START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
1
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
2
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
The products described in this manual are subject to continuous development and improvement and it is therefore
acknowledged that this manual may contain errors or omissions. EiUK Eurotron Instruments (UK) ltd encourage
customer feedback and welcome any comments or suggestions relating to the product or documentation. These
should be forwarded to the Customer Feedback Department at the address given below.
Eurotron Instruments (UK) ltd
Unit 13 Riley Close,
Royal Oak Industrial Estate,
Daventry, NN11 8QT, England
Tel: +44 (0)1327 871044
Fax: +44 (0)1327 301255
Email: [email protected]
www.eurotronuk.com
This manual is intended as a guide to use this product.
Eurotron Instruments (UK) ltd shall not be liable for any loss or damage whatsoever arising from the wrong comment / interpretation of information’s from this manual or any misuse come out of this manual.
Attention!
Inspect shipments in the presence of deliverer. If necessary remove packaging material and have damages to packaging and goods confirmed on the packing slip. Any such notice must be received by EiUK Eurotron Instruments
(UK) ltd within 3 days upon receipt of package.
Otherwise they could not be admitted!
Save the original box and the packing material to protect in case you have to ship or transport it.
3
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
1
Content
1
Content ............................................................................................................................................................. 4
2
Measuring principle........................................................................................................................................... 7
3
4
2.1
Electrochemical measuring principle ........................................................................................................ 7
2.2
NDIR- measurement ................................................................................................................................. 8
2.3
Gas flow chart RASI 900 .......................................................................................................................... 9
2.4
Gas flow chart RASI 901 ........................................................................................................................ 10
Device views.................................................................................................................................................... 11
3.1
Right side view RASI 900 ...................................................................................................................... 11
3.2
Left side view RASI 900.......................................................................................................................... 11
3.3
Right side view RASI 901 ...................................................................................................................... 12
3.4
Left side view RASI 901.......................................................................................................................... 12
3.5
Condensate separator RASI 900............................................................................................................. 13
Prepare for measurement................................................................................................................................ 14
4.1
Printer .................................................................................................................................................... 15
4.1.1 Printing procedure ............................................................................................................................. 15
4.1.2 Handling the printer........................................................................................................................... 15
4.1.2.1
How to open the cover assembly.............................................................................................. 15
4.1.2.2
How to load paper rolls ............................................................................................................ 15
4.1.2.3
How to close the cover correctly and cut the paper.................................................................. 15
5
6
7
Base of calculation .......................................................................................................................................... 16
5.1
Analysis and calculation......................................................................................................................... 16
5.2
Combustion calculations ........................................................................................................................ 16
Technical specifications................................................................................................................................... 17
6.1
RASI 900 ................................................................................................................................................ 17
6.2
RASI 901 ................................................................................................................................................ 18
Probes RASI 900.............................................................................................................................................. 21
7.1
Probe unheated...................................................................................................................................... 21
7.2
Probe unheated with sintered metal filter .............................................................................................. 22
7.2.1 Change of the sintered metal filter .................................................................................................... 23
7.2.2 Handling of the quick hook-up connector.......................................................................................... 23
7.3
Probe with heated glass wool filter........................................................................................................ 24
7.3.1 Replace glass wool filter elements of industrial gas sampling probe................................................. 25
7.3.2 Probe with heated sintered metal tube filled with glass wool material (option)................................ 25
8
Probes RASI 901.............................................................................................................................................. 26
8.1
Probe unheated...................................................................................................................................... 26
8.1.1 Handling of the quick hook-up connector.......................................................................................... 27
8.2
Probe unheated with sintered metal filter .............................................................................................. 28
8.2.1 Change of the sintered metal filter .................................................................................................... 29
8.3
Probe with heated glass wool filter........................................................................................................ 30
8.3.1 Replace glass wool filter elements of industrial gas sampling probe................................................. 31
8.3.2 Probe with heated sintered metal tube filled with glass wool material (option)................................ 32
4
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.4
Probe with heated glass wool filter and heated sample line .................................................................. 33
8.4.1 Connection of the heated hose.......................................................................................................... 34
8.4.2 Handling of the quick hook-up connector.......................................................................................... 34
9
Default ............................................................................................................................................................ 35
9.1
10
11
Error diagnosis regarding the analyser................................................................................................... 35
Maintenance ................................................................................................................................................... 36
10.1
Insideview .............................................................................................................................................. 36
10.2
Sample gas filter (RASI 900)................................................................................................................... 37
10.3
Sample gas filter / PTFE (RASI 901) ........................................................................................................ 37
10.4
Servicing fine filter # 56066 ................................................................................................................... 38
10.5
NOx protection filter (PURAFIL).............................................................................................................. 40
10.6
Servicing Acrodisc # 51513 (option)....................................................................................................... 40
10.7
Change the mains fuses ......................................................................................................................... 41
10.8
Change the lithium battery for sensor BIAS voltage supply.................................................................... 42
10.9
Servicing gas pump # 55513 .................................................................................................................. 43
10.10
Servicing CO purging pump # 54151 ................................................................................................. 45
10.11
Servicing CO purge solenoid valve # 59687....................................................................................... 46
10.12
Replace O2 sensor # 58048A............................................................................................................. 47
10.13
Dismount the gas cooler (RASI 901) .................................................................................................. 48
Adjustment...................................................................................................................................................... 53
11.1
Adjust temperatures............................................................................................................................... 53
11.1.1
Adjust offset of the thermocouple T-air amplifier.......................................................................... 54
11.1.2
Adjust offset of the thermo couple T-gas amplifier ....................................................................... 54
11.1.3
Adjust span of the thermo couple T air amplifier .......................................................................... 54
11.1.4
Adjust span of the thermo couple T gas amplifier......................................................................... 54
11.1.5
Adjust span of the temperature reference..................................................................................... 55
11.1.6
Adjust span of the temperature reference T-air amplifier.............................................................. 55
11.1.6.1 Adjust span of the temperature reference T-gas amplifier........................................................ 55
11.2
Adjust Pressure....................................................................................................................................... 56
11.2.1
Adjustment of the lower range ..................................................................................................... 56
11.2.2
Adjustment of the upper range ..................................................................................................... 56
11.3
AUX channel adjustment........................................................................................................................ 57
11.3.1
Adjustment thermocouple AUX amplifier...................................................................................... 58
11.3.2
Adjustment 0-10V AUX ................................................................................................................. 58
11.3.3
Adjustment 4-20mA ...................................................................................................................... 58
11.4
Analog output (4 - 20mA) ...................................................................................................................... 59
12
Service maintenance plan................................................................................................................................ 60
13
Self diagnosis .................................................................................................................................................. 60
13.1
Sensors: ( to fresh air ) ........................................................................................................................... 60
13.2
Battery voltage....................................................................................................................................... 60
13.3
Temperatures ......................................................................................................................................... 61
13.4
Draft....................................................................................................................................................... 61
13.5
Bias-voltage ........................................................................................................................................... 61
5
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14
13.6
KTY-chart ............................................................................................................................................... 61
13.7
PT2000-chart.......................................................................................................................................... 61
Appendix ......................................................................................................................................................... 62
14.1
Manual for firmware update of H8-Industrial analyzers......................................................................... 62
14.1.1
System requirements..................................................................................................................... 62
14.1.2
Important hints and warnings ....................................................................................................... 62
14.1.3
Update procedure ......................................................................................................................... 62
14.1.4
Possible causes and solutions in case of update failure: ............................................................... 62
14.2
Electrical connections............................................................................................................................. 63
14.2.1
Analog output ............................................................................................................................... 63
14.2.2
RS 232 connector .......................................................................................................................... 63
14.2.3
RS 485 connector .......................................................................................................................... 63
14.2.4
RS 485 converter connection to PC ............................................................................................... 64
14.2.5
RS 232 connection to PC ............................................................................................................... 66
14.3
Transmission protocol ............................................................................................................................ 67
14.3.1
Online data transmission protocol ................................................................................................ 67
14.3.2
data storage transmission protocol ............................................................................................... 68
14.4
Display messages ................................................................................................................................... 69
14.5
Lists of countries and fuel types ............................................................................................................. 72
14.5.1
General information ...................................................................................................................... 72
14.5.1.1 CO2, Lambda, losses, efficiency................................................................................................ 72
14.5.1.2 Dew point ................................................................................................................................. 72
14.5.1.3 CO- conversion of CO [ppm] ..................................................................................................... 72
14.5.1.4 NO-conversion of NO [ppm] ..................................................................................................... 73
14.5.1.5 NO2-conversion of NO2 [ppm] ................................................................................................. 73
14.5.1.6 NOx-conversion of NO [ppm] (possible also of NO2 [ppm]) ..................................................... 73
14.5.2
Fuel type list Great Britain............................................................................................................. 74
6
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
2
Measuring principle
The measurement principle of toxic gases is based on the electrochemical principle or optionally on infrared bench
(NDIR).
Temperatures are measured by means of thermocouples Type K (NiCrNi) , optionally type S ( PtRhPt) for very high
temperature probe and Platinum RTD PT 2000.
2.1
Electrochemical measuring
measuring principle
The oxygen content of the sample gas will be measured with a 2 electrodes electrochemical sensor.
Toxic gases like carbon monoxide CO, nitrogen oxide NO, nitrogen dioxide NO2, sulfur dioxide SO2 ,hydrogen sulphide H2S will be measured with 3-electrodes sensors.
The electrochemical sensors are based on the gas diffusion technology.
The advantages of this technology are: The permanently generated signal is direct proportionally and linear to the
volume concentration (% or ppm) of the analysis gas components.
The 3 electrodes are: S (sensing electrode), C (counter electrode) and R (reference electrode).
Pos.
1
2
3
4
5
6
7
8
9
Messgas
Staubfilter
S-Elektrode
R-Elektrode
C-Elektrode
Anschlusspin
Elektrolytspeicher
Elektrolyt
Kapillarmembrane
Sample gas
Particle filter
Sensing electrode
Reference electrode
Counter electrode
Connecting pin
Electrolyte reservoir
Bilk electrolyte
Capillary diffusion barrier
The gases reaches the sensing electrode reacts on the electrode surface either through oxidation ( for example CO,
H2S, SO2, NO, H2 ) or reduction (like CO, H2S, SO2, NO, H2 )
Example: CO sensor:
CO reacts at the sensing electrode as follows:
CO + H2O - > CO2 + 2H+ + 2eotherwise: at the counter electrode oxygen from air will be re-oxidize to water:
½ O2 + 2H+ + 2e
2 e - - > H 2O
The generated current (µA-range) will be measured and analysed by the microprocessor.
7
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
2.2
NDIRNDIR- measurement
By means of the non-disperse infrared bench of the RASI 901, it is possible to measure carbon monoxide (CO),
carbon dioxide (CO2) and hydro-carbons (HC).
Optional it is possible to measure sulphur dioxide SO2 in higher ranges as it is possible with the electrochemical SO2
sensor.
Principle of the IRIR- bench
A regulated infrared source (IR) provides a photon stream in the range of 2 to 8 microns (µm). The infrared light is
directed through the sample cell to the optical detector. The source is electronically modulated with 2Hz ( no chopper wheels, no moving parts).
Infrared transparent calcium fluoride lens are provided at each end of the gas sample cell. The pyro-electric detector is temperature regulated. Additionally a measurement compensation for the sample gas temperature and pressure variations is provided.
The residual infrared energy, not absorbed by the sample gas, is guided to the optical detector. It passes four optical narrow band pass filter that are each aligned on the infrared absorption spectrum of the target gases (one filter
is the reference filter; it is used for ageing and soiling compensation).
Infrared light passed through the filter is collected by four pyro-electric detectors, which share the same housing.
They produce a voltage that is proportional to light intensity. The detector block temperature measurement and
regulation facilitates compensation for IR filter and IR detector temperature variances. The detector out-put is preamplified and sent via a multiplexer to an analog or to a digital converter (ADC) and to the u-controller.
8
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
2.3
Gas flow chart RASI 900
9
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
2.4
Gas flow chart RASI 901
90 1
10
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
3
3.1
Device views
Right side view RASI 900
Position
1
2
3
3
5
6
3.2
Thermoprinter
RS 232 interface
(8x) Analog outputs 4 – 20 mA (option)
Mains fuse ( 2x 3,16A slow)
Universal power supply 90 – 240 Vac
SD Card (option)
Left side view RASI 900
Position
1
2
3
4
5
6
7
8
11
Connector combustion air temperature
Condensate outlet
Sample gas inlet
Connector heated probe
Differential pressure port P Differential pressure port P +
Condensate seperator
Gasfilter
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
3.3
Right side view RASI 901
901
Position
1
2
3
4
5
6
7
8
9
3.4
Thermoprinter
RS 232 interface
(8x) Analog outputs 4 – 20 mA (option)
Mains fuse ( 2x 3,16A slow)
Universal power supply 90 – 240 Vac
SD Card (option)
External keyboard
External 12 Vdc power supply
RS 485
Left side view RASI 901
90 1
Position
1
2
3
4
5
6
7
8
9
10
12
Connector combustion air temperature
Condensate outlet
Sample gas inlet
Connector heated probe
Differential pressure port P Differential pressure port P +
not connectred
Ventilation gas cooler
AUX connector
Filter (PTFE)
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
3.5
Condensate separator RASI 900
Pos.
1
2
3
4
5
6
7
8
9
10
11
12
13
#
56439
11165
51517
62279
62269
56124
51516
59169
62278
51317
51516
52359
cover
filter
O-ring
O-ring
fitting
center part
2 x O-ring
high-grade steel pipe
high-grade steel pipe
condensate vessel
O-ring
plug
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
4
Prepare for measurement
It is necessary to check before even measurement the condensate separator. The separator must be clean
(without
(without condensate) and the filter has to be white coloured!
coloured! If the filter is contaminated, it changes the
colour from white to grey/brown and is has be washed up to 5 times (e.g. ultrasonic bath) or replaced!
(Start-up with other available probes see § 7)
• Take out analyser from the transport case
• Mount gas sample probe and connect it to the analyser.
Connect RASI to mains power supply
It can be power supplied by:
1. 90-240 Vac mains power supply, 100 W, with mains power cable
2. Internal battery 12V, 1800mAh
and switch on by pressing
• Zero calibration is carried out for at least 2 minutes, max. 4 minutes. During this period of time, the measuring
cells of the analyser are purged with ambient air. Flue gas measurement cannot be started. All other menu
items can be selected.
• If the option NDIR bench is installed, you have to consider the warm-up time of the bench
• After the zero calibration the RASI is ready to carry out flue gas measurements
Long term nonnon- operating and storage:
1. charge battery every 4 weeks
2. store in a dry place
14
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
4.1
4.1.1
Printer
Printing procedure
The printing procedure can be started in the menus:
Measurement, View Last Values, Save Measurement, View Stored Data, Service or Adjustment.
Press printer key in the according menu
4.1.2
4.1.2.1
Handling
Handling the printer
How to open the cover assembly
Lift the lever, acting as indicated by the arrow, until the cover is released from its locking position. To avoid damages to the lever, do not use excessive force
4.1.2.2 How to load paper rolls
4.1.2.3
How to close
close the cover correctly and cut the paper
Press on both sides of the cover group (simultaneously) (A).
Alternatively: Press on the middlle of cover group
(near the paper exit) (B).
Do not close the cover by pressing only one
side!
How to cut the paper:
Pull the paper towards the tear bar from one side
to the other.
15
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
5
Base of calculation
5.1
Analysis and calculation
Continuously measured items
Unit
Stan
Standard
O2
[%]
X
CO
[ppm]
X
NO
[ppm]
option
NO2
[ppm]
option
SO2
[ppm]
option
H 2S
[ppm]
option
Air temperature (thermocouple)
[°C]
X
Flue gas temperature (thermocouple)
[°C]
X
Draft
[hPa] and others
option
Differential pressure
[hPa] and others
option
Calculations:
Continuous transformation
[ ppm ]
X
[ ppm ] in reference to O2 reference value
X
3
X
3
X
[ mg/m ]
[ mg/m ] in reference to O2 reference value
Other calculated values
Unit
CO2
[%]
Efficiency (ETA) *
[%]
Losses *
[%]
Excess air
-
Dew point
[°C]
PI = Poison index (ratio from CO to CO2)
[%]
Losses and efficiency are calculated by means of net calorific value. These values are than referenced for the gross
calorific value for condensing boilers only. (efficiency > 100)
Formula for calculation poison index :
PI [%] = 100 x CO [ppm] / CO2 [ppm]
5.2
Combustion calculations
The calculation of efficiency and the exhaust losses is carried out according to the formula of Siegert:
(see chapter 14.5.1)
14.5.1
16
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
6
6.1
Technical specifications
RASI 900
electrochemical cell
O2
Range
Accuracy
CO (H2 comp.)
Range
Accuracy
NO [Option # 59699]
Range
Accuracy
NO2 [Option # 59731]
Range
Accuracy
SO2 [Option # 59732]
Range
Accuracy
H2S [Option # 59730]
Range
Accuracy
with NDIR MultiMulti -gas bench
CO
Range
Accuracy
CO2
Range
Accuracy
CXHY equiv CH4
Range
Accuracy
with NDIR MultiMulti -gas bench
CO
Range
Accuracy
CO2
Range
Accuracy
CXHY equiv C3H8
Range
Accuracy
temperaturee TA
Stack gas temperatur
Range
Range
Range
Accuracy
0 - 21,0 Vol-%
± 0,2 Vol-% abs.
0 - 2.000 ppm, overload up to 4.000 ppm
± 10 ppm or 5 % reading 4.000 ppm
10 % reading > 4.000 ppm
0 - 1.000 ppm, overload up to 5.000 ppm
± 5 ppm or 5 % reading 1.000 ppm
10 % reading > 1.000 ppm
0 - 200 ppm, overload up to 1.000 ppm
± 5 ppm or 5 % reading 200 ppm
10 % reading > 200 ppm
0 - 2.000 ppm, overload up to 5.000 ppm
± 10 ppm or 5 % reading 2000 ppm
10 % reading > 2000 ppm
0 - 50 ppm, overload up to 500 ppm
± 5 ppm or 5 % reading 50 ppm
10 % reading > 50 ppm
# 58118, 65 mm, CO, CO2, CH4
min. 0 - 3 % up to max.0 - 10 %
± 0,03 % or 5 % reading
min. 0 - 3 % up to max. 0 - 30 %
± 0,5 % or 5 % reading
min. 0 - 1 % up to max. 0 - 3 % equiv. CH4
± 0,01 % or 5 % reading
# 58986, 65 mm, CO, CO2, C3H8
min. 0 - 3 % up to max.0 - 10 %
± 0,03 % or 5 % reading
min. 0 - 3 % up to max. 0 - 30 %
± 0,5 % or 5 % reading
min. 0 - 2.000 ppm up to max. 0 - 5.000 ppm equiv. C3H8
± 30 ppm or 5 % reading
0 - 650 °C by means of stainless steel sampling tube
0 - 1.100 °C by means of Inconel steel sampling tube
0 - 1.750 °C by means of ceramic sampling tube
± 2 °C 200 °C
17
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Combustion air temperature T L
Range
Accuracy
Draft
Range
Accuracy
Differential pressure (option)
Range
Accuracy
Flow
Range
Accuraccy
0 - 100 °C
± 1 °C
± 100 hPa
± 0,02 hPa or 1% reading
± 100 hPa
± 0,02 hPa or 1% reading
1 m/sec up to 100 m/sec
± 1 m/sec or 1 % reading
Suction capacity
capa city of gas pump
150 hPa
Calculated values
(depends on fuel type)
CO2
Range
Accuracy
Dewpoint
Losses qA
Efficiency ή
0 - CO2 max
± 0,3 Vol-% abs.
°C
0 - 99,9 %
0 - 120 %
Measured values as
mg/Nm3, O2 referenced, NOx as mg/Nm3NO2, mg/sec
General specifications
Operation temperature
Storage temperture
Power supply
Protection class
Weight
Dimensions
+ 5 - + 45 °C, max. 95 % humidity, not condensing
- 20 - + 50 °C
Universal grid power 100 - 250 Vac, 47 - 63 Hz
12 V / 3,6 Ah internal battery, operation 8 hour
IP 21
approx. 7kg (without case and accessories)
530 x 490 x 310 mm
Technical changes without notice!
6.2
Date of issue 20110530
RASI 901
90 1
Measured
Measu
red values
electrochemical cell
O2
Range
Accuracy
CO (H2 comp.)
Range
Accuracy
NO [Option # 59699]
Range
Accuracy
NO2 [Option # 59731]
Range
Accuracy
0 - 21,0 Vol-%
± 0,2 Vol-% abs.
0 - 2.000 ppm, overload up to 4.000 ppm
± 10 ppm or 5 % reading 4.000 ppm
10 % reading > 4.000 ppm
0 - 1.000 ppm, overload up to 5.000 ppm
± 5 ppm or 5 % reading 1.000 ppm
10 % reading > 1.000 ppm
0 - 200 ppm, overload up to 1.000 ppm
± 5 ppm or 5 % reading 200 ppm
18
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10 % reading > 200 ppm
SO2 [Option # 59732]
Range
Accuracy
H2S [Option # 59730]
Range
Accuracy
H2 [Option # 60270]
Range
Accuracy
CO very high [Option # 59939]
Range
Accuracy
with NDIR MultiMulti -gas bench
CO
Range
Accuracy
CO2
Range
Accuracy
CXHY equiv CH4
Range
Accuracy
with NDIR MultiMulti -gas bench
CO
Range
Accuracy
CO2
Range
Accuracy
CXHY equiv CH4
Range
Accuracy
with NDIR MultiMulti -gas bench
CO
Range
Accuracy
CO2
Range
Accuracy
CXHY equiv C3H8
Range
Accuracy
Stack gas temperature TA
Range
Range
Range
Accuracy
0 - 2.000 ppm, overload up to 5.000 ppm
± 10 ppm or 5 % reading 2000 ppm
10 % reading > 2000 ppm
0 - 50 ppm, overload up to 500 ppm
± 5 ppm or 5 % reading 50 ppm
10 % reading > 50 ppm
0 - 1 %, overload up to 2 %
± 0,02 % or 5 % reading 1 %
10 % reading > 1 %
0 - 4,00 %, overload up to 10,00 %
± 0,02 % or 5 % reading 4 %
10 % reading > 4 %
# 61447, 103 mm, CO, CO2, CH4
min. 0 - 10.000 ppm up to max.0 - 30.000 ppm
± 40 ppm or 5 % reading
min. 0 - 3 % up to max. 0 -20 %
± 0,5 % or 5 % reading
min. 0 - 10.000 ppm up to max. 0 - 30.000 ppm equiv. CH4
± 60 ppm or 5 % reading
# 58118, 65 mm, CO, CO2, CH4
min. 0 - 3 % up to max.0 - 10 %
± 0,03 % or 5 % reading
min. 0 - 3 % up to max. 0 - 30 %
± 0,5 % or 5 % reading
min. 0 - 1 % up to max. 0 - 3 % equiv. CH4
± 0,01 % or 5 % reading
# 58986, 65 mm, CO, CO2, C3H8
min. 0 - 3 % up to max.0 - 10 %
± 0,03 % or 5 % reading
min. 0 - 3 % up to max. 0 - 30 %
± 0,5 % or 5 % reading
min. 0 - 2.000 ppm up to max. 0 - 5.000 ppm equiv. C3H8
± 30 ppm or 5 % reading
0 - 650 °C by means of stainless steel sampling tube
0 - 1.100 °C by means of Inconel steel sampling tube
0 - 1.750 °C by means of ceramic sampling tube
± 2 °C 200 °C
1 % reading > 200 °C
19
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Combustion air temperature T L
Range
Accuracy
Draft
Range
Accuracy
Differential pressure (option)
Range
Accuracy
± 100 hPa
± 0,02 hPa or 1% reading
Suction capacity of gas pump
150 hPa
Flow
Range
Accuraccy
1 m/sec up to 100 m/sec
± 1 m/sec or 1 % reading
Calculated values
(depends on fuel type)
CO2
Range
Accuracy
Dewpoint
Losses qA
Efficiency ή
0 - CO2 max
± 0,3 Vol-% abs.
°C
0 - 99,9 %
0 - 120 %
Measured values as
mg/Nm3, O2 referenced, NOx as mg/Nm3NO2, mg/sec
0 - 100 °C
± 1 °C
± 100 hPa
± 0,02 hPa or 1% reading
General specifications
Operation temperature
Storage temperture
Power supply
Protection class
Weight
Dimensions
+ 5 - + 45 °C, max. 95 % humidity, not condensing
- 20 - + 50 °C
Universal grid power 100 - 250 Vac, 47 - 63 Hz
12 V / 3,6 Ah internal battery, operation 2 hour
IP 21
approx. 7kg (without case and accessories)
530 x 490 x 310 mm
Technical changes without notice!
Date of issue 2010530
20
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
7
Probes RASI 900
7.1
Probe unheated
! Without filter elements, to be used only for clean combustion site, e.g. natural gas!
Gas sample probe for low dust site, with exchangeable probe tube, continuous stack draft measurement, gas temperature measurement up to 650°C ( up to 1100°C with Inconel tube)
Item
Probe handle with
Probe handle with
Description
Temperature
range
2,7 m unheated sample line
5,0 m unheated sample line
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
300 mm, Ø 10 mm
750 mm, Ø 10 mm
1.000 mm, Ø 10 mm
1.500 mm, Ø 10 mm
2.000 mm, Ø 10 mm
Probe tube
Probe tube
Inconel 1.000 mm, Ø 10 mm
Inconel 1.500 mm, Ø 10 mm
21
#
62222
62223
up to 650°C
up to 650°C
up to 650°C
up to 650°C
up to 650°C
55671
55672
55673
55674
55464
up to 1.100°C
up to 1.100°C
56737
56738
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
7.2
Probe unheated with sintered metal filter
Gas sample probe for increased dust site, with exchangeable probe tube, continuous stack draft measurement, gas
temperature measurement.
Item
Probe handle with
Probe handle with
Probe tube
Probe tube
Probe tube
Description
Description
Temperature
range
2,7 m unheated sample line
5,0 m unheated sample line
750 mm, Ø 10 mm with sintered metal filter # 56748
1.000 mm, Ø 10 mm with sintered metal filter # 56748
1.500 mm, Ø 10 mm with sintered metal filter # 56748
22
#
59774
59778
up to 500°C
up to 500°C
up to 500°C
60813
60814
60815
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
7.2.1 Change of the sintered metal filter
The sintered metal filter should be controlled daily (dependent on the dusty condition in the exhaust gas) around
exact results of measurement to attain. If the filter is dirty, it can be cleaned with the help of an ultrasonic bath
(unlimited re-use of the cleaned filter)
Unscrew dirty filter!
Wear protective gloves!
Filter element with compressed air (1) the inside the
filter blow out or in an ultrasonic bath clean,
Do not damage the thermocouple (2) when from/screwing the filter elements in!
7.2.2 Handling of the quick hookhook- up connector
Connect quick connector from hose end by pushing it to gas inlet port of analyser.
To remove the probe, press the metal plate of the quick connector and pull it out.
23
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
7.3
Probe with heated glass wool filter
Industrial gas sample probe for high dust site, with heated, easy replaceable glass wool filter, exchangeable probe
tube, (no stack draft measurement integrated) gas temperature measurement up to 650°C
(up to 1100°C with Inconel tube).
Stack draft measurement possible with option “Stack draft probe with 3 m silicone hose ”
Art-No. 59798)
Item
Probe handle with
Probe handle with
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Description
Temperature
range
2,7 m unheated sample line
5,0 m unheated sample line
300 mm, Ø 12 mm
750 mm, Ø 12 mm
1.000 mm, Ø 12 mm
1.500 mm, Ø 12 mm
2.000 mm, Ø 12 mm
Art. No.
62224
62225
up to 650°C
up to 650°C
up to 650°C
up to 650°C
up to 650°C
11620
11621
11622
11623
11624
Inconel, 1.000 mm, Ø 12 mm up to 1100°C
Inconel, 1.500 mm, Ø 12 mm up to 1100°C
Inconel, 2.000 mm, Ø 12 mm up to 1100°C
59552
59904
59622
24
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
7.3.1 Replace glass wool filter elements of industrial gas sampling probe
probe
(only for probes # 62224 and 62225)
If the filter element changes from white to grey/brown the glass wool must be replace.
Press the filter cartridge 3 mm against the spring, then turn left 45° (bayonet lock) Take out the glass wool by
means of a tweezers.
Use protection gloves, the exhausted glass wool can be acid-containing!
The probe handle can be hot!°
Of new glass wool approx. 2 g weight out, insert and into the filter cartridge by means of tweezers.
The drillings (1) of the filter cartridge must be on the catch side!
If the O-rings (2) # 61066 and / or (3) # 59799 are damaged, then these O-rings has to be changed!
The filter cartridge by means of press and 45° clockwise rotation fasten into the probe handle.
7.3.2 Probe with heated sintered metal tube filled with glass wool
wool material (option)
1.
2.
3.
4.
Probe handle
Filter holder
Filter
O-ring
Recommendation:
The sintered metal filter should be inspected every day (depending on the dust content of sample gas) to attain precise
measuring results. If the filter is dirty it can be cleaned by
means of an ultrasonic bath ( unlimited reusing).
25
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8
Probes RASI 901
901
8.1
Probe unheated
! Use only for clean combustion site, e.g. natural gas!
Gas sample probe for low dust site, with exchangeable probe tube, continuous stack draft measurement, gas temperature measurement up to 650°C ( up to 1.100°C with Inconel tube)
Item
Probe handle with
Probe handle with
Description
Temperature
range
2,7 m unheated sample line
5,0 m unheated sample line
59774
59778
Hose coupling
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Art.Art.- No.
60236
300 mm, Ø 10 mm
750 mm, Ø 10 mm
1.000 mm, Ø 10 mm
Inconel 1.000 mm, Ø 10 mm
1.500 mm, Ø 10 mm
2.000 mm, Ø 10 mm
26
up to 650°C
up to 650°C
up to 650°C
up to 1.100°C
up to 650°C
up to 650°C
55671
55672
55673
56737
55674
55464
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.1.1
Handling of the quick hookhook- up connector
Connect quick connector from hose end by pushing it to gas inlet port of analyser.
To remove the probe, press the metal plate of the quick connector and pull it out.
27
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.2
Probe unheated with sintered metal filter
Gas sample probe for increased dust site, with exchangeable probe tube, continuous stack draft measurement, gas
temperature measurement.
Item
Probe handle with
Probe handle with
Probe tube
Probe tube
Probe tube
Description
Temperature
range
2,7 m unheated sample line
5,0 m unheated sample line
750 mm, Ø 10 mm
1.000 mm, Ø 10 mm
1.500 mm, Ø 10 mm
28
Art. No.
59774
59778
up to 600°C
up to 600°C
up to 600°C
60813
60814
60815
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.2.1
Change of the sintered
sintered metal filter
The sintered metal filter should be controlled daily (dependent on the dusty condition in the exhaust gas) around
exact results of measurement to attain. If the filter is dirty, it can be cleaned with the help of an ultrasonic bath
(unlimited re-use of the cleaned filter)
Unscrew dirty filter!
Wear protective gloves!
Filter element with compressed air (1) the inside the
filter blow out or in an ultrasonic bath clean,
Do not damage the thermocouple (2) when from/screwing the filter elements in!
29
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.3
Probe with heated glass wool filter
Industrial gas sample probe for high dust site, with heated, easy replaceable glass wool filter, exchangeable probe
tube, (no stack draft measurement integrated) gas temperature measurement up to 650°C (up to 1.100°C with
Inconel tube and up to 1.700°C with ceramic tube).
Stack draft measurement possible with option “Stack draft probe with 3 m silicone hose ”
Art-No. 59798)
Item
Probe handle with
Probe handle with
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube with S-type
thermocouple
Description
Temperature
range
2,7 m unheated sample line
5,0 m unheated sample line
300 mm, Ø 12 mm up to 650°C
750 mm, Ø 12 mm up to 650°C
1.000 mm, Ø 12 mm up to 650°C
Inconel, 1.000 mm, Ø 12 mm up to 1100°C
1.500 mm, Ø 12 mm up to 650°C
2.000 mm, Ø 12 mm up to 650°C
Inconel, 2.000 mm, Ø 12 mm up to 1100°C
Ceramic, no temperature reading, up to 1.700°C
1.000 mm, Ø 10 mm
Ceramic tube , with temperature reading, up to 1.700°C
1.000 mm, Ø 10 mm
30
Art. No.
59782
59784
11620
11621
11622
59552
11623
11624
59622
59948
50863
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
NOTE:
Mounting of probe with ceramic sampling tube requires very slow insertion
(some 10 cm/min) of this tube into the duct or pipe.
This will avoid the thermal shocks on the fragile ceramic tube, which will cause its
mechanical dam
damage!
8.3.1
Replace glass
g lass wool filter elements of industrial gas sampling probe
(only for probe handles # 59782, 59784, 59785 and 60119)
If the filter element changes from white to grey/brown the glass wool must be replace.
Press the filter cartridge 3 mm against the spring, then turn left 45° (bayonet lock) Take out the glass wool by
means of a tweezers.
Use protection gloves, the exhausted glass wool can be acid-containing!
Consider safety data
sheet in the appendix!
The probe handle can be
hot!°
Of new glass wool approx. 2 g weight out, insert and into the filter cartridge by means of tweezers.
The drillings (1) of the filter cartridge must be on the catch side!
If the O-rings (2) # 61066 and / or (3) # 59799 are damaged, then these O-rings has to be changed!
The filter cartridge by means of press and 45° clockwise rotation fasten into the probe handle.
31
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.3.2
Probe with heated sintered metal tube filled with glass wool material (option)
5.
6.
7.
8.
Probe handle
Filter holder
Filter
O-ring
Recommendation:
The sintered metal filter should be inspected every day (depending on the dust content of sample gas) to attain
precise measuring results. If the filter is dirty it can be cleaned by means of an ultrasonic bath ( unlimited reusing ).
32
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.4
Probe with heated glass wool filter and heated sample line
Industrial gas sample probe for high dust site, with heated, easy replaceable glass wool filter, exchangeable probe
tube, gas temperature measurement and temperature regulated sample line.
Stack draft measurement possible with option “Stack draft probe with 3 m silicone hose ”
Art-No. 59798)
Denomination
Probe handle with
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Probe tube
Description
3,0 m temperature regulated,
heated gas sample line
5,0 m temperature regulated,
heated gas sample line
300 mm, Ø 12 mm
750 mm, Ø 12 mm
1.000 mm, Ø 12 mm
Inconel 1.000 mm, Ø 12 mm
1.500 mm, Ø 12 mm
2.000 mm, Ø 12 mm
Inconel, 2.000 mm, Ø 12 mm
33
TemperatureTemperature- Art.Art.- No.
range
59785
60119
up to 650°C
up to 650°C
up to 650°C
up to 1.100°C
up to 650°C
up to 650°C
up to 1.100°C
11620
11621
11622
59552
11623
11624
59622
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
8.4.1
Connection of the heated hose
Connect electrical connector from hose end to connector RASI 901.
Turn left or right the connector (1) until it snaps in the socket (2).
Than screw clockwise back nut (3)
8.4.2
Handling
Handling of the quick hookhook- up connector
Connect quick connector from hose end to gas inlet port of analyser, by withdrawing the ring (4) and release after
plugging (5).
34
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
9
Default
9.1
Error diagnosis regarding the analyser
1. Malfunction
Analyser cannot charge
the battery in spite of
connecting to mains
Analyser cannot be
switched off by pressing
the OFF key.
key
Inside of the analyser is
too cold, analyser not
ready for operation.
2. Error Indication
No mains indication in the
display
3. Cause
Cause
Fuse (3,15 A T) in the mains
socket is burned out
none
Analyser does not react on
any key..
Self test
Temperature indication:
- - - , - °C
Measuring values are not
correct
No measurement possible
Measurement without
Self test
exact temperature values.
Temperature indication:
- - - , - °C
Wrong measuring values. Measuring range exceeded:
Value O2 too high,
Values CO and CO2 to low.
No measurement possible Display indication “close
probe”
Wrong temperature
measuring values
Gas temperature is too hot or
alternates
35
4. Solution
Replace the fuse(s).
Disconnection of the main
power plug and press
simultaneous the keys F1
and F4.
e. g. analyser was stored at a Put the analyser to a warm
cold place during winter.
room and connect it to the
mains and wait for some
minutes.
Calibrate analyser .
Call our after-sales service
Analyser cannot be switched Connect the analyser to
on or does not react after
the mains in order to
being switched on.
charge the battery.
Battery discharge.
Thermo element defective,
Call our after-sales serbalancing network intervice.
rupted or not connected
Remove probe from the
gas duct and condensate
from the probe tube
Connection probe – analyser Effect tightness test!
not correct. Leakage at
By visual control of probe,
probe / tube /, pump does
tubes, leaking parts could
not suck correctly.
be found.
Probe tube too hot.
Let probe tube cool down,
Probe is not fixed respecrespectively fix probe
tively screwed correctly.
tight, measurement is
possible now.
Probe is not plugged in cor- Check probe plug respecrectly, defective cable in the tively probe line regarding
probe line, formation of
damages (loose conneccondensate at the probe tip. tion), remove condensate
from the probe tip.
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10 Maintenance
10.1
Insideview
Item
1
2
3
4
Description
not used
not used
not used
condensate separator with PTFE filter
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
not used
not used
dust filter
not used
sample gas pump
diff. pressure sensor
Acrodisc PTFE filter (option)
NDIR bench (option 1-gas or 4-gas)
CO purging pump
electrochemical sensors manifold
CO purge solenoid valve
COlow sensor
O2 sensor long-life
NO sensor (option)
NO2 sensor (option)
SO2 sensor (option)
H2S sensor (option) or CO high (option)
NOx protection filter
see chapter
3.5 and
10.2
10.2
10.9
10.6
10.10
10.11
10.12
10.5
36
remarks
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.2
Sample gas filter (R
(RASI 900)
900)
Replace the dust and particle filter of the analyser if the colour is changing from white to brown.
Solve the filter case and replace the filter cartridge.
Do not loose or forget the sealing ring.
The filter elements inside the analyser should be changed at the early service in our service agency.
10.3 Sample gas filter / PTFE (RASI
(RASI 901
90 1 )
Replace the dust and particle filter of the analyser if the colour is changing from white to brown.
Solve the filter case and replace the
filter cartridge.
Do not loose or forget the sealing ring.
OK
The filter elements inside the analyser should be changed at the early service in our service agency
37
DIRTY
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.4
Servicing fine filter # 56066
If it is necessary to replace the internal filter elements earlier than service interval, open the front panel of the
analyser (4 hexagon socket screws):
Take the fine filter (1,2 and 3) out of the clamp.
Screw up the filter housing and replace the filter pads.
The metal screens at the in- and output of the filter housing should be cleaned with compressed air.
After cleaning the metal screens and replacing the filter tabs, build in the filter and observe the marked flow direction.
Remove the hoses, NOTE: Consider installation direction of the filter (marking)!
38
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Filter housing screw on
and filter tablet take out
Take out the rougher filter (1) from of the gas outlet and finer filter (2) from the gas inlet side and clean them with
a brush or white spirits.
Insert cleaned filters, secure them with space washers and insert a new filter tablet.
Screw on the filter housing. NOTE: O-ring (3) do not damage!
If the filter housing or a connector is damaged, use new filter housing!
Replace the hoses.
Flow direction:
The Flow direction of the fine filter is marked with an arrow on the filters housing.
39
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.5
NOx protection filter (PURAFIL)
The NOx filter is to be renewed after discoloration of violet to dark-brown / black. The hoses if necessary clean or
replace
10.6
Servicing Acrodisc # 51513 (option)
Remove the hoses
Change Acrodiscfilter – replaced the hoses.
40
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.7
Change the mains fuses
If the analyser cannot charge the battery in spite of connecting to mains (no mains indication in the display ), than
the fuses ( 3,15 A slow) of mains connector is burn out and has to be replaced!
Note:
If the fuse will burn out a second time, it is necessary to look for the cause.
Hints:
1. The optional heated sample line has a short-cut.
2. The internal switched mode power supply has a short-cut
41
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.8
Change the lithium battery for sensor BIAS voltage supply
Note:
We recommend to let this job be done by a service department !
If it is necessary to replace the BIAS battery earlier than service interval, open the front panel of the analyser (4
hexagon socket screws).
Remove the 6 screws and the marked cable (1).
Lift the board with the soldered battery and pull it out of the connectors of the complete board.
Remove the battery by means of a dissolver pump and insert the new battery (# 11313).
After the change of the battery and assembling of the device, the U-BIAS values displayed during the self test
should be correct – approx. 3.000 mV.
42
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.9
Servicing gas pump # 55513
Remove the hoses, NOTE: Consider installation direction of the hoses!
Whether the service of this pump is possible in the inserted condition in RASI, depends on the options of the analyzer
If the disassembly of the hose pump is necessary,
unscrew the marked fixing screws. At the same time
fix the nuts/mothers with an open end wrench on
the lower surface of the mounting plate!
Unscrew the 4 marked screws by means of a TORX screwdriver size TX10 and remove the cover plate inclusive the
rubber sealing.
43
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Remove the fixing ring (2) of the valve membrane # 11212 (1) by means of a small screwdriver.
Renew the valve membrane or clean it with white spirit and replace the fixing ring.
Turn the pump head housing on the other side and renew the second valve membrane or clean it with white spirit.
In the case of different coating of the valve membrane, the smoothly side is inserted to the valve seat.
Set and on the pump attach the rubber seal with metal plate on the pump head housing. The four fixing bolts first
strongly tighten and afterwards the screws crosswise tighten, until that the rubber seal straight is not laterally outsqueezed.
Replaced the hoses.
44
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.10
Servicing CO purging pump # 54151
Diaphragma clean ob renew
45
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.11 Servicing CO purge solenoid valve # 59687
Clean channels
46
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.12 Replace O2 sensor # 58048A
ATTENTION: Sensors can contain acid. Pay attention to security indications for manipulation!
In case of a leakage of the sensor, wear protection gloves!
The condition of the electro-chemical sensors is checked at every auto-zero cycle. If the sensor is not ok, an alarm
message is indicated in the display and system alarm relay will be triggered. The measurement will continue but
without O2 measurement.
To find out whether a sensor is faulty or not, please pay attention to the chapter 13.1
"service-values"!!
Unlock connection cable of oxygen cell (# 58048), unscrew the O2-cell out of the adapter plate mounted on the
sensor manifold.
The O-ring sealing and the filter paper (# 56872) of the sensor manifold should be changed too!
47
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
10.13 Dismount the gas cooler (RASI
(RASI 901
90 1 )
1) Before you disassemble the gas cooler, you should
mark the two parts as shown!!
Screw out the four screws, by turning anti clock wise
48
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Screw out the four screws, by turning anti clock wise
Now you can pull up the white housing, it can happen, that the black cylinder will stay in the housing.
49
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Take out the cylinder and the O-Ring out of the
housing
Replace the housing if necessary,
Insert the O-Ring
Insert the cylinder into the housing, there is only one
way to fit the cylinder, because of the sensor wire….
Be careful not to break the KTY sensor!!
50
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
In case you just replace the white housing, you use
thermolube between the Peltier Element and the
cooler block and also between the Peltier Element
Aluminium block.
IN CASE THAT THERE IS NO THERMO-LUBE BETWEEN THESE TWO PARTS; THEN WE HAVE USED A
SPECIAL HEAT TRANSFER FOIL; THIS FOIL CAN BE
REUSED IF NOT DAMAGED; OR YOU CAN ORDER
NEW HEAT TRANSFER FOIL. (Part Number: 61345)
In case you just replace the white housing, you use
Thermolube between the Peltier Element and the
Cooler block and also between the Peltier Element
Aluminium block.
Please insure yourself, that you have placed the
Peltier Element the correct way. (the red and black
wire have to be placed in the same direction as seen
on the picture!!!)
IN CASE THAT THERE IS NO THERMO-LUBE BETWEEN THESE TWO PARTS; THEN WE HAVE USED A
SPECIAL HEAT TRANSFER FOIL; THIS FOIL CAN BE
REUSED IF NOT DAMAGED; OR YOU CAN ORDER
NEW HEAT TRANSFER FOIL. (Part Number: 61345)
51
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Just a thin layer of Thermolube
Please insure yourself, that you have placed the
Peltier Element the correct way. (the red and black
wire have to be placed in the same direction as seen
on the picture!!!)
IN CASE THAT THERE IS NO THERMO-LUBE BETWEEN THESE TWO PARTS; THEN WE HAVE USED A
SPECIAL HEAT TRANSFER FOIL; THIS FOIL CAN BE
REUSED IF NOT DAMAGED; OR YOU CAN ORDER
NEW HEAT TRANSFER FOIL. (Part Number: 61345)
Now you can assemble the parts again
(the red and black wire have to be placed in the
same direction as seen on the picture!!!)
You screw in the screws… try to get the same pressure on the springs at the same time.
After you have screwed all the 4 screws tight (until
the 4 springs are pressed together) you release the
screws again for 1,5 turns.
52
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
11 Adjustment
Adjustment
To select adjustment menu, please access main “Xtras” menu by pressing key
Scroll the cursor by means of
Press
,
keys and select the adjustment function.
key and analyzer will ask for the PIN code
The adjustment menu is protected against use of non-authorized persons by a PIN code.
At wrong input of PIN code, the analyser will return to main settings menu.
For PIN code ask Service department
Adjustment of electrochemical sensors and IRIR -bench see user man
manual chapter 8.5 .
11.1
Adjust temperatures
temperatures
Note: These adjustments are factory made, it is generally not necessary to repeat them!
Testtools:
1. Plug with thermo couple short-circuit and adjusted 2KOhm reference
2. 250°C emulator for thermo couple with 2kOhm reference cable
,
:
:
select the “Temperatures” menu by moving the cursor
enter selected menu (highlighted line)
53
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
In this menu it is possible to correct offset and span errors of the different thermocouple amplifiers:
o for combustion air measurement (T-air)
o for gas temperature measurement (T-gas)
o for cold junction compensation of thermocouples (T-ref)
o
11.1.1 Adjust offset of the thermocouple TT- air amplifier
1. Select T-air offset with the cursor
2. connect short-circuit / 2kOhm plug to the T-air socket of the analyzer
3. wait at least 5 minutes
4. press
- key = zero
5. the value “T- air span” must be 0.0 (± 0,1) now!
11.1.2 Adjust offset of the thermo couple TT-gas amplifier
1. Select T-gas offset with the cursor
2. connect short-circuit / 2kOhm plug to the T-gas socket of the analyzer
3. wait at least 5 minutes
4. press
- key = zero
5. the value “T- gas span” must be 0.0 (± 0,1) now!
11.1.3 Adjust span of the thermo couple T air amplifier
1. The offset of the thermocouple amplifier is already adjusted!
(see chapter 11.1.1 and 11.1.1
2. Select T-air span with the cursor
3. connect plug of the 250°C emulator short-circuit / 2kOhm plug to the T-air socket of the analyzer
4. Switch on emulator and set to 250°C
5. wait at least 5 minutes
6. modify the adjust factor by means of
,
or
keys, until the displayed value is 250°C (± 0,3°C)
11.1.4 Adjust span of the thermo couple T gas amplifier
1. The offset of the thermocouple must be adjusted before!
(see chapter 11.1.1 and 11.1.1
2. Select T-gas span with the cursor
3. connect plug of the 250°C emulator short-circuit / 2kOhm plug to the T-air socket of the analyzer
4. Switch on emulator and set to 250°C
5. wait at least 5 minutes
6. modify the adjust factor by means of
,
or
keys, until the displayed value is 250°C (± 0,3°C)
54
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
11.1.5 Adjust span of the temperature reference
Press
.key to access adjustment of T- Ref.
,
:
select T-air / T-gas span
11.1.6 Adjust span of the temperature reference
reference TT- air amplifier
1. Select T-air span with the cursor
2. connect short-circuit / 2kOhm plug or 250°C emulator to the T-air socket of the analyzer
3. wait at least 5 minutes
4. modify the adjust factor by means of
,
, until the displayed value is 0,0°C (± 0,1°C)
11.1.6.1 Adjust span of the temperature reference TT- gas amplifier
1. Select T-gas span with the cursor
2. connect short-circuit / 2kOhm plug or 250°C emulator to the T-gas socket of the analyzer
3. wait at least 5 minutes
4. modify the adjust factor by means of
,
, until the displayed value is 0,0°C (± 0,1°C)
Final check:
If all 3 steps described before are accomplished, then the measured values must show the following values:
With short-circuit 2kOhm plug:
With 250°C emulator :
0,0°C (± 0,2°C)
250,0°C (±0,5°C)
55
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
11.2
Adjust Pressure
Note: These adjustments are factory made, it is generally not necessary to repeat them!
:
activate pressure adjustment
:
end:
return to adjustment menu
Test tools:
Pressure calibrator
Manual calibration:
After setting the zero point at the pressure calibrator, the adjustment pressure is carried out in two steps:
:
zero:
set value to zero
11.2.1 Adjustment of the lower range
The pressure at the reference device will be changed up to approx. - 10 hPa,
hPa however not more as 20 hPa. Following the pressure factor of the display has to be modified by means of the
corresponds to the value of the reference device.
,
- keys until the displayed value
11.2.2 Adjustment of the upper range
The pressure of the reference device will be raised still the displayed value is approx. 90 hPa, however no less than
80 hPa.
If the displayed value is smaller, it is not possible to correct them.
Following the pressure factor of the display has to be modified by means of the
value corresponds to the value of the reference device.
:
end:
end
return to adjustment menu
56
,
- keys until the displayed
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
11.3
AUX channel adjustment
Note: These adjustments are factory made, it is generally not necessary to repeat them!
The analog input (AUX) allows user to data log an additional thermocouple or transmitter with voltage output
signal (0-10V) or current signal (0/4-20mA).
,
:
select the line “AUX-channel”
Adjustment menu
AUX channel
:
activate AUX channel adjustment
AUX input adjustment
,
:
select the item
,
:
set the factor of the selected item
:
zero:
set value to zero
.:
not active
:
modify: modify the values
:
not active
:
end:
return to adjustment menu
57
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
11.3.1 Adjustment thermocouple AUX amplifier
Connect to AUX socket the calibrator, turn switch for zeroing ( pos. 1),
select line “ AUX offset” and press key
to set to zero the amplifiers offset.
Turn switch to position 2 (thermocouple voltage) , select line “T-AUX span” and adjust the span factor of thermocouple amplifier with keys
,
. Temperature value must show:
+250°C for 10,151 mV ( K type thermocouple )
+500°C for 20,64 mV ( K type thermocouple )
Select line “ T-ref span” and compare value of the cold junction compensating sensors for the thermocouples with
a reference temperature device ( e.g. a precision glass thermometer).
Adjust the factor to have the same temperature displayed as the reference thermometer.
11.3.2 Adjustment 00 -10V AUX
Connect to AUX socket the calibrator , turn switch for zeroing ( pos. 1),
select line “ AUX offset” and press key
.
Turn switch to position 3 ( 10,00 V) , select line „V span“ and adjust the span factor with keys
reading on the display.
,
for 10V
,
for 20mA
11.3.3 Adjustment 44 -20mA
Connect to AUX socket the calibrator , turn switch for zeroing ( pos. 1),
select line “ AUX offset” and press key
.
Turn switch to position 4 ( 20mA) , select line „mA span“ and adjust the span factor with keys
reading on the display.
58
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
11.4
Analog output (4 - 20mA)
Note: These adjustments are factory made, it is generally not necessary to repeat them!
This function allows user to calibrate the analog current output.
,
:
:
select the line “Analog output (4-20 mA)”
activate analog output adjustment
,
:
select the output channel (1 to 8)
,
:
select Offset 4 mA or 20 mA
Test tools:
Digital multimeter.
Test plug with 100 Ohm resistors, 0,1% tolerance, for each of the 8 channels.
Manual calibration:
Connect the test plug to the analog output socket of the analyzer.
Measure with the digital multimeter the voltage above the resistor of the selected output channel.
100 Ohm resistor measured voltage 0,4 V 4 mA value at the display: 4,00
100 Ohm resistor measured voltage 2,0 V 20 mA value at the display: 20,00
,
:
adjust the output until on the digital multimeter 4,00 is displayed for 4 mA and 20,00 for 20 mA
return to adjustment menu
59
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
12 Service maintenance plan
A check of the analyser at our service department is recommendable:
• after 1000 working hours
• if the last check of the analyser was effected 11 month ago.
As soon this period is exceeded, the analyser displays 2 times after switching on:
This message can be skipped by pressing the
- key.
For more information’s about the service plan, please call our service department.
13 Self diagnosis
The analyser is equipped with a self-diagnosis program.
All internal functions are monitored and displayed if necessary.
Switch on analyser:
Error message:
all values are checked
In the “Menu” e. g. “NO sensor not OK / analyser not ready for measurement” is displayed. Measuring analyser can be activated in all menus except for menu “Measurement” (measurement cannot be started).
Service program:
13.1
Serrvice. The mV
In case of an error message, change to the Extra Menu and select Se
values of e. g. the sensors, the flue gas temperature sensitive element etc. are displayed. Afterwards, the mV values are indicated, which restrict the “OK” range:
range
Sensors: ( to fresh air )
O2 sensor :
CO sensor :
CO sensor^(H2 compensated
CO sensor up to 100.000 ppm
SO2 sensor
:
NO sensor :
NO2 sensor :
H2S sensor:
H2 sensor:
13.2
+200 mV up to + 500 mV
-13 mV up to + 25 mV
-35 mV up to + 50 mV
-13mV up to + 25 mV
-13mV up to +25mV
-13mV up to +25mV
-13mV up to +25mV
-13mV up to +25mV
-13mV up to +25mV
Battery voltage
Battery according to state of charge:
Battery fully charged
13.000mV = 13V (voltage)
Warning limit (the battery symbol on the dis11.100mV = 11,1V (voltage)
play starts flashing)
Cut off limit of the analyser, protection against 10.500mV = 10,5V (voltage)
total discharge
60
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
13.3
Temperatures
TH Gas: (thermo element gas)
-500mV to +500mV with room temperature: app. 0mV
KTY Gas: (refer. temperature gas) -500mV to +500mV with room temperature: app. 200mV
TH Air: (thermo element air)
-500mV to +500mV with room temperature: app. 0mV
KTY Air: ( refer. temperature air) -500mV to +500mV with room temperature: app. 200mV
KTY Sensor ( temp. of sensor):
0 to 5000mV room temperature: app. 2000mV
( Appendix 1: chart in mV regarding temperature in °C )
13.4
Draft
Draft:
-500mV to 500mV not pressurized:
-20 mV to + 20mV
13.5
BiasBias -voltage
Bias: (indication of bias voltage)
+2.999mV to + 3.250mV
If the above mentioned values are exceeding the allocated ranges, a defect is present.
If no sensitive elements are connected, the corresponding fields in the service print are marked with
- - - - - ( negative service value) or with + + + + + (positive service value ).
13.6
KTYKTY- chart
AD [ mV ]
T [ °C ]
13.7
163
0
170
5
177
10
185
15
192
20
200
25
208
30
216
35
225
40
233
45
242
50
5
10
15
20
25
30
35
40
45
50
PT2000PT2000- chart
AD [ mV ]
T [ °C ]
0
Scroll service data:
Activate printer with:
,
and
,
.
print of the current service data, analyser number, EPROM version number
Note:
The number of prints can be determined by pressing the printer key (e. g. pressing printer key twice = 2 prints).
Quit service program:
get back to Extra Menu with key
61
.
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14 Appendix
14.1
Manual for firmware update of H8H8 - Industrial analyzers
14.1.1 System requirements
•
Operating system MS Windows 98SE/ME/2000/XP or later
•
Pentium PC 450MHz or better
•
One avalable RS232 port. It should be a real RS232 port. A RS232-USB adaptor may work, but may fail as well.
14.1.2 Important
Important hints and warnings
• Most probably stored measurements and sites get lost during a firmware update!
•
After the update the analyser must be switched off, as soon as it has reached the self test window! Please
disconnect mains (unplug cable or switch off the master switch), analysers with battery additionally need to be
switched off by pressing the keys F1 and F4 simultaneously. Now you can restart the analyser and use it with
its new firmware.
•
Please start the update only with a reliable mains connection.
14.1.3 Update procedure
•
Please do not switch-on the analyser yet and disconnect mains! Connect the analyser with a 0-modem-cable
(at least five-wire crossed) to the PC’s RS232 port.
•
Start the PC-program “MRU_UPLOADER.EXE”.
•
Select the COM-port, which is connected with the analyser by clicking on “RS232” and then on the appropriate “COMx”.
•
Load a firmware file by clicking on “file” / “load file” and selecting a file.
•
Now press and hold down the F1 key and connect mains (plug cable or switch on the master switch). Hold the
F1 key down until the PC indicates the message “MRU unit (H8-generation) recognized”. Now you can release
the F1 key.
•
The update usually takes less than one minute. The unit’s display stays empty during the uploading process.
When the update has finished, the unit starts its program as usual. Please bear in mind now the important
hints and warnings written above concerning switching-off and data loss!
14.1.4 Possible causes and solutions in case of update failure:
•
Cable problems: cable defective, wrong cable or cable not properly plugged
•
Another software runs on the used PC reserving the selected RS232 port: Close all running programs and restart MRU_UPLOADER.EXE.
•
A serial mouse uses the one and only RS232 port: Use another PC or another mouse.
•
You have not selected the COMx port which is connected to the analyser: Change the COM-port selection.
62
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14.2
Electrical connections
14.2.1 Analog output
All 8 analog outputs are user settable. For this, all measured and calculated values can be used.
! The analog outputs have no galvanic isolation !
14.2.2 RS 232 connector
-
+
Pin 9 leads a voltage of ( +5V ) to supply external units ( for example a Bluetooth module for wireless data transfer).
14.2.3 RS 485 connector
This connector is used only for supply and data transfer to external smart transmitters ( e.g. TOM420 ).
63
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14.2.4 RS 485 converter connection to PC
Connection RASI 901 to RS 485 converter
SWG 300300 -1
R+
RTT+
GND (housing)
RS 485 converter
1 = T+
2 = R3 = T4 = T+
GND not connected
Switch settings at the converter:
SW1 = DCE
SW2 = T ON/ R ON
Max. length of transmission cable: 1.000 m with screened cable
The screen has to be connected at the PE mounting at the PC housing.
64
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
SWG 300300 -1
R+
RTT+
GND (housing)
RS 485 converter
1 = T+
2 = R3 = T4 = T+
GND not connected
Switch settings at the converter:
SW1 = DCE
SW2 = T ON/ R ON
Max. length of transmission cable: 1.000 m with screened cable
The screen has to be connected at the PE mounting at the PC housing.
65
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14.2.5 RS 232 connection to PC
RS 232 cable from analyzer (RS232) to PC (COM/RS232)
length
#
3m
60003
5m
51532
10 m
51562
20 m
54367
66
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14.3
Transmission
Transmission protocol
14.3.1 Online data transmission protocol
67
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14.3.2 data storage transmission protocol
68
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14.4
Display messages
Message 0
SET TO ZERO, LAUNCHED
Message 1
Analyser configuration is being saved
Message 2
Set to zero in progress
Message 3
?? SENSORS ?? ? NOT READY ?
Message 4
SET TO ZERO FINISHED, SENSORS READY, ANALYSER READY FOR MEASUREMENT
Message 5
MEASUREMENT VALUES STORED
Message 6
DATA STORAGE FULL
Message 7
NO MEASUREMENT VALUES STORED
Message 8
“not used”
Message 9
ZERO SETTING DRAFT SENSOR PLEASE WAIT
Message 10
“not used”
Message 11
CO-SENSOR not ok! Limited measurement
Message 12
H2-SENSOR not ok! Limited measurement
Message 13
NO-SENSOR not ok! Limited measurement
Message 14
O2-SENSOR not ok! Limited measurement
Message 15
REMINDER! PLEASE RECHARGE THE BATTERY PERIODICALLY!
Message 16
This fuel type is currently selected for measurements. Therefore it can’t be disabled.
Message 17
-Attention- Administrator PIN-Code is 6xF1! PIN-Codes are disabled!
Message 18
69
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
!! CAUTION !! TRANSMISSION WAS NOR SUCCESSFUL
Message 19
-CONNECTION INTERRUPTED – WRONG PC-COM SELECTED – WRONG CABLE – PC-RECOURCES CONFLICT
Message 20
DATA TRANSMISSION FROM PC WAS SUCCESSFUL
Message 21
PC NOT READY
Message 22
CUSTOMER’S MEMORY IS FUL
Message 23
DATA TRANSMISSION TO PC WAS SUCCESSFUL
Message 24
BATTERY LOAD INSUFFICIENT FOR PRINT-OUT CONNECT MAINS CABLE
Message 25
“not used”
Message 26
“not used”
Message 27
ANALYSER –ERROR POSSIBLE LOSS OF TIME/DATE AND STORED DATA
Message 28
MEASUREMENT STOPPED
Message
Messag e 29
MEASUREMENT STARTED
Message 30
!!! ERROR !!! POSSIBLE LOSS OF CONFIGURATION DATA
Message 31
SEARCH WITHOUT SUCCESS
Message 32
NDIR bench not ready! Limited measurement
Message 33
DRAFT/PRESSURE WILL BE SET TO ZERO ALL NOZZLES PRESSURELESS?
Message
Mes sage 34
!! CAUTION !! DUE FOR SERVICE CONTACT YOUT LOCAL DEALER
Message 35
Configuration transmission error! Configuration rejected!
Message 36
DATA TRANSMISSION IN PROGRESS PLEASE WAIT
70
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
Message 37
A new customer was created.
Message 38
Sensors are not rinsed yet.
Message 39
Automatic zeroing Please wait…
71
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
14.5
Lists of countries and fuel types
14.5.1 General information
O 2 max = 20.96%
O2 content
O 2 = measured O2 value [%]
O2bez = O2-reference value [%] (fuel specific)
CO 2 max = CO2max-value [%] (fuel specific)
V AG min, tr = Flue gas volume, dry and O2-free (fuel specific)
H u , n = calorific value , normality (fuel specific)
14.5.1.1 CO2, Lambda, losses, efficiency

O2
CO 2 = CO 2 max ⋅  1 −
O 2 max

λ=



CO 2 max
O 2 max
=
CO 2
O 2 max − O 2
(prefer the 2nd O2 formula!)


A2
losses = (T fluegas − Tair ) ⋅ 
+ B  (temp. in °C, A2 and. B fuel specific)
 O 2 max − O 2

efficiency = 100% − losses
14.5.1.2 Dew point
(after DIN 4705, part 1, page 17)
tP =
tP
pD
pL
4077,9
− 236,67
23,6448 − ln ( p D )
: dew point temp. in °C
: steam part. pressure
: atmospheric pressure
with
fW
CO2
pD =
H 2O
* pL
100
with
H 2O = 1,1 +
100
f
1+ W
CO 2
H2O : water vapour content [%]
: humidity value of fuel [%]
: carbon dioxide content [%]
At standard pressure (1013 mbar) the dew point is calculated:
4077,9
4077,9
tP =
− 236,67 =
− 236,67








100 
100 


23,6448 − ln 1,1 +
− ln (1013)
16,7241 − ln 1,1 +
fW 
f 


1+
1+ W 



CO 2 
CO 2 


14.5.1.3 COCO- conversion of CO [ppm]
COunv[ ppm] = CO[
CO[
ppm
] = CO[ ppm] ⋅ λ
0%O 2
ppm
O 2 max − O 2bez
] = CO[ ppm] ⋅
O 2bez
O 2 max − O 2
72
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
mg
] = CO[ ppm] ⋅ 1,249
m3
CO[
mg / m 3
mg O 2 max − O 2ref
CO[
] = CO[ 3 ] ⋅
O 2ref
O 2 max − O 2
m
V AG min, tr
mg
mg / m 3 V AG min, tr
mg
] = CO[
]⋅
= CO[ 3 ] ⋅ λ ⋅
kWh
0%O 2
H u ,n
H u ,n
m
CO[
mg
]=
MJ
CO[
mg
]
kWh
3,6
CO[
14.5.1.4 NONO- conversion of NO [ppm]
[pp m]
NOundel [ ppm] = NO[
ppm
] = NO[ ppm] ⋅ λ
0%O 2
NO[
ppm
O 2 max − O 2ref
] = NO[ ppm] ⋅
O 2ref
O 2 max − O 2
NO[
mg
] = NO[ ppm] ⋅ 1,339
m3
NO[
mg / m 3
mg O 2 max − O 2ref
] = NO[ 3 ] ⋅
O 2ref
O 2 max − O 2
m
NO[
V AG min, tr
mg
mg / m 3 V AG min, tr
mg
] = NO[
]⋅
= NO[ 3 ] ⋅ λ ⋅
kWh
0%O 2
H u ,n
H u ,n
m
NO[
NO[
mg
]=
MJ
mg
]
kWh
3,6
NO[
ppm
O 2 max − 1,8%
] = NO[ ppm] ⋅
14%CO 2
O 2 max − O 2
14.5.1.5 NO2NO2- conversion of NO2 [ppm]
No conversions of NO2!. All the values of interest are in the NOx conversion.
14.5.1.6 NOxNOx- conversion of NO [ppm] (possible also of NO2 [ppm])
NOx[ ppm ] = NO[ ppm ] ⋅ 1,05
(if no NO2-sensor exists)
NOx[ ppm ] = NO[ ppm ] + NO 2[ ppm ]
(if a NO2-sensor exists)
NOx undel [ ppm] = NOx[
NOx[
ppm
] = NOx[ ppm] ⋅ λ
0%O 2
ppm
O 2 max − O 2ref
] = NOx[ ppm] ⋅
O 2ref
O 2 max − O 2
If NOx is specified in mg, it has to be considered, that all the NO in the atmosphere will become to NO2.
Therefore the conversion factor ppm => mg is factor of NO2.
mg
NOx[ 3 ] = NOx[ ppm ] ⋅ 2,053
m
73
START-UP AND SERVICE MANUAL RASI 900 / RASI 901
EiUK Eurotron Instruments (UK) ltd
NOx[
mg / m 3
mg O 2 max − O 2ref
] = NOx[ 3 ] ⋅
O 2ref
O 2 max − O 2
m
NOx[
V AG min, tr
mg
mg / m 3 V AG min, tr
mg
] = NOx[
]⋅
= NOx[ 3 ] ⋅ λ ⋅
kWh
H u ,n
H u ,n
0%O 2
m
mg
NOx[
]=
MJ
mg
]
kWh
3,6
NOx[
ppm
O 2 max − 1,8%
] = NOx[ ppm] ⋅
NOx[
14%CO 2
O 2 max − O 2
14.5.2 Fuel type list Great Britain
Some analyzer disposes only of a subset of this list.
England
Fuel type
Test gas
Oil light
Oil heavy
Nat. gas heavy
Nat. gas light
Nat. gas H blow
Nat. gas L blow
Coal gas
Coal gas blow
Coal
Wood dry
CO2max
0,0
15,3
15,8
11,7
12,2
11,7
12,2
10,0
10,0
19,1
19,4
A2
0,00
0,68
0,66
0,66
0,63
0,66
0,63
0,73
0,73
0,65
0,65
B
0,000
0,007
0,007
0,009
0,009
0,009
0,009
0,011
0,011
0,009
0,009
Fuel type lists for further countries on demand!
Data subject to change without notice!