Download Specs - TRS RenTelco
Transcript
6-2 Buffer Model 6485 Picoammeter Instruction Manual Buffer operations Model 6485 has a buffer to store from one to 2500 readings. It also stores overflow readings. Each reading has a timestamp. The timestamp for each reading is referenced to the time the measure/store process is started. In addition, recalled data includes statistical information (maximum, minimum, peak-to-peak, average and standard deviation). The buffer fills with the specified number of readings and stops. Readings are placed in the buffer after any filters and/or math operations have been performed. Math operations include relative, mX+b, m/X+b, LOG, or limits. Buffered data is overwritten each time the storage operation is selected. The data is volatile — it is not saved through a power cycle. Measurement function changes are permissible during the storage process. Note however, that the statistics will be based on the readings of the different measurement functions. Store Perform the following steps to store readings: 1. 2. 3. 4. 5. NOTE Set up the instrument for the desired configuration. Press CONFIG (CONFIGURE: will be displayed). Press STORE. The present buffer size (in readings) is displayed. Use the cursor keys and , and the RANGE keys and to set the number of readings to store (1 to 2500). Press ENTER to save the buffer size. Press the STORE key. If in the immediate trigger mode, the storage process will start immediately. If in the external input trigger mode, each input trigger (or press of TRIG key) will store a reading. See Section 7 for information on triggering. The asterisk (*) annunciator turns on to indicate that the data storage operation is enabled. It will turn off when the storage process is finished (buffer full).