected to PC to transfer the weighing results.
1.
Start Hyper Terminal
Start Hyper Terminal program from clicking: Start Menu → Programs → Accessories → Communication →
Hyper Terminal.
2.
New Connection Description
Give new connection a name then click OK.
3.
Select Your COM Port
Click Connect to select your COM port. Usually there’s only one option for select. Then click OK.
4.
Port Settings
Click Bits per second to set up rate at 9600, Data bits at 8, Parity at None, Stop bits at 1 and Flow control at
Hardware. Then click OK to complete your setting.
5.
Output Data
When the patient has been weighed and BMI calculated, simply press the PRINT key to output data from scale
to PC or an Optional Printer.
Parameters of RS232 interface
Set parameters of the scale interface on the connected device. It is not possible to change the scale parameters.
o
Baud rate: 9600 bps
10
USER’S MANUAL
MA13-02/B
o
o
o
o
o
Parity check: None
Data length: 8 bits
Stop bit: 1 bit
Handshake: RTS/CTS
Data code: ASCII.
11
USER’S MANUAL
MA13-02/B
5. MAINTENANCE AND CLEANING
The scale does not require any routine maintenance. However, we recommend checking the scale’s accuracy at
regular intervals. The regularity of these checks is dependent on the level of use and the state of the scale.
o
o
o
We would recommend using alcohol based wipes or similar when cleaning the scale.
Please do not use large amounts of water when cleaning the scale as this will cause damage to the scale’s
electronics, you should also refrain from using corrosive liquids or high pressure washers.
Always disconnect the scale from the main power supply before cleaning.
5.1.
5.1. INSTRUCTIONS FOR CHARGE AND CONNECT
If “Lo” is displayed on the LCD, you should charge scale with AC adaptor.
Locate adaptor plug at the rear of indicator.
CAUTION: Always connect the AC adaptor to the indicator before connecting to the main power supply. Please
disconnect the adaptor from main power supply before indicator.
5.2.
5.2. INSTRUCTIONS
INSTRUCTIONS FOR REPLACING BATTERIES
o
o
o
o
o
o
Remove the battery cover
Take the battery housing out
Replace new batteries
Make sure batteries are well installed
Reinstall the battery housing
Reinstall the battery cover
5.3.
5.3. ENVIRONMENTAL
All batteries contain toxic compounds; disposal of batteries should be delegated to a competent organization,
complying with the deposit of Poisonous Waste Regulation 1972.
Please do not incinerate batteries.
The optimum operating temperature for the scale is 5ºC to +35ºC; although it will operate at higher and lower
temperatures the scale’s battery life will adversely be affected.
5.4
5.4. EQUIPMENT DISPOSAL
All electrical and electronic products should be disposed of separately from the
municipal waste stream via designated collection facilities appointed by the
government or the local authorities.
This crossed-out wheeled bin symbol on the product means the product is covered
by the European Directive 2002/96/EC.
The correct disposal and separate collection of the equipment will help prevent
potential negative consequences for the environment and human health. It is a
precondition for reuse and recycling of used electrical and electronic equipment.
12
USER’S MANUAL
MA13-02/B
6. TROUBLESHOOTING
Error Message
Reason
Low Battery: This warning shows
that the voltage of battery is too
low to use
Overload: The total load exceeds
the maximum capacity of the scale
Action
Please replace a new battery or plug the AC
adaptor for operation
Counting error (too high): Indicates
that the signal from the load cell is
too high
Counting error (too low): Indicates
that the signal from the load cell is
too low
Zero count over calibration zero
range +10% while power on
This error is normally caused by a serious fault on
the scales such as faulty load cell or wiring. Please
contact the local service representative.
This error is normally caused by a serious fault on
the scales such as faulty load cell or wiring. Please
contact the local service representative.
Please recalibrate the instrument
Zero count under calibration zero
range -10% while power on
Please recalibrate the instrument
EEPROM Error: Indicates that
there is a fault with the scale’s
software
This error is normally caused by a serious fault on
the scales such as faulty load cell or wiring. Please
contact the local service representative.
13
Please reduce the loading and try again
USER’S MANUAL
MA13-02/B
7. HEIGHT ROD ACCESSORIES
AC2080D
AC2080D is a digital height rod to be mounted on the scale
with measuring range of 35~80cm. AC2080D have LCD
display for convenient, accurate and quick height
measurement.
AC2080M
AC2080M is a mechanical height rod to be mounted on the
scale with measuring range of 35~80cm, AC2080M don’t
have LCD display.
7.1.
7.1. HEIGHT ROD ASSEMBLY
1.
Fix bracket on the tray with two screws
2.
14
Assemble the height rod to the bracket
USER’S MANUAL
MA13-02/B
3.
Press the buckle to secure the height rod is well
installed
4.
Assembly complete
7.2.
7.2. BATTERY INSTALLATION
AC2080M don’t need batteries to operate
AC2080D
Installing battery AAA 1.5V x 2 in battery housing next to LCD display.
WARNING: If the height rod is not in use for long time, take out the batteries and store them separately, in case of
any leakage from battery would damage the measuring rod.
7.3.
7.3. KEY FUNCTIONS
Keys
ON/OFF
UNIT
HOLD
RESET
Description
Press to switch on and off.
To select the measurement unit between cm and inch.
To enable value store function during measuring. Current height value of user will be stored
and displayed.
Press again to disable the function.
RESET key is for user to calibrate false reading. Slide right feet piece to the left at length of
35cm. Press RESET key to calibrate the height rod. LCD will display 35.0cm which means
the height rod has been well calibrated.
15
USER’S MANUAL
MA13-02/B
7.4.
7.4. CALIBRATING THE HEIGHT ROD
AC2080D
1.
Installing battery
2.
Press ON/OFF to switch on LCD
3.
Sliding feet piece left to the end.
4.
Press RESET button to calibrate at 35.0cm
7.5.
7.5. USING THE HEIGHT ROD
AC1200D
1.
2.
3.
Extend the height rod for use
Place the baby with his head touching the left top of the height rod
Straighten baby’s feet
16
USER’S MANUAL
MA13-02/B
4.
5.
6.
7.
Slide the right piece of the height rod until it touches the baby’s feet
The LCD is displaying the baby’s lenght
Press HOLD button to memorize the length reading.
Press HOLD again to disable the function and return to normal mode.
AC1200M
1.
2.
3.
Extend the height rod for use
Place the baby with his head touching the left top of the height rod
Straighten baby’s feet
4.
5.
Slide the right piece of the height rod until it touches the baby’s feet
Take reading from Picture A
17
USER’S MANUAL
MA13-02/B
6.
7.
Pull out to use
Remove baby
7.6.
7.6. TROUBLESHOOTING
If feet piece is loose please re-assemble feet piece as below.
NOTE: If the reading has error, please follow RESET procedure for calibration.
7.7
7.7. TECHNICAL SPECIFICATIONS
Model
AC2080D
AC2080M
Type
Keys
Scale mounted height rod
Hold, Reset, Unit
Measurement range
35~80cm
Graduations
1mm
Dimensions
(W x H x L) 29 x 7 x 62 cm
Weight
Power Supply
700 g
2 * AAA-size batteries
Temperature range
+10ºC up to +40ºC
18
USER’S MANUAL
MA13-02/B
8. TECHNICAL SPECIFICATIONS
Model:
Capacity:
Display:
Tray dimensions:
Power supply:
Key Functions:
Operation temperature:
BS1015C
15 kg x 2g <6 kg > 5g
LCD display with 5 and 1/2 digits. Digit size: 24mm
550 x 290 mm
AC adapter 15V 300mA /Rechargeable Battery
ON/OFF,ZERO, PRINT, BMI, HOLD, PRE-TARE,TARE, CLEAR, ENTER, 0~9, M1-5
5ºC / 35ºC
19
USER’S MANUAL
MA13-02/B
9. WARRANTY CERTIFICATE
We provide a 1-year guarantee from the date of purchase. The guarantee covers manufacturing defects or
defects occurring as a result of normal use and operation.
It excludes all removable parts such as batteries, wires, power supply devices, etc. Faults falling within the
guarantee period shall be rectified free of charge upon presentation of the purchase receipt for the appliance. No
other rights can be taken into consideration.
Return shipment is at the customer's expense if the appliance is anywhere other than at the customer's
headquarters.
In the event of damage during shipment, guarantee rights will be granted only if the appliance is shipped in the
original packaging. You are therefore recommended to keep the original packaging.
Following expiry of the guarantee period, all services requested will be dealt with by our specialist technicians at
the user's expense.
20
USER’S MANUAL
MA13-02/B
10. CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
FABRICANTE PRODUCTO:
PRODUCT MANUFACTURER:
DAVI & CIA S.L.
DIRECCIÓN:
ADDRESS
C/ Murcia, 35, Nave G
Polígono Industrial Can Calderón
E08830 Sant Boi de Llobregat
BARCELONA - ESPAÑA
DECLARAN BAJO SU RESPONSABILIDAD QUE EL PRODUCTO:
DECLARE UNDER THEIR RESPONSIBILITY THAT THE PRODUCT
Marca
Brand
DAVI & CIA
Nombre
Name
BS1015C
Tipo
Type
EQUIPO ELECTRONICO
Electronic equipment
Finalidad Prevista
Intended Use
Equipo pesapersonas.
Weighing equipment.
CUMPLE LOS REQUISITOS DE LAS DIRECTIVAS:
CONFORMS WITH THE REQUISITES OF THE DIRECTIVES
EC Directive 2004/108/EC
Directiva de Compatibilidad Electromagnética (Electromagnetic
Compatibility Directive)
EC Directive 2006/95/EC
Directiva de Baja tensión (Low Voltaje Directive).
EC Directive 93/42/EEC
Aplicable únicamente a las básculas con aprobación médica
(Applies only to scales with medical approval).
FECHA: 12 de Diciembre de 2010
Nombre:
Name
Firmado:
th
DATE: 12 December 2010
J.Masana
Signed
Cargo:
Function
Gerente
DAVI & CIA
21
USER’S MANUAL
MA13-02/B
This manual and other support documents are available on the web
www.davicia.com
C/ Murcia, 35 Nave G Pol.industrial Can Calderón.
E 08830 Sant Boi de Llobregat. Barcelona. Spain
Tel. +34 93 654 32 32
Fax. +34 93 654 16 62
22