Download B-RAX 3100 User Manual

Transcript
Instruction Manual
7.6
B-RAX 3100
Clearing Errors
Once the cause of the IG error has been determined and resolved, the IG error must be cleared before the IG
filament can be turned on again. To clear errors, follow the steps outlined below:
1) Go to the IG CONTROL selection of the SETUP IG menu and make sure the IG control mode is set to MANUAL
MODE.
2) Next, go to the IG UNIT ON/OFF selection of the SETUP IG menu and select OFF then press the ENTER key to
save the setting.
Tips and visual key sequence to clear error :
The symbol  implies using the up or down arrow keys to select choice, etc.
You may need to press the MENU key one or more times to see the SETUP IG selection in the LCD display.
Follow this key sequence to perform the two steps outlined above: Press MENU  SETUP IG  ENTER 
IG CNTL  ENTER MANUAL MODE  ENTER :: Now, Press MENU  UNIT ON/OFF  OFF  ENTER
See the user manual for either the IGM400 or CCM500 for further instructions on troubleshooting and
disassembly of these devices.
InstruTech, Inc.