Download User manual
Transcript
Luminex 200 System User Manual 40 Wrong control lot number or target values are entered in Update CON Targets box. Verify that the correct lot number and target values are used. The system controls are in the wrong well on the plate. Verify that control microspheres are in the correct well. Not enough control microspheres added to the well. Add at least five drops of control microspheres to the well. For accurate drop volume, hold the vial upside down at a 90 degree angle to the plate while dispensing them. Control lot is expired. Use a fresh bottle of control microspheres. Control microspheres have been diluted. Do not dilute the control microspheres. The sample probe height is incorrect. Adjust the sample probe height. See page 18. The sample probe is clogged. Clean the sample probe. See page 20. There is air in the system. Verify sample probe height. Run three Prime commands, two Alcohol Flush commands, then three Washes with water The waste container is not properly vented. Verify that the waste container is properly vented and that the seal is not wet or soiled. The waste line was moved during system operation, resulting in unstable flow rate. Verify that the waste line is not moved during system operation. Possible problem with the lasers. Check the system control trend report for consistent failures. If any are evident on the report, contact Technical Support.