15
Future work
With the purpose of studying a possible relationship between their fine structures, the threshold and DPOAE were screened. Measurements of the hearing threshold require of concentration of the subjects and therefore, the time
duration of this test should be kept short. This limits the frequency range of
the measurement. However, measurements in a wider frequency range will
allow a more complete comparison of fine structures, since both threshold
and DPOAE are very frequency dependent. Hence, the design of a test procedure that allows a longer measurements time without biasing the subject’s
performance could be studied.
Although the implemented high resolution audiometer is a fast and precise
method for tracking threshold fine structure, it is not accurate on the estimation of absolute threshold level. Therefore, it could be interesting to
analyze and implement other high resolution audiometric methods and test
their performance for screening hearing thresholds.
The screening of DPOAE is highly dependent on the choices of measurement
parameters. It has been reported in the literature that the choice of the
primary frequency ratio and levels has a significant influence on the DPOAE
level and the presence of fine structure. The choices made can even produce
a shifting of the fine structure minima and maxima along the frequency axis
in a DP-gram. This constitutes an obvious limitation to use DPOAE as an
objective audiometry and therefore an appropriate line of research.
Similarities between the threshold and DPOAE fine structure periodicity are
observed in the experimental results, thus supporting the analytical models.
However, the models present certain restrictions to explain the generation
mechanisms responsible of the threshold and DPOAE fine structure. Further investigation of the different parameters involved in the simulation of
the cochlear behavior could lead to a more realistic relation between measurements and cochlear models.
70