ERRATA CARD
131220
GasAlertMicro 5 Series
X5 Series
Which pump do I have?
The best way to differentiate the Generation 2 pump from the Generation 1 pump is the inline filter that is visible on the
Generation 2 pump on the left-hand side. The product labels are also different.
Generation 1 Pump
Generation 2 Pump
Inline filter
Part number 116885 (yellow) and 118933 (black)
Part number 130916 (yellow) and 130917 (black)
Operation Information for the Generation 1
pump
Using the sintered filter
Operation Information for the Generation 2
pump
Using the sintered filter
For temperatures -10°C to +0°C (14°F to 32°F)
For temperatures -20°C to +50°C (-4°F to 122°F)
For temperatures 0°C to +50°C (32°F to 122°F)
Using the sample probe
The maximum tube length when using the metallic sintered filter is 10 ft. (3 m).
The maximum tube length when using the metallic sintered filter is 30 ft. (9.1 m).
Using the sample probe
The maximum tube length when using the metallic sintered filter is 67 ft. (20 m).
For temperatures -20°C to +50°C (-4°F to 122°F)
The maximum tube length when using the sample probe
is 10 ft. (3 m)
For temperatures -10°C to +0°C (14°F to 32°F)
The maximum tube length when using the sample probe
is 10 ft. (3 m).
a Warning
When using the sample probe at -10°C to +0°C
(14°F to 32°F), keep the sample probe in your
hand .
For temperatures 0°C to +50°C (32°F to 122°F)
The maximum tube length when using the sample probe
is 10 ft. (3 m).
see over for more important pump
information
Information for both Generation 2 and Generation 1 pumps
a Warning
Do not upgrade your X5 Series unit unless you have
contacted BW Technologies by Honeywell first.
The following information has been changed for the
GasAlertMicro 5 Series/X5 Series.
Quick Reference Guide (128900 D6258/1)
Proper Pump Operation
When using the pump module, attach it and the pump accessories prior to activating the detector.
a Warning
If the pump module is installed on the detector, the following three things must occur during start-up. If any one of
the conditions below does not occur, discontinue use
of the detector and contact BW Technologies by Honeywell immediately.
• The detector prompts for a pump test during start-up
• The pump module passes the pump test at start-up when
the pump inlet or sample chain inlet is blocked
• The J icon displays on the LCD
Pump Features
The sections regarding Replacing the Pump Filter and Replacing the Pump Nozzle are for the Generation 2 pump. To
replace those components, refer to Pump Module Auxiliary
Filter Instruction Card.
a Warning
To ensure accurate gas detection, the sensors must
be calibrated immediately when the pump module is
replaced by the diffusion cap and vice versa.
a Caution
A demand flow regulator must be used to manually
calibrate the GasAlertMicro 5/PID/IR detector when
the pump module is installed.
Calibration cap
Diffusion cap
Note
The calibration cap is designed for use with the
diffusion cap only. It cannot be used with the pump
module.
Pump Alarm
The external pump draws air over the sensors continually. If
the pump stops operating or becomes blocked, the detector
activates the pump alarm and the pump alarm latches. The
following screens display.
a Caution
Ensure the blockage is cleared before pressing C
to acknowledge the latched pump alarm.
When C is pressed, the detector automatically launches a
pump test to reset the pump module.
Refer to Pump Test in the GasAlertMicro 5/PID/IR User
Manual for more information. If the pump test is successful, the detector returns to normal operation, otherwise the
pump alarm continues. If the pump alarm persists, refer
to the Pump Operation section in Troubleshooting in the
GasAlertMicro 5/PID/IR User Manual.
Thank you for reading this data sheet.
For pricing or for further information, please contact us at our UK Office, using the details
below.
UK Office
Keison Products,
P.O. Box 2124, Chelmsford, Essex, CM1 3UP, England.
Tel: +44 (0)1245 600560
Fax: +44 (0)1245 808399
Email: [email protected]
Please note - Product designs and specifications are subject to change without notice. The user is responsible for determining the
suitability of this product.