Download User Manual
Transcript
CS99 Series User Manual Allwin Instrument the parameters set as follows: Test step: The current test step number; range: 1 ~ 30. When testing step is greater than one step, you can press the "shortcut" key to insert, delete, forward, backward and exchange, specifically refer to 7.4.4 test steps. Test mode: The current test procedure test mode; selected as ACW test mode. Test voltage: the current test voltage; ACW current range, current upper-limit and low-limit parameter setting interface is shown below: Current range: the current test step current range; Current upper-limit: current upper-limit alarm value of the test step; current upper-limit alarm range set value based on current gear. As follows: 200μA , the upper-limit: 0.1 ~ 200.0μA; 2mA , upper limit: 0.001 ~ 2.000mA; 20mA , the upper-limit: 0.01 ~ 20.00mA; 50mA , the upper-limit: 0.01 ~ 50.00mA, 100mA, the upper-limit: 0.1 ~ 999.9mA Current low-limit: current low-limit alarm value of the test step; current low-limit alarm range: 0 to the current upper-limit alarm, when the current lower limit alarm value is set to 0, the lower limit alarm judgment is off. ACW real current, arc detection, the output frequency parameter setting interface is shown below: - 51 -