Download Spectrum™ Continuous Flow Airline Respirator Instruction Manual

Transcript
Spectrum™ Continuous Flow Airline Respirator
Instruction Manual
Type C NIOSH Approval No.
TC-19C-322 Continuous-Flow Class and
TC-19C-321 Pressure-Demand Class.
READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS
RESPIRATOR. SAVE THIS MANUAL FOR FUTURE USE.
Table of Contents
APPROVAL LABELS ...................................................................................2,3
GENERAL INFORMATION................................................................................. 4
COMPONENT CONCEPT....................................................................................4
MASK/FITTING ASSEMBLIES.......................................................................... 5
AIR SUPPLY HOSE OPTIONS........................................................................... 5
WARNINGS
................................................................................ 5, 6
OPERATIONS .....................................................................................7
BREATHING AIR PRESSURE TABLE.................................................................. 8
RESPIRATOR ASSEMBLY
Putting Respirator On.......................................................................9
Taking Respirator Off........................................................................9
INSPECTION, CLEANING AND STORAGE...................................................... 10
FIT TESTING
................................................................................... 11
PARTS AND ACCESSORIES FOR
SPECTRUM RESPIRATORS............................................................. 11
RETURN AUTHORIZATIONS........................................................................... 12
www.bullard.com
SA/CF
19C-322
M45-CF-ML
SPEC-CF-ML
SPEC-CF-S
X X X X
M45-CF-S
www.bullard.com
B102
B100
B101
B135
BT100
BT108
BT102
BT101
BT135
X X X X X X X X X X
X
5454
3
Alternate Hoses
X
100’
100’
25’
50’
25’
50’
(7.6 M) (15.2 M) (30.5 M) (7.6 M) (15.2 M) (30.5 M)
5457
Flow Control Device
X
5458
X
X
X
Spectacle
Kit
X
PL
Anti-Fog
Lens Wipe
X
AFW45
Lens
Cover
X
LC
Accessories
Nose Cup
X
X
X
X
X
LNK 45PL 45LC SNA45 AMP
Amplifier
B. Not for use in atmospheres immediately dangerous to life or health.
C. Do not exceed maximum use concentrations established by regulatory standards.
D. Air-line respirators can be used only when the respirators are supplied with respirable air meeting the requirements of CGA G-7.1 Grade D or higher quality.
E. Use only the pressure ranges and hose lengths specified in the User’s Instructions.
J. Failure to properly use and maintain this product could result in injury or death.
M. All approved respirators shall be selected, fitted, used and maintained in accordance with MSHA, OSHA and other applicable regulations.
N. Never substitute, modify, add or omit parts. Use only exact replacement parts in the configuration specified by the manufacturer.
O. Refer to User’s Instructions, and/or maintenance manuals for information on use and maintenance of these respirators.
2. CAUTIONS & LIMITATIONS
1. PROTECTION
SA-SUPPLIED AIR
CF-CONTINUOUS FLOW
Protection1
TC#
B108
Respirator Components
4696
Spectacle
Kit
2
V2050ST
Lens
Cover
1
Alternate
Facepiece
V20100ST
Small
Nose Cup
Approved only in the following configurations:
Coupler Assemblies
X
CA30
Spectrum-CF
Type C Continuous Flow, Supplied Air Respirator
X
CA31
Spectrum Series
Bullard
1898 Safety Way
Cynthiana, Kentucky 41031
800-827-0423
X
CA32
Approval Labels
X
CA38
2
Cautions &
Limitations 2
Approval Labels
www.bullard.com
SPEC-CF-S
X X X X
SPEC-CF-ML
2
BT135
X
B135
X
X
X
100’
50’
(15.2 M) (30.5 M)
3
Alternate
Hoses
Spectacle
Kit
X
PL
Anti-Fog
Lens Wipe
X
AFW45
Lens
Cover
X
LC
Accessories
X
X
LNK 45PL
X
Small
Nose Cup
X
X
SNA45 AMP
Amplifier
Cautions &
Limitations 2
B. Not for use in atmospheres immediately dangerous to life or health.
C. Do not exceed maximum use concentrations established by regulatory standards.
D. Air-line respirators can be used only when the respirators are supplied with respirable air meeting the requirements of CGA G-7.1 Grade D or higher quality.
E. Use only the pressure ranges and hose lengths specified in the User’s Instructions.
J. Failure to properly use and maintain this product could result in injury or death.
M. All approved respirators shall be selected, fitted, used and maintained in accordance with MSHA, OSHA and other applicable regulations.
N. Never substitute, modify, add or omit parts. Use only exact replacement parts in the configuration specified by the manufacturer.
O. Refer to User’s Instructions, and/or maintenance manuals for information on use and maintenance of these respirators.
2. CAUTIONS & LIMITATIONS
1. PROTECTION
SA-SUPPLIED AIR
PD-PRESSURE DEMAND
SA/PD
M45-CF-S
Alternate
Facepiece
M45-CF-ML
Protection1
V2050ST
Nose Cup
TC#
V20100ST
Spectacle
Kit
Respirator Components
Lens
Cover
Approved only in the following configurations:
Spectrum-CF-P
Type C Pressure Demand, Supplied Air Respirator
Spectrum Series
Bullard
1898 Safety Way
Cynthiana, Kentucky 41031
800-827-0423
Spectrum™ Continuous Flow Airline Respirator
Instruction Manual
3
GENERAL INFORMATION
Bullard’s Spectrum-CF airline respirator, when properly used, provides a
continuous flow of air from a remote air source to the respirator wearer.
Spectrum-CF respirators offer protection from airborne contaminants that
are not immediately dangerous to life or health (IDLH), or that do not exceed
concentrations allowed by applicable OSHA, EPA, NIOSH or ACGIH regulations and recommendations for continuous-flow or pressure-demand class
airline respirators.
Spectrum-cf respirators are compatible with breathing air sources such as
air compressors or Bullard Free-Air® Pumps. Bullard offers the appropriate approved breathing tube assembly and air supply hose to connect the
Spectrum-CF respirator to these breathing air sources. This respirator is
available in two mask sizes. A probed mask kit is available for quantitative fit
testing. Contact Bullard or our local authorized distributor for more information about other accessories for Spectrum-CF respirators.
Spectrum-CF airline respirators are approved by NIOSH (TC-19C-322 Type
C, Continuous-Flow Class and TC-19C-321, Pressure-Demand Class) to provide respiratory protection in general purpose applications, including spray
painting, tank cleaning, chemical and pesticide handling, and other industrial
or agricultural applications. The Spectrum-CF is NOT to be used in confined
spaces or in IDLH conditions.
COMPONENT CONCEPT
The Spectrum-CF airline respirator consists of three components: full-face mask, breathing tube assembly and air supply hose. All must be present and properly assembled to constitute a complete NIOSH approved respirator.
1. Full-Face Mask with Headstrap.
2. Breathing Tube Assembly
3. Air Supply Hose
General Information / Component Concept
➊
4
➋
➌
Figure 1
Clean Breathable
Air Source
Supplying Grade “D”
or Higher Air Quality
(See Breathing Air
Requirements on
page 8)
� WARNING
Failure to use complete NIOSH-approved Bullard components and replacement parts voids approval of entire assembly. Basic parts are
listed on the NIOSH Approval Label on Pages 1 and 2.
www.bullard.com
Spectrum™ Continuous Flow Airline Respirator
Instruction Manual
MASK/FITTING ASSEMBLIES
FULL-FACE MASK RESPIRATOR WITH BREATHING TUBE ASSEMBLY:
Comfortable silicone full-face mask with breathing tube that connects the mask to air supply hose.
For use with breathing air compressors/continuous flow
or pressure demand
spec40, 1/4” Industrial Interchange
For use with ambient air pumps/continuous flow
or pressure demand
SPEC35, 1/2” Industrial Interchange
SPEC41, 1/4” Schrader
SPEC42, 1/4” Snap-Tite
SPEC48, 1/4” Bayonet
AIR SUPPLY HOSE OPTIONS
Connects breathing tube to air source supplying clean breathable air.
Hose for High Pressure
Compressed Air Source
Hose for Low Pressure
Ambient Air Source
V10
3/8” I.D. Hose
V20
1/2” I.D. Hose
4696 Starter Hose
545 Series Extension Hose
V20 Starter/
Extension Hose
Extension hose available in 25, 50 and 100 foot lengths with a variety of
quick-disconnect fittings, styles and materials. See parts list for details.
Available in 50 and 100 foot lengths with a variety of quick-disconnect fittings, styles and materials. See parts list for details.
2. Before using this respirator, be sure your employer has determined that airborne contaminant concentrations do not exceed those allowed by applicable
OSHA, MSHA, EPA, NIOSH or ACGIH regulations and recommendations, or any other applicable regulations for continuous-flow airline and pressuredemand airline respirators. Federal law requires that your employer measure and monitor airborne contaminant levels in the work area.
3. Improper respirator use may damage your health and/or cause your death. Improper use may also cause certain life-threatening delayed lung diseases
such as silicosis, pneumoconiosis or asbestosis.
4. DO NOT wear this respirator if any of the following conditions exist:
- Atmosphere is immediately dangerous to your life or health (IDLH).
- You CANNOT escape without the aid of the respirator.
- Atmosphere contains less than 19.5% oxygen.
- Work area is poorly ventilated.
- Unknown contaminants are present.
- Contaminant concentrations are in excess of regulations or
recommendations (as described in item 2 above).
5. Bullard recommends that you DO NOT wear this respirator until you have passed a complete physical exam (perhaps including a lung x-ray), conducted
by qualified medical personnel, and have been trained in the respirator’s use, maintenance and limitations by a qualified individual (appointed by your
employer) who has extensive knowledge of the Bullard Spectrum respirator.
www.bullard.com
Mask/Fitting Assemblies / Air Supply Hose Options
� WARNING
1. This respirator, when properly fitted and used, significantly reduces, but does not completely eliminate, the breathing of contaminants by the respirator
wearer. You may obtain better respiratory protection from other types of respiratory protection equipment such as a valve-operated pressure-demand
airline respirator or a pressure-demand self-contained breathing apparatus respirator.
5
� WARNING
6. DO NOT modify or alter this respirator in any manner. Use only NIOSH approved Bullard Spectrum components and replacement parts manufactured by
Bullard for use with this respirator. Failure to use NIOSH approved components and replacement parts such as hoses and flow-control devices, voids
NIOSH approval of the entire respirator, invalidates all Bullard warranties, and may cause death, lung disease or exposure to other hazardous or lifethreatening conditions.
7. Inspect all components of this respirator system daily for signs of wear, tear or damage that might reduce the degree of protection originally provided.
Immediately replace worn or damaged components with NIOSH approved Bullard Spectrum components or remove respirator from service. (See
INSPECTION, CLEANING AND STORAGE section for proper maintenance of the Spectrum respirator.)
8. Be certain your employer has determined that the breathing air source provides at least Grade D breathable air. This respirator must be supplied with
clean breathable air at all times.
9. DO NOT connect the respirator’s air supply hose to nitrogen, oxygen, toxic gases, inert gases or other unbreathable, non-Grade D air sources. To prevent
this, airline couplings used for this respirator shall be incompatible with outlet for other gas systems. Check the air source before using the respirator.
Failure to connect to the proper air source may result in serious injury or your death.
10. DO NOT use this respirator in poorly ventilated areas or confined spaces such as tanks, small rooms, tunnels or vessels, unless the confined space is
well ventilated and contaminant concentrations are below the upper limit recommended for this respirator. In addition, follow all procedures for confined
space entry, operation and exit as defined in applicable regulations and standards, including 29 CFR 1910.146.
11. If you have any questions concerning the use of this respirator, or if you are not sure whether the atmosphere you are working in is immediately dangerous to life or health (IDLH), ask your employer. All instructions for the use and care of this product must be supplied to you by your employer as
recommended by the manufacturer and as required by Federal Law (29 CFR 1910.134).
12. DO NOT use this respirator for abrasive blasting or underwater diving.
cautions and limitations
Warnings / Cautions and Limitations
B. Not for use in atmospheres immediately dangerous to life or health.
C. Do not exceed maximum use concentrations established by regulatory standards.
D. Air-line respirators can be used only when the respirators are supplied with respirable air meeting the requirements of CGA G-7.1 Grade D or higher
quality.
E. Use only the pressure ranges and hose lengths specified in the User’s Instructions.
J. Failure to properly use and maintain this product could result in injury or death.
M. All approved respirators shall be selected, fitted, used and maintained in accordance with MSHA, OSHA and other applicable regulations.
N. Never substitute, modify, add or omit parts. Use only exact replacement parts in the configuration specified by the manufacturer.
O. Refer to User’s Instructions, and/or maintenance manuals for information on use and maintenance of these respirators.
For technical assistance, call or write:
6
Bullard
1898 Safety Way
Cynthiana, KY 41031-9303
Toll-Free: 800-827-0423
Phone: 859-234-6611
Facsimile: 859-234-6858
www.bullard.com
Spectrum™ Continuous Flow Airline Respirator
Instruction Manual
OPERATIONS
The requirements for Grade D breathable air include:
Limitations of Use
RESPIRATORY
This respirator is NIOSH approved (TC-19C-322, continuous-flow class and
TC-19C-321, pressure-demand class) for Type C applications. It can be worn
for general purposes including: spray painting; tank cleaning; chemical and
pesticide handling; and other industrial or agricultural applications.
This respirator is not approved for use in any atmosphere immediately
dangerous to life or health (IDLH), or from which the wearer cannot escape
without the aid of the respirator. This respirator is not approved for abrasive
blasting.
HEAD
Spectrum respirators DO NOT provide head protection. Wear approved head
protection if head protection is required.
FACE
This respirator’s lens meets ANSI Z87.1-1989 faceshield requirements
for impact, penetration and optics. The lens provides LIMITED FACE
PROTECTION. The Spectrum lens is not shatterproof.
EYES
Spectrum respirators provide LIMITED eye protection from flying particles
and splash per ANSI Z87.1979. Primary eye protection should be worn when
necessary.
Point-ofAttachment
Pressure
Gauge
-Oxygen................................................. 19.5-23.5%
-Hydrocarbons (condensed)
in mg/m3 of gas.......................5 mg/m3 max.
-Carbon monoxide...............................10 ppm max.
-Carbon dioxide..............................1,000 ppm max.
-Odor................................ lack of noticeable odor*
-No toxic contaminants at levels that make air unsafe to breathe.
*Specific measurement of odor in gaseous air is impractical. Air may normally have a slight odor. The presence of a pronounced odor should render
the air unsatisfactory.
Contact the Compressed Gas Association (1725 Jefferson Davis Highway,
Arlington, VA 22202) for complete details on commodity Specifications G-7.1
AIR SOURCE
Locate the source of supplied air, whether it is an air compressor or an
ambient air pump, such as a Bullard Free-Air® pump, in a clean air environment. Locate the air source far enough from your work site to ensure the air
remains contaminant-free. Always use an inlet filter on the air source. Use
suitable after-cooler/dryers with filters, and carbon monoxide monitors and
alarms, like Bullard’s Alert-1™ CO monitor, as necessary to assure clean,
breathable air at all times. The air should be regularly sampled to be sure
that it meets Grade D requirements.
Breathing Air Pressure
Air pressure must be continually monitored
at the point-of-attachment while operating this respirator. A reliable air pressure gauge must be present to permit you to continually monitor pressure
during actual respirator operation.
Respirator
� WARNING
Air Supply
Hose
Figure 2
Grade D
Breathable
Air Source
Breathing Air Requirements
AIR QUALITY
� WARNING
This respirator MUST be supplied with clean, breathable air,
Grade D or better, at all times. This respirator does NOT purify
air or filter out contaminants.
Respirable, breathable air must be supplied to the point-of-attachment of
the approved Bullard air supply hose. The point-of-attachment is the point
at which the air supply hose connects to the air source. A pressure gauge
attached to the air source is used to monitor the pressure of air provided to
the respirator wearer (See Figure 2).
The Breathing Air Pressure Tables (See page 10) define the air pressure
ranges necessary to provide Spectrum-CF respirators with a volume of air
that falls within the required range of 4-15 cfm (Ref. 42 CFR, Part 84, Subpart
J.84.150) when operated in continuous-flow mode. Meets pressure-demand
class requirements when operated within the pressure ranges listed in Table
2 (Ref. 42 CFR 84.157). Make sure you understand the information in the
Breathing Air Pressure Table (see page 10) before using this respirator.
1. Determine the type of air source you are using (column 1), then find the
breathing tube assembly (column 2).
2. Be sure your Bullard air supply hose(s) (column 3) is approved for use
with your breathing tube assembly.
3. Determine that your Bullard air supply hose is within the approved length
(column 4).
4. Make sure you have not exceeded the maximum number of hose sections
(column 5).
5. Set the air pressure at the point-of-attachment within the required pressure range (column 6) for your breathing tube assembly and air supply
hose type and length. Accurate pressure readings can only be attained
when air is flowing into the respirator.
www.bullard.com
Operations
Supplied breathing air must AT a minimum meet the requirements for Type
1 gaseous air described in the Compressed Gas Association Commodity
Specification G-7.1 (Grade D or higher quality), as specified by Federal
Regulations 42 CFR, Part 84.141 (b) and 29 CFR 1910.134 (i).
Failure to supply the minimum required pressure at the pointof-attachment for your hose length and type will reduce airflow and may Cause injury, disease or death.
7
Breathing Air Supply Hoses and Hose Fittings
NIOSH-approved Bullard air supply hose(s) MUST be used between the breathing tube connection fitting on the wearer’s belt and the point-of-attachment to
the air supply (see Figure 1). NIOSH-approved Bullard quick-disconnect fittings MUST be used to connect V20 hose lengths together. When connecting lengths
of V10 hose, only use Bullard V11 hose-to-hose adaptors. Secure connection(s) until wrench-tight and leak-free.
Total connected hose length and number of hoses MUST be within the ranges specified on the Breathing Air Pressure Table (see page 10) and the respirator’s
NIOSH approval labels (see pages 1 and 2). The breathing tube connection fitting MUST be secured to the belt that is supplied with this respirator. Securing
the air entry connection fitting helps prevent the air supply hose from snagging, disconnecting or pulling the respirator off your face.
Breathing Air Pressure Table
Continuous-Flow Class
This table defines the air pressure ranges necessary to provide Spectrum-CF Series respirators with a volume of air that falls within the required range
of 4-15 cfm according to U.S. Government regulations (Ref. 42 CFR 84.150).
1
2
3
4
5
6
Air Source
Breathing Tube
Assembly
Air Supply Hose
Air Supply Hose Length
(Feet)
Maximum Number of
Hose Sections
Required Pressure
Range (psig air)
Stationary or portable
air compressor
BT100, BT101,
BT102, BT108
V10
25-50
100-150
200-300
2
3
5
29-57
33-62
37-65
Bullard Free-Air®
Pumps
BT135
V20
25-50
100-150
250-300
2
3
3
6-18
8-22
10-27
Pressure-Demand Class
This table defines the air pressure ranges necessary to provide Spectrum-CFP Series respirators with a volume of air that meets the pressure-demand
requirements according to U.S. Government regulations (Ref. 42 CFR 84.157).
2
3
4
5
6
Air Source
Breathing Tube
Assembly
Air Supply Hose
Air Supply Hose Length
(Feet)
Maximum Number of
Hose Sections
Required Pressure
Range (psig air)
Bullard Free-Air®
Pumps
BT135
V20
25-50
100-150
250-300
2
3
3
9-10
11-12
13-14
Breathing Air Pressure Tables
1
8
www.bullard.com
Spectrum™ Continuous Flow Airline Respirator
Instruction Manual
SPECTRUM RESPIRATOR ASSEMBLY
Putting Respirator On
11. Adjust the air pressure at the “point-of-attachment” to within the approved
pressure range. See the Breathing Air Pressure Table on page 10 for the
approved pressure range.
1. Remove the Spectrum-CF respirator from the plastic shipping bag. Except
for the belt, the respirator is shipped completely assembled. Lace the belt
through the metal slide or adjustable-flow control fitting that is attached at
the end of the breathing tube. Do not wear this respirator without the belt.
2. Connect the NIOSH-approved Bullard air supply hose to the air source
supplying Grade D breathing air. Clear the air supply hose by moving air
through it for a few minutes before connecting it to the quick-disconnect
fitting at the end of the air entry hose.
3. Fasten the belt at your waist or hip level and adjust for comfort.
12. With the air flowing into your respirator, you are now ready to enter the
work area.
Mask
1. When finished working, leave the work area wearing the respirator and with
the air still flowing.
2. Once completely outside the contaminated area, remove the respirator by
lifting the tabs on the latches. Disconnect the air supply hose using the
quick-disconnect coupling.
The mask cover must be used when operating this respirator.
This part protects the exhalation valve from outside
interference.
Taking Respirator Off
NOTE:
If using V20 Series (1/2” I.D.) air supply hose, the quick-disconnect coupler does not have a shut off valve. Therefore, the air will continue to flow
freely after it has been disconnected from the respirator.
Figure 5
� WARNING
If you cannot OBtain A proper fit, try another mask size and go
through the same steps outlined above. if a proper fit is still
not achieved,
10. With the air flowing, connect the quick-disconnect fitting on the respirator’s
air entry hose to the quick-disconnect coupler on the air supply hose (See
Figure 6). Once fitting is secured, release coupling sleeve to lock fitting
together. Pull on both hoses to make sure they are attached securely.
Figure 6
www.bullard.com
Spectrum Respirator Assembly
4. Release the headstraps to the full outward position by pulling the headstrap
quick release tabs forward. When fully
extended, tabs should be located at the Figure 3
headstrap latches.
5. Grasp the headstrap harness with
thumbs through the straps. Spread
outward.
6. Push the top of the facepiece flange
up the forehead, brushing hair upward
from the face seal area (See Figure 3).
Continue up and over the head until
the harness is centered at the rear
of the head, and the chin rests in the
chin cup.
7. Pull both lower straps at the same time towards the rear (See Figure 4).
Tighten the two temple straps. Tighten the top head strap if necessary.
8. Adjust headstraps until the
facepiece fits securely and evenly.
9. Perform a negative pressure fit
check:
A) With facepiece on and secured,
place your finger over the
quick-disconnect nipple located
at the end of the breathing
tube (See Figure 5).
B) Inhale until the mask collapses
Figure 4
inward slightly (indicating
there is negative pressure). Hold your breath for five seconds.
C) The mask is deemed to be in proper
position if it remained collapsed while the breath was held, and no
inward leakage of outside air was detected.
D) If the mask doesn’t collapse or if an inward leak is detected, re-adjust
mask on face and repeat above steps until the test is passed.
� WARNING
9
CONTAMINANTS.
� WARNING
LEAVE WORK AREA IMMEDIATELY IF:
- Any respirator component becomes damaged.
- Airflow into respirator stops or slows down.
- Air pressure gauge drops below the minimum specified in
Breathing Air Pressure Table (page 10).
- Breathing becomes difficult.
- You become dizzy, nauseous, too hot, too cold or ill.
- You taste, smell or see contaminants inside respirator facepiece.
- Your vision becomes impaired.
� WARNING
IN a contaminated ENVIRONMENT. Respirable contaminants can
remain suspended in air for more than one hour after work
activity ceases, even though you may not see them. Proper work
practice requires you to wear the respirator until you are
outside the contaminated area. If you set the respirator down
in a contaminated environment, contaminants, dirt and dust
could get into the respirator. When you put the respirator
back on, you CoUld breathe in contaminants UPON REUSE.
INSPECTION, CLEANING AND STORAGE
This respirator and all of its component parts and assemblies should be
inspected for damage or excessive wear before and after each use to ensure
proper functioning. Immediately remove the respirator from service and
replace parts or assemblies that show any sign of failure or excessive wear
that might reduce the degree of protection.
Use only complete NIOSH-approved Bullard Spectrum-CF components and
replacement parts on this respirator. Refer to the parts list for correct part
numbers.
The respirator should be cleaned, inspected and sanitized at least weekly or
more often if subjected to heavy use. Respirators used by more than one person must be cleaned, inspected and sanitized after each use. If not cleaned,
contamination may cause illness or disease.
REMEMBER, THE AIR YOU BREATHE WILL NOT BE CLEAN UNLESS THE
RESPIRATOR YOU WEAR IS CLEAN.
Inspection Cleaning and Storage
Mask
10
Immerse the facepiece in warm water (about 120 degrees F) with mild
detergent or a
germicidal disinfecting detergent. The respirator body and parts may be
scrubbed gently with a cloth or soft brush. All foreign matter must be
removed carefully from all surfaces of the exhalation valve flap and seat.
Rinse the respirator in clean, warm water (about 120 degrees F). Shake to
remove excess water, and allow to air-dry away from direct heat, sunlight or
contaminants.
Breathing Tube
Inspection: Inspect the breathing tube for tears, cracks, holes or excessive wear that might reduce the degree of protection originally provided. Be
sure the quick-disconnect fitting is screwed tightly into the breathing tube so
no air can escape.
If any signs of excessive wear are present, replace the breathing tube
assembly immediately or remove the respirator from service.
Cleaning: Hand-sponge breathing tube with warm water and mild detergent, rinse and air-dry. Do not get water inside the breathing tube. After
cleaning, once again carefully inspect breathing tube for signs of damage.
Air Supply Hose
Inspection: The hose(s) should be inspected closely for abrasions, corrosion, cuts, cracks and blistering. Make sure the hose fittings are crimped
tightly to the hose so that air cannot escape. Make sure the hose has not
been kinked or crushed by any equipment that may have rolled over it.
If any of the above signs are present or any other signs of excessive wear
are detected, replace the air supply hose(s) immediately or remove the respirator from service.
Cleaning: The air supply hose(s) should be hand-sponged with warm water
and mild detergent, rinsed and air-dried. Do not get water inside the air supply hose. After cleaning, once again carefully inspect air supply hose(s) for
signs of damage.
� WARNING
Only use hoses that are approved by NIOSH for use with this
respirator. Other hoses could reduce airflow and protection,
and expose the wearer to life-threatening conditions.
Storage
After reusable respirator components have been cleaned and inspected,
place them in a clean plastic bag or an airtight container. Store the respirator and parts where they will be protected from contamination, distortion
and damage from elements such as dust, direct sunlight, heat, extreme cold,
excessive moisture and harmful chemicals. Store the respirator so it is protected from distortion from the weight or pressure of surrounding objects.
Wipe any areas still showing accumulations of foreign matter with a cloth
moistened in a detergent or a solvent such as mineral spirits or naphtha,
until clean.
Stubborn accumulations of paints, lacquers or enamels may be removed
with a cloth containing a paint, enamel or lacquer stripping agent. Once
the dirt or paint is loosened, it may be gently rubbed or brushed off.DO
NOT USE VOLATILE SOLVENTS FOR CLEANING THIS RESPIRATOR OR ANY
PARTS OR ASSEMBLIES. STRONG CLEANING AND DISINFECTING AGENTS,
AND MANY SOLVENTS, CAN DAMAGE THE SILICONE RUBBER AND PLASTIC
PARTS. DO NOT LEAVE SOLVENTS AND STRONG CLEANING AND SANITIZING
AGENTS IN CONTACT WITH SILICONE RUBBER OR PLASTIC SURFACES ANY
LONGER THAN NECESSARY TO LOOSEN THE ACCUMULATIONS OF DIRT OR
www.bullard.com
Spectrum™ Continuous Flow Airline Respirator
Instruction Manual
fit testing
According to Osha’s revised Respiratory Standard, 29 CFR 1910.134, all
tight-fitting facepieces must now be fit tested, regardless of the mode of
operation. This includes all respirator models in the Spectrum Series. Users
must pass either a qualitative or quantitative fit test, and fit testing must
be performed in the negative pressure mode. Bullard’s QNFT45 fit test kit
converts the Spectrum facepiece to the negative pressure mode, and can be
used for either type of fit testing. The instruction sheet that accompanies the
kit provides guidance on its proper use.
The options for qualitative challenge agents include: isoamyl acetate (banana
oil), irritant smoke, saccharin and Bitrex (denatonium benzoate). Quantitative
options include generated aerosol (like the fit test chamber in Bullard’s lab),
ambient aerosol CNC (Portacount method), or controlled negative pressure
(Dynatech Nevada Fit Tester 3000 method). The minimum fit factor for a fullface respirator is 500.
The qualitative fit test protocols that had existed in the substance-specific
standards (except Bitrex, which is new) have been updated in the revised
standard with minor changes and will now be used for all substances. Fit
testing shall be performed prior to initial use, whenever a different respirator is used, and at least annually thereafter. An additional fit test must also
be performed whenever there are changes in the employee’s physical condition that could affect respirator fit, such as dental changes or an obvious
change in body weight.
PARTS AND ACCESSORIES FOR spectrum RESPIRATORs
Cat. No. Description
4. ACCESSORIES AND KITS
1. RESPIRATOR ASSEMBLIES
facepiece component replacement parts
QNFT45Quantitative Fit Test Kit – Includes fir test adapter, facepiece,
sampling adapter, 1/8” barbed mask probe and HEPA filter cartridge
LCKLens Clamp Replacement Kit – Includes upper and lower clamps
with screws and nuts
RLS
Lens replacement
FK-LMedium/large facepiece flange
FK-S
Small facepiece flange
HSKIncludes headstrap with 5 buckles and 5 slides
MCK
Includes mask cover
SEKIncludes speaker diaphragm , outer locking ring, o-ring for speaker
diaphragm, o-ring for speaker/exhalation body and speaker/exhalation body
EVOIncludes exhalation valve seat, o-ring for exhalation valve seat
LNKIncludes nosecup and 2 inhalation valve flaps
Includes headstrap, breathing tube assembly with flow control device and
nylon belt.
For use with breathing air compressors or breathing air cylinders
SPEC40LMedium/Large Spectrum Full-Face Mask with adjustable flow valve.
SPEC40SSmall Spectrum Full-Face Mask with adjustable flow valve.
For use with Bullard Free-Air pumps
SPEC35LMedium/Large Spectrum Full-Face Mask with constant flow valve.
SPEC35SSmall Spectrum Full-Face Mask with constant flow valve.
2. BREATHING TUBE ASSEMBLIES
For use with breathing air compressors or breathing air cylinders
BT100
Breathing tube assembly with 1/4” Industrial Interchange
fitting, flow control device and nylon belt.
BT101
Breathing tube assembly with 1/4” Schrader fitting, flow control
device and nylon belt.
BT102
Breathing tube assembly with 1/4” Snaptite fitting, flow control
device and nylon belt.
For use with Bullard Free-Air pumps
BT135
Breathing tube with 1/2” Industrial
Interchange fitting and nylon belt.
3. BREATHING AIR SUPPLY HOSES
V10 Series Starter Hose Kit.
Includes V1025ST hose with 1/4” Industrial Interchange quick-disconnect coupler and V13 adaptor fitting (3/8” hose to 3/8” pipe). 4696 25 ft. Industrial
Interchange starter hose kit.
V10 Series EXTENSION Hose KitS.
Includes V11 hose-to-hose adaptor fitting and V13 hose-to-pipe fitting (3/8”
hose to 3/8” pipe).
5454
25 ft. V1025EXT extension hose kit
5457
50 ft. V1050EXT extension hose kit
5458
100 ft. V10100EXT extension hose kit
Accessories
AFW45 Anti-Fog Lens Wipes (10/pkg.)
SPEC-PL Spectacle Frame Assembly
SPEC-LC Mylar Lens Covers (25/pkg.)
FTR45Fit Test Refill - Includes 15 feet of Tygon tubing, 25 suction cups
and 25 clips
HFC45HEPA cartridges for quantitative fit testing (6/carton)
Cat. No. Description
For use with Bullard Free-Air pumps
V20 SERIES STARTER/EXTENSION HOSE KITS
Includes 1/2” Industrial Interchange Male/Female quick-disconnect fitting.
V2050ST 50 ft. 1/2” I.D. hose
V20100ST 100 ft. 1/2” I.D. hose
www.bullard.com
Fit Testing / Parts and Accessories
For use with breathing air compressors or breathing air cylinders
Facepiece Component Replacement Packages
6040O-ring, Speaker Diaphragm (5/pkg.)
6041O-ring, Speaker Exhalation (5/pkg.)
6042O-ring, Exhalation Valve (5/pkg.)
6043
Inhalation Valve Flap (4/pkg.)
6059
Exhalation Valve, CF (5/pkg.)
6060
Breathing Tube Adaptor (5/pkg.)
11
RETURN AUTHORIZATION
spectrum AIRLINE RESPIRATOR
IMPORTANT: THE FOLLOWING STEPS MUST BE COMPLETED BEFORE
BULLARD WILL ACCEPT ANY RETURNED GOODS. PLEASE READ CAREFULLY.
15
Follow the steps outlined below to return goods to Bullard for repair or
replacement under warranty or for paid repairs:
14
13
12
1
11
1. C ontact a Bullard Customer Service Coordinator by telephone or in writing
at:
10
9
8
Bullard
1898 Safety Way
Cynthiana, KY 41031-9303
Toll-Free: 800-827-0423
Phone: 859-234-6611
Fax: 859-234-1303
2
4
7
6
3
5
In your correspondence or conversation with a Customer Service
Coordinator, describe the problem as completely as possible. For your
convenience, your Coordinator will try to help you correct the problem
over the telephone.
19
18
2.Verify with your Coordinator that the product should be returned to
Bullard. Customer Service will provide you with written permission and a
return authorization number as well as labels you will need to return the
product.
20. Mylar lens cover
17
16. Constant-flow valve
for use with all high
pressure models
16. Constant-flow valve for use with
low pressure model
Cat. No. Description
1 . LCK Lens Clamp Replacement
2. RLS Replacement Lens
3. FK-L & FK-S Facepiece Flange
4. HSKHeadstrap with Buckles and Slides
5. MCK
Mask Cover
6. SEK
Speaker Diaphragm
7. SEKOuter Locking Ring
8. SEKO-ring Speaker Diaphragm*
9. SEKO-ring Speaker/Exhalation*
10. SEK
Speaker/Exhalation Body
11. 6059
Exhalation Valve*
12.EVO O-ring for Exhalation Valve Seat*
13. EVO
Exhalation Valve Seat
14. LNK
Nosecup
15. LNK
Inhalation Valve Flaps*
16. BT100 Series Flow Control Valve and Belt
17. BT100 Series Hose Clamp
18. BT100 Series Breathing Tube
19. BT100 Series Breathing Tube Adaptor*
20. SPEC-LC
Mylar Lens Covers (25/pkg.)
*Packaged as replacement parts in quantities of
4 or 5 (See page 16).
3.Before returning the product, decontaminate and clean it to remove any
hazardous materials which may have settled on the product during use.
Laws and/or shippingregulations prohibit the shipment of hazardous or
contaminated materials.
Products suspected to be contaminated will be professionally discarded
at the customer’s expense.
4. Ship returned products, including those under warranty, with all transportation charges pre-paid. Bullard cannot accept returned goods on a freightcollect basis.
5.Returned products will be inspected upon return to the Bullard facility.
Your Customer Service Coordinator will telephone you with a quote for
required repair work which is not covered by warranty. If the cost of
repairs exceeds stated quote by more than 20%, your Coordinator will
call you for authorization to complete repairs. After repairs are completed
and the goods have been returned to you, Bullard will invoice you for actual
work performed.
ISO 9001
certified
Americas:
E.D. Bullard Company
1898 Safety Way
Cynthiana, KY 41031-9303
Toll free: 877-BULLARD (285-5273)
Tel: 859-234-6616
Fax: 859-234-8987
www.bullard.com
Europe:
Bullard GmbH
Lilienthalstrasse 12
53424 Remagen • Germany
Tel: +49-2642 999980
Fax : +49-2642 9999829
www.bullardextrem.com
Asia-Pacific:
Bullard Asia Pacific Pte. Ltd.
LHK Building
701, Sims Drive, #04-03
Singapore 387383
Tel: +65-6745-0556
Fax: +65-6745-5176
www.bullard.com
©2011 Bullard. All rights reserved.
Free-Air is a registered trademark of Bullard.
Spectrum is a trademark of Bullard.
6081986062B (1111)