4-9
(f)
How to install the
Phototherapy
Lamp Head
Slide the support block (900IW303) from the top of the CosyCot™ Infant
Warmer column down to the desired height.
ii Tighten the Shelf Mounting Block knob securely.
iii Slide Support Pin into Shelf Mounting Block.
iv Lamp can be installed on both sides of CosyCot™Infant Warmer by
undoing both grubscrews and reversing support pin in both 900IW501
and Swing Limit.
v Connect the Lamp head power cable to Phototherapy Power Module
900IW502 and tighten the connector securely.
vi Move the Lamp head if not position as needed. Do not put the lamp head
directly under the CosyCot™ Infant Warmer heater.
(g)
Using Phototherapy
Position the Medela® Phototherapy Lamp head as close to the baby as
possible, but not directly under the heat path of the CosyCot™ Infant
Warmer.
ii Ensure the baby’s eyes and genitals are protected from the phototherapy
light.
iii Make sure the Phototherapy Power Module fitted to the lower segment
of the CosyCot™ is switched on. Use the switches at the rear of the
Phototherapy Lamp to control the light output. Do not look directly at the
light.
iv Regularly monitor the baby.
(h)
Light output monitoring
i
i
i
ii
When the Lamp head turns on, electronic monitoring of the blue
light is activated. If the light output falls below 75% the warning light
illuminates.
If the warning light stays on for more than 45 minutes after the tubes
have been switched on, ALL the blue tubes must be replaced.
WARNING
This product in intended for use by authorized and qualified personnel who are aware of currently
known hazards and benefits of phototherapy for treatment of hyperbilirubinemia.
All personnel must be familiar with the operation of this phototherapy system before using the
device with patients.
Protect baby’s eyes and genitals from Phototherapy light.
Do not look directly at the blue light.
Ensure Phototherapy Lamp head is securely fitted to the mounting bracket.
Electric shock hazards exist beneath the control panels. Ensure qualified personnel carry out all
service and maintenance issues.
Only qualified personnel should carry out service and maintenance procedures.
To avoid the possibility of electric shock hazard when performing cleaning and maintenance
procedures, disconnect the Infant Warmer from the mains power.
Accessories