Download Untitled - Menarini Diagnostics
Transcript
DISPLAYS, MEANINGS, ACTIONS Display Meter does not turn on after inserting a test strip. 44 DISPLAYS, MEANINGS, ACTIONS What it Means� What to Do Display What it Means What to Do Used Strip Error: Used or damaged strip. Retest with a new strip. Blood Sample Error. Incorrect application of blood sample or control solution onto the test strip, or the test strip may be damaged. Review your sampling technique. The blood sample quantity may be incorrect or hematocrit may be out of the allowed range for glucose measurement. Meter does not begin test countdown after applying a blood sample. •Not enough blood sample. •Sample applied after meter automatically turned off. •Meter may not be working properly. •Test strip may be damaged. Repeat the test with a new test strip again. After 3 attempts, call A. Menarini Diagnostics for assistance: (UK) 0800 243667 (Ireland) 1800 709903. No test results in the last 30 days. No action required. •Battery is dead. •Battery is installed incorrectly or there is no battery in the meter. •Test strip is inserted upside down or not completely in. Replace the battery. Check that the battery is correctly installed with the “+” sign facing you. Insert the test strip correctly with the GlucoMen name and white tip facing up and out. Call A. Menarini Diagnostics for assistance: (UK) 0800 243667 - (Ireland) 1800 709903. GlucoMen® LX - 38973 - 09/07 GlucoMen® LX - 38973 - 09/07 UK 45