Download INOmax® DS Troubleshooting

Transcript
INOmax® DSIR Troubleshooting
Warning: Use caution when troubleshooting the INOmax DSIR while in use for a patient. When possible replace
the unit in question, and perform troubleshooting procedure once the unit is removed from the patient.
If the system fails to operate properly:
1. Check the patient condition and take appropriate action.
2. Verify that the system is set up as detailed in Section 2/ Setup of the Operation and Maintenance Manual.
3. Use the INOblender® (see INOblender Operation
Manual) or backup mode if necessary (see page 29 of manual).
4. Find a symptom or alarm condition in the
troubleshooting table which best describes the
problem and follow the recommended actions to resolve the problem.
If the problem can’t be corrected:
Contact Customer Service on 1300 198 565.
Troubleshooting Guide
High Priority Alarms
Symptom/Alarm
1. High NO Alarm
High NO
Possible Cause
Recommended Action
A.Note: A newly installed NO sensor will give high readings until fully conditioned (about 5 hours) and calibrated.
a. After installation of the NO sensor perform a low and high calibration.
B.The High NO alarm level may be
inappropriately set.
a. Make sure the High NO alarm is set greater
than the Set NO value.
C.The NO calibration may have drifted.
a. Perform a low and high range calibration of the NO sensor.
b. Wait 5 hours and repeat both the low and
high calibration.
b. Check calibration sample tee for leaks.
D.Circuit setup incorrect.
a. Check circuit setup for correct use of adapters and/or check valves.
High Priority Alarms
Symptom/Alarm
2.LowNOAlarm
Low NO
Possible Cause
Recommended Action
A.ThePatientGasSamplelinemaybe
disconnected.
a. ReconnectthePatientGasSampleline.
B.TheLowNOalarmsettingmaybe
inappropriatelyset.
a. MakesuretheLowNOalarmissetlessthan
theSetNOvalue.
C.TheNOcalibrationmayhavedrifted.
a. Performalowandhighrangecalibration
oftheNOsensor.
D.TheNOsensormaynotbeproperlyseated.
a. Makesurethesensorsarecorrectly
seatedwiththeO-ringsandthesensor
coverisfullyclosed.
E. LossofNOdelivery.
a. IftheINOblender®isavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
b. TurnthebackupmodeON(seepage29of
OperationandMaintenanceManual).
3. HighNO2Alarm
A.Incompletesystempurge.
a. Performasystempurge.
SeeSection3/
Pre-UseCheckout.
B.Ventilatorflowstopped.
a. AllowtheventilatorgastoflushNOandNO2
fromthebreathingcircuitbeforeconnecting
tothepatient.
C.Twocylindervalvesareopen.
a. Closeoneofthecylindervalves.
D.TheNO2alarmlimitmaybesettoolow.
a. MakesuretheNO2alarmlimitisappropriate
fortheSetNOlevel.
E. TheNO2calibrationmayhavedrifted.
a. Performalowandhighrangecalibrationof
theNO2sensor.
High NO2
b. Checkcalibrationsampleteeforleaks.
F. Outofdateorthewrongcalibration
gaswasused.
a. Verifythecalibrationgasexpirationdate.
G.Thepatientcircuitsetupmaybeincorrect.
a. Makesurethepatientcircuithosesandlengths
arecorrect(seeSection4/PatientApplication
ofOperationandMaintenanceManual).
b. Ifneededreplacethecalibrationgasand
performalowandhighrangecalibration
oftheNO2sensor.
b. Verifythehumidifierchamberisless
than480mL.
H.Samplelineocclusion.
Inthiscase,thisalarmmayoccurwith
aSampleLineBlockalarm.
I. TheINOmax®DSIRmayhavefailed.
a. ConfirmwhethertheHighNO2alarmoccurs
concurrentlywithasamplelineblockalarm.
b. Ifso,thisalarmwillclearwithin10seconds
afterthesamplelinefaultisremedied.
a. ContactCustomerService.
b. Replacethedeliverysystemifinuse.
c. Donotusethedeliverysystemuntilserviced.
High Priority Alarms
Symptom/Alarm
4. HighO2Alarm
Possible Cause
Recommended Action
A.TheO2alarmsettingmaybeinappropriate.
a. MakesuretheHighO2alarmisset
appropriatelyfortheO2settingbeingused
ontheventilator.
B.TheO2calibrationmayhavedrifted.
a. Performalowandhighrangecalibration
oftheO2sensor.
High O2
b. ChangetheO2sensorifthemonitorfails
tocalibrate.
c. ContactCustomerService.
d. Replacethedeliverysystemifinuse.
e. Donotusethedeliverysystemuntilservice.
5.LowO2Alarm
Low O2
A.TheO2concentrationsettingattheventilator
wasreduced.
a. MakesuretheO2alarmsettingiscorrectfor
thesettingattheventilator.
B.TheO2alarmsettingmaybeinappropriate.
a. TheINOmaxDSIRcandilutetheO2
concentrationsetattheventilatorby
upto10%.
b. Verifythatthealarmissetappropriatelyfor
theO2settingbeingusedontheventilator.
C.TheO2sensormaynotbeproperlyseated.
a. Makesurethesensorsarecorrectlyseated
andthesensorcoverisfullyclosed.
D.TheO2calibrationmayhavedrifted.
a. Performalowandhighrangecalibration
oftheO2sensor.
b. ContactCustomerService.
6.CylinderNotDetected
INOmax
INOmax
20
10
5
40
NO
60
(ppm)
1
80
0
or
CylinderValveClosed
INOmax
20
10
5
40
NO
60
(ppm)
1
0
80
a. RemoveobstructionbetweentheINOMAX
A.InterferencewiththeInfraredcommunication
linkbetweentheINOMAXcylinderandthe
cylinderandtheINOmaxDSIR .
INOmaxDSIR.(DeliveryStoppedwilloccur
b. Movetheinterferinglightorthecartto
onehourfromthepointwhencommunication
reducethehighintensitylightintheareaof
islost).
theINOmeter®.
B.INOMAXcylindervalveisclosed.(Delivery
Stoppedwilloccuronehourfromthepoint
whenthecylindervalveisclosed).
a. OpenINOMAXcylindervalve.
C.TransportCapnotconnectedtotheINOmeter. a. AttachtheTransportCaptotheINOmeter
(DeliveryStoppedwilloccuronehourfromthe
ontheINOMAXcylinder.
pointwhencommunicationislost).
b. ConnecttheTransportRegulator/Cap
Assemblycabletotheinfraredconnectoron
thebackoftheINOmaxDSIR .
D.INOMAXcylindernotpresentonthe
INOmaxDSIR .(DeliveryStoppedwill
occuronehourfromthepointwhen
communicationislost).
a. LoadanINOMAXcylinderontothe
INOmaxDSIRcart.
E. INOmetermayhavefailed.
a. ReplacetheINOMAXcylinderonthe
INOmaxDSIRcart.
F. INOmaxDSIRinfraredcartcableisnot
connectedorhasfailed.
a. Connectinfraredcartcabletotheinfrared
connectoronthebackoftheINOmaxDSIR .
b. ReplacetheINOmaxDSIR .
High Priority Alarms
Symptom/Alarm
7. DeliveryFailure
Possible Cause
A.MonitoredNOlevels≥100ppm
or
20
40
10
5
B.TheINOmaxDSIRhasfailed.
60
a. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
b. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
1
0
Recommended Action
80
c. PowertheINOmaxDSIRtoSTANDBYand
thenbackONtoresetthedeliverysystem.If
thisdoesnotworkcontactCustomerService.
d. Replacethedeliverysystemifinuse.
e. Donotusethedeliverysystemuntilserviced.
8.DeliveryStopped
INOmax
INOmax
20
10
5
40
NO
60
(ppm)
1
0
A.Infraredcommunicationlinkbetweenthe
INOMAXcylinderandtheINOmaxDSIR
hasbeenlostforonehour.
a. RemoveobstructionbetweentheINOMAX
cylinderandtheINOmaxDSIR .
b. Movetheinterferinglightorthecartto
reducethehighintensitylightintheareaof
theINOmeter.
c. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
80
or
d. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
B.INOMAXcylinderisexpiredandcylinder
valvehasbeenopenfortwominutes.
a. RemovetheINOMAXcylinderfromthe
INOmaxDSIRcart.
b. ConnectanINOMAXcylindertothe
INOmaxDSIRwithavalidexpirationdate.
C.INOMAXcylinderisthewrong
concentrationandcylindervalvehas
beenopenfortwominutes.
a. RemovetheINOMAXcylinder
withthewrongconcentrationfromthe
INOmaxDSIRcart.
b. ConnectanINOMAXcylindertothe
INOmaxDSIRwithavalidconcentration.
D.INOmetermayhavefailed.
a. ReplacetheINOMAXcylinderonthe
INOmaxDSIRcart.
E. INOMAXcylindervalveisclosed.
a. OpenINOMAXcylindervalve.
b. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
c. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
F. INOMAXcylindernotpresentonthe
INOmaxDSIR .
a. LoadanINOMAXcylinderontothe
INOmaxDSIRcart.
b. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
c. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
High Priority Alarms
Symptom/Alarm
9. DrugPastExpiryDate
or
DrugConcentrationMismatch
Possible Cause
A.INOMAXcylinderisexpired(Delivery
Stoppedwilloccurtwominutesfromthe
pointwhenthecylindervalveisopened).
Recommended Action
a. Closethecylindervalve.
b. RemoveexpiredINOMAXcylinderfrom
theINOmaxDSIRcart.
c. ReplacetheexpiredINOMAXcylinderon
theINOmaxDSIRcart.
INOmax
INOmax
20
10
5
40
NO
60
(ppm)
1
0
80
B.INOMAXcylinderisthewrong
concentration.(DeliveryStoppedwill
occurtwominutesfromthepointwhen
thecylindervalveisopened).
a. Closethecylindervalve.
b. RemovetheINOMAXcylinderwiththe
wrongconcentrationfromtheINOmax
DSIRcart.
c. ReplacetheexpiredINOMAXcylinder
withthewrongconcentrationonthe
INOmaxDSIRcart.
10. InjectorModuleFail
A.TheInjectorModuleelectricalcable
maybedisconnected.
a. ReconnecttheInjectorModuleelectrical
cable.
b. TurnOFFtheINOmaxDSIRsetdoseto
silencethealarm.
c. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
d. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
B.TheInjectorModulemayhavefailed.
a. TurnOFFtheINOmaxDSIRsetdoseto
silencethealarm.
b. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
c. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual)
d. ReplacetheInjectorModule.
e. SetthedelivereddoseandturnOFFthe
INOblenderorthebackupmode.
C.TheInjectorModuleelectricalcable
mayhavefailed.
a. TurnOFFtheINOmaxDSIRsetdoseto
silencethealarm.
b. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
c. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
d. ReplacetheInjectorModuleelectrical
cable.
e. ResetthedelivereddoseandturnOFF
theINOblenderorthebackupmode.
High Priority Alarms
Symptom/Alarm
11.LowBatteryAlarm
Possible Cause
A.Batteryisrunninglow(approximately
10minutesorlessuntilbatterydepletion).
Recommended Action
a. Checkmainpowerindicator.
b. ConnecttoACmainpowersource.
c. Makesurethepowercordisfullyinsertedinto
thePowerCordInletandthatthepowercord
clampissecure.
d. Checkandreplacefuseifnecessary.
e. ContactCustomerService.
12. LowNO/N2Pressure
A.TheNOcylindersupplymaybelow.
a. MakesuretheNOcylinderisturnedON.
b. Ifthehighpressurecylindergaugereadsless
than200psig,changethecylinder.
INOmax
INOmax
20
10
5
NO
60
(ppm)
1
c. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
40
0
or
80
d. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
B.Thesupplylinemaynotbeconnected.
a. Ifthecylindergaugereadsgreaterthan
200psig.
b. Verifythelowpressurehosesareconnected
correctlytothebackoftheINOmaxDSIR.
c. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
d. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
13.ServiceRequired
A.TheINOmaxDSIRhasfailed.
Service Required
a. IftheINOblenderisavailable,manually
ventilatethepatient(seeINOblender
OperationManual).
or
b. TurnthebackupmodeON(seepage29
ofOperationandMaintenanceManual).
Manual Delivery Available
c. TurntheINOmaxDSIRtoSTANDBYand
thenbackONtorestthedeliverysystem.
PleasecontactCustomerServiceandreplace
thedeliverysystemassoonaspossible.
Utilize the INOblender or Backup Delivery
14. SetNOandBackupOn
INOmax
INOmax
20
10
5
60
(ppm)
0
80
a. TurntheINOmaxDSIRsetdosetozero.
b. Correctthereasonforinitiatingthe
backupmode.
c. TurnONtheINOmaxDSIRsetdose.
40
NO
1
A.ThebackupmodehasbeenturnedON
andthesetdoseisstillset.
d. TurnthebackupmodeOFF.
e. IfthisdoesnotworkcontactCustomerService.
Low Priority Alarms
Symptom/Alarm
Possible Cause
A.ThebackupmodehasbeenturnedON
andthesetdoseiszero.
15. BackupOn
250ml/min
INOmax
INOmax
20
10
5
c. TurnthebackupmodeOFF.
40
60
(ppm)
0
a. Correctthereasonforinitiatingthe
backupmode.
b. TurnONtheINOmaxDSIRsetdose.
NO
1
Recommended Action
80
16. FailedNOSensor
A.Thewrongcalibrationgasmayhavebeenused.
a. Makesurethecorrectcalibrationgasisused
andensuresampletubingconnectionsare
secureanddonotleak.
B.Thereisaleakaroundthesensors.
a. Makesurethesensorsarecorrectly
seatedwiththeO-ringsandthesensor
coverisfullyclosed.
C.NOsensorabsentorfailed.
a. Completealowcalibrationfirst,andthen
repeatthehighcalibration.
b. Sensorneedstobereplaced.
c. Replacethedeliverysystemifinuse.
d. ContactCustomerService.
17. FailedNO2Sensor
A.Thewrongcalibrationgasmayhavebeenused.
a. Makesurethecorrectcalibrationgasis
usedandensuresampletubingconnections
aresecureanddonotleak.
B.Thereisaleakaroundthesensors.
a. Makesurethesensorsarecorrectly
seatedwiththeO-ringsandthesensor
coverisfullyclosed.
C.NO2sensorabsentorfailed.
a. Completealowcalibrationfirst,andthen
repeatthehighcalibration.
b. Sensorneedstobereplaced.
c. Replacethedeliverysystemifinuse.
d. ContactCustomerService.
18. FailedNO2Sensor
A.Thewrongcalibrationgasmayhavebeenused.
a. Makesurethecorrectcalibrationgasisused
andensuresampletubingconnectionsare
secureanddonotleak.
B.Thereisaleakaroundthesensors.
a. Makesurethesensorsarecorrectlyseated
andthesensorcoverisfullyclosed.
C.O2sensorabsentorfailed.
a. Completealowcalibrationfirst,andthen
repeatthehighcalibration.
b. Sensorneedstobereplaced.
c. Replacethedeliverysystemifinuse.
d. ContactCustomerService.
Low Priority Alarms
Symptom/Alarm
A.Monitorisfailingtocommunicatecorrectly
orisreportingafault.
19. MonitoringFailure
O
(%)
2
21
NO
(ppm)
2
3.0
NO
Possible Cause
a. DoesnotstopdeliveryofINOMAXto
thepatient.
b. ContactCustomerService.
5
(ppm)
20.SampleLine/FilterBlock
Recommended Action
A.Thesamplelinemaybeblocked.
a. Makesurethesampleinletlineandoutlet
portsarenotobstructed.
b. Changethesampleline.
21. TwoCylindersOpen
INOmax
INOmax
20
10
5
A.Twocylindervalvesareopen.
a. Closeoneofthecylindervalves.
A.Thewatertrapbottleonthesideofthe
INOmaxDSIRisfull.
a. Emptythewatertrapbottle.
B.Watertrapbottleisemptybutthemessage
remainsinthealarmmessagebox.
a. Removethewatertrapbottleandclean
theopticalsensorlevelindicatorwithan
alcoholswab.
C.TheINOmaxDSIRmayhavefailed.
a. ContactCustomerService.
60
(ppm)
0
a. Replacethewater
separatorcartridge.
40
NO
1
B.Thewaterseparatorcartridgemaybeblocked.
80
22.WaterTrapBottleFull
b. Replacethedeliverysystemifinuse.
Indicators
Symptom/Indicator
23.PatientInfoIncomplete
Possible Cause
Recommended Action
A.Patientidentifierhasnotbeenentered.
a. Enterpatientidentifier.
A.Thesetdosehasbeensettozeroandthe
INOMAXcylindervalveisstillopen.
a. ClosetheINOMAXcylindervalveif
treatmenthasbeenstopped.
A.Deviceisoperatingonthebattery.
a. ConnecttoACmainpowersource
whenavailable.
Patient Info Incomplete
24. SetDoseisZero,Close
CylinderValve
Set Dose is Zero,
Close Cylinder Valve
25.RunningonBattery
b. Checkmainpowerindicator.
c. Makesurethepowercordisfully
insertedintothePowerCordInlet
andthatthepowercordclampissecure.