Download Attachment - abbott

Transcript
LUNAIRE ENVIRONMENTAL
Conditioner Blower Wheel:
Remove All Power From Chamber!
Inspect and clean the conditioner fan in the conditioning plenum. Make sure the fan spins freely and that it is
tight on its shaft.
Inspection Period: 6 Months
Evaporator Cooling Coil:
Remove All Power From Chamber!
Clean the evaporator cooling coil in the conditioning plenum.
Inspection Period: 6 Months
General Electrical Connections:
Remove All Power From Chamber!
Inspect inside the control panel and the machinery compartment for loose electrical connections, frayed
wires, loose components, or other potential problems.
Inspection Period: 6 Months
Electric Heater(s):
Remove All Power From Chamber!
Inspect the electric heater(s) inside the chamber conditioning plenum and look for defects.
Inspection Period: 6 Months
Electrical Supply Voltage:
Measure the power supply voltage to your oven and verify that it is within the ±10% tolerance established for
the nameplate rating of your oven.
Inspection Period: 6 Months
Controller Calibration:
The main temperature and high limit controllers should be checked for temperature indicating accuracy, and
for the proper activation of limit or alarm outputs. Please reference the controller user manual for more
information.
Inspection Period: 1 Year
Steady State / Stability Test Chambers,
Models CEO9xx - 1,- 2,- 3,- 4: W96s
Page 37