4
Hybridization
Step 2. Hybridize the gDNA library and SureSelect or ClearSeq Capture Library
Program the thermal cycler
1 Pre-program the thermal cycler for the Hybridization workflow by
entering the thermal cycling program shown in Table 28 below.
It is critical to pre-program the thermal cycler before starting the
automation protocol for Hybridization, in order to maintain the required
sample and reagent temperatures during the workflow.
Table 28
Thermal cycler program for Hybridization*
Segment
Number
Purpose
Number of Cycles Temperature
Time
1
Denaturation
1
95°C
5 minutes
2
Blocking
1
65°C
10 minutes
3
Hold for NGS Bravo
steps†
1
65°C
Hold
4
Hybridization
60
65°C
1 minute
37°C
3 seconds
65°C
Hold
5
Hold until start of
Capture‡
1
* When setting up the thermal cycling program, use a reaction volume setting of 35 l (final volume
of hybridization reactions during cycling in Segment 4).
† Samples are transferred to the NGS Bravo during this Hold step when prompted by the VWorks
software.
‡ Samples are held at 65°C until they are processed in the Capture & Wash automation protocol that
begins on page 83.
CA U T I O N
70
The lid of the thermal cycler is hot and can cause burns. Use caution when working
near the lid.
SureSelectQXT Target Enrichment for Illumina Sequencing (NGS Bravo Option A)