9 Controller Operation
Serial Interface Parameters
Press “UP” to enter the menu. The operating parameters menu contains the
cutout reset mode setting.
9.11.1
Cutout Reset Mode
The cutout reset mode determines whether the cutout resets automatically when
the bath temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,
Cutout reset mode parameter
Press “SET” to access the parameter setting. Normally the cutout is set for
manual mode.
Cto=rSt
Cutout set for manual reset
To change to automatic reset mode press “UP” and then “SET”.
Cto=Auto
9.12
Cutout set for automatic reset
Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SErIAL
Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to baths fitted with
the serial interface. The parameters in the menu are — BAUD rate, sample period, duplex mode, and linefeed.
9.12.1
Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting determines the serial communications transmission rate.
The baud rate parameter is indicated by,
BAUd
Serial BAUD rate parameter
Press “SET” to choose to set the baud rate. The current baud rate value will
then be displayed.
1200 b
Current baud rate
49