Download VIMOS User Manual
Transcript
26 VIMOS User Manual VLT-MAN-ESO-14610-3509 Table 3.2: VIMOS Calibration Plan: MOS Mode Calibration Bias Darks Screen Flats Screen Arcs Screen Arcs Number 5 3 5 1 1 Specphot. Standard Star 4 quadrants Frequency When? Daily Day time Monthly Day time Daily Day time Daily Day time Attached to science OB with HR or LR blue See text Night time Products Master Bias Dark current MOS Flat Dispersion coeff. Dispersion coeff. Accuracy1 RON Comment 5% 0.3 pix 0.3 pix All used masks All used masks All used masks Response function 10% All grism+filter setups used 1 Internal accuracy, not considering instrument flexures (see Sect. 2.7). IMPORTANT NOTE The execution time for all attached night time calibrations (screen-flats and arcs) is subtracted from the total allocated time. 3.4.10 MOS mode: Exposure Time Calculator The approximate limiting magnitudes in MOS mode can be obtained from the exposure time calculator available on the ESO WEB pages http://www.eso.org/observing/etc/. 3.5 3.5.1 Integral Field Unit (IFU) Acquisition accuracy In IFU mode acquisition is “blind”, i.e., limited by the accuracy of the telescope pointing (∼ 300 accuracy). 3.5.2 Dithering/jitter pattern With the large field of view of the VIMOS-IFU, techniques equivalent to ”shift-and-add” classically used in imaging, are a possible mode of operations. Dithering/jitter pattern is a list of differential pointing positions around a central position. Several short images (with exposure times set to have sky background limited exposures) are taken at each position of a dithering pattern, which are combined to reconstruct an image with the total required exposure time. This allows to eliminate the sky contribution, cosmic rays, the CCD fringing (particularly strong in the red), cosmetic defects on the detectors, and to produce very accurate flat fielding for correction of pixel to pixel variations. The dithering pattern depends on the mean size of the observed objects: it should be larger than the mean size of the objects for the method to work efficiently. 3.5.3 IFU mode: grism + filter setup In IFU mode, only one filter+grism combination per OB is permitted. Users who want to observe the same targets with different filter+grism combinations are requested to submit separate OBs for different filter+grism combinations and to consider the respective overheads. 3.5.4 IFU mode: attached night time calibrations To better take into account the effects of instrument flexure the use of attached nighttime calibrations, arcs and flats (see Sect. 3.5.5), is MANDATORY for all IFU OBs. This template will provide 1 or 3 flat-field images (for HR grism: 3 screen-flats are mandatory, for the other cases, the user can choose the number) as well as 1 arc, taken at the same rotator position of the science target. The execution time for these nighttime calibrations is always subtracted from the total allocated time.