Download Directed Electronics G4203P User manual
Transcript
7 Error Information Module Error Messages Module Error Messages These errors are pump specific. Restart Without Cover The module was restarted with the top cover and foam open. The sensor on the main board detects when the top foam is in place. If the module is restarted with the foam removed, the module switches off within 30 s, and the error message is generated. Probable cause Suggested actions 1 The module started with the top cover and Please contact your Agilent service representative. foam removed. 2 Rear of the module is exposed to strong direct sunlight. Ensure that the rear of module is not directly exposed to strong sunlight. Pressure Above Upper Limit The system pressure has exceeded the upper pressure limit. Probable cause Suggested actions 1 Upper pressure limit set too low. Ensure the upper pressure limit is set to a value suitable for the analysis. 2 Blockage in the flowpath (after the damper). Check for blockage in the flowpath. The following components are particularly subject to blockage: inline filter frit, needle (autosampler), seat capillary (autosampler), sample loop (autosampler), column frits and capillaries with small internal diameters (e.g. 50 µm ID). 92 3 Defective damper. Please contact your Agilent service representative. 4 Defective main board. Please contact your Agilent service representative. 1260 Infinity QP VL User Manual