Transcript
Perfomance Testing and Calibration Introduction 4 Warning Service procedures should be performed by qualified service personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. Introduction 4-1. This chapter contains performance testing procedures that can be used to verify GMM operation within customer specifications. A separate set of calibration procedures is also included; use these procedures for closed-case calibration of the GMM. Equipment connections for performance testing and calibration are shown at the end of this chapter. The appropriate connection configuration is referenced in each procedure. Required Test Equipment 4-2. Equipment required for performance testing (“P”), calibration (“C) is listed in Table 4-1. The 5700A is used to calibrate dc volts, ac volts, dc/ac mA/µA, and Ohms. The 5725A driven by the 5700A is used to calibrate dc and ac amps. Warning Ensure that the calibrator is in standby mode before making any connection between the calibrator and the GMM. Dangerous voltages may be present on the leads and connectors. Alternative Test Equipment (Fluke 5500A) 4-3. The Fluke 5500A can also be used as the source where a calibrator is called for in performance testing and calibration. Some performance test points change when a 5500A is used; these changes are noted in the following tables: • Table 4-5. AC Volts RMS Performance Test • Table 4-15. AC Amps RMS Performance Test • Table 4-16. AC Amps Average Performance Test In ohms calibration, the 5500A can provide the values called for by the GMM; no value modification is necessary. For a general discussion of calibrator use and accuracy, refer to Guardbanding with Confidence by David Deaver. This publication is available from the Fluke Corporation, Everett, WA and is published in the 1994 NCSL (National Conference of Standard Laboratories) Workshop and Symposium Proceedings, pages 383-394. 4-3