Download SureSelect Automated Target Enrichment for Illumina Multiplexed
Transcript
4 Hybridization Step 2. Hybridize the gDNA library and SureSelect or ClearSeq Capture Library Program the thermal cycler 1 Pre-program the thermal cycler for the Hybridization workflow by entering the thermal cycling program shown in Table 28 below. It is critical to pre-program the thermal cycler before starting the automation protocol for Hybridization, in order to maintain the required sample and reagent temperatures during the workflow. Table 28 Thermal cycler program for Hybridization* Segment Number Purpose Number of Cycles Temperature Time 1 Denaturation 1 95°C 5 minutes 2 Blocking 1 65°C 10 minutes 3 Hold for NGS Bravo steps† 1 65°C Hold 4 Hybridization 60 65°C 1 minute 37°C 3 seconds 65°C Hold 5 Hold until start of Capture‡ 1 * When setting up the thermal cycling program, use a reaction volume setting of 35 l (final volume of hybridization reactions during cycling in Segment 4). † Samples are transferred to the NGS Bravo during this Hold step when prompted by the VWorks software. ‡ Samples are held at 65°C until they are processed in the Capture & Wash automation protocol that begins on page 83. CA U T I O N 70 The lid of the thermal cycler is hot and can cause burns. Use caution when working near the lid. SureSelectQXT Target Enrichment for Illumina Sequencing (NGS Bravo Option A)