Download RAE Systems QRAE 3 Manual - Enviro
Transcript
General Alarms Message Condition Alarm Indications HIGH Gas exceeds “High Alarm” limit 3 beeps/flashes per second OVR Gas exceeds sensor’s measurement range 3 beeps/flashes per second MAX Gas exceeds electronic circuit’s maximum range 3 beeps/flashes per second LOW Gas exceeds “Low Alarm” limit* 2 beeps/flashes per second TWA Gas exceeds “TWA” limit 1 Beep/flash per second STEL Gas exceeds “STEL” limit 1 Beep/flash per second Crossed pump icon flashes Inlet blocked or pump failure 3 beeps/flashes per second Empty battery icon flashes Low battery 1 flash, 1 beep per minute CAL Calibration failed, or needs calibration 1 beep/flash per second NEG Zero gas reading measures less than number stored in calibration 1 beep/flash per second * For oxygen, “low alarm limit” means a concentration is lower than the low alarm limit. Testing Alarms Under normal operation mode and non-alarm conditions, the audible, visible, and vibration alarms can be tested at any time by pressing the [Y/+] key. 43