TRANSIENT-2000
The following Spike Rate Limits are valid for the TRANSIENT-2000:
Nominal Voltage
4500
4000
3500
3000
2500
2000
1500
1000
500
0
0
2000
4000
6000
8000
Spike Rate max.(Spikes/ s)
6.2.2.1.2 Spike Rate = number of spikes/ seconds
Selection of coupling path
When pressing "More" the display below will be showed
loaded test
EUT power v, i
coupling paths
coupling paths
The coupling paths can be selected as follows:
1. Set the cursor to the coupling path to editing
2. select ON or OFF
3. Confirm that the values entered are correct by pressing the ENTER - button.
L
N
PE
L-GND
N- GND
PE-GND
Reference ground, (metallic plate)
GND
6.2.2.1.3 Definition coupling path
When several coupling paths are activated „ON“, they will be chosen step by step. The sequence
corresponds with the sequence listed in the „Main“ menu. When Impulse -Out = On, the burst impulses stay
at the high voltage. The capacitive coupling clamp can be connected on the high voltage EFT output
(superimposing EFT on data line).
Random Spikes
55/142