Download Fieldpiece - Allied Supply

Transcript
Fieldpiece
Quick Start
1 Power on your SMAN by holding the ON/OFF
button for 1 second.
2 Select desired units (English or Metric) by
pressing the UNITS button (SMAN3) or holding the ENTER button for 1 second (SMAN2).
3Calibrate temperature and/or pressure as
needed. See calibration sections for more
details.
4 Connect to the system.
5 Press the circular button to select between
Actual Superheat and Subcooling, Target
Superheat, Direct Temperature Display (T1 T2
Direct) or Saturation Temperature Display..
Digital Manifold &
Vacuum Gauge
OPERATOR'S MANUAL
Models SMAN2 SMAN3
SMAN3
Psig
Kpa
inHg
cmHg
Psig
Kpa
inHg
cmHg
°F°C
SHT1
VSAT
IDWB
°F°C
RMicrons
Target SH
StableSet
Alarm
HiLo
°F°C
SCT2
LSAT
ODDB
HH:MM:SS
AUTO
OFF
Superheat
Subcooling
Target
Superheat
r
Saturation
CAL
CAL
Atmospheric
Pressure
Test
Pressure
APO
APO
APO
APO
AUTO
OFF
AUTO
OFF
AUTO
OFF
AUTO
OFF
Psig
Kpa
inHg
cmHg
°F°C
SHT1
VSAT
IDWB
°F°C
RMicrons
Target SH
01
Stable Set
Alarm
HiLo
°F°C
SCT2
LSAT
ODDB
HH:MM:SS
Controls
AUTO
OFF
5
1
6
2
10
Target
Superheat
Ente
CAL
12 Units
13
HIGH
2
CAL
3
4
Atmospheric
Pressure
Test
Pressure
1 Insert K-type thermocouple plugs here.
2 Temperature calibration pots.
3 Press to zero atmospheric pressure.
4 Press to linear adjust pressure curve. (See Advanced Pressure Calibration section.)
5 Press to engage backlight. Hold when powering on to toggle Auto Power Off.
6 Press/hold to cycle through refrigerants.
7 Hold 1 second to toggle power on/off.
8 Press up or down arrow to adjust values.
9Press to confirm selection. For the
SMAN2, hold 1 second to adjust units.
10 Press to turn on/off the Hi/Lo vacuum alarms
and hold to change alarm settings.
11 Press to toggle through temperature tests.
12 Press to adjust units. (SMAN3)
13 Turn clockwise to close Low side port.
14 Turn clockwise to close High side port.
06
Target Superheat
04
AUTO
OFF
AUTO
OFF
Alarm
Alarm
Target Superheat is only used for charging fixed orifice air
conditioning systems. Your SMAN allows you to input indoor
wet bulb (IDWB) and outdoor dry bulb (ODDB) temperatures
manually, in real-time using K-type thermocouples, or a
combination of both to calculate the Target Superheat. Your SMAN
will detect if a thermocouple is plugged in or not. Unplugging the
thermocouple lead(s), at any time, sends the SMAN into manual
input mode and plugging them in, at any time, automatically
sends the SMAN to input directly from the thermocouples.
Saturation
Shows vapor (VSAT) and liquid (LSAT) saturation temperatures
calculated from the pressures measured and the refrigerant selected.
07
CAL
CAL
CAL
Test
Pressure
Test
Pressure
Test
Pressure
Test
Pressure
AUTO
OFF
CAL
1 Press circular button to cycle through temperature modes until SH
and SC are shown. The SMAN can show superheat and subcooling
simultaneously.
2 Choose the appropriate refrigerant using the REFRIGERANT
button.
3 Connect the K-type thermocouple and EPA approved refrigerant
hoses to your SMAN.
4 Connect your SMAN to the system:
Superheat: Hand tighten low side hose to suction line service
port. Place the thermocouple on the suction line between the
evaporator and compressor, no closer than 6 inches to compressor.
Subcooling: Hand tighten high side hose to liquid line service
port. Attach the thermocouple on the liquid line between the
condenser and expansion valve (TXV), as close to the service port
as possible.
5 After turning the system on or making any adjustments to
the system wait 15 minutes before charging by superheat or
subcooling to ensure that the system is stable and running
normally. Add or remove refrigerant using the low and high side
valves as needed. Let system stabilize again for 15 minutes.
HIGH
CAL
CAL
Superheat and Subcooling
Saturation
LOW
1 Press circular button until T1 and T2 are shown.
2 Plug in any K-type thermocouple into T1 and/or T2.
3 Read the temperature directly from the display.
1 Connect your SMAN to the system using EPA approved 1/4in
refrigerant hoses.
2 Read pressures directly from the upper display.
14
CAL
T1
T1
T1 T2 Direct
Standard Pressure
1
7
11
T1 T2 Direct
LOW
03
08
Input temperatures with K-type thermocouples
1 Press circular button until Target SH is shown in the bottom,
center of display.
2 Plug ATWB1 Wet bulb K-type thermocouple into T1 and ATA1 Dry
bulb K-type thermocouple into T2.
3 Press ENTER. The IDWB will display the real-time temperature
measured from T1. Use the ARROW button to select the
temperature you want to input first, either IDWB or ODDB.
4 For IDWB input, wet the sock of the ATWB1. Clip it to the return
of the evaporator between the filter and coil. When the wet bulb
temperature reaches its lowest point and stabilizes, press ENTER
to lock the reading. If the inputted temperature is below the wet
bulb range, “-OL” will blink. If the inputted temperature is above
the wet bulb range, “OL” will blink. Retake the measurement if
“OL” or “-OL” is displayed.
AUTO
OFF
CAL
CAL
CAL
Atmospheric
Pressure
Atmospheric
Pressure
Atmospheric
Pressure
Atmospheric
Pressure
CAL
CAL
CAL
CAL
AUTO
OFF
SMAN3
SMAN3
CAL
CAL
CAL
SMAN3
SMAN3
CAL
AUTO
OFF
AUTO
OFF
AUTO
OFF
APO
CAL
CAL
CAL
SMAN3
APO
CAL
CAL
CAL
CAL
Atmospheric
Pressure
Atmospheric
Pressure
Atmospheric
Pressure
Atmospheric
Pressure CAL
T1
APO
AUTO
OFF
CAL
CAL
CAL
APO
APO
CAL
CALAPO
CAL
APO
CAL
CAL
Superheat
Subcooling
APO
APO
T1 T2 Direct
APO
CAL
CAL
CAL
Test
Pressure
Test
Pressure
Test
Pressure
Test
Pressure
CAL
CAL
APO
Saturation
05
T2
CAL
Target
Superheat
APO
AUTO
OFF
SMAN3
APO
CAL
AUTO
OFF
AUTO
OFF
CAL
SMAN3
SMAN3
SMAN3
APO
APO
APO
APO
AUTO
OFF
CAL
AUTO
OFF
Functions
T2
r9
Superheat
Subcooling
Alarm
02
ON/OFF
Refrigerant
8
CAL
RoHS Compliant
APO
Digital Manifold + Vacuum Gauge
T1
CAL
CAL
CAL
CAL
APO SMAN3
CAL
SMAN3
CAL
CAL
SMAN3
Psig
Kpa
inHg
cmHg
SMAN3
APO
APO
AUTO
OFF
CE
SMAN3
CAL
AUTO
OFF
APO
SMAN3
APO
APO
APO
SMAN3
SMAN3
SMAN3
SMAN3
AUTO
OFF
AUTO
OFF
AUTO
OFF
AUTO
OFF
CAL
CAL
CAL
C-Tick (N22675)
SMAN3
SMAN3
AUTO
OFF
AUTO
OFF
AUTO
OFF
SMAN3
SMAN3
SMAN3
SMAN3
SMAN3
SMAN3
CAL
Units
T1 T2 Direct
Psig
Your SMAN is the portable digital refrigerant
Kpa
Psig
Psig
manifold and vacuum gauge (SMAN3) for A/C
inHg
Kpa
Kpa
and refrigeration service. Your SMAN combines
cmHg
inHg
inHg
high pressure manifold gauges, a superheat/
cmHg
cmHg
Psig Psig 2°F°C
subcooling calculator, micron gauge for vacuum
Pressure
(pounds/in
) SMAN3
SMAN3
Psig
Psig SHT1
Kpa
Kpa
2
Psig
Psig
°F°C (Kilopascals)
Psig Psig °F°C
Kpa
Kpa
(SMAN3), and dual temperature measurements
Pressure (pounds/in
)
Pressure
inHg
inHg
Psig Psig
Kpa
Kpa
Psig cmHg
Psig
Psig
Psig SHT1
Kpa
Kpa
inHg
inHg
SCT2
cmHg
Kpa
Kpa VSAT
Psig
Psig
inHg
inHg
Psig
Psig
Kpa
Kpa
Pressure
(Kilopascals)
Negative
Pressure
(inches
of mercury)
into one easy-to-use instrument. Your SMAN
inHg
inHg
cmHg
cmHg
Psig
Psig
Kpa
Kpa
inHg
inHgIDWB
Kpa
Kpa VSAT
cmHg
cmHg
inHg
inHg
°F°C
°F°C
cmHg
cmHg
Kpa
Kpa
Kpa
KpaLSAT
cmHg NegativePressure
cmHg
inHg
inHg
calculates and displays target superheat and
Negative
Pressure
(inches
of
mercury)
(cm
of
mercury)
°F°C
°F°C
inHg
inHg
cmHg
cmHg
SCT2
SHT1
inHg
inHgIDWB
ODDB
°F°C
°F°C °F°C
Stable
Set
cmHg
cmHg
°F°C
°F°C
SCT2
SHT1
inHg
inHg
LSAT
VSAT
°F°C
°F°C
cmHg NegativePressure
cmHg
SCT2
SHT1
cmHgIDWB
cmHg R(cm of mercury)
actual superheat to ensure proper charging of a
°F°C
°F°C
SCT2
SHT1
HH:MM:SS
LSAT
VSAT
ODDB
Alarm
SCT2Microns
SHT1
°F°C
°F°C
°F°C
Stable
Set
SMAN3
LSAT
VSAT
cmHg
cmHg
SCT2
SHT1
LSAT
VSAT
ODDB
IDWB
°F°C
°F°C
HH:MM:SS
RAPO
LSAT
VSAT
SCT2
SHT1
unit. Your SMAN can also display the liquid and
°F°C
StableSet
ODDBTarget SH
IDWB HiLo
LSAT
VSAT
ODDB
IDWB
RHH:MM:SS
SCT2Microns
SHT1 Alarm
°F°C Microns
ODDB °F°C
IDWB
°F°C
Stable
Set
LSAT
VSAT
Alarm
HH:MM:SS
ODDB
IDWB
R-°F°C
HH:MM:SS
°F°C
Stable
Set
LSAT
VSAT HiLo
vapor saturation temperatures inStable
Saturation
°F°C
°F°C
Set
Alarm
HH:MM:SS
Microns
ODDB
IDWB
RHiLo
Target
SH
Target
SH
Psig
Psig
°F°C
Stable
Set
SCT2
SHT1
RHH:MM:SS
Alarm
Microns
ODDB
IDWB
R°F°C
Stable
Set
Alarm
SMAN3
Microns
HiLo
Target
HH:MM:SS
Psig
Psig Psig
PsigSH
Psig Manifold
Psig
Psig Psig
RAlarm
Microns
SCT2 Kpa Digital
SHT1
°F°C
Stable
Set
+ Vacuum Ga
HiLo
Target
mode to help you trouble shoot your
system.
Alarm
HH:MM:SS
Microns
PsigSH RPsig Kpa
HiLo
Psig
PsigSH
Target
LSAT
VSAT
°F°C
Kpa
Kpa
Stable
Set
Digital
Manifold
+ Vacuum Gauge
Kpa
Kpa
Kpa
Kpa
Kpa
Kpa
HiLo
Target
SH
Alarm
Microns
Psig
Psig + Vacuum
HiLo
Target
PsigSH RPsig
Kpa
Kpa
Kpa
Kpa
T1
LSAT
VSAT
Digital
Manifold
Gauge
Microns
T1
T2
inHg
inHg
inHg
Digital
Manifold
+
Vacuum
Gauge
inHg
inHg
inHg
inHg
inHg
HiLo
Target
SH
ODDB
IDWB
inHg
inHg
Your SMAN is designed to meet the Alarm
demands
Digital Manifold
KpaT1
Kpa + Vacuum Gauge
Kpa SH
Kpa cmHg
inHg
inHg
Digital Manifold
+ Vacuum Gauge
inHg
inHg
T2
HiLo
Psig
Psig
Target
cmHg
cmHg
cmHg
cmHg
cmHg
Digital cmHg
Manifold
ODDBGauge
IDWB
T1
T2
T1
T2cmHg HH:MM:SS
inHg
inHg + Vacuum
Manifold
inHg + Vacuum Gauge
inHg T2
cmHg
cmHg
cmHg
cmHg
cmHg
of the HVAC/R technician with T1
aT1 ruggedized Digital
T1
T2
Psig
Psig
Digital Manifold + Vacuum Gauge Stable Set cmHg
°F°C
ON/OFF
Kpa
Kpa
cmHg
cmHg
Refrigerant
cmHg + Vacuum Gauge
cmHg T2
°F°C
°F°C
HH:MM:SS
°F°C
°F°C°F°C
°F°C °F°C°F°C
Digital
Manifold
RAPO
Kpa
Kpa
ON/OFF
T1
T2
°F°C
Stable
Set
°F°C
°F°C
Refrigerant
°F°C
°F°C
SCT2
SHT1
SCT2
SHT1
SCT2
SHT1
Alarm
SCT2
SHT1
Microns
ON/OFF
rubber boot for durability, a strong
metal
Superheat
T1
Direct
Subcooling
T2
Direct
Refrigerant
T1
T2
RAPO
inHg
inHg
Refrigerant
ON/OFF
inHg
inHg
°F°C
°F°C
°F°C
°F°C °F°C
SCT2
SHT1
Refrigerant
°F°C
SCT2
SHT1
Psig
ON/OFF
LSAT
VSAT
LSAT
VSAT
Alarm
LSAT
VSAT
Microns
LSAT Psig
VSAT
Refrigerant
ON/OFF
ON/OFF
HiLo
cmHg
cmHg
SCT2
Target
SH
SHT1
Refrigerant
Kpa
KpaEnter
SCT2
SHT1
VSAT
LSAT
LSAT
VSAT
ON/OFF
cmHg
cmHg
hanger for easy storage and a high quality,
Vapor
Saturation
Temperature
Liquid
Saturation
Temperature
Refrigerant
ODDB
IDWB
ODDB
IDWB
ODDB
IDWB
ODDB
IDWB
Superheat
Target SCT2
SHT1
Refrigerant
inHg ODDB
inHgEnterr
HiLo
Target
SH
ON/OFF
LSAT
VSAT
LSAT IDWB
VSAT
ODDB
IDWB
Subcooling
Superheat
Refrigerant
HH:MM:SS
HH:MM:SS
Superheat
Target
HH:MM:SS
r °F°C
cmHg
cmHgEnteHH:MM:SS
°F°C
°F°C Outdoor
°F°C
StableSet blue
°F°C
StableWet
SetEnteBulb
Psig
°F°C
Psig
StableRSet
Stable
Set
ODDB
IDWB
IDWB
ODDBVSAT
Superheat
Target
Subcooling
Superheat
HH:MM:SS
water resistant nylon pouch. A bright
Indoor
Dry
RHH:MM:SS
nterBulb
LSAT
RRE
r
Superheat
Target
°F°C
Stable
Set
A
°F°C
te
Stable
Set
la
SHT1 Alarm
Psig
Psig
rmKpa
Superheat
Alarm
Microns Alarm
Alarm Ente
Microns
Superheat
Target
Microns
Kpa
Units
R- Target
°F°CSubcooling
°F°C
R- Psig
°F°C
°F°C
HH:MM:SS
r Microns
CAL
Subcooling
Superheat
HH:MM:SS
Digital
+SCT2
Vacuum
Gauge
Superheat
A
°F°C
Stable
Set
Subcooling LSAT
Superheat
En rTarget
°F°C SH HiLo
Stable
Set
larmManifold
Psig
Alarm
Microns
Psig
Psig
Superheat
Target
Psig
Psig
Alarm
Microns
ODDB
IDWB
HiLo
Kpa
Kpa SCT2
VSAT
Units
Target
HiLo
HiLo
SHT1
Target
SH
Target SH
RSH
Superheat
Target
Subcooling
Superheat
Rte
inHg
inHg
Psig
Psig
backlight and very large, easily viewable
display
Vacuum
Stopwatch
A
r
n
la
Digital
Manifold
+
Vacuum
Gauge
E
e
its
rm
t
Alarm
Alarm
UnSubcooling
Microns
s
SCT2
Alarm
SHT1
HiLo
Microns
n
Target
SH
Superhe
it
Kpa
Kpa
HiLo
Subcooling
Superheat
Target
SH
Kpa
Kpa
n
Kpa
Kpa
Superheat
Target
LSAT
VSAT
T1
T2
E
U
ODDB
IDWB
A
inHg
inHg
la
s
CAL
CAL
rm
cmHg
cmHg
nit
Kpa
T1 T2 Direct Kpa
Saturation
U
A
HH:MM:SS
Subcooling
Superheat
la
s
HiLo
Superheat
Target
it
Target
SH
ODDB
IDWB
rm
HiLo
Target
SH
Un
inHg
inHg
inHg
inHg
inHg
inHg
T2
HH:MM:SS
lets you see your measurements in any
Hours:Minutes
or
Minutes:Seconds
°F°C
Stable
Set
Digital Manifold
+ Vacuum
Gauge
Awork
cmHg
cmHg
+ Vacuum
Gauge
T2 Direct
Saturation
DigitalT1Manifold
+ Vacuum
Gauge
Digital
Manifold
+ Manifold
Vacuum
Gauge
larm T1
its Digital
inHg
inHg
LSAT
VSAT
°F°C
Stable
Set
n
HH:MM:SS
U
Psig
Psig
Subcooling
Superheat
A
°F°C
Stable
Set
RT1 T2 DirectR-Psig
Saturation
APO
Alarm
RDigital
Manifold
+ Vacuum
Gauge
larm
Manifold
Gauge
cmHg
cmHg
cmHg
cmHg
cmHg
cmHg
°F°C
°F°C
Psig + Vacuum
T1 T2 Direct
Saturation
T1
T2
T1
T2 Saturation
T1
T2
Units Digital
T1
PsigAlarm
Psig
Alarm
Microns
cmHg
T1 T2T2
Direct cmHg
Alarm °F°C
Microns
Kpa
Kpa
Microns
Manifold
Gauge
Manifold
+ Vacuum
Gauge SHT1 Digital
ODDB
IDWBTarget
T1
T2Kpa + Vacuum
°F°C
environment. SMAN has a very intuitive
user Digital
T1
T1 T2 Direct
Saturation
SCT2
SH Kpa T2 °F°C
HiLo
SH HiLo
Kpa
Kpa Target
A
inHg
T1 T2 DirectinHg
Saturation
ON/OFF
°F°C
la
s
°F°C
°F°C
°F°C
°F°C
T1
T2
T1
T2
SCT2
SHT1
it
r
m
n
Psig
Psig
LSAT
Refrigerant
°F°C SH
°F°C U
Target
inHg
inHg
T1 T2 DirectinHg HiLo
Saturation ON/OFFVSAT
HH:MM:SS
inHg
ON/OFF
ON/OFF Kpa
SCT2
SHT1
cmHg
cmHg
ON/OFF
SCT2
ON/OFF
SHT1IDWBSet
interface and extremely long battery life.
SCT2
SHT1
Refrigerant
Refrigerant
LSAT
VSAT
Digital
Manifold +
Vacuum
Refrigerant
°F°C
Stable
Refrigerant
ODDB
Kpa
SCT2 Gauge
SHT1
Refrigerant
cmHg
cmHg
ON/OFF
ON/OFF
Digital
Manifold
+ VSAT
Vacuum
Gauge
cmHg
cmHg
T1 T2
Direct
Saturation
Refrigerant
RAPO
Refrigerant
LSAT
LSAT
VSAT
LSAT
VSAT
ODDB
IDWB
HH:MM:SS
inHg
inHg
LSAT
VSAT
ON/OFF
ON/OFF
°F°C
°F°C
°F°C
Stable
Set
AlarmHas
T1 IDWB
T2
Microns
Refrigerant
Your SMAN comes pre-programmed with
Micron
Reading
Stabilized
Refrigerant
ODDB
IDWB
ODDB
IDWB
ODDB
RHH:MM:SS
°F°C
°F°C
cmHg
cmHg
ODDB
IDWB
°F°C
°F°C
°F°C
Stable
Set
Digital
Manifold
+
Vacuum
Gauge
r
r
r
T1 T2
Direct
Saturation
SCT2
SHT1
r
Alarm
Microns
nte
ON/OFF
E
REnteSH
HH:MM:SS
Ente
HH:MM:SS
EnteTarget
SCT2 HH:MM:SS
SHT1
HiLo
r SH
°F°C
rRefrigerant
°F°C
Stable
SetHiLo Set
°F°C
Stable
Set
Superheat
Target
HH:MM:SS
Superheat
Target
SCT2
Superheat
Target
SHT1
nter
Alarm
Superheat
Target
Microns
nte
the most accurate P-T charts for 39 of the most Subcooling
Mode
°F°C
Alarm
Mode
Target
LSAT
E
VSAT
EStable
R-ON/OFF
Ente r
Stable
Set
R-Set
RCAL
CAL
Superheat
T1
T2
r
Subcooling
Superheat
Subcooling
Superheat
°F°C
°F°CCAL
Subcooling
Superheat
Superheat
Target
SuperheatVSAT
Target
RrAlarm
e
Microns
te EntLSAT
Alarm
Microns
Alarm
Superheat
Target
Microns
te
Refrigerant
n
HiLo
Target
SH
n
LSAT
VSAT
E
IDWB
ODDB
E
Alarm
Microns
Superheat
Subcooling
Superheat
CALSubcooling
CAL
r
Superheat
Target
SCT2
SHT1 its ODDB
Superheat
IDWB
Target
SH
Alaare
HiLo
Ente
HiLo
Subcooling
Superheat
Target
SH
AManifold
Target
SH
A
common refrigerants in the field so
you
High
Alarm
Low
Alarm
Battery
Life
ATarget
sHH:MM:SS
Superheat
Target
la
larm
larm
ODDB
IDWB
rm
its
rm
nitHiLo
its
n
n
Subcooling
Superheat
U
n
Digital
+
Vacuum
Gauge
HiLo
U
Subcooling
Superheat
U
Target
SH
U
°F°C
StableSet
Alarm
Alarm
HH:MM:SS
nits
Manifold
+s Gauge
Vacuum Gauge LSAT
Subcooling
Superheat
CAL
nit
U
°F°C Digital
Digital Manifold
+ VSAT
Vacuum
StableSet
T1Microns
T2UCAL
°F°C
Stable
Set
Alarm nitsHH:MM:SS
Alarm
Alarm
R- Alarm + Vacuum
ODDB
IDWB
nteitrs
always prepared for any job.
RVacuum
(Microns
ofGauge
Mercury)
Saturation
Auto
Power
OffUEnabled
EUnGauge
AGauge
RDigital
Manifold
U
Digital
Manifold
+
Vacuum
laManifold
Digital
Manifold
+T1
Vacuum
Units
Alarm
Microns
T2
rSaturation
m
nits
Superheat
TargetGauge
ON/OFF
T1 T2 Direct
Saturation
+
Vacuum
T1 T2
Direct
Saturation
Alarm
T1 T2 Direct
Microns
T1 T2 Direct Digital
T1
T2
HiLo
Target
SH
Refrigerant
A rSaturation
Atmospheric
Subcooling
Superheat
T1
T2 HH:MM:SS
T1T1T1 T2 Direct
T2T2°F°C
s T2Selected
SetSH Saturation
HiLo
Target
it
T1
T2Stable
Direct
mT1 Superheat
n
HiLo
Target SHlaTarget
U
R
Refrigerant
of System
ON/OFF Saturation
Pressure
P
T1 T2 Direct
Alarm
Microns s
T1 T2 Direct
Saturation
Ala Refrigerant
CAL
Ente
Alarm
SMAN3
SMAN3
ON/OFF
Refrigerant
CAL
T2
Display
SMAN3
Certifications
APO
Digital Manifold + Vacuum Gauge
T1
Description
AUTO
rm
OFF
Unit ON/OFF
AUTO
Digital
Manifold + Vacuum Gauge
Test
CAL
CAL CALAtmospheric
CAL OFF Manifold
CAL Target
CAL
CAL
OFF
HiLo +CAL
SH
VacuumRefrigerant
Gauge
ON/OFF
ON/OFF
ON/OFF
Digital
Manifold + Vacuum Gauge
PressureDigital
Pressure
CAL
CALRefrigerant
T1 T2
Direct
Saturation
CAL
CAL
T2
CAL
CAL
Refrigerant
Refrigerant
ON/OFF
Refrigerant
ON/OFF
T1
T2
r
r
CAL T1 T2 Direct
CAL Refrigerant
T2
CAL
CAL
e
T1 T2 Direct
nte Saturation
t
ESaturation
n
CAL
CAL
E EGauge
Digital
Manifold + Vacuum
r
CAL
SuperheatAUTO Atmospheric Target
AtmosphericCAL
Test
Test
Atmospheric
Test
Atmospheric
Test
Atmospheric
nteTest
Superheat
Target
CAL
CAL
Pressure
Pressure
PressureT1 Superheat rPressure
Subcooling
Pressure
Pressure
Pressure OFF
Pressure
Atmospheric
Test
ON/OFF
r Pressure
rSuperheat Pressure
T2
Atmospheric
Test
Target
Refrigerant
nEtente
Ente
ERefrigerant
Pressure
PressureCAL
ON/OFF
CAL
CAL
CAL TargetnSuperheat
Pressure
Pressure
CAL
CAL
ter
Subcooling
Superheat
Subcooling
ON/OFF
Atmospheric
Test
Superheat
E
Superheat
Target
Superheat
Target
Atmospheric
Test
CAL
CAL
CAL
CAL
Alarm
Refrigerant
s Pressure Superheat
AUTO CAL
Superheat Subcooling
Target
CAL
Pressure
Pressure
Pressure
Unit
Subcooling
Superheat
Subcooling
Superheat
OFF
Subcooling
Superheat
A
la
s
it
rm
r
n
UON/OFF
CAL Refrigerant
AlaCAL
Alarm
Arm
Ala
Ente
larm
s
rm CAL AEla
itsnits
nter
nU
UTarget
UnUitnsitSuperheat
r
CAL
rmr Atmospheric
Test
Superheat
Target
T1 T2 Direct
Saturation
EnteUnits
CAL
CAL
Ente
Pressure
Superheat
Target Pressure
CAL
Subcooling
Superheat
AUTO
AUTO
AUTO
OFF OFF
Pulling a Vacuum (model SMAN3)
5 For ODDB input, clip the ATA1 thermocouple to the side of the
condenser. Once reading is stable, press ENTER. If the inputted
Follow all manufacturers’ evacuation proceduresCAL
over those in
temperatureTarget
is below the dry bulb range,
“-OL” will Subcooling
blink.
IfSuperheat
the
Superheat
T1 T2 Direct
Saturation
Superheat
Subcooling
this
manual.
Subcooling
Superheat
Atmospheric
Test
r
T1 T2
Direct
Saturation
T1
T2
Direct
Saturation
T1
T2
Direct
Saturation
te
n
s dryT1bulb
inputted temperature
is Aabove
range, “OL”
will
blink.
E Pressure
T2 Direct
Saturation
Unitthe
Pressure
Superheat
Target
larm
Atmospheric
Test
Units
T2laDirect
Saturation
Note:
larger systems
may take
much
s A
nitT1
Subcooling
Superheat
its longer to reach a deep vacuum
rmis displayed. For accurate
Pressure ifU“OL”
Pressure
Un
Re-take the measurement
or “-OL”
and
a
different
evacuation
method
or equipment might be preferred.
Alarm
T1 T2 Direct
Saturation
s
it
n
U
T1 T2 Direct
Saturation
the ATA1 shaded from direct
sunlight.
T1 T2 Directresults, keep
Saturation
1
Connect
your
SMAN3
in
between
your vacuum pump and the
6 Once both IDWB and ODDB are
CALlocked in, the Target Superheat
CAL Saturation
T1 T2 Direct
Saturation
T1 T2 Direct
system,
then
power
on
your
SMAN3.
will show in the lower left corner of the display. If the Target
2 Draw a vacuum on the system. The SMAN will automatically
Superheat calculated from the IDWB and ODDB temperatures is
Atmospheric
Test
sense the negative
CAL
CAL pressure and display the vacuum in microns.
outside the range in whichPressure
air conditioning is recommended then
Pressure
Indicators
for
increasing
or decreasing pressure will be shown.
“OL” is displayed. Usually, raising the indoor temperature will
3
Press
Alarm
button
to
toggle
between the High (Hi) alarm, the
bring the conditions within manufacturer’s recommendations.
Atmospheric
Test
Low
(Lo)
alarm
and
no
alarm.
No Display of Lo or Hi means alarm
Pressure
Pressure
If your target superheat is out of the recommended range, press
is
off.
Note:
The
stopwatch
will
start when an alarm is selected.
ENTER to re-take both your measurements.
Pressing
Alarm
will
reset
the
stopwatch.
7 To re-take either IDWB or ODDB, press the ARROW button until
4 When the rate of change in pressure is less than 1 micron per 10
the label (IDWB or ODDB) you wish to re-take is blinking. Press
seconds, Stable will appear in the lower left corner of the display.
ENTER. Re-do the measurement and press ENTER to lock it in. The
Additional
Tips:
newly calculated target superheat will be displayed.
To
achieve
a deep vacuum as quickly as possible:
8 To clear both IDWB and ODDB simultaneously, press ENTER.
1
Use
the
shortest
hoses with the largest diameter available.
Input temperatures manually
2
Remove
Schrader
cores and core depressors. Core removal tools
1 Press circular button until Target SH is shown at the bottom,
like
the
"MegaFlow
Valve Core Removal Tool" can be purchased to
center of display. Unplug thermocouples from T1 and/or T2.
help
with
this
process.
2 Press ENTER. Far left digit of IDWB or ODDB will begin blinking
3 Inspect the rubber seals at both ends of your hoses for damage
indicating manual input mode is ready. Use the ARROW button
that may result in leakage.
to change values and press ENTER to lock in each digit. Once the
4
Do not use hoses with low loss fittings when evacuating or
last digit for ODDB or IDWB is locked in, the calculated target
pulling a vacuum on a system.
superheat will show in the lower left corner of the display.
Note: When pressing ENTER to lock in the last digit for IDWB or
ODDB, if the inputted temperature is below the wet bulb or dry
1 Hold the ALARM button for one second to enter Alarm Set mode.
bulb range, “-OL” will flash once. If the inputted temperature is
2 Press the ALARM button to toggle between Alarm Hi and Lo.
above the wet bulb or dry bulb range, “OL” will flash once. In both
3 Use ARROW button to select the value and press ENTER to lock in
instances you must re-input the measurement.
each digit. Once last digit in Alarm Lo is locked, your SMAN will
3 See steps 6-8 above.
automatically revert to normal testing mode.
09
10
CAL
CAL
CAL
CAL
CAL
CALCAL
Atmospheric
Pressure
CAL
CAL
Atmospheric
Pressure
Atmospheric
Pressure
Atmospheric
Atmospheric
Pressure
Pressure
CAL
CAL
Test
Pressure
Test
Pressure
CAL
CALCAL CAL
Test
Pressure
Atmospheric
Pressure
Atmospheric
TestTest
Atmospheric
Pressure
Pressure
CAL Pressure
Pressure
CAL
Atmospheric
Pressure
CAL
Test
CAL
Pressure
CAL
Test
Pressure
Atmospheric
Pressure
CAL
Test
Test Pressure
Pressure
CAL
Test
Pressure
Set Vacuum Alarms (SMAN3)
What is Superheat and Subcooling?
Why Do I Need to Measure It?
11
Superheat is the difference between the
actual temperature of the refrigerant (gas) as
it leaves the evaporator and the boiling point
of the refrigerant. After boiling, the refrigerant
continues to heat up. The number of degrees it
“heats up” after boiling is called the superheat.
Under worst case conditions (low load for fixed
orifice systems), the refrigerant in the evaporator
boils off near the end of the evaporator coil. To
make sure liquid doesn’t enter the compressor
under the worst case condition (low load),
the refrigerator or A/C manufacturers publish
charts indicating what the superheat should be
at a given indoor wet bulb measurement and
outdoor air temperature.
Measuring superheat is your best indication
on a fixed orifice system of the proper
refrigerant charge and operating conditions.
If everything else is working properly and the
actual superheat is too high, add refrigerant. If
it’s too low, evacuate refrigerant.
Subcooling is the difference between the
boiling point of the refrigerant in the condenser
and the actual temperature of the refrigerant
Low Side
High Side
12
Location of Subcooling Test
Location of Superheat Test
13
Advanced Pressure Calibration
Temperature
To calibrate your SMAN temperature
thermocouples, adjust the pot on the front of
the meter labeled T1 Cal or T2 Cal. The best way
to calibrate is to match to a known temperature.
Ice water is very close to 32°F and is readily
available. Accuracies of one degree or better
are easily obtained.
Note: If the P1 and P2 pressure readings are not at zero, perform a
Pressure Zeroing adjustment before performing this advanced
calibration.
1 Press circular button until your SMAN enters T1 T2 Direct mode.
2 Plug in a K-type thermocouple into T1. (A bead type
thermocouple, like the ATB1, is recommended.)
3 Connect the SMAN to a refrigerant cylinder of a known, single
refrigerant using an EPA approved service hose. Be sure to open
both HIGH and LOW side valves on your manifold and cap the
unused ports. (If caps are not available you can connect both ends
of a refrigerant hose to the two unused caps. Note you will lose
some refrigerant into the hose using this method.)
4 Press the REFRIGERANT button to match the refrigerant of the
cylinder you are using.
5 Attach bead-type thermocouple to the side of the cylinder using
tape. It is recommended to attach in the middle of the cylinder.
Important: Let the temperature of the thermocouple stabilize
to the refrigerant temperature for 1 to 2 minutes or until stable.
Note: For best results, the refrigerant cylinder should have been
stored in a stable ambient condition environment for at least 24
hours.
6 Open the refrigerant cylinder. The pressure inside cylinder should
now be displayed on both HIGH and LOW side pressure sensors.
7 Press the CAL Test Pressure button.
1 Stabilize a large cup of ice water by stirring. Pure, distilled water
will be the most accurate.
2 Press the circular button until your SMAN enters Direct
Temperature (T1 T2) mode.
3 Immerse the temp probe in ice water from T1 and adjust the
T1 Cal pot with a flathead screwdriver and let it stabilize, keep
stirring.
4 Repeat Step 3 for temp probe in T2.
Pressure Zeroing
To calibrate your SMAN pressure sensors to
atmospheric pressure, ensure that your SMAN
is disconnected from any pressure source and
at equilibrium with the ambient pressure.
1 SMAN must be in Superheat/Subcooling, Saturation, or T1 T2
Direct mode.
2 Press the CAL Atmospheric Pressure button and your SMAN will
set the zero point of pressure to the ambient pressure.
Cleaning the Sensors (SMAN3)
Over time, the vacuum sensor of the SMAN3
may become contaminated with dirt, oil, and
other contaminents introduced from pulling
vacuums.
The P-T charts of the following refrigerants come pre-programmed
into your SMAN. In your SMAN the refrigerants are listed in order of
most commonly used. Here, they are listed in numerical order for easy
reference.
R11, R113, R114, R12, R123, R124, R125, R13, R134A, R22, R23,
R401A(MP39), R401B, R402A, R402B, R404A, R406A, R407A, R407C,
R408A, R409A, R410A, R414B (Hotshot), R416A, R417A, R420A,
R421A, R421B, R422A, R422B(NU22B), R422C(Oneshot), R422D, R424A,
R434A(RS-45), R438A(MO99), R500, R502, R503, R507A
1 Never use an object such as a cotton swab to clean the sensor,
you may cause damage to the sensor.
2 Open all knobs/valves, and cap all the ports except for the center
port.
3 Drop enough Isopropyl (rubbing) alcohol into the uncapped
port using an eye dropper or funnel so that it can flush out
contaminents.
4 Cap central port and gently shake your SMAN upside down to
clean sensor.
5 Turn right side up. Open a port to pour out the rubbing alcohol
and open all ports to allow sensors to dry out; usually an hour or
so.
15
Limited Warranty
Obtaining Service
This meter is warranted against defects in
material or workmanship for one year from
date of purchase. Fieldpiece will replace or
repair the defective unit, at its option, subject
to verification of the defect.
This warranty does not apply to defects
resulting from abuse, neglect, accident,
unauthorized repair, alteration, or unreasonable
use of the instrument.
Any implied warranties arising from the sale of
a Fieldpiece product, including but not limited
to implied warranties of merchantability and
fitness for a particular purpose, are limited to
the above. Fieldpiece shall not be liable for loss
of use of the instrument or other incidental or
consequential damages, expenses, or economic
loss, or for any claim of such damage, expenses,
or economic loss.
State laws vary. The above limitations or
exclusions may not apply to you.
Safety Information
!
The battery must be replaced when the
battery life indicator
is empty. Turn your
SMAN off and replace with 6 AA batteries.
Sensor type: K-Type thermocouple
Range: -76°F to 999.9°F (-60°C to 537.0°C)
Resolution: 0.1°F/°C
Accuracy: ±(1.0°F) -76°F to 199.9°F; ±(0.5°C) -60°C to 93°C
±(2.0°F) 200°F to 999.9°F; ±(1.0°C) 93°C to 537.0°C
Note: All accuracies are after a field calibration.
14
to the vapor saturation temperature. If the
pressures on your SMAN are within 3psi of the
pressure corresponding to the vapor saturation
temperature, the SMAN will adjust the pressure
sensor linearity to match the P-T chart.
Your SMAN has the ability to perform a linear
adjustment of the pressure sensors based on
refrigerant type, temperature, and pressure.
Battery Replacement
Refrigerants
Connector Type: Standard 1/4 NPT male flare fitting
Range: 29” HgV to 500Psig (English), 74 cmHgV to 0 to 4000KPa
(Metric)
Resolution: 0.1 psi/inHg; 1 kPa/cmHg
Accuracy: 29” HgV to 0” HgV: ±0.2” HgV
74 cmHgV to 0 cmHgV: ±0.4 cmHgV
0 to 200 Psig: ±1 Psi,0 to 1378 KPa: ±7 KPa
200 to 500 Psig: ±0.3%+1 Psi, 1378 to 3447KPa: ±0.3%+7 Kpa
Maximum Overload pressure: 800 psig
Units: Psig, kPa, inHg, and cmHg
Compressor
Clean the exterior with a dry cloth. Do not
use liquid.
Temperature
Pressure
Condenser
Calibration
16
Direction of
Refrigerant
Flow
Maintenance
Connector Type: Standard 1/4 NPT male flare fitting
Range: 50 to 9999 microns of mercury
Resolution: 1 micron (50 to 2000 microns), 250 microns (2001 to
5000 microns), 500 microns (5001 to 8000 microns), 1000 microns
(8001 to 9999 microns)
Accuracy: ±10% or ±10 microns, whichever is greater (50 to 1000
microns)
Maximum Overload pressure: 500 psig
Units: Microns of mercury
Temperature coefficient: 0.1 x (specified accuracy) per °C (0°C to
18°C, 28°C to 50°C), per 0.6°F (32°F to 64°F, 82°F to 122°F)
Operating environment: 32°F to 122°F (0°C to 50°C) at <75%
relative humidity
Storage temperature: -4°F to 140°F (-20°C to 60°C), 0 to 80% RH
(with battery removed)
Display size: 5 inches (diagonal)
Backlight: Blue (On for 1 minute unless turned off manually)
Battery: 6 x AA (Battery life below based on alkaline type)
Battery life (SMAN2): Approx. 500 hours (without backlight)
Battery life (SMAN3): Approx. 135 hours (without backlight)
Low battery indication:
is displayed when the battery
voltage drops below the operating level
Auto Shut off: 30 minutes of inactivity when APO is activated
Over range: "OL" or "-OL" is displayed
Weight: 3.05 lbs (1.38 kg)
Evaporator
TXV
Microns for Vacuum (SMAN3 only)
Specifications
Accuracy: Stated accuracy @ 73°F±9°F (23°C±5°C), <75%RH
as it leaves the condenser. The degrees that the
refrigerant “cools down” below the boiling point
is the subcooling. Under worst case scenario
(low load for TXV) the subcooling will continue
to rise. If the subcooling rises too high, liquid
may be backed into the compressor causing
damage and catastrophic failure. See www.
fieldpiece.com for more technical articles.
WARNINGS
DO NOT APPLY MORE THAN 800 PSI TO ANY PORT ON THE MANIFOLD.
FOLLOW ALL EQUIPMENT MANUFACTURER'S TESTING PROCEDURES
ABOVE THOSE IN THIS MANUAL IN REGARDS TO PROPERLY
SERVICING THEIR EQUIPMENT.
Call Fieldpiece Instruments for current costs
on our fixed price warranty service. Send
check or money order for the amount quoted.
Send your digital manifold, freight prepaid, to
Fieldpiece Instruments. Send proof of date and
location of purchase for in-warranty service.
The meter will be repaired or replaced, at the
option of Fieldpiece, and returned via least cost
transportation.
www.fieldpiece.com
Your SMAN refers to the built-in P-T charts
to compare the temperature of the refrigerant
17
© Fieldpiece Instruments, Inc 2010; v21
18
19
20