Installing Your Instrument
Operating Requirements
CAUTION. To ensure proper cooling, keep sides of the instrument clear of obstructions.
Place the instrument on a cart or bench, observing clearance requirements:
Top: 20 mm (0.8 in)
Left and right side: 150 mm (5.9 in)
Bottom: 20 mm (0.8 in)
Rear: 75 mm (3 in)
Environmental Requirements
Before operating the instrument, ensure that the instrument has warmed up for 20 minutes and meets the environmental
requirements listed in the following table.
Requirement
Description
Temperature (operating)
10 °C to 40 °C (+50 °F to +104 °F)
Relative humidity (operating)
5% to 95% relative humidity at up to 30 °C (86 °F)
5% to 45% relative humidity above 30 °C (86 °F) up to
+40 °C (104 °F) noncondensing, and as limited by a
maximum wet-bulb temperature of 29 °C (84.2 °F)
Altitude (operating)
Up to 3,000 m (approximately 10,000 feet)
Power Supply Requirements
WARNING. To reduce the risk of fire and shock, ensure that the mains supply voltage fluctuations do not exceed 10% of the
operating voltage range.
2
Source Voltage and Frequency
Power Consumption
100 VAC to 240 VAC, 47 Hz to 63 Hz
Less than 450 W (AWG7000 series)
Less than 560 W (AWG5000 series)
AWG5000 and AWG7000 Series Quick Start User Manual