Download Xuri™ Cell Expansion System W25
Transcript
GE Healthcare Life Sciences Data file 29-0618-52 AA Cell Expansion Xuri™ Cell Expansion System W25 The single-use Xuri Cell Expansion System W25 is a reliable and intuitive cell expansion device for working volumes up to 25 l (Fig 1). The system is based on the well-known WAVE™ rocking technology that provides mixing and aeration to the culture. With the Xuri Cell Expansion System W25 , the rocking is enhanced with features such as advanced sensors and intelligent control strategies. System operation is easily managed via the intuitive and user friendly UNICORN™ system control software. The system is designed for fast installation and convenient handling, and delivers a reliable and accurate performance suitable for research, process development, and manufacturing operations. Xuri Cell Expansion System W25 offers the following benefits: • Minimize the risk of cell contamination with a functionally closed system • Reach high cell densities with confidence • Be in charge of your cell expansion with process monitoring and remote control • Designed for use in a regulated environment Applications Xuri Cell Expansion System W25 is intended for a variety of cell culture applications incuding: • Rapid production of clinical grade T-lymphocytes • Perfusion culture of T-lymphocytes • Perfusion culture of human natural killer cells System benefits Minimize the risk of cell contamination with a functionally closed system The Xuri Cell Expansion System W25 is a single-use, Fig 1. Xuri Cell Expansion System W25 bioreactor system. functionally closed system meaning that the chance of contamination between different patient samples or with adventitious agents is minimized. Reach high cell densities with confidence Perfusion is easy on the Xuri Cell Expansion System W25. Perfusion allows high cell densities to be obtained in the Cellbag. This means that an entire therapeutic dose can be obtained in one closed bioreactor system, eliminating the difficulties with combining culture vessels as is required with manual systems. High cell density and reduced manufacturing footprint allows you to maximize use of your limited manufacturing space. Be in charge of your cell expansion with process monitoring and remote control The Xuri Cell Expansion System W25 is designed to put complete control of the culture in your hands whether you are standing next to the bioreactor or at home in the evening. By giving you control of process parameters such as rocking speed, DO, pH and perfusion rate, you can optimize your cell culture protocol to obtain therapeutic cells in the most efficient manner for you. Once the culture is setup, remote monitoring gives you the peace of mind to leave it knowing that you will be alerted if anything changes from the limits you set. Lid Graphical user interface Xuri Cell Expansion W25 Pump Peristaltic pump Client computer Installed with UNICORN software Tray Xuri Cell Expansion System W25 Base Unit Rocker, temperature sensors, scale, embedded controller, and dual hard drives Xuri Cell Expansion W25 Environmental Control Unit Gas flow/mix controller, pH, and DO sensors System setup and integration of subunits Xuri Cell Expansion System W25 Base Unit Xuri Cell Expansion W25 Pump • Acid/base flow • Feed/harvest flow • Media addition • Rocking • Weight measurement • Temperature measurement UNICORN software • • • • Xuri Cell Expansion W25 Environmental Control Unit Administration System control Method editor Evaluation • Gas flow/mix • pH measurement • DO measurement Controller with dual hard drives LAN OPC UNICORN database SCADA (not included) Firewall UNICORN software At home or within enterprise network E-mail notifications Fig 2. Xuri Cell Expansion System W25 overview. The System Control module on the client computer is used to start and monitor the cultivation process. One UNICORN client can control up to three Xuri Cell Expansion W25 systems simultaneously. SCADA = supervisory control and data acquisition, OPC = open platform communications, LAN = local area network. 2 29-0618-5 AA Designed for use in a regulated environment Xuri Cell Expansion W5, base unit The Xuri Cell Expansion System W25 is designed to meet the demands and standards that are required in a regulated environment. The documentation is extensive and includes material certificates, system specifications, installation and operational qualification (IQ/OQ) protocols, and a detailed user manual. The UNICORN software is suitable for use in a manner that complies with 21 CFR Part 11 and Good Automated Manufacturing Practice (GAMP) 5. The base unit is the main hardware unit of the system and provides mixing through the rocking, reliable temperature measurement from integrated sensors, and accurate weight measurement from integrated load cells. The ergonomic design makes handling convenient and easy. The ability to handle the bioreactor in a tilt position facilitates sampling and harvest (Fig 3). The minimal footprint facilitates placement when space is limited. System overview Trays and lids Xuri Cell Expansion System W25 components The Xuri Cell Expansion System W25 consists of a base unit, a gas mixer, and a pump, all operated by the UNICORN software installed on a client computer. The base unit is the main unit and is used together with a tray and a disposable Cellbag bioreactor. The base unit has multiple functions, amongst them heating, culture mixing, and weight measurement. The gas mixer delivers gas of a defined composition to the culture and is used, together with the optical sensors of the Cellbag bioreactor, for online control of culture pH and DO. Liquid is delivered accurately to the culture by the Xuri Cell Expansion W25 pump. The pump unit has a flow range covering multiple applications, from additions of acid and base to fresh medium for perfusion culture. An overview of the Xuri Cell Expansion System W25 is shown in Figure 2. The individual components are described in detail below. A) Trays are available in three different sizes for culturing of up to 5 l, 10 l, and 25 l. The trays are easily attached to the base unit when in the tilt position. A snap lock mechanism allows Cellbag bioreactors to be correctly installed and rapidly changed (Fig 4). To protect light-sensitive components of the culture medium and to prolong the life of the optical sensors, lids are available for all tray sizes. Snap locks Fig 4. The tray has an easy lock mechanism for convenient and proper attachment of Cellbag bioreactors. Temperature control Reliable, efficient, and evenly distributed heating is provided by the tray heater plate. The temperature measurement is managed by sensors integrated in the rocker base unit (Fig 5). To minimize the risk for overheating, heating is only enabled when the base unit is in motion. For an accurate, stable, and fast temperature control, the heater power output is automatically adjusted based on the Cellbag bioreactor size and the media volume. B) Temperature sensors Fig 5. Integrated temperature sensors for convenient handling and reliable temperature control. Fig 3. (A) Xuri Cell Expansion System W25 in tilt position. (B) The ergonomic design makes activities such as sampling and harvest convenient and easy. 29-0618-5 AA 3 Mixing and gas transfer The adjustable rocking parameters are speed, angle, and motion. The parameter settings, in combination with the cell culture volume, affect the mixing and gas transfer rate in the Cellbag bioreactor. The speed parameter determines the number of rocking cycles per minute, and the angle parameter relates to the tray’s degree of tilting at the turning Rocking profile: 15% Angle (°) 10 5 0 -5 Angular velocity (°/s) -10 0 0.5 1.0 1.5 Time (s) 2.0 2.5 3.0 0 0.5 1.0 1.5 Time (s) 2.0 2.5 3.0 20 10 0 -10 -20 Rocking profile: 100% Angle (°) 10 Data storage and communication All run data and control parameters are stored on two independent, mirrored solid state drives integrated in the base unit. Thus, once cultivation has been started and data from the instrument modules have been collected, the run can continue without the need for a network connection. The use of two solid state drives enables culturing without losing any data even if only one drive is functional. After a run, data is stored in a UNICORN database on an external computer. For facilitated integration into a larger manufacturing operation, the base unit also contains software for enabling direct communication with the system via an open platform communication (OPC) link. Xuri Cell Expansion W25 Environmental Control Unit 5 0 Table 1. Environmental Control Unit overview -5 Feature -10 0 Angular velocity (°/s) Weight measurement Load cells integrated in the base unit provide accurate and continuous weight measurement of the culture and eliminate the need for external scales. The load cells communicate with the system control function for proper media handling and for tailored regulation of pH and temperature. The weight measurement functionality is also used for automatic calibration of the pump in perfusion cultures. Adjustable feet enables equal weight distribution between the load cells also on non-level surfaces, for increased precision of the weight measurement. 0.5 1.0 1.5 Time (s) 2.0 2.5 3.0 20 10 0 -10 CO2/O2/air mix Depending on the configuration, air is mixed controller with CO2 and/or O2 according to the setpoints. Nitrogen may be used instead of air for maintaining a low oxygen environment for nearanaerobic applications. Gas flow controller The gas-mix flow is measured and controlled by a mass flow controller (MFC). A correct volumetric flow is achieved by compensation for CO2 concentration. Quick filling of gas into the Cellbag bioreactor, enabled by the fast-fill function, significantly reduces start-up time. An alarm will inform the user if pressure sensors detect underor overpressure at the gas inlets or within the Cellbag bioreactor. pH measurement The pH is measured with optical pH sensors preinstalled in Cellbag bioreactors of pHOPT type. The sensor is connected to the Environmental Control Unit via an optical fiber cable. DO measurement The DO level is measured with optical DO sensors preinstalled in Cellbag bioreactors of DOOPT type. The sensor is connected to the Environmental Control Unit via an optical fiber cable. -20 3.0 1.5 2.0 2.5 Time (s) Fig 6. A user-defined rocking motion can be selected in UNICORN system control software. The wave can be aggressive or gentle, with subsequent impact on mixing and gas transfer in the culture. 0 0.5 1.0 points. The motion parameter determines the acceleration profile. The lowest motion parameter setting, 15%, gives an almost constant speed throughout the rocking cycle, whereas the highest setting, 100%, gives a faster speed in the middle of the cycle and slower at the turning points (Fig 6). A lower percent setting will give a more aggressive mixing with higher gas transfer rates. A higher setting will result in a smoother wave suitable for more delicate applications, such as adherent cultivation on microcarriers. 4 29-0618-5 AA Description The Xuri Cell Expansion W25 Environmental Control Unit is a compact unit with multiple functions. The unit is primarily used for providing gas to the culture and for monitoring of pH and DO. The unit contains a controller for mass flow, sensors for gas pressure and O2 and CO2 concentrations, and transmitters for pH and DO (Table 1). Two Environmental Control Unit configurations are available: • CO2, O2, and pH • CO2, O2, pH, and DO The most suitable Environmental Control Unit configuration will depend on the specific application. Xuri Cell Expansion W25 Pump The Xuri Cell Expansion W25 pump is a peristaltic unit incorporating two roller pumps for feed, harvest, and pH control. The pump design makes tubing installation convenient and easy. Tubing sizes from 0.5 mm (1/50”) to 4.8 mm (3/16”) internal diameter can be used to support flow rates fom 0.07 to 100 ml/min. Flow rate is regulated by weight feedback from the scale and is controlled by automatic adjustments of the pump speed. Manual adjustment of the flow rate is also possible. changed to simplify the editing process. Individual changes can be saved for later use on systems having the same instrument and component configuration. System control The System Control module is used to connect to the system as well as to start, view, and control a run (Fig 7). Default parameter values for an instrument can be viewed and edited in the System Settings dialog before a run is started or during an ongoing run using manual instructions. It is also possible to connect to other systems. The user-friendly and intuitive Process Picture pane enables manual interaction with the system and provides status of the run parameters. Data is shown in the process picture but can also be viewed as curves in the Charts tab and as numerical values in the Run Data pane. Curves and information about the run are saved in a database, which can be opened in the Evaluation module. The default view shows the most commonly curves used. The user may customize which curves to display and the color and style of the displayed curves. Media handling and perfusion using calibrated pumps Calibration of pumps can be performed manually or automatically during an ongoing perfusion process. Automatic calibration is possible for flow rates above 3.5 l/d and is easily managed by entering the tubing diameter. Automated pH control To minimize fluctuations in pH during adjustments with CO2/ base or acid/base, the flow rate of the acid/base pumps needs to be rigorously monitored. The Xuri Cell Expansion System W25 uses combined information from the load cells, the current process pH value, the pump tubing diameter, and the acid/base molarity for calculation of the needed flow rate during the run. UNICORN software UNICORN system control software consists of four modules: Administration, Method Editor, System Control, and Evaluation. A comprehensive help tool is also included. Administration The Administration module is used to administer all functions of the UNICORN software. Available functions include user and email setup, controlling access to groups and network users, defining and editing system properties, database management, and logging records of usage and activity. Method editor Instructions to control a bioreactor run can be defined in a method. The Method Editor module is used to create, edit, save, and work with methods. An existing method can be Fig 7. Monitoring and controlling all aspects of the culture is managed through the user-friendly interface. Evaluation In the Evaluation module, content of the result files can be viewed, analyzed, and compiled as reports. Reports can be customized, saved, and printed. Automation of process parameters in pH and DO control The UNICORN software automatically sets the desired PID* control parameters based on Cellbag bioreactor size and gas flow set-point. The system also adapts the parameters during the run, for optimization to reach a new set-point or to maintain a current set-point. * Proportional-integral-derivative Alarms, notifications, and data logging The UNICORN software allows easy access to all operations, data, and alarm conditions. A dynamic, graphical user interface shows the real-time status of the run, while data are automatically saved. From the UNICORN software, the system is easily configured to trigger an alarm when certain conditions are met. Individual deviation alarms can be set for all essential 29-0618-5 AA 5 parameters (heating, gas flow, CO2, pH, and DO). Alarm sensitivity and delay is defined by the user. If such an alarm condition is triggered, the source of the alarm is highlighted in the graphical user interface and an alarm dialog is displayed. The alarm dialog displays information about the alarm, such as date and time of occurrence, as well as a help text describing the cause and how to solve the problem. The system is easily configued to send email notifications of alarms and errors Regulatory readiness The Xuri Cell Expansion System W25 is suitable for biomanufacturing of regulated products under various Quality Management Systems. The use of UNICORN software in a 21 CFR Part 11 and GAMP 5 compliant manner enables use of the system in a regulated environment. Individual user access permissions can be set and individual users are password-protected. Active processes can be locked to enable unattended operations without the risk of unauthorized interference. All records are stored in a single, unalterable database, including results and extended run documentation. Additional validation support includes comprehensive documentation on control system validation as well as IQ/OQ services. Available validation support documentation includes: • Detailed description of the development model used for UNICORN software Cellbag bioreactors Cellbag bioreactors are single-use bags for the noninvasive mixing of culture medium and cells during cultivation (Fig 8). Cellbag bioreactors require no sterilization or cleaning steps. The disposable bioreactors provide a suitable environment for cell growth while minimizing the cross-contamination risk. The bags are manufactured from multilayer, laminated, clear USP Class VI plastics and are easily connected to the full suite of Xuri Cell Expansion cell culture, purification, and fluid handling products. The selection of Cellbag bioreactors is listed in Table 2. For more information about Cellbag bioreactors, see data file 29-0618-52. Optical DO sensor embedded in bag Air outlet filter Optional tubing Air inlet filter Inoculation/ harvest lines Optical pH sensor embedded in bag Cellbag rod Optional tubing Optional tubing Needleless sampling port Fig 8. Presentation of typical Cellbag bioreactor fittings. • 21 CFR Part 11 system assessment in checklist format • Audit report and 21 CFR Part 11 conclusion on functionality by an external, independent expert Table 2. Selection of Cellbag bioreactor sizes Cellbag bioreactor* Working volume Networking capabilities UNICORN software operates in Windows XP® and Windows® 7 environments and the network ability allows real-time control from a remote or local computer. Communication is Ethernet-based and each instrument is controlled by a dedicated instrument server. One database can be connected to 32 systems and up to three instruments can be controlled simultaneously from one UNICORN client. Results are saved locally, in the base unit, during the run and then stored on the database server. Because results and instructions are saved locally on the embedded drives, a run can continue even in the event of a network communication failure. Cellbag 2 l 300† ml to 1 l Cellbag 10 l 500 ml to 5 l Cellbag 20 l 1 to 10 l Cellbag 22 l 1 to 10 l Cellbag 50 l 5 to 25 l UNICORN software can be integrated in a supervisory control and data acquisition (SCADA) system such as the Emerson DeltaV™ control system using the UNICORN OPC server. 6 29-0618-5 AA * All Cellbag bioreactor sizes are available with internal perfusion filter for perfusion cultures and/or can be equipped with single use optical pH (pHOPT) and DO (DOOPT II) sensors. † Cellbag bioreactors with optical sensors require 300 ml minimal working volume. Depending on application and configuration, it might be possible to cultivate below the recommended minimum volume. However, it is highly recommended to stay above this volume for applications that require high agitation and pH and DO control. The temperature, pH, and DO sensors need to be submerged in liquid throughout the complete rocking cycle to function correctly. System specifications Xuri Cell Expansion W25 Environmental Control Unit System specifications are listed in Table 3. Gas flow control range 0 to 1000 ml/min Table 3. Xuri Cell Expansion W25 specifications Gas flow total accuracy (process value to reference flow) ± 10 ml/min + 3% of read value Fast-fill flow ~ 3 l/min CO2 control range 0% to 15% CO2 measurement accuracy ± 0.3% + 3% of read value when mixed only with air/N2 General system specifications Control system UNICORN 6.3 or later version Dimensions, W × H × D Base unit Tray 10 Tray 20 Tray 50 Lid 10 Lid 20 Lid 50 Environmental Control Unit Pump 25 404 × 560 × 205 mm 475 × 430 × 60 mm 740 × 480 × 70 mm 800 × 610 × 70 mm 475 × 430 × 230 mm 740 × 480 × 245 mm 800 × 610 × 260 mm 276 × 360 × 117 mm 275 × 280 × 115 mm Weight Base unit Tray 10 Tray 20 Tray 50 Lid 10 Lid 20 Lid 50 Environmental Control Unit Pump 25 24 kg 4.5 kg 7.3 kg 9.5 kg 1.7 kg 3.3 kg 3.9 kg 4.8 kg 3.8 kg Power supply 100 to 240 V, ~ 50 to 60 Hz Power consumption 1500 VA Enclosure protective class IP 21 Total CO2 accuracy (process value to reference CO2 content) ± 0.7% + 3% of read value O2 control range 0% to 50% when mixed with N2, 21% to 50% when mixed with air O2 measurement accuracy ± 0.6% + 1% of read value within 0% to 50% when mixed only with air/N2 Total O2 accuracy ± 1.2% + 1% of read value pH measurement range 4.5 to 8.5 pH measurement accuracy ± 0.05 for pH 6.75 to 7.25 ± 0.1 for pH 6.5 to 7.5 pH control range 6.0 to 8.0 Total pH accuracy (process value to reference pH) ± 0.1 for pH 6.75 to 7.25 ± 0.15 for pH 6.5 to 7.5 DO measurement range 0% to 250% air saturation DO measurement accuracy ± 5% air saturation (excl. atmospheric pressure variations) Gas supply Normal capacity Fast-fill capacity External air supply, 1.0 to 1.5 bar 1.3 l/min 3.5 l/min External CO2 supply, 1.0 to 1.5 bar 0.2 l/min 0.5 l/min External O2 supply, 1.0 to 1.5 bar 0.7 l/min 1.7 l/min DO control range 0% to 100% air saturation Total DO accuracy (process value to reference DO) ± 10% air saturation (excl. atmospheric pressure variations) Environmental aspects Xuri Cell Expansion W25 Pump Operating ambient temperature range 15°C to 32°C Operating humidity range 20% to 80% (noncondensing) Xuri Cell Expansion System W25, base unit Pump flow rate range 100 ml to 144 l/d (0.07 to 100 ml/min) Pump flow accuracy ± 0.1 ml/min + 5% of read value after calibration Rocking speed control range1 2 to 40 rpm Accumulated pumped volume accuracy Rocking angle control range1 2° to 12° Supported tubing dimensions Rocking motion control range 15% to 100% Media weight control range 0.5 to 25 kg Scale, absolute accuracy ± 0.050 kg + 1% of load kg Temperature sensor Pt100 Class A Temperature measurement range 2°C to 50°C Temperature control range 20°C (or ambient temperature + 5°C) to 40°C Temperature control accuracy (incl. measurement error) ± 10% of measured volume i.d. 0.5 to 4.8 mm (1/50” to 3/16“) wall thickness: 1.6 mm (1/16”) ± 0.5°C When cultivating in a 50 l Cellbag bioreactor at maximum working volume of 25 l, rocking speed and angle multiplied should not exceed 240 rpm degrees. For example, if the rocking angle is set to 12 degrees, the rocking speed should not be set higher than 20 rpm. 1 29-0618-5 AA 7 System components Product Description Xuri Cell Expansion System W25 Base Unit Base unit Xuri Cell Expansion System W25 Environmental Control Unit pH Gas flow/mix and pH Xuri Cell Expansion System W25 Environmental Control Unit Full Gas flow/mix, pH, and DO Xuri Cell Expansion System W25 Pump Pump UNICORN 6.3.2 WrkStn-pure-BP Software and license UNICORN 6.3 Remote License – UNICORN 6.3 Dry license – Tray 10 – Tray 20 – Tray 50 – Lid 10 – Lid 20 – Lid 50 – Filter heater – Bag sensor adaptor 2.5 m assembly Fiber cable for pH and DO control Related literature Code number UNICORN 6.3 Manual Package 29-0234-67 UNICORN 6 control software, data file 28-9573-46 Validation Support File UNICORN 6.3 11-0029-94 GE, imagination at work, and GE monogram are trademarks of General Electric Company. Cellbag, UNICORN, WAVE and Xuri are trademarks of GE Healthcare companies. DeltaV is a trademark of Emerson Electric Co. Windows and Windows XP are registered trademarks of Microsoft Corporation. Any use of UNICORN software is subject to GE Healthcare Standard Software End-User License Agreement for Life Sciences Software Products. A copy of this Standard Software End-User License Agreement is available on request. For local office contact information, visit www.gelifesciences.com/contact www.gelifesciences.com GE Healthcare UK Limited Amersham Place Little Chalfont Buckinghamshire, HP7 9NA UK Cellbag bioreactors with integrated optical sensors are sold under a sublicense from Sartorius Stedim Biotech under US patent numbers 6,673,532, 7,041,493, and/or its foreign equivalents. © 2013 General Electric Company—All rights reserved. First published Jul. 2013 All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information. GE Healthcare UK Limited Amersham Place Little Chalfont Buckinghamshire, HP7 9NA UK GE Healthcare Europe, GmbH Munzinger Strasse 5 D-79111 Freiburg Germany GE Healthcare Bio-Sciences Corp. 800 Centennial Avenue, P.O. Box 1327 Piscataway, NJ 08855-1327 USA GE Healthcare Japan Corporation Sanken Bldg., 3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073 Japan imagination at work 29-0618-5 AA 07/2013