Download AssayMaxTM Human Factor VII ELISA Kit

Transcript
AssayMaxTM
Human Factor VII
ELISA Kit
Assaypro LLC
3400 Harry S Truman Blvd
St. Charles, MO 63301
T (636) 447-9175
F (636) 395-7419
www.assaypro.com
For any questions regarding troubleshooting or performing the assay, please contact our
support team at [email protected].
Thank you for choosing Assaypro.
Assay Summary
Step 1. Add 50 µl of Standard or Sample per well.
Incubate 2 hours.
Step 2. Wash, then add 50 µl of Biotinylated Antibody per well.
Incubate 1 hour.
Step 3. Wash, then add 50 µl of SP Conjugate per well.
Incubate 30 minutes.
Step 4. Wash, then add 50 µl of Chromogen Substrate per well.
Incubate 30 minutes.
Step 5. Add 50 µl of Stop Solution per well.
Read at 450 nm immediately.
Symbol Key
Consult instructions for use.
H
G
F
E
D
C
B
A
1
2
3
4
5
6
7
8
9
10
11
12
Assay Template
Human Factor VII (FVII) ELISA Kit
Catalog No. EF1007-1
Sample insert for reference use only
Introduction
Factor VII (FVII) is a vitamin K-dependent plasma glycoprotein that is
synthesized in the liver and circulates in blood as a single-chain inactive
zymogen with a molecular mass of 50 kDa (1). Upon tissue damage and
vascular injury, the cell surface receptor and cofactor tissue factor binds and
allosterically activates FVII to its active form, FVIIa. The tissue factor/FVIIa
complex catalyzes the conversion of both factor IX to factor IXa and factor X to
factor Xa to initiate coagulation via the extrinsic pathway (2, 3).
Principle of the Assay
The AssayMax Human Factor VII (FVII) ELISA (Enzyme-Linked Immunosorbent
Assay) kit is designed for detection of human factor VII and factor VIIa in
plasma, serum, and cell culture samples. This assay employs a quantitative
sandwich enzyme immunoassay technique that measures total FVII in less
than 4 hours. A monoclonal antibody specific for FVII has been pre-coated
onto a 96-well microplate with removable strips. FVII in standards and
samples is sandwiched by the immobilized antibody and the biotinylated
antibody specific for FVII, which is recognized by a streptavidin-peroxidase
conjugate. All unbound material is washed away and a peroxidase enzyme
substrate is added. The color development is stopped and the intensity of the
color is measured.
Caution and Warning






This product is for Research Use Only and is Not For Use In Diagnostic
Procedures.
Prepare all reagents (working diluent buffer, wash buffer, standard,
biotinylated antibody, and SP conjugate) as instructed, prior to running
the assay.
Prepare all samples prior to running the assay. The dilution factors for
the samples are suggested in this insert. However, the user should
determine the optimal dilution factor.
Spin down the SP conjugate vial and the biotinylated antibody vial before
opening and using contents.
The Stop Solution is an acidic solution.
The kit should not be used beyond the expiration date.
1
Reagents









Human FVII Microplate: A 96-well polystyrene microplate (12 strips of 8
wells) coated with a monoclonal antibody against human FVII.
Sealing Tapes: Each kit contains 3 precut, pressure sensitive sealing
tapes that can be cut to fit the format of the individual assay.
Human FVII Standard: Human FVII in a buffered protein base (150 ng,
lyophilized).
Biotinylated Human FVII Antibody (50x): An 50-fold concentrated
biotinylated antibody against FVII (120 l).
MIX Diluent Concentrate (10x): A 10-fold concentrated buffered protein
base (30 ml).
Wash Buffer Concentrate (20x): A 20-fold concentrated buffered
surfactant (30 ml, 2 bottles).
Streptavidin-Peroxidase Conjugate (SP Conjugate): A 100-fold
concentrate (80 l).
Chromogen Substrate: A ready-to-use stabilized peroxidase chromogen
substrate tetramethylbenzidine (8 ml).
Stop Solution: A 0.5 N hydrochloric acid to stop the chromogen
substrate reaction (12 ml).
Storage Condition






Upon arrival, immediately store components of the kit at recommended
temperatures up to the expiration date.
Store SP Conjugate and Biotinylated Antibody at -20°C.
Store Microplate, Diluent Concentrate (10x), Wash Buffer, Stop Solution,
and Chromogen Substrate at 2-8°C.
Unused microplate wells may be returned to the foil pouch with the
desiccant packs and resealed. May be stored for up to 30 days in a
vacuum desiccator.
Diluent (1x) may be stored for up to 30 days at 2-8°C.
Store Standard at 2-8°C before reconstituting with Diluent and at -20°C
after reconstituting with Diluent.
Other Supplies Required



Microplate reader capable of measuring absorbance at 450 nm.
Pipettes (1-20 l, 20-200 l, 200-1000 l, and multiple channel).
Deionized or distilled reagent grade water.
Sample Collection, Preparation, and Storage
2



Plasma: Collect plasma using one-tenth volume of 0.1 M sodium citrate
as an anticoagulant. Centrifuge samples at 3000 x g for 10 minutes.
Dilute samples 1:10 into MIX Diluent and assay. The undiluted samples
can be stored at -20°C or below for up to 3 months. Avoid repeated
freeze-thaw cycles (EDTA and Heparin can also be used as an
anticoagulant).
Serum: Samples should be collected into a serum separator tube. After
clot formation, centrifuge samples at 3000 x g for 10 minutes, and
remove serum. Dilute samples 1:10 into MIX Diluent and assay. The
undiluted samples can be stored at -20°C or below for up to 3 months.
Avoid repeated freeze-thaw cycles.
Cell Culture Supernatants: Collect cell culture media and centrifuge at
3000 x g for 10 minutes at 4°C to remove debris and assay. Samples can
be stored at -20°C or below. Avoid repeated freeze-thaw cycles.
Refer to Sample Dilution Guidelines below for further instruction.
Guidelines for Dilutions of 1:100 or Greater
(for reference only; please follow the insert for specific dilution suggested)
1:100
1:10000
A)
4 ul sample: 396 µl buffer(100x)
= 100 fold dilution
A)
B)
Assuming the needed volume is less than
or equal to 400 µl.
Assuming the needed volume is less than
or equal to 400 µl.
1:1000
A)
B)
4 µl sample : 396 µl buffer (100x)
24 µl of A : 216 µl buffer (10x)
= 1000 fold dilution
Assuming the needed volume is less than
or equal to 240 µl.
4 µl sample : 396 µl buffer (100x)
4 µl of A : 396 µl buffer (100x)
= 10000 fold dilution
1:100000
A)
B)
C)
4 µl sample : 396 µl buffer (100x)
4 µl of A : 396 µl buffer (100x)
24 µl of B : 216 µl buffer (10x)
= 100000 fold dilution
Assuming the needed volume is less than
or equal to 240 µl.
Reagent Preparation



Freshly dilute all reagents and bring all reagents to room temperature
before use.
MIX Diluent Concentrate (10x): If crystals have formed in the
concentrate, mix gently until the crystals have completely dissolved.
Dilute the MIX Diluent Concentrate 1:10 with reagent grade water. Store
for up to 30 days at 2-8°C.
Standard Curve: Reconstitute the 150 ng of Human FVII Standard with
0.75 ml of MIX Diluent to generate a 200 ng/ml standard stock solution.
Allow the standard to sit for 10 minutes with gentle agitation prior to
making dilutions. Prepare duplicate or triplicate standard points by
3
serially diluting the standard stock solution (200 ng/ml) 1:2 with equal
volume of MIX Diluent to produce 100, 50, 25, 12.5, and 6.25 ng/ml
solutions. MIX Diluent serves as the zero standard (0 ng/ml). Any
remaining solution should be frozen at -20°C and used within 30 days.
Standard
Point
P1
P2
P3
P4
P5
P6
P7



Dilution
[FVII] (ng/ml)
1 part Standard (200 ng/ml)
1 part P1 + 1 part MIX Diluent
1 part P2 + 1 part MIX Diluent
1 part P3 + 1 part MIX Diluent
1 part P4 + 1 part MIX Diluent
1 part P5 + 1 part MIX Diluent
MIX Diluent
200.0
100.0
50.00
25.00
12.50
6.250
0.000
Biotinylated Human FVII Antibody (50x): Spin down the antibody briefly
and dilute the desired amount of the antibody 1:50 with MIX Diluent.
Any remaining solution should be frozen at -20°C.
Wash Buffer Concentrate (20x): If crystals have formed in the
concentrate, mix gently until the crystals have completely dissolved.
Dilute the Wash Buffer Concentrate 1:20 with reagent grade water.
SP Conjugate (100x): Spin down the SP Conjugate briefly and dilute the
desired amount of the conjugate 1:100 with MIX Diluent. Any remaining
solution should be frozen at -20°C.
Assay Procedure





4
Prepare all reagents, standard solutions, and samples as instructed. Bring
all reagents to room temperature before use. The assay is performed at
room temperature (20-25°C).
Remove excess microplate strips from the plate frame and return them
immediately to the foil pouch with desiccants inside. Reseal the pouch
securely to minimize exposure to water vapor and store in a vacuum
desiccator.
Add 50 l of Human FVII Standard or sample per well. Cover wells with a
sealing tape and incubate for 2 hours. Start the timer after the last
addition.
Wash five times with 200 l of Wash Buffer manually. Invert the plate
each time and decant the contents; hit 4-5 times on absorbent material
to completely remove the liquid. If using a machine, wash six times with
300 l of Wash Buffer and then invert the plate, decanting the contents;
hit 4-5 times on absorbent material to completely remove the liquid.
Add 50 l of Biotinylated Human FVII Antibody to each well and incubate
for 1 hour.






Wash the microplate as described above.
Add 50 l of Streptavidin-Peroxidase Conjugate per well and incubate for
30 minutes. Turn on the microplate reader and set up the program in
advance.
Wash the microplate as described above.
Add 50 l of Chromogen Substrate per well and incubate for 30 minutes
or till the optimal blue color density develops. Gently tap the plate to
ensure thorough mixing and break the bubbles in the well with pipette
tip.
Add 50 l of Stop Solution to each well. The color will change from blue
to yellow.
Read the absorbance on a microplate reader at a wavelength of 450 nm
immediately. If wavelength correction is available, subtract readings at
570 nm from those at 450 nm to correct optical imperfections.
Otherwise, read the plate at 450 nm only. Please note that some
unstable black particles may be generated at high concentration points
after stopping the reaction for about 10 minutes, which will reduce the
readings.
Data Analysis



Calculate the mean value of the duplicate or triplicate readings for each
standard and sample.
To generate a standard curve, plot the graph using the standard
concentrations on the x-axis and the corresponding mean 450 nm
absorbance (OD) on the y-axis. The best-fit line can be determined by
regression analysis using log-log or four-parameter logistic curve-fit.
Determine the unknown sample concentration from the Standard Curve
and multiply the value by the dilution factor.
Typical Data

The typical data is provided for reference only. Individual laboratory
means may vary from the values listed. Variations between laboratories
may be caused by technique differences.
5
Standard Point
ng/ml
P1
200.0
P2
100.0
P3
50.00
P4
25.00
P5
12.50
P6
6.250
P7
0.000
OD
2.014
1.961
1.471
1.467
0.818
0.797
0.470
0.457
0.373
0.357
0.181
0.192
0.081
0.087
Sample: Normal,
Sodium Citrate Plasma (10x)
0.771
0.770
Average OD
1.988
1.469
0.807
0.463
0.365
0.187
0.084
0.771
Standard Curve

The curve is used for illustration only. A standard curve should be
generated each time the assay is performed.
OD 450 nm
hFVII Standard Curve
1.0
0.1
1.0
10.0
100.0
[hFVII] (ng/ml)
6
Reference Value


Normal human factor VII plasma levels range from 170 ng/ml to 700
ng/ml.
Human plasma and serum samples from healthy adults were tested
(n=40). On average, factor VII level was 479 ng/ml.
Sample
Human Pool Normal Plasma
Human Normal Plasma
Human Pool Normal Serum
n
10
20
10
Average Value (ng/ml)
467
456
513
Performance Characteristics




This assay recognizes FVII and FVIIa.
The minimum detectable dose of factor VII as calculated by 2SD from the
mean of a zero standard was established to be 2.5 ng/ml.
Intra-assay precision was determined by testing replicates of three
plasma samples in one assay.
Inter-assay precision was determined by testing three plasma samples in
twenty assays.
Sample
n
CV (%)
Average
CV (%)
Intra-Assay Precision
1
2
3
20
20
20
4.8%
5.1%
4.7%
Inter-Assay Precision
1
2
3
20
20
20
8.8%
8.6%
8.4%
4.9%
8.6%
Spiking Recovery

Recovery was determined by spiking two plasma samples with different
factor VII concentrations.
7
Sample
Unspiked
Sample
(ng/ml)
1
36.5
2
18.0
Spike
(ng/ml)
Expected
Observed
Recovery
(%)
40.0
20.0
10.0
40.0
20.0
10.0
76.5
56.5
46.5
58.0
38.0
28.0
74.2
52.6
46.4
55.0
37.8
26.3
97%
93%
100%
95%
99%
94%
Average Recovery (%)
96%
Linearity

Plasma and serum samples were serially-diluted to test for linearity.
Average Percentage of Expected Value (%)
Sample Dilution
Plasma
1:5
90%
1:10
97%
1:20
106%
Serum
92%
99%
103%
Cross-Reactivity
Species
Canine
Bovine
Monkey
Mouse
Rat
Swine
Rabbit
Human

Cross Reactivity (%)
None
None
<10%
None
None
<80%
<5%
100%
10% FBS in culture media will not affect the assay.
Troubleshooting
Low
Precision
Issue
8
Causes
Course of Action
Use of expired
components
 Check the expiration date listed before use.
 Do not interchange components from different lots.
 Check that the correct wash buffer is being used.
 Check that all wells are dry after aspiration.
 Check that the microplate washer is dispensing properly.
 If washing by pipette, check for proper pipetting
technique.
Improper wash step
Splashing of reagents
while loading wells
Inconsistent volumes
loaded into wells
Insufficient mixing of
reagent dilutions
Deficient Standard Curve Fit
Unexpectedly Low or High Signal
Intensity
Improperly sealed
microplate
Microplate was left
unattended between
steps
Omission of step
Steps performed in
incorrect order
Insufficient amount of
reagents added to
wells
Wash step was skipped
Improper wash buffer
Improper reagent
preparation
Insufficient or
prolonged incubation
periods
Non-optimal sample
dilution
Contamination of
reagents
Contents of wells
evaporate
Improper pipetting
Insufficient mixing of
reagent dilutions
 Pipette properly in a controlled and careful manner.
 Pipette properly in a controlled and careful manner.
 Check pipette calibration.
 Check pipette for proper performance.
 Thoroughly agitate the lyophilized components after
reconstitution.
 Thoroughly mix dilutions.
 Check the microplate pouch for proper sealing.
 Check that the microplate pouch has no punctures.
 Check that three desiccants are inside the microplate
pouch prior to sealing.
 Each step of the procedure should be performed
uninterrupted.
 Consult the provided procedure for complete list of steps.
 Consult the provided procedure for the correct order.
 Check pipette calibration.
 Check pipette for proper performance.
 Consult the provided procedure for all wash steps.
 Check that the correct wash buffer is being used.
 Consult reagent preparation section for the correct
dilutions of all reagents.
 Consult the provided procedure for correct incubation
time.
 Sandwich ELISA: If samples generate OD values higher
than the highest standard point (P1), dilute samples
further and repeat the assay.
 Competitive ELISA: If samples generate OD values lower
than the highest standard point (P1), dilute samples
further and repeat the assay.
 User should determine the optimal dilution factor for
samples.
 A new tip must be used for each addition of different
samples or reagents during the assay procedure.
 Verify that the sealing film is firmly in place before placing
the assay in the incubator or at room temperature.
 Pipette properly in a controlled and careful manner.
 Check pipette calibration.
 Check pipette for proper performance.
 Thoroughly agitate the lyophilized components after
reconstitution.
 Thoroughly mix dilutions.
9
References
(1)
(2)
(3)
Davie, E.W. et al. (1979) Adv. Enzyme. 48:277
Bajaj, S.P. et al. (1981) J. Biol. Chem. 256:253
Kisiel, W. et al. (1975) Biochemistry 14:4928
Version 8.0R
www.assaypro.com • e-mail: [email protected]
10