Calibration Beam Specifications
Molybdenum/Rhodium Anode
NIST-Traceable Techniques
Filtration
Equivalent
Beam Code
Potential
(kV)
mm Mo
Mo/Mo 28
28
0.032
0.33
Mo/Mo 35
35
0.032
0.39
Mo/Rh 28
28
0.029
Rh/Rh 25
25
0.029
0.35
Rh/Rh 40
40
0.029
0.56
Mo/Mo28x
28
Rh/Rh/35x
35
mm Rh
mm Al
HVL
mm Al
0.41
0.030
0.029
2
0.63
2
0.898
mm Al
HVL
mm Al
PTB-Traceable Techinques
Filtration
Equivalent
Beam Code
Potential
(kV)
mm Mo
MV20
20
0.030
0.223
MV25
25
0.030
0.282
MV30
30
0.030
0.337
MV35
35
0.030
0.374
MV40
40
0.030
0.402
MV50
50
0.030
0.440
MV20
20
0.030
2
0.450
MV25
25
0.030
2
0.580
MV30
30
0.030
2
0.670
MV35
35
0.030
2
0.749
MV40
40
0.030
2
0.825
MV50
50
0.030
2
0.968
mm Rh
http://www.elso.sk
Locations
Service Center/Repair/
Calibration US
Fluke Biomedical
6045 Cochran Road
Cleveland OH 44139-3303
Tel: 440-498-2560
Toll free: 800-850-4608
ext 2564
Email: globalcal@flukebiomedical.
com
Service Center/Repair/
Calibration Europe
Fluke Biomedical Europe
Science Park Eindhoven 5110
5692EC Son, The Netherlands
Tel: +31 (40) 267 5435
Fax: +31 (40) 267 5436
Email: [email protected]
www.flukebiomedical.com/
service
Service and Calibration
67