Download OrthoVet (european voltage stated in manual)

Transcript
MEDIZINTECHNIK FÜR TIERÄRZTE
Operation manual
Surgical motor
Or t h o V ET
Read the operation manual before starting the device!
Pay attention to the safety instructions!
Surgical motor Or t hoVET
All brands and product names are trademarks or
registered trade-marks of the
respective companies.
Surgery motor O r t h o V ET
Edition: 2004-05-17
© 2004
Eickemeyer Medizintechnik für Tierärzte e.K.
All rights reserved, including the right for
translation.
No part of this operation manual may be
reproduced or distributed in any manner
(print, copy, microfilm or any other means of
reproduction) without prior written consent of
Eickemeyer Medizintechnik für Tierärzte e.K.
OrthoVet
Technical changes reserved.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
OrthoVet
Surgical motor Or t hoVET
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
OrthoVet
Index
1
2
3
4
5
6
7
User tips ..........................................................................................................5
Safety tips .......................................................................................................6
Short operation manual .................................................................................7
Performance description ...............................................................................8
Device description..........................................................................................9
Funktions descripion .....................................................................................9
Normal Use....................................................................................................10
7.1
7.2
7.3
7.4
Connection of the Unit ............................................................................................... 10
Power on.................................................................................................................... 12
How to use the OrthoVet............................................................................................ 12
Power off.................................................................................................................... 12
8
Maintenance..................................................................................................13
8.1
Change the protection................................................................................................ 13
9
10
11
12
13
Troubles and how to correct them ..............................................................14
Technical Details ..........................................................................................15
Spares............................................................................................................16
Terms for the user ........................................................................................16
Transport, Installation, Demounting ...........................................................16
13.1
13.2
Positioning of the device ............................................................................................ 16
Electrical connection .................................................................................................. 17
14
15
16
Warrantee ......................................................................................................17
EG-Decleration of conformity......................................................................18
Glossary ........................................................................................................19
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
1 User tips
Particular use:
The OrthoVET is a Surgery engine and is
there for to impulse different Surgery
handpieces in the Veterinary medicine.
Unparticular use:
The use in the Humanmedicin is not allowed.
The surgery engine can not be sterilized.
This operation manual is for:
1. learned and introduced operators.
Persons who not introduced in the device
can´t use the device.
For the user are the capters 1 – 9.
2. the maintenance personal
The capters 8 and 9 also the capter 13 is
there for the maintenance personal. But
Maintenance personal must read the whole
operation manual.
3. the operator
The capter 10 – 16 are there for the
operator.
Terms are descripted in the glossar.
In this operation manual are following shortcuts
used.
UVV = accident protection instructions
IEC = International electronial commission
EN = European norm
VDE = Federation german electricians
Symbols
Χ = CE-symbol
OrthoVet
=This Symbol labels all for the safety
important instructions. Follow this instructions
always. Otherwise you can hurt yoursel or your
enviroment.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
2 Safety tips
Danger!
This are warns of danger for humans.
Danger of life is atentt throu the word„Danger of life“.
Danger!
This are warns of danger throug electrical shocks.
Danger of life is atentt throu the word„Danger of life“.
Danger of life throug electric shock!
Befor you work at the device plug of the flex!!
Attention!
Damage of the device through overheating
OrthoVet
The Ventilation slots must always be free.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
3 Short operation manual
The short operation manual I only there for to remember. Read the safety Tipp and the
operation manual.
Power on device
Power off device
Use Handpiece
Green Switch power on 1
-> Switch glows
Green Switch power off 0
-> Switch does not glow
Press the foot switch
OrthoVet
Befor you do anything at the device plug of the flex!!
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
4 Performance description
The OrthoVET is a Surgery engine and is
the for to power the Surgery handpiece in
the Veterinary medicin.
The OrhoVET award especially throug:
Easy use
Compact bulding
High performance
though steinless steel chase
brushless engine
maintenance-free
low heating
constant rotation speed
maximal rotation speed 17.500 1/min
lightweight foot switch
OrthoVet
Pic 1: Surgery engine
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
5 Device description
The OrthoVET ia a Surgery engine, wich is
constructet for the use in the veterinary medicine,
and is operated over a handpiece.
The principal constituents of the Surgery engine
is the brushless surgery engine with a EMVFiilter and the engine controler.
A ventilator is at the back of the device.
6 Funktions descripion
OrthoVet
The Surgery engine OrthoVET is powering the
Surgery hand piece over the flexible cable. The
engine rotation speed is controled over a foot
switch till 17.500 1/min.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
7 Normal Use
7.1
Connection of the Unit
Power supply connection
plug in the flex at the back of the device and
connect it with your outlet
Connection of the foot switch
plug in the foot switch cable at the back of the
unit.
turn the ring right
turn the ring left when you want to plug off the
foot switch
Connection of the flexible cable
At the front of the device there is the felxible
cable connection. Plug it in till it locks.
pull the flexible cable hull back and take the
flexible cable off
Pic 4: connection of the
foot switch
OrthoVet
Pic 5: surgical engine ready for use
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
Connection of flexible cable to the OrthoVet:
OrthoVet
The flexible cable is for the power
transmisson between the engine and
the handpiece.
Connect the bigger end with the
engine, to do this you must plug it in
till it locks. For unplug you must pull
away from the pin so you can uplug it.
The other end of the flexible cable is
connecting the handpiece. For unplug
you must pull the small button at the
flexible cable.
Before you can use the device attend
to the constable noises of the unit. If
there are some remarkeable noises
dont use the flexible cable.
When you use the flexible cable
attend to a steched cable. Don´t
flexure or snap off the flexible cable
when you use it!
The flexible shafts can be cleaned
and disinfected with all commerical
cleaning agents and disinfectants
wich are suitable for cleaning rubber
and plastic products and are
nationally approved. Follow the
manufacturer´s instructions on dilution
ratios, duration of action, etc. The
flexible shafts are cleaned with a cloth
moistended with disinfectant or with a
soft brush.
The connections on the end of the
flexible cable whitch is connected to
the drill handpiece should be
lubricated at least once a week.
Suited for this task is the oil Sterilit.
The flexible cable should be cleaned
and lubricated inside aproximatley
once a year. The frequency of
maintenance depends on the number
of times the flexible cables are
sterilized within a given period of time.
This maintenance work should be
performed only by authorized
specialists. For this work you can
contact us.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
Error detection and measures to correct them:
Engine does not work
Look at the chapter engine.
Flexible cable runs rough, knocks
Corrective maintenance at manufactures
stretch the flexible cable to correct the noise
Handpiece does not work
7.2
maybe
Driving slot of drive shaft is broken
Power on
Green power switch on 1 -> power switch
glows
7.3
How to use the OrthoVet
Pull the foot switch and the engine will have a
constant speed
7.4
Power off
OrthoVet
Green power switch off 0 -> power switch
does not glow
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
8 Maintenance
This chapter is only for people who made a
specail course in do maintenance on the device.
The surgery engine is free of Maintenance.
Danger of life through elecktric shock!
Before you do anything at the device plug it off from the
outlet.
8.1
Change the protection
Femal
connector for
flex
Protectionholder
Tip:
Always change both protections.
Compulsory protections:
2 piece IEC127 – M 2,5A 250V
Schaltvermögen L (40 A)
OrthoVet
Before you change the protection plug off the
flex!
Unlock the lock at the protection holder with a
screw driver.
Pull out the protection holder.
Plug in the protection holder that you can here
that it locks.
Pic 6: Backview
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
9 Troubles and how to correct them
Symptom
engine does not run
Possible cause
No voltage on the
device
Checks
Remedy
The flex is not
Connect the flex
connected
The power switch Switch the power switch
is off
on
The flex is broken Get a new flex
The protection is burn
through
The protection is burn
through twice
Engine broken
Protection is
broken
device is broken
Engine overheated
No connection form
the device to the foot
switch
Foot switch broken
Engine blocked
Chase is extem
warm
Foot switch not
connected
Foot switch cable
is broken
Plug is broken
Broken wire
The engine Stopps at
stress
The engine cant be
controled
OrthoVet
Power switch dont
glows but the engine
runs
The protection is burn The wrong
through
protection is Plug
in
Foot switch broken
Engine can be
controled with
another foot
switch
Surgical Engine is
Engine cant be
broken
controled with
another foot
switch
Power switch broken Power switch dont
glows
Change the protection
Contact the
Manufacturer
Contact the
Manufacturer
Let the engine cool
down
Connect the foot switch
Change the foot switch
Contact the
Manufacturer
Contact the
Manufacturer
Correct the protection
Contact the
manufacturer of the foot
switch
Contact the
manufacturer of the
engine
Contact the
manufacturer of the
engine
Handpiece broken
Look at the operating manual of the Handpiece
Flexible cable runs
rought, knock
Look at the operating manual of the Flexible cable
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
10 Technical Details
Surgery engine
Or t h o V ET
Elecktric input
400 W
Operating voltage
230 V~
Frequence
Plug in
50 Hz
Outlet with protection contact
Netzabsicherung
16 A
3 x 1,5 mm2
Cable cross-section
Protection class EN 60730
Dimesion
I
apprx. 250 x 280 x 140 mm
Weight
apprx. 6,8 kg
Ambient temperature
5 till 40 °C
Sound level
apprx. 65 dB(A)
OrthoVet
We reserve the right to change desing and
technical data without prior notice.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
11 Spares
Use only original Eickemeyer Spares
Item No.:
Descripion
000774
Foot switch with cable
000775
Flex with connector
000776
G-protection M 2,5A 250V, Schaltvermögen L
12 Terms for the user
The Personal for use, service and installation
must have the special qualifications for this work.
If the personal don´t have this qualifications, they
must be instructed that they get the qualification.
Also must the operator check that his personal
understand the Operation manual.
The operator must check the situation and get
sure that the specifications are adhered.
13 Transport, Installation, Demounting
Only let the Transport, Installation, Demounting
do by qualified personal.
13.1 Positioning of the device
At the positioning and the use u must keep
The national instalation and use instructions,
The national instructions for burn- and
explode safty
The general safty and disaster instrucions of
the Gesetzgebers.
Only alginment the device in dry and ventilated
rooms on a robust underground. A ventilated
room is very important because the device
shouldn´t geht over heated.
For a trouble-free use the device must stand
horizontal.
Attention!
Damage of the device through overheating!
OrthoVet
The ventilatory of the device must be free.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
13.2 Electrical connection
Connect the device only at an alternating current
outlet, wich have a protection with 16A.
The device is protecion groundet in the class
VDE 0730.
The Voltage declaration on the label must agree
exactly with the Voltage at the installation
location.
14 Warrantee
The Warrantee is 12 Months apart form are
Spares and wear parts and starts with the
transver, wich you must prove with the invoice.
OrthoVet
Disatvantages, wich are result from normal
abrasion, overload or improper care are, keep
imposibel of Warrantee.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
15 EG-Decleration of conformity
In terms of EG-Guid line 98/37/EG
Hereby we declerate, that the engine:
Model:
Type:
Surgical engine
OrthoVET
In the shipped accompishment fits to these terms:
98/37/EG
92/31/EWG
93/68/EWG
Engine guid line
electromagnetic compatibility
Low voltage directive
The following harmonising standarts are used:
Following national norms are used:
DIN 45635 Teil 1
Sonic earned measurement (Hüllflächenverfahren)
Tuttlingen 03.12.2004
Alexander Sprung – Bussines manager
OrthoVet
Eickemeyer Medizintechnik für Tierärzte e.K. , Eltastraße 8, 78532 Tuttlingen
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
Surgical motor Or t hoVET
16 Glossary
Operator
The Operator is somebody who uses
the engine for his nomal use. There
must not be any economical
usefulness.
User manual
OrthoVet
The user manual is a internal
instruction of the employer for the
employee. The user manual is there to
help to avoidance healt risks. She
includes the descriptions of working
sequenceses, operation istructions and
safety rules in the frame of the cycel in
a copmany.
D-78532 Tuttlingen
www.eickemeyer.de
MEDIZINTECHNIK FÜR TIERÄRZTE
MEDIZINTECHNIK FÜR TIERÄRZTE
Eickemeyer Medizintechnik für Tierärzte e.K.
Eltastraße 8
D-78532 Tuttlingen
Telefon
+49 (0 74 61) 96580-0
Telefax
+49 (0 74 61) 96580-90
Email: [email protected]
Internet:
www.eickemeyer.de
We reserve the right to change design and technical
data without prior notice.
CE-mark according to guideline 98/37/EG