Operating Instructions HPT 100 High Vacuum Transmitter PG0004KJL/G (0509) Contents Page 1. 2. Safety Instructions ............................................................................................... 3 Understanding The HPT 100 ............................................................................... 4 2.1. For Your Orientation ........................................................................................................................ 4 2.2. Product Identification ..................................................................................................................... 4 Package Contents ........................................................................................................................... 4 2.3. Product Description ........................................................................................................................ 5 Connection Options ......................................................................................................................... 5 Proper Use ........................................................................................................................................ 5 Improper Use ................................................................................................................................... 5 3. Installation ............................................................................................................ 6 3.1. 3.2. 3.3. 3.4. 3.5. Preparation For Installation ........................................................................................................... 6 Vacuum Connection ........................................................................................................................ 6 Electrical Connection ..................................................................................................................... 7 Connections Plan ............................................................................................................................. 8 Disassembly ...................................................................................................................................... 9 4. Operation ............................................................................................................ 10 4.1. General............................................................................................................................................. 10 4.2. Communication............................................................................................................................... 11 4.3. Transmitter Equalization ............................................................................................................... 13 5. 6. 7. Degas ................................................................................................................... 14 Error Signals ...................................................................................................... 14 Maintenance, Service ....................................................................................... 15 8. Technical Data ................................................................................................... 16 8.1. Data List .................................................................................................................................. 16 8.2. Dimensions ............................................................................................................................. 17 8.3. Gas Correction Factor ........................................................................................................... 17 9. Accessories ....................................................................................................... 18 10. Supplementary Information .............................................................................. 19 Please note: Current operating instructions are also available at www.Lesker.com 2 1. Safety Instructions ☞ Read and follow all the instructions in this manual. ☞ Inform yourself regarding: - Hazards which can be caused by the unit - Hazards which can arise in your system ☞ Comply with all safety and accident prevention regulations. ☞ Check regularly that all safety requirements are being complied with. ☞ Take account of the ambient conditions when installing the HPT 100. The protection type is IP40/ IP 54. The unit is protected against the ingress of foreign bodies. ☞ Adhere to the applicable regulations and take the necessary precautions for the process media used. ☞ Consider possible reactions between the materials and the process media. ☞ Consider possible reactions of the process media due to the heat generated by the product. ☞ Do not carry out any unauthorised conversions or modifications on the unit. ☞ When returning the unit to us please note the shipping instructions. ☞ Before you begin to work, find out whether any vacuum components are contaminated. ☞ Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts. ☞ Communicate the safety instructions to other users. This document is based on the component software version 010500. Pictogram Definition WARNING WARNING CAUTION Danger of an electric shock. Danger of personal injury. Danger of damage to the unit or system. 3 2. Understanding The HPT 100 2.1. For Your Orientation Abbreviations used HPT = Hot Cathode Pirani Transmitter CPT = Compact Ceramic Transmitter RPT = Resistive Piezo Pirani Transmitter PPT = Pirani Pressure Transmitter DPG = Digital Pressure Gauge DPS = Digital Power Supply Operating Instruction In The Text ➡ Here, you have to do something. Validity These operating instructions describe the installation and operation of the Digital High Vacuum Transmitter HPT 100 with the part numbers KTR34130 and KTR34330. The part number is shown on the rating plate. Technical modifications are reserved without prior notification. 2.2. Product Identification Please state the information contained on the rating plate when corresponding with the Kurt J. Lesker Company. Product Identification HPT 100 RS 232 RS 485 2 14 12 10 4 8 6 Package Contents The following items are included in the package: – HPT 100 – Protective cover – Operating instructions 4 2.3. Product Description The digital Transmitter HPT 100 enables total pressure measurements in the range of 7.5 x 10-10 ... 750 Torr (1x10-9 ... 1000 mbar). The transmitter can be operated in conjunction with the Lesker Digital Measuring and Control Units, with the Digital Power Pack and the Windows Measuring Software DokuStar (please see "Accessories") on a PC or with customer related, digital signal evaluation. The transmitter communicates optionally via a Serial Interface RS 232 or RS 485. The individual functions are described in more detail in the section "Communication" . The Transmitter HPT 100 has been tested and accepted in compliance with EN 61010/VDE 0411 ”Safety Equipment For Electrical Components” . Connection Options The HPT 100 provides connection options for: – RS 232/485 (for communication with the display unit) or – Voltage supply Proper Use The Digital Transmitter HPT 100 serves exclusively to provide total pressure measurements in the range (7.5 x 10-10 ... 750 Torr (1 x 10-9 ... 1000 mbar). It may only be connected to components specifically provided for this purpose. Improper Use The following is regarded as improper: – The use for purposes not covered above and, in particular, – The connection to components not allowed for in their operating instructions; – The connection to components containing touchable, voltage carrying parts. No liability will be accepted for claims arising from improper use. The user bears the responsibility with respect to the process media. 5 3. Installation 3.1. Preparation For Installation CAUTION Unauthorized modifications or alterations to the unit are not allowed and will void the warranty. Installation location: Indoor; max. 2000 m above mean sea level. The following is applicable for open buildings and operations rooms which are not fully air conditioned: Temperature: +5°C - +40°C (+ 41°F ... +100 °F) Relative humidity: 5 - 85%, non-condensing. Air pressure: 86 kPa - 106 kPa (12.5 psi - 15.4 psi) 3.2. Vacuum Connection CAUTION Dirt and damage have an adverse effect on the function of these vacuum components. Please take account of the instructions with regard to cleanliness and damage prevention when using vacuum components. Set-Up Configuration The gauge may be mounted in any orientation. A horizontal to upright orientation is to be preffered to keep condensates and particles from getting into the measuring chamber. Connection ➡ Remove the protective cover. Do not discard the protective cover as it is required during maintenance work. ➡ Make vacuum connection via a small flange QF16 (DN 40 KF) with version KTR34130 or DN 40 CF with version KTR34330. WARNING Where excess pressure is involved in the vacuum system > 750 Torr (1 bar): Inadvertent opening of elements under stress can lead to parts flying around. ➡ Only use stressed elements which can be opened and closed (for example QF16 (DN 16KF) clamp). 6 WARNING With excess pressure in the vacuum system of 1125 to 3000 Torr (1.5 to 4 bar): With QF (KF) flange connections elastomer washers (for example O-rings) cannot withstand the pressure and this can lead to personal injury resulting from the emission of process media. ➡ Use centering rings with an outer over pressure ring (such as the QF40-075-OPR). 3.3. Electrical Connection CAUTION Only connect/disconnect the cable to/from the gauge when the power is disconnected. Kurt J. Lesker Company model DPG 101 controller: ➡ First make the connection to the Kurt J. Lesker Company DPG 101 controller by means of Data Cable, part number KT348203T (please see "Accessories") on the RS 232 Serial Interface of the controller. ➡ Rotate the transmitter address switch to RS232 operation (factory default setting). The transmitter address switch can be found on the rating plate side of the unit. Kurt J. Lesker Company model DPS 101 power pack: ➡ To connect a transmitter to a PC, use the Kurt J. Lesker Company Power Pack Model DPS 101 and set the transmitter to RS 232. Kurt J. Lesker Company model DPG 109 controller: Up to nine transmitters in series can be connected to this controller. ➡ Integrate the transmitter using Data Cable part number KT348403T (please see "Accessories") into the measuring chain. ➡ Rotate the Transmitter Address switch to RS485 operation with the appropriate address. The transmitter address switch can be found on the rotating plate side of the unit (please also refer to DPG 109 Operating Instructions). Kurt J. Lesker Company model DPS 109 power pack: ➡ To connect up to nine transmitter to a PC, use the Kurt J. Lesker Company Power Pack Model DPS 109 and set the transmitter to RS 485. Set address switch 7 3.4. Connection Plan Connection Plan HPT 100 (optional) RS 485 or RS 232 DPS 101 or 24V DC Transmitter DPG 101 Vacuum chamber DPG 109 PC Sub-D socket arrangement on the transmitter casing 54321 9876 8 PIN Settings Pin 1: Pin 2: Pin 3: Pin 4: Pin 5: Pin 6: Pin 7: Pin 8: Pin 9: n.c. RS232, RxD RS232, TxD n.c. gnd VCC RS485, DRS485, D+ n.c. DPS 109 3.5. Disassembly WARNING Possible contaminated parts. Contaminated parts can cause personal injuries. ➡ Inform yourself regarding possible contamination before beginning work. ➡ When working with contaminated parts be sure to follow the relevant instructions and take the necessary protection measures. CAUTION Dirt and damage have an adverse effect on the function of these vacuum components. ➡ Please take account of the instructions with regard to cleanliness and damage prevention when using vacuum components. ➡ Shut down the gauge head. ➡ Disconnect the connecting cable. ➡ Detach the gauge heads from the vacuum apparatus. ➡ Place the protective cover in position. 9 4. Operation 4.1. General Once voltage has been supplied the transmitter is ready to operate. A stabilization time of 510 minutes must be considered of. It is usually better to ensure that the transmitter remains in operation all the time. Use Of The Filament Protective Panel During normal UHV applications, the transmitter can be operated without the filament protection. When applications are involved where greater levels of contamination are anticipated, the protective panel, which is included, must be used. It is important to observe the necessity for cleanliness when fitting the panel. The conductance value of the transmitter is reduced when the panel is used. In this case, operations should be restricted to 7.5 x 10-6 Torr (1 x 10-5 mbar). The measured pressure depends on the type of gas. For matching to various types of gas a respective gas correction factor for the Pirani or Bayard-Alpert unit can be set; this generates the correct measurement value in the pressure range below 7.5 x 10-3 Torr (1x 10-2 mbar). The gas type correction factor for air and N2 has been set in the works to 1.00 (please also refer to Section 8.3.”Gas Correction Factor”). The transmitter has been factory adjusted. Re-adjustment may be necessary in the event of extreme temperatures or ageing etc. and is carried out with a connected vacuum controller via the PC-Software "DokuStar" or customer specific in accordance with the communication instructions. Please note that before equalization the transmitter must have been in operation for approximately 5 minutes (warm-up time) at the respective pressure. For correct zero point equalization, the pressure in the vacuum chamber must be below 3.75 x 10-5 Torr (5 x 10-5 mbar). Baking out If the bakeout temperature is higher than 60°C (140°F) the electronic unit must be removed. To avoid some mistakes the serial number of the complete unit is fixed up to the transmitter print. 10 4.2. Communication The transmitter communicates via the serial interface with the respective display units (please see "Accessories"), but alternatively can also be controlled directly via the RS 232 or RS 485 by means of the interface protocol. The Interface protocol is in ASCII format which means that all data bytes are representable symbols with an ASCII code ≥ 32 with the exception of the telegram final symbol carriage return (CR, 13). For example the character string and the ASCII code corresponds as follows: 0 1 2 0 0 7 4 0 0 2 = ? 1 0 8 CR 48 49 50 48 48 55 52 48 48 50 61 63 49 48 56 13 The following possible telegrams are transferred with 9600 baud, 8 data bits, no parity and 1 stop bit. Write request to transmitter a1 a2 a3 1 0 p1 p2 p3 I1 I2 0 p1 p2 p3 0 2 0 p1 p2 p3 I1 I2 data c1 c2 c3 c2 c3 c2 c3 CR Read request to transmitter a1 a2 a3 0 = ? c1 CR Answer of transmitter a1 a2 a1, a2, a3: p1, p2, p3: I1, I2: data: c1, c2, c3: cR: a3 1 data c1 CR Unit address (“001” ... “015”, also see chapter 3.3 ) number (see table) data length (see table) see table or “NO_DEF”: wrong parameter number, “_LOGIC”: Access fault, “_RANGE”: Out of range checksum, sum of all ASCII up to checksum modulo (remainder) 256, for example “0120074002=?” => 48+49+50+ ... +61+63=620, 620 modulo 256 = 108 => “108” Carriage return (ASCII code 13) 11 # 040 1) Name DeGas 303 Error code • “06” 312 349 Software version Transmitter type • • “06” “06” 740 Current pressure value in mbar • “06” 741 Adjustment low/high pressure Correction factor (Pirani) Correction factor (Bayard-Alpert) 742 743 1) writable readable I1, I2 • • “01” • 2) • “03” • • “06” • • “06” Date “0”: abort respectively is not activ “1”: activate respectively is activ “000000”: no error “Err001”: transmitter defective “Err002”: memory defective character string; six signs long “.... A1” (32 32 32 32 65 49): CPT 100 “.... A2” (32 32 32 32 65 50): RPT 100 “.... A3” (32 32 32 32 65 51): PPT 100 “.... A4” (32 32 32 32 65 52): HPT 100 “aaaabb” for a.aaa · 10 (bb-20) e.g. “123423” => 1.234 · 10 (23-20) => 1.234 · 103 => 1234 mbar “000”: low pressure “001”: high pressure “abcdef” for abcd,ef e.g. “000123” => 1.23 “abcdef” for abcd,ef e.g. “000123” => 1.23 1) only HPT 100 2) see transmitter adjustment DeGas (only for HPT 100) The degassing mode can be activated/canceled with a write request and the status be queried with a read request as well. While the transmitter is in the degassing mode, no valid measuring values are transmitted. Transmitter adjustment To equalize a transmitter there must first be set a pressure adjustment point and then the actual pressure value. If the setting for the pressure adjustment point is missing or if another telegram is inserted between the two steps, an error signal is generated when writing the pressure value. Handling Set correction factor to 4.2 of transmitter 2 Read current pressure value of transmitter 12 Adjust transmitter 1 at atmosphere (high pressure) 12 Telegram to Transmitter Write request to transmitter Answer (O.K) Read request to transmitter Answer represents 4.234 10-5 mbar Write request to transmitter Answer (O.K) Write request to transmitter Answer (O.K) Telegram from Transmitter "0021074206000420028CR" "0021074206000420028CR" “0120074002=?108CR“ “0120074006423415040CR“ "0011074103001130CR" "0011074103001130CR" "0011074006100023025CR" "0011074006100023025CR" 4.3. Transmitter Equalization Automatic alignment in the "ur" range (under range) Starting with instrument version 010500, the alignment is run automatically while simultaneously the following conditions apply: – the pressure of the Bayard-Alpert component is < 5.0 • 10-5 mbar. – the measured value of the Pirani component does not change for one minute and does deviate too much from the most recent calibration value. Manually operated alignment when operations via display unit DPG 101 or DPG 109: ➡ Equalization of the transmitter as described in section 4.4. or 4.5. of the DPG instructions. Manually operated alignment communication via RS 232 or RS 485: ➡ Evacuate the vacuum system to the pressure of p<1 x 10-5 mbar. ➡ Set the adjusting pressure point [P:741] to «000» for low pressure and transmit. ➡ Transmit the actual pressure value ([P:740] with «000000» ) for low pressure (represents p < 1.0 • 10-5); ==> ‘’ur’’ (under Range). ➡ Vent to atmospheric pressure with air or N2; afterwards wait about 10 minutes. ➡ Set the adjusting pressure point [P:741] to «001» for high pressure and transmit. ➡ Set the actual pressure value to the atmospheric pressure (e.g. «100023» for 1000 mbar). 13 5. Degas Deposits on the electrode of the hot cathode tubes can cause instability in the measurement signals and lead also to increased desorption in the ultra high vacuum range which, in turn, can lead to a too high measured pressure. In such cases it is desirable, at pressures of below 1.5 x 10-6 Torr (2 x 10-6 mbar), to clean the anode of the transmitter by baking out the adsorbed gas particles. This is carried out (via the connected measuring unit) by starting the Windows Software DokuStar in accordance with the communication instructions relating to the degassing procedure. In this way the anode is warmed up by ohmic heat up to 800 °C (1500°F). Although baking out lasts approximately 3 minutes, the process can be discontinued at any time beforehand. No transmission of measurement values is possible during the time the transmitter is in degassification mode. 6. Error Signals Error Transmitter does not respond A too high measuring value is printed out in the high vacuum Possible Cause No supply voltage Elimination Connect data cable or power pack Incorrect serial interface type or incorrect address set Remove rubber stoppers on rating plate side and set address switch to the serial interface to be used and/or set the correct address (see Section 3.3.). Replace transmitter Degas (see Section 5.) Transmitter defective High transmitter degassing Adjustment necessary, contamination Transmitter defective 14 Re-adjust (see Section 4.1.) Replace transmitter 7. Maintenance, Service The unit requires no maintenance. A damp cloth can be used to wipe away any dirt. Ensure that the unit is first disconnected from the mains power supply. Do Make Use Of Our Service Facilities ➡ "Should a defect occur on the HPT 100 contrary to expectations, please send the transmitter to the authorized service center for repair and for replacement with a new exchange unit." CAUTION Cannot be field serviced. Please note: Units returned to us for repair or maintenance are covered by our general conditions of sale and supply. Products returned to the Kurt J. Lesker Company for service or repair should, if possible, be free of harmful substances (e.g. radioactive, toxic, caustic or microbiological). ➡ Adhere to the forwarding regulations of all involved countries and forwarding companies. ➡ Enclose a completed contamination declaration. Products that are not clearly declared as "free of harmful substances" are decontaminated at the expense of the customer. 15 8. Technical Data 8.1. Data List Description Measurement method Material in contact with media Sealing Feedthrough Filaments Number of filaments Measuring range Pressure Repeatability % of measurement Accuracy % of measurement Measuring cycles Temperature: Operating: Temperature: Storage Temperature: Bake-out Supply: Voltage Supply: Power consumption max. Output Serial interface Unit Torr (mbar) Torr (bar) Torr (mbar) 7.5 x 10-8 ... 0.75 (1 x 10-7 ... 1) +/- 10 % ms °C (°F) °C (°F) °C (°F) 40 +5 ... +50 (+40 ... +120) -40 ... +60 (-40 ... +140) max. 356 VDC 24 W 7.5 digital RS 232 / RS 485 9600 baud, 8 databits, no parity, 1 stopbit 9 poles Sub-D socket, screwable with display unit or software to atmosphere and zero if pressure < 3.75 • 10-5 Torr ( < 5 • 10-5 mbar) IP 40/IP 54 0.4 (0.88) 1.0 147 x 85 x 35 (5.8 x 3.4 x 1.4) 156 x 85 x 35 (6.2 x 3.4 x 1.4) QF16 (DN 40 KF) DN 40 CF max. 1000 (3280), only for RS 485 Electrical connection Adjustment Pirani Protection type Weight Internal volume, approx. Dimensions: KTR34130 KTR34330 Flange: KTR34130 KTR34330 Measuring cable length 16 Value Comments Pirani / Bayard-Alpert Tungsten, Stainless Steel, Nickel Metal Glass, Ceramics Yttriated Iridium, double 2 7.5 x 10-10 ... 750 (1 x 10-9 ... 1000) max. 4 7.5 x 10-8 ... 0.75 (1 x 10-7 ... 1) +/- 5 % kg (lbs.) cm3 mm (inch) mm (inch) m (ft) without electronics incl. flange 8.2. Dimensions Dimensions HPT 100 35.0 mm (1.4") ø 55.0 mm (2.2") 85.0 mm (3.4") 147.0 mm (5.8 ") 8.3. Gas Correction Factor Relative sensitivity in the molecular range (1x10-1 mbar) Measuring diagram HPT 100 (Pirani) HPT 100 1E3 Correction factor Pirani N2 1.00 Air 1.00 H2 0.58 He 1.02 Ar 1.59 CO2 0.89 displayed pressure (mbar) 1E2 1E1 1E0 1E-1 H2 He 1E-2 N2 Ar 1E-3 1E-4 1E-4 Kr 1E-3 1E-2 1E-1 1E0 real pressure (mbar) 1E1 1E2 1E3 Correction factor BayardAlpert He 5.93 H2 2.39 Ar 0.80 CO2 0.74 C3H8 0.32 Ne 3.50 Kr 0.60 Xe 0.41 R12 0.28 17 9. Accessories Description DPS 101, plug power pack with PC connecting cable for one transmitter DPS 109, switching power pack with integrated PC adapter for up to 9 transmitters DPG 101 Controller DPG 109 Controller Data cable for connection to Controller DPG 101 (RS 232) Data cable for connection to Controller DPG 109 or Switching Power Pack DPS 109 (RS 485) Bus termination RS 485 TIC 002 Level converter Software “Dokustar “ Software “Dokustar Plus“ 18 Unit Pieces Number Comments 1 KTG20011 US version 3m (10 ft) 1 1 1 1 KTG25010 KTG10010 KTG15010 KT348203T 3m (10 ft) 1 KT348403T 1 1 1 1 KT348100 KT051519AT KT882500 KT882501 10. Supplementary Information Depending on the configuration of your components further operating instructions are included in the package (please see the table). If, despite every effort by us, information on your products is missing please get in touch with your local Kurt J. Lesker Company representative or contact us using the information on the back cover page. The following operating instructions are available for digital transmitters and accessories: Product CPT 100 PPT 100 RPT 100 HPT 100 DPG 101 DPS 101 DPG 109 DPS 109 TIC 002 Software “DokuStar Plus” Definition Rough vacuum transmitter Medium vacuum transmitter Combination vacuum transmitter High vacuum transmitter Display unit Switching power pack Display unit Switching power pack Level Converter Software for PC data evaluation No. Operating Instructions PG0001KJL PG0002KJL PG0003KJL PG0004KJL PG0006KJL PG0007KJL PG0005KJL PG0008KJL PT0076KJL PG0012KJL 19 Kurt J. Lesker Company 1925 Worthington Ave Clairton, PA 15025 www.Lesker.com