Download TDX® Wheelchairs

Transcript
SECTION 11—FORMULA INVISIBLE SUPER LOW TILT BATTERIES
SECTION 11—FORMULA INVISIBLE
SUPER LOW TILT BATTERIES
WARNING
Read and understand the information and warnings in Handling and Replacing
Batteries on page 98 before performing these procedures.
Unless otherwise indicated, make sure power to the wheelchair is OFF before
performing these procedures.
After ANY adjustments, repair or service and BEFORE use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may occur.
NOTE: If there is battery acid in the bottom of the battery tray or on the sides of the battery(ies),
apply baking soda to these areas to neutralize the battery acid. Before reinstalling the existing or
new battery(ies), clean the baking soda from the battery tray or battery(ies) being sure to avoid
contact with skin and eyes. Determine source of contamination. Never install/reinstall a battery
with a cracked or otherwise damaged case.
Removing/Installing Front/Rear Shrouds
NOTE: For this procedure, refer to FIGURE 11.1 on page 116.
Removing
CAUTION
Place the wheelchair in a well ventilated area where work can be performed without risking damage to carpeting or floor covering.
1. Ensure the wheelchair seat is completely down.
2. Verify the joystick ON/OFF switch is in the OFF position.
3. Remove the two thumb screws that secure the rear shroud to the wheelchair.
4. TRRO and TRBKTS Options Only - Remove the top controller shroud.
5. Remove the rear shroud from the wheelchair.
6. Remove both mounting screws from side of battery door.
7. Remove battery door from front of wheelchair.
8. TRRO and TRBKTS Options Only - Perform the following steps:
A. Remove the four hex screws that secure the battery support bracket to the battery
box.
B. Remove the battery support bracket.
9. Tilt seat back approximately halfway.
Part No 1114809
115
TDX™ Wheelchairs