Integral T Process Thermostats 7.8.3 Pump motor monitoring − In case of pump motor overload or a blockage, the heating, the pump and the refrigeration system are switched off. 7.8.4 A double tone sounds. − The display PUMP 1 appears and the fault triangle is flashing. − Rectify the fault, e.g. clean the pump or check the viscosity, then − if necessary allow the equipment to cool down approx. 30 min. − reset with the key. WARNING! Equipment and pump start up! − If several faults appear simultaneously, they have to be reset individually. − In case of overload of the pump for the external circuit the message PUMP 1 is displayed on model T 4600 W. − On units with a separate pump for internal circulation (T 4600 - T 10000 W) the message PUMP 2 means that the fuse F5 has to be replaced. Only by a qualified electrician: remove the side panel on the right. − On units with a 3-phase pump (T 7000 - T 10000 W) the message PUMP 1 means that the motor protection cut-out has to be reset. Only by a qualified electrician: remove the side panel on the right. − The compressor is switched off if the refrigerant pressure is excessive. − This message is displayed. − A double tone sounds. − The refrigeration compressor starts up automatically! − Reset the fault message PRES. Refrigerant pressure 80 − Starting up YAWE 0026 / 23.08.2006