4. Using polishing mode Filling the powder drum Purging the polishing circuit Proceed as follows: 1. Put the device into standby mode to depressurize the drum. 2. Unscrew the cap (ref. 18). 3. Fill the drum with Satelec® PowderMax powder up to the level marker (ref. 19). 1. 2. 3. Release the pedal to stop polishing. Point the ProphyPen handpiece towards the cuspidor. Press the key (ref. 7) to purge the circuit. Testing the effectiveness of the powder jet 1. 2. Point the handpiece at the test card and press the pedal. Set the water flow on the handpiece connector. If there is not enough water: powder accumulates on the card. If there is too much water: the polishing effect is less effective. If the setting is correct: the powder is immediately removed by the water. Where the powder jet is not powerful enough, clean the nozzle (see chapter “Cleaning the polishing system”) or have the air pressure adjusted by an approved installer of dental equipment. Polishing Proceed as follows: 1. Use the ProphyPen polishing handpiece. 2. Press the pedal to start polishing. It is important to observe a jet/surface angle of from 45° to 60° and to keep to a spray/tooth distance of 4 to 6mm. 5