Chapter 5. Communication
Command 186 (0xBA): Write Frequency Configuration
This command is to write the frequency configuration.
Byte
Table 84: Request Data Bytes for Write Frequency Configuration
Format
Description
0
Unsigned-8
Channel Number(1/2)
1
Unsigned-8
Measurement Type
2
Unsigned-8
Frequency Value Unit
3-6
Float
Frequency Base Value
7 - 10
Float
Frequency Full Value
11 - 14
Unsigned-32
Full Frequency, unit is Hz
15
Unsigned-8
Frequency Error Handling:
0: Low;
1: High;
2: Hold;
3: Value
16 - 19
Byte
Unsigned-32
Error Handling value, unit is Hz
Table 85: Response Data Bytes for Write Frequency Configuration
Format
Description
0
Unsigned-8
Channel Number(1/2)
1
Unsigned-8
Measurement Type
2
Unsigned-8
Frequency Value Unit
3-6
Float
Frequency Base Value
7 - 10
Float
Frequency Full Value
11 - 14
Float
Full Frequency, unit is Hz
15
Unsigned-8
Frequency Error Handling:
0: Low;
1: High;
2: Hold;
3: Value
16 - 19
112
Unsigned-32
Error Handling value, unit is Hz