Download USER MANUAL - Langford IC Systems, Inc.

Transcript
Langford IC Systems
Instrument Processor
USER MANUAL
Copyright © 2014 Langford IC Systems, Inc. All rights reserved.
Instrument Processor is a trademark of Langford IC Systems, Inc. All other trademarks and registered
trademarks listed herein are the property of their respective holders.
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
Langford IC Systems has determined that the Instrument Processor (Cleaner Only
configuration) meets the applicable IEC 60601 requirements for:
IEC 60601-1:1988+A1:1991+A2:1995 (Medical Electrical Equipment, Part1: General
Requirement for Safety)
EMC Emissions and Immunity: EN60601-1-2: 2001 electromagnetic emissions and
immunity requirements for medical electrical equipment – Group 1 Equipment, Class A for
non-life supporting equipment.
There are no known electromagnetic emissions or immunity to or from other equipment in
proximity with the Instrument Processor System.
The microprocessor control system was designed and tested in accordance with FDA’s
“Guidance for the Content of Pre-market Submissions for Software Contained in Medical
Devices” (05/29/98).
CAUTION: Product will be marked with this symbol when it is necessary for the
user to refer to the instruction manual in order to prevent injury or equipment damage.
Manufactured for: Langford IC Systems, Inc., 310 S. Williams Blvd, Suite 270
Tucson, Arizona 85711U.S.A.
1-520-745-6201
www.licsystems.com
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
1
CONTENTS
INTENDED USE AND DESCRIPTION............................................................................ 4
TECHNICAL OVERVIEW OF THE INSTRUMENT PROCESSOR PARAMETERS
AND OPERATION ............................................................................................................. 5
USER START - UP ............................................................................................................. 6
Zymit LF Detergent DESCRIPTION ................................................................................ 13
USER QUALIFICATION ................................................................................................. 13
COMPONENTS ................................................................................................................ 13
LID and AUTO-LATCH MECHNISM OPERATION..................................................... 15
ENDOSCOPE TRAYS, AND BAFFLE INSERTS.......................................................... 18
INDICATIONS ................................................................................................................. 18
Contraindications............................................................................................................... 20
Warnings ........................................................................................................................... 20
Cautions ............................................................................................................................. 21
Precautions ........................................................................................................................ 22
Cleaning Operation............................................................................................................ 23
Initial Pre-Operation Check List ....................................................................................... 24
Protective Cap Attachment ................................................................................................ 25
Terminate Cycle ................................................................................................................ 27
History Screen ................................................................................................................... 29
Preventative Maintenance ................................................................................................. 30
Changing the Hot Water Filter .......................................................................................... 31
Specifications .................................................................................................................... 35
System Installation ............................................................................................................ 36
Installation ......................................................................................................................... 37
System Pre-installation Checklist ...................................................................................... 37
System Supply ................................................................................................................... 38
Space Requirements .......................................................................................................... 38
Water Supply ..................................................................................................................... 39
Electrical Supply ............................................................................................................... 39
Drain .................................................................................................................................. 40
Drain Pump Service........................................................................................................... 40
Drain Pump Circuit Breaker .............................................................................................. 41
Replacing The Impeller ..................................................................................................... 43
Chamber Full, Cover Closed, Cannot Open ...................................................................... 44
Software Set-Up Screens ................................................................................................... 45
Verifying the Installation................................................................................................... 51
Administrative Setup Functions ........................................................................................ 52
User MENU ....................................................................................................................... 53
Set Time/Date .................................................................................................................... 53
Modify Instrument ID ....................................................................................................... 54
View History ..................................................................................................................... 57
Accessories and Supplies .................................................................................................. 59
TROUBLESHOOTING .................................................................................................... 60
Exception Codes ................................................................................................................ 60
List of exceptions (errors) ................................................................................................. 61
Diagnostic System (see Service Manual) .......................................................................... 66
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
2
Appendix A – Endoscope Compatibility........................................................................... 67
Appendix B – Symbols...................................................................................................... 68
Appendix C – References .................................................................................................. 71
Appendix D – Trays and Inserts ........................................................................................ 72
Appendix E – Relief Valve Inspection and Test ............................................................... 75
Warranty ............................................................................................................................ 76
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
3
Important -- Please Read Before Use
INTENDED USE AND DESCRIPTION
The Instrument Processor System consists of an automatic Instrument Processor; Zymit LF
Detergent; endoscope trays specific to endoscopes type (bronchoscope, colonoscope,
duodenoscope, etc.) and baffle inserts specific to endoscopes type. The LIC Instrument
Processor, when used in accordance with its labeling, the Zymit LF Detergent, provides
cleaning of remaining protein to validated levels as low as < 1.3 µg/cm² and remaining total
organic carbon (TOC) levels as low as < 1.8 µg/cm².
The Instrument Processor is designed to: (1) clean one endoscope at a time during a
complete processing cycle, (2) replicate the cleaning results only obtained by following the
reprocessing instructions provided in the user’s manual of the endoscope being processed,
and (3) facilitate the healthcare facility’s compliance with reprocessing guidelines published
by SGNA, APIC, AORN, ASGE, CDC, and other professional organizations. The
Instrument Processor provides a continuous, automated cleaning process. Following the
automated cleaning process, the processed endoscope is immediately available for further
processing in accordance with the endoscope manufacturer’s reprocessing instructions to
achieve high level disinfection.
The Instrument Processor is a floor standing apparatus that is approximately the size of a
large office copying machine. It stands approximately 45-5/8 inches in height, 48 inches
wide and 34-3/8 inches deep.
Figure 1: Picture of Instrument Processor showing side and front views
In the center of the apparatus, positioned at approximately tabletop height, is a stainless steel
chamber in which the medical instrument is processed. When loading, the instrument to be
processed is placed on a plastic tray with the distal and proximal ends of each lumen
positioned on opposite sides of the baffle insert.
Baffle
Customer Service (USA): 1-508-337-6500
Scope
Distal End
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
4
Scope
Proximal End
Tray
Chamber
Endoscope
Figure 2: Top view of Instrument Processor showing chamber with tray and endoscope.
The tray includes features to hold the instrument safely in place, yet does not prevent the
processing fluids from contacting all of the instrument's external surfaces. The trays are
positioned in the Instrument Processor chamber, the endoscope is placed in the trays and
supported using the hold down features. Then specially designed baffle inserts are slid into
a slot in the baffle so that the instrument is further secured. The areas on the baffle where
the instrument is secured are also designed to allow the processing fluids to freely contact
the instrument's external surfaces. The baffle partitions the chamber into two sections. The
design of the tray and baffle are such that a predetermined restriction is introduced between
the two sides of the chamber. This restriction is used to generate the patented "push-pull"
mechanical scrubbing agitation action that facilitates instrument cleaning.
On the top, right side of the apparatus there is an interactive key pad that is positioned to be
easily viewed and accessed by a user. The interactive key pad includes a graphical liquid
crystal display that supplies general status information and prompts, illuminated status and
warning indicators, and control buttons.
Interactive Key Pad
10/22/14
10:26
Instrument Processor
User: TEST
Clean
Rinse 1
Rinse 2
Agitating
Est. Time: 34 min
WARNING
Figure 3: Picture and graphic of Instrument Processor Interactive Key Pad
TECHNICAL OVERVIEW OF THE INSTRUMENT PROCESSOR PARAMETERS
AND OPERATION
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
5
The Instrument Processor System is a self-contained apparatus that automatically cleans
endoscopes. The system uses a combination of physical and chemical actions to achieve its
function.
To clean endoscopes placed in the apparatus, the system uses two stages. The first stage
uses a detergent for cleaning. The second stage rinses the endoscope twice and cleaning is
complete. The cleaning and rinse stages reduce the endoscope's soil burden.
After rinsing, remove the endoscope from the Instrument Processor System and subject the
endoscope to whatever process the facility uses to achieve high level disinfection.
The Instrument Processor relies on four fundamental process parameters to achieve its
cleaning function. The parameters include the TIME that the instrument being processed is
exposed to each process stage, the differential PRESSURE generated across the two sides of
the bisected processing chamber during the scrubbing agitation, the CONCENTRATION of
chemical, and the TEMPERATURE of the liquid that is exposed to the instrument being
processed. The software controls all of these parameters using an intricate system of
monitoring/sensing devices located throughout the device.
USER START - UP
After turning on the machine and after initialization is complete, the Login Screen is
displayed. The user is required to enter a five-digit user identification code (see
Administrative Setup Functions) to operate the Instrument Processor. This code is the first
of a series of identifiers printed with the documentation of the Instrument Processor Systems
process for the facility’s records at the end of the Instrument Processor System process.
Login Screen
10/29/14
10:26
Instrument Processing
System
Enter ID: 00000
Then Press ENTER
If an existing user ID is entered, the Inst ID Screen is displayed.
Inst ID Screen
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
6
10/29/14
10:26
Instrument Processing
System
User: TEST
Enter Inst ID: 000000007
Then Press ENTER
Next, enter an instrument ID. An Instrument ID entry contains a total of nine (9) numeric
numbers. Entering the last few digits followed by the ENTER key displays the Select Inst
ID screen. This screen displays a list of possible ID matches, from which the user selects
the desired endoscope ID.
Select Inst ID Screen
10/29/14
10:26
Instrument Processor
Inst ID: 000000417
000000417 000179417
000001417 000695417
000029417 001837417
000053417
000148417
Select Inst ID to Wash
Then Press ENTER
If the Instrument ID is listed, the user selects the Instrument ID then presses ENTER. The
screen is prompted to the Insert Instrument Screen.
10/29/14
10:26
Instrument Processing
System
User: TEST
Inst ID: 000000417
Type: Bronchoscope
Insert Instrument
Close Lid
If the ID of the scope does not appear on the screen (the first time being processed), the user
selects Add ID and follows the screen instruction to add a new Instrument ID (see Modify
Instrument ID).
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
7
10/29/14
10:26
Instrument Processor
New Instrument
Inst ID: 000000417
Unknown Instrument ID
Verify New ID to Add
Enter Inst ID:
Then Press ENTER
The user enters 000000417 and presses ENTER; the New Inst Type screen is displayed
10/29/14
10:26
Instrument Processor
New Instrument
Inst ID: 000000417
Bronchoscope
Colonoscope
Duodenoscope
Gastroscope
Select New Inst Type
Then Press ENTER
The user selects Bronchoscope and presses ENTER; the Verification Screen is displayed.
Verification Screen
10/29/14
10:26
Instrument Processor
New Instrument
Inst ID: 000000417
Type: Bronchoscope
Inst Type Correct?
No
Yes
The user confirms the type of the endoscope and the Instrument ID then chooses YES and
presses ENTER, the Insert Instrument Screen is displayed. A bronchoscope with an
Instrument ID 000000417 has been added to the system.
Note: If the user presses NO the set-up sequence is aborted and the screen will return to the
Enter ID Inst. ID screen.
The user selects the appropriate endoscope trays and inserts (they are labeled and or color
coded) and installs the tray system into the processing chamber left and right hand sides (see
Appendix D – Trays and Inserts for more information on the combination of trays and
inserts to be used).
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
8
The user then places the endoscope into the processing trays and positions the insertion tube
and electrical body tube to pass through the baffle openings. The illustration below shows a
colonoscope correctly positioned in the tray system. The baffle would be in the center
separating the Left Hand Side (LHS) tray and the Right Hand Side (RHS) tray.
LHS Tray
The colonoscope tubes pass through the baffle
and baffle inserts
RHS Tray
Figure 4: Pictures of a colonoscope correctly positioned in the LHS and RHS trays
Once the instrument is loaded in the chamber with each lumen end positioned on opposite
sides of the baffle, the baffle inserts for the type of endoscope are inserted. There are four
different baffle inserts. One for bronchoscopes, one for colonoscopes / duodenoscopes, one
for duodenoscopes and blank inserts (see Appendix D – Trays and Inserts for more
information on the baffle inserts to be used). Figure 5, shows the assembly of the baffle
inserts into the baffle and over the endoscope being processed.
Using IPA to lubricate the insert and baffle seals to facilitate assembly.
Figure 5: Instrument Processor showing endoscope placement and baffle insert placement.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
9
The user gently positions the associated inserts: TOP, BLANK, BLANK and BOTTOM
Figure 6: Bronchoscope loaded with trays and baffles, and ready to wash
The combination of the tray, baffle inserts, and endoscope defines a unique operating system
that allows the Instrument Processor to achieve the operating parameters defined by the
software. If the Instrument Processor fails to achieve the operating parameters defined by
the software, the operation will abort.
Insert Instrument Screen
10/29/14
10:26
10/29/14
10:26
Instrument Processing
System
Instrument Processing
System
User: TEST
Inst ID: 000000417
Type: Bronchoscope
User: TEST
Inst ID: 000000417
Type: Bronchoscope
Insert Instrument
Close Lid
Lid Unlatched
Latch Lid
And
After the endoscope is loaded, the user closes and latches the lid. Once the lid is secured,
the Instrument Processor System will go through the VERIFY WASH steps of the set-up to
confirm with the user that the detergent (Zymit LF Detergent) reservoir is secured in the
chemical docking station located on the front top right hand side of the L Instrument
Processor and that the reservoir is filled.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
10
Chemical
Docking
Station
Figure 7: Instrument Processor showing chemical docking station
Figure 8: Picture showing unlatching and latching of the Zymit LF Detergent Reservoir in
the Docking Station
The Zymit LF Detergent reservoir connects to the left side chemical dock.
Zymit LF
Detergent
Reservoir in
Docking
Station
Figure 9: Picture of Zymit LF Detergent reservoir in the latched Docking Station
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
11
The user is responsible for assuring that the Zymit LF Detergent reservoir is full for each
new set-up. The interactive key pad will display the following screens in the VERIFY
WASH steps:
10/29/14
10:26
10/29/14
10:26
Instrument Processing
System
Instrument Processing
System
Confirm Detergent
Latch is closed
Confirm Detergent
Reservoir is Full
No
Yes
No
Yes
And
The user confirms that the Zymit LF Detergent reservoir is full and latched and selects YES
and presses ENTER.
If the user selects NO the prompt screen will return to the beginning of the set-up checklist.
By selecting YES, the set-up sequence continues with the Ready to Wash Screen and the
green ready light will come on in the Interactive Key Pad:
Ready to Wash Screen
10/29/14
Green Ready Light
10:26
Instrument Processing
System
User: TEST
Inst ID: 000000417
Type: Bronchoscope
Ready to Wash
Press START to Continue
Figure 10: User Interface Panel Input Keypad and Display
The set-up is complete. Press START on the Interactive Key Pad to initiate the cleaning
operation.
10/22/14
10:26
Processing Indicator
Instrument Processor
User: TEST
Clean
Rinse 1
Rinse 2
Complete Indicator
Est. Time: 34 min
Agitating
WARNING
Start Button
Figure 11: Interactive Keypad and Display
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
12
Once START is pressed the Instrument Processor cleaning operation will start and the
Processing Indicator will light. The Instrument Processor System will run until all stages are
completed. Once the stages are complete the Complete Indicator will light.
ZYMIT LF DETERGENT DESCRIPTION
Figure 12: Zymit LF Detergent (1 liter bottle) with Pour Spout
The Zymit LF Detergent, is a specially formulated low-foam enzyme detergent. The
detergent is supplied in 1-liter bottles that are used to fill the detergent reservoir.
The Zymit LF Detergent is the only detergent recommended for use with the Instrument
Processor System. Filling the detergent reservoir to the designated “Fill to Line” ensures
that the detergent will be delivered at its specified concentration range.
USER QUALIFICATION
The users of the Instrument Processor System must be trained. The user must have received
qualified training in the operation of the Instrument Processor System. The user should be
familiar with the location and function of the various components of the Instrument
Processor System.
COMPONENTS
This section provides a description / illustration of the various user accessible components
and their function.
 Chamber Latch
 Processing Chamber
 Endoscope Trays and Baffle inserts
 Keypad Controls and display/indicators
 Printer
 Hot Water Filter
 Detergent Connection
 Hot Water Hook-up
 Drain connection
 Electrical Hook-up
 IFU – Instructions For Use (Manual)
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
13
Figure 13: The Langford IC Systems Instrument Processor
Figure 14: Front view of the Instrument Processor System with processing chamber open.
Top view of the Instrument Processor System with processing chamber open and Trays
inserted on each side of the baffle and a bronchoscope secured inside.
Figure 15: Processing Chamber and Endoscope Trays
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
14
Processing Chamber Open with Endoscope, Trays, Baffle and Baffle inserts in position.
1. Cover Open
2. Pull Piston Locking Pin
3. Push & Hold Closed
4. Press “Close” while holding
5. Chamber Closed
Figure 16: Chamber Latch – Closing sequence for the Locking Systems
LID AND AUTO-LATCH MECHNISM OPERATION
1. To set the home position of the auto-latch after power-up, push the Open button, no
matter position of the latch. This internally homes the position of the latch, and will
open the lid if latched.
CAUTION: Before closing lid, make sure every part of the endoscope is in the
tray and none of it is sitting over the edge, or closing the lid may crush the
endoscope, rendering it inoperable.
2. To close the lid, pull the Piston Locking Pin on the left hand piston (if the lid is open
that far) and close the lid. DO NOT force the lid closed without checking the Piston
Locking Pin, or the mechanism may become damaged.
3. To latch the lid, press down on the top of the lid with the right hand, and press the
Green Close button with the left hand. The lid is pulled downward and is latched.
4. To unlatch the lid, press the Black Open button when instructed to do so. The lid can
also be opened and closed any time the machine is idle.
5. To open the lid after it is unlatched, lift the lid until the piston on the left clicks.
6. If the lid will not open, refer to the section labeled: Chamber Full, Cover Closed,
Cannot Open.
7. For this reason, the lid should not be left closed when the unit is powered down or in
between washes. Make sure the Shutdown procedure is used before powering down
the unit.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
15
There is an electronic interlock that prevents access to the processing chamber while the
Instrument Processor System is operating.
Close Lid
Open Lid
Figure 17: Close Lid and Open Lid Control Buttons
Figure 17, illustrates the Close and Open buttons for the Automatic Locking System.
The lid is manually lowered and the Close lid button (green) is pressed. To open the lid the
Open lid button (black) is pressed.
Keypad Controls and display indicators.
L ANGFORD IC S YSTEMS
Instrument Processing
System
Software Version: 1.0
Copyright (C) 2014
WARNING
Figure 18: Keypad Controls
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
16
Detergent
Latch
Detergent
Reservoir
Printer door
Figure 19: Printer location and detergent reservoir connection
Figure 20: Back Panel (drain, hot water inlet, power cord, power switch)
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
17
ENDOSCOPE TRAYS, AND BAFFLE INSERTS
Trays are used to position the endoscopes in the processing chamber and hold the
endoscopes in place during the cleaning and rinse cycles. There are two trays one on each
side of the baffle. To allow the endoscope to pass through the baffle, chamber baffle inserts
are removable and secure the endoscope as it passes through the baffle when inserted. The
baffle inserts are specific to the type of endoscope to be processed. There are four types of
inserts (see Appendix D – Trays and Inserts for more information on the combination of
trays and inserts to be used). This system allows the external surfaces and internal lumen
surfaces of the endoscope to be cleaned without attaching connectors to the ends of the
endoscope and facilitates removal of the cleaned endoscope.
INDICATIONS
The Langford IC Systems Instrument Processor, when used in accordance with its labeling
and the Zymit LF Detergent, provides cleaning to remaining protein levels of < 6.4 µg/cm²
on the inner lumen and outer surfaces of the endoscope in preparation for the high level
disinfection process.
The Zymit LF Detergent cleaning solution used in the Instrument Processor System is
compatible with the materials contained in the endoscopes to be processed. Reprocessed
endoscopes will contain minimal residual levels of detergent.
The volume of Zymit LF Detergent required for each cycle is conveniently measured by
filling the detergent reservoir to the prescribed fill level. The volume in the reservoir is
consumed during a normal processing cycle.
The Instrument Processor System will abort the processing and signal the user to take
corrective action if the monitored parameters are not within range.
Process parameters will be printed for Hospital Quality Assurance and / or Patient records
by the onboard printer contained inside the Instrument Processor System.
The Instrument Processor System facilitates compliance with the endoscope reprocessing
guidelines published by SGNA, APIC, AORN, ASGE, CDC, and other professional
organizations.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
18
REFERENCE GUIDELINES
Always follow the endoscope manufacturer's instructions for cleaning and disinfection of
medical endoscopes. Reference guidelines published by the following professional
organizations.
Society of Gastroenterology Nurses and Associates
401 North Michigan Av.
Chicago, IL 60611-4267
TEL: (800) 245-7462
FAX: (312) 527-6658
http://www.sgna.org/
American Society for Gastrointestinal Endoscopy
1520 Kensington Rd.
Suite 202
Oakbrook, IL 60523
TEL: (630) 573-0600
FAX: (630) 573-0691
http://www.asge.org/
Association for Professionals in Infection Control and Epidemiology, Inc
1275 K Street, NW, Suite 1000
Washington, DC 20005-4006
TEL: (202) 789-1890
FAX: (202 )789-1899
http://www.apic.org/
American Society for Testing and Materials
100 Barr Harbor Drive
West Conshohocken, PA 19428-2959
TEL: (610) 832-9585
FAX: (610) 832-9555
http://www.astm.org/
The Instrument Processor System meets the applicable UL and EN 60601 requirements.
The design and test results of the microprocessor control system meet the FDA’s “Guidance
for the Content of Premarket Submissions for Software Contained in Medical Devices”
(05/29/98).
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
19
CONTRAINDICATIONS
Appendix A lists endoscopes that have been tested by Instrument Processor in the
Qualification of the Instrument Processor System’s cleaning cycles. Use only those
endoscopes that are published in this manual as compatible with the Instrument Processor
System. Additional endoscopes will be published in supplements to this manual for other
medical instruments validated in the future for processing in the Instrument Processor
System by Langford IC Systems.
Endoscopes that should not be processed with the System
Endoscopes that are not compatible with the Zymit LF Detergent based on compatibility
testing or literature information of material compatibility with the Instrument Processor
System process.
Do not process endoscopes that are not indicated for total submersion, see Appendix A.
Certain areas of the endoscope that require protective coverings prior to submersion must
have the protective coverings in place prior to placing the endoscope in the Instrument
Processor System chamber. Endoscopes that do not have appropriate protective coverings
for the aerating or irrigating channels, or cannot withstand a temperature of 140°F (60 °C)
should not be processed with the Instrument Processor System as damage to the endoscope
may result. Refer to the endoscope manufacturer’s manual for any cautions related to
protective coverings and/or temperature during reprocessing the endoscope.
WARNINGS
Do Not Use in Patient Areas. The Instrument Processor System is equipment that shall not
have contact with patients.
Use only those product(s), supplies, parts or accessories recommended by Langford IC
Systems in this manual (see Accessories and Supplies Section). Failure to do so may result
in ineffective cleaning or disinfection, damage to endoscope, and/or system malfunction.
The Instrument Processor System must be installed according to the instructions and
specifications provided in the Installation Procedures (see System Installation Section).
Failure to do so may result in ineffective cleaning, damage to endoscopes, system
malfunction and/or non-compliance with local codes.
It is the user’s responsibility to provide the Instrument Processor System with hot water at a
minimum temperature of 100°F but not to exceed 140°F. Although the Instrument
Processor System has an on-board hot water filter, Langford IC Systems recommends that
the user also supply filtered water.
Ensure that the endoscopes to be cleaned have been tested for leaks and lumen
obstructions according to the endoscope manufacturer’s instructions and current
professional practices prior to processing in the Instrument Processor System. This
involves wiping the outer surface of the endoscope with detergent, leak testing and passing a
small amount of detergent through the working lumen(s) to ensure that the lumen is not
occluded. Failure to perform these pre-processing steps may result in ineffective
cleaning and or damage to the endoscope.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
20
If a process cycle is STOPPED (cancelled), either by the user or by the Instrument Processor
System’s operating system, the endoscope being processed is not safe to use.
Failure to properly secure the endoscope in the endoscope tray, properly select and place the
correct baffle insert and properly cover the non-submersible and pressure sensitive areas of
the endoscope may result in damage to the endoscope.
Review the instructions for use for the detergent prior to use. Contact Langford IC Systems
Customer Service at “1-508-337-6500” with any questions regarding the use of the
detergent.
Use only the 510(k) approved Zymit LF Detergent in the Instrument Processor System. The
active ingredients are ethanol, 1,2-dihydroxypropane and other proprietary chemical
substances. The concentrated solution is automatically diluted by the Instrument Processor
System to an in-use concentration of approximately 2%.
WARNING: Zymit LF Detergent is a skin irritant. Wear protective gloves and
eyeglasses. Avoid prolonged breathing of vapor. In case of contact, flush with
water. Do not mix with other detergents. Store at room temperature (5-30° C). A
Material Safety Data Sheet is provided with each shipment of Zymit LF Detergent
and is also available from Langford IC Systems Customer Service.
The Zymit LF Detergent requires no user preparation other than uncapping the
container and filling the detergent reservoir to the “Fill Line” on the reservoir.
The Zymit LF Detergent is the only agent 510(k) approved for use with the Instrument
Processor System. No other chemicals are authorized or recommended for use. The use of
other agents may result in ineffective cleaning or damage to the instrument being processed
or damage to the Instrument Processor System.
DANGER! Risk of EXPLOSION if the Instrument Processor System is operated in the
presence of flammable anesthetics!
A failure to follow the instructions in this manual along with the endoscope equipment
manufacturers cleaning procedures after each use can compromise patient safety.
Patient debris and reprocessing chemicals (detergent) are hazardous. Wear personal
protective equipment to safeguard against potentially infectious material and dangerous
chemical solutions. Follow universal precautionary standards and facility policies for
handling potentially infectious material and hazardous chemicals.
CAUTIONS
Ensure that all personnel who will be using the Instrument Processor System read and
understand the User Manual and are familiar with the safe operation of the Instrument
Processor System prior to use.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
21
The Instrument Processor System continuously monitors the processing cycles to ensure
cleaning within the parameters validated for release of the processed instrument. If an error
occurs in the monitored parameters, the cycle will be ABORTED by the Instrument
Processor System controls.
The operator may also stop the operational cycle. Upon cycle cancellation, the cleaning
solution in the Instrument Processor System chamber drains and the chamber latch
deactivates.
If a cycle is either aborted or cancelled, endoscopes should be reprocessed or rinsed
manually immediately upon cancellation of the cycle in order to prevent detergent from
drying (concentrating) on the endoscope(s).
If an operational error occurs that results in the “Service Required” screen display, then
further processing by the machine will be suspended. Call the Customer Service at
Langford IC Systems at “1-508-337-6500”.
An incoming water temperature of greater than 140°F (60°C) will cause the system to
automatically shutdown to protect heat sensitive endoscopes from being damaged.
Do not process endoscopes that are not recommended to be processed by the Instrument
Processor System. See Appendix A.
Damage may occur to endoscopes if they are not properly secured in the appropriate
processing tray and the correct baffle insert selected.
Zymit LF Detergent has been developed for use with the Instrument Processor System only.
Do not use with other washer.
Do not expose the thermal printer paper to direct sunlight or overhead florescent lighting.
Premature color development will occur if the paper is exposed to high temperature, light or
scratched with a sharp object.
DANGER! Do not attempt to open the processing chamber during a cycle, as it is under
pressure. You must press and acknowledge the STOP button if access to the
endoscope is required prior to the completion of the cycle. Pressing and
acknowledging the STOP button will cause the machine to stop the cycle, drain
and release the interlock so that the chamber may be opened.

Specific safety information regarding the safe handling, use, storage and disposal of
the detergent and used containers is contained in each product’s instructions for use.
Review the safety material thoroughly and contact Customer Service at Langford IC
Systems at “1-508-337-6500” if you have any questions.
PRECAUTIONS
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
22
1. Users must use proper handling of the devices according to facility policy and
universal handling of biohazardous materials.
2. Endoscopes must be pre-cleaned with detergent solution, inspected for leaks
and patency. Check for patency by passing a small amount of detergent through
the working lumen to ensure lumen is not occluded prior to placing the
endoscope in processing chamber.
3. Endoscopes must be totally immersible with protective coverings properly attached.
Only use the protective caps supplied by Langford to protect the endoscopes from
damage during cleaning.
DO NOT USE CAPS SUPPLIED BY THE
MANUFACTURER OF THE ENDOSCOPE
4. Endoscopes must be able to withstand a temperature range of 100°F-140°F (38°C60C°).
5. An endoscope is considered completely reprocessed only when it has been manually
pre-cleaned (pre-cleaning means wiping down the insertion tube section of the
scope), leak checked according to the manufacturer’s recommendations, patency
checked by passing a small amount of detergent through the working lumen
and has completed the full Instrument Processor System automated cycle with
the approved detergent.
6. If the Instrument Processor System is used in a manner not specified in this
instruction manual, its performance may be impaired.
7. Operation, maintenance and service of the Instrument Processor System must only
be performed by properly trained individuals. Those not properly trained in the
operation, maintenance and/or service of the unit could cause damage to endoscopes
being processed.
8. The Instrument Processor System must be connected to an electrically grounded
receptacle.
9. The Instrument Processor System must be supplied with hot water at a minimum
temperature of 100° F and not to exceed 140° F.
10. It is recommended that the user supply filtered hot water to the Instrument Processor
System.
11. For endoscopes, follow the endoscope/device manufacturer's instructions and
professional guidelines (Ref: SGNA, APIC, AORN, ASGE, CDC, and other
professional organizations) for guidance concerning the pre-processing steps, precleaning (pre-cleaning means wiping down the insertion tube section of the scope)
and leak testing, that must be performed prior to placing an endoscope into an
automated re-processor. Do not use the detergent beyond its expiration date.
12. Factory-supplied replacement parts and filters must be used to assure proper
operation and efficacy of the Instrument Processor System, and to maintain warranty
protection (see Accessories and Supplies section of this manual).
13. Use only water, soap, mild detergents, isopropyl alcohol or mild disinfectants to
clean the exterior of the Instrument Processor System. Call Customer Service if you
have any questions.
CLEANING OPERATION
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
23
This section assumes that the Instrument Processor System has been properly installed and
the System Administrator has performed the initial setup steps (see section on Installation
and Administrative Setup Functions).
The following gives a general overview of the pre-operation and operational steps for the
Langford IC Systems Instrument Processor.
INITIAL PRE-OPERATION CHECK LIST
1.
2.
3.
4.
5.
Water supply valves for HOT feed water are ON.
Power Switch is ON.
Drain hose is connected to the drain.
Insure the Zymit LF Detergent reservoir is attached securely.
After pre-cleaning and leak testing the endoscope per the manufacturer’s
instructions and patency checked by passing a small amount of detergent through
the working lumen, attach any protective coverings required for fluid submersion of
the endoscope (see section Protective Cap Attachment).
Note: Use only protective caps supplied by Langford to protect endoscopes
from water invasion. Caps supplied by the manufacturer are not built to
handle the pressure in the processing chamber during processing.
Flexible endoscopes must be leak tested prior to cleaning process per the
manufacturer’s instructions. The User must verify that the internal
lumens of the endoscope are not occluded prior to loading the endoscope
into the Tray.
6. After the User is prompted by the Login Screen to enter the User’s ID CODE, and
then enter the Instrument ID CODE, the Instrument Processor System screen display
will prompt the User to place the Endoscope Trays in the Processing Chamber by
indicating: “Insert Instrument”
7. Select the Endoscope Trays as per their labeling and put one on each side of the
baffle. Secure the mechanical end of the endoscope in the holding nest attached to
the tray and attach the holding strap. Take the distal end of the endoscope and lay it
in the correct slot in the baffle allowing the proximal end of the endoscope to reside
on one side of the chamber and the distal end to reside in the other. Select the proper
baffle inserts and push them into the proper locations in the baffle. The endoscope is
now ready for processing.
Note: Throughout this manual if the Instrument Processor System detects a condition
requiring the user to take action an audible alarm will sound and the condition will
be listed on the screen. These are referred to as exception (warning or error) screens
and are detailed in the Troubleshooting Section of this manual.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
24
PROTECTIVE CAP ATTACHMENT
This section details the attachment of a cap to endoscopes that have an electronic connection
that must be protected during processing. The figures below show the cap installation
procedure:
Alignment
Groove
O-Ring &
Gasket
Slots
Alignment
Groove
Figure 21: Endoscope Cap, top and bottom views
Follow these steps to install the cap (Figure 21) on the endoscope:
1. Inspect condition of the Endoscope Cap O-ring and gasket. If damage and/or wear
are visible, DO NOT use the Endoscope Cap.
2. Wet a glove-covered finger with Isopropyl Alcohol (IPA) (Figure 22).
3. Lubricate the O-ring of the cap and the electronics endoscope ring with IPA (Figure
23).
4. Align the slots on the Endoscope Cap (Figure 21) with the pins of the endoscope
electrical connector (Figure 23, on right).
5. Tighten the cap clockwise, approximately 1/4 full turn, until the Alignment Groove
on the Endoscope Cap lines up with the yellow line on the endoscope electrical
connector (Figure 24).
CAUTION: If the cap is not installed as shown, water will get into the electronics while
cleaning the endoscope, rendering the endoscope inoperable.
Figure 22: Wetting finger with Isopropyl Alcohol (IPA)
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
25
Pin
Pin
Figure 23: Lubricate the Endoscope Cap and the electrical endoscope ring
connector with IPA. The Endoscope Electrical Connector and the
Endoscope Cap are keyed and only fit in on position
Alignment Groove
Alignment Groove
& Yellow Line are
Adjacent
Figure 24: Align the Endoscope Cap slots with the Electrical Connector pins.
Tighten the Endoscope Cap until Alignment Groove and yellow
line are adjacent
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
26
TERMINATE CYCLE
1. Perform the following steps to terminate a cycle.
If the user terminates the cycle by pressing STOP while the device is being processed, the
following Confirm Stop Screen is displayed. If the user selects NO, the machine returns
to the Processing Screen. If the user selects YES, the machine stops processing, and the
Drain Screen is displayed. The following screens will allow the user to drain (clear) the
fluid from the aborted cycle chamber and restart the cleaning cycle.
If the process is aborted the user will need to refill the detergent reservoir with
detergent to the “Fill Line” before a new cleaning cycle can be started.
Confirm Stop Screen
10/29/14
10:26
Instrument Processor
User: TEST
Stop Processing Device?
No
Yes
After canceling a wash, the following sequence of screens is displayed. The first is the
Drain Screen and the user must press START to drain the chamber.
10/29/14
10:26
10/29/14
10:26
10/29/14
10:26
Instrument Processor
User: TEST
Instrument Processor
User: TEST
Instrument Processor
User: TEST
Press START to Drain
Chamber Draining...
Process Canceled
Chamber Not Safe
DO NOT Open Lid
Chamber Not Safe
DO NOT Open Lid
WARNING:
Not Cleaned
Press ENTER to Continue
Draining Screen
Process Canceled Screen
Drain Screen
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
27
Press ENTER. The Chamber Safe Unlatch Lid screens appears. Press Enter.
10/29/14
10:26
10/29/14
10:26
Instrument Processor
User: TEST
Instrument Processor
User: TEST
Process Canceled
Process Canceled
Chamber Safe
Unlatch Lid
Chamber Safe
Open Lid
The user Login Screen is displayed and a new process cycle may be started.
10/29/14
10:26
Instrument Processing
System
Enter ID: 00000
Then Press ENTER
Login Screen
2. If the cycle is terminated due to power failure, the following steps should be performed.
a. If power is restored, the Instrument Processor System will enter the Initialization
step. Follow the prompts to initialize the machine
CAUTION: If a cycle is either aborted or cancelled, endoscopes should be
reprocessed or rinsed manually immediately upon cancellation of the cycle in
order to prevent detergent from drying (concentrating) on the endoscope(s). The
endoscope being processed is not safe to use.
b. If power cannot be restored, the following steps should be performed.
CAUTION: The endoscope and water in the cleaning chamber are considered a
biohazard.
Use universal precautions and follow institutional
procedure when handling biohazardous materials.
i.
REMOVE POWER CONNECTION by turning off the main power
switch on the Instrument Processor System and disconnecting the
power cord from the mains outlet.
ii.
IDENTIFY prominently on the machine a tag that states: DO NOT
USE - BIOHAZARD.
iii.
Phone LIC Customer Service to dispatch trained service personnel.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
28
HISTORY SCREEN
View History Screen
The user can display a list of history records, sorted by date. Pressing MENU selects VIEW
HISTORY, View History Screen will display. Pressing ENTER prints the record. Pressing
CANCEL returns to the main menu.
View History Screen
10/29/14
10:26
Instrument Processor
View History
10/28/14 18:56 Ex 28
10/28/14 18:55 Ex 17
10/28/14 18:55 Ex 11
10/28/14 18:55 Ex 107
10/28/14 18:55 Ex 114
Press ENTER to Print
Print History Screen
10/29/14
10:26
Instrument Processor
Print History
History Record Printed
Press ENTER
Press “CANCEL” as required to return to the Enter ID: 00000 screen.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
29
PREVENTATIVE MAINTENANCE
Note: Certain Areas of the Instrument Processor System require Service Engineering for
repair. All components, except for the hot waters filter and relief valves require a
trained Service Engineer for maintenance or repair.
Cleaning of external surfaces: Standard surface disinfectants can be used to clean the
exterior surface of the Instrument Processor System.
Periodic replacement of:
1. Printer paper—when indicated by the end of roll indicator (red stripe). Note: If red
strip is present prior to initiating the cycle, replace printer paper.
Replacing Paper
(1) Press the paper cover open button, and open the paper cover, handling the paper cutter
carefully not to cut your hand.
(2) Set a paper roll as shown in the figures below.
(3) Close the paper cover by pressing both ends of the cover so the tip end of the paper will be
emerged from the printer.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
30
Thermal papers used
 Product No.: TP-211C-1 (10 rolls per box)
 Paper width: 58 mm
 Outer diameter of roll paper: φ50 mm or less
 Core: included
Use the specified thermal paper. If you use another type of thermal paper, the
quality of print and the useful service life of the thermal head may not be guaranteed.
2. Hot Water filter
The Instrument Processor System uses one water filter, a 2.0 micron filter. The 2.0
micron filter serves to remove large particles, sediment, from the incoming water hot
supply. The Instrument Processor System monitors time relevant to filter operation
and provides alarms when the filter needs replacement. However, it is
recommended that the user facility routinely monitor the microbial quality of the
rinse water used during endoscope reprocessing to assure the effectiveness of the
water filtration process. Filter life of the onboard filter can also be extended if you
connect the Instrument Processor System to one of the external filtration systems
normally located in the area where endoscopes are processed.
Suggested maintenance for the 2.0 micron hot water filter is as follows:
a. Every 7-10 days or when the Instrument Processor System indicates that the
filter needs replacement by listing an Exception #23
EX_C_CHAMBER_NOT_FULL, or #78 EX_C_BYPASS_TIMEOUT.
Check water pressure as these errors are also displayed when the chamber
does not fill because of low water pressure.
Note: Frequency of hot water filter replacement varies and depends on the
quality of the incoming water as well as whether or not pre machine user
facility filtration is being used. Slow fill times will generate an error
warning (see Exception #23 or #78 in the troubleshooting section)
requiring the hot water filter to be replaced.
CHANGING THE HOT WATER FILTER
The hot water filter should be changed when the fill rate warning prints out, lights up
or is indicated by the Screen Display of Exception #23 or #78. Use only LIC
authorized filters (see Accessories and Supplies section of this manual).
Make sure the Instrument Processor System is in the following condition prior to
changing the filter:
1. Ensure that there is no wash in progress (power is OFF or green Ready light
is ON on the control panel).
2. Turn the POWER to the Instrument Processor System OFF (if ON).
3. Turn the incoming hot water supply OFF.
Steps in changing the filter:
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
31
1. Shut off the hot water supply.
2. Obtain a strap wrench large enough (at least 4” O.D.) to accommodate the
diameter of the hot water filter housing.
3. Obtain a small tray to collect water that will fit under the hot water filter.
4. Open the front service door by using a tool (i.e. pen) to depress the door latch
button (see Figure 25).
5. Place the small tray under the hot water filter to collect the water runoff.
6. Using the strap wrench, turn the hot water filter housing to the left until loose
enough to turn by hand (see Figure 27).
7. Continue turning several turns by hand until the housing and filter separate
from the filter cover.
8. Empty the housing of water and discard the old filter.
9. Center and drop the new filter into the center of the empty filter housing (it
does not matter which end is up, as the filter has no polarity).
10. Reinstall the filter housing, turning to the right by hand (while keeping the
filter centered within the housing) until snug.
11. Using the strap wrench, tighten until fully seated (DO NOT OVER
TIGHTEN).
12. Remove the small tray, and discard the collected water runoff.
13. Turn the incoming hot water supply ON.
14. Keeping the door switch closed (see Figure 26), perform the following steps:
a) Turn the POWER to the Instrument Processor System ON
b) Follow the prompts to initialize the machine.
c) Check for water leaks under and around the hot water filter housing.
d) If the filter housing is leaking, tighten it using the strap wrench until the
leak stops. (DO NOT OVER TIGHTEN!)
e) Run a wash without an endoscope (use Inst ID #000000000, using the
“Other” instrument type).
f) While the chamber is filling during the first cycle, check for water leaks
under and around the hot water filter housing.
g) If the filter housing is leaking at any time during the fill, tighten it using
the strap wrench until the leak stops. (DO NOT OVER TIGHTEN!).
h) If the leak does not cease by the time the chamber has filled, turn off the
water and contact Langford IC customer service.
15. If leaks are not detected and the cycle entered the “Agitation” step release
the switch that detects the service door open, and close the service door.
16. Acknowledge the exception that occurs from having released the switch, and
allow wash recovery to complete.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
32
Figure 25: Opening the service door
Door
Switch
Drain Pump
Circuit Breaker,
Service Switch
Hot Water Filter
Drain Pump
Impeller
Figure 26: Inside Panels (hot water filter, main chamber pump) – Hot water filter center, 2
micron, door switch
Figure 27: Replacing the Hot Water Filter (with strap wrench and by hand)
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
33
Routine Maintenance of the Instrument Processor System Processing Chamber:
Periodic cleaning of the Instrument Processor System processing chamber is
recommended as a general maintenance procedure. The processing chamber is
constructed of stainless steel which is generally kept clean by the scrubbing action of
the Instrument Processor System cleaning cycle. Langford I C Systems recommends
that the chamber be reviewed at the end of each cleaning process to determine if any
staining or residue has occurred. Because of the continuous scrubbing action of the
process residue build up is not anticipated. If, however, it does occur or minor
staining occurs Langford IC Systems recommends using the cleaning methods
described in the “Designer Handbook, The Care and Cleaning of Stainless Steel”
published by the Specialty Steel Industry of North America. A copy of the handbook
is provided in Appendix C.
Summary of Maintenance Schedule
Frequency
Action
Every 7-10 days of
use (see note)
Once a month
Change 2.0 micron
water filter
Perform chamber
disinfection
Each cleaning
cycle
Visually inspect
processing chamber for
any staining or residues.
Inspect and test relief
valve
Every 12 months
Every 12 months
Inspect and test drain
pump impeller
LIC Number
Performed by
0006
User
Not applicable.
Use EPA-registered
sanitizer (e.g. SporKlenz® or diluted
Actril® Cold Sterilant)
Not applicable.
See Appendix C.
User
0007
Authorized
Service
Technician
Authorized
Service
Technician
Not applicable.
User
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
34
SPECIFICATIONS
Langford IC Systems Instrument Processor
Chassis Dimensions (H x W x D):
45-5/8 x 48 x 34-3/8 inches (115.9 x 121.9 x 87.3 cm)
Overall Height with Station Lid Raised:
60.25 inches (153 cm)
Weight ():
1500-1700 lbs. (680-771 kgs)
Designed for Use:
Indoors
Degree of Mobility:
Casters
Environmental Operating:
Altitude: Up to 15,000 feet (4,572 meters)
Humidity: 20 - 80%, non-condensing
Temperature: 80°F ± 20°F (27°C ± 12°C)
Environmental Storage and Transport:
Altitude: Not restricted
Humidity: 20 - 80%, non-condensing
Temperature: 14°F to 122°F (-10°C to 50°C)
Mains Supply Voltage Fluctuations:
Not to exceed ± 10% of the nominal voltage
Classification:
Class I
No patient applied parts
Ordinary Equipment
Sterilization not recommended nor required
Do not operate in the presence of flammable anesthetics
Mode of Operation:
Continuous
Electrical Rating:
240 VAC, 60 Hz, 40 amp, 1Ø
Facility Power Requirements:
240 VAC, 60 Hz, 50 amp, 1Ø
Water Requirements:
- Potable water from building (hot water supply)
- Minimum water pressure: 40 psi (276 kPa)
- Minimum flow rate: 3.2 gal/min (12.1 liters/min)
- Maximum temperature: ; 140°F (60°C) hot water
-Minimum temperature hot water 100°F (38°C)
Waste Drain:
Standard garden hose fitting
Drain Fitting Location:
Back panel of the unit
Fluid Capacities:
- Detergent Reservoir: Use Container (340 mL ± 10 mL)
- Chamber 4.1 gallons (~17 liters)
Rinse Water Consumption:
- Cleaning Cycle – 17 liters
- Two Hot Water Rinses – 34 liters
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
35
SYSTEM INSTALLATION
The Instrument Processor System must be properly installed and its electromechanical
performance verified prior to use . An overview of the installation process is provided
below.
Warning: Failure to properly install the Instrument Processor System according to the
instructions and specifications provided in the Installation / Service Manual may result in
not properly cleaning the endoscopes, damage to the endoscopes, or system malfunction.
It is recommended that the user consider connecting the Instrument Processor System to the
facility’s filtered water system to provide additional assurance of feed water quality.
Initial Setup.
Location / Initial Service Connections
 Area of endoscope use
 Special procedures suite
 Appropriate Endoscope Lab
 Central Sterile Services
 Private Practice – designated endoscope reprocessing area
Utility Requirements (within 5 feet of the Processor)
 Electrical
240 VAC, 60 Hz, 50 Amp, single phase
Hospital grade connection
 Water
Quality: Tap or other potable
Pressure: 40-50 PSI (276-345 kPa)
Flow: 4 GPM (15 LPM) @ 40 PSI (276 kPa)
Temperature: Hot Water 100°F-140°F (38°C-60°C)
Machine Connection: ¾ inch (1.9cm) female hose fitting
 Drain
Machine Connection: ¾ inch (1.9cm) male hose fitting
Sink, user facility waste water system or sanitary sewer system (no back
pressure).
 Space and Weight Requirements
Space requirements are the physical space that is taken by the device without
regard to the additional space needed for a person to walk around it.
Cover Open: Dimensions (height/width/depth) 60.25 x 48 x 343/8 inches (153 x
121.9 x 87.3 cm)
Weight: 1500-1700 lbs. (680-771 kg)
Operating Weight: 1500-1700 lbs (680-771 kg)
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
36
INSTALLATION
Move the Instrument Processor System to the installation location using a hand pallet
truck/dolly and ensure that the Instrument Processor System is not damaged.
CAUTION: The Instrument Processor System must be installed on a level surface or be
adjusted to level after installation. Adjust the leveling pads after moving the
Instrument Processor System into its final location. During installation, be
sure that the protective covers remain in place, to prevent installation debris
from entering the system.
This section contains the information needed to properly install the Instrument Processor
System. The section covers water and electrical supply, filtration, drainage, space
requirements, specifications, drawings and schematics.
Prior to installation of the Instrument Processor System, the prospective site should be
checked to ensure that it meets the needs of the unit. See the " Instrument Processor System
Pre-installation Checklist" below.
SYSTEM PRE-INSTALLATION CHECKLIST
This checklist should be consulted to determine that the installation site meets the
requirements. If all requirements are met, the site is suitable for the installation of the
Instrument Processor System unit.
1.
2.
3.
4.
5.
6.
Does the installation area have a level floor?
Is the incoming water pressure sufficient (40 psi minimum)?
Does the water supply provide hot water in accordance with Instrument Processor
System requirements (Hot Water 100°F-140°F (38°C-60°C)?
Is there a water line that provides 4 gal/min (~15 L/min)?
Is the electrical outlet accessible for connecting the unit?
Will the installation site accommodate the unit's dimensions?
- Is there adequate space to operate around the unit for refilling the detergent
reservoir?
- Is there adequate space to open front door of the unit to allow for water filter
maintenance?
- Is there adequate space to open the Chamber Lid?
Proper site preparation, installation, and service of the Instrument Processor System will
ensure trouble-free use. Langford Systems field service personnel will install the machine
and provide basic training to hospital personnel.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
37
SYSTEM SUPPLY
SPACE REQUIREMENTS
The Instrument Processor System must be installed on a level surface. However, the unit
has leveling pads so that minor adjustments may be made after installation, if necessary.
To prevent damage to the wheels, AFTER LEVELING THE WHEELS MUST BE OFF
THE FLOOR. See Figure 28 below.
Leveling Pad Adjustment
Wheel
Leveling Pad
Figure 28: Caster Leveling Pad Adjustment
The dimensions of the Instrument Processor System, with the chamber lid closed and open,
are as follows:
Chamber Lid Closed: Dimensions (height/width/depth) 45-5/8 x 48 x 343/8 inches
(115.9 x 121.9 x 87.3 cm)
Chamber Lid Open: Dimensions (height/width/depth) 60.25 x 48 x 343/8 inches (153 x
121.9 x 87.3 cm)
Space requirements are the physical space that is taken by the device without regard to the
additional space needed for a person to walk around it. In addition to these measurements,
the Instrument Processor System has external components, such as the water hook-ups.
Accordingly, the following clearances are recommended: The water inlet line should have
8-10 in. clearance on the back side of the unit, so that the water hose can be connected and
are accessible.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
38
WATER SUPPLY
For optimum cycle performance, the incoming water supply must be at least 3/4in. dia. on a
dedicated line with a minimum flow rate of 4 gal/min (15 liters/min), at minimum of 30-80
psi (207-552 kPa). Use a hot water supply with a minimum of 100°F (38°C) and a
maximum temperature of 140°F (60°C).
Water Inlet
Water Inlet
Detail
Figure 29: Water Inlet Connection
Install an incoming water shut-off valve in the supply line before the system. Cold water is
not required
To improve water quality provided to the system, the facility may consider connecting the
Instrument Processor System to a pre-filtered water supply consisting of a coarse filter and
bacterial retentive filter in series. The incoming water supply needs to comply with the
above requirements for proper operation of the system.
ELECTRICAL SUPPLY
The Instrument Processor System must be electrically grounded and plugged into or
connected to a fused, branched circuit. Power supply requirement is 240 VAC, 60Hz, 50A
single phase.
Main power switch
Main power cord
Figure 30: electrical Supply Connection
WARNING: Install the unit so that the power cord, electrical outlet or main circuit breaker
is accessible at all times. Instrument Processor System’s for use in the U. S.
are supplied with a hospital-grade grounding plug that can be connected to any
standard 240 VAC outlet.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
39
DRAIN
CAUTION: The system drain operates via active pumps from the rear of the Instrument
Processor System.
Drain outlet – detail
Drain outlet
Figure 31: Drain Outlet Connection
The Instrument Processor System's drain connection is located on the back of the cabinet.
Consult your facilities department if you are installing multiple units. The user facility drain
system must have a capacity of at least 5 gal/min (18.9 L/min).
DRAIN PUMP SERVICE
The drain pump may need servicing after many hours of operation. There are two types of
servicing needs that may occur:
1. Circuit Breaker – Usually caused by the pump stalling, drawing too much current,
and tripping the internal circuit breaker. This results in drain time-out exception (i.e.
EX. #29), because the pump is disabled until the circuit breaker is reset.
2. Impeller Damage – Wear that causes individual vanes of the impeller to break off.
This causes the drain pump to run slower, resulting in longer cycle times from taking
longer to drain, and eventually resulting in drain time-out exception exception (i.e.
EX. #29).
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
40
DRAIN PUMP CIRCUIT BREAKER
When the Drain Pump Circuit Breaker breaks, there is usually a loud momentary vibrating
sound that the pump makes (which is much louder than the noise pump normally makes
when it’s operating and lasting a few seconds) before tripping the circuit breaker. Follow
these steps to reset the circuit breaker (Drain Time-out Exception):
CAUTION: Power to the Instrument Process System is “ON”.
1. Open the front service door.
2. Figure 32, Figure 33 and Figure 35 show where the Drain Pump Circuit Breaker is
located in the machine. Reach around inside and reset the pump circuit breaker as
shown in Figure 36.
3. Acknowledge the exception(s), and allow wash recovery to complete.
4. If the machine displays a service exception, enter the Service User ID to clear the
exception.
5. Endoscope in the chamber must be reprocessed.
Circuit
Breaker is
behind panel
Service
Switch is
behind panel
Aluminum
panel front
Figure 32: Behind service door and behind aluminum panel are Drain Pump Circuit Breaker
and Service Switch
Circuit
Breaker
side view
Service
Switch
Aluminum
panel side
Figure 33: Side view behind service door showing Service Switch, as well as side of Circuit
Breaker
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
41
Service
Switch
Aluminum
Panel
Service
Switch
Aluminum
Panel side
Figure 34: Reaching in to press Service Switch shown from front (left) and side (right)
Push to
turn ON
Circuit
Breaker
Tripped
Circuit
Breaker ON
Aluminum
Panel Back
Figure 35: Back view of aluminum panel showing drain pump Circuit Breaker tripped (left)
and ON (right)
Circuit
Breaker
Circuit
Breaker
Figure 36: Reaching in to turn on drain pump Circuit Breaker shown from front (left) and
side (right)
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
42
REPLACING THE IMPELLER
The impeller needs to be checked periodically (see table), or when the chamber fails to
properly drain within the time-out by producing an exception, and the drain appears to be
free of blockages. The following shows a damaged impeller:
Broken vane,
replace impeller
Missing vane, most
likely sent to drain,
replace impeller
Cracked vane,
replace impeller
Figure 37: Damaged Drain Pump Impeller
If the Drain Pump Impeller needs inspecting or replacing, make sure the Instrument
Processor System is in the following condition prior to inspecting / changing the impeller:
1. Ensure that there is no wash in progress (power is OFF or green Ready light is ON
on the control panel).
2. Turn the POWER to the Instrument Processor System OFF (if ON).
To inspect or replace the Drain Pump Impeller, follow these steps:
1. Obtain a small tray to collect water that will fit under the Drain Pump Impeller
housing.
2. Open the front service door by using a tool/pen to depress the door latch button.
3. Place the small tray under the Drain Pump Impeller housing to collect the water
runoff.
4. Remove the four screws to remove the impeller housing cover (see Figure 26) and
examine the Drain Pump Impeller using the guideline shown in Figure 37.
5. If the impeller needs to be replaced, remove the damaged impeller and replace with a
new impeller.
6. Replace the impeller housing cover and four screws.
7. Remove the small tray, and discard the collected water runoff.
8. Keeping the door switch closed perform the following steps:
a) Turn the POWER to the Instrument Processor System ON
b) Follow the prompts to initialize the machine.
c) If the impeller cover is leaking, tighten it until the leak stops.
d) Run a wash without an endoscope (use Inst ID #000000000, using the “Other”
instrument type).
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
43
e) While the chamber is performing bypass and draining during the first cycle,
check for leaks by visually inspecting the impeller cover.
f) If the impeller cover is leaking at any time during the drain, tighten it until the
leak stops.
g) If the leak does not cease by the time the chamber has completed draining,
contact Langford IC Systems customer service.
9. Release the switch that detects the service door open, and close the service door.
10. Acknowledge the exception that occurs from having released the switch, and allow
wash recovery to complete.
11. Any endoscope left in the chamber must be reprocessed.
CHAMBER FULL, COVER CLOSED, CANNOT OPEN
This can happen for a variety of causes, but specifically from the chamber being left full
when the machine is shut off, as the auto latch won’t allow a full chamber to open upon
power up:
1. There is a Drain Pump Circuit Breaker failure or some other fault that results in a
Service Exception, for which the user is forced to turn off the machine without being
able to drain the chamber.
2. There is a power failure or the user shuts off power during a wash, leaving the
chamber full when the machine powers up.
If this happens, follow these steps after powering on the machine:
1. Power cycle the machine “OFF” and “ON”. The machine will enter the Initialization
step, follow the prompts to initialize the machine.
2. Press the “OPEN” switch to set the Auto-latch home position and open the chamber
lid.
3. Call Service if initialization does not reset the machine to operational mode
CAUTION: If a cycle is either aborted or cancelled, endoscopes should be
reprocessed or rinsed manually immediately upon cancellation of the cycle in order
to prevent detergent from drying (concentrating) on the endoscope(s). The endoscope
being processed is not safe to use.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
44
SOFTWARE SET-UP SCREENS
Once the Instrument Processor System hardware is in place, the Instrument Processor
System’s electronic control system needs to be set up. Figure 38, provides an example of
the User Interface Control Panel:
Figure 38: User Interface Control Panel
The following sequence of screens will guide the user through the process of initially
configuring the operating system.
If adding a new user go to page 51, and enter the Administrator ID (22222).
Note : This ID may be changed by the administrator.
Turn on the power switch on the back panel of the Instrument Processor System. The
following screen will be displayed. (Software Version may be different than shown)
L ANGFORD IC S YSTEMS
Instrument Processing
System
Software Version: 1.0
Copyright (C) 2014
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
45
Press Enter or wait 5 seconds to advance to the next screen.
10/29/14
10:26
Instrument Processor
Exception:
Electronics Failure
#67
Press ENTER to Clear
The following screen appears because the clock has not been initialized. Press Enter to
advance to the next screen.
10/29/14
10:26
Instrument Processor
Exception:
Clock not Running
#9
Press ENTER to Clear
The following Time and Date Menu screen appears. Press Enter to begin entering the
month, day and year and time in hours and minutes using the numeric key pad.
10/29/14
10:26
Instrument Processor
Time and Date Menu
10/29/14
10:26
mm/dd/yy
hh:mm
Save new time?
No
Yes
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
46
Once the Date and Time have been entered the Cursor will move to highlight No. Use the
arrow keys to toggle to the Yes as indicated below and press Enter.
10/29/14
10:26
Instrument Processor
Time and Date Menu
10/29/14
10:26
mm/dd/yy
hh:mm
Save new time?
No
Yes
Press Enter to Continue.
10/29/14
10:26
Instrument Processor
User: TEST
Process Canceled
Press ENTER to Continue
The following screen appears and a Service Code must be entered.
10/29/14
10:26
Instrument Processor
Service Required
Service Exception: 67
Electronics Failure
Turn Off Power, OR
Enter Service ID: 00000
Then Press ENTER
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
47
Use the keypad to enter the Service Code “99998”, and then press Enter as indicated on the
following screen.
10/29/14
10:26
Instrument Processor
Service Required
Service Exception: 67
Electronics Failure
Turn Off Power, OR
Enter Service ID: 99998
Then Press ENTER
The system requires a user password to start a processing or access menu functions. At
installation the system Administrator password is required to be changed from the default
password. This is indicated by “Service Required” displayed. This then allows the user's ID
number (and users name/initial) to be set-up.
10/29/14
10:26
Instrument Processing
System
Service Required
Enter User ID: 00000
Then Press ENTER
The factory default Administrator password of 99999 must be entered, and the
Administration Screen is then displayed.
10/29/14
10:26
Instrument Processing
System
Service Required
Enter User ID: 99999
Then Press ENTER
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
48
To change the Administrator’s Password, use the keypad to enter the Administrator’s
Password consisting of 5 unique digits.
10/29/14
10:26
Instrument Processor
Administration
00: 99999 Admin
01: 00000
02: 00000
03: 00000
04: 00000
Press ENTER to Edit
Press ENTER to edit the field. In this example the entry 22222 is entered as the new
Administrator password.
10/29/14
10:26
Instrument Processor
Administration
00: 22222 Admin
01: 00000
02: 00000
03: 00000
04: 00000
Press ENTER to Change
Press ENTER to accept the new Administrator password. The screen advances to the next
entry to allow the Administrator to assign unique user ID numbers to new user.
10/29/14
10:26
Instrument Processor
Administration
00: 22222 Admin
01: 00000
02: 00000
03: 00000
04: 00000
Press ENTER to Edit
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
49
Press ENTER to edit the field. Use the keypad to assign unique user ID numbers and user
name or initials. In the following example, the user ID number ‘11111’ and user name
‘SFT’ is entered. The cursor then moves to the next line for the next entry.
10/29/14
10:26
Instrument Processor
Administration
00: 22222 Admin
01: 11111 SFT
02: 00000
03: 00000
04: 00000
Press ENTER to Change
Once the user ID numbers/Names have been entered, then press CANCEL to Return to the
Login Screen.
After pressing Enter the following Splash screen appears for about 5 seconds followed by
the Starting Machine Screen.
L ANGFORD IC S YSTEMS
Instrument Processing
System
Software Version: 1.0
Copyright (C) 2014
Wait for 5 seconds or pressing ENTER, the screen is prompt to initialize the system. If the
screen displays an exception, see troubleshooting section for further actions. Initializing the
system requires user to close and latch the lid. Do not insert any endoscopes.
10/29/14
10:26
10/29/14
10:26
10/29/14
10:26
Instrument Processing
System
Instrument Processing
System
Instrument Processing
System
Close Lid
Initializing Machine
Do not insert instrument
Initializing Machine
Latch Lid
Initializing Machine
Press START to Continue
,
, And
After user closes and latches the lid, and presses START, the Instrument Processor System
will begin to operate by performing a self-test of the operating systems. The following
screens appears.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
50
10/29/14
10:26
10/29/14
10:26
Instrument Processing
System
Instrument Processing
System
Please Wait...
Initializing Machine
Press STOP to Cancel
Enter ID: 00000
Then Press ENTER
VERIFYING THE INSTALLATION
CAUTION
If you cannot verify the operation of the Instrument Processor System, do not use the unit;
contact Customer Service.
To confirm that the system is ready to operate, perform a test processing without any
chemicals.
Place the Instrument Processor Trays into the processing chamber and insert four blank
inserts into the baffle.
Follow the steps under the Cleaning and Disinfection Operation Section above.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
51
ADMINISTRATIVE SETUP FUNCTIONS
The System Administrator can control access to the operation of the system by establishing
User ID for each user. Once the User IDs are set up the system cannot be operated unless
assigned User ID is entered. The Administrator can override certain operations and has a
special password called the Administrator password that is established at the time of
installation.
Administration Screen
The Administration Screen is displayed when the user enters the administration password.
This screen allows the administrator to change or enter user passwords.
Administration Screen
10/29/14
10:26
Instrument Processor
Administration
00: 22222 Admin
01: 11111 TEST
02: 11112 Jeremy
03: 11113 Steve
04: 11114 Dara
Press CANCEL to Return
Note: When a user password ID is added or changed, the printer will print a receipt of all
the user ID’s. The administrator should either keep the receipt in a safe place or
shred it, but it should not be left on the printer.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
52
USER MENU
In the User Menu, the system allows user to Set Date/Time, Modify Instrument ID’s, View
Records, and Shutdown. The USER MENU cannot be accessed during a processing cycle.
When pressed the MENU key on the keypad, USER MENU Screen is displayed.
USER MENU Screen
1/29/14
10:26
Instrument Processor
User: TEST
Set Date/Time
Modify Inst IDs
View History
Shutdown
Exit
SET TIME/DATE
The user can set the time and date
Set Time /Date Screen
10/29/14
10:26
Instrument Processor
Time and Date Menu
10/29/14
10:26
mm/dd/yy
hh:mm
Save new time?
No
Yes
10/29/14
10:26
Instrument Processor
Time and Date Menu
Please Wait...
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
53
Time and date changed. The screen returns to USER MENU.
USER MENU Screen
10/29/14
10:26
Instrument Processor
User: TEST
Set Date/Time
Modify Inst IDs
View History
Shutdown
Exit
Modify Instrument ID
The user can modify instrument ID. Select the Modify Inst IDs and press ENTER. Modify
Instrument ID screen is displayed. Select the instrument ID needs to be changed then press
ENTER. The screen displays 3 options, Modify Type, Delete ID and Change ID.
Modify Instrument IDs Screen
10/29/14
10:26
Instrument Processor
Modify Instrument IDs
Find IDs
000001417
Modify ID 000003692
Add ID
000004850
000000325 000029417
000000417 000044605
Select Inst ID to modify
Then Press ENTER
When an Instrument ID is selected, the user can then modify the instrument type, delete the
instrument ID and change the instrument ID.
Modify Instrument ID Screen
10/29/14
10:26
Instrument Processor
Modify Instrument ID
Inst ID: 000029417
Modify Type
Delete ID
Change ID
Select Inst ID Operation
Then Press ENTER
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
54
If Modify Type is selected, choose a new type then press ENTER.
10/29/14
10:26
Instrument Processor
Modify Instrument Type
Inst ID: 000029417
Bronchoscope
Colonoscope
Duodenoscope
Select new type
Then Press ENTER
If Delete ID is selected, confirm the ID number to be deleted, and select YES. The screen
prompts to the USER MENU Screen.
10/29/14
10:26
Instrument Processor
Inst ID: 000029417
Delete Inst ID?
No
Yes
If Change ID is selected, enter the new ID number and press ENTER. If the new ID number
is not 0 or existing ID number, the screen prompts to Modify Instrument IDs Screen. The
user can then select Add ID to add a new instrument ID.
10/29/14
10:26
Instrument Processor
Change Instrument ID
Enter Inst ID: 000000456
Then Press ENTER
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
55
Modify Instrument IDs Screen
10/29/14
10:26
Instrument Processor
Modify Instrument IDs
Find IDs
000001417
Modify ID 000003692
Add ID
000004850
000000325 000029417
000000417 000044605
Select Inst ID to modify
Then Press ENTER
Select Add ID then press ENTER. Add Instrument ID screen is displayed.
Add Instrument ID Screen
10/29/14
10:26
Instrument Processor
Add Instrument ID
Enter Inst ID: 000000000
Then Press ENTER
Enter a valid unique 9 digits number as instrument ID then press ENTER. Modify
Instrument Type Screen is displayed. Select the correct type listed then press ENTER.
10/29/14
10:26
Instrument Processor
Modify Instrument Type
Inst ID: 000029417
Bronchoscope
Colonoscope
Duodenoscope
Gastroscope
Select new type
Then Press ENTER
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
56
VIEW HISTORY
The user can view and print out history record.
10/29/14
10:26
Instrument Processor
View History
10/28/14 18:56 Ex 28
10/28/14 18:55 Ex 17
10/28/14 18:55 Ex 11
10/28/14 18:55 Ex 107
10/28/14 18:55 Ex 114
Press ENTER to Print
10/29/14
10:26
Instrument Processor
Print History
History Record Printed
Press ENTER
Shutdown
The user can shutdown the system. Select the SHUTDOWN and press ENTER.
10/29/14
10:26
Instrument Processor
User: TEST
Shutdown Machine?
No
Yes
Select YES and Press ENTER.
10/29/14
10:26
Instrument Processor
User: TEST
Shutdown Machine?
Remove Instrument
Close and Latch Lid
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
57
Close and latch the Lid.
10/29/14
10:26
Instrument Processor
User: TEST
Ready to power off
The machine’s power can be switched off. The power switch is located in rear panel of the
system.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
58
ACCESSORIES AND SUPPLIES
Langford IC
Systems
Part Number
0003
0001
0004
0005
0006
0007
0002
Description
User Manual
O-ring, Chamber Seal
Detergent Reservoir (empty)
Zymit LF Detergent (12 x 1 liter case)
Printer Paper (10 rolls per box)
Water Filter 2.0 micron Absolute
Pressure Relief valve
Detergent Reservoir (empty)
Zymit Pour Spout
Tray Assembly, Bronchoscope Left
Tray Assembly, Bronchoscope Right
Tray Assembly,
Colonoscope/Duodenoscope Left
Tray Assembly,
Colonoscope/Duodenoscope Right
Insert, Blank
Insert, Bronchoscope
Insert, Colonoscope / Duodenoscope
Insert, Duodenoscope
Insert, Auxiliary
Baffle Seal, Insert
Baffle Seal, Lower
Baffle Seal, Upper
Protective Cap
Proven Process
Medical Devices
Part Number
0592-20015
0143-03297
0592-07672
0592-20012
0169-07484
0169-07320
0143-04320
0592-07672
0592-20014
0169-08399
0169-08400
0169-08401
0169-08404
0169-08405
0169-08406
0169-08407
0169-08984
0169-10142
0169-07239
0169-07237
0169-07238
0143-03854
All other Repairs are performed by an authorized
Service Engineer. See Service Manual
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
59
TROUBLESHOOTING
EXCEPTION CODES
Exception codes are numerical error codes that the Instrument Processor System uses to
provide the User with information, warnings, cautions, and equipment malfunctions during
the processing of medical endoscopes.
Four classes or levels of exceptions are defined:
Level
I
W
Name
Information
Warning
C
Critical
S
Service
Description
Cycle continues to run.
Cycle is automatically aborted and action by the
user is required.
Cycle is automatically aborted and action by the
user is required. Restart of wash is required
Cycle is automatically aborted and Customer
Service is required.
Exception Screen
If there is a warning or error, which requires user acknowledgement, an exception screen is
displayed. The user presses ENTER to continue past the exception. If Service Required
Screen is displayed the Authorized Service Code must be used (see Administrative Setup ),
Enter pressed and condition repaired. A Service Engineer may be required based on the
Exception code. See Exception Code Descriptions in the table below.
10/29/14
10:26
Instrument Processor
Exception:
Invalid User Code
#66
Press ENTER to Clear
10/29/14
10:26
Instrument Processor
Service Required
Service Exception: 88
Hardware Failure
Enter Service ID: 00000
Then Press ENTER
Service personal is required for further action.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
60
LIST OF EXCEPTIONS (ERRORS)
Legend: EX = exception; Level I, W, C or S (I = information; W = warning; C = Level C
error, restart by user required; S = Level S error, service required)
Customer Service: 1-508-337-6500
Code
Display
Description
INFORMATION
Level I
47
Inst ID Database Full
61
62
Invalid Time
Invalid Date
66
Invalid User Code
76
Change Admin Password
98
Drain Half Dry Timeout
No more Instrument ID number storage is available because
the maximum Instrument IDs that can be stored on the
machine are all assigned to valid IDs. Refer to Modify
Instrument ID section for instruction on how to delete
unused Instrument IDs.
The user entered an invalid time. Re-enter correct time.
The user entered an invalid date. Re-enter correct date.
The user entered the wrong password. Re-enter correct
password.
User must change the administration password. Enter a new
Administrator password for the machine.
Drain timeout. One drain sensor went dry, but the other one
stayed wet. Cycle automatically aborts. Endoscope is to be
considered contaminated and the cycle rerun.
WARNING
Level W
84
Chamber Pressure Too High
85
Lid Not Safe To Open
94
Lid Opened
106
Over Absolute Pressure
113
Fill Sensor Error
116
Fill Pressure Exceeded
The chamber pressure was too high during POST. Call a
service technician.
The interlock is still engaged after a wash indicating the lid
is considered not safe to open. A faulty level sensor, a faulty
ball valve, a motor fault are all possible reasons for this
warning being issued. If this condition remains, the lid will
not be allowed to open, requiring service.
The lid was opened during a wash cycle.
The chamber absolute pressure was over the SRS specified
limit during agitation. This exception is posted when the
absolute chamber pressure is too high during agitation. This
could be caused by improper placement of the endoscope in
the tray, a clogged endoscope, or a bad tray. Reposition the
endoscope in a new tray, and try the wash again.
After draining the chamber, LS5 indicated the chamber was
still full.
During a fill, the chamber pressure rose above the minimum
level. This could be caused by a failed solenoid valve, ball
valve, or level sensor. Clean the chamber level sensors. Try
the wash again, and if that doesn’t work, call a service
technician.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
61
Code
Display
117
Drain Pump Failure
123
High Wash Pressure
124
Wash Over Pressure
Description
The chamber pressure did not drop after the drain pump was
turned on.
High differential pressure detected. Verify that the tray,
instrument, and baffle are setup properly. Verify that the
instrument is not blocked. Retry the wash. If the problem
persists, call a service technician.
Very high differential pressure detected. Verify that the tray,
instrument, and baffle are setup properly. Verify that the
instrument is not blocked. Retry the wash. If the problem
persists, call a service technician.
RESTART REQUIRED
Level C
12
Detergent Unlatched
23
Chamber Not Full
28
Hardware Failure
29
Chamber Drain Timeout
50
Bypass Dry, No Water
64
Chamber Too Cold
65
Chamber Too Hot
69
Hot Water Pressure Too Low
Detergent bottle became unlatched while processing an
instrument.
Chamber not full. The chamber is not full (from water
source or from ozone tank) within the specified time limit.
This exception is posted when the chamber has not filled
within the time limit. This could be caused by a filter that
needs to be replaced, by low water pressure, or by a bad
pump. Ensure water pressure is nominal, and the hot water
filter is not clogged. Try the wash again. If the error recurs,
call a service technician.
Hardware failure. The ORP sensor check failed. The first
ORP reading was above the specified value. Contact
Customer Service.
Hardware failure. The chamber did not finish draining
within the specified time limit. This exception is posted
when the chamber has not finished draining within the
timeout, or when a drain level sensor fails. Retry. If error
continues, call Customer Service.
The LS6 or LS7 drain sensors run contiguously dry for at
least 5 seconds while bypassing.
Chamber too cold. The chamber was too cold during
agitation. Cycle automatically aborts. Endoscope is to be
considered contaminated and the cycle rerun. This exception
is posted when the chamber cools below the expected
temperature during agitation. Try the wash again. If the
error recurs, call a service technician.
Chamber too hot. The chamber was too hot during agitation.
Cycle automatically aborts. Endoscope is to be considered
contaminated and the cycle rerun. This exception is posted
when the chamber temperature is above its expected value
during agitation. This should never happen. Try the wash
again. If the error recurs, call a service technician.
Low hot water pressure. The pre-wash self test detected
water pressure was too low. Cycle automatically aborts.
Endoscope is to be considered contaminated and the cycle
rerun. This exception is posted during power on self-test or
during pre-wash self test if the hot water pressure is too low.
Ensure the hot water pressure is 30-80 psig, and try the wash
again.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
62
Code
Display
70
Hot Water Pressure Too High
78
Water Too Cold
79
Service Door Ajar
80
Water Too Hot
86
Weak Agitation
88
Hardware Failure
100
Fill Failure
103
Fill Failure
112
Interlock Opened
119
Hardware Failure
Description
High hot water pressure. The pre-wash self test detected
water pressure was too high. Cycle automatically aborts.
Endoscope is to be considered contaminated and the cycle
rerun. This exception is posted during power on self-test or
during pre-wash self-test if the hot water pressure is too
high. Ensure the hot water pressure is 30-80 psig, and try the
wash again.
Water too cold. The hot water bypass ran longer than the
specified time limit. The hot water was too cold. Cycle
automatically aborts. Endoscope is to be considered
contaminated and the cycle rerun. This exception is posted
when the inlet hot water temperature does not reach the
appropriate temperature within the time limit. Ensure the hot
water temperature is 120-140°F. Try the wash again. If the
error recurs, call a service technician.
Front panel service door is not completely closed.
Water too hot. The hot water temperature was over the
specified limit. Cycle automatically aborts. Endoscope is to
be considered contaminated and the cycle rerun. This
exception is posted when the inlet hot water temperature is
too hot. Ensure the hot water temperature is 120-140°F. Try
the wash again. If the error recurs, call a service technician.
Weak agitation. The agitation cycle had too few good
strokes. Cycle automatically aborts. Endoscope is to be
considered contaminated and the cycle rerun. This exception
is posted when the differential pressure is too low during
agitation. This could be caused by improper placement of
the endoscope in the tray, or a bad tray or incorrect fill level.
Reposition the endoscope in a new tray, , confirm that the
proper baffle insert was used, and try the wash again
Hardware failure. The I2C hardware failed, and did not
communicate. Contact Customer Service.
LS4 failed. The Level sensor #4 (LS4) is wet after the
chamber is drained. Cycle automatically aborts. Endoscope
is to be considered contaminated and the cycle rerun. This
exception is posted when the LS4 sensor indicates wet after
the chamber has been drained. This could be caused by a
drop of water on one of the chamber level sensors. Clean the
chamber level sensors. Try the wash again, and if that
doesn’t work, call a service technician.
LS5 failed. The Level sensor #5 (LS5) is wet after the
chamber is drained. Cycle automatically aborts. Endoscope
is to be considered contaminated and the cycle rerun. This
exception is posted when the LS5 sensor indicates wet after
the chamber has been drained. This could be caused by a
drop of water on one of the chamber level sensors. Clean the
chamber level sensors. Try the wash again, and if that
doesn’t work, call a service technician.
The lid lock was opened during a wash. Contact Customer
Service.
Level sensor, LS6, failed. The Level sensor #6 (LS6) is dry.
Cycle automatically aborts. Endoscope is to be considered
contaminated and the cycle rerun. This exception is posted
when the LS6 sensor indicates that it is dry during the
process of draining a filled chamber. Try the wash again. If
the error recurs, call a service technician.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
63
Code
Display
120
Hardware Failure
121
No Wash Pressure
122
Low Wash Pressure
125
Low Stroke Count
Description
Level sensor, LS7, failed. The Level sensor #7 (LS7) is dry.
Cycle automatically aborts. Endoscope is to be considered
contaminated and the cycle rerun. This exception is posted
when the LS7 sensor indicates that it is dry during the
process of draining a filled chamber. Try the wash again. If
the error recurs, call a service technician.
No differential pressure detected within SRS specified time
limit. Verify that the tray, instrument, and baffle are setup
properly. Retry the wash. If the problem persists, call a
service technician.
Low differential pressure detected. Verify that the tray,
instrument, and baffle are setup properly. Retry the wash. If
the problem persists, call a service technician.
Verify that the tray, instrument, and baffle are setup
properly. Verify that the instrument is not blocked. Retry
the wash. If the problem persists, call a service technician.
SERVICE REQUIRED
Level S
1
Electronics Failure
2
Electronics Failure
3
Electronics Failure
4
Electronics Failure
5
Electronics Failure
6
Electronic Failure
7
Electronics Failure
8
Electronics Failure
9
Clock Not Running
10
Hardware Failure
11
Hardware Failure
12
Detergent unlatched
14
Hardware Failure
15
Hardware Failure
16
Hardware Failure
17
Hardware Failure
19
Software Failure
22
LS10 Sensor
Electronics Failure. The software failed. Contact Customer
Service.
Electronics Failure. The software program is corrupted.
Contact Customer Service.
Electronics Failure. The CPU RAM failed the POST.
Contact Customer Service.
Electronics Failure. A CPU register failed the POST.
Contact Customer Service.
Electronics Failure. The NV RAM failed the POST.
Contact Customer Service.
Electronics Failure. The CPU ALU failed the POST.
Contact Customer Service.
Electronics Failure. The real time clock was not accessible.
Contact Customer Service.
Electronics Failure. The real time clock and the CPU clock
speeds do not match within 5%. Contact Customer Service.
The real time clock is not running. Contact Customer
Service.
Hardware Failure. Ball valve #1 did not open within the
specified time limit. Contact Customer Service.
Hardware Failure. Ball valve #1 did not close within the
specified time limit. Contact Customer Service.
Detergent unlatched
Hardware Failure. Ball valve #2 did not open within the
specified time limit. Contact Customer Service.
Hardware Failure. Ball valve #2 did not close within the
specified time limit. Contact Customer Service.
Hardware Failure. Ball valve #3 did not open within the
specified time limit. Contact Customer Service.
Hardware Failure. Ball valve #3 did not close within the
specified time limit. Contact Customer Service.
Software attempted to simultaneously fill the ozone tank
and chamber with cold water. Contact Customer Service.
After draining the chamber, LS10 indicated the chamber was
still full. Contact Customer Service.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
64
Code
Display
41
Hardware Failure
42
Hardware Failure
43
Transformer Too Hot
44
Hardware Failure
45
Hardware Failure
46
Hardware Failure
48
Inst ID Record Error
53
Software Failure
54
History Record(s) Lost
55
History Record(s) Lost
56
History Record(s) Lost
57
History Record(s) Lost
58
History Record(s) Lost
59
History Record(s) Lost
60
Hardware Failure
63
Software Failure
67
Electronics Failure
68
Software Failure
73
Hardware Failure
74
Hardware Failure
81
Software Failure
89
Electronics Failure
90
Hardware Failure
91
Hardware Failure
Description
The T1 sensor failed. The T1 sensor was outside of the
specified limit. Contact Customer Service.
The T2 sensor failed. The T2 sensor was outside of the
specified limit. Contact Customer Service.
The transformer has overheated and the machine will
momentarily be powered down. Contact Customer Service.
The J10 connector on the Power Board was just
disconnected. Contact Customer Service.
The J11 connector on the Power Board was just
disconnected. Contact Customer Service.
The J12 connector on the Power Board was just
disconnected. Contact Customer Service.
At least one corrupt or uninitialized Instrument ID was found
in memory or a software error occurred accessing Instrument
IDs. Contact Customer Service.
Software failure. Not enough display memory. Contact
Customer Service.
History Record(s) Lost. A history record could not be
serialized. This is a software error, and should never
happen. Contact Customer Service.
History Record(s) Lost. A history record with an invalid
header was found. This could be caused by a software error,
or by corrupted NV RAM. Contact Customer Service.
History Record(s) Lost. Two history records marked first
record were found. This could be caused by a software
error, or by corrupted NV RAM. Contact Customer Service.
History Record(s) Lost. No history record marked first
record was found. This could be caused by a software error,
or by corrupted NV RAM. Contact Customer Service.
History Record(s) Lost. Too many history records were
detected. This is caused by a software error, and should
never happen. Contact Customer Service.
History Record(s) Lost. A corrupted history record was
detected. This is caused by an NV RAM failure. Contact
Customer Service.
Hardware failure. The motor did not reach 5Hz within the
specified time limit. Contact Customer Service.
Software failed. Contact Customer Service.
Electronics failure. The NV RAM has never been
initialized, or has been corrupted. This should only happen
one time at the factory. Contact Customer Service.
Software failure. There was not enough memory. Contact
Customer Service.
J40 on the Power Board was just disconnected. The
120VAC is not being supplied to Power Board. Contact
Customer Service.
The J14 connector on the Logic Board was just
disconnected. Contact Customer Service.
Software failure. The watchdog interrupt fired. This could
be caused by a software error. Contact Customer Service.
Electronics failure. The CPU ADC hardware failed.
Contact Customer Service.
The T3 sensor failed. The T3 sensor was outside of the
specified limit. Contact Customer Service.
The J30 connector on the Power Board was just
disconnected. Contact Customer Service.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
65
Code
Display
92
Hardware Failure
93
Hardware Failure
97
Hardware Failure
99
Hardware Failure
104
Hardware Failure
105
Hardware Failure
108
Hardware Failure
109
Hardware Failure
110
Hardware Failure
111
Software Failure
114
Hardware Failure
115
Hardware Failure
126
Software Failure
Description
The J31 connector on the Power Board was just
disconnected. Contact Customer Service.
The J32 connector on the Power Board was just
disconnected. Contact Customer Service.
The PS4 sensor failed. The PS4 sensor was outside of the
SRS specified limit. Contact Customer Service.
The J15 connector on the Logic Board was just
disconnected. Contact Customer Service.
Motor fault 1-hardware failure. The motor driver hardware
asserted fault line 1. Contact Customer Service.
Motor fault 2-hardware failure. The motor driver hardware
asserted fault line 2. Contact Customer Service.
The PS1 sensor failed. The PS1 sensor was outside of the
specified limit. Contact Customer Service.
The PS2 sensor failed. The PS2 sensor was outside of the
specified limit. Contact Customer Service.
The PS3 sensor failed. The PS3 sensor was outside of the
specified limit. Contact Customer Service.
Software failure. Software ran out of stack space. Contact
Customer Service.
Hardware failure. The I2C hardware failed, and did not
communicate. Contact Customer Service.
The ORP measurement of the ozone in the ozonated water in
the ozone tank raised above the maximum allowable level
required by the SRS while first waiting for the ozone to raise
to the minimum allowable level. Contact Customer Service.
Software exception occurred that was out of range. Contact
Customer Service.
UNUSED CODES
30-40
43-52
71-72
75
77
101-102
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
UNUSED
Unused.
Unused.
Unused.
Unused.
Unused.
Unused.
DIAGNOSTIC SYSTEM (SEE SERVICE MANUAL)
A serial port is provided for communications with outside systems for Service Technician
ONLY. Utilities shall be provided to help identify and fix faulty hardware, and test
software.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
66
APPENDIX A – ENDOSCOPE COMPATIBILITY
ENDOSCOPES THAT SHOULD NOT BE PROCESSED WITH THE INSTRUMENT
PROCESSOR SYSTEM
1. Endoscopes that are not compatible with the Zymit LF Detergent based on
compatibility testing or literature information of material compatibility with the
System process.
2. Endoscopes that cannot withstand a temperature of 140°F (60°C) should not be
processed with the Instrument Processor System as damage to the endoscope may
result.
COMPARABLE ENDOSCOPES
The following tables provide a listing of endoscopes that are comparable to those tested by
Langford IC Systems during the qualification of the Instrument Processor System.
Table 1 –Bronchoscopes used in Instrument Processor System
Qualification Studies
Model No.
Olympus® BF-20D
Olympus® BF-P20D
Olympus® BF-30
Olympus® BF-P30
Olympus® BF-160
Insertion tube
Channel size
Length
5.9 mm
2.0mm
55cm
6.0 mm
5.3 mm
2.2mm
2.0mm
55cm
55cm
Table 2 –Colonoscopes used in Instrument Processor System
Qualification Studies
Model No.
Insertion tube Channel size Length
Olympus® CF-10L
13.3 mm
3.2 mm
168cm
Olympus® 1T20L
Olympus® CF-30L
13.3 mm
3.2 mm
168cm
Olympus® CF-140L
Olympus® CF-140L
12.9 mm
3.7 mm
168cm
Olympus® CF-Q160AL
12.8 mm
3.7 mm
168cm
Table F.3 –Duodenoscopes used in Instrument Processor System
Qualification Studies
Insertion
Model No.
Channel size
Length
tube
Olympus® JF-130
11.5mm
3.2 mm
125cm
Olympus® TJF-130
12.5mm
4.2 mm
125cm
Olympus® TJF-140F
12.5mm
4.2 mm
125cm
Olympus® TJF-160VF 11.3mm
4.2 mm
124cm
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
67
APPENDIX B – SYMBOLS
Symbol
Used for
SN
Serial Number
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
68
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
69
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
70
APPENDIX C – REFERENCES
Reference: “Designer Handbook—The Care and Cleaning of Stainless Steel”, published by
the Specialty Steel Industry of North America (SSINA), 3050 K Street, N.W. Washington,
D.C. 20007.
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
71
71027 SS Care&Cleaning l/o
6/15/01
2:10 PM
Page 2
DESIGNER
HANDBOOK
THE CARE AND
CLEANING
O F S TA I N L E S S
STEEL
Stainless
Steel
The
Value
Option
®
71027 SS Care&Cleaning l/o
6/15/01
2:10 PM
Page 3
TABLE OF CONTENTS
INTRODUCTION
IN
..............................................
1
Cle
clos
WHAT IS STAINLESS STEEL
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
eac
the
ALLOY TYPES
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
FORMS AND FINISHES
.......................................
2
pha
stai
(roo
CLEANING OF STAINLESS STEEL
Types of surface contaminants
. . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
stai
. . . . . . . . . . . . . . . . . . . . . . . . . . . .3
of c
to p
TYPES OF CLEANERS AND METHODS
EFFECTIVE CLEANING METHODS
. . . . . . . . . . . . . . . . . . . . . . . .4
. . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
surf
per
is e
CARE OF STAINLESS STEEL
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
to c
T
LISTING OF STAINLESS STEEL PIPE
pra
AND TUBE MANUFACTURERS
............................
6
dur
incl
con
that
met
gen
surf
han
T
that
clea
Acknowledgments
met
The Specialty Steel Industry
are
of North America (SSINA)
use
acknowledges that this new
with
handbook contains information
req
originally published by the
app
Committee of Stainless Steel
Producers, American Iron and
Steel Institute, which no longer
exists. Current SSINA member
companies were represented
on that committee.
The SSINA wishes to
acknowledge the contributions
of the Nickel Development
Institute and its consultant,
Technical Marketing Resources
(Pittsburgh, PA) for help in
preparing the contents of
this handbook.
The Specialty Steel Industry of the North
America (SSINA) and the individual
companies it represents have made every
effort to ensure that the information
presented in this handbook is technically
correct. However, neither the SSINA nor its
member companies warrants the accuracy of
the information contained in this handbook
or its suitability for any general and specific
use. The SSINA assumes no liability or
responsibility of any kind in connection with
the use of this information. The reader is
advised that the material contained herein
should not be used or relied on for any
specific or general applications without first
securing competent advice.
71027 SS Care&Cleaning l/o
6/15/01
2:10 PM
Page 4
INTRODUCTION
WHAT IS STAINLESS STEEL
Cleanliness and stainless steel are
Stainless steel is not a single alloy, but rather the name applies to a group of iron-based
closely related and, in many applications,
alloys containing a minimum 10.5% chromium. Other elements are added and the
each is dependent upon the other. In
chromium content increased to improve the corrosion resistance and heat resisting
the handling of food, chemicals,
pharmaceuticals and in the use of
properties, enhance mechanical properties, and/or improve fabricating characteristics.
stainless steel as a construction material
There are over 50 stainless steel grades that were originally recognized by the American
(roofs, wall panels, entry ways, signs, etc.),
Iron and Steel Institute (AISI). Three general classifications are used to identify stainless
stainless steel provides the degree
steel. They are:
of corrosion resistance that is necessary
1) Metallurgical structure.
to prevent product contamination or
surface rusting. However, stainless steel
2) The AISI numbering system (200, 300 and 400 series numbers).
performs best when clean — cleanliness
3) The Unified Numbering System, which was developed by the American Society
is essential for maximum resistance
for Testing Materials (ASTM) and the Society of Automotive Engineers (SAE) to apply
to corrosion.
to all commercial metals and alloys.
This handbook describes various
The various types of stainless steel are detailed in a designer handbook, “Design
practices for cleaning stainless steel
during manufacture and in use. This
Guidelines for the Selection and Use of Stainless Steel,” available from the Specialty
includes methods for removing free-iron
Steel Industry of North America (SSINA). Several other publications are also available,
contamination on stainless steel surfaces
including: “Stainless Steel Fabrication,” “Stainless Steel Fasteners,” “Stainless Steel
that may have been picked up from
Finishes,” “Stainless Steel Specifications,” and “Stainless Steel Architectural Facts,”
metalworking tools; and for removing
to mention a few.
general accumulation of dirt, grime and
surface stains that occur during normal
handling and exposure to the elements.
ALLOY TYPES
The reader should keep in mind
304 is the basic chromium-nickel
316 offers more corrosion-resistance
that there are few specific rules for a
austenitic stainless steel and has
through the addition of molybdenum.
cleaning procedure. Accordingly, the
been found suitable for a wide range
This grade is desirable where the
methods discussed in this handbook
of applications. It is the most readily
possibility of severe corrosion exists,
are suggestions. Each manufacturer or
available in a variety of product forms.
such as heavy industrial atmospheres
user, after obtaining competent advice
This grade is easy to form and fabricate
and marine environments.
with respect to their individual
with excellent resistance to corrosion.
requirements, should select methods
appropriate to those requirements.
304L is the low carbon version of
304. It is sometimes specified where
extensive welding will be done.
h
316L is the low carbon version of 316.
430 is a straight chromium ferritic
stainless steel with lower corrosion
resistance than the 300 series. It is
principally employed for interior use.
y
s
y of
k
fic
th
n
rst
1
71027 SS Care&Cleaning l/o
6/15/01
2:10 PM
FORMS AND FINISHES
Page 5
Table 1 CLASSIFICATION OF STAINLESS STEEL PRODUCT FORMS
Width
Coil and cut lengths:
Mill finishes Nos. 1, 2D & 2B
Pol. finishes Nos. 3, 4, 6, 7 & 8
under 3/16” (4.76mm)
under 3/16” (4.76mm)
24” (609.6mm) & over
all widths
—
Strip
Cold finished, coils or cut lengths
Pol. finishes Nos. 3, 4, 6, 7 & 8
under 3/16” (4.76mm)
under 3/16” (4.76mm)
under 24” (609.6mm)
all widths
—
finishes, there are numerous special finishes
Plate
Flat rolled or forged
3/16” (4.76mm) & over over 10” (254mm)
used for architectural applications. Some of
Bar
the more delicate finishes require special
Hot finished rounds, squares,
octagons and hexagons
precautions when cleaning and the advice
Hot finished flats
Table 1. Sheet finishes are shown in Table 2.
Sheet
Bar product conditions and finishes are
shown in Table 3; Plate is shown in Table 4.
In addition to the common mill and polished
of the supplier should be obtained.
Description
Tabl
Dimensions
Thickness
Sheet, strip, plate, and bar forms are shown in
Item
The most common special finishes are
Cold finished rounds, squares,
octagons and hexagons
shown in Table 5.
Cold finished flats
Wire
Pipe &
Tubing
Cold finishes only: (in coil)
Round, square, octagon,
hexagon, and flat wire
—
1/8” (3.18mm) to
8” (203mm) incl.
—
1/8” (3.18mm) to
4 1/2” (114mm)
Con
—
Diameter or Size
Ann
hea
—
1/4” (6.35mm) & over
1/4” (6.35mm) to
10” (254mm) incl.
—
Hot
—
over 1/8” (3.18mm)
3/8” (9.53mm) to
4 1/2” (114mm)
—
Ann
to h
*
**
***
under 3/16” (4.76mm)
under 3/8” (9.53mm)
—
Several different classifications, with differing specifications, are available. For information on standard
sizes, consult your local Steel Service Center or the SSINA.
Extrusions Not considered “standard” shapes, but of potentially wide interest. Currently limited in size to
approximately 6 1/2” (165.1mm) diameter, or structurals.
Tabl
Con
Hot
Hot
or h
Table 2 STANDARD MECHANICAL SHEET FINISHES
Unpolished or Rolled Finishes:
No. 1 A rough, dull surface which results from hot rolling to the specified thickness followed by annealing
and descaling.
No. 2D A dull finish which results from cold rolling followed by annealing and descaling, and may perhaps get
a light roll pass through unpolished rolls. A 2D finish is used where appearance is not of primary concern.
No. 2B A bright, cold-rolled finish resulting in the same manner as No. 2D finish, except that the annealed and
descaled sheet receives a final light roll pass through polished rolls. This is the general-purpose cold-rolled
finish that can be used as is or as a preliminary step to polishing.
Polished Finishes:
No. 3 An intermediate polish surface obtained by finishing with a 100-grit abrasive. Generally used where a
semifinished polished surface is required. A No. 3 usually receives additional polishing during fabrication.
No. 4
A polished surface obtained by finishing with a 120-150 mesh abrasive, following initial grinding with coarser
abrasives. This is a general-purpose bright finish with a visible “grain” which prevents mirror reflection.
No. 6
A dull satin finish having lower reflectivity than No. 4 finish in a medium of abrasive and oil. It is used for
architectural applications and ornamentation where a higher luster is undesirable, and to contrast with
brighter finishes.
No. 7
A highly reflective finish that is obtained by buffing finely ground surfaces but not to the extent of completely
removing the “grit” lines. It is used chiefly for architectural and ornamental purposes.
No. 8
The most reflective surface obtained by polishing with successively finer abrasives and buffing extensively
until all grit lines from preliminary grinding operations are removed. It is used for applications such as mirrors
and reflectors.
Hot
hea
or p
Hot
and
Hot
cold
opt
Hot
trea
and
*S
Tabl
Fin
Sel
Sw
Eng
Dis
Ang
Abr
Sel
Em
Elec
Epo
Spu
Pla
2
Per
71027 SS Care&Cleaning l/o
6/15/01
2:10 PM
Page 6
Table 3 CONDITIONS & FINISHES FOR BAR
e
Conditions
Surface Finishes*
Hot worked only
(a) Scale not removed (excluding spot conditioning)
(b) Rough turned**
(c) Pickled or blast cleaned and pickled
Annealed or otherwise
heat treated
over
Annealed and cold worked
to high tensile strength***
(a) Scale not removed (excluding spot conditioning)
(b) Rough turned
(c) Pickled or blast cleaned and pickled
(d) Cold drawn or cold rolled
(e) Centerless ground
(f) Polished
(d) Cold drawn or cold rolled
(e) Centerless ground
(f) Polished
CLEANING OF
STAINLESS STEEL
Stainless steels need to be cleaned for
aesthetic considerations and to preserve
corrosion resistance. Stainless steel is
protected from corroson by a thin layer of
chromium oxide. Oxygen from the
atmosphere combines with the chromium
in the stainless steel to form this passive
chromium oxide film that protects from
further corrosion. Any contamination of the
surface by dirt, or other material, hinders
m)
* Surface finishes (b), (e) and (f) are applicable to round bars only.
** Bars of the 4xx series stainless steels which are highly hardenable, such as Types 414, 420, 420F, 431, 440A,
440B and 440C, are annealed before rough turning. Other hardenable grades, such as Types 403, 410, 416 and
416Se, may also require annealing depending on their composition and size.
*** Produced in Types 302, 303Se, 304 and 316.
this passivation process and traps corrosive
agents, reducing corrosion protection. Thus,
some form of routine cleaning is necessary
to preserve the appearance and integrity of
the surface. Stainless steels are easily
Table 4 CONDITIONS & FINISHES FOR PLATE
d
ser
cleaned by many different methods. They
Condition and Finish
Description and Remarks
actually thrive with frequent cleaning, and,
Hot rolled
Scale not removed. Not heat treated.
Plates not recommended for final use in this condition.*
unlike some other materials, it is impossible
Hot rolled, annealed
or heat treated
Scale not removed. Use of plates in this condition is generally confined
to heat resisting applications. Scale impairs corrosion resistance.*
cleaning. The effect of surface/pattern
Hot rolled, annealed or
heat treated, blast cleaned
or pickled
Condition and finish commonly preferred for corrosion resisting and
most heat resisting applications.
designs that allow for maximum rain cleaning
Hot rolled, annealed, descaled
and temper passed
Smoother finish for specialized applications.
Hot rolled, annealed, descaled
cold rolled, annealed, descaled,
optionally temper passed
Smooth finish with greater freedom from surface imperfections
than the above.
Hot rolled, annealed or heat
treated, surface cleaned
and polished
Polished finishes: refer to Table 2.
to “wear out” stainless steel by excessive
roughness, grain/pattern orientation, and
(exterior applications) should be considered.
Types of surface contaminants
Dirt - Like any surface that is exposed to
the environment, stainless steel can get dirty.
Dirt and soil can consist of accumulated
dust and a variety of contaminates that
come from many sources, ranging from the
* Surface inspection is not practicable on plates which have not been pickled or otherwise descaled.
wind to everyday use. These contaminates
will vary greatly in their effect on
appearance and corrosivity and ease of
Table 5 SPECIAL STAINLESS STEEL FINISHES
ely
rs
removal. While some may be easily
Finish
Selective Polishing
Description
Selective polishing of unprotected areas to produce patterns
removed, others may require specific
Swirl
Swirl patterns obtained with grinding wheels or stainless steel wire brushes
necessary to identify the contaminate or
Engine Turn
Rings, circles, or overlapping circles created with a CNC-operated machine
experiment with various cleaners.
Distressed
Random scratch pattern covering entire surface
Angel Hair
Finer random scratch pattern
Abrasive Blast
Matte, uniform, blast media choice determines appearance
Selective Etching
Etching of selected surface areas to obtain a pattern
Embossing
Raised patterns pressed into the stainless by rolling
Electrochemical
Applied to cut sheets, numerous colors (also called INCO or light interference coloring)
Epoxy Paint
Applied to coils, full range of colors
Sputtering
Thin layer of a colored material applied to surface
Plating
Electroplating with colored and/or precious metals
Perforated Designs
Simple or elaborate patterns
cleaners for effective removal. It may be
Frequently, warm water with or without a
gentle detergent is sufficient. Next in order
are mild non-scratching abrasive powders
such as typical household cleaners. These
can be used with warm water, bristle
brushes, sponges, or clean cloths.
3
71027 SS Care&Cleaning l/o
6/15/01
2:10 PM
Page 7
Shop oil and Grease - Shop oils, which may
not a problem on dull finishes, or those
wool should be avoided as they may leave
carry grease, grit and metal chips,
surfaces finished with a coarse polishing
particles embedded on the surface which
commonly produce surface soiling after
grit. The best preventative measure is to
Job
can lead to RUSTING.
many shop operations. Greases and other
avoid using abrasive cleaners unless
Ro
contaminates may also soil surfaces in food
absolutely necessary. When abrasives are
amount of vinegar can be added to the
preparation and many other household and
needed, first experiment on an
scouring powder.
commercial situations. These soils may be
inconspicuous area. A “soft abrasive,” such
corrosive in themselves or may not allow the
as pumice, should be used. Abrasives can
surface to maintain passivity, and so
permanently damage some colored and
periodic removal is a necessity. Initially,
highly polished finishes. Advice should be
which leave water spots, it is advisable to
soap or detergent and water may be tried or
obtained from the finish supplier when
wipe the surface completely with dry towels.
a combination of detergent and water plus a
cleaning special finishes. Many cleaners
Fingerprints and Stains - Fingerprints and mild
solvent. The removal of oil and grease from
contain corrosive ingredients which require
stains resulting from normal use in
stainless steel parts by immersion in
thorough post-clean rinsing with clean
consumer and architectural applications are
chemical solvents is frequently used with
water; however, thorough rinsing is
the most common surface contaminates.
cold-formed or machined parts that are
recommended for all cleaning procedures.
Fortunately, these usually affect only
laden with lubricants. This process, in its
Clean Water and Wipe - The simplest, safest,
appearance and seldom have an effect on
simplest form, consists of bringing liquid
and least costly method that will adequately
corrosion resistance. They are easy to
solvent into contact with the surface to be
do the job is always the best method.
remove by a variety of simple cleaning
cleaned and allowing dissolution to take
Stainless surfaces thrive with frequent
methods. Fingerprints are probably the
place; for example, washing a surface with
cleaning because there is no surface
most troublesome marks to remove from the
trichloroethylene or similar liquid or stirring a
coating to wear off stainless steels. A soft
surface of smooth polished or bright
batch of small parts in a container of
cloth and clean warm water should always
finished stainless steel. Fortunately, they
solvent. Non-halgenated solvents, such as
be the first choice for mild stains and loose
can be removed with a glass cleaner or by
acetone, methyl alcohol, ethyl alcohol,
dirt and soils. A final rinse with clean water
gentle rubbing with a paste of soda ash
methyl ethyl ketone, benzene, isopropyl
and a dry wipe will complete the process
(sodium carbonate) and water applied with
alcohol, toluene, mineral spirits, and
and eliminate the possibility of water stains.
a soft rag. Once again, this should be
turpentine work well.
Solvent Cleaning - Organic solvents can be
Ordinary carbon steel brushes or steel
For more aggressive cleaning, a small
Cleaning should always be followed by
rinsing in clean hot water.
When water contains mineral solids,
followed by a thorough warm water rinse.
Many of these solvents are widely
used to remove fresh fingerprints and oils
There are several special surface finishes
used as individual cleaners, but there are
and greases that have not had time to
where fingerprints present special problems:
thousands of blended or compound cleaners
oxidize or decompose. The preferred
polished No. 6, etched, some abrasive
on the market. Users are advised to contact
solvent is one that does not contain chlorine,
blasted finishes, and light electrochemical
suppliers of solvents for information on their
such as acetone, methyl alcohol, and
colors applied over satin or brushed finishes.
applications on stainless steel.
mineral spirits. There are many
(NOTE: there are several special finishes
designed to withstand fingerprints:
embossed, swirl patterns, lined patterns, etc.).
compounded or blended organic cleaners
TYPES OF CLEANERS
AND METHODS
General Precautions
to optimize both cleanability and safety
attributes. Cleaning can be accomplished
In selecting cleaning practices, consider the
by immersing smaller articles directly into
possibility of scratching and the potential for
the solvent, wiping with solvent-impregnated
post-cleaning corrosion caused by
cloths, or by sophisticated vapor or spray
incompletely removed cleaners. Scratching
methods. The wiping technique sometimes
can occur on a bright mirror finish by
leaves a streaked surface.
cleaners that contain hard abrasives, or
even by “grit” in wash water. This is usually
4
that are commercially available and attempt
EF
Fin
Stu
and
Gre
Bu
Bak
Gre
*NO
Om
wi
Hou
into
and
ma
wel
abr
intro
surf
abr
par
pro
mo
surf
with
of c
low
ass
rins
stat
that
71027 SS Care&Cleaning l/o
6/15/01
2:10 PM
Page 8
EFFECTIVE CLEANING METHODS
Job
Cleaning Agents*
Comments
Routine Cleaning
Warm Water, Soap, Ammonia, Detergent
Apply with sponge or soft cloth.
Can be used on all finishes.
e
ch
CARE OF STAINLESS STEEL
Fingerprints and Smears 3M Stainless Steel Cleaner and Polish,
Arcal 20, Lac-O-Nu, Lumin Wash, O’Cedar
Cream Polish, Stainless Shine
Provides barrier film to minimize
fingerprints. Can be used on all finishes.
Stubborn Stains
and Discoloration
3M Stainless Steel Cleaner and Polish,
Allchem Concentrated Cleaner, Samae,
Twinkle, Cameo Copper Cleaner, Grade FFF
or Grade F Italian Pumice, Whiting or talc,
Liquid Nu Steel, Copper’s or Revere Stainless
Steel Cleaner, Household Cleaners, Lumin
Cleaner, Zud Restoro, Sta-Clean, Highlite,
Allen Polish, Penny-Brite, Copper-Brite
Rub lightly, using dry damp cloth, in
the direction of polish lines on the
stainless steel.
Scotch-Brite Power Pad 2001, Easy-Off,
De-Grease-It, 4% to 6% hot solution
of such agents as tri-sodium polyphosphate,
5% to 15% caustic soda solution
Excellent removal on acids, all finishes.
Particularly useful where rubbing
is not practical.
Any good commercial detergent or
caustic cleanser.
Apply with sponge or soft cloth in
direction of polish lines.
while in storage, being processed or during
use, the greater the assurance of optimum
corrosion resistance. Some tips on the care
of stainless steel are listed below:
1) Use paper or other protective
an
e
The cleaner stainless steel can be kept
wrapping on the surface of the
re
stainless steel until processing
is complete.*
2) Handle stainless steel with clean
gloves or cloths to guard against
stains or finger marks.
3) Avoid the use of oily rags or greasy
s.
st,
ely
Grease and Blood
Burnt-on or
Baked-on Foods
Grease and Oil
se
er
s.
e
ne,
s
mpt
d
ted
y
es
4) Do routine cleaning of exposed
surfaces. Buildings with window
*NOTE: Use of proprietary names is intended only to indicate a type of cleaner and does not constitute an endorsement.
Omission of any proprietary cleanser does not imply its inadequacy. All products should be used in strict accordance
with instructions on package.
s
cloths when wiping the surface.
washing systems can utilize this
method to clean exterior panels.
5) Where possible, after cleaning,
rinse thoroughly with water.
6) Cleaning with chloride-containing
Household Cleaners - Household cleaners fall
or low in chloride. The cleaning method
into two categories: detergent (non-abrasive)
generally employed with these cleaners is to
and abrasive cleaners. Both are effective for
apply them to the stainless surface and
can scratch or burnish a mill-rolled
many mild dirt, stain, and soil deposits, as
follow by cloth wiping, or to wipe directly
finish. On polished finishes,
well as light oils such as fingerprints. The
with a cleaner-impregnated soft cloth. In all
rubbing or wiping should be done
abrasive cleaners are more effective but
cases, the cleaned surface should be
in the direction of the polish lines,
introduce the possibility of scratching the
thoroughly rinsed with clean water and
NOT across them.
surface. However, the degree of
wiped dry with a soft cloth if water streaking
abrasiveness will vary greatly with the
is a consideration.
particular product, and some brands will
Commercial Cleaners - Many commercial
produce noticeable scratching on only the
cleaners compounded from phosphates,
most highly polished and some colored
synthetic detergents, and alkalis are
surfaces. All of these cleaners vary widely
available for the cleaning of severely soiled
with respect to their acidity and the amount
or stained stainless surfaces. When used
of chloride they contain. A neutral cleaner
with a variety of cleaning methods, these
low in chloride is preferred unless the user is
cleaners can safely provide effective
assured that the surface can be thoroughly
cleaning. Manufacturers should be
rinsed after cleaning. The fact that the label
consulted and their recommendations
states “for stainless steel” is no guarantee
followed whenever using cleaners of this
that the product is not abrasive, not acidic,
kind. The general precautions stated above
detergents must be avoided.
7) Even the finest cleaning powders
8) DO NOT USE SOLVENTS in closed
spaces or while smoking.
*Many adhesive-backed papers and plastic
sheets or tape applied to stainless steel for
protection “age” in fairly short periods of time
and become extremely difficult to remove.
Manufacturers should be contacted
regarding information as to how long
protective films or paper can be left in place.
also pertain to these cleaners.
5
APPENDIX D – TRAYS AND INSERTS
The following shows the trays and inserts to be used with various endoscopes for washing
with the Instrument Processor System.
Endoscope
Type
Bronchoscope
Colonoscope
Duodenoscope
Left Tray
Qty
Right Tray
Qty
0169-08399
0169-08401
0169-08401
1
1
1
0169-08400
0169-08404
0169-08404
1
1
1
Duodenoscope
0169-08401
1
0169-08404
1
Endoscope
Insert
0169-08406
0169-08407
0169-08407
0169-08984
0169-08407
Qty
Blank Insert
Qty
2
2
2
1
1
0169-08405
0169-08405
0169-08405
2
2
2
0169-08405
2
BRONCHOSCOPE TRAYS AND INSERTS
0169-08399
0169-08406
AND
AND
0169-08400
0169-08405
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
72
COLONOSCOPE TRAYS AND INSERTS
0169-08401
0169-08407
AND
AND
0169-08404
0169-08405
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
73
DUODENOSCOPE TRAYS AND INSERTS
0169-08401
AND
0169-08404
0169-08984
AND
0169-08405
OR
0169-08984
AND
0169-08407
AND
0169-08405
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
74
APPENDIX E – RELIEF VALVE INSPECTION AND TEST
(This inspection and test is to be performed by qualified service personnel only.)
CAUTION: Turn off power.
Remove the rear panel, top rear panel, docking station, lower right front panel, right hand
side panel and the user interface (UI) panel.
UI panel
Top Rear panel
Rear panel
RHS side
panel
Remove the relief valve from the chamber. Attach a regulated air source to the valve with a
range of 0 to 30psi. Set the air source to 20psi and slowly increase to 15psi. Confirm that
the valve opens at 15psi. Re-install the valve.
If the valve does not open at 15psi, then set the air source to 16psi and adjust the valve until
it begins to open. Again, set the air source to 20psi and slowly increase to 15psi. Confirm
that the valve opens at 15psi. Re-install the valve.
If the valve cannot be adjusted as described, then a replacement valve must be installed.
Install the (pre-tested) replacement valve and retest to 15 psi as described above.
Adjust
relief valve
here
Disconnect
valve here
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
75
WARRANTY
TBD
Customer Service (USA): 1-508-337-6500
No part of this publication may be reproduced, photocopied, stored on a retrieval
system, or transmitted without the express written consent of Langford IC Systems, Inc.
76