Download Untitled

Transcript
PersonalScopeTM
Velleman
Legen Heirweg 33
9890 Gavere
Belgium
Internet Site: http://www.velleman.be
E-mail: [email protected]
HHPS5 - 1999 - ED1
2
PersonalScopeTM
CONTENTS
CONTENTS
FCC information for the USA ............................... 4
Important............................................................... 5
GENERAL ............................................................. 6
Features ................................................................. 5
Options.................................................................. 5
Technical specifications......................................... 6
SAFETY and WARNINGS.................................... 7
Symbols displayed on the unit................................ 7
USE......................................................................... 8
Survey of the connections and controls .................. 8
Power supply ......................................................... 9
Survey of the indications on the screen .................10
OPERATION........................................................11
Turning on/off the Personal Scope ........................11
Resetting the Personal Scope.................................11
Adjusting the contrast...........................................11
Choice of input .....................................................11
Switching from one readout to another..................12
Measuring probe adjustment .................................13
Auto-setup function ..............................................13
Signal screen ........................................................14
Screen memory .....................................................15
Set-up function keys..............................................15
The markers .........................................................15
Setting up the voltage per division .......................16
Setting up the time base ........................................15
Moving the signal in Y-position............................16
Trigger functions ..................................................17
Setting up a X10 measuring probe ........................18
MAINTENANCE..................................................18
TROUBLESHOOTING .......................................19
The screen remains blank or there is no signal......19
RMS readout is incorrect ......................................19
No signal on the oscilloscope display ....................19
Incorrect frequency readout...................................19
Voltage readout is not correct ...............................19
WARRANTY ........................................................19
DIAGRAM............................................................20
3
PersonalScopeTM
FCC information for the USA
This equipment has been tested and found to comply with
the limits for a Class B digital device, pursuant to Part 15
of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to radio communications. However, there
is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be
determined by turning the equipment off and on, the user
is encouraged to try to correct the interference by one or
more of the following measures:
•
•
•
•
4
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Important
This equipment was tested for FCC compliance under conditions that include the use of shielded test leads between
it and the peripherals. It is important that you use shielded
cables and connectors to reduce the possibility of causing
Radio and Television interference. Shielded probes, suitable for the HPS5 oscilloscope can be obtained from
the authorized Velleman dealer.
If the user modifies the HPS5 oscilloscope or its connections in any way, and these modifications are not approved
by Velleman, the FCC may withdraw the user’s right to
operate the equipment.
The following booklet prepared by the Federal Communications Commission may be of help: “How to identify and
Resolve Radio-TV Interference problems”. This booklet is
available from the US Government Printing Office, Washington, DC20402 Stock No. 044-000-00345-4.
PersonalScopeTM
GENERAL
The Velleman PersonalScope is not a graphical multimeter
but a complete portable oscilloscope at the size and the cost
of a good multimeter. Its high sensitivity - down to 5mV/
div - and extended scope functions make this unit ideal for
hobby, service, automotive and development purposes.
Because of its extreme value for money, the PersonalScope
is well-suited for educational use in schools and colleges.
Suitable for measurements on audio equipment, mains
voltage applications, digital signals, all kind of sensors,
signal analysis in automotive applications, car stereo etc…
It’s ultra fast full auto setup function takes care of all the
necessary settings and makes measuring waveforms very
easy, a feature that will be appreciated both by experts and
novices.
GENERAL
Features
• High contrast LCD with wide viewing angle
• Full auto set up for volt/div and time/div
• Trigger mode: run - normal - once - slope -/+
• True RMS and dB measurements
• Readout: Vdc-Vpp-Vrms-dBm
• x1 and x10 probe calculation option
• Display mode: none - markers - grid - crosshair
• Marker readout: dt - 1/dt (frequency) - dV
• AC/DC input coupling selection
• Automatic zero reference for DC measurements
• Sceen hold fuction
• Auto power off or continuous on
• Low battery detection
• Includes battery charge circuit
Options
• Insulated measurement probe x1 / x10: PROBE60S
• Soft carry case: CHPS5
• Adapter: type PS905 for 230V / PS905USA for 110V
5
GENERAL
PersonalScopeTM
Technical specifications
Maximum sample rate
Input amplifier bandwidth
Input impedance
Maximum input voltage
Input coupling
Vertical resolution
Linearity
LCD Graphics
dBm measurement (0 dB= 0.775V)
True-rms AC measurement
Peak peak and DC range
Timebase range
Input sensitivity range (at x1)
Probe calibration output
Supply voltage
Batteries (option)
Charge current for rechargeable batteries
Battery life
Operating temperature
Dimensions
Weight
6
5MHz for repetitive signals (0.5MHz for single shot events)
1MHz ( -3dB at 1V/div setting)
1Mohm // 20pF (standard oscilloscope probe)
100Vpeak (AC + DC)
DC, AC and GND (GND for auto zero reference)
8 bit (6 bit on LCD)
± 1bit
64 x 128 pixels, 64 x 96 pixels for signal
From -73dB tot +40dB (up to 60dB with X10 probe) ±0.5dB accuracy
From 0.1mV to 80V (up to 400Vrms with X10 probe) 2.5% accuracy
0.1mV to 180V (1mV to 600V met x10 probe) 2% accuracy
20s, 10s, 5s, 2s, 1s, 0.5s, 0.2s, 0.1s, 50ms, 20ms, 10ms, 5ms, 2ms, 1ms, 0.5ms,
0.2ms, 0.1ms, 50us, 20us, 10us, 4us, 2us / div.
5mV, 10mV, 20mV, 50mV, 0.1V, 0.2V, 0.4V, 1V, 2V, 4V, 8V, 20V /div.
Approx. 1KHz / 5Vtt
9VDC/ min 300mA adapter (unregulated) 12VDC if regulated
Alkaline type AA or Ni Cd / NiMH rechargeable ( 5 pcs required)
90mA
Up to 20h with Alkaline batteries
0 to 50°C (32 to 122°F)
105x220x35mm (4.13x7.95x1.38”)
395g (14oz.) ex. Batteries
PersonalScopeTM
SAFETY and WARNINGS
SAFETY and WARNINGS
Symbols displayed on the unit
!
Important safety information, see user manual.
Ground level
⇒ The PersonalScope is ideally suited for measurements of category II installations with pollution degree 1 and using a
maximum of 600V, in accordance with the IEC1010-1 norm. The enclosure is UL 94V0 listed
⇒ Consequently, all measurements should be avoided in case of polluted or very humid air. One should also refrain from
measuring conductors or installations that use voltages that exceed 600Vrms above ground level. CAT II indicates
conformity for measurements of domestic installations.
⇒ The maximum input voltage for the connections of the unit stands at 100Vp (AC+DC)
⇒ Do NOT open the enclosure while performing measurements.
⇒ Remove all test leads before opening the enclosure in order to avoid electrical shock.
⇒ Use a measuring probe with an insulated connector ( e.g. type PROBE60S) when measuring voltages exceeding 30V.
!
Remove all non-rechargeable batteries when connecting an AC/DC adapter !
7
PersonalScopeTM
USE
USE
Survey of the connections and controls
1 - BNC input connector (max. input 100Vp AC+DC)
2 - Adapter connection (observe the polarity!)
3 - Battery compartment
4 - X10 probe testing signal behind battery cover
5 - Serial number
6 - Display
7 - Battery charge indication LED
8 - Reset push button
9 - Keyboard
1
HP
HPS5
S5
6
mmaar
rkk 11--22
8
I/O
t-V/divv
t-V/di
Res
et
Re set
Trig
Trigger
ger
mode
mo
de
Slope
Slo
pe
9V DC
Hold
Ho
ld
DDisp
layy
ispla
0
x1
5
probe
test
8
9
AC/D
/DC
AC
C
Auto
Auto
4
123456
3
Y-pos
Y-pos
probe
probe
x1/x10
x1/x1
2
!
trig
tri
g
PERSON
AL
P ESRCSO
O PN EA L
SCOP
E
Cha
rge
Charge
7
PersonalScopeTM
Power supply
The HPS5 PersonalScope can be powered by means of an
adapter, ordinary batteries or rechargeable ones.
Rechargeable batteries are advised for intensive use.
The greater the capacity of the rechargeable batteries, the
greater the autonomy of the unit.
Rechargeable batteries of the 500mA/h type will yield approx. 5 hours of autonomy, while 800mA/h-types provide
up to 8 hours of autonomy.
⇒
⇒
Attention:
Use a non-regulated adapter of 9VDC that supplies
a min. of 300mA. Observe the polarity and consult
your retailer if necessary.
When using a regulated adapter, the user should
verify whether it can supply 12VDC.
USE
Batteries (option):
⇒ The unit can be used with either ordinary alkaline batteries or rechargeable ones e.g. our HR6 type (5 x).
⇒ Open the battery cover by loosening the screw:
Insert the batteries in the holders. Mind the polarity !
probet
tes
9
PersonalScopeTM
USE
Important:
Do NOT perform measurements unless the
battery compartment is closed. The battery
cover should only be removed for the calibration of the
probe 's X10-position.
!
F
F
F
When first using the rechargeable batteries or if the
batteries are completely discharged, the user should
charge them for at least 12 hours before using the
unit.
When the unit is OFF, the charging time for
800mA/h type batteries will be approx. 12 hours.
The “Charge” indication LED on the front panel
will light up when the batteries are being charged.
ATTENTION: Remove all non-rechargeable batteries
when connecting the AC/DC adapter !
The message “Lo . bat” will flash in the bottom right corner of your display when the batteries need replacement
(or charging). Insufficient battery voltage may entail erroneous measuring
results.
Survey of the indications on the screen :
Dot indicates the vertical position of
the signal on the screen
Trigger opening indication (fixed
in the middle of the screen)
Signal window with (possibly) the
markers , cross or dot to indicate the
various divisions.
Time per division or time between markers
(when used)
Display of converted/calculated frequency
1/dt between markers (if present)
Trigger information or screen hold indication.
Input-coupling indication : ac/dc/gnd and
X10 probe setup
Measurement readouts, Vp (V peak-to-peak)
; Vr (Volt rms) ; dB (Decibels) ; V= (DC volt.)
Readout of the voltage between the markers (when present)
Selected voltage per division
Indication of the selected function (blue
keys) or battery-low indication
10
PersonalScopeTM
OPERATION
OPERATION
Turning on/off the Personal Scope
Press the “ON/OFF” switch briefly to (de)activate
the unit. The screen will briefly display the logo above the
message “timer on”, which means that the unit will deactivate after eight minutes if no other key has been pressed
during this interval.
In order to bypass automatic deactivation : Keep the
ON/OFF switch pressed down while activating the unit.
The screen will briefly display the message “timer off”.
F When the unit is deactivated, the last setup will be
saved as long as the power is not cut and as long as
the unit is not “reset” (see below).
Resetting the Personal Scope.
In case of unusual behaviour of the keypad or the display,
it is possible to reactivate the unit by using the "Reset"
function. The unit will subsequently return to the
manufacturer-programmed setup.
Press the sunk “Reset” push button for at
least 10 seconds.
F
Do not use a sharp tool !
Adjusting the contrast
Press the “contrast” key until the desired contrast
is displayed.
The user can either keep the key pressed down or press it
repeatedly.
When pressing this key, the contrast will always increase to
maximum level before starting again at the lowest available
level and working its way up again
AC/DC
Choice of input
Depending on (part of) the signal to be measured,
the input can be connected to the signal through a direct link
or by using a decoupling capacitor.
Select DC for measuring DC voltage.
Press the “AC/DC” key to select either AC or DC input
coupling (see indications on the screen).
F
When measuring the ”ripple” of a DC voltage : put
the input on AC to limit the measurement to the AC
component of the signal.
Reset
11
PersonalScopeTM
OPERATION
Switching from one readout to another
The Personal Scope offers many voltage measuring possibilities.
Press the “readout” key repeatedly to select one of four
possible readouts :
Peak voltage (Vp). The signal's peak-to-peak voltage
(difference between
highest and lowest
value) is displayed.
True RMS readout (Vr). The area of the AC wave is
calculated and converted to voltage. Please remember that
a correct readout invariably consists of at least one or two
visible periods (one can also use the auto-setup function).
dB measurement (dB). The measured signal is converted
to dBm (0dB=0.775V). Once more, the user should make
sure that at least one or two periods are displayed or select
the auto-setup function).
12
DC voltage measurement (V=). This function enables the
user to measure DC voltages (only for DC input coupling)
F
Useful tip for measuring DC voltages:
The readout can be set at zero (reference) for any
position on the screen by keeping the AC/DC key
pressed down. Always use the "run" trigger mode
for DC voltage measurement.
NOTE: If the signal goes off-screen or when the dBsignal is too small for
measurement, the
display will show.....
PersonalScopeTM
probe
x1/x10
Measuring probe adjustment
We advise the user to use a measuring probe
equipped with a X10 setting when high or unknown
voltages or high impedances need to be measured.
When using the measuring probe in the X10 position, the
measuring impedance will increase to 10Mohm, thereby
reducing the measuring instrument's
0
charge on the cou1
x
pling.
x1
Press the “x1/x10”
key in order to adx1
just the display to
the setup of the
measuring probe.
Upon
choosing
X10, “X10” will be
indicated on the
screen and all measuring values will be adjusted accordingly.
0
x1
OPERATION
Auto
Auto-setup function
The auto-setup function is ideally suited for quick
measurements as no manual setup has to be made and
everything is automatic.
F
Use the auto-setup function when the screen no
longer displays a signal after the manual setup.
• Connect the signal to be checked to the BNC input by
means of a special measuring lead or measuring probe.
• Activate the unit.
• Press “Auto” in order to (de)activate the auto-setup
function.
When “auto setup” is engaged, the Volt/div and Time/div
readouts are highlighted.
The auto setup function will provide correct measurements
for repetitive and stable signals of max. 400KHz.
The display may read as follows :
IMPORTANT : X10 measuring probes should be calibrated. Read “Setting up a X10 measuring probe".
13
PersonalScopeTM
OPERATION
!
IMPORTANT: Put the appropriate measuring
probe in the X10 position for measuring high
voltages (>100Vp + DC).
NOTE:
⇒
The Y-position does not work in auto-setup, which
is why the trigger setup remains in run or normal
mode.
⇒
The auto-setup function will deactivate when using
the Trigger , Volt and time per division and Y
position functions. (see further).
Signal screen
The signal screen is an aid in displaying the
divisions on your screen or in positioning movable markers
on the screen.
Press the “Display”key repeatedly in order to engage the
various aids:
• A grid divides the screen into reference points.
• A cross inserts a co-ordinate system into the screen.
• Markers: Moveable markers in order to measure the
signal (see below for use).
• Empty : no aids.
GRID
Display
CROSSHAIR
MARKERS
14
PersonalScopeTM
OPERATION
Screen memory
It may be useful to "freeze" certain signals on
your screen. This will allow the user to study the signal or
perform measurements with the markers.
Press ‘
Hold’to freeze the screen.
F The screen is also saved when the unit is deactivated.
F The memory will be erased if “reset” is pressed or if
the battery voltage is too low.
Hold
Set-up function keys
The markers
The user can perform measurements on a certain
signal by using the four moveable markers.
This can be useful
when measuring the
interval between two
points or the amplitude of any given
peak.
mark 1-2
The following indications will appear on the screen :
HPS 5
The time interval between two vertical
markers.
ma rk 1-2
trig
t-V/div
The selected function is displayed in the bottom right
hand corner of the screen.
Y-pos
probe
x1/x10
Rese t
Trigge
Triggr er
mode
mode
Slope
Slope
Hold
Displa y
AC/D C
Auto
Charge
The four blue function keys
work in conjunction with the
arrow keys.
The calculated frequency (primarily used for
the measurement of periods).
The voltage between two horizontal markers.
The markers can be moved through the arrow keys.
Keeping the key pressed down will move the marker
quickly, pressing it briefly will move the marker by 1
position. The “mark 1-2”key is used to select the desired
marker.
PERS
ONAL
SCOP
E
15
PersonalScopeTM
OPERATION
F
Determining the frequency of a signal requires the
measurement of a period. The easiest way to do this
would be by placing the vertical markers either on
two consecutive peaks or two identical slopes of a
signal.
Setting up the voltage per division
The user can enlarge or reduce the signal on the
screen vertically by adjusting the displayed voltage per
division. (V/div = voltage per division). The divisions can
be made visible through the Display key (see page 16).
Press the “t-V/div”key
Press the ⇑ or the ⇓ key to select the sensitivity : from
5mV to a maximum of 20V per division.
F 50mV to 200V with X10 probe selection
t-V/div
t-V/div Setting up the time base
Adjusting the time base will visualise more or
fewer periods of a signal (t/div = time per division). The
divisions can be visualised through the Display key (see
page 16).
Press the “t-V/div”key
Press the ⇐ the ⇒ key to set the time base between 20s
and 2µS per division.
ATTENTION : the positions 10µs, 4µs and 2µs only
operates correctly with periodic signals, for which the unit
will automatically select the ‘normal’ trigger mode. The
16
screen will be adjusted once the trigger level is reached.
NOTE: Use the minimum time base (2 microsec) as a
starting point when measuring a signal and select longer
time bases until the signal is displayed properly. Otherwise the display may not correctly reflect the signal
under
measurement
due
to
aliasing.
Moving the signal in Y-position
Press the “Y-pos”key
Press the ⇑ key or the ⇓ key in order to move the signal
in the direction of the arrows. Prolonged pressing will make
the Y- position change faster. A small dot in the left-hand
corner of the screen will indicate the direction in which the
signal has moved. In this manner, the user will know in
which direction the signal was going when it went offscreen.
Y-pos
PersonalScopeTM
Trigger functions
The Personal Scope is equipped with a number
of "trigger" aids that are used when the signal can be
displayed on-screen.
Press the “trig”key.
The ⇐ (trigger mode) key enables the user to select the
appropriate triggering method, the ⇒ (slope) key is used to
select the trigger slope (see below).
trig
Press the ⇐ key repeatedly to choose between :
“run” mode: The signal will always appear on the
screen, regardless of the level involved. This
position is used most frequently and should always
be used for measuring DC voltages in particular.
“norm” mode: The screen will be adjusted provided that the signal falls within the trigger level
(interruption in the vertical line on the left-hand
side of your screen). Use this mode when you want
to start displaying the signal when it reaches a
preset threshold value.
“once” mode: The screen is adjusted one single
time, viz. when the trigger level is reached. Use this
mode to detect e.g. a short, once-only voltage peak.
ATTENTION : in the positions 10µs, 4µs and 2µs the unit
will automatically select the ‘norm’ trigger mode. The
trigger level has to be reached for the screen to be adjusted.
OPERATION
F
Both the ⇑ and the ⇓ key remain active in order
to move the signal along the Y-axis, which enables
the user to programme a trigger reference
Repeatedly press the ⇒ key to choose between :
Triggering on the rising slope of the signal. The screen
will only display the signal when a positive slope is
"detected", viz. the signal has to rise in vertical direction in order to trigger.
Triggering on the falling slope of the signal : The screen
will only display the signal when a negative slope is
"detected", viz. the signal has to drop in vertical direction in order to trigger.
Trigger on rising slope
Trigger on falling slope
NOTE : There is a time difference of ±12µs between the
trigger point (the interruption in the vertical line on the
left-hand side of the screen) and the start of the actual display.
This is specifically true on higher frequencies.
17
PersonalScopeTM
OPERATION
Setting up a X10 measuring probe
When used in the X10 position, a measuring probe should
always be callibrated to the measuring instrument being
used, in this case the Personal Scope.
• Open the battery cover.
• Set the probe input to X10
• Set the voltage per division to 1V
• Set the time per division to 0.2ms
• Select AC for the input.
Use the probe to perform measurements at the preselected
point under the battery cover
0
x1
e
prob
test
Adjust the trimmer of
the measuring probe in
order to obtain a
square wave signal
with a top that is as flat
as possible.
NG
NG
OK
MAINTENANCE
Clean the display with a shammy. NEVER use a dustcloth
or paper in order to avoid scratches. The rest of the unit can
be cleaned with a soft, dry cloth. NEVER use water to clean
the unit.
18
PersonalScopeTM
TROUBLESHOOTING
The screen remains blank or there is no signal :
• No power supply
• Batteries are discharged
• Contrast adjustment is incorrect
• Press RESET for at least 10 seconds
REMARK: Temporarily remove both batteries and adapter
in case RESET is not yet operational.
RMS readout is incorrect :
• Make sure that at least 1 and preferably even 2 periods
are displayed.
• The batteries are discharged.
No signal on the oscilloscope display :
• Time/div setting is in the wrong position. Try 1ms or
choose auto-setup mode.
• The unit is in the Hold position
• Trigger function is set in the “once” position
• The programmed trigger level is not reached (choose
“run” mode)
• Y position is wrong
• The input signal is too high, change the volt/div. setting
or choose auto-setup.
TROUBLESHOOTING
Incorrect frequency readout:
• An incorrect time/div. setting has been chosen. (start at
2µs)
Voltage readout does not correspond with the
actual value :
• The measuring probe is in the X10 position
• The batteries are discharged
• The zero refrence is not set correctly for DC measurements.
• Take into account that the RMS readout only measures
the AC voltage
WARRANTY
This product carries a one-year warranty as far as the craftsmanship and
possible flaws in the materials are concerned. The warranty expires ONE
YEAR after the date of purchase.
The warranty will only apply if the unit is wrapped in the original packing
material and either presented to VELLEMAN-KIT N.V or to an official
distributor together with a copy of the original purchasing document.
VELLEMAN-KIT N.V. is under the obligation to repair defects and
flaws, but is free to either replace or repair defective parts.
The warranty does not apply to software, fuses, measuring probes and
batteries. VELLEMAN-KIT N.V will not be held responsible for any flaw
or defect which the company feels is due to negligence on behalf of the
user, to modification or opening of the unit, or to accidents or abnormal
use or treatment of the product.
VELLEMAN-KIT N.V will not reimburse the transport costs or risks, the
costs for removing and replacing the product or any other costs that are
directly or indirectly related to the defect. VELLEMAN-KIT N.V. accepts
no liability for whatever damages may be caused by a malfunctioning
product.
19
PersonalScopeTM
POWER SUPPLY SECTION
+V1
R17
10K
LOW BAT
T1
D13
1
2
PMBT2222A
BATTERY
C6
100n
GND
GND
GND
+V1 GND
LD1
PWR
T11
L3
AGND
1
GND
TP4
+V2
220µ
C24
100µ
T13 PMBT2907A
PMBT2907A
C2
C3
-V1
PMBT2222A
100µ
22K
R60
47K
PWM
1K
T4
PMBT2222A
GND
L4
R32
D2
1K
BAT54S
C4
C5
100µ
100µ
GND
GND
+
GND
100µ
D1
GND BAT54S
R26
1K
R27
20
SCREENING
AGND
T3
R33
1
+
GND
SPR
+
BAV99
LED3RL
GND
C23
100µ
AGND
0
L2
+
C20
100µ
C7
100n
220µ
R15
D12 10K
R1
15
+V6
220µ
+
JP1
0
DGND
+V1
L1
+
GND
+
DGND
3
OUT
JP2
BDP31
PRLL4001
C1
100µ
IN
T5
R2
330
T2
D7
1
R3
330
2
4
-
47K
F1
FUSE 2A
R16
10K
LM2940S5.0(3)
IC1
ZD1
+
+
PMBT2222A
DGND
R5
GND
SK2
R34
22K
PMBZ5240B
Personal Scope: Power Supply Section
+
GND
4K7
BAS85
BDP32
-V2
220µ
C25
100µ
GND
C26
+
PRLL4001 D6
PRLL4001
DC JACK
R35
PMBT2907A
100µ
+
SK1
T6
T12
D11
+V3
C27
100µ
AGND AGND
PersonalScopeTM
ANALOG SECTION
+V1
BAV99
RY1
D5
EDR301A05
+V1
BAV70
D9
R20
+V2
AC/DC
PMBT2222A
R6
47K-1%
D4
16
AGND
1
5
2
4
AGND
R40
100-1%
6
VDD
Y
A
B
2
D3
NC
IC4
AGND
EDR301A05
R49
10
33K
9
6
ATT1
4K7-1%
R37
RV2
100
ANALOG OUT
R30
AD IN
1K
C43
68p
AGND
C11
100n
R36
4K7-1%
10K
DGND
TP3
LF357N(8)
-V2
R18
T9
3
8
1
OS NULL+
IN+
OUT
IN-
2
13
+V4
AGND
22p
R51
C40
IC2
12
14
15
11
YPOS
ATT2
R48
47-1%
R53
R54
180-1%
R55
27-1%
R52
ATT3
33K
C12
100n
C13
100n
DGND DGND
C37
10n
560-1%
R45
47-1%
R56
2K4-1%
750-1%
1
5
2
4
X0
X1
X2
X3
Y0
Y1
Y2
Y3
DGND
6
R44
2K4-1%
BAV99
PMBT2222A
Y0
Y1
Y2
Y3
VSS
510-1%
50p
X0
X1
X2
X3
INH
Transient
CV2
BAV99
12
14
15
11
14
1
RY3
R19
10K
MC74HC4052
13
X
8
R43
IC3
AGND
6
C36
330p C30
47p
C29
100µ
R42
910K-1%
8
7
+V4
+
22K
5
3
18p
16
CV1
RV1
BAS45AL
D8
C39
R28
1K
R41
39K-1%
J1
OS NULL-
AGND AGND
MC14052B
VCC
C38
1n5/200V
20p
1
INH
100K
AGND
DGND
VCC -
100
+V6
13
X
+V1
3
Y
R38
4K7
AGND
R47
10
A
ATT4
33K
9
B
R46
ATT5
33K
VEE
R7
47K
AGND
+V6
4
47n/200V
T8
R50
VEE
R39
7
C35
GND
RY2
DGND
SK3
BNC
C8
100n
DGND
C9
100n
DGND DGND
7
C10
100n
8
PMBT2222A
7
EDR301A05
VCC+
10K
T7
R29
-V2
1K
+
C28
100µ
AGND
CV1: Transient adjustment at 2V/div
CV2: Transient adjustment at 1V/div
RV1: Offset adjustment (switch between 1 & 2V/div)
RV2: Gain adjust at 0.1V/div (first adjust RV3 !)
Personal Scope: Analog input section
ZD2
PLVA2656A
AGND
21
PersonalScopeTM
DIGITAL SECTION
C32
470n
RV3: Callibration at 50mV/Div
RV4: Center setting in auto setup mode
R9
R10
RV4
R11
47K
47K
47K
CENTER
J2
22K
TP2
1
Auto center
AGND
CENTER YPOS
+V2
8
C31
470n
IC8
+V3
1
+V6
DGND
PROBE TEST
AGND
AN IN
12
SW1
R59
1K
AD CLK
DGND
AC/DC
ATT1
ATT2
ATT3
ATT4
ATT5
DGND
Ypos
tVdiv
Trig
ZD5
Mark
DGND
R58
14
2
3
4
5
6
7
8
36
41
38
39
37
42
43
44
1
17
28
40
PLVA2656A
1M
+V1
Right
Up
Contr.
OSC1/CLKIN
MCLR/VPP
RA0
RA1
RA2
RA3
RA4/T0CKI
RA5/SS
RB0/INT
RB4
RB2
RB3
RB1
RB5
RB6
RB7
NC
NC
NC
NC
OSC2/CLKOUT
RC0/T1OSO/T1CKI
RC1/T1OSI/CCP2
RC2/CCP1
RC3/SCK/SCL
RC4/SDI/SDA
RC5/SDO
RC6/TX/CK
RC7/RX/DT
RD0/PSP0
RD1/PSP1
RD2/PSP2
RD3/PSP3
RD4/PSP4
RD5/PSP5
RD6/PSP6
RD7/PSP7
RE0/RD
RE1/WR
RE2/CS
DGND
15
16
18
19
20
25
26
27
29
21
22
23
24
30
31
32
33
9
10
11
DGND
+V1
R31
1K
DGND
17
16
15
4
PWR
V+
1
CENTER REF
6
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
6
7
8
9
10
11
12
13
14
5
AD OE
AD CLK
LCD1
Meter
Down
Left
BAL
DGND
LM311N(8)
PLVA2656A
ZD4
DGND
OnOff
Auto
Input
R25
10K
Display
Personal Scope: Digital Section
ZD3
PLVA2656A
DGND
DGND
22
Hold
RST
CS2
CS1
D/I
E
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
R/W
A
K
PG12864
DGND
+V1
PLVA2656A
ZD6
7
GND
V-
5
OUT
1
4
TP1
DGND
BAL/STRB
IN+
IN-
470n
PIC16C65A20/L(44)
IC6
2
3
C34
C19
100n
LOW BAT
8
ProbeX1
-V2
3
11
13
2
24
20
21
AGND
R21
10K
C41
22p
IC7
DGND
+V1
TLC5510INS
AD IN
20Mc
2
10K
DGND DGND
270K
VDD
100n
C42
22p
12
35
100µ
PLVA2656A
R14
47K
C33
470n
R8
VDD
VDD
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
C22 + C18
DGND
DGND
DGND
19
AGND
AGND
CLKIN
C17
100n
3
4
5
6
7
8
9
10
AD OE
R13
CONTRAST
100
REFB
REFBS
1
R12
47K
VEE
10K
23
22
X1
VSS
R22
10K
R4
BAS85
D10
ZD7
1
R23
D0
D1
D2
D3
D4
D5
D6
D7
REFTS
REFT
CALIBRATION
500
AGND
ZD8
BZD27-C6V8
VSS
VSS
RV3
C16
100n
OE
LM358M(8)
13
34
16
17
VDDD
VDDD
AGND AGND
7
PWM
DGND
DGND
IC5
VDDA
VDDA
VDDA
C14
100n
14
15
18
C15
100n
+
C21
4µ7
T10
PMBTA14
18
R24
10K
R57
4.7
1
IN1+ OUT1
IN1IN2+ OUT2
IN24
+V1
+V1
AGND
V-
SK4
V+
3
2
5
6
DATABUS[DB0...DB7]
-V1
19
20
YPOS
®
Legen Heirweg 33, 9890 Gavere
Belgium Europe
+32 (0)9 3843611
http://www.velleman.be
HHPS5 - 1999 - ED1
France
VELLEMAN ELECTRONIQUE
+33 (0) 320158615
Netherlands
VELLEMAN COMPONENTS
+31 (0) 765147563
USA
VELLEMAN INC.
+1 (817) 284 7785