Download Attachment - abbott
Transcript
LUNAIRE ENVIRONMENTAL Conditioner Blower Wheel: Remove All Power From Chamber! Inspect and clean the conditioner fan in the conditioning plenum. Make sure the fan spins freely and that it is tight on its shaft. Inspection Period: 6 Months Evaporator Cooling Coil: Remove All Power From Chamber! Clean the evaporator cooling coil in the conditioning plenum. Inspection Period: 6 Months General Electrical Connections: Remove All Power From Chamber! Inspect inside the control panel and the machinery compartment for loose electrical connections, frayed wires, loose components, or other potential problems. Inspection Period: 6 Months Electric Heater(s): Remove All Power From Chamber! Inspect the electric heater(s) inside the chamber conditioning plenum and look for defects. Inspection Period: 6 Months Electrical Supply Voltage: Measure the power supply voltage to your oven and verify that it is within the ±10% tolerance established for the nameplate rating of your oven. Inspection Period: 6 Months Controller Calibration: The main temperature and high limit controllers should be checked for temperature indicating accuracy, and for the proper activation of limit or alarm outputs. Please reference the controller user manual for more information. Inspection Period: 1 Year Steady State / Stability Test Chambers, Models CEO9xx - 1,- 2,- 3,- 4: W96s Page 37