Reminder: Developer Mode is intended for non-clinical use only and is NOT cleared for use on humans
KV Imaging Modes
Imaging
Mode
Dynamic
Gain
String (id= )
Acquisition
Mode
"DynamicGain"
Acquisition
Technique
Resolution
A basic mode intended
for CBCT projections. In
this mode each pixel
automatically switches
to a larger gain before
reaching saturation,
allowing a wider
dynamic range.
As above, but Images
are continuously
acquired at 11 frames
per second.
Single Image
1024 x 768
"DynamicGainFluoro"
Continuous
1024 x 768
"DynamicGainLowFra
merate”
Single Image
1024 x 768
See Dynamic Gain.
1024 x 768
As above, except that
images are continuously
acquired at 7 frames
per second.
2048 x 1536
Full pixel resolution
used.
Dynamic Gain
Dynamic
Gain
Fluoro
Dynamic
Gain Low
Framerate
Dynamic
Gain Low
Framerate
Fluoro
High
Quality
Definition
"DynamicGainLowFra
merateFluoro"
"HighQuality"
Low Dose
”LowDose”
Triggered
"Triggered"
Fluoro
“Fluoro”
Dynamic Gain
Low Framerate
Continuous
High Quality
Low Dose
Single Image
1024 x 768
Single Image
1024 x 768
Continuous
1024 x 768
2x2 pixels merged into
one to achieve higher
contrast, but lower
resolution.
As above, but the
images are triggered by
respiratory gating.
Same as Low Dose but
images continuously
readout at 15 frames
per second